wimesure MP6Plus User manual

Photos non contractuelles - Les spécications peuvent être modiées sans préavis - wimesure.fr • MB 2021
41126 Cognento (MODENA) Italy Via Bottego 33/A Tel:+39-(0)59-346441 Fax:+39-(0)59-346437 E-mail: aep@aep.it
MP6P
P
l
l
u
u
s
s
PROFESSIONAL INDICATOR
for measures of
Force
Weight
Pressure
Torque
Displacement
Temperature
USER GUIDE
MO.MP6Plus.R7

Photos non contractuelles - Les spécications peuvent être modiées sans préavis - MP6Plus R0 - wimesure.fr • AH 2022
MP6P
P
l
l
u
u
s
s
USER GUIDE MO.MP6Plus.R7
Pag. 1
INDEX
1. Declaration of Conformity
2
42. Calibration for Incremental Encoder Channel
38
2. Identification
2
43. Channel Linearization
39
3. Unauthorized uses
2
44. Printer Header
41
4. Disposal
2
45. Data Logger Menu (Option)
42
5. Introduction
3
46. Data Logger Parameters
42
6. Available Fittings
4
47. View Data Logger
44
7. Component Supplied
5
48. Copy Data Logger
44
8. Components in Options
5
49. Files for USB Flash Memory
45
9. Software Applications(optional)
6
50. Digital Input (Option)
46
10. TECHICAL DATA
8
51. Analog Output (Option)
49
11. Purchase Codes
9
52. SetPoint
52
12. Dimensions
10
53. RS232 and Printer Communication Parameters
53
13. Trasport
11
54. RS485 Communication Parameters
54
14. Positioning
11
55. Key Block
55
15. INSTALLATION
11
56. Channel Enable and TOTAL Function
56
16. Maintenance
11
57. Wireless Transmission Parameter
57
17. Power On
12
58. Service and LCD Contrast
58
18. Power Off
12
59. Save Channels
59
19. Display and Function Keys
13
60. Restore Channels
59
20. Main Page
15
61. MODBUS RTU Communication Protocol
60
21. Error Messages
16
62. RS232 Communication Protocol
62
22. ZERO Function
16
63. USB Communication Protocol
64
23. PEAK Function
17
64. Sw Examples
66
24. HOLD Function
18
65. Read Command
69
25. UNLOAD Function
18
66. Write Commands
71
26. DATALOGGER Function
19
67. Continuos transmitting mode
73
27. COUNTER Function
22
68. DataLogger Management commands
74
28. PRINTER Function
24
69. Rear Panel Connectors
75
29. MAIN MENU
25
70. RS232 Connections
76
30. Measurement Unit
26
71. RS485 Connections
77
31. Reolution
27
72. 24V Digital Input Connections
78
32. Point Position
27
73. Analog Output Connections
79
33. Digital Filter e Frequency of Acquisition
28
74. Transducers Connections
80
34. Configuration
30
75. Load cells Connections
81
35. Date and Time setting (option)
31
76. Amplified Transducers Connections
82
36. Channel Setup
32
77. PT100 Connections
82
37. Force, Pressure, Torque and Displacement
Channel Calibration
33
78. Incremental Encoder Connections
83
38. Full Scale Calibration
34
79. Power Supply and Part Identification
84
39. Known Weight Calibration
36
40. Calibration by interlation points
37
41. Calibration for Temeprature Channels
38
TABLE 1 : pag 26
TABLE 2 : pag 61
TABLE 3 : pag 61
TABLE 4 : pag 61
AEP transducers holds the right to make any change when necessary, without notice.
The data contained in this manual are just indicative and the manufacturer declines any responsibility for
errors or discrepancies with respect to this manual.

Photos non contractuelles - Les spécications peuvent être modiées sans préavis - MP6Plus R0 - wimesure.fr • AH 2022
MP6PPlluussUSER GUIDE MO.MP6Plus.R7
Pag. 1
INDEX
1. Declaration of Conformity
2
42. Calibration for Incremental Encoder Channel
38
2. Identification
2
43. Channel Linearization
39
3. Unauthorized uses
2
44. Printer Header
41
4. Disposal
2
45. Data Logger Menu (Option)
42
5. Introduction
3
46. Data Logger Parameters
42
6. Available Fittings
4
47. View Data Logger
44
7. Component Supplied
5
48. Copy Data Logger
44
8. Components in Options
5
49. Files for USB Flash Memory
45
9. Software Applications(optional)
6
50. Digital Input (Option)
46
10. TECHICAL DATA
8
51. Analog Output (Option)
49
11. Purchase Codes
9
52. SetPoint
52
12. Dimensions
10
53. RS232 and Printer Communication Parameters
53
13. Trasport
11
54. RS485 Communication Parameters
54
14. Positioning
11
55. Key Block
55
15. INSTALLATION
11
56. Channel Enable and TOTAL Function
56
16. Maintenance
11
57. Wireless Transmission Parameter
57
17. Power On
12
58. Service and LCD Contrast
58
18. Power Off
12
59. Save Channels
59
19. Display and Function Keys
13
60. Restore Channels
59
20. Main Page
15
61. MODBUS RTU Communication Protocol
60
21. Error Messages
16
62. RS232 Communication Protocol
62
22. ZERO Function
16
63. USB Communication Protocol
64
23. PEAK Function
17
64. Sw Examples
66
24. HOLD Function
18
65. Read Command
69
25. UNLOAD Function
18
66. Write Commands
71
26. DATALOGGER Function
19
67. Continuos transmitting mode
73
27. COUNTER Function
22
68. DataLogger Management commands
74
28. PRINTER Function
24
69. Rear Panel Connectors
75
29. MAIN MENU
25
70. RS232 Connections
76
30. Measurement Unit
26
71. RS485 Connections
77
31. Reolution
27
72. 24V Digital Input Connections
78
32. Point Position
27
73. Analog Output Connections
79
33. Digital Filter e Frequency of Acquisition
28
74. Transducers Connections
80
34. Configuration
30
75. Load cells Connections
81
35. Date and Time setting (option)
31
76. Amplified Transducers Connections
82
36. Channel Setup
32
77. PT100 Connections
82
37. Force, Pressure, Torque and Displacement
Channel Calibration
33
78. Incremental Encoder Connections
83
38. Full Scale Calibration
34
79. Power Supply and Part Identification
84
39. Known Weight Calibration
36
40. Calibration by interlation points
37
41. Calibration for Temeprature Channels
38
TABLE 1 : pag 26
TABLE 2 : pag 61
TABLE 3 : pag 61
TABLE 4 : pag 61
AEP transducers holds the right to make any change when necessary, without notice.
The data contained in this manual are just indicative and the manufacturer declines any responsibility for
errors or discrepancies with respect to this manual.
MP6P
P
l
l
u
u
s
s
USER GUIDE MO.MP6Plus.R7
Pag. 2
1. DECLARATION OF CONFORMITY
Manufacter: AEP transducers s.r.l
Address: Via Bottego 33/A 41126 Cognento MODENA (Italia)
DECLARES THAT THE FOLLOWING PRODUCT
Product Name: MP6Plus
Type: PROFESSIONAL INDICATOR
Year of Construction: 2014
Options: this declaration covers all the options specified in the manual.
CONFORMS TO THE FOLLOWING DIRECTIVES:
2014/30/EU - 2014/35/EU - 2011/65/EU (RoHS) - 2012/19/EU (WEEE)
CONFORMS TO THE FOLLOWING NORMS:
EN 61010-1(2013) EN 61326-1 (2013)
CONFORMS TO THE REGULATION
n° 1907/2006 (REACH)
The product has been tested in the typical installation configuration, as described in the instruction manual. Above
described product meets the requirements of mentioned Norms, basing on both test results and considerations listed
in the technical file.
I declare that the product defined above meets the requirements of the Directives, of the Norms and Regulation
above mentioned.
41126 Cognento Modena (Italy) Date: 20/09/2014
Lioi Giovanni
2. IDENTIFICATION
On the rear panel of the instrument are marked in indelible mode all the informations for the identification, the main
configuration data, the symbols of conformity and waste disposal, the production date and the serial number.
3. UNAUTHORIZED USES
Environments with explosive atmosphere.
Environments with inflammable or corrosive gas.
4. DISPOSAL
The instrument is a professional apparatus compliant to the 2011/65/EU(RoHS) and 2012/19/EU (WEEE),
then it must be disposed separately as electric and electronic waste.
In different countries of European Community, it must be disposed as waste electric and electronic in
accord to the laws of the country where the device is commercialized.
Before to remove the instrument, you disconnect first the power supply and after the cables

