Ulvac G-Tran Series Installation guide

SK00-9662-EQ-003-00
G-TRAN SERIES
Pirani Vacuum gauge SWU10-R
Quick Manual
Introduction
This quick manual is for quick check of operation and display of the product.
Please refer to instruction manual in advance for detailed information about
operation, precautions and safety for proper use. Available for download from
ULVAC website. https://showcase.ulvac.co.jp/ja
This manual is for the following gauges. Serial Nos. 00001 and higher.
Part Names and Functions
SWU10-R controller section
①
② ③
Name
Function
①
I/O connector
The I/O connector for the power supply and signals
(RJ-45 jack connector).
②
Gauge head
clamp The clamp to hold and lock the gauge head.
③
Mating screws
Mating screws (M3) for the clamp hood.
SWU10-R I/O connector
No
This
unit
Connection destination
1
Power supply +5 VDC
SH2/ST2
Multi-ionization gauge
2
GND
3
Power supply +5 VDC
4
Communication
terminal
5
Communication
terminal
6
GND
7
Power supply +5 VDC
8
GND
SWU10-R LED
Name (symbol)
Function
①POWER/ERROR LED
On : Operating normally
Flashing: Error such as a filament
break, this unit is malfunctioning
Off : Power off, this unit or the
SH2 / ST2 is malfunctioning
Attaching the Product
Loosen the clamp locking screws to loosen the clamp. Then attach the gauge
head with the procedure below.
①Align the pins for the gauge head and push it all the way in
②Clamp the gauge head with the clamp
③Tighten the locking screws to secure the gauge head
Handling Precautions
3.1. Precautions related to usage
・ Secure cables as much as possible so that force is not applied to cable
connectors.
・ Firmly insert the RJ-45 connector.
・ To perform accurate pressure measurements, allow the pressure sensor to
age for 20 minutes or longer before measuring.
3.2. Attaching this unit
・ Attach this unit so that the gauge head attachment opening surface is
parallel to the gas flow. In particular, ensure that gases do not enter the
gauge head interior like a beam.
・ As much as possible, attach the gauge head so the filament is vertical to
the force of gravity.
・ The Pirani vacuum gauge head filament is thin at φ25 µm, so avoid use in
a location with a large amount of vibration as much as possible. The
biggest cause of filament breaks is from mechanical shock, so use caution
regarding the installation location and handling.
・ Use an O-ring to attach the gauge head that releases little gas. There is a
risk of measurement errors or the gauge head lifespan will decrease if
materials that release a large quantity of gas, such as rubber tubes or grease,
are used in the gauge head connection.
Gas flow direction
Checking the Filament and Temperature Sensor for
Burnout
Filament
Approx. 5 Ω
Temperature
sensor
Approx. 1.1
kΩ
The diagram below shows the connection of the Pirani sensor head filament. Check
continuity referring to this diagram. The resistance of the filament is about 5Ω under
atmospheric pressure and that of the temperature sensor is about 1.1 kΩunder
atmospheric pressure.
Note)Beware of over-current
If the filament or temperature sensor is energized with an over-current of 2 mA or
more, it may burn out. Be alert to the current value on the measuring instrument.
Also measure the resistance value under atmospheric pressure because in vacuum, the
filament and other components are heated and an accurate resistance value cannot be
measured.
Specifications
Name Pirani vacuum gauge measuring unit
Model SWU10-R
Measurement principle Pirani vacuum gauge that uses thermal
conduction by gas molecule
Number of gauge heads 1pc
Applicable gauge head SWP-16,SWP-R1/8,SWP-P18,SWP-P15, SWP-
25, SWP-CF16, SWP-1S
Measurement pressure range
5.0×10
-
2
to 1.0×10
+5
Pa
Output pressure range 5.0×10
-
2
to 1.0×10
+5
Pa
Accuracy (N2) 1×10
-
1
Pa to 1×10
+4
Pa: ±10%
5×10-2Pa to 1×10-1Pa: ±20%
