Winters LCD11 Series User manual

LCD11 Series Plug and Play Digital Display
Operation Manual

五、Dimension.............................................................................................
三、Technical Specifications.....................................................................
四、Function & Instructions.......................................................................
一、 ...........................................................................General Description
二、Features................................................................................................
1
1
1
1
9
1、Electrical Connections......................................................................
2、Key Operations................................................................................
3、Parameter Settings..........................................................................
4、Measurement Unit............................................................................
5、Smart Error Alarm............................................................................ 8
7
3
3
1
CONTENT
LCD11 Series Plug and Play Digital Display Operation Manual LCD11 Series Plug and Play Digital Display Operation Manual
LCD11 Series Plug and Play Digital Display Operation Manual
LCD11 digital display is used to transfer 4-20mA input signal into measuring data and
display it. The product has high accuracy, good stability, easy operate features.
● 24-bit high accuracy Sigma-Delta ADC sampling
● LCD11 backlight design, can be used in dark environment
● Reverse polarity protection, protection against lighting surge
● 35 kinds of display units for option
● At most 2-channel PMOS output
● Display range: -19999~99999
● Accuracy: 0.1%F.S
● Environment temperature: -20℃~85 ℃
● Environment humidity: 20% ~ 80%
● Power: 24V/3mA
Pin configuration (Table 1):
Pin configurations: power & switch connector
DIN 43650 LCD11-B
1 PIN 24VDC 24VDC 24VDC
2 PIN 0V 0V 0V
3 PIN ALMSW1 ALMSW1 NC
PE PIN ALMSW2 SWCOM/LED1 PE
Table 1
Technical Specifications
General Description
Features
Function & Instructions
1. Electrical Connections
LCD11-A LCD11-N
1

Fig1:Connector Fig2:Panel
Wiring diagram
LCD11 Series Plug and Play Digital Display Operation Manual LCD11 Series Plug and Play Digital Display Operation Manual
The key operations are described below (Table 2):
Key Operating Instructions
M
Parameter selection and setting key: switching parameters, save the setting
parameters data.
S
Cursor selection key: When setting parameters, select the revised data bit ,click
the key once then the cursor shift left one bit, when the cursor shift to the highest
bit it will automatically return to the lowest bit, and followed by cycle.
Z
Value modification key: Modify the current data bit value when you set the
parameter, incremented by one, and automatically click the key once the value is
return to the lower limit when the value is incremented by the upper limit, followed
by cycle. Press and hold the key the value automatically successive increments,
followed by cycle.
When the system is in measurement condition, press the "M" key to enter the setup
menu, set CLK as 132, Press "M" key to set other parameters in turn, and it will turn
back to measurement recondition when all parameters setting a over.
Parameter information is as follows (Table 3):
Par Name Range Specifications
CLK
Parameters
setting
unlocked
0~255 CLK=132, Unlocked, can set other parameters.
CLK ≠ 132, locked,can't set other parameters.
SLL Display
lower limit
-19999~99999 Display lower limit setting value of
Table 2
2. Key Operations
3. Parameter Settings
32

SLH Display
upper limit -19999~99999 Display upper limit setting value of
UNI Display unit 0~33 Setting the display units (see Table 3)
DISI
LCD
Display
information
DISI=0
The main screen displays the input current value,
vice screen displays the percentage of current
data and the current unit in mA.
DISI=1
The main screen displays the converted
measured value, vice screen displays the
percentage of data and data unit
DISI=2
The main screen displays the exact percentage
of the measurement data (keep two significant
figures after the decimal point), and vice screen
displays the percentage of dat a and the symbol
"%".
DECP
Decimal
point
position
DECP=0 No decimal point: XXXXX
DECP=1 Decimal point position: XXXX.X
DECP=2 Decimal point position: XXX.XX
DECP=3 Decimal point position: XX.XXX
PB1 Display zero -19999~99999 LCD main screen display zero migration value
KK1 Display range 0~1.9999 LCD main screen display range proportion
SP1 Upper limit of
Out1 -19999~99999
Upper limit of set value in the first channel
transistor output
Par Name Range Specifications
LCD11 Series Plug and Play Digital Display Operation Manual LCD11 Series Plug and Play Digital Display Operation Manual
RP1 Lower limit
of Out1 -19999~99999
Lower limit set value in the first of channel
transistor output
DTC1 Out1 output
delay
0~255(ms) Delay time before the first channel output be valid
DTS1 Out1 reset
delay 0~255(ms)
Delay time before the first channel output
be reset
MOD1
Out1
Output
mode
MOD1=0 No output, Out1 always keep the reset state.
MOD1=1 When measured values > SP1, Out1 be valid;
When measured value < RP1, Out1 be reset.
MOD1=2
When measured values > SP1, Out1 be reset;
when measured value < RP1, Out1 be valid.
MOD1=3
When RP1 < measured value < SP1, Out1 be
valid; SP1 or measured When measured values >
value < RP1, Out1 be reset.
MOD1=4
When measured values > SP1 or measured value
< RP1, Out1 be valid ; When RP1< measured
value < SP1, Out1 be reset.
SP2 Upper limit
of Out2 -19999~99999
Upper limit set value in the second of channel
transistor output
RP2 Lower limit
of Out2 -19999~99999
Lower limit set value in the second of channel
transistor output
DTC2
Out2 output
delay 0~255(ms)
Delay time before the second channel output be
valid
Par Name Range Specifications
54

