ACP OMNISTIM FX2 User manual

COPYRIGHT © 2008 - 2021, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
OMNISTIM®FX2
PORTABLE PENS
Patient User Manual
Revised 10.14.2021
Part No. MNL-0004 Rev 8
This manual and all content therein is owned exclusively by Accelerated Care Plus ("ACP") and
is protected by copyright. This manual or any portion thereof may not be photocopied,
reproduced or translated to another language without the express prior written consent of ACP.
This manual may only be used by entities that have purchased the equipment or have
implemented the ACP program and are covered by an executed lease agreement. This manual
may not be used for any other purpose.
Any additional copies of the Manual shall be ordered from ACP. No changes or modifications
shall be made to the Manual without prior review and written authorization from ACP. No
authorization is given to market, sell, disclose, or exploit this Manual except as for purposes of
using the Equipment as contemplated by the Lease Agreement.
ACCELERATED CARE PLUS MAKES NO WARRANTY OF ANY KIND WITH REGARD
TO THIS MANUAL, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES
OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. Accelerated
Care Plus shall not be liable for errors contained herein or for incidental or consequential
damages in connection with the furnishing, performance or use of this Manual. The information
contained in this document is subject to change without notice


OMNISTIM
®
FX
2
PORTABLE PENS PATIENT USER MANUAL
- 1 -
COPYRIGHT © 2008 - 2021, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
ACP manufactures a premier line of rehabilitation technologies to assist health care professionals
with improved outcomes and quality-of-life for Patients. The ACP product line includes Pain
Control Systems, Muscle Stimulators, Interferential Therapy, Therapeutic Ultrasound, Pulsed
Shortwave Diathermy devices, and advanced Therapeutic Exercise Systems. Our OMNISTIM®,
OMNICYCLE®, OMNIVR®, OMNISTAND®, OMNIVERSA®, OMNISWD®and
SYNCHRONY represent the most recent worldwide advances available for therapeutic
application of electromedical devices and other rehabilitation technology.
ACP is internationally recognized for its contribution to research in the development of medical
applications for therapeutic rehabilitation. The Company sponsors and conducts research at
leading health care institutions and major universities throughout the world.

- 2 -
OMNISTIM
®
FX
2
PORTABLE PENS PATIENT USER MANUAL
COPYRIGHT © 2008 - 2021, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
TABLE OF CONTENTS
ELECTROTHERAPY INDICATIONS & CONTRAINDICATIONS .................... 3
Indications........................................................................................................................................... 3
Contraindications ................................................................................................................................ 3
Adverse Reactions .............................................................................................................................. 3
Warnings ............................................................................................................................................. 4
Precautions .......................................................................................................................................... 5
THE OMNISTIM® FX2 PORTABLE PENS ...................................................... 7
Delivery of the Omnistim® FX2 Portable PENS ............................................................................... 7
Introduction......................................................................................................................................... 7
Controls and Functions ....................................................................................................................... 8
Patient Operational Sequence ............................................................................................................. 9
TREATMENT PREPARATION.........................................................13
Treatment Site / Skin Inspection....................................................................................................... 13
ACP Reusable Pre-gelled Surface Electrodes .................................................................................. 13
Lead Wires ........................................................................................................................................ 13
CLEANING ..................................................................................................... 14
TROUBLESHOOTING.................................................................................... 14
TECHNICAL SPECIFICATIONS .................................................................... 15
STANDARD AND OPTIONAL ACCESSORIES ............................................ 16
TREATMENT CHART .................................................................................... 17
STANDARD LIMITED PRODUCT WARRANTY............................................ 17

