Manual for use

2 IE-COM-D4141-09
Instruction manual for use of thermo-hygro-barometer COMMETER D4141
Instrument is designed for measurement and data logging of temperature, relative humidity,
atmospheric pressure and pressure tendency. It enables to display the dew point temperature and
atmospheric pressure corrected to the sea level. Measured values are displayed on a dual line LCD
display. Data logging of measured values into internal non volatile memory in adjustable interval is
enabled. Any time after logging is finished data transfer to a personal computer is possible via an
RS232 serial link. Temperature is measured by RTD sensor Ni1000/6180ppm. Instrument is
equipped with an external temperature - relative humidity probe. Instrument compares all measured
values (except pressure tendency) with two adjustable levels for each measured quantity. Breaking
the level is indicated by blinking the proper value on display and by audio indication (switchable).
Instrument is equipped with minimum and maximum memory and Hold function. Minimum and
maximum values and Hold value are possible to display on the LCD anytime.
Technical parameters:
Parameters of measurement:
Temperature:Range of measurement: internal sensor (INT) -10 to +60 °C
external sensor (EXT) -30 to +105 °C (temperature
from +85 to +105°C with the limitation accordingly to
graph below)
Resolution: 0.1 °C
Accuracy: ±0.4 °C
Humidity: Range measurement: 0 to 100 %RH
Resolution: 0.1 %RH
Accuracy: ±2.5 % RH from 5 to 95 %RH at 23°C
Dew point temperature (calculated from temperature and humidity):
Range: -50 to +80 °C
Resolution: 0.1 °C
Accuracy: ±0.5 °C at the range of 30 to 95 % RH
Atmospheric pressure: Range of measurement: 800 to 1100hPa
Resolution: 0.1 hPa
Accuracy: ± 2hPa at ambient temperature of 23°C
Measuring interval and display reading refresh: approximately 0.7 s in FAST mode
0.7 to 5 s in dynamic mode
Interval of refresh of pressure tendency display: 0.5 h
Response time with stainless sensor cover (air flow approximately 1m/s):
temperature (EXT): t90 < 75 s (temperature step 20 °C)
relative humidity: t90 < 60 s (humidity step 30 %RH, constant temperature)
Logging interval:
Logging interval in automatic logging mode: from 10 s to 24 h
Total memory capacity: for manual logging mode 1000 value sets
for automatic noncyclic logging mode 4062 values
for automatic cyclic logging mode 3938 values
The above specified numbers are maximum possible. Capacity is slightly lower if
logging was interrupted by communication with computer, by stopping logging or
switching OFF the instrument.
Communication with computer: RS232 (serial port)
Real time clock: adjustable from computer, integrated calendar including leap years
Real time clock accuracy: < 200 ppm (i.e. 0.02 %, 17,28 s in 24 h)
Power: battery 9V or ac/dc adapter 12V with NiCd accumulator 9V
Average current consumption: 0.3 to 1.3 mA (depending on operation mode)
Protection: IP20

IE-COM-D4141-09 3
Material of the case: ABS Forsan
Operation conditions:
Ambient temperature range: instrument: -10 to +60 °C, probe: -30 to +80°C
Ambient rel. humidity range: instrument: 5 to 95 %RH, non condensing, probe: 0 to 100%
Outer characteristics in accordance with EN 33-2000-3: normal environment with
characteristics AD1, AE1, AF1, AG1, AH1, AK1, AL1, AN1, AP1, AQ1, AR1, AS1, BA1,
BE1
Not allowed manipulation: it is not allowed to touch sensors under the cover to avoid
sensors damaging or to effect calibration.
The sensors (under the cover) should not be exposed to direct contact with water or other
liquids.
Storing conditions: temperature -10 to +60 °C relative humidity 5 to 95 %RH
Dimensions without probe: 141 x 71 x 27 mm
Weight including battery: approximately 200 g
Battery life depends on selected display refresh mode (see below). In FAST mode display is
refreshed in shortest possible interval with highest current consumption. In dynamic mode display
is refreshed in interval up to 5 s in case measured values remain stable. Refresh interval is shortened
to approximately 0.7 s only if measured values change. Current consumption in this mode in usual
operation is lower, battery life is up to 4 times longer. The FAST mode is recommend to use only in
cases, when slower display response is not acceptable.
Battery voltage drop below 7 V is indicated with blinking of "BAT" in default display mode
(displaying of actual values) and FAST mode is automatically cancelled to save the battery. At the
same time audio indication of alarms is automatically switched OFF.
If instrument is supplied from external ac/dc adapter, internal 9V battery should be replaced with
rechargeable NiCd accumulator. In usual operation from adapter accumulator is charged only with
small current. If accumulator is totally discharged, its full charging in instrument takes
approximately 100 hours. Instrument with accumulator is not recommended for permanent
operation without ac/dc adapter plugged. Accumulator works only as a standby source in case of
power mains failure.
Switching ON and OFF the instrument

