weewell WMP540 User manual

I
WMP540 USER MANUEL 1
WMP540
USER MANUAL
KULLANIM KILAVUZU
EN
TR
AUTOMATIC
UPPER ARM BLOOD
PRESSURE MONITOR
ÜST KOL TIPI OTOMATIK
TANSIYON ÖLÇME ALETI

IWMP540 USER MANUEL
2
Table of Contents
MedicalDisclaimer........................................................................................................................03
Intended Use..................................................................................................................................03
AboutBloodPressure...................................................................................................................04
Precautions.....................................................................................................................................06
DeviceOverview............................................................................................................................08
SymbolDenitions........................................................................................................................10
Features............................................................................................................................................11
InstallingBatteries.........................................................................................................................14
Applying the Cu..........................................................................................................................15
Measurement Procedure............................................................................................................16
Memory Function.........................................................................................................................18
Storage and Maintenance..........................................................................................................19
Troubleshooting...........................................................................................................................20
LimitedWarranty............................................................................................................................21
Specications.................................................................................................................................22
Note...................................................................................................................................................23
Appendix.........................................................................................................................................24
BloodPressureDiary.....................................................................................................................26

I
WMP540 USER MANUEL 3
Medical Disclaimer
Intended Use
About Blood Pressure
This manual and product do not substitute for advice provided by your doctor.
You must not use the information contained herein, or this product for diagnosing or
treating a health problem or prescribing any medication. If you suspect that you have
a medical problem, promptly consult your healthcare provider.
This device uses the oscillometric method to automatically measure systolic and
diastolic blood pressure as well as heart rate.
The measurement position is at human being’s arm.
All values can be read out in one LCD panel.
The device is designed for home use and recommended for use by adults aged 18
years and older with upper arm circumference ranging from 9 ~13” (approx. 23 ~ 33 cm).
1. What is blood pressure?
Blood pressure is the measurement of the force of blood pushing against the walls of the
arteries. Arterial blood pressure is constantly uctuating during the course of the cardiac
cycle. The highest pressure in the cycle is called the systolic blood pressure, and represents
the pressure in the artery when the heart is beating. The lowest pressure is the diastolic
blood pressure, and represents the pressure in the artery when the heart is at rest. Both the
systolic and the diastolic pressure are necessary for a physician to evaluate the status of a
patient’s blood pressure.
Many factors such as physical activity, anxiety or the time of day, can inuence your blood
pressure. Blood pressure is typically low in the mornings and increases from the afternoon
to the evening. It is on average lower in the summer and higher in the winter.
2. Why is it useful to measure blood pressure at home?
Having one’s blood pressure measured by a doctor in a hospital or a clinic, is often
associated with a phenomenon called “White Coat Hypertension” where the patient
becomes nervous or anxious, thus raising his blood pressure. There are also numerous
other factors that might cause your blood pressure to be raised at a specic time of day.
This is why medical practitioners recommend home monitoring as it is important to get

IWMP540 USER MANUEL
4
readings of blood pressure during dierent times of the day to really get an idea of your
real blood pressure.
Medical practitioners generally recommend the “Rule of 3”, where you are encouraged
to take your blood pressure three times in a row (at 1 minute interval), three times a day
for three days. After three days you can average all the results and this will give you an
accurate idea of what your blood pressure really is.
About Blood Pressure
A. WHO blood pressure classications:
Standards for assessment
of high or low blood pressure
without regard to age, have been
established by the World Health
Organization (WHO), as shown in the
chart.
However this chart is not exact for
classication of blood pressure and it’s intended to be used as a guide in understanding
non-invasive blood pressure measurements. Please consult with your physician for proper
diagnosis.
B.Variations in blood pressure:
Individual blood pressures vary greatly both on a daily and a seasonal basis. These
variations are even more pronounced in hyper tense patients. Normally the blood
pressure rises while at work and is at its lowest during sleeping period.
(hyper tense: means a person who has high blood pressure symptom.)
The graph below illustrated the variations
in blood pressure over a whole day with
measurement taken every ve minutes.
The thick line represents sleep. The rise in blood
pressure at 4 PM (A in the graph) and 12 PM (B in
the graph) correspond to an attack of pain.

