WAVECONTROL SMP2 User manual

SMP2
Portable Electromagnetic Field Monitoring System
User’s Manual
V1.11
WAVECONTROL
C/ Pallars 65-71 · 0 01 Barcelona (SPAIN)
Tel. (+34) 933 20 055 Fax. (+34) 933 20 056
www.wavecontrol.com

INDEX
1 WARNINGS AND PRECAUTIONS.........................................................................................................3
2 INTRODUCTION...................................................................................................................................
3 FIRST STEPS.......................................................................................................................................... 5
GETTING STARTED............................................................................................................................... 6
4.1 Overview..........................................................................................................................6
4.2 Field probes......................................................................................................................
4.2.1 Connector type.....................................................................................................................................................................
4.2.2 Screw connector....................................................................................................................................................................
4.2.3 Push-pull connector...............................................................................................................................................................9
4.3 External connections.........................................................................................................9
4.3.1 USB.......................................................................................................................................................................................9
4.3.2 Fibre optic...........................................................................................................................................................................10
5 SMP2 – INITIAL USE........................................................................................................................... 10
5.1 Switching the apparatus on and off.................................................................................10
5.2 Main screen....................................................................................................................10
5.3 Menu screen...................................................................................................................12
5.4 Measurement configuration screen.................................................................................13
5.5 Measurement log screen.................................................................................................14
5.6 GPS screen.....................................................................................................................14
5.7 Settings screen...............................................................................................................14
6 FUNCTIONS........................................................................................................................................ 15
6.1 Broadband - time domain measurements........................................................................15
6.1.1 Measurements according to ICNIRP procedures...................................................................................................................15
6.1.2 Low frequency time domain measurement with WP50 or WP400.......................................................................................16
6.2 Selective - frequency domain measurements....................................................................16
6.2.1 Scenarios.............................................................................................................................................................................17
6.3 Time domain weighting..................................................................................................1
6.4 Alarm............................................................................................................................. 1
6.5 GPS................................................................................................................................ 1
7 BATTERY INFORMATION................................................................................................................... 19
7.1 General information........................................................................................................19
7.2 Recharging.....................................................................................................................19
8 SOFTWARE......................................................................................................................................... 20
.1 Installing the software.....................................................................................................20
.2 System requirements.......................................................................................................21
.3 Using the PC software....................................................................................................21
.3.1 Measurements screen..........................................................................................................................................................21
.3.2 Screenshots screen..............................................................................................................................................................22
.3.3 Information screen..............................................................................................................................................................22
.3.4 Spreadsheet exporting.........................................................................................................................................................22
.3.5 Control centre screen..........................................................................................................................................................23
.4 Live screen......................................................................................................................24
9 FIRMWARE UPDATING...................................................................................................................... 2
10 MAINTENANCE................................................................................................................................ 25
10.1 Cleaning the apparatus and the screen..........................................................................25
10.2 Handling the equipment...............................................................................................25
10.3 Periodic re-calibrations..................................................................................................25
11 TECHNICAL SPECIFICATIONS...........................................................................................................26
12 WARRANTY...................................................................................................................................... 27
ANNEX 1: WP 00 PROBE SPECIAL FUNCTIONS..................................................................................28
ANNEX 2: COMMUNICATION PROTOCOL..........................................................................................30
- 2 -

Congratulations for your acquisition of a Wavecontrol SMP2.
This manual provides you with instructions for using and handling your SMP2.
Please read this Manual carefully before using your SMP2.
1 WARNINGS AND PRECAUTIONS
Review this manual and become familiar with all the instructions for using and handling
your SMP2.
Your SMP2 is a portable apparatus with a rechargeable battery. Before taking any
measurement, make sure that the battery is sufficiently charged for the time that you
need for the measurement. If it is not, recharge the battery before you begin.
Your SMP2 is not watertight and it should be protected from moisture, such as rain.
Adjustment, maintenance, or repair of the equipment must be performed only by
qualified personnel. Please contact Wavecontrol or your sales contact.
Never overtighten when screwing the SMP2 on the tripod since the thread engagement
could break down.
Environmental information
Disposal of your old product:
Your product is designed and manufactured with high quality materials and components, which
can be recycled and reused.
When this crossed-out wheeled bin symbol is attached to a product it means the product is
covered by the European Directive 2002/96/EC.
Please inform yourself about the local separate collection system for electrical and electronic
products.
Please act according to your local rules and do not dispose of your old products with your normal
household waste. The correct disposal of your old product will help prevent potential negative
consequences for the environment and human health.
- 3 -

