DocNo.FM0706issueDPage1
169MHzTelecareDevices
PillDispenserAdvance
__________________________________________________________________________________________________________


CompatiblewithReachat‐homealarms,Adventxtwardencall
AltecResponseandTouchsafeProNursecallsystems
Audiblealarm&flashinglightwhenpillsarereadytotake
1to28daycapacity–1to24dosesperday
Self‐lockingpillshuttertopreventdoubledosing
Keylockablelidtopreventaccesstomedicationtray
EasytoprogramusingLCDdisplaymenus
Realtimeclockwithauto‐BSTadjustment
Lowbatterywarning
Dimensions:180mmdiameterx56mmdeep
Weight:490gramsincbatteries
ProductCode:ZXT485

HowdoesthePillDispenserAdvancework?

ThePillDispenserAdvancewillsoundanalarm,flasharedlightandthedisplaywillshow“TAKEMED”whenit’stimeforthe
usertotaketheirmedication.

Itwillmakethecorrectdosageavailableatthecorrecttimeeachday,atallothertimesthepillsaresecurelylockedinsidethe
dispenser.

Whenthealarmsoundstheshutterwillbeunlocked,thepillsareremovedfromthedispenserby
tiltingitthrough90oandpouringthemintothehand.Thetiltingactionwillsilencethealarm
thenre‐locktheshutteragainafter2minutes.

Ifthedispenserisnottiltedthealarmwillcontinuetountiltheprogrammable“after”time
hasexpired(5minutesto5hours).

Whenhalfthe“after”timeexpiresanalarmcallwillbetransmitted.Whenthefull“after”time
hasexpiredanotheralarmwillbetransmittedandtheshutterwilllocktopreventthepossibility
of2dosesbeingtakenwithinashorttime(i.e.potentialdouble‐dosing).

Pills“ready”andpills“taken”timesarealsotransmittedandloggedintheReachPlusorAdventxt.
ThisdatacanbeanalysedusingTynetec’siCaresoftware.

Note:Adventxtrequiresadditionalhardware.

LoadingthePillDispenser

ThePillDispenserisprovidedwithclearlyprinteddiscs.Thesediscsidentifythedayoftheweekandtheappropriatenumberof
dosesperday(between1and4).Ablankdiscisprovidedforthoseon5ormoredosesperday.

Placetheappropriatediscfortherequiredpilldosageontopofthetraytakingcareto
line‐uptheholesandpushdownoverthe4locatingpins.
Usethediscasaguidewhenplacingthepills.

Certainmedicationshouldnotbeexposedtotheairandthereforeshouldnot
beplacedinautomaticdispensers;ifindoubtcheckwithaPharmacist.

Afterfilling,positionthetraysothatthecompartmentwiththefirstdosetobetakenby
theuseristotherightoftheopening.ThephotohereshowsMonEveisthenextdose

ResettingtheDosesLeftDisplay

Eachtimemedicationisdispensedthenumberof“DOSESLEFT”willreduceby1.
Whenthetrayisre‐filleditisimportanttoresetthecounterbypressingbutton2until“TRAYZ”isdisplayed.
Ifthedispenserisworkinginthe“SECURE”modethetraywillmoveahalfturntolocktheshutter.
Thenumberof“DOSESLEFT”willnowbethesameasthenumberenteredinthe“FILLED”quantitysetting.
Closethelidandlockwiththekeysupplied.

Therefillfrequencywilldependonthenumberofdailydoses;onceadayrefillevery28days,twiceadayrefillevery14days,
threetimesadayrefillevery9days,fourtimesadayrefillevery7days,etc.






Tynetec andAidcallare businessunitsofLegrandElectricLtd
Unit10CowleyRoad,BlythRiversideBusinessPark,Blyth
Northumberland,NE245TF.Tel:01670352371Fax:01670362807
Email:[email protected]Web:www.tynetec.co.uk
DocNo.FM0706issueDPage2
INITIALSET‐UP

Fromnewyouwillneedtoopen
thebatterycompartmentand
removethebatteryisolationstrip.

