ACS contsys DAL-311x0x0S User manual

Technical features:
• red display of -19999…99999 digits (optional: green, orange or blue)
• installation depth: 120 mm without plug-in terminal
• min/max memory
• 30 parameter driven supporting points
• display flashing at threshold value exceedance / undercut
•[O]-key (zero key) for triggering of Hold, Tara
• permanent min/max-value recording
• volume measurement (totaliser)
• mathematical functions like e.g. reciprocal value, square root, squaring or rounding
• setpoint generator
• sliding averaging
• brightness control
• programming interlock via access code
• protection class IP65 at the front
• plug-in screw terminal
• option: sensor supply
• option: galvanic isolated digital input
• option: 1 or 2 analog outputs
• option: 2 or 4 relay outputs or 8 PhotoMos outputs
• option: interface RS232 or RS485
• accessories: PC-based configuration kit PM-TOOL incl. CD and USB-adapter for devices
without keypad and for a simple adjustment of standard devices
User manual M3
Direct current / direct voltage signals: 0/4-20 mA, 0-10 VDC
96x48
M3_11GB.pdf update: 10.04.2015
Technical features:
• red display of -19999…99999 digits (optional: green, orange or blue)
• installation depth: 120 mm without plug-in terminal
• min/max memory
• 30 parameter driven supporting points
• display flashing at threshold value exceedance / undercut
•[O]-key (zero key) for triggering of Hold, Tara
• permanent min/max-value recording
• volume measurement (totaliser)
• mathematical functions like e.g. reciprocal value, square root, squaring or rounding
• setpoint generator
• sliding averaging
• brightness control
• programming interlock via access code
• protection class IP65 at the front
• plug-in screw terminal
• option: sensor supply
• option: galvanic isolated digital input
• option: 1 or 2 analog outputs
• option: 2 or 4 relay outputs or 8 PhotoMos outputs
• option: interface RS232 or RS485
• accessories: PC-based configuration kit PM-TOOL incl. CD and USB-adapter for devices
without keypad and for a simple adjustment of standard devices
User manual M3
Direct current / direct voltage signals: 0/4-20 mA, 0-10 VDC
96x48
M3_11GB.pdf update: 10.04.2015
Operating Instructions
DAL-311x0x0S
Direct current / direct voltage signals: 0/4-20 mA, 0-10 VDC

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected]
2
Order code
power supply
0 100-240V AC
2 10...40V DC galvanic seperated
3 100-240V AC with sensor supply 24V DC/50mA and digital input 1)
4 10...40V DC galvanic seperated with sensor supply 24V DC/50mA and digital input 1)
Y other voltages
function input
0 0/4...20 mA, 0...10V DC
function output
A display + 2 relay outputs (changeover)
B display + 4 relay outputs
C display with analog output 0-10V/4-20mA, switchable
D display + 2 relay with analog output 0-10V/4-20mA, switchable
E display + 4 relay with analog output 0-10V/4-20mA, switchable
Y others outputs
0 standard conguration
S standard, protection IP65
DAL-311
1) sensor supply
only at 0/4...20 mA,
0...10V DC input
Note
S00