I
WMP540 USER MANUEL 5
Precautions
• Do not take a measurement in a low (less than 41°F/5°C) and high (more than
104°F/40°C) temperature, nor in a place outside humidity ranges (15 % ~ 93 % R.H.), or
you may get inaccurate readings.
•Wait 30 ~ 45 minutes before measurement if you’ve just consumed caein-included
beverages or smoked cigarettes.
•Rest at least 5 ~ 10 minutes before taking a measurement.
•To allow your blood vessels to return to the condition prior to taking the measurement,
please wait at least 3 ~ 5 minutes in between measurements. You may need to adjust the
waiting time according to your personal physiological situation.
•We recommend you using the same arm (preferably the left arm) and measuring
around the same time each day.
• Sit down comfortably and place your elbow on the table with your feet at on the oor.
Please do not cross your legs during measurements.
• Keep the device at heart level. Relax your hand with the palm facing up.
• Perform measurements in a quiet and relaxed environment at room temperature.
• Do not move or shake the device during a measurement. Please keep quite and do not
talk at the same time.
*Do not use this manual and product as a substitute for advice, diagnosing or
treating a health problem or prescribing any medication by your doctor. If you have
a medical problem, promptly consult your healthcare provider.
*Read the Instruction Manual thoroughly before measuring and keep it at hand for
your reference at any time.
*This device uses the oscillometric method to measure systolic and diastolic blood
pressure as well as your heart rate. It’s recommended for use by people over the age
of 18 and not to be used on infant or children.
*The device is designed for home use and not suitable for clinical use.

IWMP540 USER MANUEL
6
Precautions
• Keep in mind that blood pressure naturally varies from time to time throughout the
day and is aected by lots of dierent factors such as stress, eating, smoking, alcohol
consumption, medication, and physical activity, etc.
Normally the blood pressure rises while at work and is at its lowest during sleeping period.
• Blood pressure measurements should be interpreted by a physician or a trained health
professional who is familiar with your medical history. Using the unit and recording the
results regularly for your physician to interpret, you will keep your physician informed of
the continuing changes in your blood pressure.
• If you have one of the circulatory problems as arteriosclerosis, diabetes, liver disease,
kidney disease, severe hypertension, peripheral circulation….., please consult your
healthcare professional before using the device.
• This product is not suitable for people with arrhythmias and pregnant women.
• Blood pressure measurements taken with this device are equivalent to those obtained
by a trained observer using the cu / stethoscope auscultation method and are within the
accuracy limits prescribed by the Standard of EN 1060-4.
*Attention !
1.. Do not use the device on infants, children, or those who cannot express their
own intention.
2. The device is equipped with sensitive electronic components. While measuring,
avoid strong electrical or electromagnetic elds, e.g. mobile phones, microwave
ovens, etc; or it may lead to temporary reading error or inaccuracy.
3. Consider the electromagnetic compatibility of the device (ex. power disturbance,
radio frequency interference etc.) Please use it indoor only.
4. Over high frequency measurements may result in blood ow interference,
which is likely to cause uncomfortable sensations, such as partial subcutaneous
hemorrhage, or temporary numbness to your wrist. In general, these symptoms
should not last long. However, if you do not recover in time, please seek your
medical practitioners for help.

I
WMP540 USER MANUEL 7
Device Overview
AA (1.5V) ALKALINE
BATTERY X 4
♦ Part names and product components
*Caution !
Substitution of a component dierent from that supplied ones might result in
measurement error.
LCD DISPLAY
MODE BUTTON
START / STOP
BUTTON
MEMORY
BUTTON
SET BUTTON
USB PORT
(AT THE BACK OF THE DEVICE)
6V 1A AC Adapter
(Excluded)

IWMP540 USER MANUEL
8
Device Overview
♦ Unit display
Hour : Minute
(Also represents YEAR in Setting Mode)
Month / Date Risk Category
Indicator
Blood Pressure
Unit Systolic Rate
Low Battery
Symbol
Diastolic Rate
Memory Record
Number
Pulse Symbol
Irregular
Heartbeat
Detector
Heart Rate

I
WMP540 USER MANUEL 9
Symbol Denitions
SYMBOLS Denitions
Low Battery Symbol
This symbol appears when the battery power is excessively
low or the polarity reverses.
►We suggest you replace all batteries with new ones, and
make sure the +/- polarities are properly positioned.
Pulse Symbol
Once pulse is detected, the symbol ashes with each
pulse beat.
►Our suggestion:
Please do not talk or move during measurements.
Irregular Heartbeat
Detector
This symbol appears for 1 minute when the user was
talking, moving, shaking, or an irregular heart beat was
detected during measurements.
►Our suggestion:
Please do not talk or move during measurements.
Repeat the measurement after resting for at least 5
minutes, and restart your measurement while sitting
down comfortably and quietly.
Memory Record
Symbol
Figures beside this symbol represent the order of memory
stored.

