DT-311D Line Powered Digital Stroboscope
Operation Manual
Use in flammable environments is prohibited. Use in this manner may
result in fire or explosion.
Do not use or store in the following environments: Direct sunshine con-
densation, dust or caustic.
Do not alter or modify. Such action may cause damage and void war-
ranty.
Operate within 0-35°C (32-95°F), 35-85% RH. Use outside of this range
may alter operation of the unit.
Case may become excessively hot when used continuously for more
than 2 hours.
SPECIFICATIONS
Flashing Range: 60-12,000 fpm (flashes per minute)
Accuracy: ±0.01% F.S.
Resolution: 0.1 from 60-6,000; 1 from 6,001-12,000
Display: 5 digits 0.4” (10 mm) height
Flashtube Power/Life: Xenon ≤ 45W / 50,000,000
Flash Duration: 10-30 μs automatically adjusts with flash rate
External Trigger Input: 5 V dc ~ 12 V dc pulse signal input
(high level trigger) from sensor
Sensor Power Supply: 12 V dc up to 50 mA
Lux Rating: 8” (20 cm) Distance = 1550 lx @ 1500 fpm, 2350 lx
@ 6000 fpm with 15.5” (400 mm) irradiation diameter.
20” (50 cm) Distance = 650 lx @ 1500 fpm , 580 lx @ 6000 fpm
with 23.5” (600 mm) irradiation diameter
Sensor Power Supply: 5-12 VDC up to 50 mA
Power Requirement: 85-240 VAC 50/60 Hz
Operating temperature: 32° – 104°F (0 – 40°C)
Dimensions: 8.3 x 4.7 x 4.7” (210 x 120 x 120 mm)
Product Weight: 2.2 lb (1.0 kg)
Package Weight: 2.6 lb (1.2 kg)
Warranty: 1 Year
Certification: CE, RoHS
Included Accessories: Top & bottom handles, 9.8’ (3 m) power
cable, 3 pin female M16 threaded connector
The DT-311D is a portable, highly stable, industrial-quality strobo-
scope for non-contact inspection and observation of moving parts
by providing the user the illusion of stopped or slowed down equip-
ment movement. Constructed with a rugged, metal enclosure, the
DT-311D’s withstand harsh industrial usage. The universal pow-
er range (85-240 VAC) allows one model to be utilized on various
power voltages in a multitude of countries. The included top and
bottom handles aid in portability and usage flexibility. The control
adjustment settings allow the flash rate to be readily altered by the
user to visually inspect rotating or reciprocating machinery such as
gear teeth, rotors, shafts, rewinders, printing presses, etc. all with-
out stopping production. These features make the DT-311D ideal for
quality inspection and preventive maintenance of automated motion
equipment.
Distributed by: ABQ Industrial LP USA
Tel: +1 (281) 516-9292 / (888) 275-5772 eFax: +1 (866) 234-0451
Web: https://www.abqindustrial.net E-mail: [email protected]
Distributed by: ABQ Industrial LP USA
Tel: +1 (281) 516-9292 / (888) 275-5772 eFax: +1 (866) 234-0451
Web: https://www.abqindustrial.net E-mail: [email protected]
FEATURES & FUNCTIONS
Display
FPM Flash speed indication in flashes per minute
HZ Flash speed indication in Hz
INT Operation in internal user settings mode
EXT Operation in external sensor input settings mode
Function Keys
SIG Signal selection. Press this key to switch between
internal and external signal operation
MODE Switches between Hz and FPM
+ Increases the frequency (internal trigger mode only)
- Deceases the frequency (internal trigger mode only)
×2 Multiplies the flash rate/frequency by a factor of 2
(internal trigger mode only)
÷2 Divides the flash rate/frequency by a factor of 2
(internal trigger mode only)
EXT External trigger Input
ON Switch on button
2
OPERATION
The Use of Internal Trigger
Turn the power on. The system enters the internal trigger mode.
Turn the encoder knob, or press the “+” “-” keys to adjust the
strobe frequency or flash speed so that detected the object ap-
pears stationary. Use “× 2” “÷ 2” keys to make large changes to
the flash frequency.
To measure true revolutions per minute (RPM) by Marking:
1. “Mark” the object to be measured by either visually noting an
inherent distinguishing characteristic (such as a label, scratch,
etc.) or physically marking the object with a small piece of tape,
pencil mark, etc. If the object has a uniform shape, such as multi
blade fan or the shaft of a motor, you should give it an identifying
mark with reflective tape or paint to distinguish its direction.
2. Turn power switch on.
3. Set flash rate to the highest value and adjust downward.
4. The true RPM can be noted once the action appears frozen and
the first single image of the “mark” appears (see chart below and
accompanying diagram for further explanation).
5. To verify RPM reading, press “÷2”; a single image should ap-
pear again.
To measure true revolutions per minute (RPM) without Marking:
This example explains that if the object has a suitable, unique
shape, then the actual speed of the object may not need to be
determined by marking the object.
In this example, the rotation is below 7000 rpm, the following har-
monic image appears:
EXT AC85~240V
Harmonic images at 6000 and 4000 rpm are not a single still im-
age, but double and quadruple. Single image appears at 3000
and 1500 rpm. 1500 is half of 3000 and the same single image
appears, therefore, the actual speed of the object is 3000 rpm.
The Use of External Trigger
When in EXT mode: In this mode, the flash trigger signal frequen-
cy or flash rate speed is automatically adjusted according to an
external sensor’s input signal.
Wiring of the external sensor is as follows:
1: +12 V dc up to 50 mA
2: Ground
3: Trigger signal 5 - 12 V dc
3
TROUBLESHOOTING
Actual Speed
RPM
Strobe
Frequency
FPM
Multiple Number of
Still Images
3000 12,000 4 4
3000 9,000 3 3
3000 6,000 2 2
3000 3,000 1 1
3000 1,500 1/2 1
3000 1,000 1/3 1
Issue Troubleshooting Methods
Unit does not work
and no display
Check the power supply is proper
Power Switch is functioning
Whether internal fuse is burn out,
check and replace
Unit does not work
but displays properly
Silicon controlled rectifier / optical
coupler (IRF840/521-1) need
replacement
Strobe flashtube needs to be replaced
Check if unit is in external trigger mode
Does not flash under
external trigger mode
Check if the trigger sensors are
working properly
Flashes intermittently
Flashtube needs to be replaced
Flash but the flash
is darker or dimmer
than typical
Distributed by: ABQ Industrial LP USA
Tel: +1 (281) 516-9292 / (888) 275-5772 eFax: +1 (866) 234-0451
Web: https://www.abqindustrial.net E-mail: [email protected]
Distributed by: ABQ Industrial LP USA
Tel: +1 (281) 516-9292 / (888) 275-5772 eFax: +1 (866) 234-0451
Web: https://www.abqindustrial.net E-mail: [email protected]
DIMENSIONS (mm)
Distributed by: ABQ Industrial LP USA
Tel: +1 (281) 516-9292 / (888) 275-5772 eFax: +1 (866) 234-0451
Web: https://www.abqindustrial.net E-mail: [email protected]
Distributed by: ABQ Industrial LP USA
Tel: +1 (281) 516-9292 / (888) 275-5772 eFax: +1 (866) 234-0451
Web: https://www.abqindustrial.net E-mail: [email protected]

