TKA UV-Radiometer User manual

TKA SCIENTIFIC INSTRUMENTS
UV-Radiometer
ТКА-UV
Operation Manual
St. Petersburg
2008
Generated by Foxit PDF Creator © Foxit Software
http://www.foxitsoftware.com For evaluation only.

Table of contents
Introduction ……………………………………………………….. 3
Intended purpose and application …………………………………. 3
Basic technical characteristics ………………………. …………….3
Installation and maintenance . . . . . . . . . . . . . . . .. ………………..4
Scope of supply . . . . . . . . . . . . . . . . . . . . . . . ………………………5
Installation of the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Maintenance of the device . . . . . . . . . . . . . . . . . . . . ……………….6
Calibration of the device ………………………………………….. 8
Dark current measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . ……9
Determining spectral sensitivity of the device . . . . . ………………10
Determining range lacing ratios . . . . . . . . . ……………………….11
Calculating the total irradiance ratio ……………………………….12
Additional energy adjustment by zones . ………………………... . 13
Basic operating program . . . . . . . . . . . . . . . . . . . . . . . . ………. . . 14
Program setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . …………..15
Program operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ….16
Storage and transportation rules . . . . . . . . . . . . . . . . ………….....16
Enclosure A. Technical support . . .……………………………….. 17
Enclosure B. Agreement to use licensed software . . . . . . . . . . . . .. 18
Enclosure C. Warranty . . . . . . . . . . . . . . . . . . . . .. …………………19
Notice of title
The information contained in this manual derives from the patent and the title data of the company
LLC TKA SCIENTIFIC INSTRUMENTS.
The contents of this manual are the property of the company TKA SCIENTIFIC INSTRUMENTS
and protected by copyright. Any full or partial reproduction of this manual’s contents is prohibited.
This publication does not grant any rights to reproduce or use this manual for any purposes different
from the installation, operation, or maintenance of the device and its software. Neither part of this
manual can be reproduced, copied, stored in a data retrieval system or translated into another (includ-
ing computer) language in any form or through any means: electronic, magnetic, mechanical, optical,
manual or otherwise without prior written consent of a governing official of the company TKA SCI-
ENTIFIC INSTRUMENTS.
The copyright is owned by © LLC TKA SCIENTIFIC INSTRUMENTS, 2008, ALL RIGHTS PRO-
TECTED
ТКА® — is a registered trademark of TKA SCIENTIFIC INSTRUMENTS
The names of the actual companies and products mentioned herein can be trademarks of the respec-
tive owners.
Generated by Foxit PDF Creator © Foxit Software
http://www.foxitsoftware.com For evaluation only.

Introduction
This operation manual is designed for familiarization with the basic operation of the device, particu-
lars of its design, storage rules, and operational procedure.
Intended purpose and application
The device UV-Radiometer “ТКА – UV» (hereinafter referred to as the device) is designed for
measuring irradiance (in mW/m2) created:
in the spectral range UV - А (315 … 400) nm by UV sources,
in the spectral range UV - В (280 … 315) nm by UV sources,
in the spectral range UV - С (220 … 280) nm by UV sources.
The sphere of application of the device: sanitary and technical surveillance in residential and indus-
trial rooms, certification of workplaces, and other spheres of activity.
Basic technical characteristics
Wight, kg (not above) with power module 2.0
Operation parameters
Ambient temperature, оС:
– normal operation conditions
– operation temperature range 20±5
0…40
Relative air humidity at ambient temperature25°С, %, not above 95
Atmospheric pressure, kPa 80…110
Mean time between failures with confidence probability р = 0.8, h, not below 2000
Irradiance meas
urement
•Measurement range, mW/m21 … 60 000
•Limit of the permitted basic relative error
of irradiance measurements, % ±10,0
Continuous operation from battery, hours (not below)
from the mains, hours (not below) 2,0
8,0
Overall dimensions, mm (not over):
— photometer head
—
laptop
ASUS
Eee PC 701 (
closed
)
185х70х55
225x165x35
It is prohibited to measure high irradiance over a long time!
It is recommended to use a heat-resistance lens cover supplied with the device.
Generated by Foxit PDF Creator © Foxit Software
http://www.foxitsoftware.com For evaluation only.

