Aeroqual Series 300 User manual

Series 300, 305, 500 & 505
Monitor
User Guide

1
Quick Start Guide
1. Unpack monitor and check supplied components are correct
2. Insert Battery (see page 6)
3. Insert Sensor
4. When new, plug in the power adaptor and turn on your monitor (with
sensor head installed) for 24hrs. Note: The Ni-MH battery requires
15hrs from full discharge and Lithium 3 hrs. The sensor requires a run
in period. Thereafter, the normal warm up takes only 10 minutes. If
your unit is new or unused for a long period, a warm up message may
display for several hours, then will disappear.
5. Follow the operating instructions to understand how to make the
most of your monitor
CLOCK SETUP: To adjust the time, use the ar-
rows , pressing to alternate between hours, minutes, seconds, and the date. For more
information see page 13 of your user guide.
LOGGING SETUP: To adjust the logging frequency, use the arrows , then press to
return to the main menu. For more information see page 18 of your user guide.
ALARM POINTS: To adjust the alarm points, use the arrows , pressing to alternate
between alarm high, alarm low, and buzzer enable / disable. For more information see page 14 of your user
guide.
CONTROL POINTS: To adjust the control points, use the arrows , pressing to alter-
nate between control high and control low. For more information see page 17 of your user guide.
UNITS: To change the units of measurement, use the arrows , then press to return to
the main menu. See page 19 of your user guide for more information
MAX / MIN AV: To adjust the max / min av, press to alternate between start and stop. See page
25 of your user guide for more information
Warning:
Do not expose the monitor to gas concentrations above the recommended
maximum range
Do not place the sensor head in a gas stream –the monitor is designed to
measure ambient levels of ozone and uses „active sampling‟ to achieve this.
1. Sensor Head
2. Power and Stand By
3. Display
4. Alarm Mute
5. Enter
6. Scroll up, or hold down for data logging
(S500 only)
7. Scroll down, or hold down for max/min
average
8. Battery case. For battery installation, see
page 10 of your user guide

Contents
1. Monitor components 3
2. About your monitor 4
3. Assembly and installation 6
3.1 Installing the battery pack 6
3.2 Sensor warm up 9
3.3 Monitor and sensor versions 9
4. Operating instructions 10
4.1 Turning the monitor on and off 10
4.2 Turning standby mode on and off 11
4.3 Operating under battery power 11
4.4 Sensor head failure 11
4.5 Using the menus and key functions 12
4.6 Calibration 12
4.61 Zero calibration 12
4.62 Span calibration 12
4.7 Setting the clock 13
4.8 Setting the high and low alarms 14
4.9 Setting the control points 17
4.10 Setting up data logging time intervals 18
4.11 Changing the units of measurement 19
4.12 The max / min / average measurement cycle 19
4.13 Starting and stopping data logging 20
5.0 Aeroqual monitor software instructions 21
5.1 Setting up the software and monitor parameters 23
5.2 Gas Sensor Properties Configuration 24
5.3 Communications Port Configuration 25
5.4 Setting the Graph Default Style 26
5.5 Setting Data Logging Functions 27
5.6 Datalogging directly to a PC 28
5.7 Advanced Data Management 35
6.0 External control wiring 37
7.0 Troubleshooting 40
8.0 Diagnostic features 42
9.0 Care and maintenance 43
10.Specifications 44
11. Appendix 46
11.1 Using the RS232 serial to USB adaptor 46
11.2 Installing the R52 drivers 46
11.3 Selecting the COM port 47
11.4 Copyright 50
11.5 Warranty 50
11.6 Terms and conditions 51
11.7 Statements of compliance 52
11.8 Guidelines on how to measure ozone 53
11.9 Zero calibration additional information 54
11.10 Accessories 56
11.11.2011 V2.1

3
1. Monitor Components
Series 300 and 305 Monitor
The following components are supplied with the Series 300 and 305
Monitor:
Series 300 or 305* monitor base
Sensor head
12 VDC AC/DC adaptor or Lithium Smart Charger
User guide
Battery pack –9.6V Ni-MH or 11.1V Lithium Polymer
Series 500 and 505 Monitor
The following components are supplied with the Series 500 and 505
Monitor:
Series 500 or 505* monitor base
Sensor head
Data logging software on CD
Serial cable and USB to Serial adapter (with instructions)
12 VDC AC/DC adaptor or Lithium Smart Charger
User guide
Battery pack –9.6V Ni-MH or 11.1V Lithium Polymer
* Series 305 and 505 monitor base is supplied with a separate temperature and
relative humidity probe that is plugged in as shown below.
Please check that all these components have been supplied and contact
your dealer or Aeroqual on email at: sales@aeroqual.com if any of the com-
ponents are missing.
Replacements
Replacement batteries, sensor heads and other accessories can be pur-
chased separately.
Note:
All sensors deteriorate over time. It will be necessary to replace your sensor
head with either a new or refurbished head. This must be done on a sched-
uled basis or when the “sensor aging” or sensor failure” diagnostic appears.
Bump tests, zero and span calibration can be performed under specified
conditions with appropriate tools.

