Tektronix 7623 User manual

MANUAL
7623/R7623
STORAGE
OSCILLOSCOPE
SERVICE
MANUFACTURERS
OF
CATHODE-RAY
OSCILLOSCOPES

WARRANTY
All
TEKTRONIX
instruments
are
warrante
against
efective
materials
an
workmanship
for
one
year.
Any
questions
with
respect
to
the
warranty
shoul
be
taken
up
with
your
TEKTRONIX
Fiel
Engineer
or
representative.
All
requests
for
repair
an
replacement
parts
shoul
be
irecte
to
the
TEKTRONIX
Fiel
Office
or
representative
in
your
area.
This
will
assure
you
the
fastest
possible
service.
Please
inclu e
the
instrument
Type
Number
or
Part
Number
an
Serial
Number
with
all
requests
for
parts
or
service.
Specifications
an
price
change
privileges
reserve .
Copyright
©
1973
by
Tektronix,
Inc.,
Beaverton,
Oregon.
Printe
in
the
Unite
States
of
America.
All
rights
reserve .
Contents
of
this
publication
may
not
be
repro uce
in
any
form
without
permission
of
Tektronix,
Inc.
U.S
.A
.
an
Foreign
TEKTRONIX
pro ucts
covere
by
U.S.
an
foreign
patents
an /or
patents
pen ing.
TEKTRONIX
is
a
registere
tra emark
of
Tektronix,
inc.

7623/R7623
Service
TABLE
OF
CONTENTS
SECTION
1
OPERATING
INFORMATION
Page
Operating
Voltage
1-1
Operating
Temperature
1-1
Plug-In
Installation
An
Removal
1-1
Operation
1-1
Non-Store
Operating
Set-Up
Information
1-1
Front
Panel
Controls
An
Connectors
1-2
Rear
Panel
Controls
An
Connectors
1-3
Storage
Mo es
1-4
Variable
Persistence
Operating
Set-Up
Information
1-4
Fast
Operating
Set-Up
Information
1-4
Multi-Trace
Fast
Operating
Set-Up
Information
1-4
Bi-Stable
Operating
Set-Up
Information
1-4
Save
Mo e
Operating
Set-Up
Information
1-4
SECTION
2
CALIBRATION
Calibration
Interval
2-1
Tektronix
Fiel
Service
2-1
Using
This
Proce ure
2-1
General
2-1
In ex
2-1
Performance
2-1
Partial
Proce ure
2-1
Complete
Calibration
Proce ure
2-1
TEST
EQUIPMENT
REQUIRED
Special
Calibration
Fixtures
2-1
Calibration
Equipment
Alternatives
2-1
Test
Equipment
2-2
Preliminary
Control
Settings
2-4
CALIBRATION
PROCEDURE
2-5
Intro uction
2-5
In ex
To
Calibration
Proce ure
2-5
Power
Supply
2-5
Display
An
Z-Axis
2-5
Vertical
Deflection
System
2-5
Trigger
System
2-5
Horizontal
Deflection
System
2-5
Z-Axis
An
Auto
Focus
2-5
Calibrator
2-5
Signals
Out
2-5
Rea out
Operation
2-6
Storage
Operation
2-6
i

7623/R7623
Service
TABLE
OF
CONTENTS
(cont)
SECTION
3
CIRCUIT
DESCRIPTION
Page
Block
Diagram
3-1
Circuit
Operation
3-1
Logic
Fun amentals
3-3
Symbols
3-3
Logic
Polarity
3-3
Input/Output
Tables
3-6
Non-Digital
Devices
3-6
MAIN
INTERFACE
3-6
LOGIC
CIRCUIT
3-6
Logic
Block
Diagram
3-6
Z-Axis
Logic
3-6
Clock
Generator
3-9
Vertical
Choppe
Blanking
3-10
Chop
Counter
3-11
Vertical
Mo e
Control
3-13
Right
3-13
Chop
3-13
Alt
3-14
A
3-14
Left
3-14
Vertical
Binary
3-14
Plug-In
Binary
3-15
Output
Buffers
3-16
TRIGGER
SELECTOR
3-16
Trigger
Mo e
An
A
Signals
3-16
General
3-16
Vert
Mo e
3-16
Left
3-16
Right
3-18
Trigger
Channel
Switch
3-18
Trigger
Output
Amplifier
3-18
Vertical
Signal
Buffer
3-18
VERTICAL
INTERFACE
3-19
Vertical
Channel
Switch
3-19
Auxiliary
Y-Axis
Input
Amplifier
3-20
Delay
Line
Buffer
3-20
Delay
Line
3-20
VERTICAL
AMPLIFIER
3-20
Input
Balance
3-21
Output
Amplifier
3-21
Beam
Fin er
Network
3-21
HORIZONTAL
AMPLIFIER
3-21
Horizontal
Channel
Switch
3-21
Riÿit
Output
Amplifier
3-22
Left
Output
Amplifier
3-22
ii

