Turtle Tough DSS HandHeld FIELD Communicator User manual

TURTLE
TOUGH
®
turtletoughsensors.com
MANUAL
DSS HANDHELD FIELD COMMUNICATOR
Dissolved Oxygen
DO

Product name DSS HandHeld Field Communicator
Code TT-HFC
Mechanical
Housing ABS
Mounng Handheld
IP Class Housing IP40
Connector Quick connect plug
Temperature Usage -15 to +50 °C (Storage -35 to +75 °C)
Weight 130 grams with baery (4.6 ounces) 100 grams without baery (3.5 ounces)
Dimensions D 26 x W 60 x H 120 mm (1.0” X 2.4” X 4.7”)
Electrical
Supply 9V baery (Alkaline or Lithium)
Consumpon ~45 mA with DSS sensor ‘On’
Baery life ~6 hrs Alkaline or ~12 hrs Lithium
Auto shutdown aer 25 seconds * without communicaon
Interface Smart Digital DSS Modbus RTU pH Sensors
Baud rate 9600 or 19,200 kbps (selectable)
CE mark EN61326A
* Number of seconds unl auto shutdown starts from when sensor is disconnected from HFC
DSS Handheld Field Communicator
Turtle Tough’s Handheld Field Communicator (HFC)
calibrates and congures DSS Smart Digital RS485
Modbus RTU sensors at any locaon. All values
stored in non-volale sensor EEPROM for hot-swap
portability when installed back into eld service.
Specicaons

Programming and Navigation
The Handheld Field communicator (HFC) has 3-digit display and 16 LEDs to show readings and analyc data
as well as to calibrate and congure sensor. Programming done by 4 key front panel. ‘Mode’ key used to toggle
and navigate to each LED. ‘Up’ or ‘Down’ buons scroll available opons and adjust values. ‘Mode’ key is used to
make selecons and save entries. ‘View’ key provides addional informaon for the given LED mode (see table for
details on use of ‘View’). Once baud rate and node of connected sensor are entered all parameters are automacally
loaded for zero conguraon plug and play use in the eld.
3-digit display
Use ‘Up’ or ‘Down’ buons
to scroll available opons
and adjust values.
Press ‘View’ for addional
informaon for the given
LED mode
Use ‘Mode’ key to make
selecons & save entries.

