Tokai Hit Thermo Plate TPi Series User manual

TPi series Instruction Manual
MA-TPi-EN-01
Microscope-Stage Automatic Thermocontrol System
Index
1 For safe and comfortable use p.2
2 Components and names p.5
3 Display and parameter p.10
4 Preparation for use p.13
5 How to use p.17
6 PID calibration p.19
7 Shift calibration p.20
8 Key lock setting p.21
9 TEM –Installation p.22
10 TEM –How to operate p.23
11 Trouble shooting p.25
12 Specification p.26

1. For safe and comfortable use
Be sure to read this instruction manual carefully before using, and please use the
device correctly. Also, please keep this instruction manual.
2
These warnings enable safe and correct usage, and prevent dangerous situations. Please
follow them at all times.
We will not be responsible in the unlikely event of death, serious injury or property
damage by using deviating from the original purpose and not following the displayed
symbols.
About symbols
If device is used in manner other than specified in instruction manual, safety cannot be
assured and may result in damage to device which will not be covered under warranty.
Please follow the instructions.
For your safety
Injury or damage to equipment is possible
Must not be done
Must be done
Electric shock is possible
Symbol meanings
Please treat this precision instrument with care. Impacting it is forbidden.
Do not put any heavy things on the box.
Please store and transport under specified storage conditions.
Do not turn the box upside down.
Do not let the unit moisten.
When shipping

3
Flammable or corrosive gas or oil mist that can damage electrical insulation
Intense vibration or impact
High voltage power lines, inductive interference
Place where dew drops are present or with direct sunlight
Dusty place
Outside the designated use environment and storage environment
Do not use this equipment in the below conditions.
Regarding on the environment for usage and storage
Use only the specified electricity and cables. If other cables are used this might
result in electric shocks. Using a non-specified voltage might result in equipment
damage.
Please install this device by the method recommended by our company.
Do not place this device near a heat source or a combustible material around this
device. Abnormal temperature control or fire may be caused.
Please do not give a shock to the glass part. Glass may break.
In case of glass breakage, do not touch the glass. It might result in injury and
electric shock.
Please do not touch the glass as much as possible. It may cause a malfunction or
abnormal heating.
Do not place heavy objects on this device or add strong force. It may cause a
malfunction.
The heating range of the plate is only the heater part. Please keep in mind that
other part part will not be warmed.
When rotating the revolver, turn the aiming handle and retract the objective lens.
The objective lens may contact with this device, possibly damaging the lens or
device. It may cause abnormal heating.
Do not handle the device with wet hands. It might cause damage
to electric equipment.
Do not touch the electrical contacts such as the connector for the
plate cable.
Points to pay attention when installing to a microscope

4
Keep the room temperature at 25℃±2℃.
Do not let air flow directly over the device.
Please observe the below in order to avoid focus drift during
microscope observations.
In an emergency pull out the plug from the outlet. Do not place
any things near the outlet.
Please be careful not to catch the power cord, various cables, etc. on the human
body such as limbs.
To avoid the risk of electric shock, connect this device to a
commercial power supply equipped with a protective installation.
Do not modify the equipment or disassemble each part. It may
cause a malfunction.
Please use the uninterruptible power supply if you want to operate the equipment
without interruption even during blackout.
Blackout
Please be sure to use a PC that conforms to the EMC standard of information and
communication equipment.
Connected PC
Turn off power supply while performing maintenance.
Do not use volatile materials such as benzene or thinner for cleaning. Use of such
materials will discolor and/or damage key device surfaces.
Clean device with soft cloth using small amount of diluted neutral detergent. Do
not use organic solvents. Also, please do not scrape strongly during cleaning.
Avoid cleaning with hard or sharp objects.
Do not wash the devise with water.
Please keep at our storage condition specified. It may cause a malfunction.
Maintenance

5
2. Components and names
Components Qty
Image Description
Controller 1 Controller that regulates the Heating
Plate
Heating Plate
* Orange line: Strengthen glass
No line: standard glass
1 Designed to install on microscope
stage to maintain specimen temp.
External Temp. Sensor
(Length:300mm)1
Temp. sensor to be used for checking
the temperature of the heating plate
(p.14) or Shift Calibration (p.20)
Extension Wire for Sensor
(Length:1800mm)1 For extending the length of the
External Temp. Sensor
AC Adapter
(Length:1800mm)1
Supplies power to the controller *
Must use in original pairs
Power Cable
(Length:1800mm)1
TEM(Temp. logging software)
Installation CD 1
Installs Tokai Hit original data logging
software. The data will be saved as
CSV format
USB Cable
(Length:2000mm)1 Cable connected between the
controller and the PC when using the
above software
Mounting Hook 1 Use it for standing or hanging the
controller
Depending on the type of heating plate, fixing screws, hexagonal wrenches, etc. may be
included in the package.

