TScale RWP User manual

RWP
Weighing Indicator
V1.16
REV:Y1,May 2016


RWP Weighing Indicator Technical Manual(v1.16)
CONTENTS
1. SPECIFICATIONS….………………………………………………….……......4
2. INTRODUCTION………………………………………………………………...6
3. INSTALATION …………………………………………………………………..7
Unpacking…………………………………………………………………..7
Installation…………………………………………………………………..7
Load cell connections..………………………………………………...8
Connect Adaptor and Charging..………………………………………...8
4. DESCRIPTION……………...........……………………………………………...9
Display……………………………………………………………………....9
Key board……………………………………………………………..…….10
5. OPERATION.…………………….………………………………….………….11
5.1. Power ON/OFF……………………………………………..………..11
5.2. Zero…………………….................................................................11
5.3.Tare................................................................11
5.4. Sample Weighing………………………………………….………….11
5.5. Check Weighing…………………………………………….…………11
5.6. Enter to Menu………………………………….…………………….12
5.7. Set Limits…………………..............................................................12
5.8. Set check weighing mode..............................................................12
5.9. Accumulation……………………………………….………………….13
5.10. Accumulation automatically……………………..………………...13
5.11. Animal Weighing…………………………………………………...14
5.12. Backlight Setting…………………………………………………15
5.13. Auto Power Off Setting……………………………………………15
5.14. Peak Hold……………………………………………………………15
5.15. Subtraction scale…………………………………………………15
6. PARAMETER…………………………………………….…………………….16
Keys operation into menu…………………………………………….. .16
Parameter Block………………………………………………………....17
7. CALIBRATION…………………………………………………………………21
8. RS232 OUT PUT………………………………………………………………24
9. DRAWING…………………………………………………………..……………27
10. ERROR CODES…………………………………………………………..……29

RWP Weighing Indicator Technical Manual(v1.16)
- 4 -
1. SPECIFICATIONS
Overall View
Dimensions

RWP Weighing Indicator Technical Manual(v1.16)
- 5 -
Model
RWP
Resolution
1/30,000
Indicator housing
ABS Plastic
Stabilisation Time
1 Seconds typical
Operating Temperature
0°C ~ +40°C / 32°F - 104°F
Power supply (external)
AC Adaptor (12V/500mA) /
Ni-MH battery (1.2V/2000mAh x 6)
Calibration
Automatic External
Display
6 digits 22mm LCD display, attached backlight
Interface
RS-232 Output Optional
Zero range
0mV~5mV
Signal input range
0~15mV
ADC
Sigma delta
ADC update
Max 60 times /second
Load cell drive voltage
Max 5V/150mA
Internal counts
600,000

RWP Weighing Indicator Technical Manual(v1.16)
- 6 -
2. INTRODUCTION
The RWP series weighing indicator that amplifies signals from a load cell,
converts it to digital data and displays it as a mass value.
It is suitable for general weighing or more specialized applications such as
check weighing, animal weighing and accumulation applications.
It can connect the indicator to a printer or a PC.
Large LCD with white LED back light display

RWP Weighing Indicator Technical Manual(v1.16)
- 7 -
3. INSTALLATION
Unpacking
When you receive the scale, inspect it to make sure that it is not damaged and
that all are parts are included:
Remove the Indicator from the carton.
Remove the protective covering. Store the packaging and to use if you
need to transport the scale later.
Inspect the indicator for damage.
Make sure all components are included.
1. Indicator
2. Adaptor
3. Manual
4. Indicator holder (Optional)
5. Load cell Output connecter (Optional)
6. RS-232 Output Connecter (Optional)
Installation
Place the Indicator on a table or connect with proper stand.
Connect the plat form load cell cable in to the indicator load cell connecter.
Load cell connecter is locating back side of the indicator.
Connect the adaptor pin in to the indicator adaptor jack.
Adaptor jack is locating, back side of the indicator.
Adaptor connects into your AC power socket.
Pluggable equipment must be installed near an easily accessible socket
outlet with a protective ground/ earth contact.
Turn on the On/Off key. If you want to turn off, press the key again.
Display will be show the scale capacity and will be starting self-checking.
After self-checking, display will be come to normal weighing mode.
Load cell connecter
RS-232 port
Adaptor jack
Bracket

