toscano T-500 User manual

USER’S MANUAL
MAGMETER
T-500

Page 2
ITM-MAGMETER-T500-ED2
Contents
1. GENERAL INFORMATION
2. FLOW RANGES OF INDIVIDUAL SIZES
3. TECHNICAL PARAMETERS
4. IMPORTANT
5. LOCATION
6. ELECTRIC CONNECTIONS
7. INSTRUMENT START
8. INSTRUMENT CONFIGURATION
A. MENU OPTIONS, COMMANDS
B. MULTIFUNCTIONAL OUTPUTS
B1st. PULSE SETTING
B2nd. STATUS REPORT SETTING
B3rd. EXAMPLES OF STATUS REPORTS
C. ANALOGUE OUTPUT
D. DISPLAY (UNITS, DECIMAL PLACES)
E. DAMPING OF SUDDEN CHANGES OF FLOW
F. DEAD BAND SETTING (INSTRUMENT ZERO SETTING)
G. DATALOGGER SETTING
G1st. WITHOUT TIME MASK
G2nd. WITH TIME MASK (USING REALTIME)
G3rd. EXAMPLES OF TIME MASK SETTING
G4th. DATALOGGER SUBMENU
H. FUNCTION OF THE THIRD ELECTRODE
I. PASSWORD SETTING
J. GSM MODULE ACTIVATION

Page 3
ITM-MAGMETER-T500-ED2
Induction fl ow meter T-500 is designed for measuring both clean and waste water in water management
applications.
The instrument is characterized by high accuracy and stability of fl ow rate measurement in whole range of
fl ow rates from 0.1 to 10 m/s in both directions.
Measured values are displayed on a large graphical display, which is standard component of the instrument
as well as a keyboard and three programmable isolated outputs for remote transmission of measured values.
Two of them can be used as pulse or status outputs and the third one as an active current output. The stan-
dard delivery includes also an internal datalogger and RS232 communication port.
A GSM module for remote data reading via SMS text messages can be ordered as an option.
Setting of fl ow meter parameters is performed using the keyboard and intuitive display menu or using service
software via RS232 communication port.
1. General Information
2. Flow Ranges
DN m³/h DN m³/h
Qmin Qmax Qmin Qmax
15 0,065 6,48 200 11,52 1152
20 0,115 11,52 250 18 1800
25 0,18 18 300 25,2 2520
32 0,288 28,8 350 34,56 3456
40 0,45 45 400 45 4500
50 0,72 72 500 72 7200
65 1,152 115,2 600 100,8 10080
80 1,8 180 700 138,6 13860
100 2,88 288 800 180 18000
125 4,32 432 900 228,6 22860
150 6,48 648 1000 282,6 28260

Page 4
ITM-MAGMETER-T500-ED2
3. Technical Data
Nominal diameter DN 10 to DN 1200
Nominal pressure PN 10 to PN 16
Tubing connection Flanges DIN 2633
Measuring tube internal coating Rubber, PTFE
Sensing electrodes Stainless steel AISI 316L, Hasteloy
2 x measuring / 1 x indication of empty tube /1 x ground
Flow range 0,1 to 10 m/s
Measurement error 0,5% of measured value in range from 0,5 to 10 m/s
1% of measured value in range 0,1 to 0,5 m/s
Ambient temperature -20 to 60°C
Voltage 230V AC (optionally 15-24V AC/12-34V DC)
Input power 10 VA max.
Outputs 1 x programmable multifunctional (pulse, status), isolated
(relay, load 125 V ~ /1A or 30V/2A)
1 x programmable multifunctional (pulse, status), isolated
(transistor NPN, load 30 V/50 mA max.)
1 x programmable active current (0-20 mA/ 4-20 mA,
load ≤ 500Ω)
1 x RS232
Communication RS232, keyboard, permanently backlit graphical display,
SMS via internal GSM module
Datalogger Capable of recording up to 100 samples
Protection IP67: sensor / IP65: converter (compact version)
IP67/68: sensor / IP65: converter (remote version)
!

