GPS®III System
Processing Instructions
GPS®III System
Step 1: Blood Draw
Attach supplied 18 gauge needle to 60 ml syringe.
Withdraw 6 ml of ACD-A (citrate anticoagulant).
*Mini GPS® III System: Withdraw 3 ml of ACD-A.
54 ml of blood is drawn from the patient according
to facility protocol.
*Mini GPS® III System: 27 ml of blood is drawn from
the patient according to facility protocol.
GPS®III System
Step 2: Load Blood
Unscrew cap on center port No.1 and discard green
packaging post.
Slowly load blood into center port.
GPS®III System
Step 3: Balance
Push Open/Stop button on control panel. The
“unlocked” indicator will illuminate. Turn latch
counterclockwise to open lid. Place tube into
centrifuge.
*Mini GPS® III System: Purple buckets (7433) must
be inserted into the centrifuge.
Fill GPS® III counterbalance (800-0508) with 60
ml of sterile saline and place into opposite side of
centrifuge.
*Mini GPS® III System: Fill GPS® III mini
counterbalance (800-0505) with 30 ml of sterile
saline and place into opposite side of centrifuge.
GPS®III System
Step 4: Spin
Close the lid by rotating the lid latch clockwise.
“Latched” indicator will illuminate. Set speed to
3200 RPM and time to 15 minutes. Press green
button to start spin. Once spin is complete, press
red button to illuminate the “Unlocked” indicator.
Twist latch counterclockwise to open lid.
GPS®III System
Step 5: PPP Extraction
To extract the platelet poor plasma (PPP), remove
yellow cap on side port No. 2 and connect the 30 ml
syringe. Slowly tilt the tube while withdrawing PPP.
Replace yellow cap.
GPS®III System
Step 6: PRP Suspension and Extraction
Remove red cap on side port No.
3 and connect 10 ml syringe.
Withdraw 2 ml of PRP.
*Mini GPS® III System: Withdraw
1 ml of PRP.
With 10 ml syringe attached,
suspend the platelets by shaking
the tube for 30 seconds.
If the pellet between the buoys is
not suspended completely, shake
tube vigorously.
Extract the remaining platelet
rich plasma (PRP) contents into
the attached 10 ml syringe.
Plasmax®Plasma
Concentration
System
Processing Instructions
Plasmax®Plasma Concentration System
Spin paddle by hand until plasma is fully mixed into
beads. Mixing is complete when the white beads are
completely saturated with plasma.
Step 1: Load Plasma
Plasmax® Plus: Unscrew yellow cap on port No.
1 of plasma concentrator and fill with 25 ml of
PPP. Replace cap. Remove and discard protective
cover on tethered white cap. Screw white cap onto
port No 1.
*Plasmax® Regular: Fill with 15 ml of PPP.
Note: Refer to GPS®III instructions for how to process PRP and PPP.
Plasmax®Plasma Concentration System
Step 2: Spin and Extract
Plasmax® Plus: Place concentrator into centrifuge
with the empty Plasmax® Plus counterbalance
(800–0512). They should be placed on opposite
sides in the centrifuge. Spin for 2 minutes at 2000
RPM.
*Plasmax® Regular: Use Plasmax® Counterbalance
(800-0510)
Plasmax® Plus: Unscrew red cap on port No. 2.
Connectsterile10mlsyringeonportNo.2andextract
approximately 10 ml of concentrated plasma.
*Plasmax® Regular: Extract approximately 6 ml of
concentrated plasma.
All trademarks herein are the property of Biomet, Inc. or its subsidiaries
unless otherwise indicated.
This material is intended for the sole use and benefit of the Biomet Biologics
sales force and health care providers. It is not to be redistributed, duplicated
or disclosed without the express written consent of Biomet.
For product information, including indications, contraindications, warnings,
precautions and potential adverse eects, see the Biomet Biologics website
at www.biometbiologics.com.
©2018 Zimmer Biomet
1792.1-US-en-REV0118

Popular Laboratory Equipment manuals by other brands

Waters Xevo TQ-S Operator's, overview and maintenance guide

Waters

Waters Xevo TQ-S Operator's, overview and maintenance guide

Milestone PRESTO PRO Operator's manual

Milestone

Milestone PRESTO PRO Operator's manual

NuAire Labgard NU-427-400 Operation & maintenance manual

NuAire

NuAire Labgard NU-427-400 Operation & maintenance manual

Cuda Fiberoptics XLS-300 Service manual

Cuda Fiberoptics

Cuda Fiberoptics XLS-300 Service manual

Foss Kjeltec 8100 quick guide

Foss

Foss Kjeltec 8100 quick guide

Quidel Solana user manual

Quidel

Quidel Solana user manual

Keysight N778 C Series user guide

Keysight

Keysight N778 C Series user guide

Agilent Technologies IDP3GBKIT Quick reference guide

Agilent Technologies

Agilent Technologies IDP3GBKIT Quick reference guide

ThermoFisher Scientific Microliter 48 x 2 sealed instruction manual

ThermoFisher Scientific

ThermoFisher Scientific Microliter 48 x 2 sealed instruction manual

OKW STAR-KNOBS Series Assembly instruction

OKW

OKW STAR-KNOBS Series Assembly instruction

Hanna Instruments HI 2314 instruction manual

Hanna Instruments

Hanna Instruments HI 2314 instruction manual

Neuation iTHERM D150-1 Product user manual

Neuation

Neuation iTHERM D150-1 Product user manual

Leica TCS SP8 user manual

Leica

Leica TCS SP8 user manual

Qiagen QIAcube Safety Instructions and Quick Start Guide

Qiagen

Qiagen QIAcube Safety Instructions and Quick Start Guide

Procomcure Phoenix-Pure96 Operation manual

Procomcure

Procomcure Phoenix-Pure96 Operation manual

IKA Plate RCT digital manual

IKA

IKA Plate RCT digital manual

Altronix NetWay1DWP installation guide

Altronix

Altronix NetWay1DWP installation guide

Optika Italy B-150D Series instruction manual

Optika Italy

Optika Italy B-150D Series instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.