Dual Range
Precision Sound Meter
©York Survey Supply Centre 2018Ref:.. \operat98\instructions 18\343050.qxp 05-01-18
I. Safety Information
Read the following safety information carefully before attempting to operate or service the meter. Use this meter only as specified in this manual; otherwise, the pro-
tection provided by the meter may be impaired.
• Environmental Conditions
1 - Altitude up to 2000 meters
2 - Relatively humidity 90% max
3 - Operation Ambient 0-40ºC
• Maintenance & Cleaning
1 - Repairs or servicing not covered in this manual should only be performed by qualified personnel
2 - Periodically wipe the case with a dry cloth. Do not use abrasives or solvents on this instruments.
• Safety Symbols
Meter is protected throughout by double insulation or reinforced insulation
When servicing, use only specified replacements parts.
Comply with EMC
II. General Description
Thank you for selecting our Sound Level. To ensure that you can get the most from it, we recommend that you read and follow the manual carefully before use.
The unit conforms to the IEC651 type 2, ANSI S1.4 type 2 for sound level meters.
The Sound Level Meter has been designed to meet the measurement requirements of Safety Engineers, Health, Industrial safety offices and sound quality control in
various environments.
• Ranges from 30dB to 130dB at frequencies between 31.5Hz and 8 Hz.
• Display with 0.1dB steps on a 4-digit LCD.
• With two equivalent weighted sound pressure levels. A and C.
III. Specifications
Standard Applied According to IEC651 type 2, ANSI S1.4 type 2
Frequency Range 31.5Hz - 8 Hz
Measuring Level Range 30 - 130 dB
Frequency Weighting A/C
Microphone 1/2 inch Electret condenser microphone
Calibration Electrical calibration with the internal oscillator (1 Hz sine wave)
Display LCD
Digital Display 4 Digits
Resolution: 0.1dB
Display Period: 0.5sec
Time Weighting FAST (125mS), SLOW (1 sec).
Level Ranges Lo: 30 - 100dB and Hi: 60 - 130dB
Accuracy ±1.5dB (under reference conditions)
Dynamic Range 65dB
Alarm Function “OVER” is shown when input is out of range.
Maximum Hold Hold readings, with decay <1dB/3 minutes.
Auto Power Off Meter automatically shuts down after approx. 15 minutes of inactivity.
Power Supply One 9V battery 006P or IEC 6F22 or NEDA 1604
Power Life About 50 hrs (alkaline battery)
Operating Temperature 0 - 40OC (32 - 104OF)
Operating Humidity 10 - 90% RH
Storage Temperature -10 - 60OC (14 - 140OF)
Storage Humidity 10 - 75% RH
Dimensions 210(L) x 55(W) x 32(H)mm
Weight 230g (including battery)
Accessories 9V battery, carrying case, instruction manual.
4. Name and Functions
3. Power ON/OFF button
Turn the meter power ON/OFF
4. Backlight button: -
Turn the meter backlight ON/OFF
5. A-Weighting / C-Weighting select button A/C
A: A - Weighting. For general sound level measurements.
C: C - Weighting. For checking the low-frequency content of noise.
(If the C-Weighted level is much higher than the A-Weighted level, then there is a large amount of low-frequency noise)
6. Time weighting select button F/S
F(fast response): For normal measurements (fast varying noise)
S(slow response): For checking average level of fluctuating noise
7. Level range select button Lo/Hi
Lo: 30~100dB; Hi: 60~130dB
When “OVER” is indicated, the ranges switch to another range for measurement.
8. MAX/Hold button MAX/HOLD
The max. Hold position is used to measure the maximum levels of sound. The maximum measured level is updated continuously. Press once again the button, will
release the hold and allow a further measurement.
Data Hold button: Press and hold the Button for over 2 seconds to turn on or off data hold function.
The hold function freezes the reading of the display.
9. Microphone
1/2 inch Electret Condensor microphone
10. Battery Cover
Measurement Preparation
(1) Battery Loading
Remove the battery cover on the back and put in one 9V battery.
(2) Battery Replacement
When the battery voltage drops below the operating voltage, mark “BAT” appears. If it appears, battery should be replaced with new one.
Operating Precautions
(1) Wind blowing across the microphone would bring additional extraneous noise.
Once using the instrument in the presence of wind, it is a must to mount the windscreen to not pick up undesirable signals.
(2) To achieve more accurate measurement, use an extension cable to separate the Microphone from the main body so that the effect of unexpected sound
reflection can be eliminated.
(3) Calibrate the instrument before operation if the instrument was not in use for a long time or operated at bad environment.
(4) Do not store or operate the instrument at high temperature and high humidity environment
(5) eep microphone dry and avoid severe vibration.
(6) Please take the battery and keep the instrument in low humidity environment. When not in use.
Measurement
(1) Open battery cover and install a 9-volt battery in the battery compartment.
(2) Turn on power and select the desired response Time and Weighting. If the sound source consists of shorts bursts or only catching sound peak, set response to
FAST. To measure average sound, use the SLOW setting.
(3) Select desired level
(4) Hold the instrument comfortably in hand or fix on tripod and point the microphone at the suspected noise source, the sound pressure level will be displayed.
(5) When MAX (maximum hold) mode is chosen, the instrument captures and holds the maximum noise level for a long period using any of the time weightings and
ranges.
(6) When HOLD (data hold) mode is chosen, the hold function freezes the reading in the display. Press the HOLD button momentarily to activate or to exit the HOLD
function,
(7) Turn OFF the instrument and remove the battery when not in use.
Calibration Procedures
1. Windscreen
If you operate and wind speeds over 10m/sec, please put protective
accessories in front of the microphone.
2. Display
SYMBOL F NCTION
LCD 4 digits
MAX Maximum Value hold
OVER Over range
F Fast response
S Slow response
A A-Weighting
C C-Weighting
Lo Low Range (30-130dB)
Hi High Range (60-130dB)
BAT Low Battery
(1) Make the following buttons and function switch settings.
Display: dB, A, Hi or Lo, F
Function: A-Weighting
Response Time: FAST
Level Range: 30 to 100dB(Lo) or 60 to 130dB(Hi)
Measurement Mode: MAX Hold and Data Hold Mode function disable
(2) Insert the microphone housing into the insertion hole of the calibrator.
(3) Open battery cover and remove the battery to adjust the CAL94dB potenitiometer of the unit. The
level display will indicate the desired level.
The products are all well calibrated before shipment

