Vaniman PURE BREEZE FX User manual

PURE BREEZE FX
FUME EXTRACTOR
25799 Jeerson Ave. ● Murrieta, CA 92562 ● 1-800-VANIMAN (826-4626) ● www.vaniman.com ● [email protected]
10380-12/21-PAGE 1
- Alternating Current
- Caution, risk of electrical shock
Warnings and Precautions
This Fume Extractor is designed to run on 115 volt alternating current, 60 Hz supply voltage in
temperatures ranging from -10 to +40 degrees celcius. With a maximum current usage of 15 amps. The
plug on the unit labeled “Power” is for the supply line cord provided. The supply plug in the wall must have
a wired ground and if the back cover is removed for any reason, do not bypass the protective earthing.
This dust collector is to be used for capturing ne aerosols or fumes, and is not suitable for direct
liquid suction.
When placing the unit in a location be sure that the power connector is accessible at all times. Any
side of the dust collector that has a vent on it must be a minimum of 2 inches from any obstruction, and
have a vent area of 20 square inches to allow ventilation to ambient room air. This will keep the unit cool
and ensure a longer motor life.
Failure to comply with the precautions listed here or any others for this equipment may cause
electrical shock or equipment failure.

1
2
3
4
PURE BREEZE FX
FUME EXTRACTOR
25799 Jeerson Ave. ● Murrieta, CA 92562 ● 1-800-VANIMAN (826-4626) ● www.vaniman.com ● [email protected]
10380-12/21-PAGE 2
Replacement Parts
Fig. # Description Part Number
1 Caster Wheels 3140
2Extra Rigid Capture Hose 96101
3Nozzle (5 Pack) 96102
4 Foot Remote/Bellows 1443
ASSEMBLY INSTRUCTIONS
1. Once the unit is removed from its packaging, remove the four caster wheels from the bag packed with
the power cord. Lay the unit on its back on top of several pieces of foam packaging to protect it. Screw
the four caster wheels into the four threaded holes in the bottom of the unit. Use the supplied Spanner
Wrench to tighten the caster wheel to the case, BUT DO NOT OVER TIGHTEN.
2. Remove the four thumb screws on the front of the unit, as well as the lter cover panel to expose the
HEPA and Charcoal Odor Filter inside the unit. Remove both lters from the unit, and then remove
the plastic packaging from both of the lters. Return the lters back into the unit it the correct order
with the Charcoal Odor Filter sitting on top of the HEPA Filter. Replace cover panel and thumb screws.
3. Connect the Extra Rigid Capture Hose to the top of the unit by sliding the black hose adapter onto
the inlet tube (only one end of the hose will t). Slide nozzle into opposite end of the hose and lock it
by twisting the nozzle clockwise until you feel a click. Adjust the hose to the desired position. If your
hose losese rigidity and is struggling to stay in the desired position, add rigidity by holding the hose
rmly with one hand and twisting the hose in the counter clockwise direction (with regards to looking
down the hose). Keep twisting until the hose reaches the desired rigidity. If the hose is twisted in the
wrong direction, the hose will begin to bulge in that area and become even less rigid. If continued the
hose will break.
4. Plug one end of the removable power cord into the socket labeled “Power Cord” on the side of the
unit, and the other end into a standard grounded wall socket. Test the unit by ipping the On/O switch
into the On position. NOTE: If using the On/O switch to turn the unit On, you MUST use the same
switch to turn the unit O.
IMPORTANT SERVICE INFORMATION:
SAVE SHIPPING BOX AND FOAM FOR FUTURE
SERVICE. SHIPPING WITHOUT PROPER FOAM
MAY CAUSE DAMAGE TO UNIT
PACKING LIST
1 - Pure Breeze FX Fume Extractor
1 - Extra Rigid Capture Hose
1 - Nozzle
1 - Power Cord
4 - Caster Wheels
1 - Foot Remote/Bellows