OMNISTIM
®
FX
2
PORTABLE PENS PATIENT USER MANUAL
- 3 -
COPYRIGHT © 2008 - 2021, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
ELECTROTHERAPY INDICATIONS & CONTRAINDICATIONS
CAUTION: Federal law restricts this device for sale or use by, or on the order of,
a Practitioner licensed by the laws of the state in which he/she practices to use
or order the use of the device.
Please note that Accelerated Care Plus cannot provide medical advice. If you
have specific medical questions, please contact your healthcare professional.
Indications
The Omnistim®FX2Portable PENS is indicated for:
1.
Relaxation of muscle spasms.
2.
Re-education of muscle action.
3.
Prevention or retardation of disuse atrophy.
4.
Immediate post surgical stimulation of calf muscles to prevent venous thrombosis.
5.
Increased local blood circulation.
6.
Maintaining or increasing range of motion.
7.
Symptomatic relief and management of chronic intractable pain and as an adjunctive
treatment in the management of acute pain, post-surgical pain and pain associated with
post-traumatic injury.
Electrical muscle stimulator devices should be used under medical supervision for adjunctive
therapy for the treatment of medical diseases and conditions.
Contraindications
1.
Do not use this device on patients who have a cardiac pacemaker, implanted defibrillator,
or other implanted metallic or electronic device because this may cause electric shock,
Never connect lead wires to the power line or electro-surgery equipment. Use only the
lead wires recommended or approved by the manufacturer
2.
Never connect lead wires to the power line or electro-surgery equipment. Use only the
lead wires recommended or approved by the manufacturer
Note:
There is no contraindication to the application of Transcutaneous Electrical Stimulation or
Powered Muscle Stimulation over metal implants.
Adverse Reactions
Skin irritation and burns, beneath the electrodes, have been reported with the use of powered
muscle stimulators. Patients may experience skin irritation and burns beneath the stimulation
electrodes applied to the skin.

- 4 -
OMNISTIM
®
FX
2
PORTABLE PENS PATIENT USER MANUAL
COPYRIGHT © 2008 - 2021, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
Patients may experience headache and other painful sensations during or following the
application of electrical stimulation near the eyes and to the head and face.
Patients should stop using the device and should consult with their physicians if they experience
adverse reactions from the device.
Warnings
•
The long-term effects of electrical stimulation are unknown;
•
Do not apply stimulation over the patient’s neck because this could cause severe muscle
spasms resulting in closure of the airway, difficulty in breathing, or adverse effects on heart
rhythm or blood pressure
•
Do not apply stimulation across the patient’s chest because the introduction of electrical
current into the chest may cause rhythm disturbances to the patient’s heart, which could be
lethal. Stimulation should not be applied transthoracically in the vicinity of the heart, as
introduction of electrical current into the heart may cause cardiac arrhythmias.
•
Do not apply stimulation when the patient is in the bath or shower
•
Do not apply stimulation while the patient is sleeping; and
•
Do not apply stimulation while the patient is driving, operating machinery, or during any
activity in which electrical stimulation can put the patient at risk of injury.
•
Consult with the patient’s physician before using this device because the device may cause
lethal rhythm disturbances to the heart in susceptible individuals
•
Apply stimulation only to normal, intact, clean, healthy skin.
•
Do not operate this device until the User Manual, including all Indications for Use,
Contraindications, Warnings and Precautions, have been carefully read and understood.
•
Operation of this device or placement of lead wires, probes, pads and electrodes in close
proximity (less than 5 feet) to an operating shortwave or microwave diathermy unit may
produce instability in the device output or burns at the treatment site. Lead wires and
device can pick up the magnetic field output of the diathermy and through induction
convert it into an electrical field, transmit the energy into the patient increasing the current
density at the electrodes of applicators. Since the patient may not feel the 27 MHz
frequency, they lack the protective sensation and tissue burns could result. Short-wave
field could potentially damage or reset medical devices in close proximity to the drum
applicator.
•
Treatment should not be applied over the carotid sinus nerves, (located in the anterior neck
triangle), including, stellate ganglion, vagus nerve, or laryngeal or pharyngeal muscle.
Particular care should be taken for patients with a known sensitivity to the carotid sinus
reflex, as carotid sinus stimulation may alter blood pressure and cardiac contractility.
•
Do not apply treatment over testes, heart or eyes. Electrical stimulation may affect organ
function.
•
Do not apply over or in close proximity to active cancer (except in terminal / palliative /
hospice care), as therapy may increase blood flow to the tumor.
•
Treatment should not be applied when high fever is present over swollen, severe infection
(osteomyelitis, sepsis, tuberculosis, etc.) or inflamed areas/skin eruptions (phlebitis,
thrombophlebitis, varicose veins, etc.).
•
Do not apply over the lumbar or abdominal region, or over the uterus during pregnancy (to
prevent uterine contraction), or during menstruation as therapy may temporarily increase
menstrual flow.
•
Treatment should not be applied transcranially. Since the effects of stimulation of the brain
are unknown, stimulation should not be applied across the head, and electrodes should not
be placed on opposite sides of the head;
•
Stimulation should not be applied to patients connected to patient monitoring equipment,
as the stimulation may have an effect on the proper operation of the monitoring equipment.