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected] 3
Contents
1
1. Brief description 2
2. Assembly 3
3. Electrical connection 4
4. Description of function and operation 6
4.1. Programming software PM-TOOL 7
5. Setting up the device 8
5.1. Switching on 8
5.2. Standard parameterisation (flat operation level) 8
Value assignment for the triggering of the signal input
5.3. Programming interlock „RUN“ 11
Activation/Deactivation of the programming interlock or change into
professional or flat operation level
5.4. Extended parametersation (professional operation level) 11
5.4.1. Signal input parameters „INP“ 11
Value assignment for the triggering of the signal input incl. linearisation
5.4.2. General device parameters „FCT“ 15
Superior device functions like Hold, Tara, min/max permanent, setpoint value function /
nominal value function, averaging, brightness control, as well as the control of the digital input and
keyboard layout
5.4.3. Safety parameters „COD“ 19
Assignment of user and master code to lock or to receive access to defined parameters such as
analog output and alarms, etc.
5.4.4. Serial parameters „ser“ 20
Parameters for interface definition
5.4.5. Analog output parameters „Out“ and „OU2“ 21
Analog output functions
5.4.6. Relay functions „rel“ 24
Parameters for setpoint definition
5.4.7. Alarm parameters „AL1…AL4“ 26
Actuator and dependencies of the alarms
5.4.8. Totaliser (Volume metering) „tot“ 28
Parameters for calculation of the sum function
6. Reset to factory settings (default values) 29
Reset parameters onto the delivery state
7. Alarms / Relays 30
Functional principle of the switching outputs
8. Interfaces 31
Connection RS232 and RS485
9. Sensor aligment 32
Diagram of functional sequences for sensors with existing adjustable resistor
10. Technical data 33
11. Safety advices 35
12. Error elimination 36

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected]
4
1. Brief description
2
1. Brief description
The panel meter instrument M3-11 is a 5-digit device for direct current / direct voltage signals and a visual
threshold value monitoring via the display. The configuration happens via 4 keys at the front or via the
optional PC software PM-TOOL. The integrated programming interlock prevents unrequested changes of
parameters and can be unlocked again with an individual code. Optional the following functions are
available: a supply for the sensor, a digital input for triggering of Hold (Tara), two analog outputs and
interfaces for further evaluating in the unit.
With help of the galvanic isolated setpoints (optional), free adjustable limit values can be controlled and
reported to a superior master display.
The electrical connection is done via plug-in terminals on the back side.
Selectable functions like e.g. the recall of the min/max-value, an averaging of the measuring signals, a
nominal presetting or setpoint presetting, a direct threshold value regulation during operation mode and
further measuring setpoints for linearisation, complete the modern device concept.
The panel meter instrument DAL-311 is a 5-digit device for direct current / direct voltage signals and a visual
threshold value monitoring via the display. The conguration happens via 4 keys at the front or via the
optional PC software PM-TOOL. The integrated programming interlock prevents unrequested changes of
parameters and can be unlocked again with an individual code. Optional the following functions are
available: a supply for the sensor, a digital input for triggering of Hold (Tara), two analog outputs and
interfaces for further evaluating in the unit.
With help of the galvanic isolated setpoints (optional), free adjustable limit values can be controlled and
reported to a superior master display.
The electrical connection is done via plug-in terminals on the back side.
Selectable functions like e.g. the recall of the min/max-value, an averaging of the measuring signals, a
nominal presetting or setpoint presetting, a direct threshold value regulation during operation mode and
further measuring setpoints for linearisation, complete the modern device concept.

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected] 5
2. Assembly
3
2. Assembly
Sealing
48,0
3,0
139,0
M3
Gap for physical unit
1. After removing the fixing elements, insert the device.
2. Check the seal to make sure it fits securely.
3. Click the fixing elements back into place and tighten the clamping screws by hand. Then use a
screwdriver to tighten them another half a turn.
CAUTION! The torque should not exceed 0.1 Nm!
The dimension symbols can be exchanged before installation via a channel on the side!
Please read the Safety advices on page 35 before installation and keep this user manual for future reference.

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected]
6
3. Electrical connection
4
3. Electrical connection
Type M3-1VR5B.0001.W70BD supply 10-40 VDC galv. isolated, 18-30 VAC
Type M3-1VR5B.0001.S70BD supply 100-240 VAC, DC ±10%
Type DAL-311x0x0S
Type DAL-311x0x0S
DAL-311x0x0S with digital input in combination
with 24 VDC sensor supply
DAL-311x0x0S with digital input and external
voltage source

