Tufner T909 User manual

T909 User’s Manual

SAFETY PRECAUTIONS .........................................................
PREPARATION & SET UP ......................................................
FEATURES .............................................................................
SPECIFICATIONS ...................................................................
POWER SUPPLY ....................................................................
DISPLAY AND KEY DESCRIPTION ........................................
OPERATING INSTRUCTIONS ................................................
CALIBRATION PROCEDURE .................................................
INDICATOR PARAMETER SETTINGS ....................................
CONNECTORS .......................................................................
TROUBLESHOOTING ............................................................
TABLE OF CONTENTS
1
1
2
3
3
4
6
7
9
14
19

Please read and follow all the instructions before using the unit. Calibration, inspections, and
maintenance shall be performed by professionals
Make sure the indicator is on a stable surface.
DO NOT disassemble or modify unit.
DO NOT exceed the rated load limit of the unit.
DO NOT use this product if there are any signs of damage.
DO NOT use for purposes other than weighing.
Keep other electronic equipment away from the unit.
Electrical Safety
Unit must be connected to a GFCI outlet.
Disconnect power before servicing unit.
All operators MUST discharge themselves before servicing unit.
Always pull the plug – not the cord – when disconnecting from the outlet.
DO NOT use if the power cord is worn or damaged
Power cord should only be replaced by certified service technicians using the original parts
SAFETY PRECAUTIONS
PREPARATION & SET UP
If the indicator is powered with an adapter, plug the adapter directly into the “DC” pin located at
the bottom of the indicator.
Always use a wall outlet to avoid interference from other wires
Turn on the indicator when there is no load
Calibrate before weighing when the scale is initially installed or moved to a new location
- 1 -

FEATURES
LED display
Weighing units
lb/kg or lb/kg/g/oz/lb-oz (only on upgraded model)
Gross/Tare/Pre-Set Tare/Zero
Multiple Hold functions
Overload / Underload indication
Configurable with multiple printers
Splash proof keyboard and display
Connects to a Remote Display/Scoreboard
Power saving mode
Ability to modify gravity based on different geographical locations
RS232 output
Optional dual load cell ports
Main Features
Technical Parameters
Accuracy class
Zero stability error
Span stability error
Sensitivity (internal)
Input voltage
Excitation circuit
AC power
Operation temperature
Operation humidity
Storage temperature
5000 e
<
<
TK0 0.1µV//K
TKspn ± 6 ppm//K
0.3 µV / d
-30 to +30mV DC
5 VDC, 4 or 6 wire connection, 8 load cells of 350ohm max
AC 100-250V (use only the included 9V adapter supplied)
14-104°F (-10 °C ~ +40 °C)
90%RH
-40-158°F (-40 °C ~ +70 °C)
≤
- 2 -

2.3 in
7.8 in
2.3 in
5.5 in
SPECIFICATIONS
POWER SUPPLY
8.2 in
8.3 in
INDICATOR MEASUREMENTS
- 3 -
AC Adapter
If the indicator is powered with an adapter, plug the adapter directly into the “DC” pin located at
the bottom of the indicator. It is recommended to plug into a wall outlet to avoid interference with
wirings.
Battery
If you have an indicator with the rechargeable battery option, please charge the internal battery
fully before first time use. A 110 to 220V AC adapter should be provided with your indicator. Please
use only the AC adapter provided. To keep the battery in the best condition, fully discharge the
battery every month by leaving the indicator on until the indicator powers off, and then recharge
fully. If the battery will not be used for long time, please remove it to avoid leakage.

LIGHT INDICATION AND KEYPAD
FRONT VIEW
1. LIGHT INDICATION
lb: ON when the weight unit is pound [lb]
kg: ON when the weight unit is kilogram [kg]
g: ON when the weight unit is gram [g]
oz: ON when the weight unit is oz [oz]
Batt: GREEN when the battery is fully charged; RED when it is low
GROSS: ON when the current weight is the GROSS weight
NET: ON when the current weight is the NET weight
ZERO: ON when the current weight is 0 lb (0 kg)
HOLD: ON when the weight value is being held
UNDER: ON when the weight is below the min preset value
ACCEPT: ON when the weight is within the acceptable range
OVER: ON when the weight is over the max preset value
2. KEYPAD
ON/OFF Power on or off (Press hard and hold for 2 seconds)
Weight value hold (5 Hold options; can be changed in parameter settings)
HOLD
Shift between weighing units
1. Resets the scale to zero when there is something on the scale (ex. Tare out the weight
of a pallet to weigh only the items on it)
2. Clears tare weight for gross weight
UNITS
TARE
Zero’s the scale
ZERO
Prints
Save and Exit
Enter
PRINT
esc
enter
Arrow keys
- 4 -