Installation and maintenance
UV-Radiometer of the ТКА-UV model is a grating-based polychromator with registration of decom-
posed radiation by a microprocessor-controlled linear photodiode array, sending digital data to a lap-
top computer compatible with OS Windows® PC — ASUS® Eee PC 701, where the signal is digi-
tally processed with subsequent displaying of data on the irradiation levels.
Eee PC 701 is a 7-inch laptop which is a classic notebook, with a lockless cover. The laptop has no
hard disk, its function is performed by a 4 Gb flash memory. The laptop is supplied with a compact
power unit resembling more a charger for a mobile phone than the one for a laptop.
Important! It is recommended before the beginning of operation to familiarize yourself with the Brief
operation manual of the laptop ASUS Eee PC 701.
Generated by Foxit PDF Creator © Foxit Software
http://www.foxitsoftware.com For evaluation only.

The scope of supply includes separate calibrating servicing programs (utilities) for operation system
Windows. All you require is start the calibrating utility GradUV.exe, enter reference values into the
specified fields. In the operation mode start the utility VDUV.exe, the main window of which con-
tains the signal spectrum and energy characteristics.
The contents of the scope of supply
The scope of supply (package) of your system ТКА-UV should contain all elements listed and illus-
trated below. When any element is missing, contact an employee of TKA SCIENTIFIC INSTRU-
MENTS.
• Photometer head
• Laptop ASUS Eee PC 701 with power unit
• Cable USB2.0, connecting USB A-B
• Documentation and registration materials
Installation of the device
To capture (register) a digital signal with the photometer head, it must be connected directly to one of
the serial ports of the laptop.
To install the device:
1. Remove the device from the package and unfold the cable ends.
2. Connect the main connecting plug (1) to the serial port on the side wall of the laptop.
3. Connect the second plug (2) to the photometer head.
4. Connect the battery or power unit to the laptop.
5. Switch up the laptop.
Generated by Foxit PDF Creator © Foxit Software
http://www.foxitsoftware.com For evaluation only.

Maintenance of the device
The following maintenance procedure will insure the reliable operation of your device.
Checking the serial numbers
The first thing to do after the unpacking of your system is get assured that the five digits of the fac-
tory number on the photometer head match the last five digits in the full name of the computer.
For instance, for the device’s factory number 14003 the computer’s name must be TKA-14003.
Note: To learn the name of the computer at which you are working, you should perform the follow-
ing:
1. At the icon My computer press the right button of the mouse and select the menu item Features.
2. In the opened window of the features select the Computer name, and in the field Full name you
will see your computer’s name.
2
1
Generated by Foxit PDF Creator © Foxit Software
http://www.foxitsoftware.com For evaluation only.

Your photometer head must be calibrated exceptionally with the help of the laptop attached to it. If
the number of the photometer head and the name of the computer does not match, immediately return
the device to R&DC ТКА for compete exchange or apply for technical support to make good the in-
compatibility.
Cleaning the device
If the device is not in use, store it in a dry and clean place. Clean the device once a week or when
necessary:
• Carefully wipe with a cloth (without lint) the eye on the photometer head.
Do not use any solvents or cleansers.
• Safe storage insures proper maintenance of the optical part of the device.
Do not let splashes of cleansers or a jet of compressed air to contact the optical part.
Generated by Foxit PDF Creator © Foxit Software
http://www.foxitsoftware.com For evaluation only.

Calibration of the device
Buttons to select a calibration step:
«1 – ТТ - Step 1. Removal of dark current
«2 - СЧП - Step 2. Determining the spectral sensitivity of the device
«3 - КСД - Step 3. Determining the range lacing ratios
«4 – ПК - Step 4. Calculating the total irradiance ratio
«5 – ПК addit - Step 5. Additional adjustment by zones
The buttons are not available, if the device is not switched up or not found through any reason.
Program operation.
Important! For the calibration, it is necessary to braid into the device a program of manual switching
of ranges (with the help of WSD-application file grad.hex).
Configuration of WSD-application:
Field of selecting a
calibration step
Generated by Foxit PDF Creator © Foxit Software
http://www.foxitsoftware.com For evaluation only.