Thedisplaywillshow“BOOT”
followedbythesoftwareversion
e.g.“SW‐007”thenthebatterychargelevel“B100%”(full
charge).Whenthedisplayshows“CLOCK”andflashes
“00:00”theunitisreadytoprogram.

Thereare2programmingmenus;MenuAcontainsthe
basicsettingsandMenuBcontainsmoreadvanced
functions.Seebelow;

MENUA
S
ETTING
S


Pressbutton1tocyclethroughthefollowingfunctions;
CLOCKDATEDOSESALARMS
Pressandholdbutton1for3secondstoexitthemenuand
reverttonormaloperation.

SettingtheClock
Pressbutton1onceandCLOCKisdisplayed.
Pressbutton2/3tosetthetime(00:00to23:59)
Note:pressandholdforfastforward/reverse
Pressbutton1toendandDATEwillbedisplayed.

SettingtheDate
TheDATEdisplaywillchangetoYYafterafewseconds.
Pressbutton2/3tosettheYEAR(00to99)
Pressbutton1andMMwillbedisplayed
Pressbutton2/3tosettheMONTH(01to12)
Pressbutton1andDDwillbedisplayed
Pressbutton2/3tosettheDATE(01to31)
Pressbutton1toendandDOSESwillbedisplayed.

SettingtheNumberofDosesperDay
TheDOSESdisplaywillchangetodisplaythenumberof
dosesperdayafterafewseconds.
Pressbutton2/3tosetthedosesperday(00to24)
Pressbutton1toendandALARMSwillbedisplayed.

SettingtheAlarmTimesforeachDose
TheALARMSdisplaywillchangetodisplaythetimeforthe
firstdoseafterafewseconds.
Pressbutton2/3tosetthealarmtimeforthefirstdose
Pressbutton1toendthensetalarmtime2
Repeatforeachalarmtime,thenumberofalarmtimesis
determinedbythe“numberofdosesperday”setabove.

MENUBSETTING
S


Pressandholdbutton3untilBEFOREisdisplayedthen
pressbutton1tocyclethrough;AFTERSOUNDLIGHT
FILLEDBATTSECUREHOURSDLSSTART
STOPRECORD
Pressandholdbutton1for3secondstoexitthemenuand
reverttonormaloperation.

SettingtheBeforeTime
TheBEFOREtimeallowsmedicationtobetakenupto5
hoursearlierthanthenextalarmtime.
Pressandholdbutton3untilBEFOREisdisplayed
Pressbutton2/3tosetthebeforetime(00:00=not
required,to05:00max)
Pressbutton1toendandAFTERwillbedisplayed
Ifthisoptionissettheusermayaccesstheirmedicationin
advancebyturningthepilldispenserupsidedownthenback
uprightagaintwiceinsuccession,thetraywillrotateone
stepandthenextdosewillbecomeavailable.

SettingtheAfterTime
TheAFTERtimesetshowlongthealarmsoundsandthepill
shutterremainsunlockedafterthemedicationisdue.
AfterAFTERisdisplayed;
Pressbutton2/3tosettheaftertime(00:05to05:00)
Pressbutton1toendandSOUNDwillbedisplayed.
IfhalftheAFTERtimeexpiresanalarmcallwillbe
transmitted.IfthefullAFTERtimeexpiresanotheralarm
callwillbetransmittedandthepillshutterwilllockshut.

SettingtheAlarmSound
ThealarmSOUNDwillplaytoalerttheuserwhenitistime
totaketheirmedication.Itwillsoundevery5secondsfor
thefirstminute,thenevery8secondsupto1hourthen
onceaminuteuntilthemedicationisdispensedorthe
AFTERtimeisreached,maximum5hours.
AfterSOUNDisdisplayed;
Pressbuttons2/3toselectthepreferredsound(01to04)
Select00fornosound
Pressbutton1toendandLIGHTwillbedisplayed.