2. About Your Monitor
The Aeroqual Series 300, 305, 500 and 505 Monitors have been spe-
cifically designed to incorporate Aeroqual‟s in-depth knowledge of accu-
rate ambient gas measurement. Different sensor heads are used for
specific gases or depending upon whether high or low concentrations
are to be measured. All the sensor heads are interchangeable on the
same base unit.
The following information details the operation and features of the moni-
tors:
1. The monitors are modular in design and are comprised of two
main components, a Series 300, 305, 500 or 505 monitor, and a
sensor head. This affords the user the ability to replace faulty sen-
sor heads without having to replace the monitor. The monitors can
be mains and/or battery powered and either permanently mounted
or used as portable units.
2. The sensor head is calibrated prior to delivery and does not nor-
mally need to be re-calibrated during its life. (Specific user
guides are supplied for the special operations of particular sensor
heads).
3. The Aeroqual Series 300, 305, 500 and 505 Monitors come with
on-board diagnostics, which will inform the user if the sensor is
not operating correctly.
4. High and low alarms. The high alarm is indicated by flashing
arrows pointing upwards accompanied by a fast beeping sound.
Similarly the low alarm is indicated by flashing arrows pointing
downwards accompanied by a slow beeping sound. This sets
both the monitor and external alarms.
5. Alarm mute. Two alarm mute functions are available, temporary
and permanent mute. Temporary mute is only active during the
specific alarm condition, while permanent mute will remain on until
reset by the user. The alarm arrow indications remain active dur-
ing both temporary and permanent mute.
6. Control. This enables external equipment to be controlled by the
monitor using the high and low control settings, which can be
used to control an external device.
7. Concentration unit selection. The concentration can be dis-
played in either ppm or mg/m3.
8. Min / max / average measurement cycle. The monitor will dis-
play the minimum, maximum, average and 15-minute average
gas concentrations over the measurement period. The Series 305
& 505 will also show temperature and relative humidity in addition
to gas concentration.
9. 0-5V Analogue signal. A proportional voltage output is available
from the external output. The voltage relates to the gas concen-
tration, with 0 V representing 0 concentration and 5 V represent-

5
ing the maximum concentration specified by the range. For ex-
ample, 0 V represents 0 ppm and 5 V representing 0.500 ppm for
the Low Concentration Ozone Sensor Head.
Note: See “Diagnostic Features” on page 42 for output status condi-
tions.
The Series 500 & 505 Monitors offer the following additional fea-
tures:
1. Data logging (on either the Series 500 or 505 Monitor or directly
to a PC). The monitor is able to log up to 8000 data points. The
data can be downloaded to a PC for analysis and record pur-
poses. Data can also be logged directly to a PC via a serial con-
nection while displaying real time data on the PC. The data log-
ging interval can be set in one minute increments from either the
monitor or the PC.
The Series 505 monitor also measures and data logs tem-
perature and relative humidity.
2. Real time clock (24 hr). The clock has a back up clock battery
which will maintain the real time even when all power is lost to
the monitor. Once power is restored data logging will continue if t
he unit was data logging at the time power was lost.
3. PC setup of monitor parameters. All the monitor parameters,
i.e. clock setup, alarm setup, external control setup, logging
setup and concentration units can be set up from the PC.

3.0 Assembly and Installation
Assembly
The following needs to be completed before the Monitor is ready
for use: -
Insert the sensor head into the top of the monitor. The sen-
sor head is keyed to ensure the head is inserted correctly.
If the monitor is a mains only unit it is ready for use. If the
monitor is to be used as a portable unit with batteries then
the batteries must be installed and charged prior to use.
3.1 Installing the Battery Pack
Remove the battery cover from
the monitor.
Insert the battery power connector
into the monitor power socket in
the back of the monitor.
Lay the battery cable into the slot
on
the back of the monitor and insert
the tail end of the battery pack into
the hinge points at the bottom of
the monitor back cover.
Clip the top end of the battery pack
into the top clips, taking note to
orientate the battery pack per the
instructions on the battery pack.
Replace the battery cover on the
monitor.
Notes on the battery pack
For more information on the correct use of the Ni-MH battery
pack, please refer to ‘Battery Information’ on Page 8.