DSS Handheld Field Communicator for Calibraon, Conguraon,
Spot Measurement and Troubleshoong of DSS DO Sensors RS485 Modbus RTU
Led Label Parameter Descripon and Method to Access Range Default
Sensor Type Measurement type Load opons for connected sensor
‘View’ key shows soware revision
pH or ORP or DO or ISE or CON
(autodetected) Per Type
Reading***** Process Parameter****
Display current calibrated value
See Display Features for a detailed
explanaon of how to use this LED mode
0.0 to 150.0 ppm*
0 to 1,500 percent (%) saturaon*
Use ‘View’ key to show % sat
Per
Measured
Soluon
Abs mV / mS Process Parameter Display the absolute mV value from
connector dissolved oxygen sensor
0 to +250 (Negave mV values are not
possible for galvanic DO cell)
Per Sensor
& Media
Cal Temp. Oset calibraon of
temperature in °C **
Adjust temp reading up & down
‘View’ key shows current temp cal. ±25.0 °C * from raw value 0.0
Cal Oset Not Used for DO
Measurements N/A N/A N/A
Cal Slope Denes span (gain) for
DO Sensor **
mV per DO ppm units
‘View’ key shows current slope 0.70 to 6.00 mV per ppm unit 1.60
Dampener Smoothing dampener
& output delay ***
Sets number of seconds to be used
for dampener for process value(s) 1, 2, 3, 4, 5, 8, 10, 15, 20 or 30 Seconds 10- Dampen
1 - Delay
Step Change Sensivity for
‘Up’ & ‘Down’ buons
mV increment for each me the
‘Up’ or ‘Down’ buon is depressed
Choices:
0.05, 0.10, 0.20, 0.5, 1.0 or 2.0 0.10
Special Salinity to compute %
saturaon value
User entered salinity value in PSU to
compute percent (%) saturaon value 0.0 to 50.0 PSU 0.0
Baudrate Sets Baudrate for Com Toggle between 9600 or 19,200 kbps 9600 or 19,200 kbps Per Network
Node Sets Address for Com Chose a valid address on network From 001 to 247 Per Sensor
Item/Serial
Number
Sensor Item Number
and Serial Number
Item # denes model; ‘View’ shows
Serial # (unique traceable idener)
Item # from 1-9,999 with >999
Serial # per DSS Serial Number Scheme Per Sensor
Days in Use Total me DSS sensor
is energised
Increments me in use aer dispatch
from factory to track sensor lifeme
and predicve maintenance purposes
0-65,535 in units of days
(>999 displayed ashing)
Within ±2% accuracy
Per Sensor
Field Use
Max/Min °C Displays max & min
temp in eld use
The max temp in eld use is shown;
Push ‘View’ buon for min temp -40 to +210 °C * Per Sensor
Field Use
* Negave values are always shown as ashing.
Color notes
Parameters in green are dened by factory at dispatch me or determined from eld use.
Parameters in grey can be adjusted as desired.
Parameters in blue are obtained from wet calibraons done with DSS sensor in the eld.
*HFC display maxes out at 99.9 for reading for DO ppm (actual max to 150) and 999 for DO % saturaon (actual max to 1,500)
** Holding the ‘View’ key for 3 to 5 seconds in this LED mode shows the ‘Days in Use’ since this calibraon was performed
** Holding both the ‘View’ and ‘Up’ keys for 3 to 5 seconds in this LED mode will reset all calibraon values back to default
*** Holding the ‘View’ key for 3 to 5 seconds allows for the delay from boot value to be shown as well as adjusted
**** Holding both ‘Up’ & ‘Mode’ keys shows soware rev or both ‘Down’ & ‘Mode’ keys shows build date in Reading LED mode
***** User Adjustable Timeout Feature: Press ‘Down’ + ‘View’ in ‘Reading’ mode to set minutes before automac shuto occurs
Note on Dissolved Oxygen Sensors When Values Are Greater Than 99.9 Ppm Or 999 % Saturaon
When the ppm value exceeds 99.9 it will not be shown (actual max value supported in output is 150ppm). When the computed percent (%)
saturaon value exceeds 999 it will not be shown on display (actual max value supported in output is 1,500%).
Note on DSS Dissolved Oxygen sensor slope calibraon
The DSS Smart Digital Dissolved Oxygen sensors will always have just a single slope and no oset. No oset calibraon is required for the
true galvanic cell since it has a true zero (0mV at 0ppm). The use of just a single slope is possible since the response of the industrial Turtle
tough’s dissolved oxygen (DO) smart digital sensor with Modbus RTU digital output is linear throughout the enre range dened by the mV
per ppm slope calibraon to dene the eecve span.
‘Modbus Com’ & ‘Baery Low’ LED
• The “Modbus Com” LED is illuminated briey each me that a communicaon packet is sent or received.
• The “Baery Low” LED will at rst ash as warning & then illuminate connuously when the 9V baery should be replaced.
o Must change 9V baery when LED is illuminated to ensure valid readings and calibraons.

