Yamazen W-Prep 2XY User manual

Service manual
YAMAZEN CORPORATION
Index
■
Cleaning of UV Flow Cell (A420, A421, A400, A401)
・・・・・・・・・1
■
How to Replace UV Flow Cell (A420, A421, A400, A401)
・・・・・・・・・1
■
How to Replace the UV Lamp (A390)
・・・・・・・・・2
■
How to Replace Pump Head Module (A217, A218) of AI-700X
・・・・・・・3-4
■
How to Replace Pump Head (A219 Piston Cylinder Group)
・・・・・・・・・5
■
How to Replace RH Assembly (A219 + P030 + P044)
・・・・・・・6-7
■
How to Replace PC boards in fraction collector
・・・・・・・・・8
■
How to Replace EEPROM
・・・・・・・・・9
■
How to Replace Solenoid Valve (A763) on Fraction Collector
・・・・・・・・10
■
How to Make Tubing with Fitting at its end
・・・・・・・・11
■
How to Make Flare Type Tubing
・・・・・・・・12
W-Prep 2XY
AI-580S
AI-700X
Cleaning of UV Flow Cell (A420, A421, A400, A401)
How to Replace UV Flow Cell (A420, A421, A400, A401)
To clean the clogged flow cell, flushing
in reverse works better. To start, prepare
a container for waste, then connect the
tubing from pump/column to the outlet
tubing of the UV detector, and let the
inlet tubing of the UV detector work as
the waste tubing.
Loosen the screw then pull the knob
to remove the flow cell.
Tubing going to fraction collector
Outlet
Inlet
Tubing coming from pump/column
Knob
Screw
- 1 -
How to Replace the UV Lamp (A390)
Remove the cover (magnet)
as shown in the photo.
Install a new UV lamp.
Loosen 4 screws, and trip the fixing bracket.
Remove the connectors at both ends of
the UV lamp, and take out the UV lamp.
Connector
A New UV lamp
Fixing bracket
Connector
- 2 -
How to Replace Pump Head Module (A217, A218) of AI-700X
Step 1:
Remove front cover by loosening 4 screws
as shown by (A).
Step 2:
Loosen 2 fixing dials.
(A)
(A)
(A)
(A)
(A)
Fixing Dials
- 3 -
Step 3:
Remove the pump head module from
the pivot. When removing the pump
head, do not pull out the piston from
cylinder more than 45mm.
Step 4:
Install a new pump head module.
Step 5:
Fix the pump head module by tightening
the two dials.
MAX45mm
Pivot
Piston
- 4 -
How to Replace Pump Head (A219 Piston Cylinder Group)
on AI-580S & W-Prep2XY
Step 1:
Loosen 2 screws (A), remove the cylinder cap
(P030), then pump head (A219) will appear.
Step 2:
Do not pull out the piston more than 25mm
from the cylinder.
Step 3:
To install the pump head, move the pivot hole
inside to 2 o’clock position with baby finger,
align the hook and put it into the pivot hole.
Put cylinder cap (P030) on pump head (A219).
(A)
Cylinder cap (P030)
Pivot
Pump head (A219)
MAX25mm
Pivot hole
Hook
- 5 -
How to Replace RH Assembly (A219 + P030 + P044)
Step 1:
Loosen screw A (2 ea) and screw B (2 ea),
and remove the sheet metal.
Step 2:
Remove 4 ea of screw C and pull out RH
assembly.
Sheet metal has been removed.
(A)
(A)
(B)
(B)
(C)
(C)
RH Assembly
- 6 -
Step 3:
Install a new RH Assembly and fix it
with 4 C screws.
Step 4:
Put the sheet metal back on using A
screws and B screws.
Holes for C screws
Holes for C screws
(A)
(A)
(B)
(B)
- 7 -
How to Replace PC boards in fraction collector
Step 1: Remove 2 screws at
both sides of the board.
Step 2: Pull out board.
Step 3: Insert a new board.
Step 4: Put screws back on the
board bracket and tighten.
A880 Control board N022
A881 Interface board N020
Insert a new board on the guide slot at
both sides.
Slot
Screws
A880 control board
for Ch-1
A880 control board
for Ch-2
A881 interface board
- 8 -
How to Replace EEPROM
Necessary tools to pull out EEPROM (firmware)
Step 1:
Pull out EEPROM (firmware) using a puller.
Step 2: Put new EEPROM (firmware) back on the board. Be careful not to bend
or break any pin on the EEPROM. Put the EEPROM on the board in a right
direction. There is a half circle recession at one end of EEPROM and its socket.
The correct direction can be determined by this half circle recession.
Half circle recession
Firmware Type Puller
N260-137r DIP40P EXTRACTOR GX7
N360-137r DIP40P EXTRACTOR GX7
MX120-02 DIP20P EXTRACTOR GX3
MX520-01 DIP20P EXTRACTOR GX3
- 9 -
EXTRACTOR GX7
EXTRACTOR GX3
MX120-02
MX520-01
(DIP20P)
N260-137r
N360-137r
(DIP40P)
How to Replace Solenoid Valve (A763) on Fraction Collector
Step 1: Loosen 2 screws (A) on solenoid valve bracket.
Step 2: Loosen these 2 screws (B) and remove signal cable.
Step 3: Prepare a new solenoid valve.
Insert the signal cable and tighten 2
screws (B).
Step 4: Fix the new solenoid to bracket
by tightening two screws (A).
(A)
(B)
(A)
Signal cable
- 10 -
(B)
Signal cable
How to Make Tubing with Fitting at its end
Fit Set:
B540 1.6mm O.D.
B543 2mm O.D.
B555 3mm O.D.
B556 3mm O.D.
B557 4mm O.D.
Insert tubing all the way
in to fitting
Tighten female screw and male screw together to make fitting,
- 11 -
How to Make Flare Type Tubing
Step 1: Insert tubing through flare nut.
Step 2: Heat the tubing end using a soldering bit.
Step 3: Push the tubing against flare fixture after it is heated well.
Step 5: Remove the flare fixture. Flare tubing has been made.
Step 4: Screw the flare nut on to the flare fixture and wait until tubing is cooled down.
- 12 -
HEAD OFFICE
RECRUIT SHINOSAKA BLDG. 3F, 5-14-22
NISHINAKAJIMA,
YODOGAWA-KU, OSAKA 532-0011, JAPAN
TEL: +81-6-6304-5839 FAX: +81-6-6304-3681
R & D
SANWA BLDG. 101, 4-6-10 NISHINAKAJIMA,
YODOGAWA-KU, OSAKA 532-0011, JAPAN
TEL: +81-6-6304-7284 FAX: +81-6-6304-7283
TOKYOOFFICE
AKIMOTO BLDG, 17 KANDATOMIYAMA, CHIYODA-KU,
TOKYO 101-0043, JAPAN
TEL: +81-3-5256-6481 FAX: +81-3-5256-6480
E-mail
[email protected]
WEB SITE
www.yamazenc.co.jp
YAMAZEN CORPORATION
January 2012