OMNISTIM
®
FX
2
PORTABLE PENS PATIENT USER MANUAL
- 5 -
COPYRIGHT © 2008 - 2021, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
•
Stimulation should not be applied directly over external stimulator systems with lead wires
•
Neuromuscular electrical stimulation (NMES) should not be applied directly over or in
close proximity to Deep Vein Thrombosis (DVT), as it activates the muscle and causes
muscle contractions. This should be avoided in tissue following an acute DVT when the
thrombosis is not completely resolved. Therapists should follow the guidelines provided
by the referring physician on recommended activity level and modality use. If the patient
is not permitted exercise, NMES therapy should be avoided. Generally, NMES over a
DVT of six weeks or less should be avoided altogether.
Precautions
•
The safety of electrical stimulation during pregnancy has not been established;
•
Some patients may experience skin irritation or hypersensitivity due to the electrical
stimulation or electrical conductive medium. The irritation can usually be reduced by using
an alternate conductive medium, or alternate electrode placement.
•
Application site and settings should be based on the guidance of the prescribing practitioner.
•
All equipment and accessories should be kept out of the reach of children or unqualified
persons.
•
Do not connect this device to any wall outlet that has not been properly grounded, or to any
electrically non-isolated medical device Powered muscle stimulators should be used only
with the leads and electrodes recommended for use by the manufacturer. Use this device
only with the leads, electrodes, and accessories recommended by the manufacturer. Use only
ACP specified accessories and/or supplies with ACP devices. Do not use any power cords,
or power supplies, other than the ones provided or specified for this device. Use of any other
power supply could seriously damage the device and will void the warranty.
•
The use of conductive mediums other than specifically approved pre-gelled or self-adhering
electrodes such as ultrasound gel or lotion, hand or body lotion, electrolyte spray mist, paper
towels, non-approved reusable or disposable pre-gelled or self-adhering electrodes— are not
advised for use with Omnistim Systems.
•
When cleaning the device, never immerse them or wash them with water. See the infection
control section in this manual for cleaning instructions. Devices should not be submerged in
water or other liquids.
•
Failure to follow the manufacturer’s prescribed maintenance for this device may lead to
device failure and transient or unreliable performance. State and federal survey and JCAHO
require all equipment to be maintained and calibrated according to the manufacturer
recommended schedules.
•
A potential electric shock hazard exists once the device outer casing has been in part, or
fully, removed. Only qualified service personnel should perform Service and repairs.
Warranty will be voided if the outer casing has been removed or tampered with.
•
Use caution when the patient has a tendency to bleed internally, such as following an injury
or fracture. Do not apply over areas of hemorrhage or active bleeding.
•
Inspect and cleanse the skin prior to application. Following treatment check the skin for
evidence of irritation or burns, and if present, treat as appropriate. If the patient has, or
complains of, skin irritation following treatment; shorten the treatment time on the next
treatment session, or use an alternative type of therapy or electrode placement.
•
Gradually increase the output intensity/power to required dose or patient tolerance while
monitoring the device display.

