Models 3000 and 3010: Turn power off. Turn power back on and momentarily press the system
button.
-------------------------------------------------------------------------------------------------------------------------------
Models 3020 T, P & M: Turn power off. Turn power back on and momentarily turn the
enter/escape switch to the escape position.
-------------------------------------------------------------------------------------------------------------------------------
Model 3160: See pages 4 - 5 of the model 3160 manual.
-------------------------------------------------------------------------------------------------------------------------------
Model 6000: Turn power off. Turn power back on and while on the introduction screen press the
system button for at least 15 seconds.
-------------------------------------------------------------------------------------------------------------------------------
Model 6020: Turn power off. Turn power back on and while on the introduction screen hold the
enter/escape switch to the escape position for at least 15 seconds.
-------------------------------------------------------------------------------------------------------------------------------
Model 7300 and 7320: Turn power off. Turn power back on and while on the introduction
screen press the enter button for 7 seconds to erase the calibration data such as zero and span.
If you continue to hold the enter button for at lease 15 seconds all data will be erased.
-------------------------------------------------------------------------------------------------------------------------------
Model 9070: See page 44, section 6.1 of the model 9070 manual.
-------------------------------------------------------------------------------------------------------------------------------
Model 9950: See pages 4 - 8 of the model 9950 manual.
-------------------------------------------------------------------------------------------------------------------------------
Model MAX 5: (Clearing the Ram) Push the up & down arrow keys simultaneously.
Display: 1 thru 8
Push the down arrow key once. Push the enter key twice.
Display: O2=15%, Comb=0.00%, CO=0.00 PPM or %.
Remove the O2 cell and perform the procedure in the manual for electronic zeroing of the O2
sensor, page 20.
Re-install the O2 cell and perform AUTOCAL.