Accora Configura Advance SC3 User manual

Configura Advance Chair
Instruction Manual
IFU-SC3-001US Rev 02
March 2021

2
Contents
Title Page
Welcome 2
General warnings 3
1. Means of delivery 4
2. Safety instructions 4
3. Intended use 4
4. Technical specification 4
5. Accessories 5
6. Assembly 6
7. Chair adjustment 8
8. Footplate operation & adjustment 11
9. Controls 12
10. Using the castor brakes 14
11. Functionality check 16
12. Cleaning & disinfection 16
13. Troubleshooting 16
14. Storage 16
15. Daily inspection 17
16. General maintenance 17
17. Guarantee 17
18. Disposal 17
19. Table of symbols 18
20. Contact details 18
Welcome
Dear Customer,
Thank you for purchasing an Accora healthcare
product. We feel sure that this product will
exceed your expectations.
This manual is important for understanding how to
use your new Configura Advance chair; please
read it before use. If you need further information,
or require the information in a different format,
please contact Accora.
See Contact Details in Section 20.
General
For USA, the Configura Advance chair is classified
as a Class 1 Medical Device.
For Europe, the Configura Advance chair is
classified as a Class 1 Medical Device in
accordance with the Medical Device Regulation
2017/745.
Notice to User
If a serious incident occurs in relation to this
medical device, affecting the user or the patient,
then the user or patient should report the serious
incident to the medical device manufacturer (or
distributor) and, in the European Union, the user
should also report the serious incident to the
Competent Authority in the member state where
they are located.
Accora Ltd, 38 Main Street,
Swords, Co. Dublin, Ireland, K67 E0A2
T: +353 (0)1 695 0614
Design Policy and Copyright
® and ™ are trademarks belonging to Accora
Ltd unless otherwise stated. As our policy is one of
continuous improvement, we reserve the right to
modify designs without prior notice. © Accora Ltd
2020.
Before using the Configura Advance chair,
you must read and understand the instructions
in this user manual. All actions and handling of
the chair must be performed in accordance
with the instructions in this manual.
Any actions that are inconsistent with the
manual are performed at your own risk and
Accora shall not be liable for any injury or
damage. Please ensure that the manual is
available to users and operators throughout
the chair’s service life.

3
GENERAL WARNINGS
1. Keep this Instruction Manual available for
future reference.
2. These instructions must be observed to ensure
the safe and effective use of this chair and the
safety of users and carers.
3. This chair must be assembled, positioned and
used in accordance with these instructions.
4. The safety features for operating the chair and
instructions concerning the chair must be
strictly observed.
5. The chair must not be exposed to smoke,
naked flame, extreme temperature, flammable
gases or other hazardous substances or
situations.
6. Accora shall not be held liable for any
damage, injuries or accidents arising from
unauthorised modifications, non-genuine spare
parts, negligence or use that is at variance with
this manual which can result in serious injury or
death.
7. Keep children and pets away from this chair
unless supervised by an adult as there is a risk of
injury and/or choking on small parts.
8. When routing cables for other electronic
equipment used with the chair (e.g. air cushion
pump), ensure cables cannot be squeezed,
crushed or damaged by the moving parts of
the chair.
9. It is essential to consult Accora in advance if
you wish to use the chair for any purpose
outside the use detailed in this manual.
10. When repairing the chair, only original
materials and components may be used,
otherwise Accora cannot guarantee against
any damage that might occur.
11. Inappropriate routing of accessory cables, e.g.
cushion pump cable, could lead to dangerous
electrical hazards if squeezed or crushed
between moving parts. The chair must not be
used if there is any visible damage to any
cables.
12. Do not move the chair when it is in the forward
tilt position.
13. This chair is not recommended for users outside
the weight specifications detailed in Section 4.
14. Do not modify this chair without the
authorisation of Accora.
15. Before operating this chair, ensure the patient
is safely positioned to reduce the risks of falling,
entrapment and imbalance.
16. Patients, or users, should be risk assessed to
ensure they are able to understand this manual
and to operate the Configura Advance chair
safely without risk to themselves or others.
17. Patients or users should only be allowed to
operate the chair independently if they are
able to understand the safety instructions in this
manual and have been risk assessed as
appropriate to do so.
18. Patients, or users, must be risk assessed before
using the chair to make sure that the chair is
suitable for the patient/user.
19. Surface temperatures can increase when
exposed to sunlight.
20. If the medical staff state that the patient’s
medical condition is inappropriate for the
patient to be able to adjust the chair
independently, the chair’s position must only
be adjusted by the carer.
21. The chair may not be used where there is a
danger of explosion or in the presence of
uncontained flammable liquids.
22. Damaged or torn fabric may cause injury.
23. It is recommended that a patient does not
smoke while using the chair.
24. The chair is not suitable for transporting patients
outside e.g. in vehicles or aircraft.