Other ABQINDUSTRIAL Test Equipment manuals

ABQINDUSTRIAL Horiba Scientific LAQUA 6583S-10C User manual

ABQINDUSTRIAL

ABQINDUSTRIAL Horiba Scientific LAQUA 6583S-10C User manual

ABQINDUSTRIAL Shimpo DT-700 Series User manual

ABQINDUSTRIAL

ABQINDUSTRIAL Shimpo DT-700 Series User manual

ABQINDUSTRIAL Shimpo DT-315AEB User manual

ABQINDUSTRIAL

ABQINDUSTRIAL Shimpo DT-315AEB User manual

ABQINDUSTRIAL DTX2 User manual

ABQINDUSTRIAL

ABQINDUSTRIAL DTX2 User manual

Popular Test Equipment manuals by other brands

BASETech BT-200 operating instructions

BASETech

BASETech BT-200 operating instructions

Fluke T5-600 user manual

Fluke

Fluke T5-600 user manual

TREND NETWORKS FiberTEK IV Quick reference guide

TREND NETWORKS

TREND NETWORKS FiberTEK IV Quick reference guide

Klein Tools NCVT-6 instructions

Klein Tools

Klein Tools NCVT-6 instructions

VeEX MTTplus-420 user manual

VeEX

VeEX MTTplus-420 user manual

CryptOn CSL640 Operating Instruction and Documentation

CryptOn

CryptOn CSL640 Operating Instruction and Documentation

Sperry instruments STOP SHOCK HGT6120 operating instructions

Sperry instruments

Sperry instruments STOP SHOCK HGT6120 operating instructions

Hanna Instruments HI 38078 instruction manual

Hanna Instruments

Hanna Instruments HI 38078 instruction manual

Sanwa SP-6D Operator's manual

Sanwa

Sanwa SP-6D Operator's manual

AEMC 6418 user manual

AEMC

AEMC 6418 user manual

Global Specialties LD-200P user manual

Global Specialties

Global Specialties LD-200P user manual

clare safe check 8 operating manual

clare

clare safe check 8 operating manual

MULTI MEASURING INSTRUMENTS 240+ instruction manual

MULTI MEASURING INSTRUMENTS

MULTI MEASURING INSTRUMENTS 240+ instruction manual

NRS AT 818 manual

NRS

NRS AT 818 manual

AEMC instruments 8510 user manual

AEMC instruments

AEMC instruments 8510 user manual

Phenix PM6 Operation and application manual

Phenix

Phenix PM6 Operation and application manual

Koehler K47001 Operation and instruction manual

Koehler

Koehler K47001 Operation and instruction manual

PNI AT6000 user manual

PNI

PNI AT6000 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.