RWP Weighing Indicator Technical Manual(v1.16)
- 8 -
Warm-up time of 15 minutes stabilizes the measured values after
switching on.
Calibrate with exact calibration weights, minimum 1/3 of the scale capacity
want to use for calibration. For calibration see details in parameter.
Then you can start your operation
Load cell connections
Connect the load cell cables to the terminal as shown below.
It can connect four 350 ohm load cells.
The load cell drive voltage is 5V DC ±5% between
Excitation + and Excitation -.
Connect Adaptor and Charging
To charge the battery insert the adaptor pin to jack. Adaptor simply plug into
the mains power. The scale no needs to be turned on.
The battery should be charged 12 hours for full capacity.
The symbol status of the battery
Battery voltage has dropped
Low voltage
Fully charged
.
Do not use any other type of power adaptor than the one supplied with the
scale.
Verify that the AC power socket outlet is properly protected.
Note: Please charge the battery before using the scale for the first time.
5Pin
Connection
Pin 1
Signal +
Pin 2
Signal -
Pin 3
Shield
Pin 4
Exc -
Pin 5
Exc +

RWP Weighing Indicator Technical Manual(v1.16)
- 9 -
4. DESCRIPTION
Display
DISPLAY
FUNCTION
HI
OK
LOW
Check weighing
ZERO
Indicator for Zero display
TARE
Indicator for Tare display
GROSS
Indicator for Gross weight
NET
Indicator for Net weight
STABLE
Indicator for Display stability
AUTO
Indicator for Auto Accumulation
M+
Indicator for Accumulation
ANIMAL
Indicator for Animal Weighing Mode
HOLD
Indicator for Hold/ Lock
Indicator for Charging status of battery.
FROUNT
BACK
FROUNT
BACK

RWP Weighing Indicator Technical Manual(v1.16)
- 10 -
Key Board
KEY
FUNCTION
Turn the power On/ Off
Used to reset to Zero.
In setting mode can use to confirm entry
Used to recording tare values and change the value from gross
value to net value.
In setting mode can use to increase the value and scroll
forward in menu.
When the scale has been tared and display is in gross or net
mode.
When using the settings mode, can use to move active digits
right.
For print the results, to the PC or printer using the optional RS-
232 interface.
It also adds the value to the accumulation memory if the
accumulation function is not automatic.
When using the settings mode,can use to clear active digits
Switch to unit weight. In setting mode,
escape back to menu/ weighing mode.
When using the settings mode, can use to move active digits
left.

RWP Weighing Indicator Technical Manual(v1.16)
- 11 -
5. OPERATION
Initial Start – Up:
Warm-up time of 15 minutes stabilizes the measured values after switching on.
5.1. Power ON/OFF:
Switch on the balance by pressing key.
The display is switched on and the test is started and if want to switched
off, press again the key.
5.2. Zero
Environmental conditions can lead to the balance exactly zero in spite of
the platform not taking any strain. However, you can set the display of
your balance to zero any time by pressing key and therefore
ensure that the weighing starts at zero.
5.3. Tare
The weight of any container can be tared by pressing button so
that with subsequent weighing the net weight of the object being weighed
is always displayed.
Load weight on the platform.
Press key. Zero is displayed, and tare is subtracted.
Remove weight on the platform. Tared weight is displayed. It can set only
one tare value. It can display with a minus value.
Press G/N to change between gross weight and net weight.
To clear the tare value, remove the load and press key. Zero is
displayed, tare weight is cleared.
5.4. Sample weighing
Place goods to be weighed on the platform.
Wait few seconds for stability display.
Read the result.
Avoid overloading. When display appears “ol” reduce the load or
unload.
5.5. Check Weighing

RWP Weighing Indicator Technical Manual(v1.16)
- 12 -
It can set an upper or lower limit when weighing with the limits range.
During the limit controls dividing the unit will indicate whether a value
upper or lower limits with an alarm sound . For details see the parameter F3 oFF.
Check mode 1: No beep sound in the limits. Function turned off.
Check mode 2: When the weight is between the limits.OK will shown and
beeper will be sounded.
Check mode 3: When the weight is out of the limits, the beeper will be
sounded and OK will shown.
5.6. Enter to Menu
In the weighing mode, press key and key together.
Display will be appear
5.7. Set limits
Press key to enter the function.
Press key to select the limit.
Display will appear or
Press key key to enter, press key to move active digits.
Press key to change the value. After enter the value press key to
sure.
Press key to escape.
5.8. Set check weighing mode.
After entering the settings mode,
Press key until display will be appear
Set hi
F0 H-L
F3 oFF
Set Lo