4
1. Means of delivery
The Configura Advance chair is supplied boxed. An
inspection must take place upon receipt to ensure the
delivery is complete and undamaged.
Any missing parts, faults or damage must be reported
immediately to the carrier and Accora in writing.
2. Safety instructions
1. Before using the chair, you must read the
instruction manual and use the chair in
accordance with it.
2. The chair must not be used if faults have been
detected on it that may injure the patient, staff or
a third person.
3. The chair must only be operated by persons who
are able to operate it in accordance with this
manual.
4. Users must make the patient aware of the control
functions that apply to the patient, subject to an
assessment by a professional.
5. Before using the chair, the user should understand
the chair and its functionality.
6. The safe working load, as specified in Section 4,
must never be exceeded.
7. If a patient is left alone in the chair, the chair
castors must be locked as an unlocked chair can
cause injury to a patient who leaves the chair or
changes position.
8. When operating the moving parts of the chair,
care must be taken to ensure that the patient,
other people and objects do not become
trapped.
9. The chair may not be used where there is a
danger of explosion or in the presence of
uncontained flammable liquids.
3. Intended use
1. The Configura Advance chair is intended to be
used as a portable chair with built in size, pressure
and posture options in the acute, long-term care
and homecare environments. The chairs are
intended for patients who do not exceed the safe
working load and height recommendations for the
device.
2. The Configura Advance chair range is intended for
use in primary healthcare environments where
supervision can be provided if necessary, for
example, long-term residential care, nursing homes
or convalescent units. The Configura Advance
chair may also be placed in private homes where
there is unlikely to be 24-hour supervision by clinical
staff, but where the patient or caregiver is
considered competent in its use.
3. A typical patient / user ranges from those who are
physically dependent on caregivers for their
mobility, social and daily care needs, through to
rehabilitating patients.
4. Patients using the chair may also have declining
cognitive awareness, with symptoms ranging from
mild forgetfulness and poor concentration through
to end-stage dementia.
5. Patients may require postural support for the
accommodation, encouragement and
management of good posture.
4. Technical specification
The table below lists the Configura Advance chair part
numbers. Part numbers may be suffixed (US etc) to
show regions.
General specification
Environmental information:
Note: Always store out of direct sunlight. If the chair is
stored in conditions outside the normal operating
range, it should be allowed time to stabilise, in normal
operating conditions, before use.
Description
Value
Model name
SC3
Overall dimensions
700mm W × 880mm D x 1300mm H
27.6in W x 34.6in D x 51.2in H
Chair castor
2x 125mm (5”) and 2x 150mm (6”)
with brake
Safe Working Load
160kg / 352lbs / 25st
Maximum Patient
Weight
160kg / 352lbs / 25 st
Mass of Chair
85.1kg / 188lbs
Flammability
Interliner fabric: TB117 - 2013
Expected service life
Typically 7 years
Condition
Temperature
Range
Relative
Humidity
Atmospheric
pressure
Operating
+10°C to +25°C
(+50°F to +77°F)
30% to 75%
(Non-
condensing)
700 hPa to
1060 hPa
Transport/
storage
-20°C to +50°C
(-4°F to +122°F)
Part number
Description
CHAIR-0-SC3-050
Advance chair, 125/150mm (5/6”)
castors, standard backrest, black
vinyl/interliner, manual