7623/R7623
Service
TABLE
OF
CONTENTS
(cont)
SECTION
3
CIRCUIT
DESCRIPTION
(cont)
Page
CALIBRATOR
An
FRONT
PANEL
SWITCHING
3-22
Mo e
Switch
Logic
3-23
Calibrator
3-23
Oscillator
3-23
Output
Amplifier
3-23
Output
Divi er
3-23
CRT
CIRCUIT
3-24
Z-Axis
Amplifier
3-24
Auto
Focus
Amplifier
3-25
High
Voltage
Oscillator
3-25
High
Voltage
Regulator
3-25
High
Voltage
Supplies
3-26
CRT
Control
Circuits
3-26
LOW-VOLTAGE
POWER
SUPPLY
3-27
Power
Input
3-27
-
50
Volt
Supply
3-27
—
15
Volt
Supply
3-30
+5
Volt
Supply
3-30
+15
Volt
Supply
3-30
+50
Volt
Supply
3-30
+130
Volt
Supply
3-31
Graticule
Light
Supply
3-31
SIGNAL
OUT
BOARD
3-31
Vert
Sig
Out
3-31
+Gate
Out
3-31
+Sawtooth
Out
3-31
READOUT
SYSTEM
3-31
Display
Format
3-32
Developing
The
Display
3-32
Circuit
Analysis
Of
The
Rea out
System
3-35
Timer
3-35
Time-Slot
Counter
3-40
Wor
Trigger
3-40
Channel
Counter
3-41
Single-Shot
Lockout
3-41
Enco ing
The
Data
3-41
Column
An
Row
Data
Switches
3-43
Display-Skip
Generator
3-44
Column
An
Row
Deco er
3-44
Zeros
Logic
An
Memory
3-44
Character
Generators
3-47
Decimal
Point
Logic
An
Character
Position
Counter
3-47
Format
Generator
3-48
Y-Output
Amplifier
3-48
X-Output
Amplifier
3-48
Display
Sequence
3-48

7623/R7623
Service
TABLE
OF
CONTENTS
(cont)
SECTION
3
CIRCUIT
DESCRIPTION
(cont)
Page
STORAGE
CIRCUITRY
3-50
Erase
An
Timing
Circuits
3-50
Sweep
Lockout
Circuit
3-50
Auto
Erase
Circuit
3-50
Rea out
Shut
Down
Circuit
3-54
Persistence
Pulse
An
Pump
Pulse
Generator
Circuits
3-54
Floo
Gun
Ano e
An
CE2
Circuits
3-55
CE3
An
CE
2
Circuits
3-55
Collector
Circuit
3-56
High
Spee
Mesh
Circuit
3-56
Bi-Stable
Mesh
Circuit
3-56
SECTION
4
MAINTENANCE
Panel
Removal
4-1
Cabinet
Mo el
4-1
Rack
Mo el
4-1
Power-Unit
Removal
4-1
PREVENTIVE
MAINTENANCE
General
Cleaning
Exterior
CRT
Interior
Lubrication
Semicon uctor
Checks
Recalibration
4-1
4-1
4-1
4-2
4-2
4-2
4-2
4-2
4-2
TROUBLESHOOTING
Intro uction
Diagrams
Circuit
Boar s
Multi-Pin
Connector
Color-Co e
Wiring
Color-Co e
Resistor
Color-Co e
Capacitor
Marking
Dio e
Color-Co e
Semicon uctor
Lea
Configuration
Troubleshooting
Equipment
Troubleshooting
Techniques
4-2
4-2
4-3
4-3
4-3
4-3
4-3
4-3
4-3
4-3
4-3
4-6
iv

7623/R7623
Service
TABLE
OF
CONTENTS
(cont)
SECTION
4
MAINTENANCE
(cont)
Page
CORRECTIVE
MAINTENANCE
4-10
Obtaining
Replacement
Parts
4-10
Special
Parts
4-10
Or ering
Parts
4-11
Sol ering
Techniques
4-11
Component
Replacement
4-11
General
4-11
Circuit
Boar
Replacement
4-11
Access
To
The
Fan
Motor
4-13
Access
To
The
Fan
Motor
For
The
R7623
4-13
Semicon uctor
Replacement
4-14
Access
To
The
Power
Transistors
4-14
Interconnecting
Pin
Replacement
4-14
Catho e
Ray
Tube
Replacement
4-15
Switch
Replacement
4-16
Graticule
Bulb
Replacement
4-16
Power
Transformer
Replacement
4-16
High-Voltage
Compartment
4-16
Fuse
Replacement
4-16
Recalibration
After
Repair
4-16
Instrument
Repackaging
4-17
SECTION
5
ELECTRICAL
PARTS
LIST
SECTION
6
DIAGRAMS
SECTION
7
MECHANICAL
PARTS
LIST
CHANGE
INFORMATION

The
TEKTRONIX
76 3/R76 3
Storage
Oscilloscope
is
a
solid-state
instrument
designed
for
fast
writing
rate
storage
applications.
The
76 3
has
four
basic
modes
of
operation:
Fast,
Bi-Stable,
Variable
Persistence,
and
Non-Store.
The
in
strument
is
designed
to
accept
TEKTRONIX
7-Series
plug-in
units
to
form
a
complete
measurement
system.
The
flexibility
of
this
plug
in
feature
and
the
variety
of
plug-in
units
available
allow
this
system
to
be
used
for
many
measure
ment
applications.
The
76 3
features
a
large
cathode-ray
tube
(CRT)
screen,
8
X
10
divisions,
with
small
spot
size,
and
fast
stored
writing
rate.
76 3/R76 3
Service

