manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. TransLite
  6. •
  7. Laboratory Equipment
  8. •
  9. TransLite Veinlite R User manual

TransLite Veinlite R User manual

OR
Selecting Colors:
Depress the End Button [B] to cycle
through these modes:
Orange Optimal wavelength for highlighting the
contrast between the venous blood
vessels and surrounding tissue.
Green Assist in finding arteries and veins.
White Exam light.
Off
Cleaning Directions
Cleaning the device will not be necessary if properly
used with the disposable plastic cover.
In the event that the device needs to be cleaned,
use germicidal wipes sparingly.
Do not submerge the unit in liquid.
Protective Cap:
The purpose of the Protective Cap [E] is to protect
the End Button [B], Side Button [C], and LED
Window [A] while the device is not in use.
Keep a disposable plastic cover on device to
ensure a proper fit.
Attaching an optional Lanyard:
Attach lanyard to the Cap [E]
then snap it over the
End Button [B].
The Protective Cap and lanyard should be cleaned
regularly with alcohol or germicidal solution.
Installing/Replacing Batteries:
The Veinlite R is powered by a single
AAA battery.
1Press firmly on the Battery Cover [D]
near the seam and slide it off.
2Remove the used battery.
3Install a new AAA battery,
oriented as indicated by the
markings inside the battery
compartment.
4Slide the Battery Cover into place.
Warranty
The Veinlite R has a one year parts & labor warranty.
Troubleshooting
Before returning the device to TransLite
for repair, please call or e-mail to obtain an
authorization number and shipping details.
No refunds will be given for
devices shipped without
authorization.
LED Failure
If one of the LED fails to turn on please replace the
batteries. If the problem persists contact TransLite
for assistance.
No Light From Any LEDs
Replace the battery, then try turning the unit on
again. If the device still will not turn on, please
contact TransLite for assistance.
Specifications
Number of LEDs: 1 Orange
1 Green
1 White
Weight with battery: 29g
Dimensions: 96.1×19.0×16.5mm
Battery: AAA 1.5V
Explanation of Symbols
Model Number
Serial Number
Manufacturer
EC Representative (EU Only)
Do not dispose of as unsorted municipal waste
M. Devices Group
Health Education Centre
The Church, Portland Street
Southport PR8 1HU
United Kingdom
+44 1704 544 944
Operating Directions
Disposable Covers:
The Veinlite R should always be used with a
single-use disposable plastic cover to protect it from
contamination and to prevent the spread of
infection.
1Place the Veinlite R inside
the disposable plastic cover.
Check the cover to make sure it
is intact and the flap is on the
side away from the LED
Window [A].
2These covers are not sterile.
They should be wiped
with a germicidal wipe
prior to use.
3Discard the single-use
disposable plastic cover after each
subject.
Disposable plastic covers are not included and
must be ordered fromTransLite or any of its
authorized dealers.
Device:
1 Install a new Cover [F].
2 Slide Large Tail Holder [G] or Small Tail
Holder [H] over the LED Window [A] to
allow illumination of tail veins.
3 Place Veinlite R against the
tail with the Tail Holder in
contact with the tail.
4Depress the End Button [B] to
turn on the LED.
5Apply gentle pressure to the Veinlite R to
maintain optimal contact.
6Device will turn off after cycling
through the colors or you can
depress and hold the End Button [B] for one
second to turn device off.
7When you turn on a device with a low battery the
device will shut itself off.
8Device will shut off after approximately 4 minutes.
Brightness Control:
There are 3 brightness levels. Depress the
Side Button [C] to change the brightness. Each
depression of the button will cycle through low,
medium, and high brightness levels. The lowest
brightness level is the default setting.
Caution
Veinlite R is designed for external use only.
To avoid contamination of the device, Veinlite R must
be used with the disposable plastic cover.
To avoid cross contamination, always change the
disposable plastic cover between subjects and wipe
the new cover with a recommended germicidal wipe.
Always turn the LEDs on after the device is in
position. Do not look directly at the LEDs when they
are illuminated or shine the light from them directly
into anyone’s eyes.
Do not immerse the device in liquid.
Do not autoclave the device.
Introduction
The Veinlite R is a hand-held battery powered
transillumination device for visualizing veins and
arteries in the tails of rats and mice.
Applications
To assist in venous and arterial access in rats and
mice tails.

