1. PRECAUTION
The Marine Salt testa is for the sole purpose of measuring the
Temperature, Salinity and Specific Gravity of natural seawater and /
or synthetic seawater. This meter has been factory calibrated and
performance tested to ensure accurate readings.
Please read the entire operation manual before using this Marine
Salt testa. Failure to read this manual prior usage may result in
equipment damage and will void the warranty.
1.1 Notes on safety
Please keep this product away from children. Ingestion of small
parts may lead to choking and serious injury. If the sensor
protection cap or battery is ingest; please consult a physician
immediately.
Be careful not to scratch the surface of the platinum-plated
sensors.
Do not use chemicals such as acetone or benzene for cleaning as
it may cause deterioration of the plastic housing.
Never attempt to disassemble or modify this product, this will void
warranty.
1.2 Notes during measurement
Always rinse the Probe Sensor before and after each test.
Make sure sensor do not contact any of the walls of the
aquarium or sump while in measurement mode. Keep sensors
at least 1cm away from all sides and bottom of the measuring
container.
1.3 Notes on storage
Avoid keeping Marine Salt testa
in high humidity environments,
direct exposure to sunlight, high temperatures, dust and
moisture to prevent damage.
Please keep sensor cap in-
place to protect the sensors when not
in use.
Always remove the batteries if tester is not use over a long
period of more than one month.
!
!!
!
!
!!
!
!
!!
!

2. Product Diagram and Functions
2.1 Buttons Definition:
The ON-OFF button is use to turn the power on and off. Once
turned on, Temperature measurement is display in the LCD
Display window.
The MODE button is to select display of temperature, Salinity
or Specific Gravity measurement. Press the button once will
switch it to Salinity measurement (ppt), pressing again will
switch to Specific Gravity measurement and a third time back
to Temperature measurement in a cyclical mode. This button
is also use to change the display units of temperature (°
°°
°C/°
°°
°F,
see section 6)
The HOLD button is use to “hold” the measurement value
during a test. Pressing it again will resume continuous
measurement. This button is also use for performing
Calibration in Salinity mode. (See section 5)
MODE
(SEL)
HOLD
(CAL)
ON
OFF
HOLD
(CAL)
Button
MODE
(SEL)
Button
ON/OFF
Button
LCD
DISPALY
BATTERY CAP
Turn anti-
clockwise to
open and clockwise to
close
PROBE SENSOR
Rinse sensor with water
before and after each test.
After a series of tests, wash
sensor with mild
dishwashing liquid and rinse
before storage.

2.2 Display Definition:
°C / °F
Indicates the Temperature of the sample.
ppt
Indicates the Salinity of the sample.
HOLD
Is displayed in the LCD display window when the
measured value is put on hold or freezed.
ERROR
Appears when the measurement is outside the normal
testing range.
Temp. range (0°C∼
∼∼
∼100°C)
Salinity Range (0∼5
∼5∼5
∼50 ppt)
Specific Gravity (1.000∼
∼∼
∼1.037)
Is displayed when battery level is low and replacement
of battery is needed.
3. Product Specification
Salinity
Specific Gravity
Temperature
Measurement
Range
0-50ppt 1.000-1.037sg
0-70°C
32-158°F
Resolution
1ppt 0.001sg 0.1°C / 0.2°F
Accuracy
±2ppt ±0.002sg ±1°C / ±2°F
Operating
Temperature
0-70°C
32-158°F
0-40°C
32-104°F
0-70°C
32-158°F
Auto-off
Around 1 minute
Battery type
2 x 3Volt Button Cell (CR2032H or equivalent)
Battery life
Approx. 150 hours continuous use
Size (LxWxH)
210 × Dia.32 mm
Weight
Approximately 61grams (including batteries)