Photos non contractuelles - Les spécications peuvent être modiées sans préavis - MP6Plus R0 - wimesure.fr • AH 2022
MP6P
P
l
l
u
u
s
s
USER GUIDE MO.MP6Plus.R7
Pag. 3
5. INTRODUCTION
MP6Plus is an instrument with 1 (up to 4 as option) analog inputs (defined in this manual as channels), individually
programmable; each input is able to receive signals from strain gauge (standard version) full-bridge (4 or 6 wires) or
transducers with amplified output voltage (+/- 10V or +/- 5V), current (4- 20mA or 0-20mA) or PT100.
It is suitable for the measurement of forces, weight and pressure, displacement and temperature (PT100). It is realized
to be used in systems of static and dynamic measurement in an industrial environment where it is necessary to
acquire more independent transducers simultaneously and in a synchronized manner.
The instrument, in its basic version has, for each channel, an accuracy of 0.010% due to an 24bit internal resolution.
In the case of strain gauge inputs each channel has a standard resolution of 100.000 divisions.
MP6Plus can use load cells or dynamometers of 350 / 700with 4-wire or 6 wires connection.
Load cells power supply is realized using a switching voltage at a frequency equal to the sampling rate. This ensures a
better suppression of interference due to offset drift and to the connecting cables.
In the case of voltage input 10V or 5V MP6Plus has a resolution of 0.1mV respectively and 0.05 mV
(100000divisions) while in the case of current input 0-20mA or 4-20mA has a resolution of 0.1 uA (respectively
200000divisions and 160000divisions).
To measure the temperature one input can dedicated to sensors PT100 in a range from -50 ° C to 250 ° C with a
resolution of 0.1 ° C and accuracy +/- 1 ° C.
Each channel has a dedicated converter Sigma-Delta type 24bit and all conversions take place at the same time
(channel synchronization).
The sampling frequency (common to all channels) can be set from 2.5 samples per second up to 4800 samples per
second therefore the instrument meets the needs of applications that require a considerable speed of response.
The instrument is equipped with 4 digital inputs to realize simple automations.
The instrument can be equipped with up to 4 analog outputs (optional). Each input channel may be associated with
one or more analog voltage ouput (± 10V, 0 / 5V, 0 / 10V, ± 5V) or current (4-20mA).
MP6Plus allows you to enable/disable the display of each individual channels. In this way suits the needs of use in
different situation of use.
MP6Plus can be optionally equipped with a powerful datalogger that can be either internal (in an internal nonvolatile
memory, which allows to exploit the maximum speed of acquisition), that external using a conventional USB stick. The
files created can be directly exported to Microsoft Excel being in csv format.
Data communication can take place simultaneously and concurrently through USB ports, RS232 and RS485.
MP6Plus main features are:
Large size and high resolution graphic display.
1 (up to 4 as option) external connections to transducers: 2mV/V, +/-10V, +/-5V, 4-20mA, 0-20mA ,PT100
Selection of a large number of measurement unit for each of the connected channels.
Data Logger with internal non-volatile memory that can store up to 130,000 measurement points when
configured with a single channel (Option)
External Data Logger that uses a usual USB stick for easy portability of data on a PC (Option)
ZERO Function
HOLD Function
PEAK Function
COUNTER Function
UNLOAD Function to measure the quantity of material unloaded from a tank
TOTAL Function: to performs the cumulative sum of channels
AUTOZERO Function: to clear a channel output if is below for a certain time under a programmed threhold
Key Block : in order to avoid changes to the settings by unauthorized persons
Clock Calendar Function with date and time (Option)
Printer (Option) connected to the RS232 serial port through which it is possible to print the measurement
points and the data of the company which carried out the survey.
USB communication port through which you can transfer measurements in real time to a PC.
RS232 Serial communication port (Option)
Wireless Transmission (Option)
RS485 communication port (Option)
4 24Vdc digital inputs with programmable function (Option)
Up to 4 analogue outputs (option): 4-20mA, +/- 10Vdc, 0-10Vdc, 0-5Vdc, +/- 5VDC. The refresh rate of the