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected] 7
3. Electrical connection
5
Connection examples
Below please find some connection examples that show practical applications. For devices with current inputs
/ voltage inputs, without sensor supply.
M3 devices
With current respectively voltage input in combination with a 24 VDC sensor supply.
DAL-311x0x0S in combination with a
2-wire-sensor 4-20 mA
DAL-311x0x0S in combination with a
3-wire-sensor 0/4-20 mA
DAL-311x0x0S in combination with a
3-wire-sensor 0-10 V
DAL-311 devices

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected]
8
6
4. Description of function and operation
4. Description of function and operation
Level Key Description
Menu-level
Change to parameterisation level and deposited values.
Keys for up and down navigation in the menu level.
Change into operation mode.
Parameterisation-
level
To confirm the changes made at the parameterization level.
Adjustment of the value / the setting.
Change into menu level or break-off in value input.
Menu-group-level
Change to menu level.
Keys for up and down navigation in the menu group level.
Change into operation mode or back into menu level.
Operation
The operation is divided into three different levels.
Menu level (delivery status)
This level was designed for the standard settings of the device. Only menu items which are sufficent to set
the device into operation are displayed. To get into the professional level, run through the menu level and
parameterise prof under menu item RUN.
Menu group level (complete function volume)
Suited for complex applications as e.g. linkage of alarms, setpoint treatment, totaliser function etc. In this
level function groups which allow an extended parameterisation of the standard settings are availabe. To
leave the menu group level, run through this level and parameterise uloc under menu item RUN.
Parameterisation level:
Parameter deposited in the menu item can here be parameterised. Functions, that can be changed or
adjusted, are always signalised by a flashing of the display. Settings that are made in the parameterisation
level are confirmed with [P] and thus saved. Pressing the [O]-key leads to a break-off of the value input and
to a change into the menu level. All adjustments are saved automatically by the device and changes into
operating mode, if no further key operation is done within the next 10 seconds.

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected] 9
7
4. Description of function and operation
Function chart:
4.1 Parameterisation software PM-TOOL:
Part of the PM-TOOL are the software on CD and an USB-cable with device adapter. The connection
happens via a 4-pole micromatch-plug on the back side of the device, to the PC-side the connection
happens via an USB plug.
System requirements: PC incl. USB interface
Software: Windows XP, Windows VISTA
With this tool the device configuration can be generated, omitted and saved on the PC. The parameters can
be changed via the easy to handle program surface, whereat the operating mode and the possible selection
options can be preset by the program.
CAUTION!
During parameterisation with connected measuring signal, make sure that the measuring signal has no
mass supply to the programming plug. The programming adapter is galvanic not isolated and directly
connected with the PC. Via polarity of the input signal, a current can discharge via the adapter and destroy
the device as well as other connected components!

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected]
10
8
5.1. Switching on
Once the installation is complete, start the device by applying the voltage supply. Before, check once again
that all electrical connections are correct.
Starting sequence
For 1 second during the switching-on process, the segment test (8 8 8 8 8) is displayed followed by an
indication of the software type and, after that, also for 1 second the software version. After the starting
sequence, the device switches to operation/display mode.
5.2. Standard parameterisation: (Flat operation level)
To parameterise the display, press the [P] key in operating mode for 1 second. The display then changes to
the menu level with the first menu item TYPE.
5. Setting up the device
Menu level Parameterisation level
Selection of the input signal, TYPE:
Default: sens.u
Available as measuring input options are 0-20 mA, 4-20 mA or 0-10 VDC signals as works
calibration (without application of the sensor signal) and SenSU (voltage) or SENSA (current) as
sensor calibration (with the sensor applied). Confirm the selection with [P] and the display
switches back to menu level.
Setting the end value of the measuring range, END:
Default: 10000
Set the end value from the smallest to the highest digit with [▲] [▼]and confirm each digit with
[P]. A minus sign can only be parameterised on the highest value digit. After the last digit, the
display switches back to the menu level. If Sens was selected as input option, you can only
select between noca and cal. With noca, only the previously set display value is taken over, and
with cal, the device takes over both the display value and the analogue input value.
Setting the start/offset value of the measuring range, offs:
Default: 0
Enter the start/offset value from the smallest to the highest digit with [▲] [▼]and confirm each
digit with [P]. After the last digit the display switches back to the menu level. If Sens was
selected as input option, you can only select between noca and cal. With noca, only the
previously set display value is taken over, and with cal, the device takes over both the display
value and the analogue input value.
5. Setting up the device