Section
1
—
76 3/R76 3
Service
OPERATING
INFORMATION
Operating
Voltage
The
76 3/R76 3
can
be
operated
from
either
a
1
10-volt
line
or
0-volt
line
source.
In
addition,
three
operating
ranges
can
be
selected
by
a
jumper
located
on
the
Rectifier
board.
See
Fig.
1-1.
Select
a
range
that
is
centered
about
the
average
line
voltage
to
which
the
instrument
is
connected.
See
Table
1-1
for
ranges
and
proper
fuses.
Table
1-1
Pins
Regulating
Range
Selected
1
10-volt
nominal
0-volt
nominal
Low
90
to
110
volts
180
to
0
volts
Med
99
to
1 1
volts
198
to
4
volts
Hi
108
to
13
volts
18
to
6
volts
Line
Fuse
3.
A
Slow-blow
1.6
A
Slow-blow
Operating
Temperature
The
76 3/R76 3
can
be
operated
where
the
ambient
air
temperature
is
between
0
°C
and
+50°
C.
The
76 3
is
cooled
by
air
drawn
through
the
instrument.
Maintain
about
two-inches
clearance
for
proper
ventilation
on
all
sides
and
the
top.
The
feet
on
the
bottom
of
the
instrument
provide
the
necessary
clearance.
The
R76 3
is
cooled
by
air
drawn
through
the
air
filter
on
the
rear
of
the
instrument
and
blown
out
through
the
holes
on
the
right
side.
Maintain
two-inches
of
clearance
on
the
right
side
for
proper
ventilation.
A
thermal
cutout
switch
in
the
instruments
provides
thermal
protection,
and
interrupts
the
power
to
the
instrument
if
the
internal
temperature
exceeds
a
safe
operating
level.
Plug-in
Installation
and
Removal
To
install
a
plug-in
unit
into
one
of
the
plug-in
compartments,
align
the
slots
in
the
top
and
bottom
of
the
plug-in
unit
with
the
associated
guide
rails
in
the
plug-in
compartment.
Push
the
plug-in
unit
firmly
into
the
plug-in
compartment
until
it
locks
into
place.
To
remove
the
plug-in
unit,
pull
the
release
latch
on
the
plug-in
unit
to
disengage
it,
and
pull
the
unit
out
of
the
plug-in
compartment.
Plug-in
units
can
be
removed
or
installed
without
turning
off
the
instrument
power.
Special
purpose
plug-in
units
may
have
specific
restrictions
regarding
the
plug
in
compartments
in
which
they
can
be
installed.
This
information
will
be
given
in
the
instruction
manual
for
these
plug-in
units.
Operation
The
following
information
provides
the
necessary
controls
and
control
settings
to
obtain
a
display,
which
can
be
used
to
verify
basic
operation
or
calibration
without
removing
the
covers
or
making
any
internal
adjustments.
In
the
NON
STORE
mode
the
76 3
functions
the
same
as
a
conventional
oscilloscope.
The
CALIBRATOR
signal
is
a
convenient
signal
for
verifying
basic
operation
and
calibration.
Use
the
Operating
Set-Up
Information
as
a
guide
for
setting
the
front
panel
controls.
Non-Store
Operating
Set-Up
Information
INTENSITY
READOUT
GRATICULE
ILLUM
FOCUS
TRIG
SOURCE
VERT
MODE
POWER
As
desired
(midrange)
As
desired
As
desired
Well
defined
display
VERT
MODE
Selected
for
the
plug-in
compartment
with
the
plug-in
unit
installed
and
with
signal
input
to
be
displayed.
Pressed
in
Spare
jumper
Fig.
1-1.
Location
of
Voltage-selector
jumper,
spare
jumper,
an
230
V
fuse
in
power-unit
(shown
remove ).
1-1

Operating
Information
—
7623/R7623
Service
1.
INTENSITY
—
Controls
the
brightness
of
the
isplay.
2.
READOUT
—
Turns
on
the
Rea out
is
play,
an
controls
the
brightness
of
the
Rea out
isplay.
3.
FOCUS
—
Screw river
a justment
provi es
a justment
for
optimum
isplay
efinition.
4.
GRATICULE
ILLUM
—
Controls
graticule
illumination.
5.
BEAM
FINDER-When
presse
the
isplay
is
limite
to
within
the
graticule
area.
6.
Camera
Power-Three
pin
connector
on
the
CRT
bezel,
(top)
+15
volt
power
source,
(mi le)
receives
remote
single
sweep
reset
signal
from
compatible
camera
system,
an
(bottom)
groun
pin.
7.
VERT
MODE
—
Selects
vertical
mo e
of
operations.
o
«X
10.
NON-STORE
—
Non-store
operation.
LEFT:
Signals
from
the
left
plug-in
compartment
are
is
playe .
ALT:
Signals
from
both
vertical
plug-in
compartments
are
isplaye
( ual
trace).
Display
is
switche
from
one
vertical
to
the
other
after
each
sweep.
ADD:
Signals
from
both
vertical
plug-in
compartments
are
algebraically
a e ,
an
the
sum
is
isplaye .
11.
VAR-PERSIST-Variable
persistence
storage
mo e.
The
AUTO
VIEW
TIME
or
PERSISTENCE
control
setting,
controls
the
retention
of
the
store
isplay.
12.
FAST
—
For
storing
fast
signals.
13.
MULTI-TRACE-IMULTI-FAST
in
early
instruments)-For
stor
ing
multiple
fast
signals.
For
each
new
signal
to
be
store
the
ERASE
(PREP
IN
MULTI
TRACE
FAST)
must
be
presse .
CHOP:
Signals
from
both
vertical
compartments
are
is
playe
( ual
trace).
Display
switches
from
the
left
vertical
compartment
signal
to
the
right
vertical
compartment
signal
at
a
one
megahertz
rate.
RIGHT:
Signals
from
the
right
plug-in
compartment
are
isplaye .
8.
TRIG
SOURCE
—
Selects
the
source
of
the
internal
trigger
signals
for
the
horizontal
compartment.
LEFT:
Trigger
signals
are
from
the
left
vertical
plug-in
compartment
only.
VERT
MODE:
Trigger
signals
are
from
the
vertical
compart
ment
being
isplaye
except
in
the
CHOP
an
ADD
mo es;
then
the
trigger
signal
is
the
algebraic
sum
of
the
trigger
signals
from
the
left
an
right
plug-in
compartments.
RIGHT:
Trigger
signals
are
from
the
right
vertical
plug
in
compartment
only.
14.
INTEG
—
For
storing
very
fast
repetitive
signals.
15.
BI
STABLE
—
For
storing
slow
signals
for
long
perio s
of
time.
16.
SAVE
—
Prevents
acci ental
erasure
of
the
store
isplay.
17.
SAVE
INTEN
(STORED
INTEN
in
early
instruments)-Controls
the
brightness
of
the
store
isplay
in
the
Save
Mo e.
18.
AUTO
ERASE-Automatically
erases
the
store
isplay.
AUTO
VIEW
TIME
setting
controls
the
time
between
erase
cycles.
19.
AUTO
VIEW
TIME-IVIEW
TIME
in
early
instruments)-
Controls
the
time
between
erase
cycles
in
the
Auto
Erase
Mo e.
20.
MAN
ERASE
—
Erases
the
store
isplay.
21.
STORAGE
LEVEL-(LEVEL
in
early
instruments)
—
Controls
the
writing
rate
in
the
FAST
an
BI
STABLE
storage
mo es.
Controls
the
writing
rate
an
brightness
of
the
store
isplay
in
the
Variable
persistence
Mo e.
9.
POWER
—
Switch
an
in icator;
switch
turns
on
instrument,
an
the
in icator
is
on
when
the
instrument
is
connecte
to
a
power
source
an
turne
on.
22.
CALIBRATOR
—
Positive
going
square
wave
or
DC
voltage
selecte
by
changing
an
internal
jumper.
Output
voltage
at
pin
jacks
(4
V.
0.4
V
an
40
mV).
Fig.
1-2.
Front
Panel
Controls
an
Connectors.
1-

