Turtle Tough TT-ISE-CL2 User manual

TURTLE
TOUGH
®
1
TURTLE
TOUGH
®
MANUAL
TT-ISE-CL2 ANALYSER
Reagentless Amperometric Free or Total Chlorine Residual Analyser
turtletoughsensors.com
Chlroine Analyser

TURTLE
TOUGH
®
2

TURTLE
TOUGH
®
3
General Informaon 4
4
4
5
6
7
9
10
11
12
13
14
15
2 Sensor Preparaon, Assembly, Maintenance
16
18
19
20
Cleaning The Electrode Finger Tip 21
22
3 System Component Idencaon and Descripon
23
24
25
Installaon
26
Unpacking 27
27
27
28
Electrical Installaon
29
29
30
31
Chlorine Sensor Input 31
31
Power Input 31
32
32
32
Startup 33
Touch Screen Interface and Navigaon
35
Menu Structure 36
37
38
39
39
41
41
42
43
Troubleshoong
44
44
45
45
Maintenance 46
Parts List 47
Warranty 48
Return Goods 50
Support 51
Content

TURTLE
TOUGH
®
4
e
policyTurtle Tough
changes
obligation
follow precautions before attempting to start up the
analyser

TURTLE
TOUGH
®
5
analyser
The Turtle Tough model TT-ISE-CL2 uses a membrane covered amperometric sensor to.
continuously monitor and control free or total residual chlorine without any reagents or moving parts.
The difference between the two models is the type of sensor used and corresponding measurement it
produces, free or total chlorine. The analyser interface is the same for both sensor types except for
the available operating ranges for each sensor type.
The measurement system does not alter the sample and does not add chemicals to the sample
stream.
TT-ISE-CL2
Wastewater that has undergone tertiary treatment may be possible to monitor if final water product is
at or near drinking water standards.
For the measuring methods used here, chlorine diffuses through the membrane from the
measurement medium and, combined with the electrolytes, triggers an electrical signal at the working
electrode. This signal is proportional to the concentration of chlorine and is amplified by the
electronics. The measurement signal is independent of the temperature of the media due to an
integrated temperature compensation.
pH has an effect on chlorine measurement accuracy since the sensor measures only the
hypochlorous acid (HOCl) form of chlorine, and not the hypochlorite ion (OCl
-
) form of chlorine. The
ratio of these two different types of chlorine varies with the pH value of the analyte. This behavior is
described by the hypochlorous acid dissociation curve. In the 3-electrode sensor there is a linear loss
of slope, or decrease in residual value, of about 5% for every 1 pH unit increase after pH 7. In the 2-
electrode sensor there is a 65% loss of slope at pH 8.
The sensor is not intended or recommended to measure or ensure the absence of chlorine.
The sensor is not suitable for determining organic chlorination agents such as cyanuric acid based
products.
The sensor cannot distinguish between various species of chloramines.
The analyser requires a constant and consistent sample flow at no more than atmospheric pressure
as well as a waste drain.
The monitoring of reservoirs, basins or tanks requires a pump to deliver sample to the analyser at the
required pressure and flow rate. Intermittent operation is possible within certain time limitations.
Residual measurements are determined and output continuously, the analyser does not use a sample
and hold method for residual determination.
The microprocessor based electronics are designed with additional capacity to accept both software
and hardware upgrades so that the analyser can be converted over time into a multi-function and
multi-parameter measurement system.