The device should be calibrated in the specified order. The program does not control the operator’s
actions on the selection of the step, therefore when the order is infringed, failures in operation are
possible.
The selection of the calibration step is effected through pressing the respective button.
Dark current measurement
Generated by Foxit PDF Creator © Foxit Software
http://www.foxitsoftware.com For evaluation only.

When fulfilling the requirements of the program, it is necessary to close the lens of the device and
install with the range selection button (the button on the photometer head increases the value succes-
sively) with the recommended program (a black digit).
On the panel on the right the current measurement range is displayed (a green digit) and the highest
value. If the current range does not match the required one, then the digit of the current range will be
displayed in the red color and the button of starting the measurement in the current window of the
program will be unavailable. Specifying the required range with the button on the photometer head,
start the measurement by pressing the button Measure in current range in the current window of the
program.
When the measurement is completed, the program will offer to make a measurement on the next
range.
There are 9 ranges altogether, the integral action time is from 4 seconds (range 0) to 16 msec (range
8).
When the measurement is completed in all ranges the program will preserve the matrix of dark cur-
rents and return to the selection of a calibration step.
Determining spectral sensitivity of the device
Place the photometer head in parallel to the plane of the item which is being measured. See to it that
the shadow of the operator who conducts the measurement does not cast itself onto the photohead
window, as well as the shadow of any irrelevant objects that are temporarily present.
Generated by Foxit PDF Creator © Foxit Software
http://www.foxitsoftware.com For evaluation only.

To operate on this step, install the photohead on the photometer bench and switch up the reference
source DNK-90.
The basic operation is the same as in step 1.
Measurement in each range should be conducted with the steady-state value of the maximum close to
4,000 units.
Do not allow overfilling (4095)!
If it is impossible to reach the normal level of the spectral sensitivity on the last ranges, the button
Low Level should be pressed in the current window of the program; the remaining ranges will be
filled with the values of the last one which was successfully measured.
Determining the range lacing ratios
The ratios are calculated following the same scheme as in steps 1 and 2.
Place the photometer head in parallel to the plane of the item which is being measured. See to it that
the shadow of the operator who conducts the measurement does not cast itself onto the photohead
window, as well as the shadow of any irrelevant objects that are temporarily present.
To begin the measurements, place the photohead on the bench so that the maximum signal
3,800 … 4,000 units might be obtained. After the measurement in the selected range, not changing
the position of the photohead, increase with the button on the head the range to the next one and re-
peat the measurement.
Generated by Foxit PDF Creator © Foxit Software
http://www.foxitsoftware.com For evaluation only.

The quotient of the measured signals for each will be preserved. Then it is necessary to reach again
the maximum level of the signal 3,800 … 4,000 units, shifting the photohead on the bench, and to
make two measurements on adjoining ranges. In case it is impossible to increase the signal to the ac-
ceptable value, press the button Low Level, and the remaining ratios will be calculated on the basis of
the available measurements which were successfully made.
Generated by Foxit PDF Creator © Foxit Software
http://www.foxitsoftware.com For evaluation only.

Calculating the total irradiance ratio
Place the photometer head in parallel to the plane of the item which is being measured. See to it that
the shadow of the operator who conducts the measurement does not cast itself onto the photohead
window, as well as the shadow of any irrelevant objects that are temporarily present.
In this mode the device is installed on the bench so that on the zero range the maximum signal might
be obtained in the area of 3,900 … 4,000 units. The position on the bench is fixed, the photohead re-
moved. Then, in the fixed point reference receivers determine irradiance in the wavelength range 220
… 400 nm (zones А, В, С). After that, the photohead returns to the reference point, and the irradiance
value is entered into the entry field, and the total irradiance ratio calculated.
Generated by Foxit PDF Creator © Foxit Software
http://www.foxitsoftware.com For evaluation only.