5
Configura Advance Chair position setting ranges:
Key parts of the chair:
5. Accessories
The following accessories are black vinyl or vapour
permeable fabric and have an interliner.
Part number
Description
BRPOSKT-0-SC1-050
Postural backrest includes all
metal frames and cushions
HRPF-0-SC1-060
Backrest profiled headrest
DRPSIDKT-0-SC1-050
Dropdown sidepieces and
covers, set of 2
DRPARM-0-SC1-250
Dropdown armrests, 400-550mm
(16-22”), set of 2
DRPARM-0-SC1-050
Dropdown armrests, 350-500mm
(14-20”), set of 2
DRPARM-0-SC1-150
Dropdown armrests, 400-550mm
(18-22”), narrow, set of 2
EXTLAT-0-SC1-060
External laterals, set of 2
TRAY-0-SC1-000
Tray
For more information, refer to the relevant instruction
manual or contact Accora.
Standard position range settings
Forward tilt position setting
1
2
3
4a
5
4b
Seat
Leg Rest
Foot Plate
Right Sidepiece
Left Sidepiece
Backrest
1
5
2
3
4b
4a

6
6. Assembly
1. Check that the delivery is complete and whether
any visible damage has occurred to the chair
during transport.
2. Remove the cardboard box and other packaging
and the chair should appear as fig 1.
3. The following components should be present:
A. Chair frame assembly
B. Right Sidepiece
C. Left Sidepiece
D. 3 x Backrest cushions
E. Footplate
F. Instruction manual
4. Remove the R-clip (Fig 2, Item 1), then remove the
clevis pin (Fig 2, Item 2).
5. Rotate the backrest upward and hold in place.
Align the mounting hole at the bottom of the gas
strut (Fig 3, Item 2) and the mounting hole at the
bottom of the backrest lockout slide (Fig 3, Item 3)
with the clevis at the back of the main frame (Fig
3, Item 1). Ensure the Tilt-in Space operating cable
(Fig 3, Item 4) is routed between the backrest and
the backrest actuator assembly.
6. Insert the clevis pin (Fig 4, Item 1) through the clevis
at the back of the main frame, the gas strut and
the backrest lockout slide. Secure the pin with the
R-clip (Fig 4, Item 2).
WARNING
Assembly MUST be carried out by suitably
trained and qualified personnel.
All functions MUST be tested and approved
after assembly by suitably trained and
qualified personnel.
Assembly MUST take place in a clear,
uncluttered area and children and pets
should be kept away.
If chair has become soiled or contaminated
during transit refer to cleaning and
disinfection instructions.
2
1
Figure 1 - Chair as delivered (side pieces are loose)
Figure 2 –Backrest clevis pin
Figure 3
4
1
2
3

7
7. There is a cover attached to the bottom of the
backrest to cover the backrest hinge. To fit this
cover, place the first part of the cover (Fig 5, Item
1) over the rear part of the hinge protection plate
(Fig 5, Item 2), then pull the front part of the cover
(Fig 5, Item 3) down over the hinge.
8. Figure 6 shows the backrest hinge cover from the
inside. Place the lower pocket of the backrest
hinge cover (Fig 6, Item 1) over the lower part of
the hinge protection plate (Fig 5, Item 4) to secure
the hinge cover in place.
9. Repeat steps 7 and 8 to secure the cover on the
other side.
10. Secure the backrest cushions to the backrest with
the Velcro flaps. Each cushion is attached to the
backrest with the Velcro flap at the top of the
cushion. The top cushion has a double width
Velcro flap and is marked Pillow –Top. The middle
and bottom cushions are identical.
11. Ensure the female poppers on the overlay are
engaged on the male poppers on the frame, 3 EA
on left- and right-hand side. (Fig 7, Item 1)
12. To fit the sidepieces, pull out the plunger located
under the seat on one side of the chair (Fig 8, Item
1) and rotate it through 90 degrees to lock it open.
Fig 9 shows the plunger locked open.
13. Undo the hand screw (Fig 8, Item 2).
Figure 7
Figure 4
2
1
Figure 5 –(Seat cushion shown removed for clarity.)
2
1
3
4
1
Figure 6
Figure 8
2
1
1