Page 5
ITM-MAGMETER-T500-ED2
Nominal diameter DN 10 to DN 1200
Nominal pressure PN 10 to PN 16
Tubing connection Flanges DIN 2633
Measuring tube internal coating Rubber, PTFE
Sensing electrodes Stainless steel AISI 316L, Hasteloy
2 x measuring / 1 x indication of empty tube /1 x ground
Flow range 0,1 to 10 m/s
Measurement error 0,5% of measured value in range from 0,5 to 10 m/s
1% of measured value in range 0,1 to 0,5 m/s
Ambient temperature -20 to 60°C
Voltage 230V AC (optionally 15-24V AC/12-34V DC)
Input power 10 VA max.
Outputs 1 x programmable multifunctional (pulse, status), isolated
(relay, load 125 V ~ /1A or 30V/2A)
1 x programmable multifunctional (pulse, status), isolated
(transistor NPN, load 30 V/50 mA max.)
1 x programmable active current (0-20 mA/ 4-20 mA,
load ≤ 500Ω)
1 x RS232
Communication RS232, keyboard, permanently backlit graphical display,
SMS via internal GSM module
Datalogger Capable of recording up to 100 samples
Protection IP67: sensor / IP65: converter (compact version)
IP67/68: sensor / IP65: converter (remote version)
DN L (mm) Weight
(Kg)
15 150 2
20 150 2
25 150 2,5
32 150 3,5
40 150 5
50 200 6
65 200 9
80 200 11
100 250 13,5
125 250 20
150 300 25
200 350 35
250 400 41
300 500 55
350 500 65
400 600 110
500 600 120

Page 6
ITM-MAGMETER-T500-ED2
If the instrument has been disconnected from power supply for period longer than 6 months, the internal clock
battery can be empty. It is recommended to check time and date when putting the instrument in operation after
longer downtime.
The battery will be automatically charged after the instrument has been connected to power supply.
To ensure the proper function of the instrument, the measuring section of the induction fl ow meter sensor –
measuring tube – has to be completely fl ooded with the measured liquid and the measuring section cannot be
infl uenced by any disturbing elements as valves, pumps, bends or sharp deviations in the tubing section.
Therefore, observe the following instructions when installing the fl ow meter into the tubing:
If the system includes pumps, never place the fl ow meter sensor into the pump intake (A)! 1.
Place the fl ow meter sensor into the lowest point of the horizontal part of the tubing or into the ascendant 2. tubing (B); never place the sensor into the tubing in the fl ow top-down direction (C)!
Ensure that steady (straight) parts of tubing are min. 5 x DN before the instrument and 3 x DN behind the 3. instrument (D).
Tubing reductions with slope up to 8° are considered as straight (E). 4.
Prevent the instrument from being exposed to vibrations or direct sunlight (F). 5.
4. Important
5. Location
!

Page 7
ITM-MAGMETER-T500-ED2
1. . Ensure that the fl ow sensor frame is connected to ground. Use the ground terminal at the outside of the
sensor for that purpose. This will prevent from undesirable interference, especially for detached versions when
shielding of the link cable between sensor and electronic unit is connected to the frame.
Remove upper cover of the instrument to get access to terminals for power cable and for signal cables.
CAUTION!
Keep on mind that induction fl ow meter is an electric device; so next steps can be performed only by
a qualifi ed person!
6. Electric connections
!

Page 8
ITM-MAGMETER-T500-ED2
!
Blindaje del cable de conexión!
BLOQUE DE TERMINALES
FLOW C tube fl ooding indication electrode
FLOW A measuring electrode
FLOW B measuring electrode
FLOW ground terminal of the sensor
PWR D solenoid
PWR E solenoid
BIN1 programmable multifunctional output
BIN2 programmable multifunctional output
ANAL programmable analogue output
FREQ frequency output
(for calibration only)
230 V L phase conductor 230 V/50 Hz
230 V PE protective conductor
230 V N neutral conductor 230 V/50 Hz
AC/DC L supply conductor +/-
AC/DC PE protective conductor
AC/DC L supply conductor +/-
PO1 protective fuse 100 mA
design with 230 V/50 Hz supply
protective fuse 1 A
design with AC/DC supply
Compact version
Remote version
!
!
!
Control module
Sensor

