manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Yamazen
  6. •
  7. Laboratory Equipment
  8. •
  9. Yamazen W-Prep 2XY User manual

Yamazen W-Prep 2XY User manual

This manual suits for next models

2

Popular Laboratory Equipment manuals by other brands

Radleys Carousel 6 Plus instructions

Radleys

Radleys Carousel 6 Plus instructions

Hemotherm 400MR Operation & technical manual

Hemotherm

Hemotherm 400MR Operation & technical manual

mrclab DVO Series Operation manual

mrclab

mrclab DVO Series Operation manual

CoolLED pE-800 quick start guide

CoolLED

CoolLED pE-800 quick start guide

Luminex Guava easyCyte Maintenance Guide

Luminex

Luminex Guava easyCyte Maintenance Guide

IKA A10 basic operating instructions

IKA

IKA A10 basic operating instructions

Belden HIRSCHMANN RPI-P1-4PoE installation manual

Belden

Belden HIRSCHMANN RPI-P1-4PoE installation manual

Koehler K1223 Series Operation and instruction manual

Koehler

Koehler K1223 Series Operation and instruction manual

Globe Scientific GCM-12 quick start guide

Globe Scientific

Globe Scientific GCM-12 quick start guide

Getinge 86 SERIES Technical manual

Getinge

Getinge 86 SERIES Technical manual

CORNING Everon 6000 user manual

CORNING

CORNING Everon 6000 user manual

Biocomp GRADIENT MASTER 108 operating manual

Biocomp

Biocomp GRADIENT MASTER 108 operating manual

Cole Parmer Stuart SB3/1 instruction manual

Cole Parmer

Cole Parmer Stuart SB3/1 instruction manual

BEVS 2501/2A user manual

BEVS

BEVS 2501/2A user manual

Qiagen DML 3000 user manual

Qiagen

Qiagen DML 3000 user manual

Queensgate NANOSCAN OP400 Quick start instructions

Queensgate

Queensgate NANOSCAN OP400 Quick start instructions

Parr Instrument 4560 Operating instructions manual

Parr Instrument

Parr Instrument 4560 Operating instructions manual

Integra DOSE IT operating instructions

Integra

Integra DOSE IT operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Service manual
YAMAZEN CORPORATION
Index
■
Cleaning of UV Flow Cell (A420, A421, A400, A401)
・・・・・・・・・1
■
How to Replace UV Flow Cell (A420, A421, A400, A401)
・・・・・・・・・1
■
How to Replace the UV Lamp (A390)
・・・・・・・・・2
■
How to Replace Pump Head Module (A217, A218) of AI-700X
・・・・・・・3-4
■
How to Replace Pump Head (A219 Piston Cylinder Group)
・・・・・・・・・5
■
How to Replace RH Assembly (A219 + P030 + P044)
・・・・・・・6-7
■
How to Replace PC boards in fraction collector
・・・・・・・・・8
■
How to Replace EEPROM
・・・・・・・・・9
■
How to Replace Solenoid Valve (A763) on Fraction Collector
・・・・・・・・10
■
How to Make Tubing with Fitting at its end
・・・・・・・・11
■
How to Make Flare Type Tubing
・・・・・・・・12
W-Prep 2XY
AI-580S
AI-700X
Cleaning of UV Flow Cell (A420, A421, A400, A401)
How to Replace UV Flow Cell (A420, A421, A400, A401)
To clean the clogged flow cell, flushing
in reverse works better. To start, prepare
a container for waste, then connect the
tubing from pump/column to the outlet
tubing of the UV detector, and let the
inlet tubing of the UV detector work as
the waste tubing.
Loosen the screw then pull the knob
to remove the flow cell.
Tubing going to fraction collector
Outlet
Inlet
Tubing coming from pump/column
Knob
Screw
- 1 -
How to Replace the UV Lamp (A390)
Remove the cover (magnet)
as shown in the photo.
Install a new UV lamp.
Loosen 4 screws, and trip the fixing bracket.
Remove the connectors at both ends of
the UV lamp, and take out the UV lamp.
Connector
A New UV lamp
Fixing bracket
Connector
- 2 -
How to Replace Pump Head Module (A217, A218) of AI-700X
Step 1:
Remove front cover by loosening 4 screws
as shown by (A).
Step 2:
Loosen 2 fixing dials.
(A)
(A)
(A)
(A)
(A)
Fixing Dials
- 3 -
Step 3:
Remove the pump head module from
the pivot. When removing the pump
head, do not pull out the piston from
cylinder more than 45mm.
Step 4:
Install a new pump head module.
Step 5:
Fix the pump head module by tightening
the two dials.
MAX45mm
Pivot
Piston
- 4 -
How to Replace Pump Head (A219 Piston Cylinder Group)
on AI-580S & W-Prep2XY
Step 1:
Loosen 2 screws (A), remove the cylinder cap
(P030), then pump head (A219) will appear.
Step 2:
Do not pull out the piston more than 25mm
from the cylinder.
Step 3:
To install the pump head, move the pivot hole
inside to 2 o’clock position with baby finger,
align the hook and put it into the pivot hole.
Put cylinder cap (P030) on pump head (A219).
(A)
Cylinder cap (P030)
Pivot
Pump head (A219)
MAX25mm
Pivot hole
Hook
- 5 -
How to Replace RH Assembly (A219 + P030 + P044)
Step 1:
Loosen screw A (2 ea) and screw B (2 ea),
and remove the sheet metal.
Step 2:
Remove 4 ea of screw C and pull out RH
assembly.
Sheet metal has been removed.
(A)
(A)
(B)
(B)
(C)
(C)
RH Assembly
- 6 -
Step 3:
Install a new RH Assembly and fix it
with 4 C screws.
Step 4:
Put the sheet metal back on using A
screws and B screws.
Holes for C screws
Holes for C screws
(A)
(A)
(B)
(B)
- 7 -
How to Replace PC boards in fraction collector
Step 1: Remove 2 screws at
both sides of the board.
Step 2: Pull out board.
Step 3: Insert a new board.
Step 4: Put screws back on the
board bracket and tighten.
A880 Control board N022
A881 Interface board N020
Insert a new board on the guide slot at
both sides.
Slot
Screws
A880 control board
for Ch-1
A880 control board
for Ch-2
A881 interface board
- 8 -
How to Replace EEPROM
Necessary tools to pull out EEPROM (firmware)
Step 1:
Pull out EEPROM (firmware) using a puller.
Step 2: Put new EEPROM (firmware) back on the board. Be careful not to bend
or break any pin on the EEPROM. Put the EEPROM on the board in a right
direction. There is a half circle recession at one end of EEPROM and its socket.
The correct direction can be determined by this half circle recession.
Half circle recession
Firmware Type Puller
N260-137r DIP40P EXTRACTOR GX7
N360-137r DIP40P EXTRACTOR GX7
MX120-02 DIP20P EXTRACTOR GX3
MX520-01 DIP20P EXTRACTOR GX3
- 9 -
EXTRACTOR GX7
EXTRACTOR GX3
MX120-02
MX520-01
(DIP20P)
N260-137r
N360-137r
(DIP40P)