Power On
Turn on the power by pressing the power button for 2 seconds. Once it is on, the unit begins to
auto-check and count down from 0-9 before entering the weighing mode
Note: Anything on the scale before powering on will automatically be tared out.
Zero
Zero is only used when the scale is supposed to be empty
Pressing the ZERO key will reset your scale to 0
Depending on what your manual zero range parameter is set to, you can zero out any number
within the range, after that you will receive an error and will need to tare out the weight
Unit Selection
To switch between measuring units (lb/kg or lb/kg/g/oz/lb:oz ) press the UNITS key
Tare Function
Tare is used when you only wish to see the current change in weight, not the total weight
When the indicator is in gross mode (gross light is shown) pressing the TARE key will Tare the
current weight on the scale and enter the net mode (net light shown)
For example, add a container to the scale, press tare and the display
will show the tare symbol and reset back to 0. Now add the product to the container, it will only
display the weight of the product, not the container.
To exit Tare mode, press the TARE key again to enter gross mode and you will see the total
weight of the container and the product
BOTTOM VIEW
LOAD CELL: Port for connecting the scale.
RS-232 PORT: Serial interface port. (printer, second display, PC, etc.)
DC POWER: Port for DC power. (DC 9V adapter available)
LOAD CELL RS-232 PORT DC POWER
OPERATING INSTRUCTIONS
- 5 -
Note: If you remove the container, the scale will display weight less that of the container.

Hold
There are 3 different hold functions you can choose from in the C11 parameter
1. Peak Hold: Grabs the highest weight (for materials testing, ie. tension and pulling force)
Press the HOLD key then add weight to the scale
The indicator will show the highest weight it recorded and hold it on the screen until a higher
weight is placed on the scale
2. Manual Hold: Grabs the current weight and holds it so it will not change/fluctuate
While weighing, press HOLD and the indicator will hold the current weight on the screen until
HOLD is pressed again
3. Average Hold: Used for weighing animal, the indicator will display the average weight sampled
from 3 seconds
Add livestock to scale and press HOLD
Indicator screen will show “LOC” for 3 seconds, then display the average weight from those
3 seconds
Press HOLD again to exit holding mode
Print
If the indicator is connected to a printer and the weight on the scale is stable press the PRINT
key to print the current weight
CALIBRATION PROCEDUER
How to enter
Available Keys
1. Turn on the scale by holding ON/OFF for 2 seconds
2. Press HOLD and PRINT together to access the setup menu.
Move to the right by 1 place of the set value
Increase the set value
Reduce the set value
Exit current menu
Move into next menu
esc
enter
- 6 -

Calibration Menu
].
1. Turn on the scale by holding ON/OFF for 2 seconds.
2. Press HOLD and PRINT together to access the setup menu.
3. If done correctly, the display should now show C01.
4. Press PRINT to access the C1 channel. The display should show
5. Press ZERO to choose which unit you want to calibrate in (1 = kg, 2 = lb).
6. Press PRINT to set the value. The display will now show C02.
7. Press PRINT to access the C2 channel. The display should show
8. Press ZERO
to change the setting to the decimal places desired (The C2 channel is
used to adjust the decimal point on the scale. A value of 1 means there is one digit
behind the decimal point.)
9. Press PRINT to set the value. The display will now show C03.
10. Press PRINT to access the C3 channel. The display should show
11. Press ZERO to cycle through the values until the desired graduation appears.
(The C3 channel adjusts the divisions on the scale. A value of 1 selected and C2 set
to 1, the scale will read in 0.1 lb. increments.)
12. Press PRINT to set the value. The display will now show C04.
13. Press PRINT to access the C4 channel. The display will show
14. Enter in the maximum capacity you want to use for this scale by using UNITS
cursor right, and TARE and ZERO to move the values down and up. (The C4 channel is
to move the
used to enter in the max capacity of the scale; Make sure this doesn’t exceed the max capacity
of the scale; Max capacity divided by the increment set in C02 and C03 above cannot
exceed 5000.)
17. The C5 channel calibrates zero on the scale. Make sure the scale is empty.
18. Press ZERO to change the value to 1.
19. Press PRINT . The display will count down from 10-1 while the scale is calibrating zero.
When the display shows 0 the zero calibration is complete.
The indicator offers 2 calibration methods, Single Point which uses one weight
to calibrate or Linear Calibration, which uses multiple (2-7) weights for a more
accurate calibration.
To Calibrate using only 1 calibration weight (Single Point Calibration)
20. Press PRINT to continue. The display will now show C06.
21. to access the C06 channel. The display will show
22. The C6 channel is used to calibrate the scale with a known weight. Press ZERO
. The display will flash
to
set the value of C6 to C6 1 Press PRINT SPAN , and
then show ######
enter
enter
enter
enter
enter
15. Press PRINT to set the value. The display will now show C05.enter
enter
Press PRINT enter
enter
enter
enter
enter
esc enter
C1 # ].[
C2 # ].[
###### ].[
C3 # ].[
16. Press PRINT to access the C5 channel. The display should showenter C5 # ].[
C6 0
[
].[
].[
- 7 -