Page 9
ITM-MAGMETER-T500-ED2
3. The induction fl ow meter is powered by mains voltage 230V/50Hz connected to power supply terminal block
labelled as L - PE - N, or by low voltage 12 V DC or 24 V AC connected to terminals labelled L – PE – L.
The polarity of supply conductors makes no difference in case of using the 12 V DC power supply.
Connect the power cable to terminals.
The induction fl ow meter does not have its own On-Off switch so it must be fused and switched on/off in other
device (e.g. switchboard).
The protection of electric circuits is ensured by protective fuse T100mA, or 1A for the 12 – 24 V version, located
next to power terminals.
Switch power supply on only after the sensor link cable and external device signal cables are connected!
4. Connect external devices that utilize current output or pulse outputs to terminals BIN1, BIN2 and ANAL. All
outputs are isolated.
2. Connect the fl ow sensor link cable to ABCDE . terminals. (Not applicable for compact version, where the
sensor was connected in factory) .
Terminal blocks on the sensor and on the electronic unit are labelled identically, so terminals A - A , B -B , C - C
, D - D , E – E, - - GND have to be interconnected. Connect the shielding of the link cable only on the side of
the sensor to the ground terminal on the terminal block of the sensor. Standard length of the link cable between
the sensor and the electronic unit is 6 meters. The link cable is included in delivery. La longitud estándar del cable
de interconexión entre el sensor y la unidad electrónica es de 6 metros. El cable de interconexión está incluido
en el envío.
Maximum length of the link cable is 25 meters (optional). The link cable should be always connected before
power supply is connected to the transducer.
If the power cable has been already connected to the transducer, connecting and disconnecting of the link cable
when the transducer is under voltage is prohibited.
!
!

Page 10
ITM-MAGMETER-T500-ED2
1. When power supply is switched on, the instrument display shows step by step messages POWER ON and
TEST INT CL 1 to TEST INT CL 30.
2. After completing internal tests, the instrument enters to measuring mode. Basic data are immediate fl ow rate
and total volume in positive and negative fl ow directions.
Press button to display the next page showing difference between total volumes DELTA V, total operating
hours from initial start of the instrument and value of the maximum fl ow rate 100%FLOW. Next page displays
the highest and the lowest immediate fl ow rate reached at particular measuring point.
3. Press EXE button shortly to display the last line of the datalogger. The line number is displayed in upper right
corner and its value can be max. 100. Press button or to scroll the display by one line up or down.
Datalogger line (one screen of display) shows time data (date, hour, minute) and total volumes in particular
time.
Besides these data, it can show one of following optional values (see Chapter 8.G):
Immediate fl ow rate Q
Maximum fl ow rate reached during particular period MAX
Minimum fl ow rate reached during particular period MIN
Average fl ow rate reached during particular period AVR (Average)
Press button shortly once again to return the instrument to basic mode. The instrument will return to basic
mode automatically if it has not been operated for approx. 60 seconds.
7. Instrument start
This manual describes confi guration of only those parameters that are useful to be changed when the instru-
ment is used in normal way.
We do not recommend changing of parameters other than mentioned here.
Detailed description of software is included in technical documentation for service engineers.
Press EXE button and hold it for approx. 3 seconds to switch the instrument from measuring mode to main
menu mode.
WARNING!
If the instrument is password protected, you will have to enter the password to open the main menu.
Message PASSWORD is displayed and the password consisting of up to 8 characters has to be ente-
red.
8. Instrument confi guration

Page 11
ITM-MAGMETER-T500-ED2
Main menu options are displayed in sixteen lines.
You may select the English, Italian, Spanish and/or German language for communication. English is pre-set
as standard.
Use following keys to select an option:
scroll down by one line
scroll down by one line
move to the left by one position; it returns to the beginning of the menu in the main menu
move to the right by one position; it shifts to the end of the menu in the main menu
confi rm selection
Cursor is blinking in the position of the option selected.
A. Menu options, commands
Main menu options
RETURN Return to previous LANGUAGE Language selection (English, Ita-
lian, Spanish, German)
ANALOG
OUTPUT Configuration of analogue output PULSE OUTPUTS Configuration of multifunctional
(pulse) outputs
COMPARATOR Selection of logical functions for
multifunctional (pulse) outputs FREQUENCY
OUT. Configuration of frequency output
DAMPING Instrument damping COMMUNICATION Configuration of communication
ports (RS232, GSM modem)
DATALOGGER Configuration of internal datalo-
gger SIMULATION Flow simulation
SENSOR Downloading of sensor calibration
parameters, setting of the flow
range measured
MEASUREMENT Permit to measure the flow in
negative direction, setting of the
zero flow zone, configuration of the
empty tubing detection function
TIME Configuration of internal real-time
clock PASSWORD Password setting
DISPLAY Selection of immediate flow rate
units and displayed decimal places
of counters
SERVICE Counter reset, check of the AD
transducer function, information on
software version