Setup of DSS RS-485 Modbus RTU Sensor to Handheld Field Communicator
1. Instrucons for scanning and changing nodes on page 10.
2. Press the ‘Mode’ buon to turn on the Handheld Field Communicator. Once on, the HFC will aempt to
communicate with the last used baudrate and node address. If either no sensor is connected or available
at the last used baudrate and node address then three dashes “---” are shown on display. If no buons
are pressed for 25 seconds the Handheld Field Communicator will automacally turn itself o to conserve
baery life.
a) If previous baudrate and node address are valid for connected sensor the HHC will automacally load
all relevant LED opons and addressable parameters for that sensor type.
3. Pressing ‘Mode’ buon navigates to ‘Node’ LED mode. Use ‘Up’ & ‘Down’ keys to scroll to node of the con-
nected sensor. Node informaon is typically found on label of sensor. If this informaon is not available, the
DSS Windows soware can be used scan the sensor in queson to determine the current node address.
The baudrate and node address of the DSS sensors can only be changed by the Windows soware. When
the desired node address is reached press the ‘Mode’ key enter the value.
a) Default nodes: pH is 1, ORP is 2, Wide-Range ORP is 3, DO is 4, ISE is 5, Conducvity is 6. If mulple
sensors of same type are used on one Modbus RTU network then node address for each same type
sensor must dier from default to ensure a unique and valid node address.
4. If baudrate needs to be adjusted (9600 or 19,200 kbps) then the HFC automacally navigates to this LED
mode next.
5. HFC returns to reading mode aer selecng node and baudrate. If the selecons are valid then the process
reading is shown otherwise three dashes “---” are shown.
6. Press ‘Mode’ buon aer reading LED to toggle to sensor type LED which shows type of sensor that is
connected.
Sensor Serial Number, Item Number and Total Time in Field Service
Systemac tracking achieved with factory stamped sensor serial and item number. The internal clock on the
DSS sensor board is incremented when energized to monitor the total number of days in acve eld service.
If the sensor is disconnected the incremenng of the me in service will stop. When the sensor is energized
the incremenng of me in service will once again resume. The number of days in service is always the actual
real-me total usage. The total days in use is shown in days and equally accurate for connuous or intermient
service such that the me in service is accurate even if the sensor is taken in and out of use for cleaning and
re-calibraon and/or swapped between dierent installaons. The total me in service since each calibraon
was performed is shown when the ‘View’ key is pressed for 3 to 5 seconds in the given calibraon LED mode.

Important note before performing calibraons
The me averaging dampener is always on even when performing calibraons. It can be desirable to adjust
dampener to a short value when performing calibraons to make the calibraon process quicker and then reset
the dampener back to a higher value before reinstalling the sensor back into connuous use in eld service
(be sure to remember this last step!)
Temperature Calibraon Instrucons
The temperature is calibrated by pushing the ‘Up’ or ‘Down’ buons when in the temperature display (°C) mode. *
Calibraon of Dissolve Oxygen DSS Sensors with Handheld Field Communicator
1. Use the ‘Mode’ buon to toggle to ‘Cal Slope’ LED.
2. Use ‘Up’ and ‘Down’ keys unl the display reads ‘CaL’.
3. Press the ‘Mode’ buon to invoke the automac calibraon roune. The display will show ‘YES’. Press the
‘Mode’ buon to complete the auto-calibraon. The HFC will return back to the ‘Reading’ LED mode upon
compleon.
a. The Dissolved Oxygen DSS Modbus RTU sensor automacally computes the dissolved oxygen ppm
levels that constutes the 100% saturaon condion to perform a dry in air calibraon including the
measured current temperature as well as the user entered barometric pressure.
b. When you complete the autocalibraon the percent saturaon reading will be 100 percent (%) saturaon.
The percent saturaon units can be seen by pressing the ‘View’ key in the ‘Reading’ LED mode.
4. All calibraon values are stored inside the Dissolved Oxygen DSS sensor in EEPROM such that sensor can
be powered down or moved without loss of calibraon values resulng in a true plug and play measurement
system with seamless hot-swap of sensor in eld.
5. Results of the performed DO-DSS sensor calibraons can be viewed by pressing the ‘View’ key in each
calibraon LED mode. The units are mV per Dissolved Oxygen ppm for the slope value used throughout the
enrety of the range.
*Negave values shown as ashing.