RWP Weighing Indicator Technical Manual(v1.16)
- 13 -
Press key to enter, press key until display show
Press key to enter, press
Check mode 1
Check mode 2
Check mode 3
Select desired setting by pressing key and press key to confirm,
press key to escape.
Note: The load weight must greater than 20 scale divisions for the check
weighing operations.
To disable the check weighing function, enter zero into both limits.
5.9. Accumulation
To enable accumulation function, select parameter F5 ACC >ACC on
Place the goods on the platform to be weigh
Wait few seconds for display stable, then press key.The value will
be saved and printed (if the printer is connected).
Display will be appear
this display will appear two seconds only.
Remove the load and wait few seconds for display return to zero.
Place the second goods on the platform.
Wait few seconds for display stable. Then press key .The value will
be saved.
Followed by the total number of weight will be displayed
It can continue the process until the maximum capacity or value.
bP 1
bP 3
bP 2
ACC 1
ACC 2
bEEP

RWP Weighing Indicator Technical Manual(v1.16)
- 14 -
Note: When you change the weighing unit this saved values will be clear.
Accumulated Total
Manually, the scale can be set to accumulation by pressing key ,when an
optional printer is connected. See details in F4 Prt.
Memory Recall
When display of Zero, you can see the number of weighing and total
weight by pressing key ,display will be shown for two seconds.
Memory Clear
When display of Zero, you can see the number of weighing and total
weight by pressing key , display will be shown for two seconds.
Press key during this display. The memory data are deleted and
display will be shown
5.10. Accumulation Automatically
In this function the individual weighing values are automatically added
into the memory. No need to press any keys.
For this function, set to parameter F4 Prt and select P AutO.
After select this function, display indicator AUTO will be shown.
Place the goods on the platform to be weighed
After the stable, will be follow beep sound twice.
Unload the goods, the weighing value will be saved automatically and will
be follow beep sound once.
It can continue the process until the maximum capacity or value.
5.11. Animal Weighing
RWS can use for vibrate loads.
For this function, set to parameter P4 CHk to ModE 2
After select this function, display indicator ANIMAL will be shown.
ACC 0

RWP Weighing Indicator Technical Manual(v1.16)
- 15 -
Bring the load on to the platform.
When the load few seconds get stable, the reading will be locked for few
seconds and will be follow beep.
It can add or remove loads also update the weighing locked values.
5.12. Peak Hold
RWS can operate peak hold function, maximum reading will be hold and
will update automatically when adding the goods.
For this function, select parameter P4 CHk to ModE 4
In the normal weighing mode press and key together to turn on
Peak hold operations, display will be indicate HOLD.
If want to turn off peak function, press and key together again
5.13. Backlight Setting
In the weighing mode. Press key and key to select
parameter F3 OFF >BL. press key to select (BK AU/BK
OFF/BK ON), After select the back light option press key to confirm
and press key to escape from the settings.
5.14. Auto Power Off Setting
In the weighing mode. Press key and key to select
parameter F3 OFF >SET OF , press key to change auto power
off time: 0/3/5/15/30. (of 0: always on, of xx: auto power off after
standby xx minutes), press key to sure, press key to escape.
5.15subtraction scale
This is used for hopper scale, you need set auto zero range to 0 (see
detail in section 6 and set scale mode to mode 3
Turn on power, scale will show “err4”, then show current total weight on
platform, press key , display show 0.00, then remove goods in
hopper, display will show it’s weight in “-“ mode, press key, scale
will print out weighing ticket。

RWP Weighing Indicator Technical Manual(v1.16)
- 16 -
6. PARAMETERS
KEYS OPERATIONS INTO THE MENU
Enter the menu
In weighing mode, press key and key together.
Select the menu
Press key, it can change the menu block one by one.
Using increase the digit.
Enter the selected menu
Press key , it can confirm, which will be shown displayed.
Change the digit
Press key , it can change the active digit.
Return to weighing mode
Press key , exit from the menu.
Enter into Prog
When display shows Prog , press and keys to enter the
function