4. Measurement of Salinity, Temperature and Specific
Gravity
4.1 Principle of salinity measurement
The salinity
measurement principle of this product is base on the
electric conductivity measurement method. Salt exists as sodium ion
and chloride ion within aqueous solution. As the quantity of sodium ion
and chloride ion increases, the solution conductivity also increases
proportionately to the increased salt concentration. Using this principle,
salinity is determined by calculating the conductivity of a given sample.
4.2 Testing Temperature, Salinity and Specific Gravity
1. Press ON/OFF button once to switch on, temperature reading will
appear on the display.
2. Dip the PROBE SENSOR into the liquid and shake vigorously to
remove bubbles on probe.
*Note: Presence of bubbles will affect reading accuracy
3. Wait for the temperature reading untill it stops changing and for a
stablized temperature to establish. Now you can press the MODE
button to switch to Salinity measurement.
4. Wait 10 seconds and confirm the Salinity (ppt) measurement.
5. Press MODE button to display Specific Gravity measurement.
Wait 10 seconds for the Specific Gravity before reading results.
6. Press the HOLD/CAL button at anytime to “Hold” or “Freeze” the
displayed reading. Press it again to “Release” for another test.
7. After each test, rinse the sensors with distilled water between each
test and wipe it dry with a clean cloth or tissue paper.
8. After a series of tests, clean the PROBE SENSOR in mild detergent
then rinse with water throughtly. Wipe the PROBE SENSOR dry
before storing away.
9. Press the ON/OFF button to switch off. The battery saving function
will automatically shut off the meter if no activity for about 1 minute.

5. Calibration
1. Perform calibration regularly or before a series of tests.
2. Use only 30ppt standard solution for this procedure.
3. Make calibration only in liquid of between 15 to 30°C
4. Switch on unit and press MODE button once to switch to salinity
mode (ppt).
5. Dip the sensor into 30ppt solution centrally; making sure the sensor
does not contact the bottom or sides of the bottle.
6. While the sensor is in the solution, press and hold-down
HOLD(CAL) button for about 15 seconds or until it display shows
C A L
flashing.
7. In a while, display will show 30ppt and return to normal measuring
mode.
8. Calibration is completed
9. If display show 31, 32 or any other values, repeat the calibration.
Note: Never perform calibration if unit is not dip into 30ppt solution, as this
will cause inaccuracy to the unit. If reading is wrong, repeat this calibration
procedure.
6. Temperature display Selection (°C / °F)
1. Switch on the unit. Ensure display is in the temperature mode.
2. Press and hold-down MODE button until display shows °F.
3. Release the button, °F reading will be displayed everytime.
4. To switch to °C temperature display, repeat step 1 to 3.

7. Replacement of the Battery
When the sign appears on the LCD display, the batteries will need
replacement. The unit will automatically shut-off when battery life is out.
How to replace the batteries
1. Uscrew the battery cap and open it. Pull out the battery holder, Take
a note at the directions of each battery placed and remove all of
them.
2. Replace with two (2) new batteries (CR2032H or equivalent)
ensuring each battery’s polarity are placed in direction (positive ‘+’
sign face up as shown below).
3. Switch on power to confirm whether the batteries were place
correctly, replace the cover.
8.Troubleshooting
Status Cause Remedy
Low battery /
Worn out batteries Replace all batteriesLCD display
becomes dim
Or stop working Wrong battery
placement
Check all batteries are placed
correctly accordingly
Contamination of
sensors
Wash sensors with neutral
detergent and rinse with water. Dry
with cloth or tissue paper
Unstable measured
value
Salinity or temperature
is not homogeneous
Please refer to 4.2 (salinity
measurement) and
1.3 (cautions on measuring)
+
+

(Singapore) Pte Ltd
email: sales@transinstruments com
Website: www transinstrumnets com
ISO9001 Certified Firm
Rev-3 Nov2010
PRINTED IN SINGAPORE
Quality checked in Singapore
WARRANTY:
Trans Instruments (Singapore) Pte. Ltd., warranties this product for a
period of 12 months and 3 months on probes and electrode from date of
purchase; against all defects in material and workmanship.
This warranty does not apply to the abuse or misuse of the instrument. If
repairs or adjustments are required, please return the defective product
freight prepaid. Instrument within warranty will be repaired at no charge.
Make sure that the product is properly packed and insured against possible
damage or loss in shipment.
Purchase invoice MUST be accompanied in returned product or else
warranty is considered void.
Please obtain authorization from Trans Instruments (Singapore) Pte Ltd.
Directly or through your local sales representatives prior to returning the
product.
Trans Instruments staff can be contacted at the following email address or
through our web-page contacts:
Table of contents
Other Trans instruments Measuring Instrument manuals

Trans instruments
Trans instruments RW0025 User manual

Trans instruments
Trans instruments KoiMedicPro User manual

Trans instruments
Trans instruments BP3001 User manual

Trans instruments
Trans instruments ELECTROLYZER User manual

Trans instruments
Trans instruments AquaCOMBO HM3070 User manual

Trans instruments
Trans instruments RBX0010 User manual

Trans instruments
Trans instruments WalkLAB HP 9010 User manual

Trans instruments
Trans instruments RSA0100 User manual