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected] 11
9
5. Setting up the device
Menu level Parameterisation level
Setting the decimal point, dot:
Default: 0
The decimal point on the display can be moved with [▲] [▼]and confirmed with [P]. The
display then switches back to the menu level again.
Setting up the display time, SEC:
Default: 1.0
The display time is set with [▲] [▼]. The display moves up in increments of 0.1 up to 1 second
and in increments of 1.0 up to 10.0 seconds. Confirm the selection by pressing the [P] button.
The display then switches back to the menu level again.
Selection of analog output, Out.rA:
Default: 4-20
Three output signals are available: 0-10 VDC, 0-20 mA and 4-20 mA, with this function, the
demanded signal is selected.
Setting up the final value of the analog output, Out.En:
Default: 10000
The final value is adjusted from the smallest digit to the highest digit with [▲] [▼]and digit by
digit confirmed with [P]. A minus sign can only be parameterised on the highest digit. After the
last digit, the device changes back into menu level.
Setting up the initial value of the analog output, Out.OF:
Default: 00000
The final value is adjusted from the smallest digit to the highest digit with [▲] [▼]and digit by
digit confirmed with [P]. A minus sign can only be parameterised on the highest digit. After the
last digit, the device changes back into menu level.
Threshold values / limits, LI-1:
Default: 2000
This value defines the threshold, that activates/deactivates an alarm.
then

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected]
12
10
5. Setting up the device
Menu level Parameterisation level
Hysteresis for limit values, HY-1:
Default: 00000
The delayed reaction of the alarm is the difference to the threshold value, which is defined by
the hysteresis.
Function for threshold value undercut / exceedance, Fu-1:
Default: high
A limit value undercut is selected with Louu (for LOW = lower limit value), a limit value
exceedance with High (for HIGH = higher limit value). If e.g. limit value 1 is on a threshold level
of 100 and allocated with function High, an alarm is activated by reaching of the threshold level.
If the threshold value was allocated to Low, an alarm will be activated by undercutting the
threshold value, as long as the hysteresis is zero.
The same applies to LI-2 !
User code (4-digit number-combination, free available), U.CodE:
Default: 0000
If this code was set (>0000), all parameters are locked for the user, if LOC has been selected
before under menu item run. By pressing [P] for 3 seconds in operation mode, the display
shows COde. The U.Code needs to be entered to get to the reduced number of parameter sets.
The code has to be entered befor each parameterisation, until the A.Code (master code) unlocks
all parameters again.
Master code (4-digit number-combination, free available), A.CodE:
Default: 1234
All parameters can be unlocked with this code, after LOC has been activated under menu item
run. By pressing [P] for 3 seconds in operation mode, the display shows COde and enables the
user to reach all parameters by entering the A.codE. Under run the parameterisation can be
activated permanently by selecting ULOC or ProF, thus at an anew pushing of [P] in operation
mode, the code needs not to be entered again.