7
Removing the Battery Pack
Remove the battery cover from the monitor.
Insert your thumb under the top left hand side of the battery
pack and carefully lever the battery pack out from the retain-
ing clips.
Unclip the connector from the monitor.
Replace the battery cover.
Rechargeable Batteries
The Ni-MH (Nickel Metal Hydride) or optional Lithium Polymer bat-
tery pack is used to power the monitor as a portable unit. Please
read the following information carefully before using the battery pack
to operate the Aeroqual monitor.
Safety
Do not use the battery pack for any other purpose than oper-
ating the Aeroqual monitor
Do not disassemble or mutilate the batteries as this may
cause burns
Do not incinerate or heat as this may cause burns, the batter-
ies may burst or cause the release of toxic materials
Do not short circuit as this may cause burns
Use only batteries approved by Aeroqual and recharge your
battery only with the AC/DC adaptors supplied by Aeroqual.
Never use any charger or battery that is damaged or worn
out.
Batteries must be recycled or disposed of properly. They
must not be disposed of in municipal waste.
Never charge the battery in a hermitically sealed container.
Using the battery pack
The life of the battery pack will be affected by the way it is handled.
Take good care of the battery pack and follow the guidelines sug-
gested in the following section.

Charging and Discharging the Batteries
Ni-MH batteries
Ni-MH batteries must be fully charged prior to first-time use. A new Ni-MH
battery's full performance is achieved only after a number of complete
charge and discharge cycles. A fully charged battery will run for approxi-
mately 4-6 hours depending on the sensor head being used.
Charging Instructions
With the adaptor plugged in allow 15 hours for the battery to fully charge.
Note 1:
During charging the battery symbol on the display flashes slowly.
The battery can be charged with the sensor head installed or removed
from the base. (Remember to switch off the monitor before removing
the sensor head).
Unplugging the main power will reset the above charging conditions.
The batteries can be charged and discharged many times but will even-
tually wear out. When the operating time is noticeably shorter than nor-
mal, it is time to replace the batteries.
It is not necessary to fully discharge the batteries. When not using the
Series 300, 305, 500 or 505 Monitor as a portable unit, always plug in
the mains adaptor to recharge the batteries.
If left unused a fully charged battery will discharge itself over time.
Note 2:
If you have difficulty charging the battery, either:
A) Replace the battery
B) Contact Aeroqual with the software version number
(see section 3.3 to locate the version number.)
Additional Information for Lithium Polymer Batteries
A new battery's full performance is achieved only after a number of com-
plete charge and discharge cycles. A fully charged battery will run for ap-
proximately 10 hours depending on the sensor head being used.
Charging Instructions
With the adaptor plugged in allow 3 hours for the battery to fully charge.
There is no battery symbol on the display when using a lithium battery.
The battery can be charged with the sensor head installed or removed
from the base. (Remember to switch off the monitor before removing
the sensor head).
The Lithium Smart Charger displays a red LED light when charging
and a green LED light when the battery is charged.

9
Installing the AC/DC adaptor
Connect the AC/DC adaptor to the mains supply. Ensure that
the adaptor is correctly rated for the power supply.
Plug the adaptor into the monitor. The monitor can now be
used as a mains unit. If the monitor is to be used as a port-
able unit the batteries will need to be charged.
3.2 Sensor warm up
Prior to operation the monitor must be warmed up to burn off
contaminants on the sensor. When the monitor is first switched
on it will warm up for 3 minutes. Then the reading will flash for
the next 7 minutes to indicate that the sensor is still in the warm
up phase. It is recommended that the monitor is kept in Stand By
mode when not being used to keep the sensor heated and
prevent the build up of contaminants. If the sensor is new (or
unused for long periods) it can take up to 24 hours for the
message “sensor warming up” (or “sensor failure” in older
models) to disappear. Should the message remain after this time,
a genuine sensor failure may have occurred. See the trouble-
shooting guide.
24hrs is a conservative time for an ozone sensor to burn off con-
taminants and achieve its stated performance specification. In
most instances this will be achieved well within 24hrs.
Since O3 is an oxidising gas the warm up period can be reduced
by subjecting the sensor to elevated O3 levels for a short period.
For example, 100 ppb of O3 for 5 minutes typically reduces the
warm up time to 30 minutes.
Warning: Do not remove the sensor head while the monitor
is switched on. This may damage the unit. If the sensor head is
removed under these conditions, the unit will automatically shut
down.
3.3 Monitor and Sensor Versions
Turning the monitor on will display “AEROQUAL MONITOR
S300 / S305 / S500 / S505 V x.x”, where "V x.x" represents
the version of the monitor base. The display will then show
“SV6.0”or similar, which indicates the software version installed
on the sensor head