Operating
Information
—
7623/R7623
Service
1.
FUSE
—
Line
voltage
fuse.
2.
REMOTE
ERASE
IN
—
Provi es
external
connection
for
remote
erase.
3.
EXT
S
S
RESET
IN
—
Remote
single
sweep
reset.
4.
EXT
Z
AXIS
IN
—
Input
for
intensity
mo ulation
of
the
CRT
isplay.
5.
VERT
SIG
OUT-Vertical
signal
selecte
by
TRIG
SOURCE
switch
(LEFT.
RIGHT.
ALT
an
ADD).
6.
+GATE
OUT
—
Gate
signal
selecte
by
gate
selector
switch
(Main,
Auxiliary,
an
Delay).
7.
+SAWTOOTH
OUT-Positive-going
sawtooth
from
time-base
unit.
Fig.
1-3.
Rear
Panel
Controls
an
Connectors.
1-3

Operating
Information
—
7623/R7623
Service
After
obtaining
a
display
in
the
NON
Store
mode
continue
to
the
Storage
modes.
Storage
Modes
Set
the
controls
as
given
in
the
Operating
Set-Up
Information
for
the
Non-Store
mode.
Set
the
storage
controls
as
given
for
Variable
Persistence,
Fast,
Bi-Stable
and
Save
mode.
MULTI
TRACE
In
ERASE
Pressed
and
released.
(Only
one
sweep
is
stored,
press
the
ERASE
button
and
another
sweep
will
be
stored
without
erasing
the
original
display.)
SAVE
Out
(released).
Variable
Persistence
Operating
Set-Up
Information
Bi-Stable
Operating
Set-Up
Information
VAR
PERSIST
In
(Pressed
in)
BI-STABLE
In
(pressed
in)
NON
STORE
Out
(released)
FAST
Out
(released)
STORAGE
LEVEL
PERSISTENCE
SAVE
As
desired
(no
fading
positive
of
the
display)
As
desired
(counterclockwise
in
the
detent
is
maximum
retention;
in
the
clockwise
direction
the
retention
decreases.)
Out
(released).
ERASE
Press
ERASE
button
SAVE
Mo e
Operation
Set-Up
Information
Obtain
a
stored
display
in
any
storage
mode.
SAVE
In
(Pressed
in)
ERASE
FAST
Operating
Set-Up
Information
Press
ERASE
button
SAVE
INTEN
FAST
VAR
PERSIST
AUTO
ERASE
STORAGE
LEVEL
In
(pressed
in).
Out
(Released).
Out.
Pressed
and
released.
(Only
one
sweep
is
stored.
Press
the
MAN.
ERASE
button
to
erase
the
display,
and
to
start
another
sweep
for
storing.)
As
desired
(adjusted
so
stored
display
does
not
fade
into
the
background).
Turning
the
SAVE
INTEN
control
clockwise
increases
the
intensity
of
the
stored
display.
There
is
no
erase
signal
generated,
and
all
sweep
signals
are
locked-out.
Press
the
SAVE
button
out
(release).
Auto
ERASE
Pressed
in
Auto
VIEW
TIME
As
desired
(counterclockwise
position
maximum
display
time
before
erase
cycle
starts.
Clockwise
direction
decreases
time
that
a
stored
display
is
present
before
an
erase
cycle).
AUTO
ERASE
Pressed
and
released
(out).
MULTI
TRACE
FAST
Operating
Set-Up
Information
Obtain
one
stored
display
in
the
fast
mode.
For
more
detailed
operating
instructions,
see
the
76 3/R76 3
Operators
Manual.
1-4