Photos non contractuelles - Les spécications peuvent être modiées sans préavis - MP6Plus R0 - wimesure.fr • AH 2022
MP6PPlluussUSER GUIDE MO.MP6Plus.R7
Pag. 3
5. INTRODUCTION
MP6Plus is an instrument with 1 (up to 4 as option) analog inputs (defined in this manual as channels), individually
programmable; each input is able to receive signals from strain gauge (standard version) full-bridge (4 or 6 wires) or
transducers with amplified output voltage (+/- 10V or +/- 5V), current (4- 20mA or 0-20mA) or PT100.
It is suitable for the measurement of forces, weight and pressure, displacement and temperature (PT100). It is realized
to be used in systems of static and dynamic measurement in an industrial environment where it is necessary to
acquire more independent transducers simultaneously and in a synchronized manner.
The instrument, in its basic version has, for each channel, an accuracy of 0.010% due to an 24bit internal resolution.
In the case of strain gauge inputs each channel has a standard resolution of 100.000 divisions.
MP6Plus can use load cells or dynamometers of 350 / 700with 4-wire or 6 wires connection.
Load cells power supply is realized using a switching voltage at a frequency equal to the sampling rate. This ensures a
better suppression of interference due to offset drift and to the connecting cables.
In the case of voltage input 10V or 5V MP6Plus has a resolution of 0.1mV respectively and 0.05 mV
(100000divisions) while in the case of current input 0-20mA or 4-20mA has a resolution of 0.1 uA (respectively
200000divisions and 160000divisions).
To measure the temperature one input can dedicated to sensors PT100 in a range from -50 ° C to 250 ° C with a
resolution of 0.1 ° C and accuracy +/- 1 ° C.
Each channel has a dedicated converter Sigma-Delta type 24bit and all conversions take place at the same time
(channel synchronization).
The sampling frequency (common to all channels) can be set from 2.5 samples per second up to 4800 samples per
second therefore the instrument meets the needs of applications that require a considerable speed of response.
The instrument is equipped with 4 digital inputs to realize simple automations.
The instrument can be equipped with up to 4 analog outputs (optional). Each input channel may be associated with
one or more analog voltage ouput (± 10V, 0 / 5V, 0 / 10V, ± 5V) or current (4-20mA).
MP6Plus allows you to enable/disable the display of each individual channels. In this way suits the needs of use in
different situation of use.
MP6Plus can be optionally equipped with a powerful datalogger that can be either internal (in an internal nonvolatile
memory, which allows to exploit the maximum speed of acquisition), that external using a conventional USB stick. The
files created can be directly exported to Microsoft Excel being in csv format.
Data communication can take place simultaneously and concurrently through USB ports, RS232 and RS485.
MP6Plus main features are:
Large size and high resolution graphic display.
1 (up to 4 as option) external connections to transducers: 2mV/V, +/-10V, +/-5V, 4-20mA, 0-20mA ,PT100
Selection of a large number of measurement unit for each of the connected channels.
Data Logger with internal non-volatile memory that can store up to 130,000 measurement points when
configured with a single channel (Option)
External Data Logger that uses a usual USB stick for easy portability of data on a PC (Option)
ZERO Function
HOLD Function
PEAK Function
COUNTER Function
UNLOAD Function to measure the quantity of material unloaded from a tank
TOTAL Function: to performs the cumulative sum of channels
AUTOZERO Function: to clear a channel output if is below for a certain time under a programmed threhold
Key Block : in order to avoid changes to the settings by unauthorized persons
Clock Calendar Function with date and time (Option)
Printer (Option) connected to the RS232 serial port through which it is possible to print the measurement
points and the data of the company which carried out the survey.
USB communication port through which you can transfer measurements in real time to a PC.
RS232 Serial communication port (Option)
Wireless Transmission (Option)
RS485 communication port (Option)
4 24Vdc digital inputs with programmable function (Option)
Up to 4 analogue outputs (option): 4-20mA, +/- 10Vdc, 0-10Vdc, 0-5Vdc, +/- 5VDC. The refresh rate of the
MP6P
P
l
l
u
u
s
s
USER GUIDE MO.MP6Plus.R7
Pag. 4
analog signals is equal to the frequency of acquisition of the respective channels in input.
REPEATER Function: The instrument can be configured to display (in the form passive as Slave) measures
from the RS232 serial port (for example from another MP6Plus -Master) to a remote view of the measures.
In this case all the features enabled on the MP6Plus Slave will be active (Setpoint, USB, printer, logger etc).
The REPEATER function is active for one channel.
MP6Plus can be accompanied by the PC program MP Supervisor (Option) which allows immediate interface via the USB
port with the instrument and allows you to display graphs, export to Microsoft Excel.
The program also allows you to download the data logger performed using the internal memory and those on
performed on USB stick and view their acquisition curves.
6. AVAILABLE FITTINGS
KAL KAL CLF C8S CLB TCETM
System for the measurement of FORCE (compression and tension)combined with force transducers with ranges from
10N to 5000kN.
CLT CLS C2S F1
System for the measurement of WEIGHT in compression and tension matched to the load cells with ranges from 1kg
to 500t.
TP16 TP1
System for the measurement of PRESSURE and VACUUM (Empty) combined with pressure transducers.
Normalized ranges from 5 bar to 2000bar (29000psi) for use in gas and liquid.
TRX TRS TOR RT2
System for the measurement of TORQUE clockwise and counterclockwise combined with torque transducers static or
dynamic. Normalized ranges from 0.5 Nm to 5000 Nm, ability to record high-speed continuous PEAKS up to 4.8kHz.
.
LDT
System for the measurement of DISPLACEMENT transducers coupled with transducers with normalized ranges from 5
to 100 mm.

Photos non contractuelles - Les spécications peuvent être modiées sans préavis - MP6Plus R0 - wimesure.fr • AH 2022
MP6P
P
l
l
u
u
s
s
USER GUIDE MO.MP6Plus.R7
Pag. 5
7. COMPONENTS SUPPLIED
Power Cord DB9 Male Connector CD with
for transducer Manual and USB Driver
8. OPTIONAL COMPONENTS (purchased separately)
USB cable
RS232 Serial Cable
Pair of mounting
brackets for panel
Desktop Printer
RS232 Interface
DB9 Male
Connector
For each
transducers
Calibration Report
ACCREDIA Certificate
(MP6+Transducer)
Calibratorfor mV/V
signals
Case for transport

