ABB PGC5000 Series Original operating manual

—
ABB MEASUREMENT & ANALYTICS | 892J009MNAE
PGC5000 Gas Chromatograph
Dielectric Barrier Discharge Ionization
Detector (DBDID)
SERVICE INSTRUCTION
Measurement made easy

2| DBDID | 892J009MNAE
—
Contents
Contents ...........................................................................................................................2
List of Figures ...................................................................................................................2
Safety ...............................................................................................................................3
Safety symbol conventions ................................................................................................3
Potential safety hazards ....................................................................................................4
1. Introduction .................................................................................................................5
1.1 Usage....................................................................................................................5
1.2 Measurement Principle .............................................................................................5
2Service and maintenance ..............................................................................................5
2.1 Detector removal ....................................................................................................6
2.2 Detector installation.................................................................................................8
2.3 Detector operation...................................................................................................8
2.4 Scheduled maintenance............................................................................................9
2.5 Recommended spare parts........................................................................................9
3Troubleshooting............................................................................................................9
3.1 Symptom – Plasma won’t light................................................................................. 10
3.2 Symptom – No signal............................................................................................. 10
3.3 Symptom – Low signal ........................................................................................... 10
3.4 Symptom – Excessive Noise .................................................................................... 11
3.5 Symptom – Elevated baseline.................................................................................. 11
3.6 Symptom – Baseline drift........................................................................................ 11
3.7 Symptom – Spikes ................................................................................................ 11
4Specifications ............................................................................................................. 12
4.1 Physical ............................................................................................................... 12
4.2 Power.................................................................................................................. 12
4.3 Modes of operation ................................................................................................ 12
4.4 Detector sensitivity................................................................................................ 12
4.5 Gas consumption................................................................................................... 12
5Drawings ....................................................................................................................12
List of Figures
Figure 1-1: Dielectric Barrier Discharge Ionization Detector.........................................................5
Figure 2-1: PGC5000 Electronics compartment open ..................................................................6
Figure 2-2: Disconnect lines...................................................................................................6
Figure 2-3: DBDID amplifier assembly .....................................................................................7
Figure 2-4: DBDID power supply board....................................................................................7
Figure 2-5: Feedthrough tube.................................................................................................8
Figure 2-6: DBDID drawing....................................................................................................9
Figure 5-1: Detector port identification .................................................................................. 13
Figure 5-2: Power supply dip switch settings........................................................................... 13
Figure 5-3: Diagram flow PGC5000 air and carrier ................................................................... 14
Figure 5-4: PGC5000 oven electronics internal connections diagram sheet 1 ................................ 15
Figure 5-5: PGC5000 oven electronics internal connections diagram sheet 2 ................................ 16

892J009MNAE | DBDID | 3
—
Safety
Read these instructions carefully before installation and commissioning. These instructions do not
contain all details on all types of products and do not explain all assembly, operating, or maintenance
scenarios. Ask the manufacturer for further information.
The content of these instructions is neither part of nor provided for changing a previous or existing
agreement, promise, or legal relationship. All obligations of ABB result from the respective sales
contract, which also contains the full and solely valid warranty clauses. These are neither limited nor
extended by the content of these instructions.
Observe warning signs on packaging and on the device.
Assign only qualified and authorized specialists for the assembly, electrical connection, commissioning,
and maintenance of the equipment. Specialist qualifications include:
–Training or instruction and/or authorization to operate and maintain devices or systems according
to safety engineering standards for electrical circuits, high pressures, and aggressive media
–Training or instruction in accordance with safety engineering standards regarding maintenance
and use of adequate safety systems
WARNING:
According to IEC 60900, use only sufficiently insulated tools for the electrical
connection.
Also consider the following regulations:
–The applicable standards and safety regulations concerning the construction and operation of
electrical installations
–The regulation on technical working materials (safety guidelines for tools)
–The regulations and recommendations relating to explosion protection
–The recommendations for safe working in the case of installation in a Safety Integrity Level (SIL)
loop.
–The regulations that apply in the country of use
Safety symbol conventions
The following conventions appear throughout this document:
DANGER and WARNING hazards relate to personal injury and NOTICE hazards are associated with
equipment or property damage. However, under certain operating conditions, operating damaged
equipment can result in a degraded system or process performance leading to serious or life-
threatening injuries. Therefore, compliance with all DANGER, WARNING and NOTICE hazards is
required at all times.
DANGER
–Serious damage to health / risk to life. These symbols, and the signal word
"DANGER", indicate imminent danger or electrical hazard. Failure to observe this safety
information will result in death or severe injury. The text may state the hazard, how to avoid
the hazard, and the result if not followed. The
lightning bolt is for electrical danger; the
exclamation point is for general dangers.
DANGER
– Serious damage to health / risk to life. This symbol, and the signal word
"DANGER", indicate
s imminent danger or explosive hazard. Failure to observe this safety
information will result in death or severe injury. The text may state the hazard, how to
avoid the hazard, and the result if not followed.
!