TURTLE
TOUGH
®
6
analyser
NOTE: Using the sensor without mounting it in the supplied flow cell will lead to incorrect
measurement results; it cannot simply be submerged in a pipe, basin, channel or tank.
analyser
If no chlorine is to be measured for over 24 hours the sensor must be disconnected from the
analyser, the membrane cap emptied, rinsed and the sensor stored per instructions.
Follow the instructions closely for
filling the membrane cap with electrolyte.
The membranes can be damaged by high pressureThe sensors should be operated under as little
pressure as possible, with the measurement media able to flow freely. If this is not possible, the
sensors can be operated under a constantpressure of up to 1 bar (relative pressure) or 2 bar
(absolute pressure). Fluctuations in pressure must be avoided.
The membrane cap must only be filled and fully screwed onto the sensor immediately before
inserting it into the flow cell to prevent salt or gel deposits on the inside surface of the membrane.
Touching and contaminating the electrode finger can damage it, making the sensor unusable.
The measurement media must not contain hydrophobic substances (e.g. oil or grease), which can
damage the membrane cap.
The measurement media must not contain surfactants (surface-active substances e.g. from
detergents, cleaning agents or disinfectants).
The presence of air bubbles in the measurement medium in front of the membrane may produce
incorrect measurement results.
If there is no chlorine in the measurement medium for more than 24 hours this will lead to incorrect
measurement results due to biofilm deposits on the membrane. You should avoid operating the
sensors with measurement medium which does not contain chlorine. After operation in a chlorine-free
medium, a settling time period is required before valid measurements can resume.
Using the sensors to measure media containing other oxidants in addition to chlorine, reducing
agents or corrosion protection agents may lead to incorrect measurement results. Refer to the
specifications for each sensor for interfering oxidants and substances.

TURTLE
TOUGH
®
7
Contaminant Secondary MCL
Aluminum 0.05 to 0.2 mg/L
Chloride 250 mg/L
Color 15 color units
Copper 1.0 mg/L
Corrosivity Non-corrosive
Fluoride 2.0 mg/L
Foaming agents 0.5 mg/L
Iron 0.3 mg/L
Manganese 0.05 mg/L
pH 6.5 - 8.5
Silver 0.1 mg/L
Sulfate 250 mg/L
Total Dissolved Solids (TDS) 500 mg/L
Zinc 5 mg/L
Turbidity not to exceed 5 NTU
NO CORROSION INHIBITORS PERMITTED
3-Electrode Total Chlorine Sensor Description & Operation
This membrane-covered, amperometric sensor is used to measure the total chlorine concentration in
water.
The sensor detects "free chlorine" liberated from organic chlorination products (chlorine gas,
hypochlorite, etc.) and "bound chlorine", for example chloramines. It cannot distinguish between
various forms of chloramines.
This sensor can only be used in media similar in quality to drinking or swimming pool water. Typical
applications include the monitoring of filtered, treated drinking water and swimming pools.
Iron and manganese levels above USEPA Secondary Drinking Water Standards MCL will prevent the
operation of the 3-electrode chlorine sensor.
The sensor is not suitable for detecting the absence of chlorine.
This sensor has a 3-electrode measuring system. Three electrode sensors work on the potentiostatic
principle. The reference and counter electrode are separated. The working electrode is membrane-
covered. Together with the reference electrode, it is located in an electrolyte chamber that contains a
special pH buffered electrolyte and is separated from the process medium. The potential of the
reference electrode is especially constant due to its high resistance. The current flows through the
counter electrode. The sensor is loop powered. This loop power is generated by the analyser
electronics and supplied through the cable that transmits the residual signal.
The working electrode (cathode) is made of gold (Au) while the reference electrode is made of
silver/silver chloride (Ag/AgCl). The counter electrode is made of stainless steel.
In this measurement method the chlorine diffuses out of the process medium and through the
selective membrane, inducing an electrical signal in combination with the electrolyte on the working
electrode. The signal, which is proportional to the concentration of chlorine, is amplified by the
electronics. Due to an integrated temperature compensation system, the measurement signal is
independent of the medium temperature.