Additional energy adjustment by zones
Place the photometer head in parallel to the plane of the item which is being measured. See to it that
the shadow of the operator who conducts the measurement does not cast itself onto the photohead
window, as well as the shadow of any irrelevant objects that are temporarily present.
By the method stated at the previous step, irradiance is determined separately by zones А1, А2, В and
С.
Measurement can be made in any range. It is recommended for calibration to select the level of ir-
radiance at which the maximum value is at the level of 3,000 … 3,900 units.
Into the field of the respective UV zone the measured value is entered and calculation made. Above
the entry field it is necessary to place a flag for confirming the entry of changes. When the flag is
removed, the changes for a given field will not be rewritten.
The values of the obtained ratios are used thereafter as supplements to the ratio of total energy ad-
justment. The button Value Reset serves for zeroing the calculated ratios. In this case irradiance is
calculated only with taking the total ratio into account.
Output data
After the fulfillment of the necessary operations the calibration program retains all the calibration
data in the Windows register into the branch
«HKEY_CURRENT_USER\Software\TKA\VDUV\1.0».
Generated by Foxit PDF Creator © Foxit Software
http://www.foxitsoftware.com For evaluation only.

Basic operating program
By default the main window of the program contains the signal spectrum and energy characteristics.
Output of energy characteristics only can be adjusted (see item Setup).
Generated by Foxit PDF Creator © Foxit Software
http://www.foxitsoftware.com For evaluation only.

Program setup
The user zone range specifies the irradiance measurement range within the user-assigned boundaries.
Options of Graph group control the spectrum display.
Display group allows the format of the main window to be set up. The switch Show text only
switches the main window of the program to the mode of either display of irradiance only (turned
on), or the mode of simultaneous display of irradiance and signal spectrum (turned off).
In the mode of simultaneous display the irradiance output can be controlled for each zone.
Important! For the operation mode it is necessary to braid into the device a program of manual
switching of ranges (with the help of WSD-application file work.hex).
Configuration of WSD-application:
Generated by Foxit PDF Creator © Foxit Software
http://www.foxitsoftware.com For evaluation only.

Program operation
Place the photometer head in parallel to the plane of the item which is being measured. See to it that
the shadow of the operator who conducts the measurement does not cast itself onto the photohead
window, as well as the shadow of any irrelevant objects that are temporarily present.
Upon being started, the program scans the USB bus and in case the device is detected displays Ready
in the status line and starts automatically to display the data received from the photohead. When there
is no connection, the program displays Searching… in the status line.
The four buttons of the control panel are designed for storing data being received into a text file, the
spectrum graph into a file, setting up the program, and outputting additional information correspond-
ingly. The spectrum graph is stored in BMP format.
For the user’s convenience data are stored into a text file also through pressing the button ENTER.
The name of the file in this case is formed automatically in the format
tka_data_YYYYMMDD_HHMMSS. The obtained file can be downloaded into any word processor
or electronic table for further processing.
Storage and transportation rules
The device must be stored in a dry heated room under the conditions corresponding to the group 1
GOST15150-69 at the temperature +5 to +40 0С and relative humidity not above 85% at 250С.
The air in the room should not contain admixtures of aggressive vapors or gases.
The transportation of the devices in the manufacturer’s packing can be made by any closed transport
without speed limitations GOST 15150-69 at the ambient temperature from 10 below zero to 60 0С
above zero and relative humidity up to 98% at the temperature 35 0С. After transportation at tempera-
ture below 00С it should be unpacked should only be effected after holding it in the packed condition
at the temperature (20±5) 0С for at least 12 hours.
Air transportation is not allowed.
An acceptance certificate
The device UV-Radiometer ТКА-UV, factory No.
is acknowledged ready for operation.
Date of release: “____” _______ 200
Stamp QCD: ___________________________________
Date of sale: “____” _____________ 200___
Generated by Foxit PDF Creator © Foxit Software
http://www.foxitsoftware.com For evaluation only.