SettingtheLightOn/Off
TheLIGHTwillflashtoalerttheuserwhenitistimetotake
theirmedication.Itwillflashevery5secondsforthefirst
minute,thenevery8secondsupto1hourthenonceevery
16secondsuntilthemedicationisdispensedortheAFTER
timeisreached,maximum5hours.
AfterLIGHTisdisplayed;
Pressbutton2fornolight(displayshowsOFF)
Pressbutton3forlight(displayshowsON)
Pressbutton1toendandFILLEDwillbedisplayed.

SettingtheFilledQuantity
Thisisthenumberofdosesplacedinthemedicationtray
whenfilledandallowstheDOSESLEFTdisplaytobeshown.
AfterFILLEDisdisplayed;
Pressbutton2/3tosettheFILLEDquantity(00to28)
Pressbutton1toendandBATTwillbedisplayed.

SettingtheBatteryType
Itispossibletouse4xAAsizealkalineor4xAAsize
rechargeablebatteriesinthe‘Advance’PillDispenser.
AfterBATTisdisplayed;
Pressbutton2toselectalkaline(ALC)
Pressbutton3toselectrechargeable(NIMH)
Pressbutton1toendandSECUREwillbedisplayed.

SettingtheSecureMode
WithSECUREmodeONthepillshutterwillbelockedexcept
atalarmtimeandforthesubsequentaftertime.With
SECUREmodeOFFtheshutterdoesnotlock.
AfterSECUREisdisplayed;
Pressbutton2toselectsecuremodeON
Pressbutton3toselectsecuremodeOFF
Pressbutton1toendandHOURSwillbedisplayed.

SettingtheHoursMode
Theclockcanbesettodisplaythetimein12hourAM/PM
modeor24hourmode.
AfterHOURSisdisplayed;
Pressbutton2toselect12hourclock(AM/PM)
Pressbutton3toselect24hourclock
Pressbutton1toendandDLSwillbedisplayed.

SettingtheDaylightSavingMode
Toclockcanbesettoautomaticallyadjust+/‐1hourevery
March/OctoberforGMT/BST.
AfterDLSisdisplayed;
Pressbutton2/3toselectOFF/EU/US
SetEUforEurope,USforAmericaorOFFtodisable.
Pressbutton1toendandSTARTwillbedisplayed.

DocNo.FM0706issueDPage3
MENUBSETTING
S
‐CONTINUED

SettingtheStartTime
Itispossibletodelaythetimeatwhichthedispenserwill
starttooperate(usefulifthedispenserisbeingfilledoff
siteandbeingdeliveredtotheuseratalaterdate).
AfterSTARTisdisplayedthedisplaywillchangetoYYafter
afewseconds;
Pressbutton2/3tosettheYEAR(00to99)
Pressbutton1andMMwillbedisplayed
Pressbutton2/3tosettheMONTH(01to12)
Pressbutton1andDDwillbedisplayed
Pressbutton2/3tosettheDATE(01to31)
Pressbutton1andHHwillbedisplayed
Pressbutton2/3tosettheHOUR(01to23)
Pressbutton1toendandSTOPwillbedisplayed.
Thetime(HH)shouldbeinadvanceofthefirstdosetobe
dispensedfortheuser.

SettingtheStopTime
Itispossibletostopthedispensersoperationatafuture
dateandtime.Thisisusefuliftwodispensersarebeing
usedbyoneuserinsequenceandthisensuresonlyone
dispenserisactiveatonetime.
AfterSTOPisdisplayedthedisplaywillchangetoYYaftera
fewseconds;
Pressbutton2/3tosettheYEAR(00to99)
Pressbutton1andMMwillbedisplayed
Pressbutton2/3tosettheMONTH(01to12)
Pressbutton1andDDwillbedisplayed
Pressbutton2/3tosettheDATE(01to31)
Pressbutton1andHHwillbedisplayed
Pressbutton2/3tosettheHOUR(01to23)
Pressbutton1toendandRECORDwillbedisplayed.
Thetime(HH)shouldbeafterthelastdosetobedispensed
fortheuser.Ifthefunctionisnotrequiredsettheyearto99.