6
Controller
Plate Connector
Name Plate
USB Connector
Stand-by Switch
DOWN Key
UP Key
Screen Shift Key
SET Key
Display
Sensor Connector Power Supply Connector
Hook Mounting Hole
Name Plate(Sample)
For details, please refer to the actual name plate attached on the product.
Serial Number
Model Name
Dimension of Controller
(Excluding protrusions):
W85×D135×H30 mm

TPi-TS2X Plate Dimension:238×122mmTPi-TCSX Plate Dimension:127.5×85.5mm
TPi-SQMX Plate Dimension:165×105mm
TPi-SQX Plate Dimension:160×110mm
TPi-110RX Plate Dimension:φ110mm
TPi-108RX Plate Dimension:φ108mm
Plate Sensor
Plate Cable
Plate Lamp
Glass Heater
Plate Lamp
Glass Heater
Metal Plates
Height Adjustable Leg
Depending on the illuminator base design and size, the glass
heater is too large and some of the glass area will be protruded
from the base. To maintain Heating Plate temp. uniformly, it is
recommended to install the metal plate at the bottom of plate.
Fixing Spring
TPi-SQFTX Plate Dimension:160×110mm
Glass Heater
Plate Lamp
Glass Heater
Plate Lamp
(Adjust the screws for height adjustment)
Glass Heater
Plate Lamp
Glass Heater Glass Heater
TPi-UNIX Plate Dimension:435×220mm, Leg Adjustable Range:75~100mm
TPi-CKX53X
Dish Fixing Blocks
Glass Heater
Plate Dimension:190×138mm
7
* Exclusively designed for Integra 3
(not compatible to use with Integra Ti)
*
Glass Heater
Inverted
Inverted
Inverted
Inverted
Stereo
Inverted
Inverted
Inverted
* Incl.:Fixing Springs
* Plate Lamp is attached to some major models.
TPi-URI2X Plate Dimension:φ110mm

TPi-TB300X
TPi-STMX Plate Dimension:272×227mm
TPi-SMZ25XTPi-SMZSLX
TPi-SZ2 Plate Dimension:278×175mm
TPi-SZX2BXTPi-SZX2AX
8
Glass Heater
Spot Facing for
Mounting Screw Plate Lamp
(Please use the
attached screws )
Glass Heater
* Incl.:Fixing Screws(M4×6)×4, Hexagonal Wrench
Glass Heater
* Incl.:Fixing Screws(M4×20)×2,(M4×15)×4,
Hexagonal Wrench, Slider
TPi-TLBaseX
Glass Heater
Plate Dimension:219.5×169.5mm
* Incl.:Fixing Screws(M4×6)×2, Hexagonal Wrench
Glass Heater
Plate Lamp
Glass Heater
Plate Lamp
TPi-STKX Plate Dimension:155×204mm
Glass Heater Glass Heater
Glass Heater
Glass Heater * Plate Lamp is attached to some major models.
TPi-SX Plate Dimension:142×115mm
Glass Heater
Mounting Plate
(Please use by peeling off double-sided tape)
Plate Dimension:222×240mmPlate Dimension:238×227mm
Plate Dimension:254×263mm Plate Dimension:251×238mm
Plate Dimension:280×266mm
Plate Lamp
Stereo
Stereo
Stereo
Stereo
Stereo
Stereo
Stereo
Stereo
Stereo
Upright
Spot Facing for
Mounting Screw
Spot Facing for
Mounting Screw
(Please use the
attached screws )
(Please use the
attached screws )
* Incl.:Fixing Screws(M4×6)×4, Hexagonal Wrench

