PCB Piezotronics ICP 352M221 User manual

Model 352M221
ICP® Accel., 10 mV/g, 500 g, 2nd order filter, 1 Hz to 10 kHz response, calibration from 5 Hz to
upper 5% frequency, overmolded 40 foot hydrotested cable to 500 psi
Installation and Operating Manual
For assistance with the operation of this product,
contact the PCB Piezotronics, Inc.
Toll-free: 716-684-0001
24-hour SensorLine: 716-684-0001
Fax: 716-684-0987
E-mail: [email protected]
Web: www.pcb.com
Manual 21354 Rev E
ECN 50523
Repair and Maintenance
PCB guarantees Total Customer Satisfaction through its
“Lifetime Warranty Plus” on all Platinum Stock Products
sold by PCB and through its limited warranties on all other
PCB Stock, Standard and Special products. Due to the
sophisticated nature of our sensors and associated
instrumentation, field servicing and repair is not
recommended and, if attempted, will void the factory
warranty.
Beyond routine calibration and battery replacements
where applicable, our products require no user
maintenance. Clean electrical connectors, housings, and
mounting surfaces with solutions and techniques that will
not harm the material of construction. Observe caution
when using liquids near devices that are not hermetically
sealed. Such devices should only be wiped with a
dampened cloth—never saturated or submerged.
In the event that equipment becomes damaged or ceases
to operate, our Application Engineers are here to support
your troubleshooting efforts 24 hours a day, 7 days a
week. Call or email with model and serial number as well
as a brief description of the problem.
Calibration
Routine calibration of sensors and associated
instrumentation is necessary to maintain measurement
accuracy. We recommend calibrating on an annual basis,
after exposure to any extreme environmental influence,
or prior to any critical test.
PCB Piezotronics is an ISO-9001 certified company whose
calibration services are accredited by A2LA to ISO/IEC
17025, with full traceability to SI through N.I.S.T. In
addition to our standard calibration services, we also offer
specialized tests, including: sensitivity at elevated or
cryogenic temperatures, phase response, extended high
or low frequency response, extended range, leak testing,
hydrostatic pressure testing, and others. For more
information, contact your local PCB Piezotronics
distributor, sales representative, or factory customer
service representative.
Returning Equipment
If factory repair is required, our representatives will
provide you with a Return Material Authorization (RMA)
number, which we use to reference any information you
have already provided and expedite the repair process.
This number should be clearly marked on the outside of