- 6 -
OMNISTIM
®
FX
2
PORTABLE PENS PATIENT USER MANUAL
COPYRIGHT © 2008 - 2021, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
•
Caution should be taken with patient exhibiting psychological or physical hypersensitivity
to the therapeutic treatment. Several attempts should be made to place them at ease so that
their confidence and cooperation can be gained during the treatment.
•
The treatment area should be checked from time to time, and if there is evidence of, or if the
patient complains of, pain during treatment, adjust the output downward until it is tolerated
by the patient. If the patient continues to complain of pain, discontinue the treatment and
shorten the treatment time on the next treatment session, or use an alternative type of therapy
or electrode placement.
•
Do not apply treatment directly over/under hot or cold packs. Caution is recommended when
treatment follows the application of hot or cold therapy, which may alter the patient’s
sensation. Application of thermal agents over areas of impaired circulation should be
performed with caution as the circulation may be insufficient to heat or cool the tissue,
altering the patient’s perception of warmth and pain, and burns or tissue necrosis may result
from subsequent treatment.
•
Caution is recommended when treatment follows the application of medicated patches,
salves, or creams which may alter the patient’s sensation. If there is a medical necessity to
perform such treatments, these patients should be monitored diligently during application.
The effect of electrical stimulation may be altered by the presence of these materials on the
patient’s skin.
•
Caution should be used over areas of body where circulation is impaired, or which lack
normal sensation. Absent or diminished sensation should be avoided or, if unavoidable,
treated with caution. Establishment of acceptable intensity levels for desensitized areas may
be related to the intensity levels tolerated on normal skin in opposite or related body parts.
•
Caution should be used in the presence of recent surgical procedures, fractures or healing
bone and soft tissue when muscle contraction may disrupt the healing process.
•
Caution should be used for patients with suspected or diagnosed epilepsy. Patients with
suspected or diagnosed epilepsy should follow precautions recommended by their
physicians.
•
Electrodes should not be placed in direct contact or in close proximity (one inch or less) of
each other during treatment. Electrodes placed in contact or in close proximity can lead to
high energy density and skin burns under or between the electrodes.
•
Care should be used when removing electrodes after treatment, in order to minimize the
potential for skin tearing. Skin should be inspected after removal of electrodes for any signs
of tearing or irritation.
•
Do not connect the stimulator to any electrical equipment for combination therapy except
the Omnisound® family of ultrasounds.

OMNISTIM
®
FX
2
PORTABLE PENS PATIENT USER MANUAL
- 7 -
COPYRIGHT © 2008 - 2021, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
THE OMNISTIM®FX2 PORTABLE PENS
Delivery of the Omnistim®FX2 Portable PENS
Upon receipt of your Omnistim®FX2Portable PENS,
inspect the shipping container and contents for any obvious
or concealed damage. All ACP products are packaged
carefully for rapid, safe delivery. We guarantee delivery in
perfect condition to the postal or delivery services. However
any damage or loss incurred during transportation or
delivery is the postal or delivery company responsibility. If
damage or loss to the product and/or container is obvious or
suspected, appropriate notation must be made on the signed
freight bill at the time of delivery. All damage claims should
be promptly filed with the delivering carrier and must be
initiated by the addressee where the package was to be
delivered. Retain the original shipping container and inserts
for validation of damage claim or use at a later date.
Introduction
The Omnistim®FX2Portable PENS provides Patterned Electrical Neuromuscular Stimulation
(PENS), which is a form of stimulation that replicates the correct firing patterns of muscles
(agonist and antagonist or reciprocal muscle pairs) in Upper and Lower Extremity Triphasic
(ballistic), or Upper Extremity Biphasic (reciprocal) patterns. This approach to neuro re-ed
provides a comfortable, precisely timed sensory input, which duplicates the firing activity of
sensory nerves and muscles during voluntary activity.
The Omnistim®FX2Portable PENS also provides Medium Frequency Alternating Currents
(MFAC) for pain management applications. Its generator produces medium frequency 5000 Hz
current. Two output circuits with independent intensity controls are provided. The output of
each circuit is easily determined in milliamps through the display screen. The digital timer
allows the operator to select the length of the total treatment time in minutes.

