Turtle Tough TT-MA-DSS User manual

TURTLE
TOUGH
®
2
Content
Safety Precauons 3
Introducon 6
System Contents 7
Main Features 8
Technical Specicaons 9
Benets of using mang Smart Digital DSS RS-485 Modbus RTU Sensors 10
Field Commisioning of Transmier 11
Wiring Schemac 11
Menu Navigaon 12
Funcon and Programming 13
Parameters Explained 14
Display Features 15
Warranty 17
Return Goods 19
Support 20
Scaling Charts
pH 21
ORP 22
Wide ORP 23
DO ppm 24
DO % Sat 25
ISE 26
Cond in Micro Siemen units 27
Cond in PSU or MegOhm Standard 28
Cond in TDS or MegaOhm UPW 29
Frequently Asked Quesons 30

TURTLE
TOUGH
®
3
Safety Precautions
Before aempng to unpack, set up, or operate this instrument, please read this enre manual.
• Make certain the unit is disconnected from the power source before aempng to service or remove any
component.
• Make certain the unit is disconnected from other sources of force or pressure (for example, pneumac or
hydraulic), before aempng to service or remove any component.
• Failure to follow these precauons could result in personal injury and damage to the equipment.
General
This manual contains basic informaon to be noted during installaon, operaon and maintenance. It is therefore
essenal that this manual be read by the contractor before installing and commissioning the system, as well as by
the relevant operang personnel/owner of the unit. It must remain available for reference at all mes. In addion
to the general safety instrucons under this main heading Safety Precauons, the special safety precauons
outlined in other secons must also be observed.
Warnings used in this manual
This manual contains vital informaon relang to the safety of people and the environment, the analyser and any
equipment aached. These statements are idened by the following symbols:
Danger
DANGER
Refers to an imminent danger.
Non-compliance can lead to death or extremely serious injury.
Warning
WARNING
Refers to a potenal hazardous situaon.
Non-compliance can lead to death or extremely serious injury.
Cauon
CAUTION
Refers to a potenal hazardous situaon.
Non-compliance can lead to minor injury or property damage.
Noce
NOTICE
Appears in conjuncon with safety instrucons which may endanger the analyser and its
operaon if disregarded.
Important
IMPORTANT
Draws aenon to supplementary informaon to make the work easier and ensure
trouble-free operaon. Markings which are axed directly to the equipment mut be
observed without fail, and must remain full legible at all mes.
Disconnect electrical supply before working on this equipment.

TURTLE
TOUGH
®
4
Qualification and Training of Personnel
The personnel employed for installaon, operaon, inspecon, and maintenance, must be suitably qualied for
this work. The areas of responsibility, competence and supervision of the personnel must be precisely dened
by the owner. Personnel who do not have the required know-how must be trained and instructed. In addion,
the owner of the system must ensure that the relevant personnel are fully familiar with and have understood the
contents of this manual. Should training be required please contact your Turtle Tough distributor
.
Important Safety Instructions
When installing and using this electrical equipment, basic safety precauons should always be observed, including
the following:
NOTICE
Read and follow all instrucons. Save these instrucons.
WARNING
To reduce the risk of injury, do not permit children to use this product.
WARNING
Risk of Electric Shock. Connect only to a suitable isolated, hard wired electrical outlet. It is
recommended that the outlet is protected by an RCD (Residual Current Detector) or is in any
event in compliance with all local electrical regulaons. Do not bury electrical supply cable.
WARNING
To reduce the risk of electric shock, replace damaged electrical cable immediately.
WARNING
To reduce the risk of electric shock, do not use an extension cable to connect the unit to an
electric supply; provide a properly located outlet
Hazards due to Non-compliance
Failure to comply with the safety instrucons may endanger not only people, but also the environment and the system.
The following hazards in parcular may arise:
• Failure of major system funcons.
• Failure of specied methods for maintenance and repair.
• Danger to people due to electrical, mechanical and chemical eects.
Safe Operation
The safety instrucons contained in this manual must be observed.
The owner is responsible for ensuring compliance with local safety regulaons.
Safety Instructions for the Owner/Operator
Danger due to electric current must be excluded. Refer to local electrical safety standards and regulaons.
Danger
Warning
Electric Shock
Hazard
Electric Shock
Hazard
Danger
Electric Shock
Hazard