DTS2
Out2 reset
delay
0~255(ms)
Delay time before the second channel output be
reset
MOD2
Out2
Output
mode
MOD2=0 No output,Out2 always keep the reset state.
MOD2=1 When measured values>SP2, Out1 be valid;
When measured value <RP2, Out2 be reset.
MOD2=2 When measured values>SP2, Out2 be reset;
When measured value <RP2, Out2 be valid.
MOD2=3
When RP2<measured value <SP2, Out2 be valid;
When measured values > SP2 or measured value
< RP2, Out2 be reset.
MOD2=4
When measured values>SP2 or measured value
<Rp2, Out2 be valid ; When RP2 < measured
value < SP2, Out2 be reset.
Note :
● When modify any parameters, the current data modification bit will be flashing,
through the "S" key to switch.
● If the parameter values need to be modified from a positive number to a negative,
change the sign bit at the highest bit. For example, to modify parameter value
from 100 to -100, LCD -A/B main screen will display 11 “-0100”.
● When the set value of the parameter exceeds the valid range, the system
automatically save the befsetting data ore. For example, SLL parameter
Table 3
Par Name Range Specifications
LCD11 Series Plug and Play Digital Display Operation Manual LCD11 Series Plug and Play Digital Display Operation Manual
range is -19999~99999, SLL is 0 before modification, then modify SLL to 2000,
the system will display SLL in 0.
● When the parameters value higher data bit is 0 value, the system automatically
optimizes the visual effects (not display), press the "S" key to switch the data
modifications data bit to restore the display. For example, SLL is 0, LCD11 to this
main screen displays "0" instead of "0000".
Par Val Unit
Note : When measurement unit is set to "mH2O" or "mmH2O" or “cmH2O"
influenced by the LCD -digit lim11 it, they are actually displayed as“mHO"
"mmHO" "cmHO".
Table 4
UNI=0 kPa UNI=9 mmHg UNI=18 ℉ UNI=27 mm
UNI=1 MPa UNI=10 TOR UNI=19 K UNI=28 inch
UNI=2 Pa UNI=11 atm UNI=20 RH UNI=29 m/s
UNI=3 bar UNI=12 Kg UNI=21 VOL UNI=30 ohm
UNI=4 mbar UNI=13 g UNI=22 PPM UNI=31 Kohm
UNI=5 psi UNI=14 mg UNI=23 LEL UNI=32 mV
UNI=6 mH2O UNI=15 N UNI=24 PH UNI=33 V
UNI=7 mmH2O UNI=16 KN UNI=25 m
UNI=8 cmH2O UNI=17 ℃ UNI=26 cm
Par Val Unit Par Val Unit Par Val Unit
4. Measurement Unit (Table 4):
76

LCD has a power-on self11 -check function, a numeric code to reflect the run error
and "M" key to clear the alarm. The error code as follows flashing alarm, press the
(Table 5):
Err Code Specifications
ERR06 The wrong parameters value relationship between SLL and
SLH. Should meet: SLH ≠ SLL
ERR07 The wrong parameters value relationship between RP1 and
SP1. Should meet: RP1 ≤ SP1.
ERR08 RP2 relationship with SP2 parameter values error,
Should meet: RP2 ≤ SP2.
ERR10 CLK parameter value exceeds the valid range,
Should meet: 0 ≤ CLK ≤ 255.
ERR011 SLL parameter value exceeds the valid range,
Should meet: -19999 ≤ SLL ≤ 99999.
ERR012 SLH parameter value exceeds the valid range,
Should meet: -19999 ≤ SLH ≤ 99999.
ERR013 UNI parameter value exceeds the valid range,
Should meet: 0 ≤ UNI ≤ 33.
ERR014 DISI parameter va lue exceeds the valid range,
Should meet: 0 ≤ DISI ≤ 2.
ERR015 DECP parameter val ue exceeds the valid range,
Should meet: 0 ≤ DECP ≤ 3.
ERR016 PB1 parameter value exceeds the valid range,
Should meet: -19999 ≤ PB1 ≤ 99999.
ERR017 KK1 parameter value exceeds the valid range,
Should meet: 0 ≤ KK1 ≤ 1.9999.
5. Smart Error Alarm
ERR019 SP1 parameter value exceeds the valid range,
Should meet: 0 ≤ SP1 ≤ 99999.
ERR020 RP1 parameter value exceeds the valid range,
Should meet: 0 ≤ RP1 ≤ 99999.
LCD11 Series Plug and Play Digital Display Operation Manual LCD11 Series Plug and Play Digital Display Operation Manual
Err Code Specifications
ERR021 DTC1 parameter value exceeds the valid range,
Should meet: 0 ≤ DTC1 ≤ 255.
ERR022 DTS1 parameter value exceeds the valid range,
Should meet: 0 ≤ DTS1 ≤ 255.
ERR023 MOD1 parameter value exceeds the valid range,
Should meet: 0 ≤ MOD1 ≤ 4.
ERR024 SP2 parameter value exceeds the valid range,
Should meet: 0 ≤ SP2 ≤ 99999.
ERR025 RP2 parameter value exceeds the valid range,
Should meet: 0 ≤ RP2 ≤ 99999.
ERR026 DTC2 parameter value exceeds the valid range,
Should meet: 0 ≤ DTC2 ≤ 255.
ERR027 DTS2 parameter value exceeds the valid range,
Should meet: 0 ≤ DTS2 ≤ 255.
ERR028 MOD2 parameter value exceeds the valid range,
Should meet: 0 ≤ MOD2 ≤ 4.
Table 5
Dimensions (Unit: mm)
98