2 INTRODUCTION
The SMP2 Portable Electromagnetic Field Monitoring System can measure electromagnetic field
levels generated by any source within the frequency range of the probe attached to the system.
It combines 2 instruments in 1: Broadband measurements and selective measurements (FFT),
including % direct EMF exposure assessments to limit lines by using the weighted peak exposure
(WPE) technique.
The SMP2 has been developed to meet demanding customer requirements in the RF safety
measurements domain. Its versatility makes it a perfect measurement tool for many different
sectors and industries such as Railway, Power, Manufacturing, Chemical, Medical, Defence, RFID
and EAS, Telecommunications, etc.
Those measurements can be made quickly and easily anywhere because the SMP2 is portable
and simple to handle and use.
The SMP2 can sample the field level continuously and store that data automatically in its built-in
memory at user-configured intervals with a capacity of up to 1 000 000 samples.
The data stored in the memory can be easily downloaded to a computer via a USB or fiber optics
port for later processing.
The SMP2 is also available with a built-in GPS receiver to allow each measurement to be
associated automatically with the geographic location where it was taken.
- 4 -

3 FIRST STEPS
This section contains instructions for unpacking and checking the equipment when you receive it.
1) When you receive the equipment, check the packaging to see if it has been damaged in any way.
Make a description of any damage on the delivery note before you sign it.
2) Check that all the components of the SMP2 are there. The SMP2 is delivered in a compact
briefcase that should contain the following components:
1. SMP2
2. 1 or more electromagnetic field probes
3. Battery charger
4. USB key containing the User’s Manual and SMP2 software
5. USB – mini USB cable
6. Fibre optic link (optional)
If any of those components is missing, please contact Wavecontrol.
Figure 1: Case
- 5 -
SMP2
and USB cable
below
Battery
charger
Field Probe
Flash drive
Field Probe
Optional
fibre optic
link below

GETTING STARTED
4.1 Overview
The front panel of the SMP2 contains the user interface, made up of the screen and keypad.
Figure 2: Front
The keyboard presents standard buttons, 4 virtual buttons (below the screen) and 1 ON/OFF
button.
: These arrows allow you to move through the menu options and the data
stored in memory.
: allows to validate an option or enter in a menu.
: always allows to come back to the main screen (measuring screen).
: allows to get back to an earlier menu or to cancel a selection.
: allows to take print-screens anywhere in the program.
- 6 -

The top of the SMP2 holds the field probe connector.
Figure 3: Top
Probe connector: (see 4.2)
On / Off button: switches the equipment on and off.
Status LED: when the unit is off, red means charging, green charged. When the unit is on,
green LED should be always on.
Light sensor: allows to control the automatic screen back-light intensity depending on
ambient light.
The bottom of the SMP2 holds the other connections of the equipment.
Figure : Bottom
Mini USB connector and fibre optics: allows connection to a PC in order to:
oDownloading data stored in the SMP2 memory.
oUpdating firmware for the equipment.
oControlling the unit remotely
Battery charger connector: allows the SMP2 battery to be recharged using the
charger supplied with the equipment.
¼’ screw connector: takes a ¼’ screw, the standard screw used for most tripods,
allowing the SMP2 to be mounted on a tripod.
- 7 -
Charger
connector
Mini USB
connector
Tripod
¼' screw
Probe
connector
On/off
button
Fiber optic
connector
Status
LED Light
sensor
Hidden
GPS
antenna