Other York Measuring Instrument manuals

York Elite 1600 ARC User manual

York

York Elite 1600 ARC User manual

York Multi-Digit Pro+ User manual

York

York Multi-Digit Pro+ User manual

York 354690 User manual

York

York 354690 User manual

Popular Measuring Instrument manuals by other brands

TURNER DESIGN AQUAFLUOR user manual

TURNER DESIGN

TURNER DESIGN AQUAFLUOR user manual

Electro-Sensors SG1000A manual

Electro-Sensors

Electro-Sensors SG1000A manual

Graco HG6000 Instructions and parts manual

Graco

Graco HG6000 Instructions and parts manual

Stabila Laser Distancer LD 320 operating instructions

Stabila

Stabila Laser Distancer LD 320 operating instructions

ATAGO PAL-55S instruction manual

ATAGO

ATAGO PAL-55S instruction manual

socomec COUNTIS E15 quick start

socomec

socomec COUNTIS E15 quick start

GREISINGER GCO 100 operating manual

GREISINGER

GREISINGER GCO 100 operating manual

ABB PFEA 111 user manual

ABB

ABB PFEA 111 user manual

XC-Tracer Mini II user manual

XC-Tracer

XC-Tracer Mini II user manual

Levenhuk Wezzer Air PRO DM50 user manual

Levenhuk

Levenhuk Wezzer Air PRO DM50 user manual

Hanna Instruments HI 701 quick start guide

Hanna Instruments

Hanna Instruments HI 701 quick start guide

Teledyne 3350 operating instructions

Teledyne

Teledyne 3350 operating instructions

Xylem WTW OxiTop-IDS Multi 3630 IDS operating manual

Xylem

Xylem WTW OxiTop-IDS Multi 3630 IDS operating manual

Kikusui TOS9301PD user manual

Kikusui

Kikusui TOS9301PD user manual

Analytik Jena SPEKOL 1500 manual

Analytik Jena

Analytik Jena SPEKOL 1500 manual

LumaSense IN 300/5 Operation manual

LumaSense

LumaSense IN 300/5 Operation manual

Emerson CSI 2140 user guide

Emerson

Emerson CSI 2140 user guide

Prosensor N1500LC manual

Prosensor

Prosensor N1500LC manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.