8
14. Locate the two sidepiece fixing bars into the chair
frame. Push the sidepiece all the way in (Fig 10)
and release the plunger by rotating it until the cap
engages with the base (Fig 11).
15. Ensure the plunger is engaged by moving the
sidepiece in and out –the sidepiece should not
move.
16. Tighten the hand screw to secure the sidepiece.
17. Repeat steps 12 to 16 to fit the other sidepiece.
18. Carry out a functional check (see section 11).
7. Chair adjustment
The Configura Advance chair can be adjusted in the
following areas:
1. Chair Width range –The chair width has 4 widths
that can be set, adjusted using the armrest locking
plungers and hand screws. The available width
dimensions are:
400 mm (16”)
450 mm (18”)
500 mm (20”)
550 mm (22”)
2. Chair Depth range –The chair depth can be
adjusted in 9 increments from approximately 422
mm (16”) to 512 mm (20”). The depth is adjusted
using a lever and slide mechanism.
3. Armrest Height range –The armrest height can be
set at 165mm (6.5”), 200mm (8”) or 225mm (9”).
The height is adjusted by changing the armrest
bracket fitting position.
4. Footplate Height range –The footplate height can
be adjusted from 330mm (13”) to 533mm (21”).
The height is adjusted using two plungers.
5. Chair Seat Height range –The Advance chair
frame can be adjusted to give seat heights from
the floor of 615mm (24.2”) and 540mm (21.2”).
Chair Width adjustment
1. Make sure a patient is not in the chair.
2. Make sure the brakes on all 3 locking castors are
locked on (See section 10).
3. On one side of the chair, underneath the base of
the sidepiece, locate the locking plunger (Fig 8,
Item 1).
4. Pull the plunger out and rotate 90 degrees to lock
the plunger open (Fig 9).
5. Loosen the hand screw (Fig 8, Item 2).
6. With two hands, pull or push the sidepiece until the
desired measurement is achieved. The
measurement can be read just inside the
sidepiece at the front between the seat and the
sidepiece (See Fig 12).
7. Pull the plunger out, rotate 90 degrees and let go
to lock the plunger in position (Fig 11).
Figure 9
Figure 10
Figure 11
WARNING
Adjustment MUST be carried out by suitably
trained and qualified personnel.
A clinical assessment and risk assessment MUST
be carried out by suitably trained and
qualified personnel before adjustments are
made to the chair.
Do not adjust the width, depth or height of
the chair when a patient or user is in the chair.
①
②
②
①

9
8. Move the sidepiece in and out to make sure that
the plungers have locked the sidepiece in position.
9. Tighten the hand screw.
10. Adjust the opposite sidepiece in a similar manner
to match the same measurement on both sides.
Chair Depth adjustment
1. Make sure a patient is not in the chair.
2. Make sure the brakes on all 3 locking castors are
locked on (See section 10).
3. At the back of the chair, under the seat on the left-
hand side, locate the depth adjustment lever (Fig
13, Item 1)
4. Turn the lever clockwise to release the backrest
slide mechanism and move the backrest to the
desired measurement with the other hand. The
depth setting can be read off the scale on the
back part of the seat frame (Fig 14, Item 1).
Figure 13
5. Release the depth adjustment lever and move the
backrest in and out slightly to ensure the locking
mechanism has engaged.
Armrest height adjustment
1. Make sure a patient is not in the chair.
2. Undo the hand screw and sidepiece locking
plunger (Figure 8).
3. Pull the sidepiece from the chair seat frame, undo
the screws (Figure 15) from the fixing bars.
Figure 15
4. Locate the fixing bars to the desired position for
armrest height, secure the fixing bars with screws.
Refer to Figure 15 for armrest height position and
fixing bar location.
5. Fit the sidepieces to the chair. Pull and rotate the
locking plunger to lock the fixing bar, tighten the
hand screws.
Figure 12
Figure 14
1
1
8”
6.5”
9”