Photos non contractuelles - Les spécications peuvent être modiées sans préavis - MP6Plus R0 - wimesure.fr • AH 2022
MP6PPlluussUSER GUIDE MO.MP6Plus.R7
Pag. 5
7. COMPONENTS SUPPLIED
Power Cord DB9 Male Connector CD with
for transducer Manual and USB Driver
8. OPTIONAL COMPONENTS (purchased separately)
USB cable
RS232 Serial Cable
Pair of mounting
brackets for panel
Desktop Printer
RS232 Interface
DB9 Male
Connector
For each
transducers
Calibration Report
ACCREDIA Certificate
(MP6+Transducer)
Calibratorfor mV/V
signals
Case for transport
MP6P
P
l
l
u
u
s
s
USER GUIDE MO.MP6Plus.R7
Pag. 6
9. SOFTWARE APPLICATION (purchased separately)
To complete the system of measurement AEP transducers has developed several software applications that interface
directly to the instrument MP6Plus and support the user in the various functions of calibration, testing, analysis, data
storage, transfer of measures on Microsoft Excel etc. ...
MP Supervisor is a software dedicated to MP6Plus . Through this software you can download the data logger and
operate directly on MP6Plus to change parameters and create graphics test.
Quick Analyzer is a general purpose acquisition software where MP6Plus can be associated to other AEP instruments .
For more information download the manuals of the software on the site:
www.aeptransducers.com
www.aep.it

Photos non contractuelles - Les spécications peuvent être modiées sans préavis - MP6Plus R0 - wimesure.fr • AH 2022
MP6P
P
l
l
u
u
s
s
USER GUIDE MO.MP6Plus.R7
Pag. 7
QuickAnalyzer
Dedicated to recording and graphical analysis of up to 16 different AEP transducers instruments to measure:
force, weight, pressure, torque and displacement.
MPSupervisor
A dedicated program that allows an immediate interfacing through the USB port with the MP6Plus and allows you to
view graphs, export data to Microsoft Excel directly from the PC and set all configuration parameters.
The program also allows you to download a Data Logger carried out using the internal memory or the USB Flash
Memory and display the respective curves of acquisitio

Photos non contractuelles - Les spécications peuvent être modiées sans préavis - MP6Plus R0 - wimesure.fr • AH 2022
MP6PPlluussUSER GUIDE MO.MP6Plus.R7
Pag. 7
QuickAnalyzer
Dedicated to recording and graphical analysis of up to 16 different AEP transducers instruments to measure:
force, weight, pressure, torque and displacement.
MPSupervisor
A dedicated program that allows an immediate interfacing through the USB port with the MP6Plus and allows you to
view graphs, export data to Microsoft Excel directly from the PC and set all configuration parameters.
The program also allows you to download a Data Logger carried out using the internal memory or the USB Flash
Memory and display the respective curves of acquisitio
MP6P
P
l
l
u
u
s
s
USER GUIDE MO.MP6Plus.R7
Pag. 8
10. TECHNICAL DATA
ACCURACY
NON LINEARITY ERROR
≤± 0.010%
≤± 0.010%
STANDARD RANGE FOR STRAIN GAUGES TRANSDUCERS
STRAIN GAUGES RESOLUTION AT 2mV/V
RESOLUTION FOR +/-10V AND +/-5V TRANSDUCERS
RESOLUTION FOR 4-20mA AND 0-20mA transducers
TEMPERATURE RESOLUTION : PT100 2 WIRES (range -50°C –250°C)
INTERNAL RESOLUTION (ALL CHANNELS)
MAX NUMBER OF STRAIN GAUGE TRANSDUCERS CONNECTABLE
AMPLIFIED TRANSDUCERS (+/-10V or 4-20mA) CONNECTABLE
STRAIN GAUGE POWER SUPPLY
AMPLIFIED TRANSDUCERS POWER SUPPLY
ACQUISTION FREQUENCY (Programmable) (samples per second)
(for all channels)
2mV/V
100.000 div
0.1mV e 0.05mV (100000 div)
0.1 µA (resp. 160000 div and 200000div)
0.1°C accuracy: 1°C
24bit
Each channel: 2 @350or 4 @ 700
4
5V switching (3%)
20Vdc
2.5–5–10–20–50-100-200-400–600-1200–
2400-4800
INTERNAL DATALOG
MAX STORING POINTS
MAX STORING SPEED
Note: The maximum storage speed is however limited by
the acquisition frequency set
MAX TIME
4 channels + Total : 26000
4 channels: 32000
3 channels: 43000
2 Channels: 65000
1 Channel:130000
4 o 5 channels: 1200 points per second
3 Channels: 2400 points per second
1 o 2 Channels: 4800 points per second
100 days
EXTERNAL USB Flash Memory (Option)
MAX STORING POINTS
MAX STORING SPEED
MAX TIME
See chapter 26
See Chapter 26
100 days
ZERO FUNCTION
PEAK FUNCTION
UNLOAD FUNCTION
KEY BLOCK FUNCTION
TOTAL FUNCTION
AUTOZERO FUNCTION
COUNTER FUNCTION
PROGRAMMABLE RESOLUTION
PROGRAMMABLE DIGITAL FILTER
POINT POSITION
100% (all measurement range)
POSITIVE AND NEGATIVE PEAK
YES
YES
YES
YES
YES
1 –2 - 5 –10 –20 –50 - 100
0 … 5
Programmable from 0 to 5
24V DIGITAL INPUT (OPTION)
4
RS232 INTERFACE (OPTION)
MAX Cable Length 13m
RS485 INTERFACE (OPTION)
MAX Cable Length 1000m
USB port type B for data communication
USB port type A for external flash memory (on front panel)
MAX Cable Length 3.5m
NOMINAL WORKING TEMPERATURE
MAX WORKING TEMPERATURE
STORAGE TEMPERATURE
TEMPERATURE EFFECTS on the measurements 10°C
a) on zero
b) on full scale
0... +50°C
0... +50°C
-20... +70°C
±0.005%
±0.005%
PROTECTION CLASS (EN 60529)
DESKTOP CONTAINER
DIMENSIONS (AxLxP) mm
WEIGHT
REAR AND FRONT PANEL
IP40
ALLUMINIUM
86x156x213mm
about 1.9 kg
ALLUMINIUM

