3nh TS Series User manual

Spectrophotometer
OPERATION MANUAL
Please read this manual carefully before use the spectrophotometer.

Safety Instructions
Please read carefully and strictly abide by the following provisions to avoid accidental injury or harm. The
company shall not be liable for any loss arising from the use of the instrument in violation of this manual.
Battery
1. This instrument uses built-in battery which can not be replaced with other batteries, in case of
damage to the instrument or other malfunctions.
2. Do not dismantle, squeeze, strike and heat the battery, and do not put the battery in the fire or
high temperature environment, otherwise the battery will explode and cause fire.
3. When the instrument is fully charged and not in use, cut off the external power supply to prevent
electric shock and instrument damage.
4. If the instrument is not used for a long time, charge the instrument every two weeks, or the
internal battery may be damaged and the instrument cannot be used again.
5. For the first three times of charging, it is better to get fully charged and then use it out before
next charging.
External
power
1. When external power supply is required, please use the standard power adapter of the
instrument, or it may shorten the battery life or even cause electric shock to damage the
instrument or cause fire.
2. If the instrument is not used for a long time, the external power should be cut off to prevent
burning the instrument Grating Spectrophotometer and causing fire.
Instrument
1. Do not use the instrument in a place with flammable gas (gasoline, etc.), as this may cause a
fire.
2. Do not dismantle the instrument privately, or it will be damaged. And if dust and metal foreign
objects may enter the instrument, it may cause short circuit, generate electric shock, and damage
to the instrument, even cause fire.
3. If the instrument emits peculiar smell like burning when using it, it should be stopped
immediately and sent to the maintenance point for testing and maintenance.

Content
Instruduction................................................................................................................................................................................4
Cautions.......................................................................................................................................................................................4
1. Interface Description............................................................................................................................................................. 5
2. Operating Instruction.............................................................................................................................................................6
2.1 Power On& Off.............................................................................................................................................................6
2.2 Calibration.....................................................................................................................................................................6
2.3 Measurement Instruction........................................................................................................................................... 9
2.4 Measurement............................................................................................................................................................. 11
2.4.1 Standard Measurement................................................................................................................................ 11
2.3.3 Sample Measurement...................................................................................................................................14
2.3.4 Average Measurement..................................................................................................................................16
2.5 Connecting to PC...................................................................................................................................................... 16
2.5.1 USB Connection.............................................................................................................................................16
2.5.2 Bluetooth Connection....................................................................................................................................17
2.6 Print..............................................................................................................................................................................17
3. Main Menu............................................................................................................................................................................ 18
3.1 Data Management.....................................................................................................................................................18
3.1.1 Check Record.................................................................................................................................................19
3.1.2 Delete Record.................................................................................................................................................24
3.1.3 Search Record................................................................................................................................................26
3.1.4 Standard Input................................................................................................................................................28
3.2 Calibration...................................................................................................................................................................30
3.4 Illuminant.....................................................................................................................................................................30
3.4 Average....................................................................................................................................................................... 32
3.5 Color Space................................................................................................................................................................32
3.6 Set Color Index.......................................................................................................................................................... 33
3.6.1 Set Color Index...............................................................................................................................................33
3.6.2............................................................................................................................................................................36
3.7 Display Setting...........................................................................................................................................................40
3.8 System Setting...........................................................................................................................................................42
3.8.1 Auto Save........................................................................................................................................................ 44
3.8.2 Measurement Aperture................................................................................................................................. 44
3.8.3 Bluetooth......................................................................................................................................................... 45
3.8.4 Buzzer..............................................................................................................................................................45
3.8.5 Measurement Mode.......................................................................................................................................46
3.8.6 Calibration Validity......................................................................................................................................... 46
3.8.7 Control Mode.................................................................................................................................................. 47
3.8.8 Language Setting...........................................................................................................................................48
3.8.9 Time Setting.................................................................................................................................................... 48
3.8.10 Backlight Time..............................................................................................................................................49
3.8.11 Screen Brightness....................................................................................................................................... 49
3.8.12 Restore Factory Settings............................................................................................................................50
4. Daily Maintenance...............................................................................................................................................................51
5. Technical Parameters......................................................................................................................................................... 52
5.1 Technical Features (Advance).................................................................................................................................52
5.2 Technique Specification............................................................................................................................................52
Appendix....................................................................................................................................................................................54

