5
DPI-500 Technical Manual
3)Thedisplay will show"A1"toindicate that theunitisinUser Keyword SetupMode. Thefront
panel keys work exactly as in the System Keyword SetupMode. See Figure 4 for a mapof this
setupmenu.
3)Thedisplaywillmomentarilyshow"CAL0"followed byavalue. This valueisthezerocalibration
value andshouldbe recorded. Press anykeytocontinue.
4) Thedisplay willmomentarily show"CAL 1"followed bya value. This valueis the spancalibra-
tion value. Press any key to returnto keyword F18.
To Key-in the Zero Calibration Value
Pleasenotethatthisprocedureisintendedforemergencyuse,onlyinthecaseofnonvolatile
memoryloss. Aproperzerovalue,obtainedfromasuccessfulzerocalibrationprocedure,mustbe
used.
1) While in the System Keyword Setup Mode scroll to keyword F19.
2)AtkeywordF19, scrolldownonceusingtheZEROkey.
3)Thedisplaywillmomentarilyshow"CAL0",followedbyaflashingzero. Usethefrontkeypad
directional keys,as shown in Figure 2 onpage4,tochange the displayed value to the proper
zerocalibrationvalue.
4)Aftersettingthevalue,presstheNET/GROSSkeytosavethevalue.
5)Thedisplaywillmomentarilyshow"ECAL0"andthenreturntokeywordF19.
To Key-in the Span Calibration Value
Pleasenotethatthisprocedureisintendedforemergencyuse,onlyinthecaseofnonvolatile
memoryloss. Aproperspanvalue,obtainedfromasuccessfulspancalibrationprocedure,must
beused.
1) While in the System Keyword Setup Mode, scroll to keyword F20.
2)AtkeywordF19, scrolldownonceusingtheZEROkey.
3)Thedisplaywillmomentarilyshow"CAL1",followedbyaflashingzero. Usethefrontkeypad
directional keys,as shown in Figure 2 onpage4,tochange the displayed value to the proper
spancalibrationvalue.
4)Aftersettingthevalue,presstheNET/GROSSkeytosavethevalue.
5)Iftheenteredvalueisgreaterthanzero,thedisplaywillmomentarilyshow"ECAL1"andthen
returntokeywordF19. Ifazerovalueisentered,thenthedisplaywillmomentarilyshow"Err5"
andthenreverttothedisplayinstep3ofthis procedure.