SHUNT CALIBRATION
The Model 318 provides a unique internal shunt calibration resistor that allows field calibration to be performed by
electrically simulating pressure. The connector shall be connected as shown in Fig. 5. The nylon access screws
should be removed and the pressure port shall be vented to zero pressure. For transducers measuring absolute
pressure, a full vacuum must be applied to the pressure port. Connecting (shorting) the calibration terminals (Pins 3
and 4) unbalances the transducer bridge by an amount equivalent to the application of approximately 100% of full-
scale pressure. See the performance certificate, which was supplied with the transducer, for the exact milliamp
output and adjust the span control accordingly. If the certificate is unavailable, the milliamp output can be calculated
by using the following formula: mA = ((Cal pressure / pressure range) x 16) + 4. Thus, following adjustment of zero
output for zero pressure, the calibration circuit may be activated and the gain or span adjustment screw can be turned
until the correct calibration output appears at the output terminals. The calibration terminals should then be
disconnected and the zero adjustment reconfirmed. Note - the calibration circuit shall be deactivated during normal
operation. The access screws should be reinstalled after calibrating the transducer.
ACCESSORIES - OPTIONAL EQUIPMENT
Mating Electrical Connector (118, 318) ............……...................…................................. 10MSWK0204
Mating Electrical Connector (218) ......................….......................….............................. 10MSWK0206
Mating Electrical Cable Assembly (118, 318) ...….......................…................................ 230005.001
Mating Electrical Cable Assembly (218) .........…..........................…............................... 230006.001
Mounting Bracket .................................................................................……........…........ 080002.001
Protective Cover .................................................................................…….............…..... 03XL-20RBC
Digital Meter ...................................................................................................…….....…. Consult Factory
MAINTENANCE AND REPAIR
All Viatran transmitters have been designed to function free from routine or scheduled maintenance. Simple cleaning
of the electrical connector, pressure port threads and pressure cavity on an as needed basis will provide many years
of satisfactory performance. Protecting the product from continued exposure to moisture or fluids at the electrical
connection, breather area (model dependent) will eliminate premature internal failure of the product. Generally any
time the product is removed from service the connector and pressure port threads should be cleaned and the
pressure cavity flushed with a 316SS and/or O-Ring (material model dependent) compatible cleaner to prevent media
buildup. During the cleaning process only a soft, lint-free cloth is recommended. Never use a coarse of stiff bristle
brush to clean media from the diaphragm surface.
It is suggested that the calibration be verified on a usage dependent schedule. If the product is in continuous service
7 days a week, then calibration verification may be necessary every 6 to 8 months. If the product is in a lab test
environment a more lenient verification schedule would be appropriate. In all instances the performance of the
product will depend on the individual application or process in which it is installed. More continued usage would
require a shorter period between calibration verification and product maintenance.
If a product is perceived to be exhibiting problems, it can be returned to Viatran for analysis and/or repair. It is
suggested that only Viatran personnel attempt repair of the product. Any damage resulting from customer
disassembly would result in a loss of coverage under the warranty policy. All Viatran products are able to be repaired
at minimal cost if simple cleaning and precautions are taken in the handling and application of the product. Older
products returned for repair are updated to current specifications unless the repair cost would outweigh the cost of a
new replacement model. Products returned for repair should include information on the person to contact for repair
quote approval, the individual to contact if Viatran’s technical staff requires additional information during analysis, and
a brief description of the problem associated with the product’s failure.
WARRANTY
Viatran Corporation warrants that its products shall be free from defective workmanship and/or material for a period of
twelve (12) months from the date of shipment, provided that Viatran’s obligation hereunder shall be limited to
correcting any defective workmanship and/or replacing any defective material f.o.b. factory. No allowance will be
made for any expense incurred for correcting any defective workmanship and/or material without written consent by
Viatran. This warranty is in lieu of all other warranties expressed or implied.
REPAIR INSTRUCTIONS
Viatran’s Transmitters are designed to be easily repaired and recalibrated if necessary. If a failure occurs, the
transmitter should be returned to the factory for inspection and testing. Please contact the Customer Support
Department at 1-800-688-0030, for a return tracking number and/or a repair cost estimate. A nominal inspection fee
is charged on all units returned to the factory which are not subsequently repaired.