1. OPERATOR MODE
Status bar
Under the window with video signal is status bar, which display informaon about state of system (Live Stream,
measurement, etc.) on the le, in the middle, according to sengs in tab App. Sengs, informaon about
current zoom of video signal or temperature of camera or current value of frame per second and resoluon
of camera. On the right is indicator Time to Calib. which counts down the me to the next automac image
compensaon. During automac image compensaon is video signal paused for moment.
Temperature range and palee of thermographic image
To the right of window with video signal is temperature palee of thermographic image. All temperatures be-
tween maximal and minimal limits (right of palee) are connuously allocang color. From palee can be easy
read what temperature has specific area in the thermographic image.
Buon Full Screen
This buon will launch the Full Screen Mode, which is described in Chapter 3. This buon can be pressed only
when is acve streaming image from the camera and running inspecon. Because this applicaon is only display
and it is not possible to set anything or launch.
Table of definion of measurement area (ROI)
To the right of palee is table containing sengs for each defined measurement area. Each area in image (ROI)
corresponds to one row in the table. Table contains following columns:
ROI – Numeric idenfier which assigns row from table to specific ROI in image.
Trigger – Indicate which trigger is assigned to each ROI.
Stasc – Stasc which is evaluated in ROI. Stasc is calculated for all pixels in ROI.
Current Value – Current value of stasc in ROI.
Criterion – Criterion which is in stasc checked during the measurement.
Last Result – Value of criterion aer last measurement.
Operator – Comparison operator with which is evaluated Last Result against Criterion Value during mea-
surement.
Criterion Value – Set reference value of criterion.
Outputs 1-5 – In these columns are marked by cross linkage row to digital output of system. Etc. when all
condions are met, marked output is acve.
The table in Figure 1.5 is primary used to representaon current seng criterion, reference value, performing
operator and last measured value. If the last measured value of criterion met the condion of comparison op-
erator and reference value, whole row is colored green aer each measurement. Otherwise, row is colored
red.
In the last row of table on the le is displayed seng behavior of digital output. Are used two opon All and
One. First opon All indicate that for acvate output in the column (Outputs 1-5) have to be met condions in
all rows which have the cross in corresponding column (or row aer the measurement is green). Second opon
One indicate that for acve output in the column have to be met condions in at least of one rows which have
the cross in corresponding column.
Headers of individual column of outputs labeled 1-5 is lighted up yellow when the digital output is acve.