TURTLE
TOUGH
®
8
2 & 3-Electrode Free Chlorine Sensor Description
The 2 or 3-electrode membrane-covered, amperometric sensors are used to measure the
concentration of free chlorine in drinking and swimming pool water, industrial, process and cooling
water.
The following inorganic chlorinating agents can be measured with the sensor for free chlorine:
chlorine gas (Cl2), electrolytically generated chlorine, sodium hypochlorite (NaOCl, chlorine bleach
lye), calcium hypochlorite (Ca(OCl)2) or chlorinated lime (Ca(OCl)Cl).
The sensors are not suitable for detecting the absence of free chlorine.
The integrated electronics of the sensors provides a temperature-compensated current signal
of 4 to 20 mA.
The sensor for free chlorine with reduced pH dependency is a potentiostatic 3-electrode
sensor with a micro porous, hydrophilic (moisture attracting) membrane and special electrolyte.
The 3-electrode sensor can only be used in water of drinking or swimming pool water quality. Solid
materials in the media clog up the membrane and prevent the sensors from working correctly.
Iron and manganese levels above USEPA Secondary Drinking Water Standards MCL will prevent the
operation of the 3-electrode chlorine sensor.
Suitable chlorinating agents that can be measure include inorganic chlorine compounds such as
chlorine gas (Cl2), chlorine produced by membrane electrolysis (not suitable: chlorine electrolysis
without a membrane), sodium hypochlorite (NaOCl), calcium hypochlorite (Ca(OCl)2) and
chlorinated lime CaCl(OCl).
The measured variable is the free chlorine (including "(iso)cyanuric acid chlorine") independent of
the (iso)cyanuric acid concentration. Combined chlorine (chloramine) is not measured. The sensor is
not suitable for determining organic chlorination agents such as cyanuric acid based products.
This membrane type allows both hypochlorous acid and hypochlorite from the water to be measured
to reach the electrode chamber through the membrane. A buffered electrolyte sets a specific ratio of
HOCl and OCl- here.
The membrane also allows the flow of ions from the electrolyte into the water, which eventually
causes the electrolyte to be unable to function. A measured medium containing tensides (detergents,
surfactants) may lead to an accelerated exchange between the electrolyte and the medium. The
decision whether a membrane can operate in a medium containing tensides can only be made on a
case by case basis.
The working electrode is made of gold, the high-impedance reference electrode of silver/silver halide
and the current carrying counter electrode of stainless steel.
With the 3-electrode system, the selectivity is increased by the improved stability of the potential:
the system consists of the measuring or working electrode (ME), the counter electrode (CE) and a
reference electrode (RE). The potential between ME and RE is maintained at the value required for
the flow-through reaction by means of a potentiostatic switching arrangement. The RE does not
carry any current. The current flows via the CE. Again, in the case of the 3-electrode system, the
diffusion-controlled limiting current is proportional to the concentration of the analyte.
No zero adjustment is required. The electrode chamber contains a defined electrolyte, with which the
sensor does not exhibit a zero signal. Therefore, no zero adjustment using analyte-free water is
required.

TURTLE
TOUGH
®
9
As there must be an electrical connection between the counter electrode and the measurement
medium, the measurement medium must have a minimum conductivity of approx. 10 µ S/cm. This
means that the sensors are not suitable for use in highly-purified water, or similar.
2-Electrode Free Chlorine Sensor Description
The sensor for free chlorine is a potentiostatic 2-electrode sensor with a micro porous, Hydrophobic
(moisture repellent) PTFE membrane and special electrolyte.
Suitable chlorinating agents that can be measure include inorganic chlorine compounds such as
chlorine gas (Cl2), electrolytically generated chlorine, sodium hypochlorite (NaOCl), calcium
hypochlorite (Ca(OCl)2), chlorinate lime CaCl(OCl). The sensor is not suitable for determining
organic chlorination agents such as cyanuric acid based products or combined chlorine.
The sensors are not suitable for detecting the absence of free chlorine.
The 2-electrode sensor can only be used in water of drinking or swimming pool water quality. Solid
materials in the media clog up the membrane and prevent the sensors from working correctly.
If the membrane comes into contact with tensides (detergents, surfactants) from washing, cleaning or
disinfecting agents, the hydrophobic properties of the membrane are lost, the membrane becomes
permeable, and therefore damaged. For this reason, contact with these substances must be strictly
avoided.
The membrane does not allow hypochlorite anions to penetrate the membrane to reach the
electrolyte chamber; so, the sensor does not measure OCl-, only hypochlorous acid.
The (cathode) is made of gold (Au). The anode, which performs the function of
is made of silver (Ag), and is provided with a coating of
silver chloride (AgCl).
A 2-electrode system has the disadvantage that the potential between ME and CE can vary. This
applies particularly in an open system, as the conditions at the electrodes can vary here, as a result
of fluctuations in the composition of the water. If the potential does vary, then other oxidizing
agents contained in the water having a polarization voltage in a similar range, can also react
and distort the result.
No zero adjustment is required. The electrode chamber contains a defined electrolyte, with which the
sensor does not exhibit a zero signal. Therefore, no zero adjustment using analyte-free water is
required.