Section
2
—
7623/R7623
Service
CALIBRATION
Calibration
Interval
To
assure
instrument
accuracy,
check
the
calibration
of
the
76 3
every
1000
hours
of
operation,
or
every
six
months
if
used
infrequently.
Before
complete
calibration,
thoroughly
clean
and
inspect
this
instrument
as
outlined
in
the
Maintenance
section.
Tektronix
Field
Service
Tektronix,
Inc.
provides
complete
instrument
repair
and
recalibration
at
local
Field
Service
Centers
and
the
Factory
Service
Center.
Contact
your
local
Tektronix
Field
Office
or
representative
for
further
information.
Using
This
Procedure
General.
This
section
provides
several
features
to
facili
tate
calibration
of
the
76 3.
These
are:
In ex.
An
index
is
given
preceding
the
calibration
procedure
to
aid
in
locating
a
step.
Performance
Check.
The
performance
of
this
instrument
can
be
checked
by
performing
only
the
VCHECK
steps.
The
V
preceding
a
step
indicates
that
performing
this
step
checks
the
instrument
against
the
tolerance
listed
as
a
Performance
Requirement
(see
Specification
Section
in
the
Operators
Manual).
Partial
Proce ure.
A
partial
calibration
is
often
desirable
after
replacing
components,
or
to
touch
up
the
adjustment
of
a
portion
of
the
instrument
between
major
recalibra
tions.
To
calibrate
only
part
of
the
instrument,
set
the
controls
as
given
under
Preliminary
Control
Settings
and
start
with
the
nearest
Equipment
Required
list
preceding
the
desired
portion.
To
prevent
unnecessary
recalibration
of
other
parts
of
the
instrument,
re-adjust
only
if
the
tolerance
given
in
the
CHECK-
part
of
the
step
is
not
met.
If
re-adjustment
is
necessary,
also
check
the
calibration
of
any
steps
listed
in
the
INTERACTION
—
part
of
the
step.
Complete
Calibration
Proce ure.
Completion
of
each
step
in
the
following
calibration
procedure
ensures
that
this
instrument
is
both
correctly
adjusted
and
performing
within
all
given
tolerances.
TEST
EQUIPMENT
REQUIRED
The
test
equipment
and
accessories,
or
the
equivalent,
given
in
the
Test
Equipment
table
is
required
for
complete
calibration
of
the
76 3.
Specifications
given
for
the
test
equipment
are
the
minimum
necessary
for
accurate
calibra
tion.
Therefore,
the
specifications
of
any
test
equipment
used
must
meet
or
exceed
the
listed
specifications.
All
test
equipment
is
assumed
to
be
correctly
calibrated
and
operating
within
the
listed
specification.
Detailed
operating
instructions
for
the
test
equipment
are
not
given
in
this
procedure.
Refer
to
the
instruction
manuals
for
the
test
equipment
for
more
information.
Special
Calibration
Fixtures
Special
Tektronix
calibration
fixtures
are
used
in
this
procedure
only
where
they
facilitate
instrument
calibration.
These
special
calibration
fixtures
are
available
from
Tektronix,
Inc.
Order
by
part
number
through
your
local
Tektronix
Field
Office
or
representative.
Calibration
Equipment
Alternatives
All
of
the
listed
test
equipment
is
required
to
completely
check
and
adjust
this
instrument.
This
calibration
pro
cedure
is
based
on
the
first
item
of
equipment
given
as
an
example
of
applicable
equipment.
When
other
equipment
is
substituted,
control
settings
or
calibration
setup
may
need
to
be
altered
slightly
to
meet
the
requirements
of
the
substitute
equipment.
If
the
exact
item
of
test
equipment
given
as
an
example
in
the
Test
Equipment
table
is
not
available,
first
check
the
Specifications
column
carefully
to
see
if
any
other
equipment
is
available
which
might
suffice.
Then
check
the
Usage
column
to
see
what
this
item
of
test
equipment
is
used
for.
If
used
for
a
check
or
adjustment
which
is
of
little
or
no
importance
to
your
measurement
requirements,
the
item
and
corresponding
step(s)
can
be
deleted.
The
following
procedure
is
written
to
completely
check
and
adjust
the
76 3
to
the
limits
given
in
Section
of
the
Operators
Manual
and
to
allow
interchanging
7000-series
plug-in
units
between
7000-series
mainframes
without
the
need
to
recalibrate
the
instruments
each
time.
If
applica
tions
for
which
you
will
use
the
76 3
do
not
require
the
full
available
performance
from
the
76 3
and
plug-in
units,
this
procedure
and
the
required
equipment
list
can
be
-1