RecordFunction
Thedispenserwillrecordtheactualtimestheprevious28
medicationsweredispensed.
AfterRECORDisdisplayed;
Pressbutton2/3toscrollthroughtheactualtimesthelast
28medicationsweretaken.
Pressbutton1toend.

S
ELFTESTMOD
E


Thetrayrotation,lightandalarmsoundcanallbetested
anddemonstratedbypressingandholdingall3buttons
downtogetheruntilTESTisdisplayed.
Pressbutton3tostartawholealarmsequence,thedisplay
willshow“DEMO”.
Closethelidandthetraywillturntotheopenposition.
Thealarmwillsound,thelightwillflashandthedisplaywill
show“TAKEMED”.
Pickthedispenserupandtipforwardtodemonstrate
dispensingthemedication.
Thedisplaywillshow“CLOSE”andthetraywillturntothe
lockedposition.
Pressbutton3torepeatorpressandholdbutton1fora
fewsecondstoend.

ThismodeisalsousedwhenlearningthePillDispenser
ontothewardencallsystem,at‐homealarmunitor
Nursecallsystem.
Seethefollowingpagesforhowtoputeachofthese
productsinto“LearnMode”.



ERRORALARM

Ifthetrayfailstorotateduetoanobstructionanalarmwill
soundandthedisplaywillshowERROR01,04or08.
Thiscanbeclearedbypressingbutton3.
Anyobstructionshouldberemoved.
Whentheobstructionisclearedretesttheunit.
Ifthefaultcannotberectifiedcontactyoursupplier.

LOWBATTERYWARNING

Ifthebatterylevelfallsbelow20%thedisplaywillshow
“LOWBATT”insequencewiththeDATEandTIME.
Ifthebatterieshavenotbeenreplacedby00:05:10(five
minutesand10secondsaftermidnight)theunitwill
transmitalowbatterywarningatthattime.Thiswillrepeat
everynightat00:05:10untilthebatteriesarereplaced.
Ifthebatterypercentagefallsto0%thepilldispenserwill
ceasetooperate,itwillbleepandthedisplaywillshow
“ERROR10”.

Thebatterylevelcanbecheckedatanytimebytiltingthe
dispenserforward(between15and40degrees);afterthe
“NEXTDOSE”and“DOSESLEFT”informationthebattery
levelremainingwillbedisplayede.g.“B75%”.

Allprogrammingisretainedwhenthebatteriesarechanged.
Thetimemayneedcorrectingifthebatteriesareleftoutor
arecompletelyoutofcharge.
Replacementbatteries:4x1.5VAlkalineAA
Rechargeablebatteriesmayalsobeusedbutensurethe
“BATT”functioninMenuBissetas“NIMH”.

Detailedinformationonbatterymanagementforall
Tynetecproductsisavailableinaseparatedocument.
RequestacopyofDocNo.FM0630.

DISPENSERLEFTUPSIDEDOWN

Ifthedispenserisupsidedownwhenanalarmoccurs,the
alarmwillsoundasnormalbutthetraywillnotturnuntil
thedispenserisreturnedupright.

Ifmorethanonealarmtimehaspassedwhilstthe
dispenserhasbeenupsidedownthedispenserwillpresent
thecorrectmedicationforthattime.

CLEANING

DispenserUnit:donotplaceinwater,wipecleanwitha
dampclothonlyandpolishwithadryduster.Avoidusing
strongdetergentsorfurniturepolish.
InternalTray:maybewashedinwarmwaterorinadish
washer(max70C).

