1. Press PRINT enter to continue. The display will now show C06.
2. Press PRINT enter to access the C06 channel. The display will show
3. The C6 channel is used to calibrate the scale with a known weight. Press ZERO to set the
value of C6 to C6 2 . Press PRINT enter. The display will flash SPAN, and then
show [ LnE 2
4. Press ZERO to enter the number of weights you want to use (2-7); the more you use, the
more accurate the calibration will be. (an example of 2 will be used).
5. Press PRINT enter to set your value. The screen will flash [ dbno01 ] then [ 000100]
6. Enter the lowest calibration weight value you will use (at least 10% of max capacity you set
in C04) by using UNITS to move the cursor right and TARE and ZERO move the values
down and up.
7. Place the calibration weight you have on the empty scale and press PRINT enter .
8. The scale will count down from 10 to 0. Once 0 has been reached, the display will show the
screen will flash [ dbno02 ] then [ 000200 ].
9. Enter the next calibration weight value you will use by using UNITS to move the cursor right,
and TARE and ZERO move the values down and up .
10. Place the calibration weight you have on the empty scale and press PRINT enter .
11. The scale will count down from 10 to 0. Once 0 has been reached, the display will show he
screen will show [ CALEnd ].
12. Then follow steps 26-31 on the single point calibration instructions.
23. Enter the calibration weight value you will use (at least 10% of max capacity you set in C04 by
using UNITS
to move the cursor right, and TARE and ZERO move the values down
and up.)
24. Place the calibration weight you have on the empty scale and press PRINT enter.
25. The scale will count down from 10 to 0. Once 0 has been reached, the display will show
CALEnD .
26.
.
Press PRINT
27. Press HOLD esc to save and exit the setup menu.
28. The scale has now been calibrated. The display will show the value of the calibration weight
on the scale.
29. If the scale does not show the value of the calibration weight, check that the feet on the platform
are not screwed in too tightly, and verify that the platform is level.
30. Unload the scale; the display should read 000000.
31. If the scale does not display 000000, check that the feet on the platform are not screwed in too
tightly, and verify that the platform is level.
to set the value. The display will now show C07.
enter
To Calibrate using only multiple calibration weights (Linear Calibration)
C6 0
[ ].
[
[
[
.
- 8 -

Function Parameter Settings/Options
Weighing Unit
Decimal Setting
Maximum Capacity
Zero Calibration
Graduation Setting
(readability of the least
significant digit)
C01
C02
C03
C04
C05
1 = kg
2 = lb
1 = kg
2 = lb
3 = gram
4 = oz
0 = no decimal
1 = 0.0
2 = 0.00
3 = 0.000
4 = 0.0000
options: 1/2/5/10/20/50
Example with no decimal places (ie. C02=0)
1 = 1 lb
2 = 2 lb
5 = 5 lb
10 = 10 lb
20 = 20 lb
50 = 50 lb
set max capacity
0 = zero calibration not needed
1 = set the zero calibration (Please ensure scale is empty and
the stable light is on)
Note: for calibration only kg or lb are allowed
Model 1 Model 2
INDICATOR PARAMETER SETTINGS
The parameter settings menu has a calibration section (C01 to C07 explained above) and a
parameter settings section (C08 and up).
To access the calibration section the seal switch (located at one corner of the PCB) must be
OFF. This will allow access to all C01 and up settings. If the seal switch is ON, then only C08
and up can be accessed by the user. If you break the official seal by opening the back of the
indicator to access the seal switch, you may need to have the indicator recertified. Be sure to
adjust the seal switch back to the original setting after calibration/configuration has been
performed.
To enter calibration/parameter settings, follow the procedure below:
1. Make sure UNITS is set to either kg or lb
2. Press and hold HOLD and PRINT key at the same time for 2 seconds
3. Navigate through the settings (C01 to C45) as shown in the table 4 below by using the arrow
keys and return keys as labeled under each indicator button
4. Press PRINT key to enter/edit the parameter setting
5. Press PRINT enter / HOLD esc key to save and exit settings at any time
- 9 -
Calibration C06
0 = calibration not needed
1 = Ready to calibrate with one calibration weight
2 = Ready to calibrate using multiple calibration weights (Linear)
3 = Sensitivity Output