IWMP540 USER MANUEL
10
*Attention !
• The pulse display is not suitable for checking the frequency of heart pacemakers.
If a certain pulse irregularity is detected during measurement often, we
recommend you to take medical advice
• As a safeguard, we recommend that if you have arrhythmias such as atrial or
ventricular premature beats and atrial brillation or any other special conditions
you should check with your physician before using your device.
• The IHB function is not designed for use by people with arrhythmias nor for
diagnosing or treating an arrhythmic problem. In order to lter the unstable
status of user and avoid aecting the detection of heart rate from any movement,
shaking or talking in the beginning of measurement, the method of averaging
heart beat intervals of subject device is calculated with the three proper heart beat
pulses detected in the beginning of measurement and that is dierent from a strict
mathematical averaging of all recorded intervals.
• At least 3 beats with at least 25% dierence from the average heart beat interval
will generate the IHB icon on the screen.
Features
♦ Irregular Heartbeat Detector
The symbol will appear on screen indicating a certain heartbeat irregularity was
detected during measurement.
The heartbeat rhythm that is more than or less than 25% from the average rhythm is
usually dened as an irregular heartbeat rhythm.
Talking, moving, shaking or an irregular pulse during the measurement can result in
the appearance of this symbol.
Usually this is not a cause for concern, however if the symbol appears often, we
recommend you seek medical advice.
And please note that the device does not replace a cardiac examination, but serves to
detect pulse irregularities at an early stage.

I
WMP540 USER MANUEL 11
This device is equipped with Risk Category Indicator which classies your blood
pressure measurements into six stages (Optimal to Severe hypertension) as shown in
below chart:
After each measurement is completed, LCD display will show your position
automatically on the six segments of the bar indicator which corresponds to Risk
Category Indicator.
Features
♦ Risk Category Indicator
Stages of Blood Pres-
sure Levels
Systolic
(mmHg)
Diastolic
(mmHg)
Recommendations by
SIGN 49: Hypertension in older
people
Grade
3Severe
Hypertension ≧180 ≧110
Conrm immediately and
repeat BP in one day and again
within one week depending
on clinical situation.
Grade
2Moderate
Hypertension 160 ~
179 100 ~ 109 Serial blood pressures repeated
within one month.
Grade
1Mild
Hypertension 140 ~
159 90 ~ 99
Provide advice about lifestyle
modication and conrm
within two months.
High-Normal 130 ~
139 85 ~ 89 Provide advice about lifestyle
modication andrecheck in
one year.
Normal 120 ~
129 80 ~ 84 Recheck in 2 - 5 years.
(patients aged > 75 years
oered annual health check)
Optimal < 120 < 80

IWMP540 USER MANUEL
12
*Note !
When a person’s systolic and diastolic pressures fall into dierent categories, the
higher category should apply.
e.g. systolic rate 181 & diastolic rate 99 Red category (Severe Hypertension)
e.g. systolic rate 110 & diastolic rate 95 Red category (Mild Hypertension)
Note !
The above table is not exact for classication of blood pressure and it’s intended
to be used as a guide in understanding non-invasive blood pressure measurements.
Usually this is not a cause for concern; however we recommend you consult
with your physician for proper diagnosis or seek medical advice according to
our recommendation mentioned above. Please note that the device does not
appropriate to diagnose hypertension, and it is only for user reference on blood
pressure monitoring.
Installing Batteries
When LOW BATTERY SYMBOL appears on the display, or nothing appears on the
display when the device is powered on, please change the batteries.
Replace all the batteries with new ones and do not mix new and old batteries. Do not
mix alkaline, standard (carbon-zinc) or rechargeable (cadmium) batteries either. It
may shorten the battery life or cause the device to per.
Remove the battery cover and insert 4 AA alkaline batteries into the battery
compartment as shown. Make sure the polarities “+” and “–” ends are properly
positioned.