Enclosure A
Technical support
UV-Radiometer ТКА-UV is used with the special software.
If you have any problems concerning directly the device or means of calibration, contact the Service
Department of LLC TKA SCIENTIFIC INSTRUMENTS.
E-mail of Technical Support: akopyan_igor@mail.ru, aduc812@list.ru
Phone/fax of the Service Department: +7(812) 274-06-42, +7(812) 274-49-01
The center LLC TKA SCIENTIFIC INSTRUMENTS can render services in technical support of
software. The technical support is performed under the conditions specified by the manufacturer, or
stated on the web-site www.tka.spb.ru, or described in the user’s manual or other documentation. In
the process of rendering technical support services the manufacturer has the right to require from you
the information relating to the technical parameters of your equipment. The manufacturer has the
right to use the above information for the purposes of business development, including rendering of
services in technical support.
The service department of technical support of users of the device UV-Radiometer ТКА-UV is fully
comprised of qualified technicians for rendering assistance to you by phone or fax. For such case
please have the following information at hand:
• Serial No. of your device and serial No. of the laptop
• Software version No.
• Name of the company and your name
• Your phone No.
• Record any error messages and what led to them
• Is the computer or software able to operate within the range of the telephone
Generated by Foxit PDF Creator © Foxit Software
http://www.foxitsoftware.com For evaluation only.

Enclosure B
Agreement to use licensed software of UV-Radiometer ТКА-UV
IMPORTANT: READ THIS AGREEMENT CAREFULLY. THIS IS A LEGAL AGREEMENT BETWEEN TKA
SCIENTIFIC INSTRUMENTS AND YOU (ACTING AS AN INDIVIDUAL OR, IF APPLICABLE, ON BEHALF
OF THE INDIVIDUAL OR ENTITY ON WHOSE COMPUTER THE SOFTWARE IS INSTALLED). IF YOU
CLICK THE "ACCEPT" OPTION OR INSTALL THE SOFTWARE PRODUCT ACCOMPANYING THIS
AGREEMENT (THE "SOFTWARE") YOU WILL BE BOUND BY THIS AGREEMENT. You are not required
to accept these terms but, unless and until you do, the Software will not install and you will not be au-
thorized to use the Software.
If you do not wish to agree to these terms, please click "DO NOT ACCEPT" and obtain a refund of the
purchase price as follows:
* If you purchased a download, follow the instructions on the confirmation email you received in con-
nection with the purchase. Please allow thirty (30) days for the credit to be applied to your account.
* For all other purchases, destroy all copies of the Software you may have made (including archival
copies) and return the Software in its original package, along with your receipt, to the point of pur-
chase.
NOTICE: THE SOFTWARE MAY CONTAIN A "TIME OUT" FEATURE THAT WILL RENDER THE SOFT-
WARE INOPERATIVE AT THE END OF THE TERM (AS DEFINED BELOW).
END USER SOFTWARE LICENSE AGREEMENT
1. License Grant; Related Provisions.
a. Grant of License. TKA SCIENTIFIC INSTRUMENTS, subject to the terms and conditions of this
Agreement, hereby grants to you a non-exclusive and non-transferable license during the Term to use
the Software only in executable or object code form solely for your personal or internal business pur-
poses. This license authorizes you to use the Software on or in conjunction with up to (i) the number
of computers specified by your order for the Software, or (ii) in the case of Software purchased on a
CD or other physical medium, the number specified on the Software package, or (iii) if you received the
Software in combination with other hardware or software, solely in conjunction with such other hard-