Instruduction
This instrument is a high-precision spectrophotometer with independent intellectual property rights independently
developed by the company. It is a leader in the field of portable spectrophotometers. It is suitable for precise color
transfer and quality control of plastic electronics, chemical coatings, ink printing, textile printing and dyeing,
automotive electronics, medical analysis, cosmetics and food analysis, etc. It is also widely used in scientific
research institutions, laboratories and other fields.
Under the condition of D / 8 ° geometric optical illumination, the instrument has accurate and stable color
measurement, large storage capacity, equipped with USB and Bluetooth dual communication modes, and its PC
color management software has powerful extension functions, which can realized accurate measurement and
presentation for different color difference formula and color indexes in a variety of color spaces.
Cautions
● The spectrophotometer is a precise measuring instrument. Please avoid drastic changes of external
environment when measuring. These changes, including the flash of surrounding light and the rapid change of
temperature, will affect measurement accuracy.
● Keep the instrument stable; make sure the measuring aperture touch the surface of the test sample placidly,
and no shaking or shifting when measuring.
●The instrument is not waterproof. Do not use it in high humidity environment or in water.
● Keep the instrument clean. Avoid dust, powder or solid particle entering the measuring aperture and the
instrument. Please prevent the colorimeter from fierce collision or crash.
● White calibration cavity should be cleaned regularly with wiping cloth to ensure the surface clean, and kept it in
a dark cool dry place.
● Please take out the battery to prevent the instrument from damage if you don’t use it for a long time.
● Please keep the instrument in a cool dry place.
● Any unauthorized changes to the instrument are not permitted, or it will affect the measuring accuracy, even
cause irreversible damage

1. Interface Description
Figure 1 Interface Description
Port Description
1. Power Switch 1/0: Push the switch to “1” to turn on the instrument.
Push the switch to “0” to turn it off.
2. USB/RS-232/Charging port: common interface. The instrument automatically judges the connection. USB
interface is used to connect to PC software. RS-232 port is used to connect to the printer. It can charge the
instrument by connecting the USB interface to power charger or PC. The specification of the external power
charger is 5V == 2A.
3. Switch / Testing button: Long press for 3 seconds to turn ON/OFF power, short press for measurement.
4. Aperture switch: used for changing the measuring aperture, setting this switch to “MAV”, turn to Φ8mm
aperture, while setting this switch to “SAV”, turn to Φ4mm aperture.
5. Indicator light: LED has green and red colors, which are used to indicate the working status of the instrument.
* During turn-on charging, when the charging capacity is greater than 90%, the green light will be on, and the red
light will be on when the charging capacity is lower than 90%.
* During the normal start-up process, the instrument is in normal state (no power adapter is connected, black and
white calibration is in the validity period, the power is more than 10%, the instrument has no fault), the green
indicator light is on for 10 seconds, the green light flashes during the measurement process, and it goes out after
the measurement.

* When the instrument is in working mode and the power of the instrument is less than 10%, the red light will
flash.
6. M5 fixed thread hole: fixed instrument thread interface. Thread type is standard metric common coarse thread,
pitch 0.8mm, depth 5mm.
7. Wrist strap fixing post: used to fix the wrist strap. When the wrist strap is put on the wrist, it can prevent the
instrument from slipping accidentally.
8.Calibration Board: including white board and black cavity. White board is used for white calibration and black
cavity is used for black calibration. Please refer to the black and white correction section for details.
2. Operating Instruction
2.1 Power On& Off
As shown in Figure 1, press the power switch to "I", to start the instrument. Press the power switch to "O", the
instrument is turned off.
If the instrument is without any operation for a long time, it will automatically enter into standby mode. At this time,
press the Testing Keyor click the LCD screen to wake up the instrument.
2.2 Calibration
In the measurement interface, click to enter main menu, others please click or to enter main menu, as
shown in Figure 2.