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected] 13
11
5. Setting up the device
Menu level Parameterisation level
Activation / deactivation of the programming lock or completion of the standard
parameterisation with change into menu group level (complete function range), run:
Default: uloc
Choose between the deactivated key lock Uloc (works setting) and the activated key lock Loc, or
the change into the menu group level ProF with the navigation keys [▲] [▼]. Confirm the
selection with [P]. After this, the display confirms the settings with "-- - - -", and automatically
switches into operating mode. If Loc was selected, the keyboard is locked. To get back into the
menu level, press [P] for 3 seconds in operating mode. Now enter the CODE (works setting 1234)
that appears using [▲] [▼] plus [P] to unlock the keyboard. FAIL appears if the input is wrong.
To parameterise further functions ProF needs to be set. The device confirms this setting with
„-- - - -„and changes automatically in operation mode. By pressing [P] for approx. 3 seconds in
operation mode, the first menu group InP is shown in the display and thus confirms the change
into the extended parameterisation. It stays activated as long as ULOC or LOC is entered in menu
group RUN.
5.4. Extended parameterisation (Professional operation level)
Menu level Parameterisation level
Selection of the input signal, TYPE:
Default: sens.u
There are several measuring input options: 0-20 mA, 4-20 mA or 0-10 VDC signals as works
calibration (without application of the sensor signal) and SenSU (voltage) or SENSA (current) as
sensor calibration (with the sensor applied). Confirm the selection with [P] and the display
switches back to menu level.
Menu group level
Menu level
5.4.1. Signal input parameters
5.3. Programming interlock „RUN“

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected]
14
5. Setting up the device
12
Menu level Parameterisation level
Setting the end value of the measuring range, END:
Default: 10000
Set the end value from the smallest to the highest digit with [▲] [▼]and confirm each digit with
[P]. A minus sign can only be parameterised on the highest value digit. After the last digit, the
display switches back to the menu level. If Sens was selected as input option, you can only
select between noca and cal. With noca, only the previously set display value is taken over, and
with cal, the device takes over both the display value and the analogue input value.
Setting up the start/offset value of the measuring range, offs:
Default: 0
Enter the start/offset value from the smallest to the highest digit with [▲] [▼] and confirm each
digit with [P]. After the last digit the display switches back to the menu level. If Sens was
selected as input option, you can only select between noca and cal. With noca, only the
previously set display value is taken over, and with cal, the device takes over both the display
value and the analogue input value.
Setting the decimal point, dot:
Default: 0
The decimal point on the display can be moved with [▲] [▼] and confirmed with [P]. The
display then switches back to the menu level again.
Setting up the display time, SEC:
Default: 1.0
The display time is set with [▲] [▼]. The display moves up in increments of 0.1 up to 1 second
and in increments of 1.0 up to 10.0 seconds. Confirm the selection by pressing the [P] button.
The display then switches back to the menu level again.
Rescaling the measuring input values, EndA:
Default: 10000
With this function, you can rescale the input value of e.g. 19.5 mA (works setting) without
applying a measuring signal. If sensor calibration has been selected, these parameters are not
available.
then

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected] 15
13
5. Setting up the device
Menu level Parameterisation level
Rescaling the measuring input values, OFFA:
Default: 0
With this function, you can rescale the input value of e.g. 3.5 mA (works setting) without
applying a measuring signal. If sensor calibration has been selected, these parameters are not
available.
Setting up the tare value/offset value, tArA:
Default: 0
The given value is added to the linearized value. In this way, the characteristic line can be
shifted by the selected amount.
Setting up the balance point, Adj.pt:
Default: 08000
The balance point for the final value can be chosen from the measuring range by Sens.u with
0…10 V or Sens.A with 0…20 mA in %. The preset 80.000% result from the widespread
detuning of the melt pressure sensors.
Setting up the physical unit, UnIt:
Default: no
One can choose between the above shown physical units. It will be displayed on the 5th digit of
the display.
Number of additional setpoints, SPCt:
Default: 00
30 additional setpoints can be defined to the initial- and final value, so linear sensor values are
not linearised. Only activated setpoint parameters are displayed.
Display values for setpoints, dIS.01 … dIS.30:
Under this parameter setpoints are defined according to their value. At the sensor calibration,
like at final value / offset, one is asked at the end if a calibration shall be activated.