DocNo.FM0706issueDPage4
RegisteringaPillDispenserontoaReachat‐homealarmunit

1.PuttheReachunitintoradiodevicelearnmode…

     









2.Activatethepilldispenser…


  













3.Exitradiodevicelearnmode…


     










4.CheckwiththeControlCentrethatitisconvenienttoperformatestcall.Checkapilldispenseralarmisreportedbylettingthe
pilldispenserdeliveradose,DONOTtilttheunittoTAKEthedose,andlethalftheAFTERtimeexpire.HINT:tospeed‐upthetest
adjusttheCLOCKtofewminutesbeforethenextdosetimeandreducetheAFTERtimeto5minutes.Don’tforgettoresetthe
clock/aftertimewhenthetestcalliscomplete.


DeletingaPillDispenserfromaReachat‐homealarmunit

Enter1670usingthekeypadontherearoftheReachunit–thefrontlightwillFLASHFASTGREENandtheunitwillannounce
“ProgramMode‐Ready”.
Enter#041andtheReachwillannounce“DeleteRadioDevice”followedbya“Beep”.
Enterthe8digitpilldispenserIDtobedeleted;afterthe“Beep”press#toexit.
PressthekeytoexitProgramMode;theunitwillsounda“Double‐Beep”andthefrontlightwillreturntoSTEADYGREEN.


The8digitIDcanbefoundonthelabelonthebackofthepilldispenser–itwillalwaysstartwith31xxxxxx





TheinstructionsgivenaboveforthestandardReachproductalsoappliestotheReachPlus,ReachPlusGSM
andReachUltramodels.Refertotheinstructionbookletsprovidedwiththeseproductsformoreinformation
onprogrammingandtestingradiodevices.





Openthelid,pressandholdall3buttonstogetheruntilTESTis
displayed.

Pressbutton3,thedisplaywillshowDEMOandtheReachwill
announce“PillDispenser”.

Pressandholdbutton1toexittheDEMOmode.
“Trigger
Radio
Device”
PressandHOLDtheGREENbuttonontheReachunit.

WhenthefrontlightFLASHESFASTGREENreleasethe
buttonandtheReachwillannounce“TriggerRadioDevice”.

Younowhave1minutetoactivatethepilldispenser.

PresstheGREENbuttonontheReachunit.

ThefrontlightwillreturntoSTEADYGREENandtheReach
isbackinnormaloperatingmode.

DocNo.FM0706issueDPage5
RegisteringaPillDispenserontoanAdventxtwardencallsystem

1.PuttheAdventxtsystemintoradiodevicelearnmode…
 
     
 












2.Activatethepilldispenser…


  













3.ExittheAdventxtsystemradiodevicelearnmode…















4.Checkapilldispenseralarmcallisreportedonthewardencallsystembylettingthepilldispenserdeliveradose,DONOTtiltthe
unittoTAKEthedose,andlethalftheAFTERtimeexpire.HINT:tospeed‐upthetestadjusttheCLOCKtofewminutesbeforethe
nextdosetimeandreducetheAFTERtimeto5minutes.Don’tforgettoresettheclock/aftertimewhenthetestcalliscomplete.


DeletingaPillDispenserfromanAdventxtwardencallsystem

PresstheLINEkeyandwaitforthe“Ready”prompt.
Enter5001followedbythekey.
Entertheflatnumberthepilldispenserisbeingdeletedfromfollowedbythekey.
Activatethepilldispensertobedeleted.
Waitforthe“PendantDisabled–EnterFlat”prompt.
Pressthe#keyandwaitforthe“Ready”prompt,thenpresstheLINEkeytocleartheconnection.





RefertotheAdventxtoperatorsmanual(DocNo.FM0420)providedwiththewardencallsystemformore
informationonprogrammingandtestingradiodevices.



UsingtheManager’sDECTtelephone;

PresstheLINEkeyandwaitforthe“Ready”prompt.

Enter5000followedbythekey.

Entertheflatnumberthepilldispenserisbeinginstalledinfollowed
bythekey.

Younowhave1minutetoactivatethepilldispenser.

“Ready”
UsingtheManager’sDECTtelephone;

Pressthe#keyandwaitforthe“Ready”prompt.

PresstheLINEkeytocleartheconnectionandthewardencallsystemis
nowbackinnormaloperatingmode.