all returned package(s) and on any packing list(s)
accompanying the shipment.
Contact Information
PCB Piezotronics, Inc.
3425 Walden Ave.
Depew, NY14043 USA
Toll-free: (800) 828-8840
24-hour SensorLine: (716) 684-0001
General inquiries: in[email protected]
Repair inquiries: rma@pcb.com
For a complete list of distributors, global offices and sales
representatives, visit our website, www.pcb.com.
Safety Considerations
This product is intended for use by qualified personnel
who recognize shock hazards and are familiar with the
precautions required to avoid injury. While our equipment
is designed with user safety in mind, the protection
provided by the equipment may be impaired if equipment
is used in a manner not specified by this manual.
Discontinue use and contact our 24-Hour Sensorline if:
Assistance is needed to safely operate equipment
Damage is visible or suspected
Equipment fails or malfunctions
For complete equipment ratings, refer to the enclosed
specification sheet for your product.
Definition of Terms and Symbols
The following symbols may be used in this manual:
DANGER
Indicates an immediate hazardous
situation, which, if not avoided, may
result in death or serious injury.
Manual 21354 Rev E
ECN 50523
CAUTION
Refers to hazards that could damage
the instrument.
NOTE
Indicates tips, recommendations and
important information. The notes
simplify processes and contain
additional information on particular
operating steps.
The following symbols may be found on the equipment
described in this manual:
This symbol on the unit indicates that
high voltage may be present. Use
standard safety precautions to avoid
personal contact with this voltage.
This symbol on the unit indicates that
the user should refer to the operating
instructions located in the manual.
This symbol indicates safety, earth
ground.
Manual 21354 Rev E
ECN 50523
PCB工业监视和测量设备 - 中国RoHS2公布表
PCB Industrial Monitoring and Measuring Equipment - China RoHS 2 Disclosure Table
部件名称
有害物质
铅(Pb)
汞
(Hg)
镉
(Cd)
六价铬(Cr(VI))
多溴联苯 (PBB)
多溴二苯醚(PBDE)
住房
O
O
O
O
O
O
PCB板
X
O
O
O
O
O
电气连接器
O
O
O
O
O
O
压电晶体
X
O
O
O
O
O
环氧
O
O
O
O
O
O
铁氟龙
O
O
O
O
O
O
电子
O
O
O
O
O
O
厚膜基板
O
O
X
O
O
O
电线
O
O
O
O
O
O
电缆
X
O
O
O
O
O
塑料
O
O
O
O
O
O
焊接
X
O
O
O
O
O
铜合金/黄铜
X
O
O
O
O
O
本表格依据 SJ/T 11364 的规定编制。
O:表示该有害物质在该部件所有均质材料中的含量均在 GB/T 26572 规定的限量要求以下。
X:表示该有害物质至少在该部件的某一均质材料中的含量超出 GB/T 26572 规定的限量要求。
铅是欧洲RoHS指令2011/65/ EU附件三和附件四目前由于允许的豁免。
CHINA RoHS COMPLIANCE
Manual 21354 Rev E
ECN 50523
Component Name
Hazardous Substances
Lead (Pb)
Mercury (Hg)
Cadmium (Cd)
Chromium VI
Compounds
(Cr(VI))
Polybrominated
Biphenyls (PBB)
Polybrominated
Diphenyl Ethers
(PBDE)
Housing
O
O
O
O
O
O
PCB Board
X
O
O
O
O
O
Electrical Connectors
O
O
O
O
O
O
Piezoelectric Crystals
X
O
O
O
O
O
Epoxy
O
O