Display detailed operating instructions
LCD11 Series Plug and Play Digital Display Quick Installation Guide
Display interface Restore factory settings
Locked
Modify lower limit
of the range
Modify upper limit
of the range
Revision process unit
Modify the
function
interface
Decimal position
Zero shift
Range shift
Identify and select parameters for the next step
Modify parameters
Interface Model
The number of
decimal places
Non decimal
1 decimal
2 decimal
3 decimal
4 decimal
Process units
Non
Parameter table
Restore factory settings
Confirm restore factory settings
Selection parameters
To confirm and return function interface
Modify upper limit of the range
Modify upper limit of the range
To confirm and into the next menu
Current parameters
Parameter settings locked
Parameter settings locked
Number of decimal places
Revision process unit
Function interface model
Display zero shift
Display range shift
Current parameters
Current parameters
Current parameters
Current parameters
Current parameters
Current parameters
Current parameters
Current parameters
Identify and select parameters for the next step
Identify and select parameters for the next step
Identify and select parameters for the next step
Identify and select parameters for the next step
Identify and select parameters for the next step
Identify and select parameters for the next step
To confirm and get into the next menu
Modify parameters
Modify parameters
Modify parameters
Modify parameters
Modify parameters
Modify parameters
Modify parameters
Modify parameters
Display range shift
Display zero shift
Number of decimal places
Function interface model
Revision process unit
Modify lower limit of the range
Modify lower limit of the range
Completion parameter unlock
Selection parameters
Selection parameters
Selection parameters
Selection parameters
Selection parameters
Selection parameters
Selection parameters
Next menu
Next menu
Next menu
Next menu
Next menu
Next menu
Return function interface
5 seconds in the next step
To confirm and get into the next menu
To confirm and get into the next menu
To confirm and get into the next menu
To confirm and get into the next menu
To confirm and get into the next menu

LCD11 Series Plug and Play Digital Display Quick Installation Guide
With 1 way on-off alarm output detailed operating instructions
Parameter table
Function output
Display interface
Set open value
Reset value
Reset delay value
Reset delay value
Set output
function
5 seconds in the next step
Next menu
Next menu
Next menu
Next menu
Next menu
Current parameters
Current parameters
Current parameters
Current parameters
Current parameters
Current parameters
Selection parameters
Selection parameters
Selection parameters
Selection parameters
Selection parameters
Identify and select parameters for the next step
Identify and select parameters for the next step
Identify and select parameters for the next step
Identify and select parameters for the next step
Modify parameters
Modify parameters
Modify parameters
Modify parameters
Modify parameters
To confirm and get into the next menu
Set open value
Reset value
Reset delay value
Reset delay value
Set output function
Modify open value
Modify reset value
Modify reset delay value
Modify reset delay value
Modify output function
To confirm and get into the next menu
To confirm and get into the next menu
To confirm and get into the next menu
To confirm and get into the next menu

LCD11 Series Plug and Play Digital Display Quick Installation Guide
With 2 way on-off alarm output detailed operating instructions
Parameter table
Function output
Display interface
Set open value
Reset value
Reset delay value
Reset delay value
Set output
function
5 seconds in the next step
Next menu
Next menu
Next menu
Next menu
Next menu
Current parameters
Current parameters
Current parameters
Current parameters
Current parameters
Current parameters
Selection parameters
Selection parameters
Selection parameters
Selection parameters
Selection parameters
Identify and select parameters for the next step
Identify and select parameters for the next step
Identify and select parameters for the next step
Identify and select parameters for the next step
Modify parameters
Modify parameters
Modify parameters
Modify parameters
Modify parameters
Set open value
Reset value
Reset delay value
Reset delay value
Set output function
Modify open value
Modify reset value
Modify reset delay value
Modify reset delay value
Modify output function
To confirm and get into the next menu
To confirm and get into the next menu
To confirm and get into the next menu
To confirm and get into the next menu
To confirm and get into the next menu