- 8 -
OMNISTIM
®
FX
2
PORTABLE PENS PATIENT USER MANUAL
COPYRIGHT © 2008 - 2021, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
Controls and Functions
6
8
2
4
(1)
Main Power Switch – is used to turn the device on and/or off.
(2)
Program Button A – is used to select the first program prescribed.
(3)
Program Button B – is used to select the second program prescribed.
(4)
Output Adjustment Channel A – is used to adjust the intensity level of channel A.
(5)
Output Adjustment Channel B – is used to adjust the intensity level of channel B.
(6)
Channel A Connection – plug the green marked lead into the green channel,
which corresponds to the Program Button A being used.
(7)
Channel B Connections – plug the blue marked lead into the blue channel, which
corresponds, to the Program Button B being used.
(8)
Display – displays program information and intensity readings. The A and B at the
top of the display mark the channels.
(9)
Battery Compartment – this is where the 9-volt battery is located.
7
3
5
1
9

OMNISTIM
®
FX
2
PORTABLE PENS PATIENT USER MANUAL
- 9 -
COPYRIGHT © 2008 - 2021, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
Patient Operational Sequence
Follow the sequence below. The number in parenthesis references the location of the
button from the control section on the previous page.
1.
Install battery in compartment located on the back of the unit (9).
2.
Connect lead wires to corresponding bar code under the display window (green lead
wire in the left input (6), blue lead in the right input (7).
3.
Wash the skin with water, (use soap if dirty).
4.
Connect black unit lead wires to white electrode wires.
5.
Press the power on/off button (1) to power on the device.
6.
At power on, the display (8) will default to the pre-set program selected for your
condition.
a.
If a second program has been ordered by your physician, press Program
Button B (3) to access the additional program.
b.
To return to the first program, press Program Button A (2).
7.
Apply electrodes to treatment area prescribed by your physician or therapist.
8.
Adjust stimulation intensity to level prescribed by your physician or therapist.
a.
Increase / decrease (4) Output A. Press (+) button to increase intensity in
channel A. Press the (-) button to decrease intensity in channel A.
NOTE: If the channel indicator flashes or you see this symbol you must
reset the output level. (See troubleshooting section).
Attention/Caution
b.
Increase / decrease (5) Output B. Press (+) button to increase intensity in
channel B. Press the to decrease intensity in channel B. See note above.
NOTE: Shortly after reaching the desired intensity setting, the unit will
lock the setting preventing an accidental increase in the intensity. A lock
icon will be indicated on the display. To unlock, you must first press the (-
) button to decrease the intensity, and then press the (+) button to increase
the intensity to the new desired level. If you see the reset output symbol
you must press (-) button to decrease the intensity, and then press the (+)
button to increase the intensity to the new desired level.
Lock Decrease

- 10 -
OMNISTIM
®
FX
2
PORTABLE PENS PATIENT USER MANUAL
COPYRIGHT © 2008 - 2021, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
9.
The treatment timer will automatically end the treatment in the time prescribed.
Treatment ends when bars show up on the display (8).
10.
Turn the unit off by pressing the Power Button (1).
11.
Remove the stimulation electrodes. Repack the re-useable electrodes into their foil
pouch electrodes and follow the directions on the label for electrode care and
storage.
NOTE: If you are using single use electrodes, discard after each use.
12.
Clean unit with ACP Germicidal Wipes.
13.
Store unit in carrying case.
14.
Record treatment activity and setting on form provided.
1). Install battery in compartment located on
the back of the unit.
2). Connect lead wires to corresponding
bar code under the display window
(green lead wire in the left input,
blue lead in the right input.
3). Wash the skin with water (use soap if
dirty).
4). Connect black unit lead wires to
white electrode wires.

OMNISTIM
®
FX
2
PORTABLE PENS PATIENT USER MANUAL
- 11 -
COPYRIGHT © 2008 - 2021, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
5). Press the power on/off button to power on
the device.
6). Select Program Button A or Program
Button B as directed by your
physician or therapist.
7). Apply electrodes to treatment area
prescribed by your physician or therapist.
8). Adjust stimulation intensity to level
prescribed by your physician or
therapist.
NOTE: To readjust stimulation intensity,
remove the lock by pressing (-) Button and
then (+) Button to preferred level.
9). Treatment is completed when bars
appear in the Display Window.

- 12 -
OMNISTIM
®
FX
2
PORTABLE PENS PATIENT USER MANUAL
COPYRIGHT © 2008 - 2021, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
10). Remove the stimulation electrodes.
Repack the re-useable electrodes into
their foil pouch electrodes and follow the
directions on the label for electrode care
and storage.
NOTE: If you are using single-use
electrodes, discard after each use.
11). Turn the unit off by pressing the
Power Button.
12). Clean unit with ACP Germicidal Wipes. 13). Store the device in carry case.
14). Record treatment activity and setting on
form provided.