Page 12
ITM-MAGMETER-T500-ED2
B. Multifunctional outputs
The fl ow meter is equipped with two multifunctional outputs. Relay or optotransistor contacts are connected to
BIN1 terminals, BIN2 terminals are used only for optotransistor. Use jumper NC-OC-NO above BIN1 terminals
to set their function.
a) Output works as optotransistor.
b) Relay contacts are normally open.
c) Relay contacts are normally closed.
Outputs can function as
a) Pulse transmitter
of total volume in positive direction imp V+•
of total volume in negative direction imp V- •
of total volume in both directions imp V •
b) Status output
!

Page 13
ITM-MAGMETER-T500-ED2
B1. Pulse setting
B2. Status report setting
Volumes 0.1 l, 1 l, 10 l, 100 l, 1 m3, 10 m3, 100 m3 and widths 10 ms, 20 ms, 40 ms, 80 ms, 160 ms, 320 ms
can be assigned to transmitted pulses.
WARNING!
The fl ow meter transmits number of pulses corresponding to liquid quantity fl owed during measuring
cycle of 640 ms. Pulse can be assigned only such volume and width so that number of transmitted
pulses during time unit is real! Incorrect setting is indicated by error message. Display shows the code
E5.
Use cursor keys to select option PULSE OUTPUTS from main menu and confi rm your selection by EXE 1. key. A submenu will be displayed.
Select pulse width under option ms/imp1(2) and assign required volume to the pulse under option 2. l/imp1(2).
Select option OUT1 (output to terminals 1 - 2) or OUT2 (output to terminals 3 - 4) and confi rm your selec-3. tion by EXE key.
Select V+ from menu (output transmits pulses only for fl ow in positive direction), V- (output transmits 4. pulses only for fl ow in negative direction) or V (output transmits pulses for both directions) and confi rm by
EXE key.
Select option RETURN to fi nish setting. 5.
The instrument will ask if you want to save new parameters SAVE SETUP?. Select YES if they are OK. The
fl ow meter will save new parameters and switch to normal display mode. If NO is selected, parameters will not
be saved.
The instrument will switch to display mode and works with new parameters until it is switched off. When the
instrument is switched on again, it works with original parameters.
Use cursor keys to select option PULSE OUTPUTS from main menu and confi rm your selection by EXE 1. key. A submenu will be displayed.
Select option OUT1 (output to terminals 1 - 2) or OUT2 (output to terminals 3 - 4) and confi rm your selec-2. tion by EXE key.
Select option COMP1(2) from menu (to assign status function to output) and confi rm by EXE key.If status 3. function COMP1(2) is assigned to outputs, further setting of this function will be performed under option
COMPARATOR in main menu.
Select and confi rm this option to display submenu with direct setting of fl ow levels for changing of output 4. status (submenu options Qa, Qb), mode indicating active status of the output COMP1(2) (e.g.. Q<Qa) and
hystheresis.
HYST. Complete setting in the same way as described in paragraph B.1.4.