I
WMP540 USER MANUEL 13
*Attention !
• Batteries are hazardous waste. Do not dispose of them together with the
household garbage. Please take the used batteries to the recycling collection point
according to your local regulations.
• Keep the battery away from small children in case they swallow it.
• To prolong the battery life and prevent damage caused by leakage, remove the
batteries from the device if the device is not to be used for a long period.
• Memories (if any) will not be deleted during battery replacement.
• After replacing the batteries, reset the date and time.
*Attention !
• Fit the cu appropriately, leaving enough space for 1 inch (2 ~ 3 cm) between
the inner elbow and the lower edge of the cu, or the measurement may not be
accurate.
• This monitor comes with one size of arm cu: 9” ~ 13” (23 ~ 33 cm).
• In case the cu kept pumping up non-stop, open the cu at once.
• Do not wrap the cu around any body part other than your arm.
• The device is not supposed to be used when your arm is wounded or injured.
Applying the Cu
• Press your brachial artery approximately 1 inch (2 ~ 3 cm) above the elbow on the
inside of your left arm to determine where your strongest pulse is.
• Slide the end of arm cu furthest from the tube through the metal ring to a loop.
The smooth cloth should be on the inside of the cu.
• If the cu is located correctly, the velcro will be on the outside of the cu and metal
ring will not touch your skin.
• Put left arm through the cu loop. The bottom of the cu should be approximately
1 inch (2 ~ 3 cm) above the inner elbow. The tube should lie over the brachial artery
on the inner part of the arm.
• Pull the cu so that the top and bottom edges
are tightened around your arm.
• When the cu is positioned properly, press the
velcro rmly against the pile side of the cu.
• Sit on a chair and lay your forearm on the
table so that the cu is at the same level as your
heart.
• Relax your arm and turn your arm upward.
• Make sure there are no kinks in the air tube.

IWMP540 USER MANUEL
14
Measurement Procedure
A. Press button to switch on the monitor.
B. All segments appear on the screen.
A. Before measurement, press +button to select ser 1, 2, or 3.
B. With the cu wrapped around your arm, press button to start measurement. All
display units appear on the screen.
A. Press button (“YEAR” ashes). Press + button to adjust YEAR value.
B. Press button (“MONTH” ashes). Use +button to adjust MONTH (1, 2, 3,…, 12).
C. Adjust DATE (1, 2, 3,…, 31), HOUR (1, 2, 3,……..12PM,1 PM,…, 12) and MINUTE
(00,01,02,03,.....59) as described in Step A above. When settings are done, press
button to conrm the entries. The device is ready for use.
♦ Switch on the monitor
♦ Taking a Measurement
♦ Setting year, date and time
*Note !
• Do not inate the cu until it is wrapped around your arm.

I
WMP540 USER MANUEL 15
Measurement Procedure
After all symbols disappear, the display will show “00”.
The monitor is “Ready to Measure” and will automatically
inate to the level that is right for you.
C. After ination of the cu, the pressure will slowly
decrease. When pulse is detected, PULSE SYMBOL
ashes.
D. LCD screen displays your systolic rate, diastolic
rate, pulse, Risk Category Indicator Bar, and Irregular
Heartbeat Detector symbol with date and time for 1
minute.
E. Without any operation for 1 minute, device
automatically shuts o.
*Note !
• If the cu does not stop inating, remove the cu at once.
• To stop measurement, press button.

IWMP540 USER MANUEL
16
Memory Function
After each measurement, the systolic and diastolic pressure, heart rate with date and
time will be automatically stored.
The monitor can store 120 memories total for 3 users, and automatically replace the
oldest data with new one.
A. Press +button to select User 1, 2, or 3.
B. Press M button to enter Memory Mode. LCD displays
average of last 3 measuring results rst.
C. Press M button again, LCD displays the latest
measuring result. Keep pressing M button, user can
scroll through following measurements in sequence.
D. To stop reading memories, press button, and
switch to Standby Mode.
A. Press +button to select User 1, 2, or 3.
B. Press M button to enter Memory Mode.
C. Press and hold and +buttons at the same time,
the data will be erased automatically.
D. To conrm the data in the selected user has been
erased, press M button and no data would appear.
Note: Once deleted, your data can NOT be restored.
♦ Storing data
♦ Recalling data
♦ Erasing data

I
WMP540 USER MANUEL 17
Storage and Maintenance
♦ General Use
• Do not twist the cu.
• Do not press button if the cu is not wrapped around your upper arm.
• Do not drop the product and avoid any strong impacts.
♦ Maintenance
• Use a piece of cloth with water or mild cleansing agent to wipe the device and dry it
immediately with a dry cloth.
• Do not use detergent or any strong chemicals to clean the device.
• Use only a dry cloth to wipe the cu.
• Do not attempt to disassemble or change any parts of the monitor, including arm
cu, due to substitution of a component dierent from that supplied might result in
measurement error.
• If any suggestion or service is requested, please consult your service station.
♦ Storage
• If the device is not to be used for a long time, please remove the batteries from the
device (leaking of battery acid can cause the device to perform improperly).
• Always store the unit in the storage case after use.
• Do not place the device under direct sunlight, in high temperature, or in humid or
dusty places.
• Do not store the device in extremely low (less than –13°F/–25°C) or high (more than
158°F/70°C) temperature, neither in a place its humidity exceeds 93% R.H.