Calibration
—
7623/R7623
Service
shortened
accordingly.
For
example,
the
basic
measurement
capabilities
of
this
instrument
can
be
verified
by
checking
vertical
deflection
accuracy,
vertical
square-wave
response,
and
basic
horizontal
timing
with
7000-series
real-time
plug-in
units
and
an
accurate
square-wave
signal.
Also,
if
the
76 3
is
to
be
used
in
a
fixed
system
without
the
need
to
interchange
plug-in
units,
all
tests
can
be
made
by
substi
tuting
vertical
plug-in
units
and
applicable
test
signals
for
the
067-0587-01
mainframe
standardizer
calibration
fix
ture.
TEST
EQUIPMENT
Description
Minimum
Specifications
Usage
Examples
of
Applicable
Test
Equipment
1.
Precision
DC
Voltmeter
Range,
zero
to
150
Volts;
accuracy,
within
0. %.
Calibrator
output
accuracy
check
and
adjustment.
a.
Tektronix
7D13
Digital
Multi
meter
(test
oscilloscope
must
have
Readout
System).
b.
Fluke
Model
8 5A
Differential
DC
Voltmeter.
.
DC
Voltmeter
(VOM)
Range,
zero
to
4000
Volts;
accuracy,
checked
to
with
in
1%
at
1500
Volts.
High
voltage
power
supply
check.
Z-Axis
DC
levels
adjustment.
a.
Triplett
Model
630-NA.
b.
Simpson
Model
6 .
3.
Time-Mark
Generator
Marker
outputs,
10
nano
seconds
to
0.1
second;
marker
accuracy,
within
0.1%;
Trigger
output,
one
millisecond.
CRT
geometry
check
and
adjustment.
Horizontal
timing
check
and
adjust
ment.
a.
Tektronix
901
Time-Mark
Gen
erator.
b.
Tektronix
184
Time-Mark
Gen
erator.
4.
Medium
frequency
con
stant
amplitude
signal
gen
erator
Frequency,
50
to
100
megahertz;
reference
fre
quency,
50
kilohertz;
out
put
amplitude,
variable
from
5
millivolts
to
5
Volts
peak-to-peak
into
50
ohms;
amplitude
accuracy,
constant
within
3%
of
ref
erence
as
output
fre
quency
changes.
External
Z-axis
operation
check.
Horizontal
and
Ver
tical
bandwidth
check.
Vertical
amplifier
isolation
check
and
storage
writing
rate.
a.
Tektronix
191
Constant
Ampli
tude
Signal
Generator.
b.
General
Radio
1 15-C
with
1 63-C
Amplitude
Regulating
Power
Supply.
5.
Low
frequency
signal
generator
Frequency,
35
kilohertz;
output
amplitude,
variable
from
50
to
100
millivolts.
X-Y
phase
shift
check
and
storage
writing
rate.
a.
General
Radio
1310-B
Oscillator
with
a
binding
post
to
BNC
adapter
( 74-QBJ
General
Radio).
6.
Test
oscilloscope
system
(dual-trace)
Bandwidth,
DC
to
50
megahertz;
minimum
de
flection
factor,
10
m
illi
volts/div;
accuracy,
within
3%.
Horizontal
limit
centering
adjustment
and
+
GATE
OUT.
a.
Tektronix
7503
or
7504
oscillo
scope
with
two
7A15A
or
7A16
Amplifier
and
7B50
and
7B53A
Time
Base
plug-in
units
and
two
P6053
Probes.
b.
Tektronix
453A
oscilloscope
with
two
P6054
Probes.
-

Calibration
—
7623/R7623
Service
TEST
EQUIPMENT
(cont)
Description
Minimum
Specifications
Usage
Examples
of
Applicable
Test
Equipment
7.
Vertical
plug-in
unit
(two
identical
units
re
quired),
and
a
dual
display
vertical
unit.
Tektronix
7A-series
65
megahertz
bandwidth
re
quired
for
complete
pro
cedure
as
written.
Used
throughout
pro
cedure
to
provide
vertical
input
to
the
instrument
under
calibration.
Identi
cal
units
required
for
only
X-Y
phase
shift
check.
a.
Tektronix
7A15A
and
an
7A18
Amplifier
(may
be
shared
with
an
7000-series
test
oscilloscope).
b.
Any
7A-$eries
plug-in
unit
(tol
erances
in
some
stpes
may
be
lim
ited
if
low-frequency
units
used).
8.
Time-base
plug-in
unit
Tektronix
7B-series.
Used
through
procedure
to
provide
sweep.
a.
Tektronix
7B53A
or
7B5
Time
Base.
b.
Any
7B-series
plug-in
unit.
9.
Mainframe
standardizer
calibration
fixture
Produces
gain-check
and
pulse-response
waveforms.
Used
throughout
pro
cedure
to
standardize
in
strument
so
plug-in
units
can
be
interchanged
with
out
complete
recalibra
tion.
a.
Tektronix
Calibration
Fixture
067-0587-01.
b.
Calibrated
7000-series
plug-in
units
with
suitable
signal
sources
may
be
substituted
if
lower
per
formance
is
acceptable.
10.
1
0X
passive
probe
Compatible
with
7B-series
external
trigger
input.
Chopped
mode
operation
check
(adjustment
pro
cedure).
a.
Tektronix
P6053
or
P6054
Probe
(may
be
shared
with
test
oscilloscope).
11.
T
connector
Connectors,
BNC.
External
Z-axis
operation
check.
a.
Tektronix
Part
No.
103-0030-00.
1 .
Termination
Impedance,
50-ohms;
accuracy,
± %;
connec
tors,
BNC.
Horizontal
timing
check
and
adjustment.
X-Y
phase
shift
check.
a.
Tektronix
Part
No.
011
0049-01.
13.
Dual-input
coupler
Connectors,
BNC.
Added
operation
check.
X-Y
phase
shift
check.
a.
Tektronix
Calibration
Fixture
067-05 5-00.
14.
Cable
(two
required)
Impedance,
50
ohms;
type,
RF-58/U;
length,
18
and
4
inches;
connectors,
BNC.
Used
throughout
pro
cedure
for
signal
intercon
nection.
a.
Tektronix
Part
No.
01 -0076-00
(18-inches).
Tektronix
Part
No.
01 -0057-01
(4 -inches).
15.
GR
in-line
termination
Impedance,
50
ohms;
accuracy,
± %;
connec
tors,
GR874
input
with
BNC
male
output.
External
Z-axis
operation
check.
Vertical
bandwidth
check.
Vertical
amplifier
isolation
check.
Horizontal
bandwidth
check.
a.
Tektronix
Part
No.
017-0083-00.
16.
Cable
Impedance,
50
ohms;
type
RG- 13/U;
electrical
length,
five
nanoseconds;
connectors,
GR874.
External
Z-axis
operation
check.
Vertical
bandwidth
check.
Vertical
amplifier
isolation
check.
Horizontal
bandwidth
check.
a.
Tektronix
Part
No.
017-050 -00.
-3

