W&H Osstell ISQ User manual

module
Si-SQ
Instructions for Use
Edition USA

2
Contents
Symbols...................................................................................................................................................................................................... 3 – 5
1. Introduction.........................................................................................................................................................................................6 – 7
2. Electromagnetic compatibility (EMC) ....................................................................................................................................................8
3. Equipment supplied..................................................................................................................................................................................9
4. Safety notes................................................................................................................................................................................... 10 – 11
5. Description ..............................................................................................................................................................................................12
6. Start-up – General...................................................................................................................................................................................13
7. Operation ....................................................................................................................................................................................... 14 – 16
8. Operation documentation.............................................................................................................................................................. 17 – 18
9. Hygiene and maintenance ............................................................................................................................................................ 19 – 23
10. W&H accessories and spare parts ........................................................................................................................................................24
11. Servicing.........................................................................................................................................................................................25 – 26
12. Technical data ................................................................................................................................................................................27 – 28
13. Disposal...................................................................................................................................................................................................29
Explanation of warranty terms.......................................................................................................................................................................31
Authorized W&H service partners..................................................................................................................................................................32

3
Symbols in the Instructions for Use
WARNING!
(if persons could be injured)
ATTENTION!
(if property could be damaged)
General explanations,
without risk to persons or property
Call customer serviceSterilizable
up to the stated temperature

4
Symbols on the Osstell ISQ module
Date of manufacture
Article number
Serial number
Direct current
Supply voltage
of the control unit
Type B applied part (not suitable
for intracardiac application)
Do not dispose of with
domestic waste
Consult Instructions for Use
The medical device with reference to electrical safety, mechanical safety
and fire prevention conforms to UL 60601-1:2006,
CAN/CSA-C22.2 No.601.1-M90:2005, CAN/CSA-C22.2 No. 60601-1:2008,
ANSI/AAMI ES 60601-1:2005 25UX (Control No.)
DC
V
Data Matrix code
for product information
including UDI (Unique Device
Identification)
CE mark
with identification number
of the Notified Body
XXXX

5
Symbols on the packaging
This way up
Caution! Federal law restricts this device to sale
by or on the order of a dentist, physician or any
other practitioner licensed by the law of the state
in which he or she practices to use or order the use
of the device.
Fragile, handle with care
Keep dry
Permitted temperature range
Humidity,
Limitation
»Der Grüne Punkt« (The Green Dot)
trademark of Duales System
Deutschland GmbH
Trademark of RESY OfW GmbH
for identification of recyclable
transport and outer packaging
of paper and cardboard
-40 °C (-40°F) Min.
+70 °C (+158°F) Max.
8 %
80 %
CE mark
with identification number
of the Notified Body
XXXX
Data structure in accordance with
Health Industry Bar Code
Data Matrix code
for product information
including UDI (Unique
Device
Identification)

6
1. Introduction
For your safety and the safety of your patients
These Instructions for Use explain how to use your product. However, we must also warn against possible hazardous situations.
Your safety, the safety of your team and, of course, the safety of your patients, are of paramount importance to us.
It is therefore essential that you observe the safety notes on page 10 – 11.
Intended use
Osstell ISQ module is indicated for use in measuring the stability of implants in the oral cavity and craniofacial region. Osstell
ISQ module can add important information to the evaluation of implant stability and can be used as part of an overall treatment
evaluation program. The final implant treatment decisions are the responsibility of the user.
Misuse may damage the Osstell ISQ module and hence cause risks and hazards for patients, users and third parties.
Qualifications of the user
The Osstell ISQ module is intended for use by suitably qualified and trained medical, technical and specialist staff only.
We have based our development and design of the Osstell ISQ module on the "physician" target group.

7
Introduction
Production according to EU Directive
The medical device complies with the regulations of Directive 93/42/EEC.
Responsibility of the manufacturer
The manufacturer can only accept responsibility for the safety, reliability and performance of the Osstell ISQ module
when compliance with the following instructions is ensured:
> The Osstell ISQ module must be used in accordance with these Instructions for Use.
> The Osstell ISQ module has no components that can be repaired by the user. Assembly, modifications or repairs
must only be undertaken by an authorized W&H service partner (see page 32).
> Unauthorized opening of the equipment invalidates all claims under warranty and any other claims.