1×10+4Pa to 1×10+5Pa: ±20%
Repeatability 1×10
-
1
Pa to 1×10
+4
Pa: ± 2%
Sampling time 50 ms, 5x moving average
Measurement value output Output by serial communication
LED display POWER/ERROR LED
Gauge head material Filament:Pt
Other:SUS304, FeNiCo, Ni, Au, Glass, Ceramic
Gauge head withstand
pressure
2×10
+5
Pa(abs)
*Take the withstand pressure for flanges,
clamps, and other components into account
separately.
Operating temperature range
10 to 40℃
Operating humidity range 15 to 85% RH (no condensation)
Storage temperature -20 to 65℃(no condensation)
Bake out 150℃(without electronics unit)
Mounting position Free
IP protection class IP30
Power supply voltage 5VDC±5%
I/O connector RJ-45 jack connector
Gauge head internal volume Pirani vacuum gauge head(SWP-16): 7.3cm3
Weight Controller: Approx. 38 g,
External dimensions φ40×51 (controller section only)
Standard Accessories
Quick manual(This paper) 1copy
Trouble shooting
Bad electrical wiring is a major cause of this unit malfunctions. If this unit
behaves in a manner you find unusual, first check the wiring between this unit and
the SH2/ST2. Refer to the SH2/ST2 instruction manual as well.
The POWER/ERROR LED is
off even though power is being supplied
Cause
Action
Miswired or broken cable
Repair the wiring and check continuity
with a tester.
The power supply voltage to the
SH2/ST2 is not within the specified
range
Check the power supply voltage with a
tester.
This unit or the SH2/ST2 is
malfunctioning
Requires inspection and repairs at
ULVAC.
The POWER/ERROR LED is flashing or the SH2
/ST2
keeps outputting the
error signal
Cause
Action
Filament break or short circuit
Measure filament continuity and
isolation between each pin and the case.
⇒Replace the gauge head when there is
a problem.
The gauge head is not attached
Attach the gauge head.
This unit or the SH2
/ST2
is
malfunctioning
Requires inspection and repairs at
ULVAC.
Pressure does not reach 5×10-2 Pa or lower even when pumping to vacuum
Cause
Action
The actual pressure is not 5×10
-
2
Pa
or lower
Check the pressure with an ionization
vacuum gauge.
Room temperature, gas temperature
is different when measuring
This is normal. Measurements are
temperature dependent.
This unit or the SH2
/ST2
is
malfunctioning
Requires inspection and repairs at
ULVAC.
The pressure reading does not show a steady value.
Cause
Action
The pressure is actually fluctuating
This is normal.
This unit is affected by
pump
vibrations and fluctuating
Suppress vibrations as much as possible.
Measurements are affected by the
principle of the Pirani vacuum gauge.
Attached to a location with a gas
flow
Attach this unit to a location with no gas
flow.
Measurements are affected by the
principle of the Pirani vacuum gauge.
The environmental temperature is
fluctuating
Wait for the environmental temperature
to stabilize.
Measurements are affected by the
principle of the Pirani vacuum gauge.
High temperature and low
temperature gases are flowing in
Implement corrective action such as
attaching the gauge head to a location
unaffected by gas temperature.
Measurements are affected by the
principle of the Pirani vacuum gauge.
Dirty gauge head interior
Replace the gauge head
Affected by moisture and dirt
Attach the gauge head to a location as
unaffected as possible
Influenced by heat of vaporization by
gradual vaporization.
There is a leak in the gauge head
or the area where the gauge head is
attached
Gauge head leak ⇒Replace the gauge
head
Other leaks ⇒Stop the leak
Bad cable contact, increased wire
resistance
Check the cable connection and
resistance.
The cable is affected by
electromagnetic induction.
(By external noise)
Change the cable installation location.
Use this unit with the device generating
the noise turned off.
Review how the cables are installed
Examine separate measures for noise
18

