Specifications
1-2 1740A Series, 1750A Series, & 1760 Series Service Manual
These instruments employ a bright, post accelerated CRT with lighted internal
graticule. The parallax free internal graticule structure contains targets and
markings for both the vector and waveform functions. The “Lightning”graticule,
for the 1760 Series, is electronic, in order to keep from complicating the
waveform/vector graticule. Option 74, which uses a white phosphor (P4) CRT, is
available for all three instrument series.
These monitors are microprocessor controlled to provide greater versatility.
Switch settings, affecting the operation of the instrument, are continuously
polled by the processor, with any change in status acted upon immediately.
Current operating conditions are preserved in Non-Volatile Random Access
Memory (NOVRAM), which returns the front panel settings to the current
settings, in the event of power interruption (either accidental or routine power
down).
Many operational measurements are performed on a repetitive basis, and these
instruments provide a method of repeating common measurements by simply
selecting stored measurement settings from a CRT menu list. The common
measurement front panel settings can be stored and named by the user. Measure-
ment specific, front-panel settings can be recalled by pushing the front-panel
Preset Menu button and selecting the desired preset by name or number.
Many functions that were formerly selected by changing internal jumpers and/or
wire straps are now accessible through on screen menu selections. Menus are
selected by pushing the appropriate front panel Menu selection. Menu items are
then selected by pushing one or more of the assignable switches and/or rotating
the assigned front-panel control. Once selected these menu choices are retained
until changed by subsequent reconfiguration.
The front panel provides both assignable switches (located next to the CRT) and
controls (located beneath the CRT), that operate with CRT readout to increase
functionality without cluttering the limited front panel area. Function of these
switches and controls is dictated by the front-panel Display and initial Menu
choices.
The 1740A Series is a full capability waveform/vector monitor that also provides
audio and time code measurements. The 1750A Series has all of the capabilities
of the 1740A Series, with the addition of SCH phase measurements.
The 1760 Series combines component measurements with the composite
capabilities of the 1740A/1750A Series. It incorporates all of the measurements
of the 1740A Series with a full set of component analog measurements. The
assignable cursors, along with the CRT readout can be used for time, voltage,
and phase measurements. The system of Menus and CRT readout simplifies the
configuration of this monitor for measurement or monitoring of signal character-
istics.