4. Operating Instructions
The following instructions detail the operation and set up of the
monitor.
4.1 Turning the monitor on and off
To turn the monitor on, press & hold the power button until
the screen activates. The monitor will turn on and the display
will show “AEROQUAL MONITOR S300 / S305 / S500 / S505
Vx.x”,. After 5 seconds the display will change to indicate the
type of sensor head. The monitor will warm up as described
in “Warm up” in section 3.2. Once warm up is complete the
main display will appear (for example):
Series 300 & 500 Series 305 & 505
To turn the monitor off, press and hold the power button for
2 seconds until the monitor beeps. The display will now be
blank and operation of the monitor will cease.
Power
button
Alarm
mute
O3 LOW
0.009
RH 35.5%
TEMP 21.6C
O3 LOW
0.009
Enter
Scroll
up
Scroll
down

11
4.2 Turning Standby Mode on and off
To activate Standby Mode with the monitor on, press the power
button once. This will stop operation of the monitor; however the
sensor will be kept warm. This mode is used to conserve power
between operations but keep the sensor ready to measure the
gas concentration with only a short warm up phase as described
in “Warm up” in section 3.2. In this mode the display will be as
shown below:
To return the monitor to operational mode press the power but-
ton once.
NOTE: When using a Carbon Dioxide sensor head in
standby mode, the sensor runs at full power and the fan
remains on.
4.3 Operating under Battery power
A battery indicator will appear at the bottom of the display when
the monitor is operating under battery power only. The indicator
shows the battery life remaining.
4.4 Sensor Head Failure
There are two possible Sensor failure modes, which will be
indicated by one of the following messages on the display: -
“Sensor Failure, Replace sensor” –The sensor head should
be replaced as this indicates that there has been component
failure.
“Sensor aging”:- This indicates that the sensor has reached
the end of its usable life and the sensor head should be replaced
as soon as possible. The measurement readings can no longer
be relied upon to be within specification.
NH3
O3 LOW
0.080 ppm

4.5 Using the Menus and Key Functions
To enter the setup menu press The following menus will appear: -
. Series 300 & 305 Menu
Series 500 & 505 Menu (V5.5 and above)
Press or to scroll to the desired menu.
4.6 Calibration
To access the calibration menu, scroll down and hold and
simultaneously for approximately 2 seconds. Once the calibration menu
has been entered, users can select „ZERO CAL‟ or „SPAN CAL‟.
4.61 Zero Calibration
The “ZERO CAL” routine resets the zero point of the monitor.
This routine must be conducted under controlled conditions:
Minimum 24 hours warm up
Clean air (charcoal filtered is best), no cross-sensitive gases
Stable and low air flow around the monitor
No vibration or movement of the monitor
Temperature at 20°C ±2°C (68°F ±3.5°F)
Relative Humidity of 50% ±5%
An incorrect "ZERO CAL" does not cause permanent damage.
The process should be repeated under the above conditions.
NOTE: Carbon Dioxide sensors MUST be calibrated in a certi-
fied Zero Air environment. If in doubt contact Aeroqual Limited.
Press and hold until the word GO appears next to Reset
Routine. The routine will run for up to ten minutes (depending on
the gas sensor installed) and then beep to indicate completion.
Talk to your supplier or distributor if you have questions.
See section 11.9 for more information
4.62 Span Calibration
Span calibration, located in the ‟CALIBRATION‟ menu, provides an
option to adjust the span of Aeroqual sensor heads. Once entered,
users can adjust the gain factor of their sensor.
EXIT
ALARM POINTS
CONTROL POINTS
UNITS
MAX MIN AV
ZERO CAL
EXIT
CLOCK SETUP
LOGGING SETUP
ALARM POINTS
CONTROL POINTS
UNITS
MAX MIN AV
CALIBRATE

13
4.7 Setting the Clock *Series 500 & 505 Monitor
Only
To enter Clock Setup press The following display will
appear:
(Hours: Minutes: Seconds)
(Month Day Year)
Changing Clock Hours
To change Hours, press or
Press to confirm hours. Note: the clock is a 24-hour
clock.
Once the change is confirmed the cursor will automatically
scroll to the next parameter.
Changing Clock Minutes
To change Minutes press or
Press to confirm minutes.
Changing Clock Seconds
To change Seconds press or
Press to confirm minutes.
Changing the Month
To change the Month press or
Press to confirm the month.
Changing the Day
To change the Day press or
Press to confirm the day.
14:30:12
Oct 07 02