TURTLE
TOUGH
®
5
Installation, Maintenance and Inspection
The owner must ensure that all maintenance, inspecon and installaon work is undertaken by
authorised and duly qualied personnel who have also studied this manual.
• Sensors must always by isolated before starng any work.
• Please be aware that the connected sensor(s) and the associated analyser may be controlling
chemical dosing and as such shung down the unit without due regard to the systems it is
controlling can lead to chemical release.
Impermissible Modes of Operation
Usage other than as described in this manual will lead to the immediate cancellaon of the warranty
and any other manufacturer’s liability.
Unauthorised Modification
Usage other than as described in this manual will lead to the immediate cancellaon of the warranty
and any other manufacturer’s liability.
Chemicals
Should the sensor(s) be in contact with hazardous chemicals, great care must be taken when handling
them.
CAUTION
When handling this equipment, the accident prevenon regulaons applicable on site must be
observed and the specied personal protecve equipment worn.
PPE: examples of protecve clothing, gloves and goggles.
DANGER Fire hazard. No parts are suitable for use in a hazardous rated area.
IMPORTANT
Please unpack the equipment and ordered accessories carefully in order not to miss small parts.
Immediately compare the scope of delivery to the delivery note. If there are any discrepancies,
contact your Turtle Tough Representave.
Sensors
Health and Safety
Before making or breaking any electrical or signal connecons, ensure that the instrument is isolated
from the electrical supply. When handling the sensor please wear the appropriate PPE.
Environmental Consideraons
Please use this product in a manner sensive to the environment and at the end of its life dispose
or recycle in a manner appropriate at that me that is in compliance with local regulaons.
Disconnect
electrical supply
before working
on this equipment
Appropiate
PPE to be
worn when
in contact
with the
sensor
Take care
when
removing
the sensor
Important
Danger
Cauon
Warning

TURTLE
TOUGH
®
6
Introduction
Congratulaons on purchasing a Turtle Tough TT-MA-DSS Display Module. Turtle Tough sensors are handmade
with care and precision, combining state-of-the-art technology and materials to deliver superior performance
under extreme condions. The TT-MA-DSS Display Module has been specically developed to achieve the
opmum performance from your Turtle Tough Digital Sensor.
This instrucon manual provides informaon for the correct installaon and use of a Turtle Tough TT-MA-DSS
Display Module to ensure you get the maximum life and performance.
Through our experience in the world’s toughest applicaons, we have developed systems that have been
opmised for high hydrouoric acid, saturated sodium, high sulphide, organic solvents, high temperature and
slurry/viscous applicaons. Our applicaon-specic sensors include industrial, wastewater, acid etching, Nickel,
Gold, Titanium Dioxide, Food and Beverage and Pharmaceucal.
Please ensure that the system you are using is suitable for your intended applicaon. You can locate your specic
sensor in the sensor selecon matrix, which can be found under support aurtletoughsensors.com

TURTLE
TOUGH
®
7
System Contents
All Turtle Tough instruments undergo thorough quality control and calibraon before delivery and always
include a detailed instrucon manual.
Your TT-MAD System includes the following items:
110-240V AC Power Supply
pH/ORP Analyser Module
DSS Analyser Module
Dissolved Oxygen Analyser Module
Conducvity Analyser Module
Ion Selecve Analyser Module
Datalogger Module
Temperature Module
TOT pH Compensaon Module
Relay Module
IP65 Enclosure
pH/ORP Digital Sensor
Dissolved Oxygen Sensor
Conducvity Sensor
Ion Selecve Sensor
User’s manual

TURTLE
TOUGH
®
8
Main Features
The TT-MA-DSS supports all of the common liquid analycal electrochemical measurements parameters of pH,
ORP, dissolved oxygen (D.O.), ion-selecve (ISE) and conducvity. Scanning feature nds node address of sensor
with the parameters automacally loaded for measurement type allowing for a single universal transmier to be
stocked for any potenal electrochemical measurement of interest at facility thereby reducing both inventory and
training costs.
Complementary modules for
TT-MA-REL: Alarm & Relay controller with simple supervision, On/O or Time Proporonal Control (TPC) Modes
TT-MA-TOT: Computes pH compensated “Total ISE” from Free ISE & pH analogue inputs, 0/4-20mA analogue &
Modbus output
RTU Sensor Input
TT-MA-DSS interfaces smart digital DSS MODBUS RTU sensors via 100% digital communicaons for signal
integrity even in noisy areas. Integral smart sensor board provides temperature compensaon & stores calibraons.
Total cable lengths to 300 meters for conducvity & 1,000 meters for pH, ORP, ISE & DO sensors with waterproof
snap connector.
TT-MA-DSS provides scalable, proporonal reversible 4-20mA or 0-20mA analogue current loop output for any
mang connected sensor input. Minimum scaling down to 2% of the full range input of sensor allowing for a very
high-resoluon signal to be sent to the mang analogue input device. Analogue output is galvanically isolated
from input with a 3KV rated optocoupler.

TURTLE
TOUGH
®
9
DSS ANALYSER MODULE
for pH, ORP, DO, ISE & CONDUCTIVITY
Specifications
Mechanical
Housing Lexan UL94V-0 (Upper part)
Noryl UL94V-0 (Lower part)
Mounng M36 for 35 mm DIN rail
IP Class Housing IP40. Connector IP20
Connector Max 16A. Max 2.5 mm2
Max torque 0,6 Nm
Temperature Usage -15 to +50°C (Storage -35 to +75°C)
Weight 75 grams (2.64 ounces)
Dimensions D 58 x W 36 x H 86 mm (2.3” X 1.4” X 3.4”)
Ce Mark EN61326A
Electrical
Power supply 24VDC ±10%
Power Consumpon 60mA max when pH/ORP/ISE/DO
80mA max when conducvity (EC)
Sensor Input Smart Digital DSS MODBUS RTU
Temperature Sensor Integral Planum TC Element
Temperature Range -40 to +210°C ± 0.3°C
Temperature Compensaon Automac Temperature Compensaon (ATC)
Analogue Output 0-20mA or 4-20mA, max. 500Ω
Output Hold Automac if sensor is not connected
DSS