4| DBDID | 892J009MNAE
WARNING
– Bodily injury. These symbols, and the signal word "WARNING", indicate a
potentially dangerous situation. Failure to observe this safety information
could result in death
or severe injury. The text may state the hazard, how to avoid the hazard, and the result if not
followed. The bolt is
for electrical warnings; the exclamation point is for general warnings.
NOTICE
– Equipment damage or loss of data or cybersecurity risk.This symbol
indicates a potential for equipment damage, loss of data or another unintended outcome.
Failure to observe this information may result in damage to or destruction of the product
and / or other system components.
IMPORTANT NOTE:
This symbol indicates operator tips, particularly useful information, or
important information about the product or its further uses
.
Potential safety hazards
DANGER – Serious damage to health / risk to life.Ensure that there are no hazardous or
flammable gases present in the immediate area of the analyzer. Protected enclosures that rely
upon air purge and pressurization will be jeopardized when the protected air is removed. During
this non-purged time, a danger exists for fire, explosion, damage
to property, and injury or death
to plant personnel.
WARNING – Bodily injury. high voltage safety:The DBDID uses a high voltage alternating
current discharge to create a plasma. The system operates at a high voltage and a low current
which is applied to the glass reactor tube at the round metal electrodes. This electrode should be
covered with the plasma shield whenever the plasma is operating.
Before performing any maintenance procedures or adjustments to the plasma cell,
make sure the
power is off. The best way to do this is to remove the power pack from the main line. This will
ensure that the plasma cell is not at voltage.
If the center rod of the plasma supply is not connected to ground, the center rod will have
voltage on it (approx. 40 Volts) and it will shock maintenance personnel. It will also spark
when connecting the ground wire to the center electrode; do not do this in a hazardous
environment. ABB recommends turning the plasma power off, allowing a minimum of 20
seconds for the DBDID to discharge and then connecting the center electrode to ground. If done
in the proceeding manner, the center electrode will not spark.
DANGER – Serious damage to health / risk to life.Ensure there is no hazardous atmosphere
present when performing maintenance on the unit. Do not separate components when energized.
This applies to all connectors and connections, cabling and wiring.
NOTICE – Equipment damage or loss of data.
Printed circuit boards and other electronics are
electrostatic discharge sensitive. Ensure that proper ESD measures are taken, as defined by
location.
!

