
2. Operating Modes
The function switch is used to select the required mode. The four
operating modes available are…
1. Set Upper Limit : Switch to Set Upper
Limit when setting the Millivolt value at which
the High LED and relay output are activated.
The relay output is not operational in this mode.
Factory-set to activate when the reading is
above the Upper Limit. Can be user set to be a
activate when the reading is below the Upper
Limit. See section 3.5.
2. Set Lower Limit : Switch to Set Lower
Limit when setting the Millivolt value at which
the Low LED and relay output are activated.
The relay output is not operational in this mode.
Factory-set to activate when the reading is
below the Lower Limit. Can be user set to be a
activate when the reading is above the Lower
Limit. See section 3.5.
3. Calibrate : Switch to Calibrate mode to
place the proCHEM-R into standby (eg. when
calibrating). This mode disables the two output
relays to avoid chemicals being added when
placing the sensor into calibration buffers.
4. Automatic Control : Switch to
Automatic Control to enable the Alarm LED’s
and relay outputs. These will be activated when
the Millivolt reading is outside the Upper and
Lower limit values. See section 3.5.