4 IE-COM-D4141-09
Connect the external probe to the connector before switching ON the
instrument. Switch ON the instrument by pressing ON/OFF key. After
switching ON the instrument all symbols on the LCD are displayed. If the
ON/OFF key is being held pressed, all LCD symbols are displayed till the
key is released.
In usual operation instrument then starts the measurement mode and actual
measured values are displayed. If instrument is ON, do not disconnect or
connect the probe to prevent storing of incorrect value to minimum and
maximum memory.
If a voltage drop below critical limit occurred since last adjustment of
internal real time clock or battery disconnection longer than approximately
1 min, new instrument initialization is performed (reading INIT could be
displayed up to 12 s !).
After initialization is finished instrument starts to display internal time to
warn to incorrect internal time. New time adjustment is performed by the
computer. Press MENU key to switch instrument to actual measurement
mode.
It is possible to switch OFF the instrument anytime, all instrument setting is kept saved. If
automatic logging was active before instrument is switched OFF, logging will continue with the
same parameters after new switching ON.
Displaying of actual measured values
In this mode is instrument anytime after switching ON. It is possible to enter this mode from
other modes by pressing or by repeating pressing of MENU key. If the probe is not connected
properly or temperature is out of the range symbol “--“ is displayed.
Temperature is displayed on the upper LCD line (°C EXT) and relative
humidity in %RH is displayed on the lower LCD line. Those readings are
from the external probe.
Press DISPLAY key to display other readings - ambient temperature from
internal sensor (°C INT) and the dew-point temperature calculated from
external probe values (symbol DP).
Press DISPLAY key to display other readings - atmospheric pressure in
hPA on upper line and dew point temperature on lower line (°C DP).
Press DISPLAY key to display pressure tendency (hPa), which is the
difference between actual pressure and pressure before 3 hours. If the
instrument is not switched ON at least 3 hours, pressure tendency is not
available yet and symbol „--„ is display.

IE-COM-D4141-09 5
Function HOLD (storing of actual measured values) and minimum a maximum memory
Press HOLD key in the default mode (displaying of actual measured value) to store actual
measured values to internal memory (indicated by short beep). Anytime it is possible to display
stored values from MENU (see below). Each pressing of the HOLD key in the default mode causes
values stored in HOLD memory are replaced with actual ones.
Switched ON instrument permanently updates minimum and maximum
memory of each measured values. Press MIN key (resp. MAX key) in the
default mode to display minimum (resp. maximum) reading. These
minimum and maximum readings are indicated by MIN (MAX) symbols on
the LCD. Press DISPLAY key to display minimum (resp. maximum) value
of other values. Pressing MIN (MAX) or MENU key again to return to
default mode. Minimum and maximum memory is cleared from menu after
confirmation selection CLR (see below). Values in HOLD, MIN and MAX
memories remain stored even after instrument is switched OFF.
Functions and settings available from menu
Press MENU key to enter mode of viewing menu items one by one. Press arrow keys up and
down to list all menu items. Press MENU key again to return to default mode (displaying of actual
measured values).
First selection is START/STOP logging. There are two possibilities: if
logging is stopped, logging start is offered (symbol Strt = START), if
logging is in progress already, logging stop is offered (reading STOP). Press
ENTER key to confirm one of two possibilities. Notice: if battery voltage is
low it is not possible to start logging.
This item indicates if audio signaling of alarm indication is switched on
(On) or switched off (OFF). Press ENTER key to change actual setting.
Notice: if the battery voltage is low, audio indication is out of operation to
reduce current consumption independently on this selection.
Clearing of minimum and maximum memory of temperature. Memory is
cleared after pressing ENTER key. Clearing is confirmed by reading YES
on the LCD lower display.
Press ENTER key to display value stored in the HOLD memory. Press
MENU key to leave this mode.