TURTLE
TOUGH
®
10
Benefits of using mating Smart Digital
DSS RS-485 Modbus RTU Sensors
• Integral RS-485 MODBUS RTU interfaces all-modern PLC controllers & data acquision systems.
• Communicator provides easy management of eld installaons without the cost of a mang transmier.
This is ideal for locaons where a local display is not necessary or possible due to installaon limitaons.
• Intelligent management of sensor calibraons and service life-cycle for ecient commissioning & maintenance.
All aspects of the installaon are completely portable from the shop to the eld site locaon.
• Days in use-value is stamped for calibraons that are performed. This allows for predicve scheduling of
maintenance in the PLC to ensure accurate measurement in the eld based upon user-dened criteria.
• All digital sensors ensure reliable operaon even in noisy process environments, unlike analogue sensors.
• No degradaon in digital communicaons even with very long cable runs. Max 1,000 meters (3,280 feet) for
pH, ORP, ISE & DO sensors and Max 300 meters (1,000 feet) for conducvity sensors with TT-MA-DSS.
• Bridging connecons & modifying installaons easily without loss of signal quality with NEMA 6P & IP67 rated
quick connect waterproof and corrosion-resistant dual quick connect plugs. Simple plug and play operaon for
intelligent maintenance planning & smart management of sensor installaons and stocking.
• Quick connect digital extension cables facilitate consolidaon of very many DSS sensors outputs into a one-pan-
el enclosure where very many remote eld installaons can all be conveniently viewed at once.
• Intelligent DSS handheld eld communicator soware idenes the type of sensor connected & autoloads
correct features. There exists no possibility of accidentally using the wrong set of opons or sengs.
• All Extension cables for DSS sensors are inter-compable. Uniform extension cables minimize stocking.
Separate eld installaon guide details available opons to commission & exchange sensors.
RTU
• The enre line of proven inline, immersion, submersible, twist lock, sanitary, HOT-TAP retractable pH & ORP,
Conducvity, Galvanic D.O & ISE sensors made by TURTLE TOUGH are ALL available for use with TT-MA-DSS
universal smart transmier
• The very rugged Ultra Tough pH glass element opmized for use in slurries & high viscosity applicaons is
ONLY available for the smart digital type sensors
• The extreme dehydraon resistant style reference technology that allows for prolonged exposure to dry
condions and intermient wet & dry use is ONLY available for the smart digital type sensors
• Waterproong Opons is recommended for any DSS smart digital sensor with integral RS-485 MODBUS RTU
digital output for immersion or fully submersible installaons.