10
Footplate Height adjustment
1. Make sure a patient is not in the chair.
2. Make sure the brakes on all 3 locking castors are
locked on (See section 10).
3. Locate the footplate height adjustment plungers
on either side of the legrest (Fig 16, Items 1&2).
4. Pull and hold the caps of the two plungers to
release the footplate height frame and move the
footplate to the desired height measurement. The
footplate height measurement can be read off the
scale located at the rear of the legrest (Fig 17).
5. Release the caps of the plungers to lock the
footplate height.
6. Ensure both plungers are locked and that they
both read the same measurement off the scale.
Seat Height adjustment
1. The chair is supplied on the high seat height
position.
2. Make sure a patient is not in the chair.
3. Remove the footplate (See section 8) and place
the chair carefully on its front.
4. Using a 13mm spanner and 5mm Allen key, undo
and remove the Nylock nut and shoulder bolt (Fig
19, Item 1 and Item 2). Keep the bolt, washers,
bushes, and nut for next use.
5. Using a 17mm spanner and 8mm Allen key, undo
and remove two bolts, nuts, and washers on left-
hand and right-hand side of base frame (Fig 18; Fig
19, Item 3).
6. Move the base frame to the new position. Figure
18 shows the relative positions of the chair supports
and the bolt positions.
7. Refit the bolts, nuts and washers taken out in step 5
in the new position (Fig 18; Fig 19, Item 3).
8. Refit the shoulder bolt, washers, bushes, and nut in
the required hole of the base frame. (Fig 20).
9. Place the chair back on its wheels and refit the
footplate (See section 8).
2
1
Figure 16
Figure 17
Figure 18 –Chair support fixings
Figure 19 –Actuator and chair mounting
Chair at Low Position
Chair at High Position
Chair front
②
①
③
③
Low seat
height position
High seat
height position

11
Note: After any of these adjustments a functionality
check must be carried out (See section 11).
8. Footplate operation & adjustment
The Configura Advance chair is supplied with a
removeable angle adjustable footplate. The footplate
has 3 angle settings: 90°, 105°and 120°.
The height adjustment system of the footplate is
separate from the footplate removal –for details on
how to adjust the footplate height refer to section 9 of
this manual. The footplate may be fitted, removed
and adjusted with the patient in the chair. Ensure the
patient is sitting securely and cannot slide out.
Operation –Fitting the Footplate
1. Ensure that the plungers that secure the footplate
(Fig 21, Items 1&2) are in the closed position (Fig
23).
2. Take the footplate (Fig 24) and insert the two
tapered spigots (Fig 24, Items 1&2) into the two
sockets either side of the footplate height
adjustment frame (Fig 21, Items 3&4). Push the
footplate into the tubes until the plungers latch
with an audible click.
3. Ensure the plungers are latched by moving the
footplate in and out slightly.
Operation - Removing the Footplate
1. Locate the plungers that secure the footplate
either side of the leg rest, at the bottom of the
footplate height adjustment frame (Fig 21, Items
1&2).
2. Turn the levers on the two plungers towards the
front of the chair until they latch. The lever arm
plungers are now locked in the open position.
(Fig 22).
3. Pull the footplate out.
Figure 20 –Actuator mounting
WARNING
Always engage the castor brakes when the
chair is stationery or left unattended.
Ensure the patient is in a stable sitting position
while the footplate is adjusted, fitted or
removed.
1
2
Figure 21
3
4
Figure 22 –Plunger Open
Figure 23 –Plunger Closed
1
2
Figure 24
Low seat
height position
High seat
height position