9
TPi-110RH26 Plate Dimension:φ110mm
Metal Heater (with a hole)
TPi-108RH26 Plate Dimension:φ108mm
Metal Heater
Plate Cable
TPi-SQH26
Plate Dimension:160×110mm
TPi-SQH26FT
Plate Dimension:160×110mm
Metal Heater
Flat Springs
Metal Heater
Flat Springs
TPi-TCSH26
Plate Dimension:127.5×85.5mm
Metal Heater
Flat Springs * The hole diameter at the center of the
Metal Heater is 26mm as a standard, but
you can specify the hole diameter between
10 and 26 mm.
Ex)TPi-110RH〇〇, TPi-SQH〇〇FT
⇒ Hole diameter is 10mm
Model:TPi-110RH10, TPi-SQH10FT
* Flat Springs are included to the Metal
Heater as standard
Flat Springs
Inverted
Inverted
Inverted
Inverted
2 in 1 Heater
TPiD-I2X Plate Dimension:160×110mm
Glass Heater
Metal Heater
* 2 channel controller is included
Inverted
Some models are not listed in this manual.
For details, please contact Tokai Hit.

3. Display and parameter
10
Operation key / Display
1SET Key 1) SET enter button
2) Starts PID Calibration by pressing and holding the button
For details, refer to 6. PID Calibration (p.19).
2Screen Shift Key 1) Shift the screen
2) Starts Shift Calibration by pressing and holding the button
For details, refer to 7. PID Calibration (p.20).
3DOWN Key Decreases the Setting Value (SV) lower than its setting. By holding the
key, the value will be dropping continually.
4UP Key Increases the Setting Value (SV) higher than its setting. By holding the
key, the value will be increasing continually.
5Stand-by Switch ON/OFF switch of the controller. Temp. control starts from the moment
pressed the Stand-
by Switch. Please follow the procedures in this manual
and use it correctly.
⑦
⑥
②
⑤
④③
⑧⑨⑩⑪⑫
6Plate Temp.
(Process Value)
The process value is displayed on the temp. display screen. On setting
mode, character strings of contents are displayed.
7OUT Lamp Lights up when the controller is outputting to the plate.
8REM Lamp Lights up when using the optional TEM-i (Temp. Management Software)
and running automatic operation at the set temp. program pattern.
9AT Lamp
Lights up during PID calibration or Shift calibration. For details, refer to 6.
PID Calibration (p.19) and 7. Shift Calibration (p.20).
10
CH Lamp Displays current CH (Channel)
e.g. Lights up [1] : Heating Plate channel
No light: External Temp. Sensor channel
11
Plate Temp.
(Setting Value)
The setting value is displayed on the temp. display screen. On setting
mode, setting value for each contents are displayed.
12
ALM Lamp Lights up when an error occurs. Refer to 11. Troubleshooting (p.25).
①

11
Lights up
Plate temp. process value (PV) = Plate temp. setting value (SV).
Lights when“the plate temp. process value (PV) = plate temp. setting
value (SV)”condition lasts longer than 5 seconds.
Turns off when the plate temp. process value (PV) is outside the ±0.3℃
range of the plate temp. setting value (SV) for 5 seconds or longer.
Blinks per 1.0 sec Shift calibration or PID Calibration is running.
Blinks per 0.2 sec An error occurred.
Lights off Temperature is rising or preparing.
Plate Lamp condition
Plate Lump
* Plate Lamp is attached to some major models.

12
Parameter
Temp. of Plate 1
Temp. of External
Sensor
Stand-by(Display OFF・Control STOP)
Plate 1
Shift Value
Password
User Mode
By setting the shift value to
zero, you can restore the
factory default settings.
All states
(Press and hold)
To User
Mode
To Main Display
To Engineer
Mode
Key lock setting PID Calibration Shift Calibration
To prevent changing setting value
accidentally, the system can be set
the key lock.
It enables to see main monitor but
cannot change the values. For
details, refer to 8. Key lock setting
(p. 21).
Carry out calibration to achieve
suitable PID parameters for one’s
room temp. to regulates the Heating
Plate optimally.
For details, refer to 6. PID Calibration
(p. 19).
Enables to correct the Setting
Value (SV) of Plate temp. with
External Temp. Sensor possible.
For details, refer to 7. Shift
Calibration (p. 20)
To Main Display
At Main Display
Press
& hold
Main Display
At Main Display
Press
& hold
At Main Display
Press
& hold
Returns to Main Display
automatically after the Shift
Calibration ends.
Returns to Main Display
automatically after the PID
Calibration ends.