892J009MNAE | DBDID | 5
—
1.Introduction
This service instruction is for reference when installing, operating, and servicing the ABB PGC5000
Dielectric Barrier Discharge Ionization Detector (DBDID). It is for use only by qualified personnel.
WARNING
– Bodily injury. Servicing of this product shall be performed by individuals who
are knowledgeable of the procedures, precautions, and hazards associate
d with equipment
containing hazardous energy circuits.
1.1 Usage
Dielectric Barrier Discharge Ionization Detector (DBDID) is used in the ABB PGC5000 Process Gas
Chromatograph.
It is used to detect parts per billion (ppb) and parts per million (ppm) concentrations of the following
applications:
Impurities in high purity gases and products
Halogenated hydrocarbons
BTEX (Benzene, Toluene, Ethylbenzene, and Xylenes)
Arsine and phosphine
Ethylene oxide
Ammonia
Nitric oxide in cold box applications
Ambient air applications
1.2 Measurement Principle
The dielectric barrier is a plasma discharge obtained using a high voltage alternating current applied
to a gas as it flows through a dielectric material. The gas is typically helium or argon, and the
dielectric material is a quartz or Pyrex tube.
Two electrodes are arranged within the detector. When high voltage is applied to the gas, a
breakdown occurs with a subsequent discharge from one electrode to the other electrode. The
presence of the dielectric barrier behaves as a capacitor in the localized region of the discharge. As
such, the dielectric barrier stores a substantial amount of energy for each discharge resulting in the
generation of highly excited state of the helium (or argon) atoms. This is called the reaction gas.
The sample components are ionized when they elute from the column. A second set of electrodes
measures the current generated from these ionized components. The output is sent to an
electrometer where it is amplified, measured, and displayed.
Figure 1-1: Dielectric Barrier Discharge Ionization Detector
2Service and maintenance
The following chapter includes a list of replacement parts, instructions for replacing spare parts, and
other maintenance procedures for the Dielectric Barrier Discharge Ionization Detector (DBDID).

6| DBDID | 892J009MNAE
Figure 2-1: PGC5000 Electronics compartment open
3
2
1
Legend: PGC5000 Electronics compartment open
Item
Description
1
DBDID amplifier assembly
2
DBDID power supply board
3
Flange nut
2.1 Detector removal
WARNING
– Bodily injury. High voltage present. Make sure that power to the PGC5000 has
been disconnected before
removing the detector.
To remove the detector from the PGC5000, perform the following steps:
Disable the isothermal oven temperature zone and allow the oven to cool to room temperature.
Power down the oven.
Disconnect the column, cell vent line and the reactor gas inlet from the cell body. See items 1-
3 in Figure 2-2.
Figure 2-2: Disconnect lines
1
2
3

892J009MNAE | DBDID | 7
Legend: Disconnect lines
Item
Description
1
Reaction gas inlet (plasma gas)
2
Detector inlet (chromatograph column connection for analytes to the detector)
3
Detector vent
Open the left side oven door and disconnect the coax cable (item 1 in Figure 2-3) from the
DBDID amplifier assembly.
Figure 2-3: DBDID amplifier assembly
1
On the DBDID power supply board, unplug the high voltage power connector (J2) (item 3), the
bias cable (RF2) (item 1), and the feedback cable (RF1) (item 2). See Figure 2-4.
Figure 2-4: DBDID power supply board
3
1
2
Legend: DBDID power supply board
Item
Description
1
Bias cable (RF2)
2
Feedback cable (RF1)
3
High voltage power connector (J2)

8| DBDID | 892J009MNAE
Loosen the 3 flange screws around the feedthrough tube.
Figure 2-5: Feedthrough tube
Unscrew the flange nut and remove the detector by pulling the detector to the inside of the
oven. The wires coming from the feedthrough tube are potted so they will need to be fed
through the opening to the inside of the oven.
2.2 Detector installation
To install the detector, perform the following steps:
WARNING
– Bodily injury. High voltage present. Make sure that power to the PGC5000 has
been disconnected before accessing the electronics compartment.
1. Remove power prior to installing the detector.
Perform the steps in section 2.1, above, in reverse order.
After reassembly, check for leaks.
2.3 Detector operation
This section details the steps to initialize operation of the detector.
IMPORTANT NOTE:
Allow the oven to heat up and stabilize with flows enabled before setting the
correct flow
.
1. Set flows by measuring the flow at the cell vent tube. Refer to the data pack for correct flows.
The detector should automatically light when power is applied. To verify, remove the front
cover of the detector and with low light you should be able to see the plasma glow a soft blue.
a. To remove the cover, take the 16 screws out of the housing.
b. Very gently pull the cover straight away from the body of the cell. Pay close attention not to
rub any of the detector components while removing the cover. Be sure to observe the
orientation of the cover and reinstall it in the same manner. Ensure the flame-proof or
explosion-proof surfaces of the cover or the body of the detector are not damaged. Reinstall
the DBDID detector cover and insert the 16 screws and tighten them to 16 in.-lb, using ABB
Tool TL1000/TL1002. There is a small cutout inside of the cover that will need to go back on
in the same orientation.
There is no need to push the DBDID ignite switch to light the plasma because when power and utilities
are added, the plasma will automatically light.