Calibration
—
7623/R7623
Service
Examples
of
Applicable
Test
Equipment
TEST
EQUIPMENT
(cont)
Description
Minimum
Specifications
Usage
17.
BNC
to
pin
jack
cable
Adapts
pin
jacks
to
BNC
male
connector.
18.
Screwdriver
19.
Low-capacitance
screw
driver
Three-inch
shaft,
inch
bit.
1
1/ -inch
shaft.
3/3 -
Added
operation
check.
Trigger
source
operation
check.
Astigmatism
adjust
ment.
Used
throughout
adjust
ment
procedure
to
adjust
variable
resistors.
Used
throughout
adjust
ment
procedure
to
adjust
variable
capacitors.
a.
Tektronix
Part
No.
175-1178-00
(one
supplied
as
standard
acces
sory).
a.
Xcelite
R-33 3.
a.
Tektronix
Part
No.
003-0000-00.
0.
Adapter
Connector,
GR
to
BNC
male.
Vertical
bandwidth
check.
Storage
writing
rate
check.
a.
Tektronix
Part
No.
017-0064-00.
NOTE
This
instrument
should
be
adjusted
at
an
ambient
temperature
of
+25°
C
±5°C
for
best
overall
accuracy.
1.
Remove
the
sides
and
bottom
covers
from
the
76 3
or
the
top
cover
and
side
panel
from
the
R76 3.
.
Connect
the
instrument
to
a
power
source
that
meets
the
voltage
and
frequency
requirements.
The
applied
voltage
should
be
near
the
center
of
the
voltage
range
marked
on
the
rear
panel
(see
Section
1
for
information
on
converting
this
instrument
from
one
operating
voltage
to
another).
NOTE
f
correct
line
voltage
is
not
available,
use
a
variable
autotransformer
to
provide
the
correct
input
voltage.
3.
Set
the
controls
as
given
under
Preliminary
Control
Settings.
Allow
at
least
0
minutes
warmup
before
pro
ceeding.
NOTE
Titles
for
external
controls
of
this
instrument
are
capitalized
in
this
procedure
(e.g.,
NTENS TY
).
nternal
adjustments
are
initial
capitalized
only
(e.g.,
CR
T
Grid
Bias).
Preliminary
Control
Settings
Set
the
76 3
controls
as
follows:
INTENSITY
Midrange
FOCUS
Adjusted
for
well-
defined
display
BEAM
FINDER
Out
GRATICULE
ILLUM
As
desired
VERT
MODE
Left
TRIG
SOURCE
VERT
MODE
READOUT
Midrange
NON-STORE
In
VAR
PERSIST
Out
FAST
Out
BI-STABLE
Out
MULTI
FAST
Out
INTEG
Out
SAVE
Out
MANUAL
ERASE
Out
AUTO
ERASE
Out
STORAGE
LEVEL
Counterclockwise
STORED
INTEN
Counterclockwise
AUTO
VIEW
TIME
Counterclockwise
(Persistance)
in
detent
POWER
ON
-4

Calibration
—
76 3/R76 3
Service
CALIBRATION
PROCEDURE
76 3
Serial
No.
_____________________________________
Calibration
Date
_____________________________________
Calibrated
By
______________________________________
Introduction
The
following
procedure
returns
the
76 3
to
correct
calibration.
All
limits
and
tolerances
given
in
this
procedure
are
calibration
guides,
and
should
not
be
interpreted
as
instrument
specifications
except
as
listed
in
Section
.
Index
to
Calibration
Procedure
Power
Supply
1.
Adjust
-50
Volt
Power
Supply
Page
-7
.
Check
Remaining
Power-Supply
Voltages
Page
-7
3.
Check
High-Voltage
Power
Supply
Page
-8
Display
and
Z-Axis
4.
Adjust
CRT
Grid
Bias
Page
-9
5.
Adjust
Astigmatixm
Page
-10
6.
Adjust
Trace
Rotation
Page
-10
7.
Adjust
Y-Axis
Alignment
Page
-10
8.
Adjust
Geometry
Page
-10
9.
Check
External
Z-Axis
Operation
Page
-11
10.
Check
Beam
Finder
Page
-11
Vertical
Deflection
System
11.
Adjust
Bias
Adjustment
Page
-1
1 .
Adjust
Vertical
Centering
Page
-1
13.
Check
Vertical
Gain
Page
-1
14.
Check
Vertical
Linearity
Page
-13
15.
Adjust
Vertical
High-Frequency
Com
pensation
Page
-13
16.
Check
Vertical
Amplifier
Bandwidth
Page
-13
17.
Check
Vertical
Amplifier
Isolation
Page
-14
18.
Check
Added
Operation
Page
-14
19.
Check
Alternate
Operation
Page
-14
0.
Check
Vertical
Chopped
Mode
Opera
tion
Page
-15
Triggering
System
1.
Check
Trigger
Source
Operation
Page
-16
Horizontal
Deflection
System
.
Adjust
Horizontal
Amplifier
Gain
and
Low
Frequency
Linearity
Page
-17
3.
Adjust
Horizontal
Amplifier
Centering
Page
-17
4.
Adjust
Horizontal
Amplifier
Limit
Centering
Page
-18
5.
Adjust
High
Frequency
Timing
Page
-18
6.
Check
X-Y
Phase
Shift
Page
-19
7.
Check
Horizontal
Bandwidth
Page
-19
Z-Axis
and
Auto
Focus
8.
Adjust
Z-Axis
Compensation
Page
- 0
9.
Adjust
Auto
Focus
Compensation
and
Operating
Levels
Page
- 0
Calibrator
30.
Adjust
Calibrator
Output
Voltage
Page
- 1
31.
Check
Calibrator
Repetition
Rate
Page
- 1
Signals
Out
3 .
Check
SINGLE
SWEEP
READY
OUT
Page
-
33.
Check
EXT
SS
RESET
IN
Page
-
34.
Check
VERT
SIG
OUT
Page
-
35.
Check
+
SAWTOOTH
OUT
Page
-
-5