ware or software.
b. Limitations. You shall not, and shall not permit any third party to, (i) duplicate the Software for
any purpose other than as reasonably necessary to use the same as contemplated by this Agreement
and for off-line archival and disaster recovery purposes, (ii) except for temporary transfer in the event
of computer malfunction, install the Software on a second computer, (iii) use any license number sup-
plied by TKA SCIENTIFIC INSTRUMENTS (each, a "License Number") in connection with more than one
(1) copy of the Software at any time, (iv) disclose any License Number to any party other than TKA
SCIENTIFIC INSTRUMENTS or, as required to obtain support services in respect of the Software, TKA
SCIENTIFIC INSTRUMENTS designated representatives, (v) publish the Software or any License Num-
ber or use the same other than for the purposes described in Section 1.a of this Agreement, (vi) except
as expressly authorized by law, reverse engineer, disassemble, decompile, translate, reconstruct,
transform or extract the Software or any portion of the Software (including without limitation any re-
lated malware signatures and malware detection routines), (vii) except as expressly authorized by law,
change, modify or otherwise alter the Software (including without limitation any related malware signa-
tures and malware detection routines), (viii) transfer, pledge, rent, share or sublicense the Software
other than in connection with the sale, lease, rental or other transfer of the computer on which it is ini-
tially installed, (ix) grant any third party access to or use of the Software on a service bureau, timeshar-
ing, subscription service or application service provider or other similar basis, or (x) defeat or circum-
vent, attempt to defeat or circumvent, or authorize or assist any third party in defeating or circumvent-
ing controls on the installation or use of copies of the Software.
c. Updates. TKA SCIENTIFIC INSTRUMENTS, from time to time during the Term, may provide up-
dates to the Software and/or the malware signatures included therein (each, an "Update"). Updates will
be deemed Software for all purposes under this Agreement. You acknowledge that you must routinely
download and permit installation of Updates in order to obtain maximum benefit from the Software. In
no event shall TKA SCIENTIFIC INSTRUMENTS be obligated to provide you with Updates following the
expiration or termination of this Agreement, and in its sole discretion may terminate updates for any
version of the Software other than the most current version, or for use of Software in connection with
versions of any third party operating systems, email programs, browser programs and other software
other than the most current version.
d. Customer Comments. TKA SCIENTIFIC INSTRUMENTS welcomes your comments concerning
the Software, including notice that you have experienced a Software failure, error or other malfunction
and suggestions for additional or different features and functions. Please email your comments and
suggestions to [email protected]b.ru. TKA SCIENTIFIC INSTRUMENTS shall have no obligation to respond
or act on to any such comments or suggestions, but you grant TKA SCIENTIFIC INSTRUMENTS a per-
petual, irrevocable, fully paid-up, royalty-free, worldwide right and license under your intellectual
property rights (if any) to implement your comments and suggestions in the Software and other prod-
ucts and services offered by TKA SCIENTIFIC INSTRUMENTS, its affiliates, and their respective licen-
sors, licensees, successors and assigns.
Generated by Foxit PDF Creator © Foxit Software
http://www.foxitsoftware.com For evaluation only.