12
Adjustments –Changing the Footplate Angle
1. Locate the footplate angle adjustment plunger
underneath the footplate (Fig 25, Item 1).
2. Pull and hold the cap of the plunger whilst moving
the footplate to the desired measurement. The
measurement setting is shown on the scale (Fig 26,
Item 1) by the indicator on the frame (Fig 26, Item
2). Setting 1 = 90°, Setting 2 = 105° and Setting 3 =
120° from the vertical.
9. Controls
The Configura Advance chair has the following
controls:
1. Backrest –The backrest can be adjusted between
90 degrees and 130 degrees from the seat angle;
see page 5 for a diagram of this. The backrest
lever (Fig 27, Item 1) is used to adjust this angle.
The backrest is also fitted with a lockout slide (Fig
27, Item 4.) that can be used to restrict the use of
the backrest angle adjustment.
1. Tilt-in-Space –The tilt-in-space position can adjust
the seat angle from 7 degrees to 37 degrees from
the horizontal position; see page 5 for a diagram
of this. The Tilt-in- space lever (Fig 27, Item 2) is used
to adjust this angle.
Figure 25
1
Figure 26
2
1
WARNING
Check for obstructions around, above and
below the chair frame and position the chair
so that it can operate through the full
adjustment range without any possibility of
obstruction or entrapment.
Always engage the castor brakes when the
chair is stationery or left unattended.
Make sure the castor brakes are in the locked
position before using the manual controls.
Be aware of entrapment risks through opening
and closing gaps as the chair is adjusted.
Ensure the patient is in a stable sitting position
while any of the following adjustments are
made.
3
1
2
4
Figure 27 –Overview of controls

13
2. Forward Tilt –The chair can tilt forwards from 7
degrees normal tilt to 10 degrees of forward tilt
with lock bolts removal.
3. Legrest –The legrest can be adjusted between 75
degrees and 160 degrees; see page 5 for a
diagram of this. The Legrest lever (Fig 27, Item 3)
can be used to adjust this angle.
Backrest Adjustment
2. Make sure the brakes on all 3 locking castors are
locked on.
3. Locate the locking lever (Fig 28, Item 2) on the side
of the backrest lockout slide (Fig 28, Item 1).
4. Release the backrest lockout slide by lifting the
locking lever away from the backrest lockout slide.
The backrest can now be adjusted using the
backrest lever.
5. Pull the locking catch on the backrest lever, then
pull the main lever (Fig 27, Item 1). Move the
backrest to the desired measurement. The
backrest angle measurement can be read off the
scale at the top of the backrest adjustment slide
(Fig 28, Item 3).
6. Release the backrest adjustment lever to secure
the backrest at the desired angle.
7. If locking out the backrest is required, return the
locking lever to the locked position. If the locking
lever is too loose or too tight it may be necessary
to rotate the whole lever clockwise or anti-
clockwise respectively to achieve the desired
locking action.
8. If locking out the backrest is not required there is a
Velcro strap that can be used to hold the lever in
the unlocked position –see Fig 29.
Tilt-in-Space Angle Adjustment
1. Make sure the brakes on all 3 locking castors are
locked on.
2. Pull and hold the tilt-in-space lever (Fig 27, Item 2)
and move the chair to the desired tilt-in-space
angle.
3. Release the lever to lock the tilt-in-space angle.
Forward Tilt Adjustment
This function enables some patients to exit the chair
more easily, however it must only be used following a
risk assessment of the patient.
1. The chair is delivered with this function locked.
Remove the lock bolts on both LHS and RHS side
on the chair using an Allen key. Ensure the correct
bolts (Fig 30, Item 1) are removed.
2. Retain the lock bolts in case they need refitting in
the future.
1
Figure 28
2
3
Figure 29
WARNING
Footplate must be removed before using the
forward tilt function.