Page 14
ITM-MAGMETER-T500-ED2
C. Analogue output
D. Display
The fl ow meter is equipped with programmable active current output on terminals ANAL. The output can be
confi gured for 4 - 20 mA or 0 - 20 mA.
Use cursor keys to select the option ANALOG OUTPUT from main menu and confi rm your selection by 1. EXE key. A submenu will be displayed.
Select option Ia and confi rm your selection by EXE key.2.
In submenu, select the mode of current output 4 - 20 mA or 0 - 20 mA or OFF (if the output is to be disa-3. bled) and confi rm by EXE key.
Use option OUTPUT to select if the current output is to be enabled only for positive fl ow direction Q+, only 4. for negative fl ow direction Q- or for both fl ow directions Q and confi rm by EXE key.
In submenu option MAX, use command EDIT and press EXE key to enable editing of fl ow rate value co-5. rresponding to current value 20 mA. Use cursor keys to set fl ow rate value in l/s and confi rm by EXE key.
Select option RETURN to fi nish setting. 6.
The instrument will ask if you want to save new parameters SAVE SETUP?. Select YES if they are OK. The
fl ow meter will save new parameters and switch to normal display mode. If NO is selected, parameters will not
be saved.
The instrument will switch to display mode and works with new parameters until it is switched off. When the
instrument is switched on again, it works with original parameters.
Measuring units and decimal point position are set by default in factory and correspond to size of the fl ow
meter.
If this default setting does not conform to your requirements, you can change it in main menu using option
DISPLAY and in submenu using options UNIT (units of immediate fl ow rate) and DECIMAL POINT (selection of
places after the decimal point for fl owed quantity).

Page 15
ITM-MAGMETER-T500-ED2
E. Damping
Adverse effect of sudden changes of fl ow rate or laminar fl ow disturbances in measuring section of the instru-
ment (e.g. due to fi ttings, bends or pumps) sensed by the instrument and causing sudden changes of displayed
values or oscillations of current output can be eliminated by setting of instrument damping.
Use cursor keys to select the option DAMPING from main menu and confi rm your selection by EXE key. 1. A submenu will be displayed.
Select option TIME and confi rm your selection by EXE key.2.
Select time constant from submenu and confi rm it by EXE key. Generally, the longer the time constant is, 3. the more stable is the value on display and on outputs; however the slower is response to fl ow changes,
i.e. with time constant of 5 s, the instrument responds to immediate drop of fl ow and displays zero fl ow
rate only after 5 seconds.
Select option MODE and confi rm by EXE key. Select LINEAR if damping is to be linear or AVERAGE if it 4. should be calculated from several samples depending on selected time constant. Confi rm your selection
by EXE key.
In option ZONE select command EDIT and press EXE key to enable editing of bandwidth above and under 5. steady fl ow mean value in which damping should occur. Use cursor keys to set numeric value in l/s and
confi rm it by EXE key.
Select option RETURN to fi nish setting. 6.
The instrument will ask if you want to save new parameters SAVE SETUP?. Select YES if they are OK. The
fl ow meter will save new parameters and switch to normal display mode. If NO is selected, parameters will not
be saved.
The instrument will switch to display mode and works with new parameters until it is switched off. When the
instrument is switched on again, it works with original parameters.

Page 16
ITM-MAGMETER-T500-ED2
F. Dead band setting
G. Datalogger
Basic dead band of the instrument near the zero fl ow rate is 0.1 l/s. This dead band can be changed so that it
would be possible to suppress undesirable fl ow rate values occurring due to liquid surges and vibrations in tubing
at the zero fl ow rate.
WARNING!
Extension of the dead band increases measurement error at low fl ow rates. Hence parasitic movements
of liquid should be eliminated fi rst.
Use cursor keys to select the option MEASUREMENT from main menu and confi rm your selection by 1. EXE key.
Under the ZERO option, select command EDIT and press EXE key to edit the band size in l/s, in which no 2. fl ow rate is to be registered. Use cursor keys to set the numeric value in v l/s and confi rm by EXE key.
Select option RETURN to fi nish setting. 3.
The instrument will ask if you want to save new parameters SAVE SETUP?. Select YES if they are OK. The
fl ow meter will save new parameters and switch to normal display mode. If NO is selected, parameters will not
be saved.
The instrument will switch to display mode and works with new parameters until it is switched off. When the
instrument is switched on again, it works with original parameters.
Internal datalogger enables recording of measured values in chosen intervals either periodically or once at a
time and their storing in instrument memory.
Stored values can be subsequently read on the display or on a PC via RS232 communication link.
The datalogger records following parameters by default: total volume in positive direction V+, total volume in
negative direction V-, instrument status at the moment of reading Exx and real time at the moment of reading t.
Alternatively you can choose recording of one of additional values: immediate fl ow rate Q, maximum fl ow rate
during chosen interval MAX, minimum fl ow rate during chosen interval MIN or average fl ow rate during chosen
interval AVR.
All samples containing above mentioned data are stored in memory as a single record.
Capacity of the sample memory is 100 records. Samples are recorded step by step and they are assigned
addresses from 1 to 100. After recording of 100 samples, the oldest sample with address 1 is deleted, memory
is readdressed and the new sample is inserted to address 100. The memory content is not corrupted if power
failure occurs.