2. Term, Termination.
a. Term. The initial term of this Agreement will commence on the date you download or otherwise
acquire the Software, and will continue (i) for the term you specified in your order for the Software, or
(ii) if you purchased the Software on CD or other physical medium, on the Software package, or (iii) if
you received the Software in combination with other hardware or software, the term specified by your
supplier. If renewal terms are available, TKA SCIENTIFIC INSTRUMENTS will provide notice offering
you the opportunity to purchase renewals at the then-current renewal price. The initial term and any
renewals purchased are referred to in this Agreement as the "Term".
b. Termination. TKA SCIENTIFIC INSTRUMENTS, in addition to such other rights may be available
at law or equity, shall be entitled to terminate the license granted by this Agreement without liability (i)
for convenience on five (5) days' prior notice, provided that TKA SCIENTIFIC INSTRUMENTS shall, in
its sole discretion, either refund to you the license fees you paid in respect of the then-current initial
term or renewal term., prorated over the applicable term, or grant license for substantially similar
product for the remainder of the Term, or (ii) for cause at any time without notice if you commit a mate-
rial breach of this Agreement.
c. Effect of Termination. On the expiration or termination of this Agreement, you will cease using
the Software, TKA SCIENTIFIC INSTRUMENTS may cease making Updates available to you, and the
Software may cease functioning. Sections 1.b, 1.d, 3, 4, 5 and 6 will survive the expiration or termina-
tion of this Agreement.
3. Ownership. TKA SCIENTIFIC INSTRUMENTS reserves all rights in the Software not expressly
granted by this Agreement. All copyrights, trademarks and other conceivable intellectual property
rights in and to the Software (including, but not limited to, malware signatures and other data files, im-
ages appearing in the Software and screen displays as well as any and all documentation relating to
the Software) are owned by TKA SCIENTIFIC INSTRUMENTS or its licensors, and are protected by
United States and foreign copyright laws, international treaties and other applicable laws. Any copy of
the Software you are allowed to make pursuant to this Agreement must contain the entire copyright
and other notices included with the original copy of the Software.
4. Warranties.
a. General. TKA SCIENTIFIC INSTRUMENTS warrants, that on delivery of the Software and for a pe-
riod of thirty (30) days thereafter, that the medium (if any) on which the Software is delivered will be
free of material defects, and that the Software will perform substantially in accordance with the appli-
cable specifications. The foregoing warranty applies only to the Software as originally delivered, and
does not apply to Updates. Your sole and exclusive remedy for breach of this Warranty is replacement
of the defective media or Software or, at TKA SCIENTIFIC INSTRUMENTS' option, return of the Soft-
ware for a full refund. In order to exercise your rights under this Section 4, you must deinstall and de-
stroy all copies of the Software you may have made (including all archival copies), and (i) if you pur-
chased the Software by download, follow the instructions on the confirmation email you received in
connection with the purchase, or (ii) for all other purchases, return the Software in its original package,
along with your receipt, to the point of purchase.
b. Beta Versions. Notwithstanding the provisions of Section 4.a, THE PROVISIONS OF THIS SEC-
TION 4.b APPLY IN PLACE OF SECTION 4.a IF (AND ONLY IF) THE SOFTWARE IS A "BETA TEST"
VERSION. In view of the evaluation nature of Software, Software is provided on an "as is", "as avail-
able" basis, without warranty and without support or other services by TKA SCIENTIFIC INSTRU-
MENTS. YOU AGREE TO USE THE Spectrofotometer Beta SOFTWARE STRICTLY FOR THE PUR-
POSES OF EVALUATION AND TESTING, AND THAT YOUR RELIANCE ON THE AVAILABILITY OR AC-
CURACY OF THE Spectrofotometer Beta SOFTWARE SHALL BE ENTIRELY AT YOUR OWN RISK.
c. Disclaimer. EXCEPT AS EXPRESSLY PROVIDED BY SECTION 4.a OF THIS AGREEMENT, TKA
SCIENTIFIC INSTRUMENTS DISCLAIMS ALL OTHER WARRANTIES WITH RESPECT TO THE SOFT-
WARE, MEDIA AND ANY OTHER SUBJECT MATTER OF THIS AGREEMENT, WHETHER EXPRESS OR
IMPLIED, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTY OF MERCHANTABILITY, THE
IMPLIED WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE AND IMPLIED WARRANTY OF
NONINFRINGEMENT. TKA SCIENTIFIC INSTRUMENTS DOES NOT WARRANT THAT THE OPERATION
OF THE SOFTWARE WILL BE UNINTERRUPTED OR ERROR FREE, OR THAT THE SOFTWARE WILL
PROVIDE 100% PROTECTION. Some jurisdictions do not allow limitations on an implied warranty, so
the above limitations may not apply to you. You may have other rights that vary from jurisdiction to
jurisdiction.
d. Hazardous Environments. You acknowledge that the Software is not designed or licensed for
use in hazardous environments, including without limitation operation of nuclear facilities, aircraft
navigation systems, aircraft communication systems, air traffic control, life support or weapons sys-
tems and any other environment in which bodily injury or death could result from failure of or inability
to use the Software. Without limiting the provisions of Sections 4.b and 4.c of this Agreement, TKA
SCIENTIFIC INSTRUMENTS and its licensors hereby disclaim any express or implied warranties of fit-
ness for such uses.
5. Limitation of Liability. TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, IN NO EVENT
SHALL TKA SCIENTIFIC INSTRUMENTS OR ITS SUPPLIERS OR RESELLERS BE LIABLE TO YOU OR
ANY THIRD PARTY FOR ANY INDIRECT, CONSEQUENTIAL, INCIDENTAL, PUNITIVE OR SPECIAL
DAMAGES WHATSOEVER, WITHOUT REGARD TO CAUSE OR THEORY OF LIABILITY (INCLUDING,
Generated by Foxit PDF Creator © Foxit Software
http://www.foxitsoftware.com For evaluation only.
Table of contents