Testec SI7000 Series User manual

INSTRUCTION MANUAL
TT-SI 7005
60MHz active differential pr be
x1 / x5 /x10


Man_7005_1019 www.testec.de 1
Safety Summary
To avoid personal injury and/or product damage, review and comply
with the following safety precautions. These precautions apply to both operating and
maintenance personnel and must be followed during all phases of operation, service,
and repair of this probe.
A
WARNING
statement calls attention to an operating procedure, practice,
or condition, which, if not followed correctly, could result in injury or death to
personnel.
A
CAUTION
statement calls attention to an operating procedure, practice, or
condition, which, if not followed correctly, could result in damage to or destruction
of parts or the entire product.
Do Not Work Alone
Do not work alone when working with high voltages.
Inspect the Probe
Inspect the probe and accessories for cracks and frayed or broken leads before each
use. If defects or damages are noted, D N T USE the probe.
Dry Conditions
Hands, shoes, floor, and work bench must be dry. Avoid making
measurements under humidity, dampness, or other environmental conditions
that might affect safety.
Do Not Remove the Probe’s Casing
Removal of the probe’s casing may expose you to electric shock. If necessary,
disconnect the inputs and outputs of the probe before opening the case.
Hazardo s Contact
To avoid injury, remove jewelry such as rings, watches, and other metallic objects. Do
not touch exposed connections and components when power is present.
Unexpected Charges
Hazardous voltages may be present in unexpected locations in circuitry being
tested when a fault condition in the circuit exists.
Capacitors inside the instrument may retain a charge even if the instrument is
disconnected from its source of supply.
Use Only in Office-Type Indoor Setting
The probe is designed to be used in office-type indoor environments.
Do not operate the probe:
•
In the presence of noxious, corrosive, flammable fumes, gases, vapors,
chemicals, or finely-divided particulates.
•
In environments where there is a danger of any liquid
spilled
on the probe.
•
In air temperatures exceeding the specified operating
temperatures.
•
In atmospheric pressures outside the specified altitude limits or
where the
surrounding gas is not air.

Man_7005_1019 www.testec.de 2
Not for Critical Applications
This probe is not authorized for use in contact with the human body or for use as a
component in a life-support device or system.
Do Not S bstit te Parts
Do not install substitute parts or perform any unauthorized modification to the
instrument.
Only Q alified Personnel
nly qualified personnel should use this probe. This differential voltage probe is
designed to be used by personnel who are trained, experienced, or otherwise
qualified to recognize hazardous situations and who are trained in the safety
precautions necessary to avoid possible injury when using such a device.
Observe Maxim m Working Voltage
Do not use any probe above its maximum working voltage ranges. See
specifications on page 7.
Use Proper Power So rce
Do not operate this probe from a power source that applies more than the voltage
specified.
M st be Gro nded
This probe is grounded by the shell of the BNC connector through the grounding
conductor of the power cord of the measurement instrument. Before making
connections to the input leads of this probe, ensure that the output BNC connector is
attached to the BNC connector of the measurement instrument, and that the
measurement instrument is properly grounded. Whenever it is likely that the ground
protection is impaired, you must make the instrument inoperative and secure it
against any unintended operation.
Terms and Symb ls
The following symbols appear on the product or in its documentation:
Direct voltage
Both direct and alternating voltage
Caution, possibility of electric shock
Caution, see documentation for details
Earth Ground

Man_7005_1019 www.testec.de 3
Definiti ns
Measurement Category II (CAT II)
refers to local-level electrical distribution, such as that provided by a standard wall
outlet or plug-connected equipment. Examples of CAT II measurements would be
household appliances, portable tools, and similar modules.
Measurement Category III (CAT III)
refers to measurements on hard-wired equipment in fixed installations, distribution
boards, and circuit breakers that form part of a building wiring installation. ther
examples are wiring, including cables, bus bars, junction boxes, switches, socket
outlets in the fixed installation, and stationary motors with permanent connections to
fixed installations
ollution Degree 2
refers to an operation environment where normally only dry, non-conductive pollution
occurs. Temporary conductivity caused by condensation can be expected.
Working CAT rating is eq al to that of the lowest rated element
within the test set- p.
C mpliance Statements
EC Declaration of Conformity
The product conforms to the applicable European Union requirements per
IEC 61010-031:2015 Safety requirements
for electrical equipment for measurement, control and laboratory use.
Part 31: Safety requirements for hand-held probe assemblies for electrical
measurement and test.
EU RoHS Compliance
The probe and accessories conform to the 2011/65/EU RoHS2 Directive.
Disposal of Old Electrical & Electronic Eq ipment
(Applicable in the European Union and other European countries with
separate collection systems). This product is subject to Directive
2012/19/EU of the European Parliament and the Council of the European
Union on waste electrical and electronic equipment (WEEE), and in
jurisdictions adopting that Directive, is marked as being put on the market
after August 13, 2005, and should not be disposed of as unsorted
municipal waste. Please utilize your local WEEE collection facilities in the
disposition of this product and
otherwise observe all applicable requirements.
This probe is in compliance with IEC 61010-031:2015 CAT III, Pollution Degree 2.