892J009MNAE | DBDID | 9
NOTICE
– Equipment damage. Ta k e g reat care to avoid any leaks associated with this
detector, as leaks can cause a variety
of problems. See Troubleshooting in section 3.
2.4 Scheduled maintenance
There are no parts within the detector power supply or electrometer which require scheduled
maintenance.
2.5 Recommended spare parts
Part Number
Description
Quantity
852A038-1
Power Supply DBDID
1
885A005-1
Detector Cell Assembly DBDID
1
851K006-1
FID/DBDID Amplifier Kit
(Order 851K006
-1 Amp Kit to build the standard
851A094-11412 per included 851J003 bulletin.)
1
Contact your local ABB sales and service representative for specific instructions for ordering spare
parts. See the last page of this instruction for contact information. Always include the information
listed in Equipment Identifications and Configuration Identification in your request.
When ordering parts for replacement, use the list of parts included in the Engineering Data Package
that was provided with the analyzer to insure the correct version of each part.
Figure 2-6: DBDID drawing
3Troubleshooting
This section describes general troubleshooting procedures, issues, and remedies.
Before proceeding with further troubleshooting procedures, first verify:
The system has power.
All switches and gas settings are correct.
Column flow rates and pressure settings are correct.
Using a VOM meter, verify the voltage between chassis ground and the test points on the
bottom of the power supply board are as follows:
•+10 volts DC Bias voltage
•+2.5 volts DC Feedback voltage
•+2.5 volts DC Plasma voltage

10 | DBDID | 892J009MNAE
To determine the source of the problem:
1. Identify whether the problem is with the chromatographic oven or the detector. In general, if
the problem is with component retention times and not component sensitivity, the problem is in
the chromatographic oven and not the detector.
Ask: “What was the last thing done to the system?” Most problems are associated with the
latest change made to the system, so this is the best place to begin look for the problem.
If the considerations above identify the detector as the primary cause of the malfunction, use the
following sections to help identify and correct the problem.
3.1 Symptom – Plasma won’t light
Possible Cause
Action(s)
No power
If the fuse is intact, contact customer support for further assistance.
Air in plasma
The presence of too much air in the plasma will often cause the
plasma not to light.
Purge the system with reaction gas for at least 15 minutes.
If the plasma still will not light, check for air leaks in all of the fittings.
Back pressure too high
If there is too high of a restriction in the exhaust line, the detector
plasma will not light.
Replace the exhaust/signal tube with a new one and see if the plasma
will light.
Wrong plasma gas
Verify that the reaction gas is either helium or argon. Replace if
necessary.
Defective electronics
See sections 3.2 and 3.3.
3.2 Symptom – No signal
Possible Cause
Action(s)
No Power
If the fuse is intact, contact customer support for further assistance.
Air in plasma
Observe the plasma tube in a dark environment. There should be a
pink or almost white glow in the
plasma tube. If the tube is blue and
the signal is low, there is nitrogen in the plasma.
Verify that the correct reaction gas (helium or argon) is being used.
Turn off the power to the plasma and check the gas supply, check for
leaks, and verify the proper flow rate.
Signal cable problem
Verify that the collector shield BNC fitting contacts the exhaust tube.
To do this, remove the BNC cable from the collector shield. Verify
contact between the center pin of the BNC and the exhaust tube using
a VOM or
continuity tester. Re-connect the BNC cable to the collector
shield. Verify the cable is connected to the collector shield and the
electrometer board.
Wrong bias voltage
Using a VOM meter, verify the voltage between the Bias test point and
chassis ground on the Power Supply Board. The voltage should be +10
volts DC.
3.3 Symptom – Low signal
Possible Cause
Action(s)
Leakage in the
chromatographic system
Verify retention times for the analytes of interest.
If OK, check for leakage on all the fittings on the detector end of the
system.
If retention times have shifted, check for leaks on the sample
introduction side of the system.
Wrong reaction gas
Using nitrogen as a reaction gas will result in a low signal. Verify that
the proper reaction gas is being used. Replace if necessary.
Plugged exhaust tube
Unique problem that will manifest itself with a lower signal and slower
retention times.
Visually
inspect the exhaust tube to identify if any materials have
plugged the tube. If plugged, insert a cleaning wire into the exhaust
tube to remove the blockage.
If these actions do not correct the problem, replace the exhaust tube.