TURTLE
TOUGH
®
11
Field Commissioning of Transmitter
The TT-MA-DSS universal smart transmier is typically supplied pre-congured with a female snap to nned leads
panel mount connector installed onto a suitable eld ready enclosure assembly. The quick-connect plug of the smart
digital MODBUS RTU sensor is interfaced with a female quick connect plug TT-MA-DSS universal smart transmier
assembly. DSS MODBUS RTU sensors are precalibrated ready for immediate plug and playeld use.
Wiring Schematic
Please consult the Turtle Tough website turtletoughsensors.com for addional wiring informaon.
4 | Page
Turle Tough
Unit 12 / 634-644 Mitcham Road,
Vermont, VIC 3133
Australia
+61 (3) 9872 5055 | http://www.turtletoughsensors.com
FFiieellddCCoommmmiissssiioonniinnggooffTTrraannssmmiitttteerr
WWiirriinnggSScchheemmaattiicc
The TT-MA-DSS universal smart transmitter is
typically supplied preconfigured with female snap
to tinned leads tinned leads panel mount connector
installed onto a suitable field ready enclosure
assembly. The quick connect plug of the smart
digital MODBUS RTU sensor is interfaced with
female quick connect plug TT-MA-DSS universal
smart transmitter assembly. DSS MODBUS RTU
sensors are precalibrated ready for immediate plug
& play field use.
IIffssooffttwwaarreelloocckk((SSeettuuppppaarraammeetteerrPP0011))iiss““OOnn””aallllooffppaarraammeetteerrssccaannoonnllyybbeerreeaadd..SSeettPP0011SSooffttwwaarreeLLoocckkttoo““OOffff““ttoocchhaannggeevvaalluueess..
TThheePP0011ssooffttwwaarreelloocckkwwiillllaauuttoommaattiiccaallllyyrreesseettbbaacckkttoo““OOnn””iiffnnookkeeyyiisspprreesssseeddffoorrsseevveerraallmmiinnuutteess..
UUsseerrSSeettuuppPPaarraammeetteerrss
NNoo
PPaarraammeetteerr
DDeessccrriippttiioonn
RRaannggee
DDeeffaauulltt
P01
Lock
Software Lock
On / Off
On
P02
Address
Address on MODbus
Off, 1…247
Per Order
P03
Baudrate
MODbus baudrate
9,600 / 19,200
Per Order
P04
Analogue Output
Type
Toggle for Current Loop Type
4-20mA, 0-20mA
Per Order
P05
Analogue Output
Mode
Select Polarity of Analogue Output
noninverted, inverted
Per Order
P06
0/4 mA Whole
Scale Low setpoint for output – Whole Percent
0% to 98%
Per Order
P07
0/4 mA Dec.
Scale Low setpoint for output – Decimal Point 0-97.XX
XX.00% - XX.99%
Per Order
P08
20 mA Set
Scale High setpoint for output – Whole Percent
2% to 100%
Per Order
P09
20 mA Set
Scale High setpoint for output – Decimal Point 2-99.XX
XX.00% - XX.99%
Per Order
P10
D.O. Units Selected
Select between ppm and % Saturation units for output
ppm or % Sat
Per Order
P11
Conductivity Units
Selected for Output
If Conductivity Sensor Type = 6 (Standard/High) then
choices are uS/cm, Salinity (PSU) or TDS
If Conductivity Sensor Type = 7 (Ultralow) then choices
are uS/cm, MΩStandard or MΩfor UPW
For Sensor Type = 6
uS/cm, PSU, TDS or For
Sensor Type = 7 uS/cm,
MΩ, UPW
Per Order
P12
Conductivity Sensor
Cell Constant (K)
Indicates nominal cell constant for connected sensor
From K=0.01/cm to K=20.00/cm
0.01 to 20.0
Per EC Sensor
P13
Conductivity Sensor
Range Mode
Indicates the range mode scaling factor for EC sensor
“UL”=2, “Std”=200, “Hi”=2,000
2..2,000
Per EC Sensor
P14
0/4mA Offset
Trim Low
±9.99% *
Per Factory Cal
P15
20mA Gain (Span)
Trim High
±9.99% *
Per Factory Cal
P16
Sampling Rate
Set sampling frequency in seconds
0.5, 1.0, 2.0 and 4.0
Per Order
P17
Recalibrate Notify
Set max time since cal last peformed before notification
1 to 999 Days
Per Order
P18
Display Sensor Type
1=pH, 2=ORP, 3=Wide ORP, 4=DO, 5=ISE,
6=Cond Standard/High Style, 7=Cond Ultralow
Style
1..7
Per Sensor
P19
Formula Weight
Formula Weight of Measured Ion – Only for ISE sensors
6.94..655.35
Per ISE Sensor
P20
Type of TDS Units
Type of TDS units which are sent from EC Sensor
0=NaCl, 1=442, 2=KCl
Per EC Sensor
P21
Back to Default
Reset to Default
Def=Reset, Par=NoReset
Par
* Negative values will be shown as flashing. Shaded portions of chart above indicate display only parameters.
PPaarr..nnoo..22set node address of sensor. IIffnnoosseennssoorriissffoouunnddaatttthheeccuurrrreennttnnooddeesseettttiinnggtthheenn‘‘SSEEnn’’wwiillllffllaasshhoonnssccrreeeenn..PPrreessssaannyy
bbuuttttoonnttooeenntteerrPP0022nnooddeesseelleeccttmmooddee..UUssee‘‘UUpp’’&&‘‘DDoowwnn’’kkeeyyssttoosseelleeccttbbeettwweeeenn‘‘SSeett’’ttoommaannuuaallllyyddeeffiinneennooddeeaaddddrreessssoorr‘‘SSCCnn’’ffoorr
aauuttoommaattiiccnnooddeessccaannnniinnggffeeaattuurree((uussee‘‘MMooddee’’ttooeenntteerr‘‘SSEEtt’’oorr‘‘SSCCnn’’ffeeaattuurree))..WWhheennnnooddeeiissffoouunndddduurriinnggssccaanntthheesseennssoorrttyyppee&&nnooddee
aaddddrreessssaarreettoogggglleedd..AAcccceepptttthheennooddeeaaddddrreessss&&sseennssoorrttyyppeeffoouunnddwwiitthh‘‘MMooddee’’kkeeyyoorrpprreessss‘‘UUpp’’oorr‘‘DDoowwnn’’ttooccoonnttiinnuueesseeaarrcchh
PPaarr..nnoo..33sets baudrate to be used.Choices are 9,600 or
19,200.
PPaarr..nnoo..44
select whether output type is 0-20mA or 4-20mA.
PPaarr..nnoo..55
select whether output is inverted or non-inverted type.
PPaarr..nnoo..66,,77,,88&&99
define 0/4mA and 20mA setpoints. Appendix
provides percentages corresponding with specific engineered
units for various sensor. Min scaling between low/high setpoints
2% full range. Excel worksheet to compute % setpoints available.
PPaarr..nnoo..1100
selects ppm or % saturations units to be used as basis
for output & main LED display for connected D.O. sensor.
PPaarr..nnoo..1111
selects measured conductivity or else computed PSU,
TDS or MΩunits as basis for analogue output & main LED
PPaarr..nnoo..1122&&1133displays cell constant & range mode of EC
sensor.
PPaarr..nnoo..1144&&1155
Trim offset for 4mA and Trim span for 20mA
PPaarr..nnoo..1166
define sampling rate for connected sensor in seconds
PPaarr..nnoo..1177
Number of days after which recalibration notification
is displayed when sensor is initially connected to transmitter. If
limit is exceeded then ‘CAL’ ‘OLd’ will be dispayed as warning.
PPaarr..nnoo..1188
Display the sensor type which is connected
PPaarr..nnoo..1199
Display the formula weight of the measured ion for
ISE Sensor. For anion selective sensor value is shown as flashing
PPaarr..nnoo..2200
Display type of TDS units which are sent by EC
sensor