TURTLE
TOUGH
®
10
Details Total Chlorine Sensor and 3-Electrode Free Chlorine Sensor
(1)
(2)
(3)
(4)
(5)
(6)
(14)
(13)
Cable gland nut
Terminal Cover
O-ring
2-pin terminal for measuring cable connection
Electrode shaft with integrated electronics
Counter electrode (stainless steel)
O-ring
Electrode finger (reference electrode)
Measurement electrode
Transparent vent seal
Membrane cap
Membrane holder (stainless steel)
Membrane
Vent hole
(7)
(8)
(9)
(10)
(11)
(12)

TURTLE
TOUGH
®
11
Details 2-Electrode Free Chlorine Sensor
(1)
(2)
(3)
(4)
(5)
(12)
(11)
(6)
(7)
(8)
(9)
(10)
(1)
Cable gland nut
(2)
Terminal
Cover
(3)
O-ring
(4)
2-pin terminal for measuring cable connection
(5)
Electrode shaft with integrated electronics
(6)
O-ring
(7)
Electrode finger (reference electrode)
(8)
Measurement electrode
(9)
Transparent vent seal
(10)
Membrane cap
(11)
PTFE membrane
(12)
Vent hole

TURTLE
TOUGH
®
12
Specifications, Total Chlorine Sensor
Analyte Total chlorine
Membrane type Hydrophilic membrane
Measuring cable
connection
2
-
pin terminal, polyamide PG7 screw
connection;
wire cross section 2x 0.25mm2, cable diameter approx. 4 mm
Voltage supply
U
B
12 to 30 V DC (electrical isolation
recommended)
Electromagnetic
compatibility
According to EN 61326-1
Interference emission: Class B
Interference immunity: To industrial requirements
Output signal 4 to 20 mA
Settling time 2 hours
Incident flow velocity Approx. 15cm/s or 30L/hr (8 GPH)
Measuring ranges 0 to 0.5 mg / l (ppm)
0 to 2 mg / l (ppm)
0 to 5 mg / l (ppm) 0
to 10 m g / l (ppm) 0
to 20 m g / l (ppm)
Response time
t
90
About 2 min
Operating temperatures /
temperature compensation
+5 to +45 °C
Zero point adjustment Not required
pH value operating range
4 to 12 pH
pH dependency
(loss of slope)
Linear decrease of approx. 5 % per each upward pH unit
(starting from pH 7)
Disruptive substances /
cross sensitivity
Chlorine dioxide,
Ozone. Calcium deposits. Iron interferes with electrode
function and must not exceed USEPA MCL
Pressure resistance
p
abs
max. 2 bar
prel max. 1 bar
No pressure fluctuations are admissible when operating under pressure.
We recommend unpressurized operation (atmospheric pressure).
Material
Shaft, cover, cap: PVC
Membrane disk holder: stainless steel
Dimensions Diameter: 25mm, length: 220mm
Weight about 125g
Maintenance Check the measurement signal: regularly, at least once a week
Replace the membrane cap: once a year (subject to water quality)
Change the electrolyte: every 3 to 6 months
Storage
Sensor: frost
-
free,
dry and without
electrolyte, can be stored for an
unlimited time at +5 to +45 °C
Electrolyte: In the original bottle and protected against sunlight at +5 to
+25 °C

