Zeiss S8 User manual

S8 Floor Stand
Instructions for use
G-30-1346-en
Issue 4.0
Printed on 05. 06. 2001


G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Contents
Functions at a glance 5
– S8 Floor Stand 6
– Key to symbols 8
Safety 9
– Directives and standards 10
– Notes on installation and use 11
– Safety devices 16
– Warning labels and notes 24
Description 27
S8 floor stand 29
– Intended use 29
– Description of the modules 30
– Design 30
– Stand base with stand column 34
– Connection panel 36
– Suspension arm 38
– Display field with control keys 40
– Halogen illumination system 42
– Xenon illumination system 46
Preparations for use 51
Attaching the equipment 52
Connections 56
– Connecting the surgical microscope 56
– Connecting the S light guide 56
– Mounting the strain relief device 58
– Connecting the stand 60
– Positioning the stand 61

G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Operation 63
Preparations for use 64
– Adjusting the balance setting of the suspension arm 64
– Adjusting the limit of downward movement 66
– Relocating the stand 68
– Checklist 70
Using the display and key field 72
– General functions 72
– Operating the OPMI® Vario 76
– Operating VISU 150 and VISU 200 96
– Operating the OPMI® PRO magis 108
What to do in an emergency 114
– Failure of a halogen lamp 114
– Failure of the xenon lamp 116
– Failure of lamp control 118
– Failure of focusing system 118
– Failure of magnetic clutches 118
Maintenance / Further information 121
– Trouble-shooting table 122
– Changing the halogen lamp 126
– Changing the xenon lamp module 130
– Care of the unit 132
– Sterilization 132
– Disinfecting the control keys 133
– Ordering data 134
– Accessories 134
– Spare parts 135
– Technical data 136
– Ambient requirements 139
Index 141

6Functions at a glance
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
S8 Floor Stand
Halogen illumination system
1Closed flap: main lamp is on - Open flap: backup lamp is on
2Filter knob
3Opening the lamp module
4Manual selection of backup lamp
Xenon illumination system
5Filter knob
6Manual selection of backup lamp
7Opening the lamp module
8Switching on the stand
9Connection panel
10 Rated voltage display
11 Connector for foot control panel or hand control panel
12 Locking the suspension arm in its horizontal position
13 Setting the limit of downward travel
14 Releasing the magnetic clutches of the stand
15 Balance setting
16 Lamp housing for halogen illumination (or xenon illumination)
17 Removing and mounting the coupling for the surgical microscope
18 Control panel
19 Brake for locking the stand in position
12
347
6
5

G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Functions at a glance 7
18
19
13
12
11
10
14 15 16 17
9
8

8Functions at a glance
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Key to symbols
Different symbols used in this user's manual draw your attention to safety
aspects and useful tips. The symbols are explained in the following.
Warning!
The warning triangle indicates potential sources of danger which may
constitute a risk of injury for the user or a health hazard.
Caution:
The square indicates situations which may lead to malfunction, defects,
collision or damage of the instrument.
Note:
The hand indicates hints on the use of the instrument or other tips for the
user.
OPMI®is a registered trademark of Carl Zeiss.
OPMI®

10 Safety
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
The instrument described in this manual has been developed and tested
in accordance with Carl Zeiss safety standards and with national and in-
ternational regulations. A high degree of instrument safety is thus en-
sured.
We would like to inform you on the safety aspects involved in operating
the instrument. This chapter contains a summary of the most important
precautions to be observed.
Further safety notes are also contained in other parts of this user's
manual; they are marked with a warning triangle containing an exclama-
tion mark as shown here. Please pay special attention to these safety
notes.
Safety is only ensured when this instrument is operated properly. Please
read through this manual carefully before turning the instrument on. Also
read through the user's manuals of the other equipment used with this in-
strument. You may obtain further information from our service organiza-
tion or authorized representatives.
Directives and standards
The instrument described in this manual has been designed in compli-
ance with the following standards:
–EN
–IEC
–UL
–CSA
In accordance with Directive 93/42/EEC, Annex II, Article 3, the quality-
management system of Carl Zeiss has been approved by TÜV Rheinland,
which is a notified body. The registration number is: 95 102 7601.
• The instrument must be connected to a special emergency backup
line supply in accordance with the regulations or directives which ap-
ply in your country.
• This is a class I instrument as defined by Directive 93/42 /EEC.
• Please observe all applicable accident prevention regulations.