TURTLE
TOUGH
®
12
TT-MA-DSS has 3 digit display and 6 LEDs to setup and display values. ‘Mode’ is used to navigate. Programming
done by 3 keys. ‘Mode’ toggles & ‘Up’ or ‘Down’ used to scroll and select. Setup Parameters entered via ‘Mode’ .
Values changed using ‘Up’ or ‘Down’. The TT-MA-DSS automacally selects and illuminates LED based upon the
type of sensor which is connected.
If sowarelock (P01) “On” no changes can be made. Set P01 to “O “ to allow for changes to scaling and
conguraon. If keys are not used for several minutes then soware lock resets back “On”.
Lock Funcon (IMPORTANT)
Please note: Parameter No. 01 is a LOCK funcon which MUST be turned OFF to change ANY parameter.
The default for the lock is ON. Once the lock is OFF, if there is a period of inacvity for 60 seconds the lock
then defaults back to ON.
Important
DSS
Mode key
to navigate.
Press to select
'Setup' Up and Down keys are used to
scroll through the parameters
The led Com is illuminated
when MODbus is acve

TURTLE
TOUGH
®
13
Funcon and Programming
No Parameter Descripon Range Default
01 Lock Soware Lock On / O On
02 Address Address on MODbus O, 1…247 Per Order
03 Baudrate MODbus baudrate 9,600 / 19,200 Per Order
04 Analogue Output Type Toggle for Current Loop Type 4-20mA, 0-20mA Per Order
05 Analogue Output
Mode Select Polarity of Analogue Output noninverted,
inverted Per Order
06 0/4 mA Whole Scale Low setpoint for output – Whole
Percent 0% to 98% Per Order
07 0/4 mA Dec. Scale Low setpoint for output –
Decimal Point 0-97.XX XX.00% - XX.99% Per Order
08 20 mA Set Scale High setpoint for
output – Whole Percent 2% to 100% Per Order
09 20 mA Set Scale High setpoint for output – Decimal
Point 2-99.XX XX.00% - XX.99% Per Order
10 D.O. Units Selected Select between ppm and % Saturaon
units for output ppm or % Sat Per Order
11 Conducvity Units
Selected for Output
If Conducvity Sensor Type = 6 (Standard/
High) then choices are uS/cm, Salinity
(PSU) or TDS If Conducvity Sensor Type
= 7 (Ultralow) then choices are uS/cm, MΩ
Standard or MΩ for UPW
For Sensor Type =
6 uS/cm, PSU, TDS
or For Sensor Type
= 7 uS/cm, MΩ,
UPW
Per Order
12 Conducvity Sensor
Cell Constant (K)
Indicates nominal cell constant for connected
sensor From K=0.01/cm to =20.00/cm 0.01 to 20.0 Per EC Sensor
13 Conducvity Sensor
Range Mode
Indicates the range mode scaling factor for
EC sensor “UL”=2, “Std”=200, “Hi”=2,000 2..2,000 Per EC Sensor
14 0/4mA Oset Trim Low ±9.99% * Per Factory Cal
15 20mA Gain (Span) Trim High ±9.99% * Per Factory Cal
16 Sampling Rate Set sampling frequency in seconds 0.5, 1.0, 2.0 & 4.0 Per Order
17 Recalibrate Nofy Set max me since cal last peformed be-
fore nocaon 1 to 999 Days Per Order
18 Display Sensor Type
1=pH, 2=ORP, 3=Wide ORP, 4=DO,
5=ISE, 6=Cond Standard/High Style,
7=Cond Ultralow Style
1..7 Per Sensor
19 Formula Weight Formula Weight of Measured Ion –
Only for ISE sensors 6.94..655.35 Per ISE
Sensor
20 Type of TDS Units Type of TDS units which are sent from EC
Sensor
0=NaCl, 1=442,
2=KCl Per EC Sensor
21 Back to Default Reset to Default Def = Reset,
Par = No Reset Par
*Negave values will be shown as ashing. Shaded porons of chart above indicate display only parameters.