4.2 Field probes
The SMP2 is a highly versatile EMF measurement device that requires connection of the proper
field probe to obtain the best results for each application. Only Wavecontrol field probes will
work with the device.
If the device is switched on with no field probe attached, it will not show any field value and
therefore no measurement will be saved to memory. However, saved data and GPS positioning
functions will be available.
The SMP2 automatically detects the type of field probe and its calibration parameters.
.2.1 Connector type
There are two types of connector: push-pull and screw.
Figure 5: Screw connector Figure 6: Push-pull connector
.2.2 Screw connector
4.2.2.1 How to connect the field probe
Connect the field probe by inserting the male connector into the female connector with the tab
aligned. Once fully inserted, screw the connector until it is snug, without forcing it.
Figure 7: How to connect the field probe
4.2.2.2 How to disconnect the field probe
First unscrew the probe connector, then pull the connector to remove it.
Important: Pull the connector.
Do not try to twist or turn the probe tube.
- 8 -
Sleeve to screw on
after inserting the
probe
Tab

.2.3 Push-pull connector
4.2.3.1 How to connect the push-pull probe
With a push-pull connector, the field probe is connected by inserting the male connector into the
female connector of the SMP2 until it stops. The polarity of the connector must be observed and
so the red marks on the male and female connectors must line up.
Figure 8: How to connect the field probe
4.2.3.2 How to disconnect the field probe
You do not need to switch the device off before you disconnect the field probe. Pull the
connector of the probe firmly with one hand (pulling with the thumb) while you hold the SMP2
with your other hand as shown in the pictures below:
Figure 9: How to disconnect the field probe
Important: Pull the connector. Don't pull any other part of the probe!
Don't try to twist or unscrew the probe.
4.3 External connections
.3.1 USB
A USB to mini USB cable is delivered with the SMP2. After installing the SMP2 software (First
install the software see .1) just plug the cable to the SMP2 and to the computer. You will be
able to download stored measurements to the computer or update the SMP2 firmware.
Note: it is not recommended to perform measurements while the USB is connected, it could affect the
quality of the readings.
- 9 -

.3.2 Fibre optic
As an option a fibre optic link is available in order to
remotely control and see live readings of the instrument
from a PC without interfering with the measurements.
To connect the fibre optic link:
•plug the fibre to the SMP2 (respect the colour
code, grey side is looking up)
•plug the fibre to the USB to fibre converter
(respect the colour code)
•plug the USB of the converter to the PC.
Note: no need to install additional software, everything has been already installed with the main
software.
Note2: in case the connector does not fit into a small waveguide, you can brea the connector in 2
parts.
5SMP2 – INITIAL USE
The SMP2 is designed for simple and intuitive operation with 2 basic functions:
-Time or frequency domain field measurement.
-Storage of results in memory.
Several additional functions make the SMP2 a very complete field measurement system.
The SMP2 is controlled via the keypad and easy access to data is available through the different
screens.
5.1 Switching the apparatus on and off
The apparatus is switched on and off with the On / Off button located on the upper part of the
keyboard.
To power on the unit, click on the On/Off button. The status LED will stay in red for some
milliseconds before turning green. A welcome screen is displayed until the software completely
boots and shows the main screen.
To power off the unit, click once on the On/Off button. A pop-up will appear on the screen to
confirm the command. F1 and F2 buttons will confirm the action, F3 and F4 buttons will cancel
it. You can also select the option with the right and left arrows and click on OK to confirm.
Note: All device’s configurations (alarm level, units, type of measurement, measurement interval, etc.)
is saved when it is switched off and recovered automatically when it is switched on.
5.2 ain screen
The main screen provides instantaneous and continuous information on the electromagnetic field
being measured.
Date and time: date and time of the apparatus.
Icons: Several icons give some information on the state of the unit
Alarm is activated USB is connected Fiber connection available
Battery : battery charge level.
Battery discharged
- 10 -
Breakable part
see Note2

Battery fully charged
robe information: serial number of the field probe attached to the apparatus and its
frequency information.
Figure 10: Main screen
•Measurement information:
• Maximum, minimum and average values: these values are initialized when beginning
a new log or can also be initialized with the virtual button “Reset”. Average value is
updated depending on the type of averaging and interval chosen in the measurement
configuration.
• Logging and average parameters:
1. Average windows time
2. type of interval (sliding or arithmetic)
3. logging time
Field information: (updated twice per second)
• Total field: instantaneous total field level. Obtained, for three-axis probes as:
•The quadratic sum of each axis field (for field intensity units):
222
ZYXTot EEEE
•The sum of each axis field (for power density units):
ZYXTot
SSSS
- 11 -
Probe information
Measurement information
Field information
Graph
GPS information
Functions
Battery
Date and time
Icons