G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Safety 11
Notes on installation and use
Safe working order
• Do not operate the equipment contained in the delivery package in
– explosion-risk areas,
– the presence of inflammable anesthetics or volatile solvents such
as alcohol, benzine or similar chemicals.
• Do not station or use the instrument in damp rooms. Do not expose
the instrument to water splashes, dripping water or sprayed water.
• Immediately unplug any equipment that gives off smoke, sparks or
strange noises. Do not use the instrument until our service represent-
ative has repaired it.
• Do not place any fluid-filled containers on top of the instrument. Make
sure that no fluids can seep into the instrument.
• Do not force cable connections. If the male and female parts do not
readily connect, make sure that they are appropriate for one another.
If any of the connectors are damaged, have our service representative
repair them.
• Do not use a mobile phone in the vicinity of the equipment because
the radio interference can cause the equipment to malfunction. The ef-
fects of radio interference on medical equipment depend on a number
of various factors and are therefore entirely unforeseeable.
• Potential equalization: If desired, the unit can be incorporated in po-
tential equalization measures. The potential equalization bolt is re-
quired for this purpose, see chapter "Description".
• Modifications and repairs on these instruments or instruments used
with them may only be performed by our service representative or by
other authorized persons.
• The manufacturer will not accept any liability for damage caused by
unauthorized persons tampering with the instrument; this will also for-
feit any rights to claim under warranty.
• Use this instrument only for the applications described.
• Only use the instrument with the accessories supplied. Should you
wish to use other accessory equipment, make sure that Carl Zeiss or
the equipment manufacturer has certified that its use will not impair
the safety of instrument.

12 Safety
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
• Only personnel who have undergone training and instruction are al-
lowed to use this instrument. It is the responsibility of the customer or
institution operating the equipment to train and instruct all staff using
the equipment.
• Keep the user's manuals where they are easily accessible at all times
for the persons operating the instrument.
• Never look at the sun through the binocular tube, the objective lens or
an eyepiece.
• Do not pull at the light guide cable, at the power cord or at other cable
connections.
•This instrument is a high-grade technological product. To ensure opti-
mum performance and safe working order of the instrument, its safety
must be checked once every 12 months. We recommend having this
check performed by our service representative as part of regular main-
tenance work.
If a failure occurs which you cannot correct using the trouble-shooting
table, attach a sign to the instrument stating it is out of order and con-
tact our service representative.
Requirements for operation
Our service representative or a specialist authorized by us will install the
instrument. Please make sure that the following requirements for opera-
tion remain fulfilled in the future:
– All mechanical connections (details in the user's manual) which are
relevant to safety are properly connected and screw connections tight-
ened.
– All cables and plugs are in good working condition.
– The voltage setting on the instrument conforms to the rated voltage of
the line supply on site.
– The instrument is plugged into a power outlet which has a properly
connected protective earth contact.
– The power cord being used is the one designed for use with this in-
strument.
Before every use and after re-equipping the instrument
• Make sure that all ”Requirements for operation” are fulfilled.
• Go through the checklist.
• Re-attach or close any covers, panels or caps which have been re-
moved or opened.

G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Safety 13
• Pay special attention to warning symbols on the instrument (triangular
warning signs with exclamation marks), labels and any parts such as
screws or surfaces painted red.
• Do not cover any ventilation openings.
For every use of the instrument
• Always position the stand in such a way that the long extension of the
stand base points in the direction of the surgical field (if the instrument
is equipped with a base of this type).
• The transport casters must be retracted to ensure that the unit is se-
curely resting on its base (with instruments with retractable transport
casters).
• Avoid looking directly into the light source, e.g. into the microscope ob-
jective lens or a light guide.
• When the illumination is on, the light guide must be connected at both
ends. Otherwise there is a risk of fire or burn injuries.
• Make sure that the instrument has been switched off before you
change the xenon lamp module. When switched on, the ignition sys-
tem generates high voltage.
• The xenon illumination system is a high-intensity light source which -
if used improperly - can cause thermal injury to skin or tissue. Keep
the exposed tissue moist and provide sufficient irrigation. Carefully
monitor the effects of the illumination on the tissue, in particular in the
following cases:
– during prolonged procedures on skin and tissue using objective
lenses with a short focal length (short working distance),
– during procedures on tissue with a low blood supply,
– with high brightness settings of the xenon lamp.
• Since the xenon lamp provides high light intensity and generates light
with a spectrum similar to daylight, it must not be used for ophthalmic
applications.
• Any kind of radiation has a detrimental effect on biological tissue.This
also applies to the light illuminating the surgical field. Please therefore
reduce the brightness and duration of illumination on the surgical field
to the absolute minimum required.
After every use of the instrument
• Always use the main power switch of the instrument to turn it off.