Calibraon Instrucons for Dissolved Oxygen DSS Sensors
Turtle Tough’s Dissolved Oxygen DSS sensors do not require oset calibraon as they have a true 0mV potenal
at 0ppm dissolved oxygen. Accordingly, the ‘Oset’ LED is skipped on the Handheld Field Communicator when
toggling through the LED modes. Calibraon of the dissolved oxygen sensor is then only performed in the
‘Slope’ LED mode. Remove the dissolved oxygen sensor from service and allow sucient me for the temperature
and sensor reading to be stable to ensure a good gain calibraon result. Perform the slope calibraon (a.k.a.
gain) when the sensor is clean and dry and exposed only to air. If the relave humidity is not 100%, suspend the
sensor in the air over a source of water for best results. Be sure to wait a sucient me unl the sensor reading
is stable before performing the span (slope) calibraon. The days in use since the calibraon was last performed
is saved in the sensor EEPROM and accessible via RS-485 Modbus RTU.
IMPORTANT: Perform precise temperature calibraon before slope (a.k.a. gain) calibraon on DO sensor.
The correct 100% saturaon value at any temperature from -5 to +50°C is programmed in the DO-DSS sensor
board soware allowing for a fully automated calibraon procedure.
Temperature Calibraon Instrucons
The temperature is calibrated by pushing the ‘Up’ or ‘Down’ buons when in the temperature display (°C) mode. *
* Negave values shown as ashing.
Important note for powering DSS Sensors
• The RS-485 Modbus RTU digital output from the DSS smart digital sensors is non-isolated.
o Power source that energizes sensor should be isolated (dedicated and separate from all other devices) or
o DC/DC converter/isolator can be added to the exisng power supply employed to accomplish the same
net result as having a dedicated separate 9V to 12V DC power source.
o Device that will serve as RS-485 MODBUS RTU master should have an isolated serial input port

Automac Calculaon of Theorecal 100% Dissolved Oxygen Saturaon
at any Temperature & Pressure for Accurate Calibraon & Measurement
The DO-DSS sensor has preprogrammed the correct 100% dissolved oxygen saturaon levels valid at any
temperature and pressure. This is important for two main purposes: 1) to ensure accurate calibraon of the
sensor which is performed dry in air and 2) when the percent (%) saturaon is displayed and output for purposes
of monitoring and control. The graph below demonstrate the impact of both temperature and pressure on the
dissolved oxygen (DO) ppm levels that constute 100% saturaon condion.
For the calibraon funcon, either the eld condion should be 100% relave humidity for best accuracy or
else the sensor should be suspended dry in air but over a water source to simulate locally the 100% relave
humidity condion. The water molecule in air (humidity) is then saturated with oxygen in manner that can be fully
described by the ambient temperature and pressure as shown above. When placed into service, the galvanic DO
sensor will measure the ppm levels at the installaon depth. To convert this measured ppm value into percent (%)
saturaon the DO-DSS sensor uses the internally stored curve visualizaon below.

Automac Calculaon of Theorecal 100% Dissolved Oxygen Saturaon
at any Temperature & Pressure for Accurate Calibraon & Measurement
The DO-DSS has preprogrammed the correct 100% dissolved oxygen saturaon levels valid at not only any
temperature and pressure but also corrected for salinity. This is important for applicaons, where not only
freshwater will be present, but also for brackish and saltwater sources in variable amounts. The graph below
demonstrates the impact of salinity on the dissolved oxygen (DO) ppm levels that constute 100% saturaon
condion at the nominal 760mm pressure condion. For simplicity of visualizaon, just one set of curves is
shown although the analyser can perform this compensaon any temperature, pressure or salinity.
This salinity correcon is only required as a correcon to the computaon of the % saturaon from the
measured DO ppm levels for the inline measurement. Since the calibraon is done dry in the air, salinity
correcon is not required for this part of the operaon. Since the impact of salinity is considerable as shown
in the graph above, it must be corrected carefully at any level of salinity and temperature. The salinity value
in standard PSU (PPT) units can be entered into the dissolved oxygen sensor to perform this correcon. The
value of the salinity can be determined by a handheld meter or else monitoring connuously using a conducvity
transmier from which one can readily convert into common salinity units.