IWMP540 USER MANUEL
18
Troubleshooting
SYMBOLS/
SYMPTOMS
CONDITIONS/
CAUSES
INDICATION/
CORRECTION
No power-on
by pressing
button.
Depleted batteries. Replace them with 2 new
AAA alkaline batteries.
Battery polarities
have been positioned
incorrectly.
Re-insert the batteries in the
correct positions.
Measuring
Error Symbol
appears
when blood
pressure
value
displayed is
excessively
low or high.
Cu has been placed
incorrectly.
Wrap the cu properly so that
it is positioned correctly.
Did you talk or move
during measurement?
Measure again. Keep wrist
steady during measurement.
Shaking of the wrist with
the cu on.
Measuring
Error Symbol
Air circuit abnormality.
Cu tube may not be
plugged into monitor
correctly.
Check cu connection.
Measure again.
Measuring
Error Symbol
Ination pressure ex-
ceeding 300 mmHg.
Switch the unit o, then
measure again.

I
WMP540 USER MANUEL 19
Limited Warranty
To ensure measurement precision, all digital blood pressure monitors require
recalibration regularly.
After 2 years from the manufacturing date, we recommend you have your monitor
recalibrated at the local distributor or importer.
Please contact your distributor/importer for the details about the recalibration
service and the charge of shipping and handling.
Please also note that this service does not cover damage caused by misuse or
abuse; accident; the attachment of any unauthorized accessory; alteration to the
product; improper installation; unauthorized repairs or modications; improper
use of electrical/power supply; loss of power; dropped product; malfunction or
damage of an operating part from failure to provide manufacturer’s recommended
maintenance; transportation damage; theft; neglect; vandalism; or environmental
conditions; loss of use during the period the product is at a repair facility or otherwise
awaiting parts or repair; or any other conditions whatsoever that are beyond the
control of importers or distributors.
Measuring
Error Symbol
Error determining
measurement data. Measure again.
Note: If “EP” appears on the display, just return the device to your local
distributor or importer.

IWMP540 USER MANUEL
20
Specications
Model Number WMP540
Measurement
Method Oscillometric
Measurement
Range
Pressure: 0 ~ 300 mmHg
Pulse: 40 ~ 199 Beats / Minute
Accuracy Pressure: ± 3 mmHg
Pulse: ± 5 % Max.
Ination Automatic Ination (Air Pump)
Deation Automatic Air Release Control Valve
Display Liquid Crystal Display
Memory 120 Memory Total for 3 Users
Unit Dimensions 25.51 x 4.09 x 1.97 inch (L x W x H)
140 x 104 x 50 mm (L x W x H)
Unit Weight 8.73 ± 0.35 oz (247.5 ± 10 g)
(Cu and Batteries Excluded)
Cu Size 23 ~ 33 cm (9 ~ 13 inch)
Storage/
Transportation
Environment
Temperature: -25 °C ~ 70 °C (-13 °F ~ 158 °F)
Humidity: ≤ 93 % R.H.
Operation
Environment
Temperature: 5 °C ~ 40 °C (41 °F ~ 104 °F)
Humidity: 15 % ~ 93 % R.H.
Power Supply 1. AA (1.5V) alkaline battery x 4
2. 6V 1A AC Adapter (Excluded)
Battery Life Approx. 300 measurements
Power-saving
Mode Without any operation for 1 minute, the device
automatically powers o.
Accessories 4 AA Alkaline Batteries, Arm Cu with Tube, Instruction
Manual, Pouch
*The contents of this manual and the specications of the device covered by
this manual are subject to change for improvement without notice.
Table of contents
Languages:
Popular Blood Pressure Monitor manuals by other brands

Omron
Omron INTELLI-SENSE HEM-780 Manual de instrucciones

SHENZHEN PUMP MEDICAL SYSTEM
SHENZHEN PUMP MEDICAL SYSTEM BF1204B instruction manual

Livongo
Livongo HT900 user guide

Kinetik
Kinetik BPM1KTL Series user manual

FarmaMed
FarmaMed UPPER ARM BLOOD PRESSURE MONITOR Instructions for use

Beurer
Beurer easyClip BM 51 Instructions for use

Omron
Omron IntelliSense HEM-650 instruction manual

Beurer
Beurer BM 58 Instructions for use

Omron
Omron M1 Compact instruction manual

Omron
Omron R2 instruction manual

Scala
Scala SC 2100 Instructions for use

Hangzhou Hua'an Medical & Health Instruments
Hangzhou Hua'an Medical & Health Instruments MW-300C Operation guide