Photos non contractuelles - Les spécications peuvent être modiées sans préavis - MP6Plus R0 - wimesure.fr • AH 2022
MP6P
P
l
l
u
u
s
s
USER GUIDE MO.MP6Plus.R7
Pag. 9
OPTION
Input Channels (up to 4)
2mV/V +/-10V, +/-5V , 4-20mA, 0-20mA, PT100
Analog Output
Current Output:
Voltage Output: max 20mA –RL min: 1kΩ
Note:
The refresh rate of the analog signals is equal to
the frequency of acquisition of the respective channels
in input.
Up to 4
4-20mA
0-5V 0-10V +/-10V +/-5V
24V Digital Input
4
Wireless Transmission of the measures
Frequency : 433MHz
Serial Communication
RS485 with protocol MODBUS and RS232
MP Supervisor communication Software
Internal Datalogger with clock Calendar
On Internal non volatile memory
Internal Datalogger with clock Calendar
USB typ
e A connector for Flash memories
Serial printer
VERSIO
N WITH HANDLE
11. PURCHASE CODES
Inputs
Power
Analog Output
Serial Output
Functions
Accessories
Digital Inputs
MP6P
X
XXX
XX
X
X
X
X
2
2 inputs
230
230 Vac
A1
1 output
S
RS232, RS485,
Printer
D
Data logger
Clock-Calendar
M
Handle
N
4 Digital Inputs
3
3 inputs
115
115Vac
A2
2 outputs
W
Wireless
Transmission
F
Datalogger
Clock-Calendar
USB Flash
Memory
4
4 inputs
24
24Vdc
A3
3 outputs
A4
4 outputs
Example: MP6P230 (MP6Plus power supply230Vac base version)
Example: MP6P224A2SM (MP6Plus 2 channels- power supply 24Vdc + 2 Analog output + Serial output + handle)
Examples: MP6P3115SF (MP6Plus 3 channels power supply 115Vac + Serial output + USB Flash Memory)
ALWAYS SPECIFY in the puchase order how to configure the input channels:
Example: CH1 = 4-20mA CH2 = 2mV/V CH3 = 10V CH4 = PT100
Note: For transducers of type potentiometer please select the channel with input +/-5V

Photos non contractuelles - Les spécications peuvent être modiées sans préavis - MP6Plus R0 - wimesure.fr • AH 2022
MP6PPlluussUSER GUIDE MO.MP6Plus.R7
Pag. 9
OPTION
Input Channels (up to 4)
2mV/V +/-10V, +/-5V , 4-20mA, 0-20mA, PT100
Analog Output
Current Output:
Voltage Output: max 20mA –RL min: 1kΩ
Note:The refresh rate of the analog signals is equal to
the frequency of acquisition of the respective channels
in input.
Up to 4
4-20mA
0-5V 0-10V +/-10V +/-5V
24V Digital Input
4
Wireless Transmission of the measures
Frequency : 433MHz
Serial Communication
RS485 with protocol MODBUS and RS232
MP Supervisor communication Software
Internal Datalogger with clock Calendar
On Internal non volatile memory
Internal Datalogger with clock Calendar
USB type A connector for Flash memories
Serial printer
VERSION WITH HANDLE
11. PURCHASE CODES
Inputs
Power
Analog Output
Serial Output
Functions
Accessories
Digital Inputs
MP6P
X
XXX
XX
X
X
X
X
2
2 inputs
230
230 Vac
A1
1 output
S
RS232, RS485,
Printer
D
Data logger
Clock-Calendar
M
Handle
N
4 Digital Inputs
3
3 inputs
115
115Vac
A2
2 outputs
W
Wireless
Transmission
F
Datalogger
Clock-Calendar
USB Flash
Memory
4
4 inputs
24
24Vdc
A3
3 outputs
A4
4 outputs
Example: MP6P230 (MP6Plus power supply230Vac base version)
Example: MP6P224A2SM (MP6Plus 2 channels- power supply 24Vdc + 2 Analog output + Serial output + handle)
Examples: MP6P3115SF (MP6Plus 3 channels power supply 115Vac + Serial output + USB Flash Memory)
ALWAYS SPECIFY in the puchase order how to configure the input channels:
Example: CH1 = 4-20mA CH2 = 2mV/V CH3 = 10V CH4 = PT100
Note: For transducers of type potentiometer please select the channel with input +/-5V
MP6P
P
l
l
u
u
s
s
USER GUIDE MO.MP6Plus.R7
Pag. 10
12. DIMENSIONS
STANDARD VERSION
VERSION WITH HANDLE

Photos non contractuelles - Les spécications peuvent être modiées sans préavis - MP6Plus R0 - wimesure.fr • AH 2022
MP6P
P
l
l
u
u
s
s
USER GUIDE MO.MP6Plus.R7
Pag. 11
13. TRANSPORTATION
The instrument is made by electronic parts.
In the case of transportation please pack the instrument with care.
Beware of strong shocks and moisture
14. POSITIONING
It shall be housed in a well ligthed position and protected by atmospherics agents.
15. INSTALLATION
This indicator has been produced in accordance with the standards for electromagnetic compatibility according to
Directive 2014/30 / EC.
To ensure that they are respected you must perform the electrical connections as described by the safety standards
and as indicated in this manual. For a correct and rapid installation to proceed in the following way:
1. Connect the transducers to the instrument
2. Verify the correct supply voltage stated on the rear panel between 230Vac, 115Vac, 24Vdc.
Supply the instrument with a wrong voltage can lead to its rupture.
3. Switch on the instrument.
The installation should be performed by trained personnel only.
16. MAINTENANCE
The instrument does not require periodic maintenance; in case of malfunctions contact the supplier.
Occasionally clean the front panel with a nonabrasive cloth of non-corrosive substances.
When the date and time indicated there is no longer correct, replace the internal battery (see cap.34).

Photos non contractuelles - Les spécications peuvent être modiées sans préavis - MP6Plus R0 - wimesure.fr • AH 2022
MP6PPlluussUSER GUIDE MO.MP6Plus.R7
Pag. 11
13. TRANSPORTATION
The instrument is made by electronic parts.
In the case of transportation please pack the instrument with care.
Beware of strong shocks and moisture
14. POSITIONING
It shall be housed in a well ligthed position and protected by atmospherics agents.
15. INSTALLATION
This indicator has been produced in accordance with the standards for electromagnetic compatibility according to
Directive 2014/30 / EC.
To ensure that they are respected you must perform the electrical connections as described by the safety standards
and as indicated in this manual. For a correct and rapid installation to proceed in the following way:
1. Connect the transducers to the instrument
2. Verify the correct supply voltage stated on the rear panel between 230Vac, 115Vac, 24Vdc.
Supply the instrument with a wrong voltage can lead to its rupture.
3. Switch on the instrument.
The installation should be performed by trained personnel only.
16. MAINTENANCE
The instrument does not require periodic maintenance; in case of malfunctions contact the supplier.
Occasionally clean the front panel with a nonabrasive cloth of non-corrosive substances.
When the date and time indicated there is no longer correct, replace the internal battery (see cap.34).
MP6P
P
l
l
u
u
s
s
USER GUIDE MO.MP6Plus.R7
Pag. 12
17. POWER ON
Connect the instrument to the power and tun on the instrument.
WARNING: For this operation, observe the safety rules
Give power to the instrument.
At power on the instrument displays the AEP transducers logo, the model of
the instrument, the software version installed and a set of information
specifying the installed options.
Cx : x=Number of channels Installed (x = 1 –2 -3 -4 )
Dx : x=1 : The DataLogger Function is installed
x=0 : The DataLogger Function is not installed
Tx : x=1: The Wireless Transmission option is installed
x=0 : The Wireless Transmission option is not installed
Sx : x =1 : The Serial Communication Option is installed
x=0 : The serial Communication Option is not installed
Ax : x = Number of analog output (x = 0 - 1 –2 –3 - 4 )
Ox : reserved
Ix : x=1: the function digital input is installed
x=0: the function digital input is not installed
This page is shown for 6 seconds, then MP6Plus is ready to shown the measures in the main page.
If you press any button this page remains blocked so you can take note of this setting.
To continue press any button again or wait about 30s.
WARNING: Power on must be done by trained personnel.
18. POWER OFF
Turn off the instrument