TURTLE
TOUGH
®
14
Parameters Explained
To access the programmable parameters see Menu Navigaon.
The following is a descripon of each parameter sengs.
Parameter Descripon
P02
Par. no. 2 set node address of sensor. If no sensor is found at the current node seng then ‘SEn’
will ash on screen. Press any buon to enter P02 node select mode. Use ‘Up’ & ‘Down’ keys
to select between ‘Set’ to manually dene node address or ‘SCn’ for automac node scanning
feature (use ‘Mode’ to enter ‘SEt’ or ‘SCn’ feature). When node is found during scan the sensor
type & node address are toggled. Accept the node address & sensor type found with ‘Mode’ key
or press ‘Up’ or ‘Down’ to connue search
P03 Sets baudrate to be used. Choices are 9,600 or 19,200
P04 Select whether output type is 0-20mA or 4-20mA.
P05 Select whether output is inverted or non-inverted type.
P06
P07
P08
P09
Dene 0/4mA and 20mA setpoints. Appendix provides percentages corresponding with specic
engineered units for various sensor. Min scaling between low/high setpoints 2% full range. Excel
worksheet to compute % setpoints available.
P10 Selects ppm or % saturaons units to be used as basis for output & main LED display for
connected D.O. sensor.
P11 Selects measured conducvity or else computed PSU, TDS or MΩ units as basis for analogue
output & main LED
P12
P13 Displays cell constant & range mode of EC sensor.
P14
P15 Trim oset for 4mA and Trim span for 20mA
P16 Dene sampling rate for connected sensor in seconds
P17 Number of days aer which recalibraon nocaon is displayed when sensor is inially
connected to transmier. If limit is exceeded then ‘CAL’ ‘OLd’ will be dispayed as warning.
P18 Display the sensor type which is connected
P19 Display the formula weight of the measured ion for ISE Sensor. For anion selecve sensor value
is shown as ashing
P20 Display type of TDS units which are sent by EC sensor
P21 Resets ALL parameters back to factory set defaults display

TURTLE
TOUGH
®
15
Display Features
For Sensor Type 1 pH - the “pH / ORP” LED will be connuous illuminated unless otherwise indicated below
• -2.00 to -0.01 displayed as 2.00 to 0.01 ashing
• 0.00 to 9.99 displayed not ashing with two decimal points
• 10.0 to 16.0 display with one decimal point
If the ‘Down’ buon is pressed, then the temperature of connected sensor is shown *
If ‘Down’ buon is held for 3 to 5 seconds, then the absolute mV will be shown *
If the ‘Up’ buon is pressed, then the mA for the current process value and scaling will be shown
For Sensor Type 2 ORP - the “pH / ORP” LED will be connuous illuminated unless otherwise indicated below
• -999 to -1 displayed as 999 to 1 ashing
• 0 to +999 displayed not ashing
If the ‘Down’ buon is pressed, then the temperature of connected sensor is shown *
If ‘Down’ buon is held for 3 to 5 seconds, then the absolute mV will be shown *
If the ‘Up’ buon is pressed, then the mA for the current process value and scaling will be shown
For Sensor Type 3 Wide ORP - the “pH / ORP” LED will be connuous illuminated unless otherwise indicated below
• -2,000 to -1,000 display as 2.00 to 1.00 with LED ashing (units are Volts)
• -999 to -1 displayed as 999 to 1 ashing
• 0 to +999 displayed not ashing
• +1,000 to +2,000 display as 1.00 to 2.00 with LED not ashing (units are Volts)
If the ‘Down’ buon is pressed, then the temperature of connected sensor is shown *
If ‘Down’ buon is held for 3 to 5 seconds, then the absolute mV will be shown *
If the ‘Up’ buon is pressed, then the mA for the current process value and scaling will be shown
For Sensor Type 4 Dissolved Oxygen (D.O.) - the “D.O.” LED will be connuous illuminated unless otherwise
indicated below
• If P10 is ‘ppm’ then 0.00 to 150.00 ppm units displayed not ashing as 0.00 to 9.99, 10.0-99.9 and 100-150 ppm
• If P10 is ‘%Sat’ then 0.0-1,500.0 percent (%) saturaon units displayed not ashing as 0.0-99.9%, 100-999%
with the special range of 1,000-1,500% displayed as 1.00-1.50 with LED ashing (kilo % Saturaon Units)
If the ‘Down’ buon is pressed, then the temperature of connected sensor is shown *
If ‘Down’ buon is held for 3 to 5 seconds, then the absolute mV will be shown *
If the ‘Up’ buon is pressed, then the mA for the current process value and scaling will be shown
If ‘Up’ buon held and P10 is ‘ppm’ (basis of 4-20mA output) then % Saturaon units are displayed
If ‘Up’ buon held and P10 is ‘%Sat’ (basis of 4-20mA output) then ppm units are displayed
For Sensor Type 5 Ion Selecve (ISE) - the “ISE” LED will be connuous illuminated unless otherwise indicated below
• 0.00-9.99, 10.0-99.9, 100-999 ppm units displayed same as per TT-MA-ISE transmier
kilo-ppm units displayed with LED ashing to signify kilo-ppm scale is in use same as per TT-MA-ISE-kilo
• 1.00-9.99 (1,000-9,990 ppm), 10.0-99.9 (10,000-99,900 ppm) and 100-999 (100,000-999,000 ppm)
If the ‘Down’ buon is pressed, then the temperature of connected sensor is shown *
If ‘Down’ buon is held for 3 to 5 seconds, then the absolute mV will be shown *
If the ‘Up’ buon is pressed, then the mA for the current process value and scaling will be shown
If ‘Up’ buon held for 3 to 5 seconds, pION value is shown with same scheme used display the pH
For Sensor Type 6 or 7 Conducvity (EC) - the “Cond” LED will be connuous illuminated unless otherwise indi-
cated below
• <1.00 mS shown as ashing from 1 to 999 uS/cm with 0.01-9.99, 10.0-99.9 and 100-999 oang decimal point
• 1.00 to 999 mS/cm shown displayed not ashing with 0.01-9.99, 10.0-99.9 and 100-999 oang decimal point