OMNISTIM
®
FX
2
PORTABLE PENS PATIENT USER MANUAL
- 13 -
COPYRIGHT © 2008 - 2021, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
Treatment Preparation
Treatment Site / Skin Inspection
Thoroughly cleanse the treated area with soap and water to remove oils, creams, dirt, and sweat; this
will ensure uniform current conduction across the skin. After cleansing, inspect and evaluate the
skin’s integrity and sensation prior to treatment. Avoid absent or diminished sensation; if
unavoidable, treat with caution. Establishment of acceptable intensity levels for desensitized areas
may be related to the intensity levels tolerated on normal skin in opposite or related body parts.
Frequently monitor the intensity level and skin response during all treatments.
Stinging, burning or other painful sensation under the electrodes on normal or desensitized areas
is an indication of increased current density under part or the entire electrode surface. In this case,
slowly but immediately reduce the current intensity to zero; remove the electrodes to inspect the
surface skin. Recheck your application techniques.
Immediately after treatment, clean and thoroughly inspect the skin under the electrode. Peripheral
vasodilatation along with systemic vasomotor responses can lead to redness (hyperemia) directly
under both electrodes. Inform the patient of this normal after effect and that the redness will
disappear within an hour or two. Apply topical agents to the reddened area under the electrodes if
needed to decrease post-treatment irritation. Persistent skin irritation could be due to repeated
stimulation of the same electrode site or a possible allergic reaction to the conductive mediums,
tapes, elastic wraps, and/or cleaning and disinfectant solutions. Therefore, use additional electrode
stimulation sites to decrease or eliminate skin irritation on electrically sensitive patients. If skin
irritation persists with alternate site applications, decrease the treatment times and lower the
intensities; if necessary, discontinue treatment. If an allergic reaction is suspected, attempt to
identify and change the allergic substance(s). If skin irritation persists, discontinue treatment until
the source of irritation is determined.
By far the most common error with reported faulty machines is inadequate or improper conductive
medium interface or lead wire breakage. Because of the increased current density available with
pulsed or continuous medium frequency currents, a proportionally greater degree of conductive
medium interface problems exists and should be monitored by the clinician.
ACP Reusable Pre-gelled Surface Electrodes
Remove the electrodes from their foil packaging. Connect the electrodes to the device. Cleanse
the skin, and then apply the electrodes over the treatment site points according to the electrode
placements techniques described in this manual. Various sizes of electrodes are available
dependent upon muscle size of the area to be treated. Follow the enclosed infection control
procedures. Review the warnings and application directions on the electrode packaging.
Lead Wires
Inspect the full length of the lead wires for signs of frayed or cut wires and loose connections
where the lead wires join the jack plug and tip pins. Insert the plug completely. Allow the lead
wires to hang freely with no excessive strain on the connector.
NOTE: The use of conductive mediums other than specifically approved pre-gelled or self
adhering electrodes such as ultrasound gel or lotion, hand or body lotion, electrolyte spray
mist, paper towels, non-approved reusable or disposable pre-gelled or self-adhering
electrodes—are contraindicated for use with Omnistim®Systems.