Openthelid,pressandholdall3buttonstogetheruntilTESTisdisplayed.

Pressbutton3andthedisplaywillshowDEMO.

Waitforthe“EnterLocationCode”promptthenpress0(nolocation)
followedbythekey.

Waitforthe“PendantEnabled–EnterFlat”prompt.

Pressandholdbutton1toexittheDEMOmode.

“Ready”
DocNo.FM0706issueDPage6
RegisteringaPillDispenserontoanAltecResponselocalcareralarmunit

1.PuttheAltecResponseunitintoradiodevicelearnmode…


     
 











2.Activatethepilldispenser…


  








 





3.ExittheAltecResponseradiodevicelearnmode…















4.CheckapilldispenseralarmcallisdisplayedontheAltecResponseunitbylettingthepilldispenserdeliveradose,DONOTtilt
theunittoTAKEthedose,andlethalftheAFTERtimeexpire.HINT:tospeed‐upthetestadjusttheCLOCKtofewminutesbefore
thenextdosetimeandreducetheAFTERtimeto5minutes.Don’tforgettoresettheclock/aftertimewhenthetestcalliscomplete.

DeletingaPillDispenserfromanAltecResponselocalcareralarmunit

PressENTERtoactivatethewake‐updisplaythenENTERagaintodisplaythemainmenu
PressENTERtoselectRadioDevices
PressthekeytoselectDeleteDevicethenpressENTER
Thedisplaywillpromptyouto“TriggertheSensor”
Activatethepilldispensertobedeleted‐thedisplaywillshowthedevicetypeandits8digitIDnumber
IfthisisthedeviceyouactivatedpressthekeytoselectDeletethenpressENTER
Ifit’sNOTthedeviceyouactivatedpresstheENTERkeytoselectRejectthentryagain
ToexitthemodeselectExitthenpresstheENTERkey.




RefertotheAltecResponseinstallationguide(DocNo.FM0548)providedwiththeportablealarmunitfor
moreinformationonprogrammingandtestingradiodevices.



Openthelid,pressandholdall3buttonstogetheruntilTESTisdisplayed.
Pressbutton3andthepilldispenserdisplaywillshowDEMO.

TheAltecResponsedisplaywillshowPillDispenserandits8digitIDnumber.
ChecktheIDiscorrect,pressthekeytoselectAcceptthenpressENTER.
Note:ifanotherdevicehasbeenpicked‐uppressENTERtoRejectandtryagain.
ThedisplaywillshowFlat,usethe&keystochoosethedwellingtype.
PressthekeytoselectNo:000000thenpresstheENTERkey.
Usethe&keystoselecteachdigitoftheflatnumberandthe&keys
tomovebetweendigitsthenpresstheENTERkey.
PresskeyseveraltimestoselectExitthenpresstheENTERkey.
Thedisplaywillpromptyouto“TriggertheSensor”again.

Pressandholdbutton1toexittheDEMOmode.
PressENTERtoactivatethewake‐updisplaythenENTERagaintodisplay
themainmenu.

PressENTERtoselectRadioDevices

PressENTERtoselectLearnDevice

Thedisplaywillpromptyouto“TriggertheSensor”.

Younowhave20secondstoactivatethepilldispenser.
FromtheLearnDevicemenuselectExitthenpressENTER.

FromtheRadioDevicesmenuselectExitthenpressENTER.

Thedisplaywillreturntothewake‐upscreenandgoblankafter10seconds.

TheAltecResponseisnowbackinnormaloperatingmode.

tynetec
tynetec
DocNo.FM0706issueDPage7
RegisteringaPillDispenserontoaTouchsafeProNursecallsystem

1.PuttheTouchsafeProMasterPanelintoTelecareregistrationmode;touchthescreenandLogonasanEngineeroraManager
(passwordrequired),thenfromthemainmenupagetouchtheTelecarebuttontoopenthescreenshownbelow…























2.EnterthepilldispenserID…









3.PresstheClosebuttontoexitTelecaremodeandtheNursecallpanelisnowbackinnormaloperatingmode.