TURTLE
TOUGH
®
16
• 1,000 to 2,000 mS/cm display as 1.00 to 2.00 with the LED ashing (kilo-mS/cm)
• If P11 is ‘PSU or MΩ’ then salinity (sensor type 6) or resisvity (sensor type 7) is shown as 0.00-9.99 and 10.0-50.0
• If P11 is ‘TDS’ then ppt is shown as 0.00-9.99 and 10.0-99.9 (mulply by 1,000 to get ppm units instead of ppt units)
If the ‘Down’ buon is pressed, then the temperature of connected sensor is shown *
If ‘Down’ buon is held for 3 to 5 seconds, then raw conducvity will be shown per scheme above
If the ‘Up’ buon is pressed, then the mA for the current process value and scaling will be shown
If ‘Up’ buon is held for 3 to 5 seconds and P11 is ‘PSU’, ‘TDS’ or ‘MΩ’ then reading in conducvity units will be
shown (see scheme above). If conducvity units selected for P11 then nothing is shown.
Producon data (yy.m) displayed by pressing ‘Down’ & ‘Mode’ “Mode” simultaneously in any main LED display
mode. The month will display as 1>9 and then A for October, B for November, and C for December. I.e. October
2011 will display as “11. A”.
Revision of soware is displayed by pressing the ‘Up’ ‘Mode’ simultaneously in any main display mode.
* Negave values will be shown as ashing.

TURTLE
TOUGH
®
17
TT-MA-DSS Analyser
Updated 111220
Warranty
Product Warranty
Every Turtle Tough product is thoroughly inspected and tested before leaving the factory and prior to shipping. In
addion to any statutory rights and remedies you may have, Turtle Tough warrants all its products against defecve
workmanship and faulty materials for 12 months from the date of purchase and undertakes, at its opon, to repair
or replace, free of charge, each product or part thereof on condion that:
• The complete product is returned to Turtle Tough or one of its authorised service agents, in person or freight
pre-paid by you, and found, on examinaon, to be suering from a manufacturing defect.
• The product or relevant part has not been subject to misuse, neglect, or been involved in an accident.
• The repairs are not required as a result of normal wear and tear.
• Damage caused by wear and tear, inadequate maintenance, improper installaon, corrosion, or by the eects
of chemical processes is excluded from this warranty coverage.
Sensor Warranty
Turtle Tough sensors are electrochemical devices and as such have a limited operang life. Life expectancy
depends on the eld of applicaon such as the medium, pressure and temperature. It can vary between a number
of weeks to several years. There are special cases in extreme environments where operang life will only be a few
days. Characterisc and response me will also change with aging. As such electrochemical sensors are arcles
of consumpon and are not subject to a common guarantee. Replacements or exchanges are generally excluded
unless a manufacturing defect is determined to be the cause. It is not possible to predict the rate of deterioraon
for a parcular process, nor can we provide a guarantee on sensor life because it is impossible to predict the
rate of exposure, contaminaon and deterioraon. Damage caused by wear and tear, inadequate maintenance,
faulty installaons, corrosion, or the eects of chemical processes is excluded from this warranty coverage. Our
agents or representaves may provide you with a life expectancy guide based on similar applicaons we have
experienced, however, this in no way constutes a warranty of performance and is a general indicator.
Shelf Life Warranty
The standard shelf life for a Turtle Tough sensor is one year from the date of shipment. Sensors stored longer
than this period may sll be funconal but are no longer under warranty. Sensors should be stored in a cool, dry
locaon with the sensor p (where the pH/ORP element is located) oriented toward the ground. All pH/ ORP/ISE
sensors come standard with a condioning soluon in the cap. This condioning soluon is 50% pH 4 buer and
50% saturated potassium chloride (mixed by volume). The sensor cap should be kept ghtly axed to the sensor
body and sealed with common piping Teon tape when the sensor is not in use. Sensors that are to be returned
for a shelf-life warranty claim must have the original sensor cap and condioning soluon intact to be eligible for
warranty replacement.
Damage to Internal Electronics
Damage to the sensor’s internal electronic components is not covered under warranty. Analogue sensors with
internal preampliers are sensive to electrostac discharge. Sensors with preamps are clearly marked and extra
care must be taken when handling these sensors as human contact with the electrical connecons can discharge
stac to the preamplier causing it to blow. This will render the product inoperable. Addionally, sensors
containing digital RS485 boards are suscepble to damage when powered incorrectly or improperly installed. Sensors
containing internal electronic boards undergo addional quality checks prior to shipment to ensure that
components are 100% operaonal upon delivery. Ground loop and 3rd party hardware problems (including but
not limited to power supplies) may also cause blown electronics or damage to the sensor components and as such
faulty installaons are not covered by warranty.