14
3. Remove the footplate (See section 8).
4. Make sure the brakes on all 3 locking castors are
locked on.
5. Ensure it is safe to move the chair into the forward
tilt position.
6. Pull and hold the tilt-in-space lever (Fig 27, Item 2)
and move the chair forward to the forward tilt
position.
Leg rest Adjustment
1. Make sure the brakes on all locking castors are
locked on.
2. Locate the leg rest adjustment lever (Fig 27, Item
3).
3. Pull and hold the lever, then move the leg rest to
the desired position.
4. Release the handle to secure the leg rest in
position.
10.Using the castor brakes
The Configura Advance has the following castor
system:
Front: One green pedal 125mm castor (directional
lock) and one red pedal 125mm castor (normal
brake).
Rear: Two linked pedal brake 150mm castors (pedal
marked green and red to show which end to press).
Care must be taken to ensure the castor brakes are
always locked when the chair is in use, being
assembled or dismantled, so that the chair does not
move accidentally.
Individually braked 125mm Castor Set
One green pedal 125mm castor (directional lock) and
one red pedal 125mm castor (normal brake).
1. To lock the castor brake, press on the outer edge
of the red lever. The lever will flip up.
2. To unlock the castor brake, press on the top of the
red lever to bring it back horizontal.
Brake on Brake off
The green castor has a directional locking function
that assists with pushing the chair in straight/forward
direction.
1. Push the chair forward so the castor is running in
the direction of travel.
2. To lock the castor swivel, press on the outer edge
of the green lever. The lever will flip up.
3. Push the chair forward, moving slightly from side to
side, to ensure
4. To unlock the castor swivel, press on the top of the
green lever to bring it back horizontal.
Directional lock on Directional lock off
1
Figure 30

15
Centrally Braked 150mm Castor Set
Two linked pedal brake 150mm castors (pedal marked
green and red to show which end to press).
The two rear 150mm castors of the chair can be
braked at the same time using the foot lever.
1. To lock the central brake, press the rear red part of
the foot lever down until the brake locks. Figure 31
shows the brake locked.
2. To unlock the brake, press the front green part of
the foot lever down until the brake disengages.
Figure 32 shows the brake unlocked.
Figure 32 –Brake unlocked
Figure 31 –Brake locked

16
11.Functionality check
These checks must be carried out after chair assembly,
adjustment or maintenance.
Checklist
1. Check overlay poppers are engaged (Fig 7, Item1)
2. Tilt the chair backwards
3. Tilt the chair forward to upright position
4. Raise the leg rest
5. Lower the leg rest
6. Lower the backrest
7. Raise the backrest
8. Check the correct function of the castors.
9. Ensure footrest is fitted correctly.
10. Ensure sidepieces are fitted correctly.
12.Cleaning & disinfection
Cleaning Information:
To disinfect the chair, only use detergents designed for
use in healthcare e.g. warm soapy water. Do not use
abrasives, scourers or other materials that could
damage the coating. Do not use corrosives, caustics or
strong acids.
Do not use detergents that could alter the structure or
behaviour of the plastics (petrol etc.).
Clean by wiping with a damp cloth. Alternatively use
Microfibre cleaning cloth without any soap.
For extreme soiling use a 5% Sodium Hypochlorite
solution.
The chair is not designed for maintenance in
automatic chair washers or for cleaning with
pressurised water, spraying, showering or steam
cleaning.
Accora cannot be liable for any damage or risk of
damage if inappropriate cleaning or disinfectant
agents are used.
Cleaning procedure:
1. Move the chair to where cleaning will take place
and lock the chair castors.
2. Remove the cushions and any accessories such as
profiled headrest, external laterals.
3. Clean as described in the “Cleaning Information”.
13.Troubleshooting
Problem
Possible solution(s)
Backrest will not
recline (manual
version).
Ensure the backrest lockout slide
is released –see section 10 of this
manual.
If the Configura Advance chair still does not function
correctly, please contact Accora for further advice.
14.Storage
For problem-free storage we recommend:
1. Remove the accessories
2. Wrap the chair and accessories or cover them so
that the coating and plastic parts are not
damaged
3. Chair should be stored in a temperature between -
20°C to +50°C (-4°F to +122°F).
4. Chair should be stored in a relative humidity (non-
condensing) between 30% and 75%
WARNING
Functionality check MUST be carried out by
suitably trained and qualified personnel.
Check for obstructions around, above and
below the chair frame and position the chair
so that it can operate through the full
adjustment range without any chance of
obstruction or entrapment.
Always engage the castor brakes when the
chair is stationary or left unattended.
WARNING
The chair must be disconnected from the
power supply when being cleaned or
disinfected.
All functions MUST be tested and approved by
a competent person after cleaning or
disinfection.
The chair MUST be cleaned and disinfected
before re-using the chair for a different
patient.
WARNING
Troubleshooting MUST be carried out by
suitably trained and qualified personnel.
Do not attempt to open any electrical part
enclosures.
All functions MUST be tested and approved by
a competent person after troubleshooting.