8
2. Electromagnetic compatibility (EMC)
Medical electrical equipment is subject to particular precautions with regards to EMC and must be installed and put into
operation in accordance with the EMC notes included.
W&H guarantees the compliance of the device with the EMC Directives only when used with original W&H accessories and
spare parts. The use of accessories and spare parts that have not been approved by W&H may lead to increased emission
of electromagnetic interference or to reduced resistance to electromagnetic interference.
You can find the current EMC manufacturer’s declaration on our website at http://wh.com
or, alternatively, you can also request a copy directly from the manufacturer.
HF communication equipment
Do not use any portable and mobile HF communication equipment (e.g., mobile telephones) during operation.
These may affect medical electrical equipment.

9
3. Scope of delivery
REF Description
30210000 Osstell ISQ module
07849900 TestPeg
077218 0 0 Universal support
07460300 SmartPeg mount

10
4. Safety notes – general
> Before using the Osstell ISQ module for the first time, store it at room temperature for 24 hours.
> Check the Osstell ISQ module and the measuring probe with cable for damage and loose parts every time before use.
> Do not operate the Osstell ISQ module and the measuring probe with cable if it is damaged.
> Perform a test measurement with the TestPeg prior to every use.
> The responsibility for the use and timely shutdown of the system lies with the user.
> Ensure that it is possible to complete the operation safely should the units or instruments fail.
Hygiene and maintenance before using for the first time
> Sterilize the measuring probe with cable.
Improper use
In addition to unauthorized assembly, installation, modification of or repairs to the Osstell ISQ module and measuring probe
with cable, dental handpieces and non-compliance with our instructions, improper use will invalidate all claims made under
warranty
or otherwise.
The Osstell ISQ module is classed as “conventional equipment” (closed equipment without protection against the ingress
of water).

11
Safety notes – general
Danger zones M and G
The Osstell ISQ module with the measuring probe is not suitable for use in potentially explosive atmospheres or with
potentially explosive mixtures of anaesthetic substances containing oxygen or nitrous oxide according to IEC 60601-1/ANSI/
AAMI ES 60601-1.
The Osstell ISQ module with the measuring probe is not suitable for use in oxygen-enriched atmospheres.
Zone M, also referred to as the “medical environment”, includes the part of a room in which potentially explosive
atmospheres may occur as a result of the use of analgesics or medical skin cleaning or disinfectants, but only
in small quantities and only for a short time.
Zone M comprises a truncated pyramid below the operating table which is tilted outwards at a 30° angle.
Zone G, also referred to as an “enclosed medical gas system”, comprises not necessarily fully enclosed cavities in
which permanent or temporary potentially explosive mixtures may be generated, supplied or used in small quantities.

12
5. Description
TestPeg
Osstell ISQ module
Measuring probe connection
USB cable Measuring probe with cable
SmartPeg mount

13
6. Start-up – General
Push in the Osstell ISQ
module until it locks audibly.
Pay attention to
the positioning
of the USB cable!
Connect USB.
Connect measuring probe.
Pay attention to
the positioning!

14
7. Operation
> The TestPeg is for testing only and for teaching in the function.
> You can purchase SmartPegs from smartpegs.wh.com or osstell.com.
> SmartPegs are for single use only.
> SmartPegs are available for a range of different implant systems and can be used in combination
with all conventionally available implants.*
> Ensure that the sterile chain is not broken.
> Only use SmartPegs with intact packaging.
Select program 6 (ISQ).
The ISQ program (P6) always appears after the last program.
Pull a thread through the SmartPeg mount.
Tie the thread to your wrist to prevent loss.
Insert the SmartPeg into the SmartPeg mount.
The SmartPeg is magnetic and is held in place by SmartPeg mount. Check that it is retained securely.
Insert the SmartPeg into the implant or abutment and screw it in hand-tight using the SmartPeg mount.
Do not overtighten the SmartPeg or the SmartPeg thread may be damaged.
* For further information, please contact an authorized W&H service partner (see page 32) or visit osstell.com