• Field unit: unit of measurement of the field value shown on the display. This unit
applies to the total field, the field on each axis and the maximum, minimum and average
values.
• Field on each axis: the field measured on each axis for isotropic three-axis probes.
Updated twice per second.
Graph:
In the time domain: It shows a graphical view of the temporal evolution of the field. It represents
the last 112 seconds of measurement.
In the frequency domain: it shows the frequency information of the signal (horizontal axes value
depend on the span). Also a vertical line represents the position of the cursor (with the
associated information on the right top side of the graph)
G S information:
Positioning information provided by the GPS receiver. This information is only displayed when the
GPS is switched on.
• lat: Latitude
• Lon: Longitude
On the right side a colour indicates the quality of the position:
red: bad orange: average green: good
Functions:
These menus depend on the probe connected but general functions are as follows:
-Menu: access to the menu screen.
-Log: start a new measurement.
-Units: change the current units (this will reset the min, max, average and graph). With the
up and down arrow you can select the unit. By pressing OK it will select the highlighted
unit.
-Reset: it reset the min, max, average and graph
-Keypad Arrows in frequency mode:
-right-left arrows shift the cursor by one step
-up-down arrows shift the cursor to the next right
or left peak found in the FFT:
Note: When there are more than 4 functions available, the button
on the right allows to show more functions available.
5.3 enu screen
This is the main screen to access sub menus.
•Configure measurement (allows to change all the parameters related to
measurement)
•Measurement log (allows to see the stored measurements)
•G S (allows to change GPS parameters)
•Settings (allows to change general parameters of the unit)
•Messages. (in case some events are available this menu will appear)
- 12 -

5.4 easurement configuration screen
This screen is used to configure different parameters related to measurement behaviour. It
appears as a menu with a scrolling bar. The parameters to be configured are shown on the left
and the different options for each parameter are shown on the right.
The keypad is used to move along the menu. The process for changing the value of a parameter
is:
-Run through the different parameters using the arrows .
-To change a parameter, press Enter when the cursor is on that parameter. A pop-up will
appear showing all the possible values.
-Run through the different values using the arrows and click on OK to select the desired
value. (you can cancel with the back button)
-When finished you can either press again back to return
to the menu screen or press the home button to come
back to the main screen
The following parameters can be configured:
Standard: List of preselect standards. This will affect
other parameters that will not be editable. For
example: if you choose ICNIRP standard then all the
parameters in “Measurement behaviour” will not be
editable as they are defined by the standard.
Measurement time: allows to define a maximum
time for the measurement. Available values are: Not
limited, 1min, 5min, 6min, 10min, 15min, 30min,
1h, 2h, or custom from 1sec to 100 hours.
Log interval: used to select the time interval for
storage of data in memory (see section 6.1). A datum
will be stored every ‘Log interval’ seconds (or minutes).
The available intervals are 0.5s, 1s, 5s, 10s, 15s, 30s, 1min and 6min.
Average type: two types of average are available: arithmetic (a new average is
calculated every average interval) , sliding (a new average is calculated every new sample)
Average interval: used to select the time interval for calculation of the average. The
available intervals are 10s, 15s, 30s, 1min, 2min, 5min, 6min, 10min, 15min and
30min.
Schedule measurement: allows to program a scheduled measurement. A delay up to 24
hours can be set. Even with the unit power off, this feature works.
Wait time before start: delay before the measurement really starts (it allows the
technician to separate himself from the equipment. Available values are No delay, 5s,
10s, 15s, 20s, 30s, 1min.
Shut down after measurement: useful option to power off the unit when
measurement is done. It avoids having to wait to power off the unit manually.
- 13 -