O
O
O
O
Teflon
O
O
O
O
O
O
Electronics
O
O
O
O
O
O
Thick Film Substrate
O
O
X
O
O
O
Wires
O
O
O
O
O
O
Cables
X
O
O
O
O
O
Plastic
O
O
O
O
O
O
Solder
X
O
O
O
O
O
Copper Alloy/Brass
X
O
O
O
O
O
This table is prepared in accordance with the provisions of SJ/T 11364.
O: Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit
requirement of GB/T 26572.
X: Indicates that said hazardous substance contained in at least one of the homogeneous materials for this part is above the limit
requirement of GB/T 26572.
Lead is present due to allowed exemption in Annex III or Annex IV of the European RoHS Directive 2011/65/EU.
Model Number
352M221 ICP® ACCELEROMETER Revision: A
ECN #: 44460
Performance ENGLISH SI
Sensitivity(± 15 %) 10 mV/g 1.02 mV/(m/s²)
Measurement Range ± 500 g pk ± 4905 m/s² pk
Frequency Range(± 10 %) .5 to 10,000 Hz .5 to 10,000 Hz
Frequency Range(± 5 %) 1 to 6000 Hz 1 to 6000 Hz
Electrical Filter Corner Frequency >15 kHz >15 kHz [1]
Electrical Filter Roll-off 12 dB/octave 12 dB/octave [1]
Resonant Frequency ≥50 kHz ≥50 kHz
Broadband Resolution(1 to 10,000 Hz) .001 g rms .01 m/s² rms [1]
Non-Linearity ≤1 % ≤1 % [2]
Transverse Sensitivity ≤5 % ≤5 % [3]
Environmental
Overload Limit(Shock) ± 7000 g pk ± 68,670 m/s² pk
Temperature Range(Operating) -65 to +250 °F -54 to +121 °C
Temperature Response See Graph See Graph [1]
Electrical
Excitation Voltage 18 to 30 VDC 18 to 30 VDC
Constant Current Excitation 2 to 20 mA 2 to 20 mA
Output Impedance ≤100 Ohm ≤100 Ohm
Output Bias Voltage 7 to 12 VDC 7 to 12 VDC
Discharge Time Constant 0.4 to 2.5 sec 0.4 to 2.5 sec
Settling Time(within 10% of bias) <10 sec <10 sec
Spectral Noise(1 Hz) 230 µg/√Hz 2256 (µm/sec2)/√Hz [1]
Spectral Noise(10 Hz) 38 µg/√Hz 373 (µm/sec2)/√Hz [1]
Spectral Noise(100 Hz) 19 µg/√Hz 186 (µm/sec2)/√Hz [1]
Spectral Noise(1 kHz) 10 µg/√Hz 98 (µm/sec2)/√Hz [1]
Physical
Size (Height x Hex) 0.71 in x 0.44 in 18 mm x 11.1 mm
Weight(without cable) 0.22 oz 6.3 gm [1]
Sensing Element Ceramic Ceramic
Sensing Geometry Shear Shear
Housing Material Titanium Titanium
Sealing Hermetic Hermetic
Electrical Connector Sealed Integral Cable Sealed Integral Cable [4]
Electrical Connection Position Side Side
Cable Termination Pigtail Ends Pigtail Ends
Cable Length 40 ft 12.2 m
Cable Type SB-42275 Twisted Shielded Pair SB-42275 Twisted Shielded Pair
Mounting Thread 10-32 Female 10-32 Female
[5]
All specifications are at room temperature unless otherwise specified.
In the interest of constant product improvement, we reserve the right to change specifications without notice.
ICP®is a registered trademark of PCB Group, Inc.
NOTES:
[1]Typical.
[2]Zero-based, least-squares, straight line method.
[3]Transverse sensitivity is typically ≤3%.
