
F−1031, Section 2414 Page 7 of 8
Calibration
AquistFoam Pump
The Aquistsystem is able to be field calibrated using the con
trol function buttons on the operating interface terminal (OIT),
see Figure 8.
Only units of measure, water flow and the foam percent
age default can be field set.
A. Entering Calibration Mode
1. Entering calibration is accomplished by using the con
trol unit functions buttons.
2. To enter this function, press and hold the SELECT but
ton for minimum of five (5) seconds. The OIT will dis
play “5 SEC”.
3. Continue to hold the SELECT button.
4. While the SELECT button is depressed, push the UP
and DOWN arrow buttons at the same time. The display
will show CAL for two (2) seconds and then display F1.
5. Use the UP button to advance to the next parameter
(F2, F3, F4).
6. Use the DOWN button to back up to a previous para
meter (F2, F1).
7. To edit a parameter, press the SELECT button at the
associated Fn prompt.
B. F1 - Default Mix Percentage
1. At the F1 prompt, press the SELECT button.
2. Use the UP and DOWN arrow buttons to select the de
fault FOAM %. The only foam % selections will be .1%,
.2%, .3%, .4%, .5%, .6%, .7%, .8%, .9%, 1.0%, 3.0%
and 6.0%.
3. When finished, press the SELECT button to save the
displayed value.
C. F2 - Units of Measurement
The units of measurement must be selected prior to any
calibration functions. Units of measurement are U.S. Gal
lons, Imperial Gallons and Liters.
1. Push the SELECT button at the F2 prompt. The display
will show the current setting.
2. Use the UP and DOWN arrow buttons to select one of
the following measurement units:
·GAL will select U.S. Gallons
·IPL will select Imperial Gallons
·LTR will select Metric Liters
3. Pushing the SELECT button will save the displayed
selection. This will allow for calibration of the water flow
in the units of measurement selected.
D. F3 - Flow Rate Calibration
1. Press the SELECT button at the F3 prompt. The
current measured flow rate will be displayed. Es
tablish a flow using an accurate flow measuring
device.
2. Once this is established, press the UP or DOWN but
ton to advance to the calibration step. If the displayed
flow rate is correct, press the SELECT button to return
back to the F3 prompt. If there is insufficient flow, the
display will show LoFL for two (2) seconds, then return
back to the F3 prompt.
3. The display will now display 50 (starting value) or the
last calibration value entered.
4. Use the UP and DOWN arrow buttons to change the
displayed value to match the actual flow rate according
to the flow measuring device. If the button is only
pressed momentarily, the display increments or decre
ments by one (1). If the button is held, the display will
increment or decrement by five (5).
5. Once the correct flow is achieved, pushing the SELECT
will save the setting and return to the F3 prompt.
E. F4 - System Lock
System can be locked out after calibration and setup to
avoid accidental changing.
WARNING
A locked out system can only be unlocked at a quali
fied Waterous Service Center.
1. Press the SELECT button at the F4 prompt and “EnAb”
(enabled) will be displayed.
2. Pressing the UP button will toggle between “LoC”
(locked) and “EnAb” (enabled). Settings will not be per
manent until the calibration settings are saved.
3. Press the SELECT button and “F4” will be displayed.
4. Pressing the FOAM button saves the settings, locks the
system and cannot be unlocked except at a qualified
Waterous Service Center.
F. Saving Calibration
To save all the calibration settings to flash memory, press
the FOAM button while at the Fn (n=1-4) prompt. The data
will be stored and the unit will return to normal operation.
Figure 8. Operator Interface Terminal
IL3437