Figure 2 Main menu
Select “Calibration” to enter white and black calibration interface as shown in Figure 3. It will show if the
calibration is valid or not, and the remaining time if it is valid.
Figure 3 White and Black Calibration
Click to continue and enter “Black Calibration” as shown in Figure 4. Fit the measuring port to the black cavity
according to the prompt, and then click the "Testing Key" for black calibration. You can click the back button
to cancel the calibration.

Figure 4 Black Calibration
After black calibration, it will prompt you to white calibration as shown in Figure 5. According to the prompt,
confirm the serial number of white calibration plate correspond to the instrument and correct aperture setting,
then put the measuring aperture on white board and press “Testing key” for white calibration, or click to cancel
and quit the calibration.

Figure 5 White Calibration
2.3 Measurement Instruction
As shown in Figures 6, 7, and 8, it is working condition area at the top of measurement interface, displaying the
status of measurement mode (SCI/SCE), Bluetooth and UV conditions. Under the measurement interface is a
shortcut display area. You can click the corresponding shortcut key to quickly switch the test data. The middle part
of the measurement interface is the data display area. The instrument displays the corresponding chromaticity
data according to the current user settings. The bottom of the measurement interface is the key operation area.
You can operate the current data by clicking the corresponding operation button. Figures 7 and 8 respectively
show the spectral area and the color index area and quickly switch by clicking the down button .

Figure 6 Standard Measurement Interface

Figure 7 Spectral Reflectance
Figure 8 Color Index Interface
2.4 Measurement
Measurement is divided into standard measurement and sample measurement.
The standard measurement is to measure the chromaticity data of standard sample, while sample measurement
is to measure the color different or contrast chroma data between standard sample and target sample.
After turning on the the instrument and finishing the white and black calibration, it is ok to do the standard
measurement and sample measurement (the users can set the corresponding color space and color index in the
main menu interface as required (please refer to Chapter 3 for details), and the system default setting is CIE lab
color space, ΔE * ab color formula, and the color index is empty). If it is not currently in the measurement interface,
continuously click the return key several times in the interface or tap the "testing key" to return to the
measurement interface.
2.4.1 Standard Measurement
In the “Standard Measurement” interface, Align the measuring aperture of the instrument to the tested sample
and press the “Testing key” tightly. Then there is a “Beep” voice, and the LED indication light will go from flashing
to off. After that, a “Beep” voice will come again, and it means measurement finished as shown in Figure 9
and in Figure 10.
Detail instructions of standard measurement interface are as following:

1) Interface Title: Indicating it is under interface of standard measurement.
2) Condition area: display system setting such as illumination,UV (on/off), Bluetooth (on/off), measurement mode
and the current time, date and power etc. Only when turn on Bluetooth or UV function, they will be highlighted.
3) Camera Locating: Click to use the camera to position the measurement, then press “Testing” key to complete
measurement.
4) Standard Name: Display the current measurement standard name, and click it to quickly modify, or the default
is No name.
5) Display mode: Click “SCI/SCE” to refresh the current measurement data in real time.
Note: The switchable of SCI or SCE model is only to display the current data, and the sample measurement mode
“SCI/SCE/I+E” is required to switch in the system setting (Refer to Chart 3.8.5). The default standard
measurement model is I+E, while the default sample measurement is SCI.
6) Illuminant: Click it to quickly switch different light source among the D 6 5,A,C,F1 ~ F12 etc.
7) Color space: Click it to quickly switch different color spaces among CIE lab,CIE XYZ,Hunterlab etc.
8) Observer angle: Click it to quickly switch the observer angle 10 and 2 degree.
9) Switch to sample measurement: Click it to sample measurement.
10) Measurement aperture: Indicate the current measurement aperture on use.
11) Tolerance setting: Click it to set the tolerance of the current sample.
12) Delete/Save: If the auto-save function opens, click it to delete the current data. If the auto-save function
closes, it will display save button, and click it to save the current data.
13) Page Turing: Click it to quickly switch the current data among data display area, spectrum display area and
color index display area (as shown in Figure 8).
14) Wavelength switch button: As shown in Figure 10, click or , the wave length and reflectance of the
current sample will switch as range of 10nm interval.