TURTLE
TOUGH
®
18
Damage to Cables and Connectors
Please note that integral sensor cables, connectors and plugs must NOT be cut, removed or modied in any way.
Sensors contain sensive internal electronics and our cables and connectors are designed to protect the integrity
of these components. Any modicaon or alteraon to cables and connectors can compromise their integrity
and will void the warranty. Always use factory-approved/manufactured cables and connectors. Addionally, the
cable contains a unique idener laminated to the cable end and this must not be removed or it will also void the
warranty.

TURTLE
TOUGH
®
19
Return Goods
For all return goods the following informaon must be included in the leer accompanying the returned goods:
• Model Code and Serial Number
• Original Purchase Order and Date
• Length of me in service and descripon of the process
• Descripon of the fault and circumstances of the failure
• Process/environmental condions that may be related to the failure of the sensor
• Statement as to whether warranty or non-warranty service is requested
• Complete shipping and billing instrucons for return of material, plus the name and phone number of a contact
person that can be reached for further informaon
• Clean Statement: returned goods that have been in contact with process uids must be decontaminated and
disinfected prior to shipment. Goods should carry a cercate to this eect, for the health and safety of our
employees. Material Safety Datasheets must be included for all components of the process to which the sensor(s)
have been exposed.
All sensor returns are to be accompanied by a completed Return Material Authorisaon Document clearly stang
the reason for the return and with the Clean Statement lled in.
See the warranty and returns secon under support on our website turtletoughsensors.com for details.

TURTLE
TOUGH
®
20
Support
For technical support please contact head oce on +61 (0)3 9872 5055 or visit our website turtletoughsensors.com
for informaon on sensor care, calibraon, wiring and installaon-related issues.
TURTLE TOUGH DIRECT SMART SENSORS (DSS) are fully supported by a range of factory hardware soluons
including SMART Analysers, Handheld Field Communicators, PC Communicaon tools and eld implementaon
devices such as juncon boxes, extensions leads and power supplies. This ecosystem provides a complete soluon
for reliable connecvity and sensor management.
TURTLE TOUGH sensors may also be directly interfaced with end-user hardware via the MODBUS RTU protocol.
Integrang sensors directly into your industrial network requires a moderate to a high level of onsite experse
in system integraon and should ONLY be undertaken by someone with the requisite experience. While direct
integraon provides some signicant benets, it also incurs a higher degree of risk as integrang with 3rd party
hardware provides a lot of opportunies for unexpected issues to arise. 3rd party hardware includes any device or
component not supplied by Turtle Tough and can include, but is not limited to, PLC’s, SCADA systems, DCS’s, data
loggers, power supplies/isolators, or any other externally sourced interfacing device or component. We suggest
you read the sensor implementaon guide carefully to understand the full scope and requirement for installaon,
because:
1. Turtle Tough Pty Ltd and its representaves can provide only limited factory support when integrang sensors
with 3rd party hardware. Unexpected behaviour or damage as the result of 3rd party hardware is not the
responsibility of Turtle Tough.
2. Product support is limited to the informaon provided in our eld implementaon notes and guidelines. The
product is fully supported for installaon requirements, eld implementaon and programming protocols in
direct relaon to our product. It does not extend to the compability of our sensor with or the subsequent
behaviour of 3rd party hardware. Due to the innite amount of 3rd party hardware available we are unable to
support issues related to 3rd party problems and it is the sole responsibility of the end-user to test and verify
the compability of our product with their equipment.
3. Turtle Tough and its representaves are unable to provide support on 3rd party hardware or provide trouble-
shoong on equipment not supplied by us. If you intend to integrate the sensor directly with your hardware,
we highly recommend the following:
• Read the Field Installaon Guide and Sensor Implementaon Guide. Make sure you understand the requirements
and please contact us if you require claricaon or further informaon.
• Seek advice from your 3rd party hardware provider and make sure this equipment meets the minimum
specicaon requirement outlined in our support documentaon.
• Purchase a handheld eld communicator (HFC) from Turtle Tough. This stand-alone device, self-powers the
sensor and provides complete diagnoscs and communicaon. The HFC allows us to independently test and
verify the sensor performance, independent of your system or hardware and is the single fastest way to rule
out any sensor performance issues and be able to idenfy eld installaon problems.
• Alternavely, you may purchase a windows PC interface box. Once again, this factory-supplied and tested
component will ensure reliable communicaon with the sensor and it can be tested independently of your
system.
• Please note that our quick connect plugs must never be removed or modied in any way. This will immediately
void the warranty. Only use Turtle Tough factory interface leads or panel mount connectors to interface with
your hardware.