Man_7005_1019 www.testec.de 4
1 Intr ducti n
Overview
The TT-SI 7005 differential probe allows safe, accurate measurement between two
voltage points where neither point is referenced to ground. The probe is designed for
high sensitivity measurements up to 60 MHz bandwidth. Compatible with
oscilloscopes from all major manufacturers, the probe is exclusively powered by the
included 9 V power adapter.
Feat res
•
Meets IEC 61010-1:2015 safety standard
•
Selectable attenuation settings of x1, x5 and x10
•
60 MHz bandwidth @ x10 attenuation setting
•
50 MHz bandwidth @ x5 attenuation setting
•
15 MHz bandwidth @ x1 attenuation setting
•Up to ±35 V (DC + AC peak) common mode
•
High accuracy (±2%)
•
Power indicator LED
•
ver range indicator LED
Initial Inspection
This unit is tested prior to shipment. It is therefore ready for immediate use upon
receipt. An initial physical inspection should be made to ensure that no damage has
been sustained during shipment. After the inspection, verify the contents of the
shipment.
Delivery Content
•
Differential probe - TT-SI 7005
•
2 x Grabber, black & red – TT-SI GR2
•
2 x test leads with insulated 4 mm banana plugs – TT-SI TL7
silicone jacket, black & red
•insulated BNC cable, 100 cm – TT-SI BN7
•9 V power adapter – EU version – TT-SI NT7
•User manual
2 Using the Pr be
WARNING
At the time of powering on the probe, the input leads must not be connected to an item
to be tested. Never operate the probe with the case open.
CAUTION
This probe is used to carry out differential measurements between two points on the
circuit under test. This probe is not designed for electrically insulating the circuit under
test or the measuring instrument.

Man_7005_1019 www.testec.de 5
Inspection Proced re
1. Connect the BNC output connector to the vertical input of the oscilloscope.
2. Power on the probe.
3. Set the attenuation setting on the oscilloscope to match the probe.
4. Connect the input of probe to a function generator. Then select a square-wave
output of 10 V amplitude and 100 kHz frequency.
5. The square-wave will be displayed on the screen of the oscilloscope. This
indicates the probe is working properly.
Note that the oscilloscope should show the same voltage and frequency as
the function generator.
Getting Started
1. Connect the BNC output connector to the vertical input of a general purposed
oscilloscope. The oscilloscope must have a ground referenced.
2. Connect power adapter.
3. Select the proper attenuation ratio. When measuring signals below 3,5 V,
switch the attenuation ratio to x1 in order to get higher resolution and less
noise ratio. For voltages up to 18 V, set the attenuation ratio to x5 and for
voltages up to 35 V set the attenuation ratio to x10.
4. The power indicator LED should turn on.
5. Plug the supplied grips onto the test leads and connect the circuit under test.
Vertical Scale on Oscilloscope
The actual vertical scale of the oscilloscope is equal to the attenuation factor
multiplied by the range of vertical scale selected on the oscilloscope.
For example, with the probe on factor x00 and the oscilloscope on 0.5 V/div, the
real vertical scale is 10 x 0.5 = 5 V/div.
With the probe on x5, the real vertical scale is 5 x 0.5 = 2,5 V/ div.
These values apply when the oscilloscope is set to the typical 1 MΩ impedance
input. When the scope is set to 50 Ω input, the actual vertical scale will be
doubled:
10 V/div for the x10 setting and 5 V/div for the x´5 setting.
Sc pe Input
Pr be
Actual
Vertical Scale
Actual Vertical
Impedance
Attenuati n
Attenuati n
Reading n
Scale f the
Setting
Setting
Oscill sc pe
Oscill sc pe
1 MΩ
x1
x1
0,5 V/div
0,5 V/div
1 MΩ
x5
x5
0,5 V/div
2,5 V/div
1 MΩ
x10
x10
0,5 V/div
5 V/div
50 Ω
x1
x2
0,5 V/div
1 V/div
50 Ω
x5
x10
0,5 V/div
5 V/div
50 Ω
x10
x20
0,5 V/div
10 V/div

Man_7005_1019 www.testec.de 6
Pr be Input
–
2 x 4mm Plug
Pr be Output
BNC C nnect r
P wer Indicat r (Red)
Over Range Indicat r (Red)
Attenuati n Select r Switch
External P wer Supply
3 Pr duct Overview
Included Accessories
TT-SI NT7 TT-SI TL7 TT-SI GR2 TT-SI BN7