Notes on adjustable smoothing dampener and output delay
• Dampener LED when Dissolved Oxygen DSS sensor is connected allows for display and modicaon of the
variable that is used to set the number of seconds used for the smoothing dampener and delay from boot
to send the output values
• For intermient operaon, it is recommended to set this dampener and output delay variable to a low number
in order to minimize power consumpon while from baery power sources and maximize sampling me of
process output
Display features available using the ‘VIEW’ key
• In ‘Node’ mode press ‘View’ key to invoke the node scanning feature *
• In ‘Baud Rate’ mode press ‘View’ key for 3 to 5 seconds to invoke node changing mode. **
• In ‘Sensor Type’ LED mode, the soware revision for the connected sensor is shown when the ‘View’ key is pressed.
• In ‘Reading’ LED mode, the displayed units are ppm. Pressing ‘View’ key returns reading in percent (%) saturaon
units. The computaon of the percent (%) saturaon units is preprogrammed in Dissolved oxygen DSS
sensors at the factory including correcons for the air pressure, temperature and salinity.
o To view the percent (%) saturaon excluding the salinity hold ‘View’ key for 3 to 5 seconds. This percent
(%) saturaon excluding the salinity is the one to be used when sensor is dry in air during calibraon procedure.
• In ‘Cal Temp.’ LED mode, the oset in °C * for current temp calibraon is shown when the ‘View’ key is pressed.
• In ‘Cal Slope’ LED mode, the current slope for the connected DSS dissolved oxygen sensor is shown when the
‘View’ key is pressed in mV per DO ppm units.
o Days since this calibraon performed shown by holding ‘View’ in the ‘Cal’ mode for 3 to 5 seconds.
If both ‘View’ & ‘Up’ pressed for 3 to 5 seconds in any ‘Cal’ mode will reset all calibraons back to factory
default
* Node scanning feature, instrucons:
• Inial node of ‘0’ will be shown (press ‘Mode’ when node address is ‘0’ to exit scan mode).
• Select starng address for scan with ‘Up’ or ‘Down’ keys. Node address scrolled in increments of 10. For example, pressing ‘Up’ key gives
address of 1102030… and so forth while pressing ‘Down’ key gives addresses of240230220… and so forth.
• Press ‘Mode’ to begin scan. Scanning is always performed in an ascending fashion. Scan will stop when sensor is found. Sensor type for node
address found displayed ashing with node address.
• Press ‘Mode’ to select this node and you will enter ‘Reading’ mode.
• Press ‘View’ to connue scanning. If no sensors found when address 247 is reached, then ‘Err’ is displayed. Press ‘Mode’ to resume scan and
repeat these procedures.
* * Node changing mode notes:
• Select the current node for the aached sensor. If the current node is not known use the node scanning feature to determine it. When in the
‘Baud Rate’ LED mode, hold the ‘View’ key for at least 3 seconds to iniate the node changing mode. The current node of the sensor will be
shown and the ‘Sensor Type’ and ‘Node’ LED will ash.
• If the ‘Mode’ key is pressed immediately aer entering this node changing mode, then no change to the address will be made since the address
displayed will equal the current node address.
• Use ‘Up’ and ‘Down’ keys to adjust the node to the modied address if desired.
• Press ‘Mode’ key to enter the new node address selected with ‘Up’ & ‘Down keys.