RWP Weighing Indicator Technical Manual(v1.16)
- 17 -
PARAMETER BLOCK
Menu
Sub-Menu
Description
F0 H-L
Weighing with
set limits
SET Lo
Lower limit value
SET Hi
Upper limit value.
F1 toL
to CLr
Clear the accumulation memory with out
printout
to P-C
Print the total accumulation memory and clear
the total memory
to Prt
Print the total accumulation and keep all the
memory.
F2 Unt
G
Weighing units.Tj and Hj can’t use at the same
time.
Lb
Oz
Tj
hj
F3 off
Set of
0/3/5/15/30
To set the auto off time
Bl
BK on
Display of back light on
BK au
Display of back light on
automatically
BK off
Display of back light off
beep
Bp 1
Beep sound off during the
check weighing
Bp 2
Beeper will be sounded within
the check weighing limits
Bp 3
Beeper will be sounded above
the check weighing limits
F4 prt
RS 232 mode
P prt
By pressing key , weighing value will be
added to the memory and print the print out
P cont
Send data continuous
Seire
Also send data continuous
Ask
Bi- direction , through PC
Commands R= Send, T= Tare, Z= Zero
P cnt 2
No documented
P stab
Send data of stable weighing values
P auto
Automatic accumulation.
Individual weighing values are automatically

RWP Weighing Indicator Technical Manual(v1.16)
- 18 -
added
wirel
To connect wireless
Set BAUD rate
After setting the RS 232 mode, display will be shown current
baud rate bXXX. Available baud rate: b600,b1200,
b2400,b4800 and b9600 If necessary change the
baud rate by pressing key and enter by pressing
key
Set print out format
If enter settings p prt,p auto,p cont and connected
optional printer
Pr X
M+ format- Date/Time
Only for p prt,p
auto format
Lab X
M+ format –
Gross/Accumulation
Cont 1
Only for p cont only
N.A
Cont 2
Cont 3
Set printer type
Ty-tp
Ticket printer
Ty 711
Label printer
Lp 50
Label printer
F5 ACC
ON
Accumulation function enable
OFF
Accumulation function disable
prog
pin
Enter the programming and calibration menus
by using password

RWP Weighing Indicator Technical Manual(v1.16)
- 19 -
PROGRAM PARAMETERS
Menu
Sub Menu
Description
P1 ref
A2n 0
OFF
Auto zero point settings
0. 5d
1d
2d
4d
0–auto
P1 0
Zero setting range.
When the display is turn on the scale is set to
zero
P1 2
P1 4
P1 10
P1 20
P1 50
0–
range
P 2 2
Manually zero setting range, by pressing
P 2 4
P 2 10
P 2 20
P 2 50
P 2
100
Speed
S 7.5
Set A/D speed
S 15
S 30
S 60
P 2
cal
Sigr
a
Deci
C
0
Decimal point settings
C
0. 0
C
0. 00
C
0. 000
C0. 000
0
Inc
1
Increment settings
2
5
10
20
50
Cap
00000
Enter the scale capacity

RWP Weighing Indicator Technical Manual(v1.16)
- 20 -
cal
Linear
Linear calibration
nonlin
Normal calibration
Crpcal
Creep calibration
Dua
ra
For Dual Range Settings
Deci
Decimal point settings
Inc
Div 1
Set the scale first increment
Div 2
Set the scale second increment
Cap
Cap 1
Enter the scale first capacity
Cap 2
Enter the scale second capacity
cal
Linear
Linear calibration
nonlin
Normal calibration
Crpcal
Creep calibration
duai
n
For Dual Interval Settings
Deci
Decimal point settings
Inc
Div 1
Set the scale first increment
Div 2
Set the scale second increment
Cap
Cap 1
Enter the scale first capacity
Cap 2
Enter the scale second capacity
cal
Linear
Linear calibration
nonlin
Normal calibration
Crpcal
Creep calibration
P3 pro
Count
This display will show XXXXX for indicating the internal
counts.
Reset
Factory default settings
gra
Set the local gravity
P4 chk
Mode 1
Normal weighing mode. (check weighing, accumulation)
Mode 2
Animal weighing mode. (scale can lock reading, when
little unstable)
Mode 3
This is a subtraction scale (print out “-“ weight)
Mode 4
Peak Hold mode. (Scale can hold maximum reading)