- 14 -
OMNISTIM
®
FX
2
PORTABLE PENS PATIENT USER MANUAL
COPYRIGHT © 2008 - 2021, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CLEANING
Cleaning / Disinfecting of the OMNISTIM®FX2PORTABLE PENS
It is recommended to clean the device and lead wire attachments after each use to control the
spread of infection.
•Clean the Omnistim®FX2Portable PENS after each use with ACP germicidal wipes. Wipe
common contact surfaces, such as control panel, lead wires, and probe tips with germicidal
disposable wipes and allow to air dry. This technique will inactivate most bacteria and viruses.
This will also facilitate removal of contaminants from the equipment and accessories.
•Disposable/reusable electrodes are for individual patient use only and should not be shared
with others.
•All disposable electrodes should be discarded after each use. Do not attempt to clean and
reuse disposable electrodes.
TROUBLESHOOTING
The following table lists machine problem symptoms and possible areas to check for the problem
causes. If these suggested measures do not correct the machine malfunction, call your physician
for assistance.
PROBLEM
CAUSE
REMEDY
Unit will not power on
•
No batteries
•
Low batteries
•
Install batteries
•
Replace batteries
•
Verify type of batteries
•
Inspect battery contacts
Display shows low
battery symbol
•
Battery voltage is too low
•
Replace batteries for future use
Channel indicator
flashing and caution
symbol is displayed
•
Batteries are too low to perform treatment
•
Lead wires not properly connected
•
Lead wire to electrode not connected
•
Improper skin preparation
•
Replace batteries
•
Replace lead wire(s) if defective
•
Properly connect leads as needed
•
Properly prepare skin prior to treatment
Patient feels surging or
spiking sensation
•
Lead wire(s) breakage
•
Electrode(s) lead wire breakage
•
Replace lead wire(s) if defective
•
Replace electrode(s)
Patient cannot detect
output
•
Failure of lead wire(s)
•
Electrode(s) failure, or electrode gel
dried out
•
Device failure
•
Replace lead wire(s) if defective
•
Remove electrode(s) and replace
if necessary
•
Contact ACP customer service
Symbols
Caution or Attention. There is a problem. No patient detected.
Output locked. Decrease output to unlock then increase output.
Reduce output before increasing the channel to desired level.

OMNISTIM
®
FX
2
PORTABLE PENS PATIENT USER MANUAL
- 15 -
COPYRIGHT © 2008 - 2021, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
TECHNICAL SPECIFICATIONS
GENERAL:
Dimensions: 1.5" (38mm) D x 2.5" (65mm) W x 5.4" (137mm) H
Weight (Including batteries): 0.35lbs. (0.16kgs)
Operating Power: 9V alkaline
Battery Life:
New alkaline batteries operate the system for 40 hours
at full output and over 100 hours at normal settings.
Battery voltage emblem is displayed when battery low
voltage is indicated.
Display System: LCD Display
Push Buttons:
Polyester embossed overlay for tactile feel and infection
control.
System Memory:
The system remembers all prior custom settings from
treatment to treatment-in non-volatile memory.
System Architecture:
CMOS integrated micro-controller with on board memory
and instruction set.
STIMULATION SYSTEM:
Output:
Constant current up to maximum preset current limit of
60mA into a 500 ohm load.
PENS (PATTERNED ELECTRICAL NEUROMUSCULAR STIMULATION) PROGRAMS:
Waveform:
Asymmetric Biphasic Pulsed Current 0-60mA average single
pulse current into a 500 ohm load.
Phase Duration: Set at 70µs; Variable from 50µs to 120µs
Pulse Rate: Set at 50 Hz burst pattern
TENS (MEDIUM FREQUENCY CURRENT) PROGRAMS:
Carrier Frequency:
Output channel A, or B fixed at 5.0 KHz. Modulated at
Burst Rate.
Burst Rate:
Three programs, Sensory 80-120 BPS, Motor 2-15 BPS,
Sensory Motor 15-2-100 BPS
Rate Scan: 0-20 seconds for rate sweep
TIMER FUNCTIONS:
Treatment Timer:
Set to 20 minutes by default; Adjustable for 1-30
minutes in one minute increments.
CAUTION: Federal law restricts this device to sale by or on the order of a physician (or other
health practitioner licensed by their State).
ACP reserves the right to change technical specifications and product availability without notice.