Function Parameter Settings/Options
Restore Default Settings 0 = do not restore
1 = restore to default settings
Warning Tone 0 = turn off warning tone
1 = turn on warning tone
Automatic
Power Off
0 = turn off auto power off
10 = power off automatically if no change within 10 minutes
30 = power off automatically if no change within 30 minutes
60 = power off automatically if no change within 60 minutes
Power Saving Mode
LED Version OP900A:
0 = turn off power saving setting
3 = turn off display if no change within 3 minutes
5 = turn off display if no change within 5 minutes
Hold Function
0 = turn off hold function
1 = Peak hold - Grabs the highest weight
2 = Manual hold - Grabs the current weight
5 = Auto Average hold - Average hold without the need to press the
hold key
Unit Conversion * See table 2
Upper Limit Alarm Set upper limit within the max. capacity
Lower Limit Alarm Set lower limit within the max. capacity
Inner Code Display check the inner code (raw data)
Set Date Set date from left to right: year/month/day
Set Time Set the time from left to right: hour/minute/second
Communication
Setting
Set the serial interface data output method:
0 = Turn off serial interface data output
1 = Continuous sending mode, for remote display
2 = Print to paper thermal ticket printer
3 = Command request mode, for computer.
4 = PC continuous sending mode, for computer
5 = PC/remote display, continuous sending mode
6 = Print to adhesive label thermal printer
7 = Print to Zebra/large adhesive label thermal printer
8 = Reserved
Baud Rate
C07
C08
C09
C10
C11
C12
C13
C14
C15
C16
C17
C18
C19
0 = 1200 (for OP-910 remote display)
1 = 2400
2 = 4800
3 = 9600 (for all printers and OP-910X, OP-910XL remote display)
4 = 14400
Manual Zero Range
0 = turn off manually zero setting
1 = ±1% max capacity
2 = ±2% max capacity
4 = ±4% max capacity
10 = ±10% max capacity
20 = ±20% max capacity
100 = ±100% max capacity
C20
- 10 -

Function Parameter Settings/Options
Initial Zero Range
0 = no initial zero setting
1 = ±1% max capacity
2 = ±2% max capacity
5 = ±5% max capacity
10 = ±10% max capacity
20 = ±20% max capacity
100 = ±100% max capacity
Zero Tracking
0= turn off zero tracking
0.5 = ±0.5d d = division
1.0 = ±1.0d
2.0 = ±2.0d
3.0 = ±3.0d
4.0 = ±4.0d
5.0 = ±5.0d
Note: zero tracking range cannot be bigger than manual zero range
Zero Tracking Time
0 = turn off zero tracking time
1 = 1 second
2 = 2 seconds
3 = 3 seconds
Overload Range 00 = turn off overload range
01-99d = overload range setting d = division
Negative Display
0 = -9d
10 = -10% max. capacity
20 = -20% max. capacity
50 = -50% max. capacity
100 = -100% max. capacity
Standstill Time
0 = quick
1 = medium
2 = slow
Standstill Range
1 = 1d d = division
2 = 2d
5 = 5d
10 = 10d
-
Digital Filter
(for filtering mov
ing items such as
animals)
C21
C22
C23
C24
C25
C26
C27
C28
0 = turn off dynamic filter
1 = 1 digital filter strength
2 = 2 digital filter strength
3 = 3 digital filter strength
4 = 4 digital filter strength
5 = 5 digital filter strength
6 = 6 digital filter strength
7 = 7 digital filter strength
8 = 8 digital filter strength
9 = 9 digital filter strength
Note: The higher the number, the higher the filter strength
0 = turn off noise filter
1 = 1 digital filter strength
2 = 2 digital filter strength
3 = 3 digital filter strength
Noise Filter C29
- 11 -