Screenshot on last sample: when this is activated it will allow to take a screenshot on
the last measurement. When using for example the ICNIRP standard it will allow to have
an image with the 6 min average results. This image will be available to be exported
directly into the report with the SMP2 software.
Save measure name: permits to save a name for each measurement. In case it is
activated, a pop-up will appear before the measurement starts in order to indicate a
name.
Delete all measurement: will delete all measurements as well as all the screenshots
related to these measurements. Depending on the number of data in memory this
operation may last some time. It is recommended to use this function from time to time in
order to free space for new measurements.
5.5 easurement log screen
This simple screen will show all available measurements. The arrows are used to select
the next or previous measurement. The arrows are used to select the next or previous
measurement located in the next page.
From this menu you can use the virtual button “view” to view a specific measurement (can be
done by clicking on OK), “delete” to delete a single measurement or “delete all measurements”
to erase all data in memory (the last action does the same as the “Delete all measurements” in
the Measurement configuration screen.
5.6 GPS screen
This screen has the same layout and behavior as the measurement configuration screen. It allows
to modify the GPS parameters. Those parameters are:
Format: allows to select 3 standard format for the latitude and longitude. You can
choose between:
• degrees minutes seconds: example: 40° 26 46 N′ ″ 79° 5 56 W′ ″
• degrees decimal minutes: example: 40° 26.7671 N′ 79° 5 .9331 W′
• decimal degrees: example: +40.446 -79.9 2°
Turn on at start up: defines the GPS state when the unit starts up.
Turn on for current session: defines the GPS state only for this session. That means that
next time you switch on the unit the GPS will be in the state defined by “Turn on at start
up” parameter.
5.7 Settings screen
This screen has the same layout and behavior as the measurement configuration screen. It allows
to modify the general parameters of the SMP2.
The following parameters can be configured:
Alarm: allows to activate or deactivate the alarm. When activated, you can define the
alarm threshold and unit for electric and magnetic field. (the arrows are used to
select the figure to be changed and the arrows are used to change the figure value.)
Date: allows to change the date of the system
Time: allows to change the time of the system
- 14 -

Zero based time plots. Allow to scale the graph between 0 and the maximum of the
measurement or between the minimum and maximum to have a precise view of the time
variation of the field.
Graph timespan: set the duration of the representation of the graph up to 12 min.
Language: set the desired language
Adjust backlight automatically: if enabled the SMP2 will adjust backlight depending on
ambient light.
Backlight level: If automatic adjustment is disable, manual level can be selected from 1
to 4 (4 is the brightest)
Screensaver delay: defines the time before screen power off. When the screensaver is
active the instrument will not stop measuring. To exit the screensaver just click any
button.
Sound: enable or disables sound.
Turn off after inactivity period: The instrument turns off after the specified period of
inactivity. (Activity may be communication with PC, keyboard pressing key). When log is
activated, automatic turn off is disabled.
System info: indicates the version of the SMP2 software. By clicking on OK, you will have
more information about the unit like disk space available.
6 FUNCTIONS
This section describes the main functions of the SMP2, in addition to instantaneous
measurement of the electromagnetic field.
6.1 Broadband - time domain measurements
One of the main functions of the SMP2 is the storage of measurements in memory for later
study or processing. Measurements in time domain contain field samples, one sample every “log
interval” as defined in 5.4. For example a 6 minutes measurement at 0.5s sampling interval will
create a 720 samples measurement.
Measurements are stored in a SQL database. The memory of the SMP2 allows storage of up to
1,000,000 samples. The maximum number of measurement is 1000.
Before starting the measurement, the parameters for the measurement interval, type of average
and average interval should be configured via the measurement configuration menu (see 5.4).
The field and average field values will be stored in the unit selected in the main screen.
To start logging samples, go in the main screen (direct access is always available by pressing the
home button) and press the “log” function. When logging starts, the device automatically
creates a new measurement with its starting date.
To stop logging samples, press the “stop function”.
Note: If the storage limit of the memory is reached, the logging of samples will be automatically
stopped
During logging parameters of the SM 2 can not be changed.
In order to store positioning information along with the field values, the GPS receiver should be
enabled before logging is started.
6.1.1 Measurements according to ICNIRP procedures
ICNIRP procedure set-up is the most widely referenced by international standards.
- 15 -