4.CheckapilldispenseralarmcallisdisplayedontheNursecallpanellikebelowbylettingthepilldispenserdeliveradose,DONOT
tilttheunittoTAKEthedose,andlethalftheAFTERtimeexpire.HINT:tospeed‐upthetestadjusttheCLOCKtofewminutesbefore
thenextdosetimeandreducetheAFTERtimeto5minutes.Don’tforgettoresettheclock/aftertimewhenthetestcalliscomplete.






Thescreenbannerisblackforapilldispenseralert(Touchthebannertodisplaynotes).
PresstheAcceptbuttononthebannertosilencethealarm‐thiswillalso
notifyallotherpanelsthecallisbeingdealtwith.
Thescreenbannerchangestopurplewhenacallisaccepted.
YoumustvisittheappropriateNursecallCallPointtocancelthecall.

DeletingaPillDispenserfromaTouchsafeProNursecallsystem

SelectthepilldispenserIDtobedeletedfromtheDeleteTelecareDevicepull‐downlistbox(prefixedwitha“T31”).
DetailsofthepilldispenserselectedwillbedisplayedintheRoom,HouseCodeandIDboxesbelow.
IfthedetailsshownarecorrectpresstheDeletebutton.
PresstheClosebuttontoexit.





RefertotheTouchsafeProDisplayPaneluserguide(DocNo.FM5505)providedwiththeNursecallsystemfor
moreinformation.




TynetecandAidcalloperateapolicyofcontinuousimprovementandreservetherighttochangeproductspecificationswithoutnotice.
IfanyvariationtothedetailscontainedinthisdocumentaresuspectedpleasecontactTynetec’scustomersupportdepton01670352371.

The8digitIDisprintedonthelabelontherearofthepilldispenser.
Note:thisIDwillalwaysstartwith31xxxxxx.

TouchtheAddTelecareDeviceboxthenusetheon‐screenkeypadtoentertheID.

Hint:enterazoneintheFilterboxtonarrowthesearchfortheroom(seebelow).

SelecttheusersCallPointIDfromtheRoompull‐downlistboxthenpressSave.
Note:thisistheroomunitthatwillbewillbeusedtocancelthecall.
Usekeypadtoenter
8digitIDhere
Selecttherelevant
CallPointIDhere
Savewhendone
Closetoexitmode
On‐screenkeypad
CALLING
CANCEL
HC345ID001
PresstheLIGHTGREY
buttontocancelthecall
ZXT485W65707
Ser:11151014
ID:31036757
EE0765V1.01A
55


Popular Health Care Product manuals by other brands

Suunto PC POD manual

Suunto

Suunto PC POD manual

General Sensing MedSense Badge user manual

General Sensing

General Sensing MedSense Badge user manual

ResMed AutoSet CS-A user guide

ResMed

ResMed AutoSet CS-A user guide

Circadiance SleepWeaver 3D quick start guide

Circadiance

Circadiance SleepWeaver 3D quick start guide

Uclan XIN.BAO Quick start user manual

Uclan

Uclan XIN.BAO Quick start user manual

Tecla Shield DOS Getting started

Tecla

Tecla Shield DOS Getting started

MIO-CARE MiCor A100 user manual

MIO-CARE

MIO-CARE MiCor A100 user manual

Detecto ProDoc DHRWM owner's manual

Detecto

Detecto ProDoc DHRWM owner's manual

eVito collect manual

eVito

eVito collect manual

Helo LX manual

Helo

Helo LX manual

Omron BF212 instruction manual

Omron

Omron BF212 instruction manual

Ossur Miami Lumbar user manual

Ossur

Ossur Miami Lumbar user manual

Direct Healthcare Dyna-Form Mercury Advance Service manual

Direct Healthcare

Direct Healthcare Dyna-Form Mercury Advance Service manual

Braemar ER920W user manual

Braemar

Braemar ER920W user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.