4. Preparation for use
13
①Place the heating plate on the microscope stage.
* For inverted microscopes, please be careful not to entangle the plate cable
with objective lens.
②Connect the heating plate to the plate connector of controller.
* When removing, please pull the cable while
pressing the convex part shown in the left figure.
③Connect the AC adapter to the power connector of controller and connect
the power cord.
④When using TEM (temperature logging software), connect the USB cable
which is included in the package to the controller.
⑤If the glass part of the heating plate gets dirty, please clean it up.
⑥Connect the External Temp. Sensor to the Extension Wire and connect the
Extension Wire to the sensor connector of controller.
Installation
Since it is a lock
type connector,
please insert it
deeply.
* Please note that the connector of the Sensor and
Extension Wire have thick pins and thin pins and
the direction of insertion is decided.
Thin pin
Thick pin
Since it is a lock
type connector,
please insert it
deeply.

14
Attach the External Temp. Sensor to the glass part of the Heating Plate.
Please be sure to follow the procedure as follows:
Adhesive
tape Attach the cable of the
External Temp. Sensor in
contact with the glass as long
as possible. However, attach it
only to the glass surface.
In order to improve the temp. acquisition property
by pasting it on the glass, it has the shape as shown
in the left.
Image of sensor tip shape
Please attach the flat surface sensor tip to center of glass plate with adhesive tape
for accurate temp. measurement. Remove the remaining air between the tape and
glass to make sure the top is in closed contact with glass.
The flat surface of the External Temp.
Sensor tip is not in close contact with
the glass.
NG NG
External Temp. Sensor is not
attached to the center of plate. NG External Temp. Sensor is attached
over the plate sensor. NG
For measuring temperature
The adhesive tape should not
touch the plate temp. sensor.
The flat surface of the External Temp.
Sensor tip is not in close contact with
the glass.

15
How to use the Mounting Hook
You can use the controller as it is on the flat surface, or use in the following way with the
attached Mounting Hook.
◎When placing the controller on the flat surface
Attach the Mounting Hook as shown above.
Please use it on a flat surface.
◎When using the controller by hanging on the wall
Please prepare a wall hook shown above and
hang the controller.
(Recommended: at intervals of 40 mm, diameter about 5 mm)
Please do not stress the Mounting Hook more than necessary, it may damage the
system. (e.g. prohibited to pull the wall-hanged controller, push the controller from
above while controller is placed on the table.)
Attach the Mounting Hook as shown above.

16
Setting Image(For measuring temp.)
AC Adapter
DC24V
Outlet
AC100-
240V
PC(USB Connection)
External Temp. Sensor
(Auto-clavable)
Heating Plate
Plate Temp.
Sensor
Extension Wire
For accurate temp. control
Before using, be sure to confirm that the serial number stated in the product name
plate on the back of the controller: ○○○○○○ 6 digits matches the cable on the
Heating Plate(near the connector).
Please do not combine controller and Heating Plate with different serial number. There
is a possibility that temp. may not be controlled correctly.
Please do not let air flow directly over Heating Plate surface. An error may occur in the
Heating Plate surface temp..
Please use a thermometer that can be traced to international or national weighing
standards and check that the temp. indicated by the attached External Temp. Sensor is
correct. Please ensure this confirmation at least once a year. Be sure to use this
thermometer which has a small and thin tip shape. A large thermometer may cause
abnormal heat heating or malfunction of the Heating Plate.
Be sure to check that the temp. at the central part of the Heating plate is near the
setting value with the attached External Temp. Sensor before placing the dish on the
Heating Plate. If there is a phenomenon such as a difference between the Heating
Plate temp and the External Temp. Sensor, refer 6. Procedure for PID Calibration (p.19)
or 7. Shift Calibration (p.20) and do temp. correction.
Be careful not to spill the specimen or the medium in the dish when placing the dish
on the Heating Plate.