Function Parameter Settings/Options
0
1
2
3
=
=
=
=
yy.mm.dd
mm.dd.yy
dd.mm.yy
yy.mm.dd
0 =
1 =
0 - 5V output
4 - 20mA output
4 - 20mA current
0 = turn off relay output
1 = turn on relay output function 1
0 = Back to zero
1 = Press ZERO and PRINT together to back to zero.
2 = turn on relay output function 2
3 = Reserved menu
9.7000 - 9.9999
9.7000 - 9.9999
0 =
1 =
2 =
auto mode
gross mode
tare mode
0 - 9 (How much space between print outs)
0 - 9 (Where the data prints on the paper: 0 = left ; 9 = right)
0
1
=
=
do not print the date
print date the date
0
1
=
=
do not print the time
print the time
Print Time and Date
Analog Output Setting
Calibrate Current
Relay Output Setting
Zero Setting
Gravity of Calibration
Location
Gravity of Destination
Version No.
Print Mode
Print Carriage Return
Space Print
Date Print
Time Print
LED Display Brightness
C30
C31
C32
C33
C34
0 = normal weight
1 = weight reversed
C39
C36
C37
C38
Weight reversed setting
C41
C42
C43
C44
C45
C49 0-7 Brightness increased
Table 2. Unit Conversion Parameter Settings
Model 1 Model 2
- 12 -
Parameter Settings Units Available
C01 = 1 C12 = 0 kg only
lb only
kg / lb
kg / lb
kg / lb
lb / kg
lb / kg
lb / kg
kg only
lb only
&
C01 = 1 C12 = 1
&
C01 = 1 C12 = 2
&
C01 = 1 C12 = 3
&
C01 = 1 C12 = 4
&
C01 = 2 C12 = 0
&
C01 = 2 C12 = 1
&
C01 = 2 C12 = 2
&
C01 = 2 C12 = 3
&
C01 = 2 C12 = 4
&
Parameter Settings Units Available
C01 = 3 C12 = 0 oz only
lb / kg
g / oz / lb-oz
lb / kg / g / oz / lb-oz
lb only
&
C01 = 3 C12 = 1
&
C01 = 3 C12 = 2
&
C01 = 3 C12 = 3
&
C01 = 3 C12 = 4
&

Table 3. Default Parameter Settings
Function Parameter Default Setting
Weighing Unit C01 1
Decimal Setting C02 0
Graduation Setting C03 1
Maximum Capacity C04 100000
Zero Calibration C05 0
Calibration C06 0
Restore Default C07 0
Warning Tone C08 1
Automatic Power Off C09 0
Power Saving Mode C10 0
Hold Function C11 0
Unit Conversion C12 1
Upper Limit Alarm C13 000000
Lower Limit Alarm C14 000000
Inner Code Display C15
Set Date C16
Set Time C17
Communication Setting C18 1
Baud Rate C19 3 (9600)
Manual Zero Range C20 10
Initial Zero Range C21 10
Zero Tracking C22 1.0
Zero Tracking Time C23 3
Overload Range C24 9
Negative Display C25 10
Standstill Time C26 1
Standstill Range C27 2
Digital Filter C28 3
Noise Filter C29 2
Print Time and Date C30 0
Analog Output Setting C31 1
Relay Output Setting C33 1
Multi-connection add. C34 0
Wireless Communication C35 6
Gravity of Calibration Location C36 9.7936
Gravity of Destination C37 9.7936
Print Mode C41 0
Print Carriage Return C42 2
Space Print C43 0
Date Print C44 1
Time Print C45 1
LED Display Brightness C49 4
- 13 -

CONNECTORS
1. Load cell connection
The indicator can work with up to 8 load cells of 350Ω
Either 4 wire or 6 wire load cell connection
Please contact us directly if you have other special needs for your applications
There are two connection methods between the load cell and indicator
Quick Disconnect as shown below:
FIGURE 2: QUICK DISCONNECT CONNECTION DIAGRAM
Hardwire (Using Inner Terminal Block Connection:
Note: Make sure you follow all the anti-static rules to avoid damage to your indicator
Excitation voltage: 5V
DC
Largest output current: 120 mA
Excitation circuit: 5 VDC, 4 wire connection, 8 load cell of 350ohm maximum
Open the back cover of the weighing indicator, and insert signal cable to the terminal block
(see figure 3); Make sure the screw on terminal block is fixed tightly
FIGURE 3: INNER TERMINAL BLOCK CONNECTION DIAGRAM
- 14 -