[4]Sensor and Cable hydrotested without connector to 500 psi [3.3 MPa].
[5]See PCB Declaration of Conformance PS023 for details.
SUPPLIED ACCESSORIES:
Model 081B05 Mounting Stud (10-32 to 10-32) (1)
Model ACS-130 5Hz to highest spec 10k Hz (1)
Model ATS-6 Hydro test (sensor/cable pressurized in water environment) (1)
3425 Walden Avenue, Depew, NY 14043
Phone: 716-684-0001
Fax: 716-684-0987
E-Mail: [email protected]
OPTIONAL VERSIONS
Optional versions have identical specifications and accessories as listed for the standard
model except where noted below. More than one option may be used.
Entered: JM Engineer: DK Sales: RWM Approved: BAM Spec Number:
Date: 7/31/2015 Date: 7/31/2015 Date: 7/31/2015 Date: 7/31/2015 61398
MOUNTNG THRE AD SEE DR AWI NG
5-40 A
M3 X 0.50
B
10-32 C
M5 X 0.80 D
1/4-28
E
M6 X 1.00
F
BOLT THREAD SEE DRAWING
10-32 C
M5 X 0.80 D
1/4-28
E
M6 X 1.00
F
M8 X 1.25
F
1
1
2
2
3
3
4
4
A A
B B
CODE
IDENT. NO.
52681
DWG. NO.
SCALE: SHEET
DRAWN CHECKED ENGINEER
TITLE
UNLESS OTHERWISE SPECIFIED TOLERANCES ARE:
DIMENSIONS IN MILLIMETERS
[ IN BRACKETS ]
XX ±0.13
ANGLES 2 DEGREES
DIMENSIONS IN INCHES
DECIMALS XX ±.01
ANGLES 2 DEGREES
FILLETS AND RADII
.003 - .005
DECIMALS X ± 0.3
FILLETS AND RADII
0.07 - 0.13
INSTALLATION DRAWING
081-XXXX-90
1 OF 1NONE
FOR STANDARD
081 SERIES MOUNTING
XXX ±.005
NJF 05/03/23 JDM 05/03/23 MJN 05/03/23
3425 WALDEN AVE. DEPEW, NY 14043
(716) 684-0001 E-MAIL: sales@pcb.com
1"≤LENGTH < 1' = +1"/ - 0
1'≤LENGTH < 5' = +2"/ - 0
5'≤LENGTH < 100' = +6"/ - 0
100'≤LENGTH = +1'/ - 0
CABLE TOLERANCES IN ENGLISH CABLE TOLERANCES IN METRIC
2.54cm ≤LENGTH < 30.5cm = +2.54cm/ - 0
30.5cm ≤LENGTH < 1.5m = +5.1cm/ - 0
1.5m ≤LENGTH < 30.5m = +15.2cm/ - 0
30.5m ≤LENGTH = +30.5cm/ - 0
"A"
5-40
MOUNTING INSTRUCTIONS
(METRIC DIMENSIONS IN BRACKETS)
MOUNTING HOLE PREPARATION:
.101[2.57]
X .20[5.1]MIN.
5-40 UNC-2B
X .15[3.8]MIN.
4.) RECOMMENDED MOUNTING TORQUE,
4-5 INCH POUNDS
[45-55 NEWTON CENTIMETERS].
"B"
M3 X 0.50
MOUNTING INSTRUCTIONS
(ENGLISH DIMENSIONS IN BRACKETS)
MOUNTING HOLE PREPARATION:
2.5[.099]
X 4.6 [.18]MIN.
M3 X 0.50-6H
X 3.3[.13]MIN.
4.) RECOMMENDED MOUNTING TORQUE,
45-55 NEWTON CENTIMETERS
[4-5 INCH POUNDS].
"C"
10-32
MOUNTING INSTRUCTIONS
(METRIC DIMENSIONS IN BRACKETS)
4.) RECOMMENDED MOUNTING TORQUE,
10-20 INCH POUNDS
[113-225 NEWTON CENTIMETERS].
MOUNTING HOLE PREPARATION:
.159[4.04]
X .23[5.8]MIN.
10-32 UNF-2B
X .15[3.8]MIN.
"D"
M5 X 0.80
MOUNTING INSTRUCTIONS
(ENGLISH DIMENSIONS IN BRACKETS)
4.) RECOMMENDED MOUNTING TORQUE,
113-225 NEWTON CENTIMETERS
[10-20 INCH POUNDS].
MOUNTING HOLE PREPARATION:
4.22[.166]
X 7.62 [.300]MIN.
M5 X 0.8-6H
X 5.08[.200]MIN.
"E"
1/4-28
MOUNTING INSTRUCTIONS
(METRIC DIMENSIONS IN BRACKETS)
4.) RECOMMENDED MOUNTING TORQUE,