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected]
16
14
5. Setting up the device
Menu level Parameterisation level
Analog values for setpoints, InP.01 … InP.30:
The setpoints are always set according to the selected input signal. The desired analog values
can be freely parameterised in ascending order.
Device undercut, dI.Und:
Default: -i9999
With this function the device undercut (_ _ _ _ _) can be defined on a definite value. Exception is
input type 4-20 mA, it already shows undercut at a signal <1 mA, so a sensor failure is marked.
Display overflow, dI.OUE:
Default: 99999
With this function the display overflow (_ _ _ _ _) can be defined on a definite value.
Input variable of process value, SIG.in:
Default: a.meas
With this parameter, the device can be controlled via the analog input signals a.meas = 0-20 mA,
4-20 mA or 0-10 VDC or via the digital signals of the interface M.bus = RS232/RS485 (Modbus
protocol). With [P] the selection is confirmed and the device changes into menu level.
Back to menu group level, rEt:
With [P] the selection is confirmed and the device changes into menu group level „–INP–“.

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected] 17
5. Setting up the device
15
5.4.2. General device parameters
Menu level Parameterisation level
Display time, DISEC:
Default: 01.0
The display is set up with [▲] [▼]. Thereby it switches up to 1 second in increments of 0.1
seconds and up to 10.0 seconds in increments of 1.0. With [P] the selection is confirmed and
the device changes into menu level.
Rounding of display values, round:
Default: 00001
This function is for instable display values, where the display value is changed in
increments of 1, 5, 10 or 50. This does not affect the resolution of the optional outputs. With [P]
the selection is confirmed and the device changes into menu level.
Arithmetic, ArItH:
Default: n0
With this function the calculated value, not the measuring value, is shown in the display. With
no, no calculation is deposited. With [P] the selection is confirmed and the device changes into
menu level.
Sliding average determination, AVG:
Default: 1.0
Here, the number of the meterings that need to be averaged is preset. The time of averaging
results of the product of measuring time SEC and the averaged metering AVG. With the selection
of AVG in the menu level DISPL, the result will be shown in the display and evaluated via the
alarms.
Menu group level
Menu level
then
Reciprocal Root
extraction
Square

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected]
18
5. Setting up the device
16
Menu level Parameterisation level
Zero point slowdown, ZErO:
Default: 00
At the zero point slowdown, a value range around the zero point can be preset, so the display
shows a zero. If e.g. a 10 is set, the display would show a zero in the value range from -10 to
+10; below continue with -11 and beyond with +11. The maximum adjustable range of value is
99.
Definite contstant value, const:
Default: 0
The constant value can be evaluated via the alarms or via the analog output, like the current
measurand. The decimal place cannot be changed for this value and is taken over by the
current measurand. Like this a setpoint generator can be realised via the analog output by this
value. Furthermore it can be used for calculating the difference. At this the constant value is
substracted from the current measurand and the difference is evaluated in the alerting or by the
analog output. Thus regulations can be displayed quite easily.
Minimum constant value, con.mi:
Default: -i9999
The minimum constant value is adjusted from the smallest to the highest digit with the
navigation keys [▲] [▼] and confirmed digit per digit with [P]. A minus sign can only be
adjusted on the highest digit. After the last digit the display changes back into menu level.
Maximum constant value, con.ma:
Default: 99999
The maximum constant value is adjusted from the smallest to the highest digit with the
navigation keys [▲] [▼] and confirmed digit per digit with [P]. A minus sign can only be
adjusted on the highest digit. After the last digit the display changes back into menu level.
Display, dISPL:
Default: actua
With this function the current measurand, min/max value, totaliser value or the process-
controlled Hold-value can be allocated to the display. With [P] the selection is confirmed and
the device changes into menu level.