- 16 -
OMNISTIM
®
FX
2
PORTABLE PENS PATIENT USER MANUAL
COPYRIGHT © 2008 - 2021, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
OMNISTIM®FX2PORTABLE PENS
STANDARD AND OPTIONAL ACCESSORIES
ITEM
ITEM NO.
DESCRIPTION
300200A
OMNISTIM®FX2
PORTABLE PENS
Offers 2 channels of
programmable
stimulation with PENS
and TENS programs.
28126
OMNISTIM®FX2a
PORTABLE PENS
Hard Carry Case
74747 9V Alkaline battery
38155
2x2 Reusable E-stim
Electrodes
(4 ea/pkg, 10 pkg/bx)
61227
2x4 Reusable E-stim
Electrodes
(4 ea/pkg, 10 pkg/bx)
https://acplus.com/acp-
technology-user-
manuals
OMNISTIM®FX2
PORTABLE PENS
Professional User
Manual
https://acplus.com/acp-
technology-user-
manuals
OMNISTIM®FX2
PORTABLE PENS
Patient User Manual
Infection Control Supplies
ITEM
ITEM NO.
DESCRIPTION
55536 Super Sani-Cloth®Wipes, Single
Use Packets (50 pkt/bx)
44425
Super Sani-Cloth®Wipes, Tub (160
wipes/tub)
96849 Sani-Cloth®Wipes w/ Bleach, Tub
(75 wipes/tub)
63574 Barrier Tubing
3" x 1200’/roll

OMNISTIM
®
FX
2
PORTABLE PENS PATIENT USER MANUAL
- 17 -
COPYRIGHT © 2008 - 2021, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
TREATMENT CHART
Mark the prescribed program, button, and intensity prescribed. Place date and time in the column
provided, and comment on the intensity level.
DAY
TIME
PROGRAM
(PENS 1, 2, 3,
TENS 1, 2, 3)
BUTTON
( A / B )
INTENSITY
( 1 to 100 )
COMMENTS
( TWITCH OR
TINGLE)
ELECTRODE PLACEMENT GUIDE
Provide the patient with the Patient Electrode Placement Guide and indicate the electrode placement
for each treatment prescribed. Mark each electrode location with (+) or (–) sign to indicate the polarity
of the electrodes, or use red and black marker to identify the color code for each wire. Red represents
(+) and Black represents (-). Mark the blue (B) and Green (A) channels.
STANDARD LIMITED PRODUCT WARRANTY

- 18 -
OMNISTIM
®
FX
2
PORTABLE PENS PATIENT USER MANUAL
COPYRIGHT © 2008 - 2021, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
The warranty information provided in this section is applicable only to products
purchased from ACP, directly or through an authorized dealer. This section does
not apply to leased products. The terms of maintenance and repair of any leased
products are detailed in the separately executed agreement between the parties.
Warranty Coverage
This warranty provides coverage, for Equipment purchased, against manufacturer’s defects in
material and workmanship, and extends to the original owner of the product during the warranty
period for that product. Only those items returned to the ACP Service Center within the warranty
period, and also within thirty (30) days after notification to ACP of the defect, shall be eligible
for repair under the Standard Limited Product Warranty. Buyer is responsible for shipping cost
associated with sending the Equipment to the ACP Service Center. ACP shall ship Equipment
to Buyer after repair at no cost to the Buyer provided repair is deemed to be under warranty.
ACP may, at its discretion and only for valid warranty claim, repair or replace any part(s) that
prove to be defective during the warranty period.
Warranty Exclusion
Any and all warranty coverage will be void if any of the following have occurred:
1.
The product contains repairs or replacement parts not furnished by ACP.
2.
The product is damaged resulting from misuse or negligence.
3.
The product has been tampered with and/or altered, including serial number alteration.
Note: Use of the Equipment with accessories and/or supplies not approved by ACPL for use with
the Equipment may void the warranty if such accessory or supply item caused damage to the
Equipment.
Warranty Period
The following coverage is provided at no additional cost to the Buyer:
New Equipment / Product. Products purchased as new from ACP are warranted against
manufacturer’s defects in material and workmanship for a period of two (2) years from the
date of purchase.
Refurbished Equipment / Product. Products purchased specifically as Refurbished
Equipment are warranted against manufacturer’s defects in material and workmanship for a
period of one (1) year from the date of purchase.
Accessories. All accessories for ACP equipment / products are warranted against
manufacturer’s defects in material and workmanship for a period of three (3) months from the
date of purchase.
Warranty Validation
The following information needs to be provided to the ACP Customer Support
representative prior to the product being returned under warranty coverage:
1.
Buyer name or account number as it appears under the “Bill TO” on the ACP or
recognized ACP Dealer invoice.
2.
Invoice Date and Number
3.
Model number, description, and serial number of equipment
4.
Detailed description of the problem
Table of contents