Modbus RTU setup of DSS sensor enables all the funconality detailed below
READ-ONLY Data Core Process Value Descripon READ-ONLY Data Analyc Sensor Value Descripon
Calibrated Process Values
for DO-DSS
Calibrated DO ppm value with range 0.00 to
150.00 sent as 0 to 15,000
Calibrated DO % saturaon with range 0.0
to 1,500.0 % sent as 0 to 15,000
Calibrated temp with the range
-40.0 to +210.0 °C sent as 0 to 2,5000
Connected
Sensor Type
1 - DSS-pH
2 - DSS-ORP Standard Range
3 - DSS-ORP Wide Range
4 - DSS-DO (Dissolved Oxygen)
5 - DSS-ISE (Ion Selecve)
6 - DSS-CON (EC Standard/High)
7 - DSS-CON-L (EC Ultralow)
Raw Process Values
Absolute {Raw} mV value with range
0 to +250 sent as 0 to 25,000
(Aux Column in Windows Datalogger)
Sensor Serial
Number
Sensor Diagnoscs
Unique Serial Number Designaon:
YY.M-AA.DDD **
Sensor Item Number
Soware Revision
Max Temp in Use
Min Temp in Use
Days in Field Use
NOTE: Dissolved oxygen values are always
automacally temperature compensated. Calibraon Values
Temperature Oset
Days since Temp Oset Cal
mV per DO ppm Slope
Days since DO Cell Calibraon
** Serial format YY is last digits of year; M is month with A=Oct, B= Nov & C=Dec; AA is leer(s) from A to nY (as permissible); DDD is value from 0 to 255
READ/WRITE Type Adjustable Calibraon Descripon READ/WRITE Type Adjustable Parameter Descripon
Oset Adjust
Temperature
Calibrated Temperature Value
Limit ±25.0 °C * from raw value
The temperature to which reading is adjusted
is sent as 0 to 2,500 corresponding to
-40.0 to +210.0 °C
Reset Calibraons
Will reset all user adjustable sensor
calibraons back to factory default
values
Adjust Dissolved Oxygen
(DO) Galvanic Cell
Working Slope
0.70 to 6.00 mV per DO ppm
Max 250mV input yielding max of 150.00
ppm | 1,500.0 % Saturaon
NOTE: Actual max measured DO ppm is
limited by slope of sensor. For example, if
calibrated slope is 2.00 mV per DO ppm, then
the actual max will be 125ppm instead of
150ppm.
Dampener
Time averaging of process value
1, 2, 3, 4, 5, 8, 10, 15, 20 or 30 Sec
Note: Dampener seng is the basis
for the seng for the autocalibraon
in the factory supplied DSS controller.
The minimum me to complete the
autobuer calibraon is always at least
the dampener sengs plus 2 seconds.
This should be considered if fast auto-
buer calibraon is important.
Reset Calibraons Resets all user adjustable sensor calibraons
back to factory default Delay from Boot
Time unl process values are sent
from boot
1, 2, 3, 4, 5, 8, 10, 15, 20 or 30 Sec
* Negave values shown as ashing.
Note 0: The maximum supported temperature for all fully submersed dissolved oxygen DSS sensors is 65°C absolute.
Note 1: For communicaon to be successful all Modbus devices on the network must use the same baudrate and have a unique node
address assigned. The handheld communicator is a Modbus master whereas all DSS sensors are Modbus slaves. In order for the
handheld communicator to be interfaced with the DSS sensor, that sensor must either be removed from the network, or else
bypassed by means of a suitable bridge box scheme. It is also possible to access any given DSS sensor on the Modbus network if the
exisng Modbus master is disconnected or powered down. If the node of the DSS sensor to be interfaced is not known, please use the
Windows soware to determine the current node address and modify if it should be necessary to ensure a valid and unique node address
seng on the network. Please see DSS installaon guide and DSS controller manual for addional recommendaons and details about
commissioning, calibraon and troubleshoong.
Note 2: Access to READ values in Core Process Value Column gained through Modbus funcon code (04).
Note 3: Access to READ parameters in the Analyc Sensor Value Column, Adjustable Calibraon Column and Adjustable Parameters Column
gained through Modbus funcon code (03).
Note 4: Access to WRITE parameters in the Analyc Sensor Value Column, Adjustable Calibraon Column and Adjustable Parameters Column
gained through Modbus funcon code (16).
This manual suits for next models
1
Table of contents
Other Turtle Tough Measuring Instrument manuals