Figure 9 Standard measurement

Figure 10 Reflectance of standard measurement
2.3.3 Sample Measurement
After testing the standard, click “Sample measurement” to enter Sample Measurement interface. It is the same
steps with standard measurement. After measurement, it will display as shown in Figure 11 and in Figure 12.
Sample measurement is similar with standard measurement but it will display the color difference between the
current standard and measured sample.
Detail instructions of sample measurement are as following:
1) Interface Title: Indicating it is under interface of sample measurement.
2) Standard Name: Display the current measurement standard name, and all sample chromatic data is based on
the current standard.
3) Sample Name: Display the current measurement sample name, and click it to quickly modify, or the default is
No name.
4) Display mode: Click “SCI/SCE” to refresh the current measurement data in real time.
Note: The SCI and SCE of the display mode is the same as those of the standard measurement.
5) Sample Chromatic Data: Display the current chromatic data.
6) Delta E (Color Difference): The color difference between standard and sample.
7) Measurement Result: Display the test result under the color formula and tolerance. If the result is larger than
the tolerance value, it will show red “Failure”, or show green “Pass”. This function only exists when you turn on the
function of “Display Measurement Result” in the system setting.
8) Color Offset: Color deviation of the current sample compared with standard. It only displays color offset when
you turn on the function in the system setting.

9) Wavelength switch button: As shown in Figure 12, click or , the wave length and reflectance of the
current sample will switch as range of 10nm interval.
Figure 11 Sample Measurement

Figure 12 Reflectance of Sample Measurement
2.3.4 Average Measurement
When the measured sample is very large or relatively not very uniform, it needs to measure several points to get
an average reflectance to show the sample true chromatic value. This instrument can realize 2-99 times average
measurement.
In the main menu, click “Average Measurement” as shown in Figure 13, input average times and click to
confirm.
If input 1 time, it measures as usual; if more than 1, the output measurement results will be calculated on average
after the set number of measurement.
Figure13 Average Measurement
2.5 Connecting to PC
The instrument has a PC software with powerful extended function, and it could realize more chromatic data
analysis. It can connect to PC through USB cable or Bluetooth (Only for the model with Bluetooth function) .
2.5.1 USB Connection
First install the software on the computer, then use USB cable to connect PC and the instrument. The software
will aromatically connect to the instrument. If it successes to connect, it will display the icon of USB icon or if not, it
won’t display USB icon. If the connection is successful, the instrument can be fully controlled by software, and
relevant samples can be tested and analyzed.

2.5.2 Bluetooth Connection
For these models with Bluetooth function, it can connect to PC software through Bluetooth. 14
After installing the PC software, open Bluetooth function in the system setting, and set Bluetooth matching in the
PC software setting. If Bluetooth matching successes, it could measure the sample and standard or analyze the
data on the computer.If the connection is successful, the instrument can be fully controlled by software, and
relevant samples can be tested and analyzed.
2.6 Print
Micro Printer is an optional accessory and is required to be purchased separately.
Users should measure the sample first, and save all records. Then connect to micro printer by USB cable. Find
the printed data from the sample or standard records as shown in Figure 14, click “Operate” to pop up menu as
shown in Figure 15, and select “Print Data” to print the current data.
Figure14 Standard Records

Figure15 Print Data
3. Main Menu
Click in the main menu interface or in other interfaces to enter the main menu and set all system functions
from each submenu.
3.1 Data Management
Click “Data Manage” in main menu to enter data management interface as shown in Figure 16. Data
management is mainly for checking, operating the measured records and inputting standard data manually.

Figure 16 Data Management
3.1.1 Check Record
1)Check Standard Record
Select “Check Record” to enter “Standard Record” interface as shown in Figure 17. The measurement condition
shows the standard light mode, UV status and the measurement mode, measurement time & date.

Figure 17 Standard Record
Table of contents
Other 3nh Measuring Instrument manuals