6 IE-COM-D4141-09
Time (hour and minute) in internal real time clock of the instrument. Press
ENTER key to check date (day, month, year).
Battery voltage of partially loaded battery is displayed. This value illustrates
battery condition.
Display refresh mode is indicated. In the FAST mode refreshment is fastest
with regular interval approximately 0.7 s. In the dynamic refresh mode
(DYN.) each 15 s refresh interval of display is doubled to maximum 5 s if
measured values are stable. If measured values change, refresh interval
decreases to approximately 0.7 s. This dynamic mode prolongs battery life
significantly. Select the desired mode by ENTER key. Notice: if battery
voltage is low, the FAST mode is out of operation to reduce current
consumption independently on this selection.
Each pressing of ENTER key causes displaying information on storing
interval, actual logging mode (MODE), memory occupation (MEMO - in
automatic logging mode memory occupation in % is displayed, in manual
mode the number of logged values is displayed) and service information on
software version (upper LCD line) together with instrument configuration
on the LCD lower line.
Correction of pressure reading to the sea level
Select menu item Calc P (calculation of pressure) and press the ENTER key.
The actual pressure reading is displayed (LCD upper line, hPa) and the
value of correction constant is displayed on the LCD lower line (DIF). Use
the arrow keys to set the appropriate value of adding constant. At the same
time the actual pressure reading on the upper line is updated. Actual
pressure value above sea level is possible to gain e.g. from nearest weather
station. The new value of adding constant will be stored after pressing
ENTER. This is indicated by YES on LCD the lower line. It is possible to
leave this mode without any change by pressing the MENU key. The above
procedure is necessary to repeat anytime after instrument is moved to the
location of different altitude above sea level. If you set the adding constant
(DIF) to zero, instrument will display the absolute pressure again.
Alarm indication and setting
It is possible to set lower and upper limit for each measured quantity.
Breaking of the limit is indicated by blinking of the appropriate value on the
display. If a new alarm was indicated (i.e. it was not active in the previous
measurement), display starts to display the value out of limits. If at least one

IE-COM-D4141-09 7
alarm is active, audio indication can be activated, if menu AUDI "On" is
selected (see setting described above). Alarm activation of each value can
be disabled by setting lower alarm limit of the desired value up to its
maximum. This is indicated by OFF reading at the position of numeric
value. Value of upper limit of the same alarm is indifferent.
To set alarms press MENU key, select ALAR from menu items and confirm
by pressing ENTER key.
The following description of alarm setting is a model for alarm setting of all
measured quantities.
Reading ALLO indicates adjusted lower alarm limit (here air temperature).
Set the desired value by means of the arrow keys. Press and hold the arrow
key UP to make value increase fast. Press and hold the arrow key DOWN to
make value decrease fast. Release the arrow key and press ENTER to
confirm new limit.
Reading ALHI indicates adjusted upper limit of the same alarm (here air
temperature). Set the desired value in the same way as in above lower limit.
If needed it is possible to get back to lower limit setting of the same alarm
by pressing DISPLAY key. Press ENTER key to confirm new upper limit.
Then you are offered to set alarm of other input value (here relative
humidity). The procedure is the same as the above temperature limit setting.
Alarm activation of each value can be disabled by setting lower alarm limit
of the desired value up to its maximum. This is indicated by OFF reading at
the position of numeric value. Value of upper limit of the same alarm is
indifferent.
Above procedure is followed by alarm setting of all other quantities with the
exception of pressure tendency, which is not alarmed.
It is possible to leave the alarm setting mode by pressing MENU key. New
adjusted limits up to pressing MENU key are stored in memory.
Logging mode
In logging mode measured values are stored to instrument internal memory. Memory is non
volatile, i.e. if battery voltage drops or battery is disconnected, stored data will not be lost.
Instrument can work in two logging modes:
automatic mode - actual measured values are stored to memory in adjustable time interval. Storing
of the first value is synchronized with internal real time clock , so the memorizing is performed
at sharp multiples of minutes, hours and days. E.g. if the storing interval is 15 minutes value will
be stored each 15., 30., 45. a 00. minute and zero second in each hour, if interval is 3 hours
values will be stored in 00:00:00, 03:00:00, 06:00:00. etc. It means after starting automatic
logging instrument waits for the nearest integer multiple of time and then stores the first value.
manual mode - actual measured value is stored after pressing ENTER key (instrument should be in
default mode of displaying actual measured values). Each storing is indicated by a short beep