READ INSTRUCTIONS BEFORE USE
TransLite LLC
345 Commerce Green Blvd
Sugar Land TX 77478 USA
+1 281 240 3111
www.veinlite.com
[email protected]
LED Window [A]
End Button [B]
Cap [E]
Cover [F]
Large Tail Holder [G]
Small Tail Holder [H]
Side Button [C]
Battery Cover [D]
Veinlite R
Selecting Colors:
Depress the End Button [B] to cycle
through these modes:
Orange Optimal wavelength for highlighting the
contrast between the venous blood
vessels and surrounding tissue.
Green Assist in finding arteries and veins.
White Exam light.
Off
Cleaning Directions
Cleaning the device will not be necessary if properly
used with the disposable plastic cover.
In the event that the device needs to be cleaned,
use germicidal wipes sparingly.
Do not submerge the unit in liquid.
Protective Cap:
The purpose of the Protective Cap [E] is to protect
the End Button [B], Side Button [C], and LED
Window [A] while the device is not in use.
Keep a disposable plastic cover on device to
ensure a proper fit.
Attaching an optional Lanyard:
Attach lanyard to the Cap [E]
then snap it over the
End Button [B].
The Protective Cap and lanyard should be cleaned
regularly with alcohol or germicidal solution.
Installing/Replacing Batteries:
The Veinlite R is powered by a single
AAA battery.
1Press firmly on the Battery Cover [D]
near the seam and slide it off.
2Remove the used battery.
3Install a new AAA battery,
oriented as indicated by the
markings inside the battery
compartment.
4Slide the Battery Cover into place.
Warranty
The Veinlite R has a one year parts & labor warranty.
Troubleshooting
Before returning the device to TransLite
for repair, please call or e-mail to obtain an
authorization number and shipping details.
No refunds will be given for
devices shipped without
authorization.
LED Failure
If one of the LED fails to turn on please replace the
batteries. If the problem persists contact TransLite
for assistance.
No Light From Any LEDs
Replace the battery, then try turning the unit on
again. If the device still will not turn on, please
contact TransLite for assistance.
Specifications
Number of LEDs: 1 Orange
1 Green
1 White
Weight with battery: 29g
Dimensions: 96.1×19.0×16.5mm
Battery: AAA 1.5V
Explanation of Symbols
Model Number
Serial Number
Manufacturer
EC Representative (EU Only)
Do not dispose of as unsorted municipal waste
M. Devices Group
Health Education Centre
The Church, Portland Street
Southport PR8 1HU
United Kingdom
+44 1704 544 944
Operating Directions
Disposable Covers:
The Veinlite R should always be used with a
single-use disposable plastic cover to protect it from
contamination and to prevent the spread of
infection.
1Place the Veinlite R inside
the disposable plastic cover.
Check the cover to make sure it
is intact and the flap is on the
side away from the LED
Window [A].
2These covers are not sterile.
They should be wiped
with a germicidal wipe
prior to use.
3Discard the single-use
disposable plastic cover after each
subject.
Disposable plastic covers are not included and
must be ordered fromTransLite or any of its
authorized dealers.
Device:
1 Install a new Cover [F].
2 Slide Large Tail Holder [G] or Small Tail
Holder [H] over the LED Window [A] to
allow illumination of tail veins.
3 Place Veinlite R against the
tail with the Tail Holder in
contact with the tail.
4Depress the End Button [B] to
turn on the LED.
5Apply gentle pressure to the Veinlite R to
maintain optimal contact.
6Device will turn off after cycling
through the colors or you can
depress and hold the End Button [B] for one
second to turn device off.
7When you turn on a device with a low battery the
device will shut itself off.
8Device will shut off after approximately 4 minutes.
Brightness Control:
There are 3 brightness levels. Depress the
Side Button [C] to change the brightness. Each
depression of the button will cycle through low,
medium, and high brightness levels. The lowest
brightness level is the default setting.
Caution
Veinlite R is designed for external use only.
To avoid contamination of the device, Veinlite R must
be used with the disposable plastic cover.
To avoid cross contamination, always change the
disposable plastic cover between subjects and wipe
the new cover with a recommended germicidal wipe.
Always turn the LEDs on after the device is in
position. Do not look directly at the LEDs when they
are illuminated or shine the light from them directly
into anyone’s eyes.
Do not immerse the device in liquid.
Do not autoclave the device.
Introduction
The Veinlite R is a hand-held battery powered
transillumination device for visualizing veins and
arteries in the tails of rats and mice.
Applications
To assist in venous and arterial access in rats and
mice tails.

VLNR DFU Rev B 0519
Made in USA
Veinlite R®is a Registered Trademark.
* Patent Pending

Popular Laboratory Equipment manuals by other brands

Newport 500B Series user manual

Newport

Newport 500B Series user manual

Digitus DN-651141 Quick installation guide

Digitus

Digitus DN-651141 Quick installation guide

Tenda PoE15F user manual

Tenda

Tenda PoE15F user manual

BioLAB BCLT-202 Operation manual

BioLAB

BioLAB BCLT-202 Operation manual

Stanford Research Systems SIM985 Operation and service manual

Stanford Research Systems

Stanford Research Systems SIM985 Operation and service manual

Medisafe SI PCF user manual

Medisafe

Medisafe SI PCF user manual

Dreve Innovation MediTech PCU 90 Instructions for use

Dreve

Dreve Innovation MediTech PCU 90 Instructions for use

Velp Scientifica BOD Sensor F102B0133 instruction manual

Velp Scientifica

Velp Scientifica BOD Sensor F102B0133 instruction manual

Hanil Ultra 5.0 quick guide

Hanil

Hanil Ultra 5.0 quick guide

Biohit PROLINE PLUS instruction manual

Biohit

Biohit PROLINE PLUS instruction manual

GYROZEN MicroCentrifuge mini user manual

GYROZEN

GYROZEN MicroCentrifuge mini user manual

Techne PrimeQ Operator's manual

Techne

Techne PrimeQ Operator's manual

Heraeus HS 9 Service manual

Heraeus

Heraeus HS 9 Service manual

BD Phoenix M50 System Quick reference guide

BD

BD Phoenix M50 System Quick reference guide

MikroTik GPEN11 quick start guide

MikroTik

MikroTik GPEN11 quick start guide

Drawell LBI-175A-N user manual

Drawell

Drawell LBI-175A-N user manual

vacuubrand PC 510 NT Instructions for use

vacuubrand

vacuubrand PC 510 NT Instructions for use

Beckman Coulter JA-25.50 Instructions for use

Beckman Coulter

Beckman Coulter JA-25.50 Instructions for use

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.