17
5. How to use
①Please turn on the Stand-by Switch and run preliminary operation.
For the preliminary operation, please leave it for about 30 minutes
to 1 hour.
When you press the Stand-by Switch, heating starts. Do not touch the heater part
beyond necessity while heating. It may cause burns.
②Use the UP or DOWN Key to input the setting temp. of Heating
Plate. Refer to 3. Display and parameter (p.10) for the operation.
The Heating Plate temp. will be stable in about 10 minutes after inputting the temp.
setting value.
* Temp. stabilization time varies depending on the plate model.
③Before placing the dish on the Heating Plate, please run temp. check (p.14)
as necessary. Measure the Heating Plate temp. with the attached External
Temp. Sensor and confirm that it matches the setting temp..
As a result of checking the temp., if the Heating Plate temp. does not stabilize in your
environment, please correct the temp.. For temp. correction, refer to 6. PID
Calibration (p.19) or 7. Shift Calibration (p.20).
Main Display
Temp. of Plate 1
Temp. of External
Sensor
To User Mode

18
④Set a dish on the Heating Plate and start observation.
Examples of dish INCORRECT installation
NG NG NG NG
Dish is on the Sensor There is an object on
the Sensor
Dish protrudes beyond the
heating surface
It may cause sensor false detection, which
may affect temperature control.
There is a risk that the sample is not heated
properly.
* When installing a dish on the Harming Plate, please be careful not to spill
the specimen and the medium.
⑤When finishing, turn off the controller by pressing and hold the Stand-by
Switch. Disconnect the plug of Power Cord from power socket after
controller is turned off completely.
⑥If the glass part of the Heating Plate gets dirty, please clean it.

19
6. PID Calibration
This system adopts PID control for stable temp. control, and the controller is calibrated under
the environment of room temperature 25 ±2℃before shipping. If the temp. of the Heating
Plate is not stable in your environment, the controller can calculate the optimum parameters
according to your environment by carrying out PID calibration so that the Heating Plate temp.
becomes stable.
Temp.
Time
37.0℃37.0℃
Temp.
Time
①Display the heating plate temperature from the main screen, set the temp.
to be used with the UP/DOWN key, and press and hold the SET key.
②PID calibration starts when the AT Lamp blinks, and it ends when the AT
Lamp lights off.
Set the temp.
AT Lamp blinks
Press and hold the
SET key
AT Lamp
lights OFF

20
7. Shift Calibration
The Shift Calibration is for correcting temp. so that the surface temp. of the Heating Plate is
equal to the temp. of the External Temp. Sensor measures.
* Since the External Temp. Sensor is not included in the Economy Type, this
operation can not be carried out.
36.0℃
Temp.
Time
37.0℃After correction
Automatic correction of
Heating Plate surface temp.
Setting Temp.
37.0℃
Plate surface
Temp.
36.0℃
You can check how much correction was made from the Plate Shift Value in the User Mode.
You can also manually do the Shift Calibration by directly inputting the correction value to
this Shift Value.
Please check the stability of the Heating Plate temp. before carrying out Shift
Calibration. The operating environment is stable within the range of 25 ±2℃, and
the air blowing should not hit the surface of the Heating Plate. It may cause sensor
temp. false detection and the Shift Calibration program may not be finished.
Shift calibration takes at least 30 minutes or more.
User
Mode Plate 1 Shift Value
Set the temp. Press and hold the
Screen Shift key
AT Lamp blinks AT Lamp
lights OFF
①Display the heating plate temperature from the main screen, set the temp.
to be used with the UP/DOWN key, and press and hold the Screen Shift key.
②Shift calibration starts when the AT Lamp blinks, and it ends when the AT
Lamp lights off.
Before correction
This manual suits for next models
26
Table of contents
Popular Laboratory Equipment manuals by other brands

Fisher Scientific
Fisher Scientific Fisher Isotemp 307C quick start guide

VistaLab
VistaLab MLA pipette MLA 1061 Operator's guide

THORLABS
THORLABS CM401 user guide

NEW BRUNSWICK SCIENTIFIC
NEW BRUNSWICK SCIENTIFIC innova 4335 Guide to operations

ISO Italia Group
ISO Italia Group UVC HUNTER user manual

Micromeritics
Micromeritics GEOPYC 1365 Operator's manual

Funke Gerber
Funke Gerber LactoFlash User and reference manual

Renfert
Renfert Safety Burner eco Series manual

Metrohm
Metrohm OMNIS 6.03001.120 product manual

3M
3M Steri-Vac GS Series Site Planning & Installation Guide

Hettich
Hettich ROTANTA 460 operating instructions

Powerfiller
Powerfiller POWERFILLER 2 user manual