This manual suits for next models

2

Popular Laboratory Equipment manuals by other brands

Pike Falcon 110-60 Series Installation and user guide

Pike

Pike Falcon 110-60 Series Installation and user guide

Agilent Technologies InfinityLab LC Series user manual

Agilent Technologies

Agilent Technologies InfinityLab LC Series user manual

Takara Bio Thermal Cycler Dice Touch quick guide

Takara Bio

Takara Bio Thermal Cycler Dice Touch quick guide

Beckman Coulter Allegra 25R instruction manual

Beckman Coulter

Beckman Coulter Allegra 25R instruction manual

Thermo Scientific CL10 user manual

Thermo Scientific

Thermo Scientific CL10 user manual

Fisherbrand HandyStep S operating manual

Fisherbrand

Fisherbrand HandyStep S operating manual

DMT 320A user guide

DMT

DMT 320A user guide

Thermo Electron FL40 Service manual

Thermo Electron

Thermo Electron FL40 Service manual

Four E's Scientific MI0102001 operating instructions

Four E's Scientific

Four E's Scientific MI0102001 operating instructions

Radleys GreenHouse instructions

Radleys

Radleys GreenHouse instructions

Leica CryoJane user manual

Leica

Leica CryoJane user manual

amplex GENIE II Instructions for use

amplex

amplex GENIE II Instructions for use

Thermo Scientific AquaSensors AquaChlor user guide

Thermo Scientific

Thermo Scientific AquaSensors AquaChlor user guide

ecodyst HYDROGEN operating manual

ecodyst

ecodyst HYDROGEN operating manual

DEMA 633GAP Series installation instructions

DEMA

DEMA 633GAP Series installation instructions

Sartorius Stedim Biotech Sartocube user manual

Sartorius Stedim Biotech

Sartorius Stedim Biotech Sartocube user manual

FARMALABOR TOP BLISTER Use and maintenance manual

FARMALABOR

FARMALABOR TOP BLISTER Use and maintenance manual

Millipore Sigma Scepter 3.0 user guide

Millipore Sigma

Millipore Sigma Scepter 3.0 user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.