Calibration
—
7623/R7623
Service
36.
Check
+
GATE
OUT
Rea out
Operation
37.
Check
READOUT
SYSTEM
Operation
38.
Check
READOUT
Gate
Triggered
Oper
ation
Storage
Operation
39.
Check
AUTO
ERASE
Function
40.
Check
SAFE
Function
41.
Check
Readout
Storage
Operation
4 .
Check
Storage
Test
Points
43.
Adjust
Bi-Stable
Operating
Level
44.
Check
Storage
Test
Points
Page
- 3
45.
Check
ERASE
and
FAST
Transfer
Pulses
46.
Check
High
Speed
Mesh
Pump
Pulses
Page
- 4
47.
Check
Pump
Pulse
Frequency
Page
- 4
48.
Check/Adjust
STORAGE
LEVEL
Range
49.
Check
FAST
Writing
Rate
Page
- 5
50.
Check/Adjust
FAST
Stability
Page
- 5
51.
Check/Adjust
Variable
Persistence
Mode
Page
- 6
and
Voltage
Levels
Page
- 7
5 .
Check/Adjust
Variable
Persistence
Pulse
Amplitude
Page
- 8
53.
Check/Adjust
CE
and
CE
3
Voltage
Page
- 9
Levels
Page
-30
Page
-30
Page
30
Page
-30
Page
-31
Page
-31
Page
-31
Page
-31
Page
-3
-6

Calibration
—
76 3/R76 3
Service
POWER
SUPPLY
Equipment
Required
.
DC
voltmeter
(VOM)
1.
Precision
DC
voltmeter
3.
Three-inch
screwdriver
Control
Settings
Set
the
controls
as
given
in
the
Preliminary
Control
Settings.
1.
Adjust
—
50
Volt
Power
Supply
a.
Set
the
INTENSITY
control
fully
counterclockwise.
Fig.
2-1.
(A)
Location
of
low-voltage
an
high-voltage
test
points
(right
si e
of
instrument)
(B)
Location
of
—
50
V
a justment
(Low
Voltage
Regulator
boar ).
b.
Connect
the
precision
DC
voltmeter
between
TP
-50
(see
Fig.
-1
A)
and
chassis
ground.
c.
CHECK-Meter
reading;
-50
volts
±0.1
volt.
d.
ADJUST-
—
50
volts
adjustment
R881
(see
Fig.
-1
B)
for
a
meter
reading
of
exactly
-50
volts.
e.
INTERACTION
—
Change
in
setting
of
R881
may
affect
the
operation
of
all
circuits
within
the
76 3.
.
Check
Remaining
Power-Supply
Voltages
a.
CHECK
—
Table
-1
lists
the
low-voltage
power
supplies
in
this
instrument.
Check
each
supply
with
the
precision
DC
voltmeter
for
output
voltage
within
the
given
tolerance
(connect
meter
ground
lead
to
chassis
ground).
Power
supply
test
points
are
shown
in
Fig.
-1
A.
NOTE
Ripple
and
regulation
of
the
individual
power
supplies
can
be
checked
using
the
procedure
given
under
Troubleshooting
Techniques
in
Section
4.
TABLE
-1
Power
Supply
Tolerance
Power
Supply
Test
Point
Output
Voltage
Tolerance
-50
Volt
Pin
8
P1
170
±0.1
Volt
-15
Volt
Pin
7
P1170
±0.1
Volt
+5
Volt
Pin
5
P1
170
±0.07
Volt
+15
Volt
Pin
4
P1170
±0.1
Volt
+50
Volt
Pin
3
P1
170
±0.3
Volt
+130
Volt
Pin
P1
170
±5.
Volt
-7

Calibration
—
7623/R7623
Service
3.
Check
High-Voltage
Power
Supply
a.
Push
in
the
power
switch
(turn
off
the
instrument).
b.
Set
the
DC
voltmeter
(VOM)
to
measure
at
least
1500
volts.
Then,
connect
it
between
the
high-voltage
test
point
(see
Fig.
2-1
A)
an
chassis
groun .
c.
Pull
out
the
power
switch
(turn
on
the
instrument).
Check
meter
rea ing;
—
1475
volts
±45
volts.
.
Push
in
the
power
switch
(turn
off
the
instrument).
Disconnect
the
DC
voltmeter.
e.
Pull
out
the
power
switch
(turn
on
the
instrument).
-8
Other manuals for 7623
1
This manual suits for next models
1
Table of contents
Other Tektronix Storage manuals
Popular Storage manuals by other brands

HP
HP StorageWorks XP48 - Disk Array Configuration guide

Intenso
Intenso 3.5" MEMORYTOWER Manual guide

Mercia Garden Products
Mercia Garden Products 0628POPA0603FGDD3TW-V1 General instructions

Seagate
Seagate ST3600002FC - Cheetah 600 GB Hard Drive product manual

Seagate
Seagate FreeAgent GoFlex Series Reviewer's guide

Texas Memory Systems
Texas Memory Systems RamSan-500 user manual