TURTLE
TOUGH
®
13
Free Chlorine Sensor Data, 2 -Electrode & 3-Electrode with Reduced pH Dependence
Analyte
Membrane type Hydrophobic PTFE membrane Hydrophilic membrane
Measuring cable
connection
2
-
pin terminal, polyamide PG7 screw
connection;
wire cross section 2x 0.25mm2, cable diameter approx. 4 mm
Suitable chlorination agents Inorganic chlorine compounds: NaOCl (sodium hypochlorite), Ca(OCl)2,
chlorine gas, chlorine produced by membrane electrolysis (not suitable:
chlorine electrolysis without a membrane)
Voltage supply UB 12 to 30 V DC (electrical isolation recommended)
Electromagnetic
compatibility
According to EN 61326-1
Interference emission: Class B
Interference immunity: To industrial requirements
Output
signal
4 to 20
mA
Settling time 1 hour 2 hours
Incident flow velocity Approx. 15cm/s or 30L/hr (8 GPH)
Measuring ranges mg / l (ppm)
0.05 to 0.5
0.05 to 2
0.05 to 5
0.05 to10
0.05 to 20
0.05 to 50
0.05-100, 0.05 to 200
0.05 to 2
0.05 to 5
0.05 to10
0.05 to 20
0.05 to 200
Resolution 0.01 mg/l for 0.5,2,5,10,20 mg/l
0.1 mg/l for 50, 100, 200 mg/l
0.01 mg/l for 2,5,10,20 mg/l
0.1 mg/l for 50, 100, 200 mg/l
Response time
t
90
About 30 sec. About 2 min
Operating temperatures /
temperature compensation
+5 to +45 °C
Zero point adjustment Not required
pH value operating range
6.0 to 8 pH
4 to 9 pH
pH dependency
(loss of slope)
No loss from pH 5-7, at pH 8 about
65%, at pH9 about 90%
No loss from pH 5-7, at pH 8 about
10%, at pH9 about 20%
Disruptive substances /
cross sensitivity
Chlorine dioxide,
Ozone, Combined
Chlorine, reducing agents,
corrosion protection agents,
Calcium deposits.
Chlorine dioxide,
Ozone, Combined
Chlorine. Calcium, Iron over
USEPA MCL prevents electrode
function. Corrosion inhibitors
Min. Sample Conductivity
50µS / cm (not suitable for highly purified water)
Pressure resistance
p
abs
max. 2 bar
prel max. 1 bar
No pressure fluctuations are admissible when operating under pressure.
We recommend unpressurized operation (atmospheric pressure).
Material
Shaft, cover, cap: PVC Shaft, cover, cap: PVC, Membrane
disk holder: stainless steel
Dimensions & Weight Diameter: 25mm, length: 220mm, Weight about 125 g
Maintenance Check the measurement signal: regularly, at least once a week
Replace the membrane cap: once a year (subject to water quality)
Change the electrolyte: every 3 to 6 months
Storage Sensor: frost-free, dry and without electrolyte, can be stored for an
unlimited time at +5 to +45 °C
Electrolyte: In the original bottle and protected against sunlight at +5 to
+25 °C

TURTLE
TOUGH
®
14
The flow cell (part no. 303500) is fastened to a wall or mounting panel with mounting bracket (part no.
303501)
(1)
(2)
(3)
(4)
(5)
Sensor
Mounting bracket (303501)
Connection G1/4 for hose ø 8mm x 6mm
Flow cell housing
Removable inspection glass

TURTLE
TOUGH
®
15
(1)
(2)
(8)
(9)
(10)
(3)
(4)
(7)
(5)
(6)
(1)
Sensor
(6)
Inspection glass
(2)
Union nut
(7)
Inlet G1/4A or DN10
(3)
Flow Cell
(8)
Stepped collar 1"
1
(4)
Outlet G1/4A or DN10
(9)
Compression ring
1
(5)
O-ring, inspection glass (10) O-ring, stepped
collar
1
1 Part of the flow cell.
The transparent inspection glass (6) can be unscrewed from the fitting housing for cleaning.
First push the O-ring (10) onto the sensor (1), then the compression ring (9)
and over it the 1" stepped collar (8) (starting from the Pg screw connection).
The stepped collar (8) must engage in the groove.
After the sensor has been prepared in this manner, insert it into the flow cell
(3) and fasten it in place with the union nut (2).