8 IE-COM-D4141-09
and position of the record in memory is displayed (reading MEMO with the record serial
number). In this mode it is possible to store max. 1000 records.
It is possible to select one of the three logging modes (MODE):
-automatic non cyclic logging, when logging stops after fulfilling the memory. Reading MEMO
FULL is displayed (reading disappears after pressing any key)
-automatic cyclic logging, when after fulfilling the memory oldest data are replaced with new
values
-manual logging, designed for manual storing of measured values by ENTER key
Logging modes are not mutually compatible and after changing logging mode memory of the
instrument should be cleared.
Logging in progress will always be interrupted by instrument connection to the computer or
switching OFF the instrument by ON/OFF key. Logging in progress will be automatically stopped
if battery voltage below 7 V is indicated by the instrument. This prevents to log values effected by
low battery voltage.
Logging start from instrument keyboard
Start logging by confirmation of the menu item START (symbol Strt) by
pressing ENTER key.
Notice: it is not possible to start logging if low battery voltage is indicated.
Instrument offers the possibility to clearing the instrument memory (ERAS).
If you do not wish to clear the memory before new logging start, confirm
selection ERAS NO by pressing the ENTER key
Now instrument displays adjusted storing interval. To modify the storing
interval use arrow keys. Interval 10 sec is displayed as MIN 0:10, minute
intervals are displayed as MIN x:00 and hour intervals as x:Ho. Confirm
adjusted interval by pressing ENTER key. Notice: if manual logging mode
is selected, this menu item is skipped (there is no interval adjustment).
Logging is started and instrument returns to default mode displaying actual
values. Logging in progress is in default mode indicated with reading LOG
on display. If memory is filled over 90 %, symbol LOG is blinking (only in
automatic logging).
Logging start with clearing memory and logging mode selection
It is possible to clear memory only if logging is stopped. Modification of logging mode is
enabled only after memory is cleared. Clearing of memory is optional part of the above procedure
logging START. Notice: there is no way to "undelete" the cleared data!
If it is needed to clear the memory before logging start, first it is necessary
to select menu item logging START (symbol Strt) and confirm by pressing
ENTER key. Notice: if low battery voltage is indicated, it is not enabled to
start logging nor to clear the memory
Now instrument offers possibility to clear the memory (ERAS). Default is

IE-COM-D4141-09 9
selection ERAS NO. If clearing of the memory is needed use the arrow keys
to select ERAS YES and confirm by pressing ENTER key.
Clearing procedure is in progress. It takes approximately 16 s and the
duration is indicated by displaying of already cleared memory in % (digits 0
to 100).
After memory is cleared it is possible to select one of three logging modes
(MODE) by using arrow keys (non cyclic logging is indicated by symbol
n.CYC, cyclic logging by symbol CYCL and manual logging by symbol
Man). If manual logging mode is selected, after confirmation by ENTER
key logging is active. In automatic logging mode (i.e. cyclic or non cyclic)
last applied storing interval is offered to confirm or modify. See above
procedure.
Logging stop from keyboard
Logging stop is enabled by confirmation of menu item STOP by pressing
ENTER key.
Reading STOP NO is displayed to enable logging stop. Use arrow keys to
modify to STOP YES. Logging is still in progress.
Logging is stopped after confirmation selection STOP YES by pressing
ENTER key. Instrument then returns to default mode of displaying actual
values.
Connection of the instrument to a personal computer
For connection of the instrument to a personal computer is designed the communication cable
(included in delivery). Connect the cable with 9pin connector Canon to computer serial port
(COM1 or COM2). Connect cable other side (JACK 3.5mm) to the instrument. During time
instrument is connected to the computer, there is a reading PC on the LCD display and
measurement nor logging is interrupted. Keyboard of the instrument is disabled with the exception
of ON/OFF key.
Result of reading data to the computer is a protocol of measurement including instrument serial
number and a table of logged values with date and time.
After disconnection of the communication cable reading PC disappears and instrument is in the
default mode as after switching on by the ON/OFF key. If logging was in progress before
connection to the computer, logging will continue with the same parameters after disconnection of
the cable (if new setting of logging was not performed during communication - including logging
stop).
The most power from the battery is needed during communication with the PC - let the