Photos non contractuelles - Les spécications peuvent être modiées sans préavis - MP6Plus R0 - wimesure.fr • AH 2022
MP6P
P
l
l
u
u
s
s
USER GUIDE MO.MP6Plus.R7
Pag. 13
19. DISPLAY and FUNCTION KEYS
The display of data and messages, is made through back-lit graphic LCD display (resolution 128x64 dots) on the front
panel of the instrument.
MP6Plus has six function keys: 2 of these, SET eZERO, have a dedicated function while the other 4, F1-F2-F3-F4, have a
programmable function.
SET:
In the main page provides access to the MENU parameters
Inside a MENU page its functions is CONFIRM and QUIT. So it returns to the MENU
ZERO:
In the main page allows to set to 0 the shown values (ZERO Function)
Inside some MENU Pages (if some value must be modified) it allows to set the
moltiplicative factor to speed up the data setting
F1-F2-F3-F4
In the main page the F1, F2, F3, F4 function is shown in the dedicated row.
Inside the MENU pages they assume the following meaning
F1: assume the + function: it increments the value of the parameter in editing
F2: assume the function . It moves to the previous parameter
F3: assume the - function: it decrements the value of the parameter in editing
F4: assume the function
. It moves to the next parameter
When a key is pressed the instruments beeps.
Press SET to enter in the programming MENU and set the various parameters.
The various menu items can be scrolled with the cursor keys and .
You confirms a selection with the SET button.
It is possible return to the main page by selecting MENU-> QUIT or using the short-cut key F1+ or F3 -

Photos non contractuelles - Les spécications peuvent être modiées sans préavis - MP6Plus R0 - wimesure.fr • AH 2022
MP6PPlluussUSER GUIDE MO.MP6Plus.R7
Pag. 13
19. DISPLAY and FUNCTION KEYS
The display of data and messages, is made through back-lit graphic LCD display (resolution 128x64 dots) on the front
panel of the instrument.
MP6Plus has six function keys: 2 of these, SET eZERO, have a dedicated function while the other 4, F1-F2-F3-F4, have a
programmable function.
SET:
In the main page provides access to the MENU parameters
Inside a MENU page its functions is CONFIRM and QUIT. So it returns to the MENU
ZERO:
In the main page allows to set to 0 the shown values (ZERO Function)
Inside some MENU Pages (if some value must be modified) it allows to set the
moltiplicative factor to speed up the data setting
F1-F2-F3-F4
In the main page the F1, F2, F3, F4 function is shown in the dedicated row.
Inside the MENU pages they assume the following meaning
F1: assume the + function: it increments the value of the parameter in editing
F2: assume the function . It moves to the previous parameter
F3: assume the - function: it decrements the value of the parameter in editing
F4: assume the function . It moves to the next parameter
When a key is pressed the instruments beeps.
Press SET to enter in the programming MENU and set the various parameters.
The various menu items can be scrolled with the cursor keys and .
You confirms a selection with the SET button.
It is possible return to the main page by selecting MENU-> QUIT or using the short-cut key F1+ or F3 -
MP6P
P
l
l
u
u
s
s
USER GUIDE MO.MP6Plus.R7
Pag. 14
How to Modify Numeric Values
To modify a numeric value please use the Keys +/-.
The + key increments its value while the –key decrements its value.
If you keep the keys + /- pressed the value increments/decrements fast.
The increment/decrement step is choosen by using the FATT value.
The FATT value is modified using the ZERO key and its actual value is always shown in the last row.
(see picture below).
The value of FATT is initialized to the minimum value of the parameter in edit (according to the decimal point position
with which it is set).
When you press the button ZERO the multiplication factor increases by a factor of 10 until it reaches its maximum
value. Then it resets again to the minimum value.
SETUP
23/03/14
11:56:47
P1: 200.00 kN 0.400000 mV/V
P2: 400.00 kN 0.800000 mV/V
P3: 600.00 kN 1.200000 mV/V
P4: 800.00 kN 1.600000 mV/V
P5: 1000.00 kN 2.000000 mV/V
ZERO KEY: FATT = 0.01

Photos non contractuelles - Les spécications peuvent être modiées sans préavis - MP6Plus R0 - wimesure.fr • AH 2022
MP6P
P
l
l
u
u
s
s
USER GUIDE MO.MP6Plus.R7
Pag. 15
20. MAIN PAGE
The main page is the operative page of the instrument MP6Plus.
It can assume different onfiguration according to the number of connected channels and to the channels enabled.
Note : The Date and Time are present only if the DATALOG option is installed
In case of just 1 channel enable (or when the ZOOM
mode is active) the main page appears as the
picture aside.
The measure is shown enlarged for a better
visibility.
In the case of 2 channels are enabled and the
funtion TOTAL is enable enabled the page looks like
in the figure.
If you do not assign any function to a function key it
is not displayed and the button is disabled. In this
case the function keys F3 and F4 have been
disabled.
The picture aside shows how the main page
appears with 2 channels enabled.
It can be seen as the character with which
measurements are showing adapts automatically to
make more evident the measures.
With the ZERO key it is possible to set or reset ZERO function.
When the ZERO function is active the dedcated LED near the button is on.
To remove the ZERO keep the key pressed for 4-5 seconds.
The ZERO led will be turned off when the ZERO is removed.
Using the F1 –F2 - F3 e F4 keys it is possible to perform different functions. The active function for each key is
selected inside the page : Main Menu
Configuration.
When the funtion is active its description will be shown in REVERSE
Each key each can assume one of the following function
PRINT –PEAK –DATALOG –HOLD –TX DATA –UNLOAD –ZOOM –COUNTER.
When you select the function as OFF, the key is disabled. It has no function and nothing is displayed.
From the MAIN PAGE it is possible to go to the MAIN MENU using the key SET.
It is not possible enter in the MAIN MENU when:
A DATALOG cycle is active .
An USB - RS232 –RS485communication is active.
DATALOG
15-04
08:17
PEAK
12.3467
kg
HOLD CH2 UNLOAD
DATALOG
15-04
08:17
UNLOAD
CH1 : 152.60 kg
CH2 : 153.72 kg
TOT : 305.32 kg
DATALOG
15-04
08:17
PEAK
CH1 :
152.60
kg
CH2 :
153.72
kg
HOLD UNLOAD