17
15.Daily inspection
Daily visual inspection is strongly recommended and
may be carried out by carer, user or another person.
The following checks must be carried out:
1. Does the chair operate as per its intended purpose
without unexpected noise or motion?
2. Are there any signs of abuse or excessive wear?
3. Is any fabric torn or damaged? Pay particular
attention to the seat overlay.
4. Check that the overlay is secured to the frame with
the press studs.
5. Are all fixtures and fittings tight and secure?
6. Does the chair frame appear stable and secure?
7. Are all accessories fitted in line with the accessory
manufacturer or accessory supplier’s instructions?
8. Are all the castor brakes in the locked position?
9. Are all electrical cables for accessories, (e.g. air
pump) secured and routed to prevent damage?
10. Is the area around, above and below chair clear of
possible obstruction?
11. Is the footplate securely attached?
12. Are the plungers securing the sidepieces locked
and secure?
13. Is there any risk of entrapment or patient injury?
14. Are any accessory electrical cables pinched,
crushed or damaged in any way?
If any damage, performance issue or cause for
concern is noted during this inspection the chair should
be withdrawn from service and appropriate steps
should be taken.
16.General maintenance
Repairs to the Configura Advance chair must be
carried by suitably trained and qualified personnel.
After any maintenance has taken place a functionality
check must be carried out (see section 11).
17.Guarantee
Model
Configura Advance chair
Warranty period
2 years
18.Disposal
In the event of the disposal of materials from the chair,
end-of-life parts must be disposed of in accordance
with current environmental regulations
WARNING
Maintenance MUST be carried out by suitably
trained and qualified personnel.
All functions MUST be tested and approved
after maintenance by suitably trained and
qualified personnel.
Do not carry out maintenance with service
user or patient on the chair.

18
19.Table of symbols
Warning, beware of potential hazard –refer to instruction for use
Refer to instructions for use
Complies with the European Medical Device Regulation 2017/745
Model number
Serial number
Manufactured date
Manufacturer
Unique Device Identification (UDI) label
The B symbol indicates this product has a degree of protection against
electric shock for type B equipment
IPX4
Degree of protection against liquid ingress
Do not dispose of in household waste
Degree of protection against electric shock: Class II Double Insulated
For indoor use only
Medical Device in accordance with EU Medical Device Regulation
2017/745.
EC Representative
20.Contact details
UK and Rest of World
USA
Address
Accora Ltd.
Charter House, Barrington Road
Orwell, Cambridge
SG8 5QP, UK
Accora Inc.
9210 Corporate Blvd. Suite 120
Rockville, MD 20850
USA
Telephone
+44 (0)1223 206100
+1 301-560-2400
Email
information@accora.care
Website
www.accora.care
www.accora.care

19
This page intentionally left blank

20
IFU-SC3-001US REV 02
This manual suits for next models
1
Table of contents
Other Accora Indoor Furnishing manuals
Popular Indoor Furnishing manuals by other brands

Argos
Argos Heart of House kent 3 door 3 drawer wardrobe... Assembly instructions

Homestead
Homestead 89-386-917 manual

Frank Olsen
Frank Olsen INTEL 1100 CORNER Assembly instructions

Anywhere Cart
Anywhere Cart AC-LOCKER Series user guide

DURAVIT
DURAVIT Vitrium VT 7551 Mounting instructions

Habitat
Habitat DERWENT DERW-BT-SS20-A Assembly instructions