892J009MNAE | DBDID | 11
3.4 Symptom – Excessive Noise
Possible Cause
Action(s)
Bad electrical connections
Check the signal cable connection and the bias voltage cable
connection.
Check the signal cable shield for fit. The fitting should be snug.
Verify connection between ground rod and ground.
Air Leak
Leak-check all of the fittings within the oven system to verify a leak-
free system.
Insufficient clearance for
the plasma electrode
The plasma O-ring on the plasma tube is designed to prevent this
issue. If the plasma electrode
appears to be too close to the
Swagelok
®fitting, remove power from the analyzer and remove the
plasma tube from the top nut.
Install or replace the plasma O
-ring on the plasma tube and re-install
the plasma tube.
3.5 Symptom – Elevated baseline
Possible Cause
Action(s)
Poor carrier gas
Background signal is directly linked to the quality of gases being used.
UHP (ultra
-high purity) gases should always be used. If gas quality is
suspected, replace gas with new UHP quality gas.
When UHP gases
cannot be used, a scrubber will need to be installed
to remove excess levels of oxygen, water, and trace hydrocarbons.
Air Leak
An air leak in the chromatographic system can also cause an elevated
baseline. Verify that there are no air leaks in the
chromatographic
system.
3.6 Symptom – Baseline drift
Possible Cause
Action(s)
Poor pressure control
Pressure controls not providing a consistent, pulse free, flow rate to
the detector can cause detector drift.
Monitor the pressure gauges on the supply cylinder and measure the
flow out of the exhaust port over a one
-hour time period to insure
correct flow conditions.
Column bleed
Re-condition or replace the column.
Plasma cell leak
Initially, this may appear as column bleed, but the signal will
equilibrate as the oven reaches an isothermal temperature.
Reset the plasma cell in the Swagelok
®
fitting and tighten to leak-free.
Contamination or late-
eluting peaks
It is possible that there is a late-eluting peak or contamination that is
coming off the column several runs after it was introduced.
Alter the cycle time of the injection to see is the baseline drift changes
location.
If this is confirmed, install filters to eliminate the contamination.
Extend the
instrument cycle time to elute the material prior to the
next analysis.
3.7 Symptom – Spikes
Possible Cause
Action(s)
Loose connections
Verify that the signal cable is securely attached to the electrometer
and the BNC fitting on the collector shield.
Missing ground leads
Verify that the ground lead from the plasma center electrode is
connected. If it is not, power the analyzer down and re
-connect the
center electrode ground lead.
Verify that the main power supply ground lead is connected to the
appropriate oven chassis ground point.