2-5 FOOT POUNDS
[3-7 NEWTON METERS].
MOUNTING HOLE PREPARATION:
.218[5.54]
X .300[7.62]MIN.
1/4-28 UNF-2B
X .200[5.08]MIN.
"F"
M6 X 0.75, M6 X 1.00, M8 X 1.25
MOUNTING INSTRUCTIONS
(ENGLISH DIMENSIONS IN BRACKETS)
4.) RECOMMENDED MOUNTING TORQUE,
3-7 NEWTON METERS [2-5 FT POUNDS].
M8 X 1.25
MOUNTING HOLE PREPARATION:
6.75[.266]
X 8.64 [.340]MIN.
M8 X 1.25-6H
X 5.00[.197]MIN.
M6 X 0.75
MOUNTING HOLE PREPARATION:
5.31[.209]
X 7.62 [.300]MIN.
M6 X 0.75-6H
X 5.08[.200]MIN.
M6 X 1.0
MOUNTING HOLE PREPARATION:
5.05[.199]
X 8.10 [.320]MIN.
M6X 1.0-6H
X 6.35[.250]MIN.
"B"
"C"
"G"
MOUNTING INSTRUCTIONS
FOR SPECIAL THREAD LENGTHS
(METRIC DIMENSIONS IN BRACKETS)
THREAD DEPTH : B= X + 1 THREAD PITCH
DRILL DEPTH : C= B + 3 THREAD PITCH
SEE A-F FOR APPROPRIATE DRILL AND TAP
THREAD PITCH= 1/TPI [P]
3.) FOR BEST RESULTS, PLACE A THIN LAYER OF SILICONE GREASE (OR EQUIVALENT) ON INTERFACE PRIOR
TO MOUNTING.
MOUNTING SURFACE SHOULD BE FLAT TO WITHIN .001(0.03) TIR OVER DIM 'A' WITH A OR BETTER FINISH FOR BEST RESULTS.
DRILL PERPENDICULAR TO MOUNTING SURFACE TO WITHIN 1.
SENSOR
SENSOR
"THRU-BOLT"
"THRU-BOLT" STUD MOUNT
INTEGRAL STUD MOUNT
SENSOR
THREAD
MOUNTING
THREAD
SENSOR
THREAD
MOUNTING
THREAD
STANDARD STUD MOUNT
"X"
"X"
SENSOR THREAD
"X"
1111
1
1
1
1
2222
2
2
2
MOUNTING HOLE PREPARATION:
DRILL DIA.
X "C"MIN.
TAP
X "B"MIN.
1
"A"
"A"
"A"
"P"
1 THREAD PITCH
SHOWN
"X"
SENSOR
INTEGRAL
MOUNTING
STUD
63[1.61]
REVISIONS
REV DESCRIPTION DIN
R CHANGED "B" HOLE TOLERANCE 53828
2
1
MOUNTING THREAD SEE DRAWING
5-40 A
M3 X 0.50 B
10-32 C
M5 X 0.80 D
1/4-28 E
M6 X 1.00 F
081-XXXX-90
PCB Piezotronics Inc. claims proprietary rights in
the information disclosed hereon. Neither it nor any
reproduction thereof will be disclosed to others
without the written consent of PCB Piezotronics Inc.
1
1
2
2
A A
B B
CODE
IDENT. NO.
52681
DWG. NO.
SCALE: SHEET
DRAWN CHECKED ENGINEER
TITLE
UNLESS OTHERWISE SPECIFIED TOLERANCES ARE:
DIMENSIONS IN MILLIMETERS
[ IN BRACKETS ]
ANGLES 2 DEGREES
3425 WALDEN AVE. DEPEW, NY 14043
(716) 684-0001 E-MAIL: sales@pcb.com
DIMENSIONS IN INCHES
ANGLES 2 DEGREES
FILLETS AND RADII
.003 - .005
FILLETS AND RADII
0.07 - 0.13
OUTLINE DRAWING
34927
1 OF 13X
ACCELEROMETER
WITH WATER PROOF CABLE
DECIMALS XX ±.03
XXX ±.010
DECIMALS X ± 0.8
XX ± 0.25
BB 6/17/15 ECB 6/17/15 DRK 6/17/15
34927
PCB Piezotronics Inc. claims proprietary rights in
the information disclosed hereon. Neither it nor any
reproduction thereof will be disclosed to others
without the written consent of PCB Piezotronics Inc.
REVISIONS
REV DESCRIPTION DIN
AUPDATED TITLE 44262
.72 [18.2]
.142 [3.6]
.43 [10.9] TYP
INTERGRAL SEALED, TWISTED,
SHIELDED 2 CONDUCTOR
CABLE Ø.12 [2.9] TERMINATING
IN PIGTAILS.
10-32 UNF - 2B .17
7/16 HEX