TURTLE
TOUGH
®
16
The electrolyte will come out of the valve opening when the membrane cap is screwed on. Wear
safety goggles and gloves.
Wash off electrolyte (an aqueous solution of an alkali halide) under flowing water.
NEVER SCREW THE MEMBRANE CAP ON OR OFF WITHOUT FIRST MOVING THE VENT
SEAL AWAY FROM THE VENT HOLE. NOT DOING SO WILL CAUSE A VACUUM OR EXCESS
PRESSURE AND DESTROY THE MEMBRANE.
DO NOT TOUCH THE BOTTOM OF THE MEMBRANE CAP. DO NOT TOUCH THE ELECTRODE FINGER
OR GET IT DIRTY.
Do not unscrew the stainless steel membrane disc holder from the membrane cap, this will
cause the membrane to be misaligned and damage it.
The steps must be performed exactly as they are described below.
In order for the sensor to function correctly, the membrane must be screwed onto the sensor.
The first screw-in resistance is the sealing O-ring. The membrane cap must be screwed on further
until it comes into contact with the sensor shaft.
Put on protective gloves for handling electrolyte which can irritate the skin.
Use a small screwdriver or similar tool to raise the vent seal covering the
vent hole and push it down out of the groove to open the vent hole circled.
Screw the membrane cap off the shaft of the sensor.
Discard the used electrolyte, clean the membrane cap with clean water and
then dry it.

TURTLE
TOUGH
®
17
To avoid forming bubbles do not shake the electrolyte. Make sure the vent
hole is open. Tilt the electrolyte bottle as shown to minimize bubbles and fill
the membrane cap up to the brim with electrolyte, if needed tap to eliminate
bubbles.
Holding the shaft of the sensor vertically (do not hold the
membrane cap by covering the vent hole), slowly screw on
the filled membrane cap. You will feel resistance when the
cap reaches the O-ring, continue tightening until the cap
touches the electrode shaft. Excess electrolyte will overflow
from the cap, and when almost completely tightened it will
escape from the open vent hole. Be sure the cap is
screwed on tightly against the sensor shaft.

TURTLE
TOUGH
®
18
After tightening the membrane cap, slide the vent seal up into the groove, so
that it covers the vent hole. Rub around the vent seal to remove any bubbles
or voids. If liquid leaks through the membrane, the membrane is faulty and
you must use a new membrane cap.
Take the stepped collar, smaller diameter up, and insert a screwdriver into
the split to enlarge the opening.
Push the stepped collar down until it snaps into position in the sensor shaft
groove.
Slide compression ring and O-ring onto the sensor shaft in the order shown.
Push the ring and O-ring up against the stepped collar.

TURTLE
TOUGH
®
19
Install the Sensor
Insert the sensor into the flow cell with the wiring terminals facing you. Slide
the union nut over the sensor and tighten onto the flow cell.
Slide the cable gland nut and terminal cover over the sensor cable in the
order shown. The cable shrink wrap should protrude through the cable gland
nut.
There are two possible sensor cable colors. Attach the blue with white dots
(or black) sensor cable to the positive terminal of the sensor. Attach the white
with blue dots (or white) sensor cable to the negative terminal of the sensor.

TURTLE
TOUGH
®
20
Screw the terminal cover onto the sensor and then tighten the cable gland
nut.
The solution ground pin on the front of the flow cell is normally not needed,
it may be used if you find a solution ground is necessary.
The sensor will not provide a constant value until after a conditioning time of 2 hours has
passed. It cannot be calibrated until then.
The calibration procedure should be repeated on the day after initial commissioning.
On the chlorine sensor, the screw coated with thread locking paint must not be
adjusted. If the thread locking paint is damaged claims under the warranty will not be
honored.