10 IE-COM-D4141-09
communication cable plugged-in the instrument only for time necessary for data transfer! If
you let instrument plugged-in the computer longer than approximately 5 minutes with no
communication, instrument will be automatically switched OFF.
Battery replacement
If battery voltage is low logging mode of the instrument is disabled. Low battery voltage is
indicated on the display with blinking reading "BAT". It is necessary to replace it with new one as
soon as possible. Battery is located under small cover on the instrument lower side. It is absolutely
necessary to replace battery with instrument switched OFF, otherwise internal time and setting of
d.REF. and AUDI (from menu selections) and data in memory HOLD, MIN and MAX will be lost.
For the same reason do not disconnect the battery for longer than 1 minute even if instrument is
switched OFF. If it happens (or if battery is totally discharged), it is necessary to set time from the
computer and in appropriate menu selection set again LCD refreshment mode (d.REF.), alarm audio
indication (AUDI) and clear the minimum and maximum memory (CLR).
The PC software for the instrument
Installation of Windows software
Software is designed for computer with operational system Windows, version Windows 95 or
newer. There is approximately 5MB free space required on the hard disk. A free serial port is
required for communication with the instrument.
Start the installation from Control Panel of Windows (Start - Settings - Control panel) -
Add/Remove programs, button Install. You will be suggested by Windows to insert first installation
diskette. After pressing Next button installation of software for COMMETER will start.
After installer start a window with License agreement occurs. Read the license agreement
carefully please. All text is possible to view by cursor movement keys up and down or PageUp,
PageDown keys. If you agree with the License agreement, press the Yes button. If you do not agree,
press No and installation will be cancelled.
In next step of installation it is possible to select path, where software will be install to. If you
agree with offered path, press Next, otherwise change the path by button Browse.
In the next window you can select folder in Start menu where software will be enable to run
from. If you agree with offered folder, press Next.
Now installation itself of software to hard disk will be carried out. Wait please for installer to
perform all necessary steps. During installation you will be invited to insert second diskette.
In the end message of successful installation will occur. Quit the Installer by button Cancel.
Now it is possible to run software from menu Start - Program Files - Comet - CommDx.
Software for Windows - instruction manual
Communication with the instrument. First connect communication cable to computer serial port.
Plug second connector of communication cable to the instrument and start data transfer to the
computer (selection from menu File - from datalogger, key F4 or button on the toolbar). Software
will find connection with the instrument and transfer data. Computer offers name of the file with
measurement to be created on disk. Implicitly files are stored to files with extension *.msb. File
name is generated from year, month and day of file creation and serial number in the day of
creation. It is possible to use different file name and path. After name confirmation it is possible to

IE-COM-D4141-09 11
insert short description with e.g. conditions or location of measurement which is stored together
with measured values.
Viewing of measured values
Measured values can be read directly from datalogger or from formerly created file on disk
(command File - From disk, key F3). After reading a table of measured values occurs. Detailed
information on file is available from context menu - menu occurs after clicking right mouse button
on the position of table. In context menu function export to text or database format is available.
Graph with measured curves it is possible to by command Display - Graph or button on the
toolbar. Graph mode enables to modify setting of axis, color of curves etc. Detailed information on
software functions is available in help.
Buttons on tool bars (left to right)
From disk opens file with measured values stored on disk
From instrument reads measured values from instrument and store them to disk
Table displays table of measured values
Graph displays time graph of measured values
Print prints table or graph to printer
Exit quits work with software
Information reads and display actual information on instrument setting
Display of file segments if file contains more segments of measured values, it is possible to
view those segments one by one
Previous segment moves to previous segment
Next segment moves next segment
Curve setting color modification, vertical axis assignment
Graph setting setting of graph parameters - axis, description etc.
Graph-channels one by oneenables/disables displaying of curves one by one
(i.e. there is only one curve on the screen at a time)

12 IE-COM-D4141-09
Graph-channel name in the Graph-channels one by one mode displays actual
channel name
Graph-next channel in the Graph-channels one by one mode switches to next
channel
Graph-previous channel in the Graph-channels one by one mode switches to previous
channel
Commeter instruments passed the following electromagnetic compatibility (EMC) tests:
radiation: EN 55022 class B
immunity: EN 61000-4-2 (levels 4/8 kV, class A)
EN 61000-4-3 (intensity of electromagnetic field 3 V/m, class B)
EN 61000-4-4 (levels 1/0,5 kV, class A)
EN 61000-4-6 (intensity of electromagnetic field 3 V/m, class B)
EN 61000-4-11 (class A)
EN 61000-4-5 (class A)
Other Wuntronic Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Niigata seiki
Niigata seiki GVC Series instruction manual

Norsonic
Norsonic nor118 instruction manual

Velleman
Velleman DVM856 user manual

Tektronix
Tektronix SDA601 Service manual

Sunrise Manufacturing International
Sunrise Manufacturing International Wings Series owner's manual

TESTO
TESTO 511 instruction manual