Other PCB Piezotronics Measuring Instrument manuals

PCB Piezotronics 353B03 User manual

PCB Piezotronics

PCB Piezotronics 353B03 User manual

PCB Piezotronics Larson Davis HVM200 User manual

PCB Piezotronics

PCB Piezotronics Larson Davis HVM200 User manual

PCB Piezotronics IMI SENSORS 683A001000 User manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS 683A001000 User manual

PCB Piezotronics ICP 337A31 User manual

PCB Piezotronics

PCB Piezotronics ICP 337A31 User manual

PCB Piezotronics J320C04 User manual

PCB Piezotronics

PCB Piezotronics J320C04 User manual

PCB Piezotronics 357B81 User manual

PCB Piezotronics

PCB Piezotronics 357B81 User manual

PCB Piezotronics IMI SENSORS M603M170 User manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS M603M170 User manual

PCB Piezotronics IMI SENSORS 66333PNZ1 User manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS 66333PNZ1 User manual

PCB Piezotronics 1515-110-03A User manual

PCB Piezotronics

PCB Piezotronics 1515-110-03A User manual

PCB Piezotronics ICP 352B70 User manual

PCB Piezotronics

PCB Piezotronics ICP 352B70 User manual

PCB Piezotronics 356M57 User manual

PCB Piezotronics

PCB Piezotronics 356M57 User manual

PCB Piezotronics IMI SENSORS TO603C11 User manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS TO603C11 User manual

PCB Piezotronics IMI SENSORS 624B11 User manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS 624B11 User manual

PCB Piezotronics IMI Sensors M602D01 User manual

PCB Piezotronics

PCB Piezotronics IMI Sensors M602D01 User manual

PCB Piezotronics 356A67 User manual

PCB Piezotronics

PCB Piezotronics 356A67 User manual

PCB Piezotronics 3741B122G User manual

PCB Piezotronics

PCB Piezotronics 3741B122G User manual

PCB Piezotronics ICP 353B31 User manual

PCB Piezotronics

PCB Piezotronics ICP 353B31 User manual

PCB Piezotronics ICP 352B10 User manual

PCB Piezotronics

PCB Piezotronics ICP 352B10 User manual

PCB Piezotronics 625B11/-0008 User manual

PCB Piezotronics

PCB Piezotronics 625B11/-0008 User manual

PCB Piezotronics P356A71 User manual

PCB Piezotronics

PCB Piezotronics P356A71 User manual

PCB Piezotronics 3711E1150G User manual

PCB Piezotronics

PCB Piezotronics 3711E1150G User manual

PCB Piezotronics IMI SENSORS 606B01/ICS-14 User manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS 606B01/ICS-14 User manual

PCB Piezotronics 357B61 Installation instructions

PCB Piezotronics

PCB Piezotronics 357B61 Installation instructions

PCB Piezotronics Triaxial ICP 356M57 User manual

PCB Piezotronics

PCB Piezotronics Triaxial ICP 356M57 User manual

Popular Measuring Instrument manuals by other brands

POOL LAB V20 user manual

POOL LAB

POOL LAB V20 user manual

Linortek ITRIXX-­WFMN Quick Instruction

Linortek

Linortek ITRIXX-­WFMN Quick Instruction

LineEye LE-650H2 Installation and operating manual

LineEye

LineEye LE-650H2 Installation and operating manual

Metrix AG 900 Operation manual

Metrix

Metrix AG 900 Operation manual

Base Line 12-0235 Standard instruction manual

Base Line

Base Line 12-0235 Standard instruction manual

PerkinElmer Frontier IR user guide

PerkinElmer

PerkinElmer Frontier IR user guide

Ono Sokki DS-2000 Series Operation manual

Ono Sokki

Ono Sokki DS-2000 Series Operation manual

Next Century RR301 quick start guide

Next Century

Next Century RR301 quick start guide

EMU Check Short manual

EMU

EMU Check Short manual

Centech 93297 Assembly & operating instructions

Centech

Centech 93297 Assembly & operating instructions

ISAFE MOBILE SENSEVEN IS-VS1A.1 Quick start manual

ISAFE MOBILE

ISAFE MOBILE SENSEVEN IS-VS1A.1 Quick start manual

AlertLabs SENTREE Safety and installation

AlertLabs

AlertLabs SENTREE Safety and installation

Master cool 89660 operating instructions

Master cool

Master cool 89660 operating instructions

FRO­NIUS Smart Meter 240V-3 UL operating instructions

FRO­NIUS

FRO­NIUS Smart Meter 240V-3 UL operating instructions

Smart Witness CP4 Configuration guide

Smart Witness

Smart Witness CP4 Configuration guide

globalwater FP101 user manual

globalwater

globalwater FP101 user manual

Applied Instruments Sat Buddy 2 User instructions

Applied Instruments

Applied Instruments Sat Buddy 2 User instructions

Pro's Kit PD-161-C manual

Pro's Kit

Pro's Kit PD-161-C manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.