same.
2. Operating instructions
(1) Zeroing—Turn the measuring knob to “OFF”(关) first, adjust the mechanical zero setter(0) in the
instrument's center, to make the pointer locate on zero line(0).
Caution: --This process should be adjusted carefully before connecting power supply, so as to ensure the
accuracy rate.
(2) Connecting power supply.
a. Connecting power supply outlet and Hall Sensor with Tesla meter
b. Press down the power supply indicator “.” On the left side.
c. Connecting power supply (220V+/-10%, 50Hz)
(3) Calibration line adjust—When the measuring knob indicates “Calibration”(校准), the pointer should
locate near the “”calibration constant"(which could be found on the probe). Warm-up for 30 minutes, then
adjust the “calibration” hole(校准), make the pointer locate exactly on “calibration constant”.
Caution:
a. this equipment adopts constant current circuit, and usually the current changes seldom. Adjust is only
needed when ambient temperature changes considerably. If the current fluctuates too greatly, please check if
the constant current apparatus is out of order.
(4) Zeroing of amplifier—adjust the measuring knob to indicate “0”, adjust mechanical zero setter(0) in the
instrument’s center, and make the pointer to locate exactly on zero line(0).
(5) “Zeroing” knob(调零)
a. When use 50~2500mT measuring range, adjust the measuring knob to indicate “50”, then adjust the
“zeroing” knob(调零), to make the pointer to locate exactly on zero line (0), then adjust the
measuring knob to the range which will be measured, after repeating the zeroing process, measuring
work could start.
b. When use 10mT, adjust the measuring knob to indicate “10”, then adjust the “zeroing” knob(调零)
to make the pointer to locate exactly on zero line (0).