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected] 19
5. Setting up the device
17
Menu level Parameterisation level
Brightness control, Light:
Default: 15
The brightness of the display can be adjusted in 16 levels from 00 = very dark to 15 = very
bright via this parameter or alternatively via the navigation keys from the outside. During the
start of the device the level that is deposited under this parameter will always be used, even
though the brightness has been changed via the navigation keys in the meantime.
Display flashing, FLASH:
Default: no
A display flashing can be added as additional alarm function either to single or to a combination
of off-limit condition. With no, no flashing is allocated.
Assignment (deposit) of key functions, tASt:
Default: no
For the operation mode, special functions can be deposited on the navigation keys [▲] [▼], in
particular this function is made for devices in housing size 48x24mm which do not have a 4th
key ([O]-key). If the min/max-memory was activated by EHtr, all measured min/max-values are
saved during operation and can be recalled via the navigation keys. The values get lost by
restart of the device. If the threshold value correction LI.12 or LI.34 is choosen, the values of the
threshold can be changed during operation without disturbing the operating procedure. With
tArA the device is tared to zero and saved permanently as offset. The device confirms the
correct taring by showing ooooo in the display. Set.tA switches into the offset value and can be
changed via the navigation keys [▲] [▼]. Via totAL the current value of the totaliser can be
displayed, after this the device changes back on the parameterised display value. If tot.rE is
deposited, the totaliser can be set back by pressing of the navigation keys [▲] [▼], the device
acknowledges this with ooooo in the display. The configuration of EHt.rE deletes the min/max-
memory. Under ActuA the measurand is shown, after this the display returns to the
parameterised display value. If AbS.UA (absolute value) was selected, the display shows the
value that has been measured since voltage connection, without consideration of a previous
taring. Via selection L1.1, L1.1-2, li.1-3, l1.1-4, in case of 8 switching points li.1-5…li.1-8, threshold
values can be addressed via the navigation keys; they can be changed digit per digit or taken
over by pushing the [P]-key. The adjustment is taken over directly, an excisting limit value
monitoring and the current measurement will not be influenced by this. If nois selected, the
navigation keys are without any function in the operation mode.

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected]
20
5. Setting up the device
18
Menu level Parameterisation level
Special function [O]-key, tASt.4:
Default: no
For the operation mode, special functions can be deposited on the [O]-key. This function is
activated by pressing the key. With tArA the device is set temporarily on a parameterised value.
The device acknowledges the correct taring with oo0oo in the display. Set.tA adds a defined
value on to the currently displayed value. Via totAL the current value of the totaliser can be
displayed, after this the device switches back on the parameterised display value. If tot.rE is
deposited, the totaliser can be set back by pressing the navigation keys [▲] [▼], the device
acknowledges this with ooooo in the display. EHt.rE deletes the min/max-memory. If HOLD has
been selected, the moment can be hold constant by pressing the [O]-key and is updated by
releasing the key. Advice: Hold is activated only, if HOLD was selected under parameter DISPL.
ActuA shows the measuring value, after this the device switches back on the parameterised
display value. The same goes for AVG, here the sliding average values will be displayed. A
sensor calibration is done by triggering of the digital input via se.cal, the flow diagram is shown
in chapter 9. The constant value const can be recalled via the digital input, or changed digit per
digit. At AL-1…AL-4 an output can be set and therewith e.g. a setpoint adjustment can be done. If
nois selected, the [O]-key is without any function in the operation mode.
Special function digital input, dIG.In:
Default: no
In operation mode, the above shown parameter can be laid on the optional digital input, too.
Function description see tASt.4.
Back to menu group level, rEt:
With [P] the selection is confirmed and the device changes into menu group level „–fct–“.
…
…
Table of contents
Other ACS contsys Measuring Instrument manuals

ACS contsys
ACS contsys DAL-311x3x0S User manual

ACS contsys
ACS contsys SAT User manual

ACS contsys
ACS contsys DAL-311-2 Series User manual

ACS contsys
ACS contsys DAP-311 0 0S Series User manual

ACS contsys
ACS contsys Sonicont USN4SP User manual

ACS contsys
ACS contsys DAL-311 5 0S Series User manual

ACS contsys
ACS contsys Flowcont UN User manual

ACS contsys
ACS contsys RCE 300 User manual