Man_7005_1019 www.testec.de 7
4 Specificati ns
All specifications apply to the unit after a temperature stabilization time of 20
minutes over an ambient range of 25°C ± 5°C.
TT-SI 7005 – order number 15176
Bandwidth
60
MHz @ x
10
50 MHz @ x5
15 MHz @ x1
Rise Time
5,8
ns @ x
10
7 ns @ x5
23 ns @ x1
Attenuation
x
1
, x5, x
10
DC
-
Accuracy
±2%
CMRR
80
dB @ 60
Hz
60 dB @ 100 Hz
50
dB @ 1
MHz
Maximum Input Voltage @ x
1
(DC + AC peak)
±
3,5
V
Maximum Input Voltage @ x
5
(DC + AC peak)
±
18
V
Maximum Input Voltage @ x
10
(DC + AC peak)
±
35
V
Absolute Maximum Input Voltage
(each side to ground)
600
Vrms
Input Impedance
(differential)
1
M hm //
2,5
pF
Input Impedance
(each side to
ground)
0,5
M hm //
6
pF
utput Voltage Swing
±
4
V
using 1 M
Ω oscilloscope input
ffset (typical)
±5
mV
Noise (typical)
2
mVrms
Source Impedance
50
hm
Power Supply
9 V power adapter (included)
Weight
250
g
Dimensions
195 x 55 x 30
mm
BNC
cable length
100
cm
Input lead length
55
cm each
peration Tempearure/Humidity
0°C to 50°C / 10% to 85%
RH
Storage Tempearure/Humidity
-
30
°C to
7
0°C / 10% to
90
%
RH
Safety Standard
IEC
6
1010
-
031:2015
CAT II
I
Pollution Degree
2
Specifications are subject to change without notice. To ensure the most current
version of this manual, please download the current version from our website:
www.testec.de

Man_7005_1019 www.testec.de 8
5
V ltage Derating Curve
The derating curve of the absolute maximum input voltage in common mode is
shown as follows:
TT-SI 7005 Voltage Derating
Figure: TT-SI 7005 Derating Curve
6 Cleaning
This probe does not require any particular cleaning. If necessary, clean the case
with a soft cloth.
WARNING
Dry the probe thoroughly before attempting to make voltage
measurements.
CAUTION
Avoid immersing or using abrasive cleaners or solvents containing
Benzene (or similar solvents) on the probe as these can cause
deterioration of the probe body and cables.

Man_7005_1019 www.testec.de 9
7 Service & Warranty Inf rmati n
Limited One-Year Warranty
Testec Elektronik GmbH warrants these products to be free from defective material
or workmanship for a period of 1 year from the date of original purchase. Under
this warranty, Testec Elektronik GmbH is limited to repairing the defective device
when returned to the factory, shipping charges prepaid, within the warranty
period.
Units returned to Testec Elektronik GmbH that have been subject to abuse,
misuse, damage, or accident, or have been connected, installed, or adjusted
contrary to the instructions furnished by Testec Elektronik GmbH, or that have
been repaired by unauthorized persons, will not be covered by this warranty.
Testec Elektronik GmbH reserves the right to discontinue models, change
specifications, price, or design of this device at any time without notice and without
incurring any obligation whatsoever.
The purchaser agrees to assume all liabilities for any damages and/or bodily injury
which may result from the use or misuse of this device by the purchaser, his
employees, or agents.
This warranty is in lieu of all other representations or warranties expressed or
implied and no agent or representative of Testec Elektronik GmbH is authorized to
assume any other obligation in connection with the sale and purchase of this
device.
Service
If you have a need for calibration or repair services, technical, or sales support,
please contact us:
Testec Elektr nik GmbH
Fritz-Klatte-Str. 6
65933 Frankfurt / Germany
Tel.: +49 (0) 69 – 9433350
E-Mail: [email protected]
www.testec.de

Man_700_0919 www.testec.de 1
TT-SI 7005 Differential Probe User’s Manual
© 2019 TESTEC Elektronik GmbH. All rights reserved.
Unauthorized duplication of TESTEC documentation materials is strictly prohibited.
Customers are permitted to duplicate and distribute
TESTEC documentations for internal educational purposes
This manual suits for next models
2
Table of contents
Other Testec Test Equipment manuals
Popular Test Equipment manuals by other brands

Hydrajaws
Hydrajaws 2008 operating instructions

Interstate Batteries
Interstate Batteries IB Pulse Testing Program Manual

M-system
M-system Mini-M M2AMS2 instruction manual

Besantek
Besantek BST-IT30 instruction manual

Bego
Bego Nautilus T Translation of the original instructions

Bacharach
Bacharach R-22id Operation and maintenance guide