Page 17
ITM-MAGMETER-T500-ED2
G1. Datalogger setting without time mask
Use cursor keys to select the option DATALOGGER from main menu and confi rm your selection by EXE 1. key. A submenu will be displayed.
Select option MASK and confi rm your selection by EXE key.2.
A time mask is displayed. The time mask is disabled if it is in format xx - xx - xx xx : xx (day – month 3. – year hrs : min). If any position shows character different from x, select command EDIT and use cursor
keys to correct it to x. Use command RETURN to return to submenu.
Select option VAR and confi rm your selection by EXE key. Select in menu one of optional values to be 4. recorded (immediate fl ow rate Q, maximum fl ow rate during chosen interval MAX, minimum fl ow rate du-
ring chosen interval MIN or average fl ow rate during chosen interval AVERAGE). Selected option will be
marked with asterisk *. Use command RETURN to return to submenu.
Select option PER and confi rm your selection by EXE key. Use command EDIT to enable interval editing 5. in minutes. Enter interval length and use command RETURN to return to submenu.
Select START. Option RUN should be marked with asterisk *; this means that datalogger is operating.6.
Select option RETURN to fi nish setting. 7.
The instrument will ask if you want to save the new parameters SAVE SETUP?. Select YES if they are OK. The
fl ow meter will save the new parameters and switch to normal display mode. If NO is selected, parameters will
not be saved.
The instrument will switch to display mode and works with new parameters until it is switched off. When the
instrument is switched on again, it works with original parameters.
Sample recording can be set with fi xed interval from 15 to 107 minutes (e.g. every 15 minutes after datalogger
start) regardless to real time or using a time mask (e.g. datalogger starts each Friday at 15:00 and measurement
interval takes 2 hours).

Page 18
ITM-MAGMETER-T500-ED2
G2. Datalogger setting with time mask (using real time)
G3. Examples of time mask setting
1. Use cursor keys to select the option DATALOGGER from main menu and confi rm your selection by EXE 1. key. A submenu will be displayed.
2. Select option MASK and confi rm your selection by EXE key.2.
3. A time mask is displayed. The time mask is disabled if it is in format xx - xx - xx xx : xx (day – month 3. – year hrs : min). Select command EDIT and use cursor keys to set the required time mask format. Use
command RETURN to return to submenu.
Select option VAR and confi rm your selection by EXE key. Select in menu one of optional values to be 4. recorded (immediate fl ow rate Q, maximum fl ow rate during chosen interval MAX, minimum fl ow rate du-
ring chosen interval MIN or average fl ow rate during chosen interval AVERAGE). Selected option will be
marked with asterisk *. Use command RETURN to return to submenu.
Select option PER and confi rm your selection by EXE key. Use command EDIT to enable interval editing 5. in minutes. Enter interval length and use command RETURN to return to submenu.
Select START. Option RUN should be marked with asterisk *; this means that datalogger is operating.6.
Select option RETURN to fi nish setting. 7.
The instrument will ask if you want to save the new parameters SAVE SETUP?. Select YES if they are OK.
The fl ow meter will save the new parameters and switch to normal display mode. If datalogger is set in this way,
it will start recording in accordance with the time mask and records measured values during interval PER; time
intervals between individual records correspond to time mask setting.
xx - xx – xx xx : 00 recording starts every round hour •
15 - xx – xx 11 : 30 recording starts every 15th day in month at 11.30 •
15 - 11 – xx 11 : 30 recording starts every 15th November at 11.30 •
15 - 11 – 07 11 : 30 recording starts only on 15.11.2007 at 11.30 •
x1 - xx – xx 11 : 30 recording starts each Monday at 11.30 •
x2 - xx – xx 11 : 30 recording starts each Tuesday at 11.30 •
x3 - xx – xx 11 : 30 recording starts each Wednesday at 11.30 •
x4 - xx – xx 11 : 30 recording starts each Thursday at 11.30 •
x5 - xx – xx 11 : 30 recording starts each Friday at 11.30 •
x6 - xx – xx 11 : 30 recording starts each Saturday at 11.30 •
x0 - xx – xx 11 : 30 recording starts each Sunday at 11.30 AM•