PURE BREEZE FX
FUME EXTRACTOR
25799 Jeerson Ave. ● Murrieta, CA 92562 ● 1-800-VANIMAN (826-4626) ● www.vaniman.com ● [email protected]
10380-12/21-PAGE 3
5. While the unit is O, plug the black foot remote tube into the round black port labeled “Remote” on the
unit. To use the foot remote instead of the On/O switch for sanitary purposes, leave the On/O switch
on the side of the unit in the O position. Test the remote function by pressing on the round bellows to
turn it On; press it again to turn it O. NOTE: If using the foot remote to turn the unit On, you MUST
use the foot remote to turn the unit O as well. Once the foot remote air switch has been activated by
the foot remote, the On/O switch will not turn the unit O, only the foot switch will.
FILTER REPLACEMENT INSTRUCTIONS
It is recommended that you replace the HEPA Filter when it becomes excessively dirty, or every 6-12
months. Replace the Charcoal Odor Filter when its eectiveness has noticably decreased.
1. Turn o the unit.
2. Remove the four thumb screws on the front of unit and set aside for later use.
3. Remove the panel to expose the HEPA and Charcoal Filter
4. Slide the Charcoal Filter out rst, then the HEPA lter by pulling on the tab. Once removed, properly
dispose of the dirty lters.
5. Remove any packaging from the replacement HEPA and Charcoal Filter. Carefully align the HEPA
lter frame with the outer walls of the unit, sit the lter onto the guide ledge, and slide until the lter
stops. Once the HEPA lter is installed, carefully place the replacement Charcoal Filter on top of the
HEPA lter.
6. Put the front panel back on the unit and secure with thumb screws.
MAINTENANCE
1. Circuit Breaker - The unit is protected by a circuit breaker on the left side of the unit. In the unlikely
event there is an overload condition the breaker may trip. If this happens reset the breaker by pushing
on the white button until it clicks into place.
2. Motor Brushes - This long life motor does not have replaceable brushes. Replacement motors can
be ordered as P/N 97041.
RECOMMENDED USE
The Pure Breeze FX Fume Extractor is designed for use in collecting toxic or odorous fumes and should
not be used for collection of large particulates or wet substances. For an exhaustive list of chemicals
compounds that are best ltered with this unit, please visit the Pure Breeze FX Fume Extractor page at
Vaniman.com.

PURE BREEZE FX
FUME EXTRACTOR
25799 Jeerson Ave. ● Murrieta, CA 92562 ● 1-800-VANIMAN (826-4626) ● www.vaniman.com ● [email protected]
10380-12/21-PAGE 4
REPLACEMENT PARTS
The following items are oered to enhance the usage of your quality Vaniman unit. These items can be
order through your supplier using the following part numbers. Please feel free to contact Vaniman Mfg.
Co. with questions regarding our products.
Fig. # Description Part Number
1 Replacement Motor 97041
2Motor Gasket 3396
3Front Cover Thumb Screws, Pair 2984
4 Remote Plug Air Switch 1442
5Power Switch ON/OFF 2904
6 Power Connector 3010
7Rotary Switch Knob 1471
8 Rotary Switch 3019
9Transformer 1419
10 Circuit Breaker, 15Amp 2869
11 Charcoal Odor Filter 97095
12 HEPA Filter 96097
Not Shown Power Cord 1422
1
2
34
5
6
7
8
9
10
11
12
Table of contents
Other Vaniman Scrubber manuals
Popular Scrubber manuals by other brands

HealthyAir
HealthyAir HA-CMSC-G2-W-AEA owner's manual

Viper
Viper FANG 20HD-EU user manual

LYSON
LYSON W4090 manual

COMAC
COMAC ANTEA 50 B Use and maintenance manual

Nilfisk-Advance
Nilfisk-Advance Viper Fang 15B Guide to Machine Operation and Maintenance

Ghibli & Wirbel
Ghibli & Wirbel RAPID 40 SM 55 Touch Faq