14 Safety
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
• The main power switch must always be turned off when the instrument
is not in use.

G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Safety 15

16 Safety
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Safety devices
1Release bar
Allows non-sterile persons to release the magnetic clutches of the
ceiling mount (floor stand).
2Adjustment screw for limiting the downward travel
Use this screw to set the minimum vertical distance (working distance)
from the surgical field. Check this setting before each surgical proce-
dure.
3Locking knob
for locking the suspension arm in its horizontal position.
Before removing or attaching a unit (microscope, tube, etc.), move the
suspension arm into a horizontal position. Pull out the locking knob
and turn it clockwise or counterclockwise through 180°, while slightly
moving the suspension arm up and down until the lock engages.
When locked, the suspension arm can no longer suddenly spring up-
ward when insufficient weight is attached. After attaching a unit, per-
form the balancing procedure.

G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Safety 17
1
1
2
3

18 Safety
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Halogen illumination system
1Activating the backup lamp
Each of the two lamp housings contains a backup lamp which will be
automatically activated if the first lamp fails. If this automatic function
fails, you can switch on the backup lamp by pressing this button.
2GG 475 retina protection filter
When operating on the eye, always use a GG 475 protection filter to
ensure that the patient's retina is not exposed to unnecessary (blue)
radiation (risk of retinal injury). The filter knobs have four positions:
3Yellow indicator lamp
– Lights when the main lamp has failed. In addition, the open flap on
the lamp module indicates that the main lamp has failed. The back-
up lamp is on.
– Blinks when the backup lamp has failed.
4Manual function
When the manual function has been activated, all electrical control
systems are inoperative. The lamp brightness is automatically ad-
justed to a fixed setting.
0 no filter
1 GG 475 filter: to protect the patient's eye during surgery against
unnecessary (blue) radiation (retinal injury).
2 KK 40 filter: to increase color temperature
3 no filter

G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Safety 19
3
2
1
4

20 Safety
G-30-1346-en S8 Floor Stand Issue 4.0
Printed on 05. 06. 2001
Xenon illumination system
1Activating the backup lamp
The lamp module contains two xenon lamps. The second lamp is used
as a backup lamp which has to be swung into the illumination beam
path when the first lamp fails.
If the xenon lamp fails, open the lamp module as follows:
Press button (4). The lamp module is slightly ejected. Pull out the lamp
module all the way. Turn knob (1) through 180° until it snaps in. This
moves the backup lamp into the illumination beam path. Push the
lamp module all the way back into the lamp housing.
2Yellow indicator lamp
Lights when the lamp has failed. After activation and ignition of the
backup lamp, the yellow indicator lamp goes out again.
3Manual function
When the manual function has been activated, all electrical control
systems are inoperative. The lamp brightness is automatically ad-
justed to a fixed setting.
Other manuals for S8
1
Table of contents
Other Zeiss Rack & Stand manuals
Popular Rack & Stand manuals by other brands

JARDINICO
JARDINICO JCP.4 Series Assembly and user's manual

Panasonic
Panasonic WV-Q169A operating instructions

Rivers Edge Treestands
Rivers Edge Treestands RELAX WIDE RE631 Operator's manual

Vicon
Vicon V2000D-PM quick guide

Konig & Meyer
Konig & Meyer 18807 manual

Classic Exhibits
Classic Exhibits visionary DESIGNS VK-2924 manual