Page 19
ITM-MAGMETER-T500-ED2
H. Function of the third electrode
G4. Datalogger submenu
The instrument is equipped with sensing electrode as a standard. It is located in upper part of the measu-
ring tube section and indicates fl ooding of the measuring tube section by liquid.
This third electrode can work in three modes.
a) If the measuring tube section is empty or incompletely fl ooded, i.e. the electrode is not in contact with
liquid, measurement is locked and instrument displays zero fl ow rate. At the same time, character Z is dis-
played in lower left corner of the display indicating that lock function has been enabled.
Select the option MEASUREMENT from main menu and confi rm your selection by EXE key. A sub-1. menu will be displayed.
Select command LOCK under the option EMPTY PIPE to activate the above described function. 2.
Note: The above described function is set by default in factory and need not to be set in most cases.
b) If the measuring tube section is empty or incompletely fl ooded, i.e. the electrode is not in contact with
liquid, measurement is not locked and this situation is only indicated. The instrument displays fl ow rate and,
at the same time, character Z is displayed in lower left corner of the display to indicate that this function
has been enabled.
Select the option MEASUREMENT from main menu and confi rm your selection by EXE key. A sub-1. menu will be displayed.
Select command ALARM under the option EMPTY PIPE to activate the above described function. 2.
c) Function of the third electrode can be completely disabled.
Select the option MEASUREMENT from main menu and confi rm your selection by EXE key. A sub-1. menu will be displayed.
Select command OFF under the option EMPTY PIPE to activate the above described function. 2.
RETURN Return to previous menu RUN* Datalogger operating
STOP Datalogger stop START Datalogger start with deleting
previous samples
PER Time interval setting MASK Time mask configuration
VAR Selecting of optional values

I. Password setting
J. GSM module activation
You can set a password to prevent unauthorized persons from access to confi guration menu.
Use cursor keys to select the option PASSWORD from main menu and confi rm your selection by EXE key. 1. A submenu will be displayed.
Select command NEW and press EXE key to enable password editing. The password can consist of up to 2. eight characters. Use cursor keys to set your password and confi rm it by EXE key.
Select option RETURN to fi nish setting.3.
The instrument will ask if you want to save the new parameters SAVE SETUP?. Select YES if they are OK.
The fl ow meter will save the new parameters and switch to normal display mode. If NO is selected, parameters
will not be saved.
The instrument will switch to display mode and works with new parameters until it is switched off. When the
instrument is switched on again, it will work with original parameters.
The instrument can be equipped with optional GSM module allowing remote data reading using SMS messages
in cellular phone networks.
Prior to activation of this function, switch off the instrument and insert a SIM card in it. Use a cellular phone to
cancel function of PIN request for this SIM card. Then install an external antenna, which is included in delivery.
After power supply is switched on, the instrument displays message POWER ON and then GSM INIT... 1. indicating that the instrument recognized the internal GSM module. After that, starting procedure is perfor-
med with standard messages from INTERNAL TEST INT CL 1 to INTERNAL TEST INT CL 30.
Use cursor keys to select the option COMMUNICATION from main menu and confi rm your selection by 2. EXE key. A submenu will be displayed.
Select command GSM and press EXE key to display next submenu. Select option INIT and confi rm it by 3. EXE key. The message GSM INIT…OK will be shown on the display of the instrument.
Select option RETURN to fi nish setting. 4.
The instrument is ready for sending SMS messages. Send SMS message in format F3(space)DATA from your
cellular phone to the number of the SIM card installed in the fl ow meter to enable reading of data. The instrument
will respond by sending SMS message to the number of the cellular phone it received the enabling SMS from.
The fi rst indication in SMS message gives total volume in positive fl ow direction, the second indication gives
total volume in negative fl ow direction, the next value is the information on the immediate fl ow rate at the moment
of reading followed by the information on total operating time and possible error.