That measurement allows an average to be taken of the field values measured over 6 minutes at
a given place without the presence of the technician affecting the result.
To perform this type of measurement select ICNIRP in the configuration measurement screen. All
the parameters will be forced to the pre-defined value:
1. The apparatus will wait 10s before starting the measurement to allow the technician to
move away. After the 10s, a beep will indicate the start of the measurement.
2. It will log samples for six minutes.
3. At the end of the 6 minutes, the measurement will be stopped automatically and the
alarm will sound. The apparatus will calculate the average, display it on the screen and
store it in memory.
This measurement can be cancelled at any time by pressing the STOP function.
6.1.2 Low frequency time domain measurement with WP50 or WP 00
When using this probes, additional functions are available other than those explained in 5.4
•Mode: select frequency or time domain mode (only available for WP400)
•Field: select electric or magnetic field
•Limit: (see 6.3) select the standard to apply
◦None. No weighting will be applied and the measurement will be displayed in the
desired units.
◦Other choice (measurement will be weighted in % compared to the chosen limit)
6.2 Selective - frequency domain measurements
This function is only available with a selective probe like WP400. When using such probe you can
decide in the main screen whether you want to work in the time or frequency domain using the
function “Mode”.
•In time domain mode you can refer to section 6.1 as it works the same way as a
broadband probe. The only additional value you will have using the WP400 is the peak
value.
•In frequency mode you will have, in the main screen, the graph representing the
frequency information (FFT) for the selected SPAN and the total field value representing
the RMS value. (Span has no effect on the total field value)
Measurements in frequency domain contain frequency information of the measured field. One
measurement contains 401 frequency points so the memory of the SMP2 allows storage of up
to 1000 frequency measurement.
In frequency mode the parameters in “measurement configuration” have no effect. All
parameters have to be chosen from the functions in the main screen:
•Mode: select the frequency or temporal mode
•Field: select electric or magnetic field
•Span: select the frequency range (only affects the FFT):
◦0 - 400 Hz (1 Hz resolution)
◦0 - 4 kHz (10 Hz resolution)
◦0 - 40 kHz (100 Hz resolution)
- 16 -

◦0 - 400 kHz (1 kHz resolution)
•Limit: (see 6.3) select the standard to apply
◦None. No weighting will be applied and the measurement will be displayed in the
desired units.
◦Other choice (measurement will be weighted in % compared to the chosen limit)
•Hold
◦Off. No Maxhold is applied. Real-time value is displayed on the screen and FFT is
update with new values every half second
◦Max. Maxhold is applied. Peak value, RMS value and FTT will keep only the max
value. This option is the best when you want to make a single measurement for
compliance with a limit. Activate the MaxHold, wait the desired time and press log to
save the maximum Peak value, RMS value and FTT in a single register.
•Axis (useful for direction finding application)
◦T. Total value of the FFT
◦X axis of the FTT
◦Y axis of the FFT
◦Z axis of the FFT
Note: When downloaded to the PC, all axis of FFT measurements can be seen on the
graph.
To save a FFT, go to the main screen (direct access is always available by pressing the home
button) and press the “log” function. The device will save the current FFT.
6.2.1 Scenarios
This paragraph intents to give you some recommendations on how to use this selective system
for several situations:
•Checking power lines harmonic content:
•Select mode: frequency domain
•Span: 400 Hz
•Filter: 10 Hz
You will get an FFT showing the 50 Hz and its main harmonics in the frequency domain
(from 1 to 400 Hz). Repeat this for electric and magnetic field.
•Power lines measurement:
•Select mode: temporal
•Filter: 10 Hz.
In this mode you will get a temporal recording of the total value like a broadband probe.
Repeat this for electric and magnetic field
•Measurements where field type is unknown:
•we recommend to always use at least the filter 10Hz because below 10 Hz any
non desired movement around the unit may affect the measurement. If you are
not sure if there is field below 10 Hz, then put the filter to 1 Hz. In this case
always measure with the SMP2 on a tripod and the fibre link in order to avoid
any external perturbation.
- 17 -