Photos non contractuelles - Les spécications peuvent être modiées sans préavis - MP6Plus R0 - wimesure.fr • AH 2022
MP6PPlluussUSER GUIDE MO.MP6Plus.R7
Pag. 15
20. MAIN PAGE
The main page is the operative page of the instrument MP6Plus.
It can assume different onfiguration according to the number of connected channels and to the channels enabled.
Note : The Date and Time are present only if the DATALOG option is installed
In case of just 1 channel enable (or when the ZOOM
mode is active) the main page appears as the
picture aside.
The measure is shown enlarged for a better
visibility.
In the case of 2 channels are enabled and the
funtion TOTAL is enable enabled the page looks like
in the figure.
If you do not assign any function to a function key it
is not displayed and the button is disabled. In this
case the function keys F3 and F4 have been
disabled.
The picture aside shows how the main page
appears with 2 channels enabled.
It can be seen as the character with which
measurements are showing adapts automatically to
make more evident the measures.
With the ZERO key it is possible to set or reset ZERO function.
When the ZERO function is active the dedcated LED near the button is on.
To remove the ZERO keep the key pressed for 4-5 seconds.
The ZERO led will be turned off when the ZERO is removed.
Using the F1 –F2 - F3 e F4 keys it is possible to perform different functions. The active function for each key is
selected inside the page : Main Menu Configuration.
When the funtion is active its description will be shown in REVERSE
Each key each can assume one of the following function
PRINT –PEAK –DATALOG –HOLD –TX DATA –UNLOAD –ZOOM –COUNTER.
When you select the function as OFF, the key is disabled. It has no function and nothing is displayed.
From the MAIN PAGE it is possible to go to the MAIN MENU using the key SET.
It is not possible enter in the MAIN MENU when:
A DATALOG cycle is active .
An USB - RS232 –RS485communication is active.
DATALOG
15-04
08:17
PEAK
12.3467
kg
HOLD
CH2
UNLOAD
DATALOG
15-04
08:17
UNLOAD
CH1 : 152.60 kg
CH2 : 153.72 kg
TOT : 305.32 kg
DATALOG
15-04
08:17
PEAK
CH1 : 152.60 kg
CH2 : 153.72 kg
HOLD
UNLOAD
MP6P
P
l
l
u
u
s
s
USER GUIDE MO.MP6Plus.R7
Pag. 16
21. ERROR MESSAGES
The instrument handles different types of error reported in the following way:
The indication -ERROR-indicates that the instrument can not display the measure. Possible causes:
A problem of connection of the load cell.
The channel calibration is not correct.
It is selected a 6-wire connection instead uses 4 wires.
The measure in the current unit of measurement has an excessive number of significant digits excessive and
can not represent
In the event that the measure is higher than the range of the transducer the value is displayed in REVERSE to make
clear that you are creating an anomalous situation. When the measure becomes greater than 135% compared to the
full scale of the transducer is given the indication of OVERLOAD to signal the situation of danger.
ATTENTION: we are in the presence of a risk to the safety of people, things and the transducer that could
break irreparably. Bring to zero (UNLOAD) immediately transducer.
CAUTION: If the transducer has suffered an overload could be altered its metrological characteristics. It is
therefore advisable Metrological Confirmation" or a new setting to make sure that the transducer still
complies with its original metrological characteristics.
IT IS RECOMMENDED that the operator records the natural zero of the transducer (in mV/V switching off the ZERO
function) in order to check possible high changes after an overload.
If the variation is important the transducer is permanently deformed.
In this case it is necessary to substitute the transducer
22. ZERO FUNCTION
The ZERO function is activated with the dedicated key ZERO (or remotely through a digital
input) and is used to set to zero the display of a channel.
The function acts on the entire field of measurement (100%).
The status of the function is stored permanently.
When the function is activated, the corresponding LED is illuminated. Each press of the button performs a new ZERO.
To remove the ZERO, hold the key for 4-5 seconds. The ZERO LED turns off when accepted the command.
Each channel can be enabled / disabled to perform the function of ZERO(in the page MENU-> Channels ENABLE).
The disabling to perform the ZERO can be linked to the nature of the transducer connected (eg. Temperature
transducer or absolute pressure) or to application needs.
In PEAK mode pressing ZERO resets the maximum registers and prepare the instrument for a new measure of PEAK.
When the ZOOM function is active the ZERO is performed only on the displayed channel.
DATALOG
15-04
08:17
PEAK
CH1 : -ERROR-kg
CH2 : 153.72 kg
TOT : OVERLOAD kg
PRINT UNLOAD

Photos non contractuelles - Les spécications peuvent être modiées sans préavis - MP6Plus R0 - wimesure.fr • AH 2022
MP6P
P
l
l
u
u
s
s
USER GUIDE MO.MP6Plus.R7
Pag. 17
23.PEAK FUNCTION
The PEAK function is used to display the
maximum (PEAK +) or minimum (PEAK-) of a
dynamic measurement.
For example, in applications where it needs to
register the torque generated by a torque
wrench after shooting, or the maximum force
generated by a body in free fall, or to record the
pressure peaks generated by a hydraulic
intensifier or a hydraulic pump unit, to measure
the maximum displacement and minimum of a
body subjected to vibrations etc ...
MP6Plus allows to select the function PEAK+ o
PEAK-.
See the picture to verify how the PEAK+
(POSITIVE) and PEAK+- (NEGATIVE) works.
The PEAK function is enabled when at least one
of the Function keys F1 –F2 -F3 or F4 is
associated to the function PEAK (see MAIN
MENU->CONFIGURATION) or remotely through a digital input.
The function is not permanently stored. If you enter in the MAIN MENU or when you turn off the intsrument the
function is disabled.
The way in which the function PEAK is displayed depends on the number of enabled channels.
If only one channel is enabled, when the function is activated, the display shown the MIN and MAX and the value of
the ACTUAL value measured by the transducer in real time.
In the case of two channels enabled, the peak function displays the ACTUAL value, the MAX value in the case of PEAK
+or the value of MIN if PEAK-.
In all other cases, you receive:
The ACTUAL value in the case of PEAK OFF
The MAX value in the case of PEAK +
The MIN value in the case of PEAK -
When the PEAK function is active the message PEAK will be displayed in REVERSE
PEAK values can be reset using the ZERO key.
DATALOG
15-04
08:17
PEAK+
CH1 : 152.60 kg
CH2 : 153.72 kg
HOLD UNLOAD