The expected pressure and measured pressure greatly differ
Cause Action
The gauge head is dirty
Sensitivity has dramatically deviated
Replace the gauge head with another
gauge head and check the phenomena.
If there is no problem with the other
gauge head, the gauge head was the
problem.
Measuring air at the nitrogen setting.
This is normal.
When air is measured at the nitrogen
setting, a pressure 60% higher is shown.
Environment with a large amount of
moisture or oil
Attach the gauge head to a location as
unaffected as possible
Influenced by heat of vaporization by
gradual vaporization.
The actual pressure differs Check with another pressure gauge.
Warranty
This product was shipped after rigid company inspection. However, in case any
failure occurs under ULVAC’s responsibility, such as defect in manufacturing and
damage during transportation, Buyer shall inform ULVAC, Inc. or the local ULVAC
representatives. ULVAC will repair or exchange it at free of charge.
Warrantable Items
1)This unit
Duration of guarantee
One (1) year after shipping date from ULVAC
Warrantee scope
1)Domestic business in Japan: Product, which has damage, caused by a failure
on delivery.
2)Direct export transaction: Product, which has damage, caused by a failure on
delivery. The warrantee scope shall confirm to the new INCOTERMS.
3)
Products not satisfying meet the standard specifications although the product
is used under the normal service conditions such as temperature range and
power etc.
Response procedure
1)Domestic business in Japan: ULVAC send a replacement or Buyer return the
defective items to ULVAC, Inc. or to the local ULVAC representatives for
repair. If field service is required, Buyer shall ask ULVAC, Inc. or the local
ULVAC representatives.
2)
Direct export transaction: ULVAC send a replacement or Buyer return the
defective items to ULVAC, Inc. or to the local ULVAC representatives for
repair. Return charge shall be paid by Buyer.
Disclaimer
1)Failure occurred after expiration of warranty period
2)Failure caused by force majeure, such as fire, storm and flood damage,
earthquake, lightning strike, war etc
3)Failure occurred due to carelessness handling or faulty usage
4)Products remodeled, disassembled or repaired without ULVAC’s acceptance
5)Failure occurred under abnormal environment, such as intense electromagnetic
field, radiation, high-temperature, high-humidity, flammable gases, corrosive
gases, dust etc.
6)Failure occurred by noise
7)Product deficiency or secondary damnification occurred to Buyer, from law
suit to ULVAC by third party for patent infringement.
8)Sensor head being used (expiration of life, measurement error, etc.)
9)
Sensor head cable in use (cable burnout due to improper installation, poor
contact, etc.)
Others
1) In case, special agreement or memorandum for specifications is made
individually, the descriptions are prior to this article “13 Product Warranty”.
2) Buyer shall inform ULVAC when this product is exported out of Japan. In
the meantime, Buyer shall take necessary procedures according to Foreign
Exchange and Foreign Trade Law.
3) As for the question and consultation, Buyer shall check the model and serial
number and ask the local representative or ULVAC, Inc.
4)
The content of this document is subject to change without notice in future.
Certificate of Decontamination
All material must be certified as decontaminated and this certificate must be
submitted to your closest local ULVAC service center or sales office prior to
shipment. The form is available for download from ULVAC website.
Network
ULVAC,Inc: http://www.ulvac.co.jp/eng/index.html
Service Centers: http://www.ulvac.co.jp/eng/support/service/index.html
Sales Offices: http://www.ulvac.co.jp/eng/support/sales_office/index.html
Drawing
ULVAC, Inc.
Components Division,
http://www.ulvac.co.jp/
Other manuals for G-Tran Series
11
This manual suits for next models
1
Other Ulvac Control Unit manuals
Popular Control Unit manuals by other brands

Panasonic
Panasonic PAN1326C2 quick start guide

Bucher
Bucher C-LRV Series Installation and startup guide

Hattersley
Hattersley Fig.1051 Installation and operating instructions

Telegesis
Telegesis ETRX2-PA product manual

Discrete Wireless
Discrete Wireless MARCUS 3G GPRS installation manual

Richter
Richter GU/F Series Installation and operating manual