•if you need only to see the compliance in % respect to a standard, temporal
mode is enough. This will give you a complete and instantaneous result to
compare with the standard
•If you need to check the frequency content of the field, then use frequency
mode. First use the higher span. If no field is detected above 40 kHz then change
span to 40 kHz. Then if no field is detected above 4kHz then change span to 4
kHz. Then again if no field is detected above 400Hz, change to the lowest span
that is 400 Hz. When you have the desired span you can save the FFT by pressing
the log function.
Note: even in frequency mode, total field value, axes values, max, min and pea value are updated
every 0.5 seconds and represent the field value in all the frequency range.
6.3 Time domain weighting
Some probes like WP50 and WP400 can assess the RF exposure by directly comparing the field to
a reference limit. It uses the Weighting Peak Exposure technique that weights the field with a
filter function which is related to the reference level.
The result is a number in % where 100% represents the reference limit line. More than 100%
will mean that the present field is exceeding the limit specified by the chosen standard.
Note: This function is available for time and frequency domain measurement and the results will be
displayed in % instead of the electric or magnetic corresponding units.
6.4 Alarm
The alarm function is very useful for the technician who makes the measurements because it
allows the apparatus to give an audible warning when the field exceeds a threshold value
defined on the settings screen. The alarm will only sound if the alarm function has been
activated and if the sound is enabled in the Settings screen.
6.5 GPS
The SMP2 is available with a built-in latest-generation GPS ublox engine 7 allowing you to
display and memorise the position of the device when taking measurements. The incorporation
of ublox functions makes the receiver extremely sensitive and precise, allowing exact positioning
even in built-up areas with poor satellite visibility.
The GPS is enabled/disabled through the configuration menu as explained in section 5.6.
When the position obtained by the GPS is not valid, the main screen displays ‘Searching for
satellites’. When the position obtained by the receiver is valid, the data for latitude, longitude are
displayed on the screen with a colour indicator representing the quality of the position.
Note: Enabling the GPS increases power consumption and reduces its autonomy. It is important to
eep it switched off when it is not in use to avoid unnecessary drain on the battery.
- 18 -

7 BATTERY INFORMATION
7.1 General information
The SMP2 contains a rechargeable Li-ion battery allowing the apparatus to run for over 15 hours
with broadband probes and over 4h30 for selective probe like WP400. The level of battery
charge can be checked at any time on the main screen of the apparatus, in the battery field.
Li-ion batteries have no “memory effect” and so there is no need to fully discharge the battery. It
is recommended to charge completely the battery before the first use.
Always switch the apparatus off when it is not in use.
Recharge the battery regularly (for example, once every 2 months) if the apparatus
goes for long periods without being used.
The battery of the SMP2 should be recharged when the battery indicator shows that it is at its
minimum level:
Although the apparatus can be fully used while the battery is being charged we
do not recommend to measure with the charger connected as mains perturbation
may alter measurement. In general measurement has to be done without the
charger or USB connected.
The battery will function normally at temperatures ranging from -20ºC to +45ºC. However, we
recommend avoiding low temperatures to prevent its useful life from being shortened.
Only the battery charger supplied by Wavecontrol should be used to recharge the battery of
the SMP2.
7.2 Recharging
The battery charge time is –5 hours. We recommend to turn off the SMP2 for charging the
battery. The procedure for charging the battery is:
• Plug the charger into an electrical outlet (110-240 V AC).
• Insert the connector of the battery charger into the connector of the SMP2. The
recharge red LED will light up. If it does not light up, it means that the charger is not
properly connected.
• The charge is complete when the recharge LED is green. Disconnect the charger from
the SMP2 and unplug it from the mains.
The SMP2 has a Li-ion battery charge controller so you can leave the apparatus charging for
more than the specified 5 hours with no danger of damaging the battery.
Note: it is recommended to do a complete charge from time to time for SMP2 to recalibrate the battery
capacity.
Note2: if during the charging, the red light blin s this means the battery is too hot or is not charging
properly. Disconnect the charger and wait at least 1 hour, connect the charger bac : in case this
happens again battery may be bro en.
- 19 -

8 SOFTWARE
8.1 Installing the software
Before connecting the SMP2 to the PC, launch the setup file (available in the usb flash key).
During the process the setup file will install the software and all the necessary drivers.
Images below show the complete process:
Figures 11: Installation of the software
When the installation is finished you can connect the SMP2, windows will install the driver
automatically.
- 20 -
Other manuals for SMP2
3
Table of contents
Other WAVECONTROL Measuring Instrument manuals