12 | DBDID | 892J009MNAE
Defective bias voltage
Using a VOM meter, verify the voltage between the Bias test point and
chassis ground on the Power Supply Board. The voltage should be a
stable +10 volts DC.
4Specifications
Note: All specifications are subject to change without notice.
4.1 Physical
Environmental (enclosure): The device is only suitable for use in clean, dry areas.
Operating Temperature Range: 50 to 150 degrees C (122 to 302 degrees F)
Installation and mounting: Integrated and configured with the PGC5000 series Smart Ovens.
4.2 Power
Voltage:
Plasma reactor: 12V, 1.0 A
Total power consumption: <20 W
4.3 Modes of operation
Helium ionization mode (HID)
In this mode, the detector is sensitive to all organic and inorganic components except neon. Operating
in HID mode, sub ppm levels of fixed gases can be measured.
Argon ionization mode (AID)
In this mode, the detector is not sensitive to fixed gases and methane but is sensitive to a wide range
of organic and inorganic volatile components. It is capable of ionizing any component with an
ionization potential less than 11.5eV.
4.4 Detector sensitivity
Sensitivity is a function of the total application design. It depends on the mode of operation and the
ionization potential of the analyte. Sensitivity values are given with a 4:1 minimum signal to noise
ratio.
HID mode: Sensitive to fixed gases down to 0.5 nanograms using hydrogen as an indicator.
AID mode: Sensitive to benzene down to 20 picograms.
4.5 Gas consumption
Reaction gas flow:
HID mode: 80 to 100 ml/min
AID mode: 5 to 20 ml/min
Carrier gas flow:
Packed columns: 20 to 60 ml/min
Capillary columns: 5 to 30 ml/min
Utility specifications:
UHP (ultra-high purity) for all carrier and reaction gas supplies.
5Drawings
This section contains drawings pertaining to the DBDID as installed in the PGC5000 gas
chromatograph.

892J009MNAE | DBDID | 13
Figure 5-1: Detector port identification
1
2
3
4
5
Legend: Detector port identification
Item
Description
1
Purge gas inlet. (Detector assembly purge not used in most applications.)
Flameproof breather must remain open or connected to optional purge system.
2
Reaction gas inlet (plasma gas)
3
Purge gas vent. (Detector assembly purge not used in most applications.)
Flameproof breather must remain open or connected to optional purge system
4
Detector inlet (chromatograph column connection for analytes to the detector)
5
Detector vent
Figure 5-2: Power supply dip switch settings
NOTICE
– Equipment damage. Dip switches on the power supply should never be
adjusted without factory support
.

14 | DBDID | 892J009MNAE
Figure 5-3: Diagram flow PGC5000 air and carrier

892J009MNAE | DBDID | 15
Figure 5-4: PGC5000 oven electronics internal connections diagram sheet 1

16 | DBDID | 892J009MNAE
Figure 5-5: PGC5000 oven electronics internal connections diagram sheet 2

892J009MNAE | DBDID | 17
—
ABB Inc.
Measurement & Analytics
Quotes: US-IAMA.inquiry@us.abb.com
Orders: US-IAMA.order@us.abb.com
Training: US-IAMA.training@us.abb.com
Support: upstream.support@us.abb.com
+1 800 442 3097 (opt. 2)
Additional free publications are available for download at:
www.abb.com/upstream
Main Office - Bartlesville
7051 Industrial Blvd
Bartlesville, OK 74006
Ph: +1 918 338 4888
Texas Office - Houston
3700 W. Sam Houston
Parkway S., Suite 600
Houston, TX 77042
Ph: +1 713 587 8000
We reserve the right to make technical changes or modify the contents of this document without prior
notice. With regard to purchase orders, the agreed particulars shall prevail. ABB does not accept any
responsibility whatsoever for potential errors or possible lack of information in this document.
We reserve all rights in this document and in the subject matter and illustrations contained therein.
Any reproduction, disclosure to third parties or utilization of its contents - in whole or in parts – is
forbidden without prior written consent of ABB.
MOXA®is a registered trademark of Moxa Inc.
Windows®is a registered trademark of Microsoft.
Swagelok®is a registered trademark of Swagelok.
892J009MNAE
Copyright©2022 ABB all rights reserved
Other manuals for PGC5000 Series
1
Table of contents
Other ABB Laboratory Equipment manuals