15
Operation
Press the foot control pedal once to start the measurement.
Press the foot control pedal again to stop the measurement early.
Hold the measuring probe about 3 to 5 mm from the tip of the SmartPeg until the
measured value is displayed.
Measure in both the mesiodistal direction (a) and the buccolingual direction (b).
Do not measure from above.
Repeat 5 and 6 to perform multiple measurements.
The measured value is underlined in colour and confirmed by a signal tone.
Remove the SmartPeg with the SmartPeg mount.
ISQ40 60 80
a
b

16
Measurement result*
The measurement result can be used as part of a general assessment program. The user bears the ultimate responsibility
for the decision for implant treatment.
This is a summary of scientific data and therefore does not represent an official recommendation.
To monitor osseointegration, measurements should be taken after implant insertion and before restoration of the implant.
Scientific studies can be found here www.osstell.com
ISQ value
The resonance frequency as a measure of implant stability is calculated from the oscillation frequency of the SmartPeg.
The results of this calculation are displayed as the ISQ value. The scale from the ISQ value ranges from 1 to 100.
* For further information, please contact an authorized W&H service partner (see page 32) or visit osstell.com
Operation
Low Stability Medium Stability High Stability
60 70
Red
ISQ <60
Implant at risk –
monitor ISQ
Yellow
ISQ 60 – 65
Full splint (immediate
loading)
2-stage
Traditional loading
Yellow
ISQ 65 – 70
Partial case
1- or 2-stage
Early loading
Green
ISQ >70
Single case
1-stage
Immediate loading
65
Indication
Surgical protocol
Restorative protocol

17
The ISQ values can only be documented in program 6.
Documentation must be activated or deactivated for each program.
A USB stick is required to save the documentation.
Never remove the USB stick during the measurement.
Record documentation
> Connect USB stick
Icon appears
> Enter ID
> Enter date
> Enter tooth quadrant
> Select tooth
> Confirm selection
Documentation starts when the motor starts.
When the motor stops, the ISQ value appears. It is automatically saved to the USB stick.
8. Operation – documentation

18
Edit documentation
A text file (csv) and a pdf file are saved on the USB stick.
The text file can be opened in Microsoft® Excel* for editing.
The pdf file can be opened in Adobe® Reader®**.
* Microsoft® Excel® is a registered trademark of the Microsoft® Corporation in the United States of America and/or other countries.
** Adobe® Reader® is a registered trademark of Adobe Systems Incorporated in the United States of America and/or other countries.
Operation – documentation

19
9. Cleaning and Sterilization
Follow your local and country-specific laws, directives, standards and guidelines for cleaning, disinfection and sterilization.
> Wear protective clothing.
> Clean and disinfect the measuring probe with cable immediately after every treatment!
> Sterilize the measuring probe with cable after cleaning and disinfection.
Medical device
> Clean and disinfect the medical device immediately after each patient, to flush out any liquid (such as
blood, saliva etc.) and to prevent debris from settling on the internal parts.
> After manual cleaning, disinfection, and lubrication, you must sterilize the (wrapped) medical device
using a steam sterilizer.
> Sterilize the reassembled medical device after each patient, following manual or automated cleaning,
disinfection, and lubrication.
> Sterilize the coolant tube.

20
Cleaning and Sterilization
> The Osstell ISQ module is not approved for mechanical cleaning (thermo washer disinfector) and sterilization.
> Do not immerse the Osstell ISQ module in water or clean under running water.
Pre-disinfection
> If heavily soiled, clean first with disinfectant cloths.
Only use disinfectants which have no protein-fixing effect.
Manual disinfection
W&H recommends wipe disinfection.
> Use only disinfectants which do not contain chlorine and which are certified by officially accredited institutes.
> Note the manufacturer’s specifications for the use of the disinfectant.
Other manuals for Osstell ISQ
1
Table of contents
Other W&H Control Unit manuals