Photos non contractuelles - Les spécications peuvent être modiées sans préavis - MP6Plus R0 - wimesure.fr • AH 2022
MP6PPlluussUSER GUIDE MO.MP6Plus.R7
Pag. 17
23.PEAK FUNCTION
The PEAK function is used to display the
maximum (PEAK +) or minimum (PEAK-) of a
dynamic measurement.
For example, in applications where it needs to
register the torque generated by a torque
wrench after shooting, or the maximum force
generated by a body in free fall, or to record the
pressure peaks generated by a hydraulic
intensifier or a hydraulic pump unit, to measure
the maximum displacement and minimum of a
body subjected to vibrations etc ...
MP6Plus allows to select the function PEAK+ o
PEAK-.
See the picture to verify how the PEAK+
(POSITIVE) and PEAK+- (NEGATIVE) works.
The PEAK function is enabled when at least one
of the Function keys F1 –F2 -F3 or F4 is
associated to the function PEAK (see MAIN
MENU->CONFIGURATION) or remotely through a digital input.
The function is not permanently stored. If you enter in the MAIN MENU or when you turn off the intsrument the
function is disabled.
The way in which the function PEAK is displayed depends on the number of enabled channels.
If only one channel is enabled, when the function is activated, the display shown the MIN and MAX and the value of
the ACTUAL value measured by the transducer in real time.
In the case of two channels enabled, the peak function displays the ACTUAL value, the MAX value in the case of PEAK
+or the value of MIN if PEAK-.
In all other cases, you receive:
The ACTUAL value in the case of PEAK OFF
The MAX value in the case of PEAK +
The MIN value in the case of PEAK -
When the PEAK function is active the message PEAK will be displayed in REVERSE
PEAK values can be reset using the ZERO key.
DATALOG
15-04
08:17
PEAK+
CH1 : 152.60 kg
CH2 : 153.72 kg
HOLD
UNLOAD
MP6P
P
l
l
u
u
s
s
USER GUIDE MO.MP6Plus.R7
Pag. 18
24. HOLD FUNCTION
The HOLD function freezes the display of the instrument at the time the button is pressed.
The HOLD function is enabled by associating this function to one of the function keys F1 - F2 - F3 or F4 (see MAIN
MENU
Configuration).
Entering the main menu or by switching the instrument off the instrument the function is automatically disabled.
In PEAK mode the HOLD function only locks the display of the current value. The values of MIN and MAX are always
updated.
The activation of the function is displayed on the LCD display of the instrument with the message HOLD in reverse.
It is possible to activate the HOLD also by a external Digital Input (see dedicated chapter) when properly programmed.
CAUTION: When the function is activated the indication is locked, so if the measurement changes may
present danger to the safety of persons, property or breaking transducer associated to a channel overload.
25. UNLOAD FUNCTION
The function UNLOAD allows to determine the amount of material that is downloaded for example from a tank.
The UNLOAD function is enabled by associating this function to one of the function keys F1 - F2 - F3 or F4 (see MAIN
MENU
Configuration).
It is possible to activate the UNLOAD also by a external Digital Input (see dedicated chapter) when properly
programmed.
The activation of the function is displayed on the LCD display of the instrument with the message UNLOAD in reverse.
When activated, the display it resets the measurements and shows the loads unloaded.
When this function is selected, you can activate one or more relay outputs to achieve a small automation.
When completed the operation UNLOAD press the function key for its deactivation.
It is possible to consider the UNLOAD measurement as a positive or negative quantity (see MAIN MENU
Configuration).
DATALOG
15-04
08:17
PEAK
CH1 : 152.60 kg
CH2 : 153.72 kg
CH3 : 153.00 kg
CH4 : 154.36 kg
TOT : 613.68 kg
HOLD UNLOAD
DATALOG
15-04
08:17
PEAK
CH1 : 152.60 kg
CH2 : 153.72 kg
CH3 : 153.00 kg
CH4 : 154.36 kg
TOT : 613.68 kg
HOLD UNLOAD

Photos non contractuelles - Les spécications peuvent être modiées sans préavis - MP6Plus R0 - wimesure.fr • AH 2022
MP6P
P
l
l
u
u
s
s
USER GUIDE MO.MP6Plus.R7
Pag. 19
26. DATA LOGGER Function (Option)
To set DATA LOGGER parameters refere to the chapter MAIN MENU
DATALOGGER MENU
Data Logger
Parameters.
The DATALOG function is enabled by associating this function to one of the function keys F1 - F2 - F3 or F4 (see MAIN
MENU
Configuration).
The activation of the function is displayed on the LCD display of the instrument with the message DATALOG in
reverse.
If you have set the data storage on FLASH MEMORY (OPTION) it is necessary that this has already been inserted or the
command is not accepted.
Additionally it is necessary that any USB cable for communication with a PC has been removed
If you have set a threshold (threshold parameter <> 0), no storage will be made until this condition has been satisfied.
Each time a new point has been stored the icon will be shown on the display .
In case of DATALOG on FLASH MEMORY (OPTION) and with the parameter
interval different from MANUAL, two different files format can be created
:
.csv files : this file can be read directly by Microsoft Excel.
The field separator can be selected inside the page MAIN MENU
DATALOGGER MENU
Files Flash memory
.bin files: the file format of this file is descrpted below and allows
smaller file size and a faster storage.
To read these file it is possible to use the software MP Supervisor
The selection between these two files is made in the page MAIN MENU
DATALOGGER MENU
Files Flash memory
In both case the name of the file created is : MP Plus DataLogger DD:MM:YYYY –HHMMSS
where:
DD:MM:YYYY is the starting date (Day-Month-Year) of the datalog cycle
HHMMSS is the starting time (Hour-Minutes-Seconds) of the datalog cycle
This convention makes the file name unique and recognizable and can not be duplicated even later when you will go
to analyze it in your PC.
The format of the file created is described below.
WARNING: It is responsibility of the operator to check that within the Pen USB there is enough free space
consistent with the cycle set.
DATALOG
15-04
08:17
PEAK
CH1 : 152.60 kg
CH2 : 153.72 kg
HOLD UNLOAD