RS232 SERIAL OUTPUT FORMAT
Follow the pin out of Table 5 below to connect the indicator the RS-232 Serial device
Table 5. DB9 Pin Description
DB9 Pin Definition Function
2 TXT Transmit Data
3 RXD Receive Data
5 GND Ground Interface
The serial output format depends on the settings for parameter C18. The serial output consists
of a string of ASCII characters. Here is a list of the serial parameters
8 data bits
1 stop bits
No parity
No handshaking
2. RS-232 Connection (DB9 9 pin Connector)
Table 4. Wiring Color Code
Signal Name Description
+Exe/ +EX Positive excitation voltage to load cell
+IN / +SIG Positive output signal from load cell
HD / SHLD Shield Wire
-IN / -SIG Negative output signal from load cell
-EXC / -EX
Color Code
RED
GREEN
YELLOW/THICK BLACK
WHITE
BLACK Negative excitation voltage to load cell
The DB9 9 pin serial connector is used for different purposes depending on the indicator model
Figure 4 shows the pin assignment on the DB9 9 pin connector
FIGURE 4: DB9 SERIAL CONNECTOR PINOUT
- 15 -

Remote Display Continuous Sending Mode (C18=1)
For use with a Scoreboard/Remote Display Note: Baud Rate must be set to 1200 (C19 = 0)
Output Continuous
Format
S
T
X
S
W
A
S
W
B
S
W
C
X X X X X X X XX X X X C
R
C
K
S
1 2 3 4 5 6
State A
Bits0,1,2
0
1
0
1
0
1
1
0
1
1
0
0
2
0
0
0
1
1
Bits3,4
10
1 0
Decimal point position
XXXXXX0
XXXXXXX
XXXXX.X
XXXX.XX
XXX.XXX
Division
X1
X2
State B
BitsS function
Bits0 gross=0, net=1
Bits1 Symbol: positive=0, negative=1
Bits2 Overload (or under zero)=1
Bits3 dynamic=1
Bits4 unit: lb=0, kg=1
Bits5 Constant 1
Bits6 Constant 0
State C
Bit2 Bit1 Bit0 unit
0 0 0 Kg or lb
0 0 1 g
0 1 0 t
Bit 3 printing=1
Bit 4 Extend display=1
Bit 5 Constant 1
Bit 6 Constant 0
- 16 -

Print Mode (C18 = 2)
For printing on a non-adhesive ticket printer. Parameters 16, 17, 30, & 42-45 all effect your
ticket print out.
Normal weighing ticket printout example:
Date: 05/01/2017
Time: 11:30:52
Net: 25.6lb
Tare: 10.3lb
Gross: 35.9lb
The R command will trigger the indicator to output the following data format:
Command Request Mode (C18=3)
In this mode, the indicator can receive ASCII commands as listed below
Command Name
TTare
Z Zero
P Print
R G.W / N.W
C Kg/lb
G G.W
Function
Save and clear tare
Zero gross weight
Print the weight
Read gross weight or net weight
Kg/lb conversion
Check gross weight at net weight mode
Computer Continuous Sending Mode (C18=4)
S1: weight status, ST=standstill, US=not standstill, OL=overload
S2: weight status, GS=gross mode, NT=net mode
S3: weight of positive and negative, “+” or “-”
Date: weight value, including decimal point
S4: “kg” or “lb”
CR: carrige return
LF: line feed
- 17 -

PC or Remote Display Continuous Sending Mode (C18=5)
Print to Adhesive Label Printers (C18=5) & (C18=6)
Printing formats:
Date:
Time:
Net:
Tare:
Gross:
05/01/2017
11:30:52
25.6lb
10.3lb
35.9lb
Date:
Time:
XX.XX. XX (yy.mm.dd)
XX.XX.XX (hh.mm.ss)
NET
TARE
GROSS
6.00kg (net weight)
2.88kg (tare)
8.88kg (gross weight)
- 18 -
Table of contents
Other Tufner Measuring Instrument manuals