Changing the Year
To change the Year press or
Press to confirm the year.
Press to scroll down to the next Menu choice.
4.8 Setting High and Low Alarms
To enter the Alarm Setup menu press .
Setting the High Alarm
After entering the Alarm Setup Menu, the „Alarm Hi‟ will display.
Press or to increase or de-
crease the setting.
Press to confirm change.
Setting the Low Alarm
Once the High Alarm setting has been confirmed, the cursor will
automatically scroll to the „Alarm Lo‟ display.
Press or to increase or decrease setting
Press to confirm change.
Note: This sets both the monitor alarm and the external alarm.
ALARM HI
35 ppm
ALARM LO
0.040 ppm

15
BUZZER
ENABLE (or)
DISABLE
Enabling or Disabling the Alarm
Once the Low Alarm setting has been confirmed, the cursor will
automatically scroll to the "Buzzer" display.
Press or to scroll between Enable or Disable.
Press to confirm choice.
Note: If the Disable option is selected, the Alarm will be per-
manently muted. A „P‟ on the right hand side of the Alarm
Mute symbol will be displayed to indicate Permanent Mute.
Screen Display of High Alarm Condition
Two flashing arrow symbols pointing upwards will appear on the
left hand side of the display, accompanied by a fast beeping
sound when the Ozone concentration rises above 8.00 ppm.
Screen Display of Low Alarm Condition
Two flashing arrow symbols pointing downwards will appear on
the left hand side of the display, accompanied by a slow beeping
sound when the Ozone concentration rises above 0.015 ppm.
O3 HIGH
8.00
ppm
O3 LOW
0.015
ppm

Screen Display of Alarm Mute
This is activated by pressing the alarm mute button, .
Muting the high and low alarms using this button only mutes the
alarm during that specific alarm condition. Once the concentra-
tion returns to within alarm limits, the alarm will reset.
A speaker symbol with a cross through the middle of it will ap-
pear on the right hand side of the display when an alarm condi-
tion exists and has been muted from the key pad. Once the
alarm condition is removed and the alarm has reset the symbol
will disappear from the screen.
Screen Display of Permanent Alarm Mute
The Alarm mute symbol, as above, will be permanently displayed
on the screen until the alarm is once again re-enabled via the
Alarm Setup Menu.
If an alarm condition arises when permanent mute has been acti-
vated, flashing arrows will appear on the display as described in
the previous sections, however no sound will accompany the
flashing arrows.
Press to scroll down to the next Menu choice.
CO
35
CO
35

17
4.9 Setting the Control Points
To enter the Control menu press .
Setting the Control High
After entering the Control Points menu, the „Control Hi‟display
will appear.
Press or to increase or decrease the setting
Press to confirm change.
Setting the Control Low
Once the change has been confirmed, the cursor will automati-
cally scroll to the „Control Lo‟display.
Press or to increase or decrease the setting
Press to confirm the change.
Press to scroll down to the next Menu choice
Note: Once the control settings have been set the monitor must
be either put into “Standby” or switched off for the new settings to
become effective. Once the monitor is put back into normal oper-
ating mode the new settings will become effective.
CONTROL HI
0.080 ppm
CONTROL LO
0.040 ppm

4.10 Setting up Data Logging Time Intervals
*500&505 Series Only
To enter the Logging Setup menu press . The Log Freq
display will appear.
Press or to change the logging frequency in one-
minute increments
Press to confirm change and following screen will be dis-
played.
You can now elect to clear any logged data by selecting YES or
NO if you wish to retain the data.
Press to toggle between NO and YES on the CLEAR LOG
screen.
Press to confirm your selection and
fol- lowing screen will be displayed.
Press to toggle between OFF and ON on the LOGGING
screen.
Press to confirm your selection and you will be returned to
the main menu.
Press to scroll down to the next Menu choice
LOG FREQ:
1 MIN
CLEAR LOG
NO
LOGGING
OFF

19
4.11 Changing Units of Measurement
To enter the „UNITS‟ menu press once for the concentra-
tion unit and again for temperature unit –see displays below.
S305 &
S505 only
Press or to scroll between „ppm‟ &„mg/m3‟ or „ºC & ºF‟
Press to confirm the unit selection
Press to scroll down to the next Menu choice
4.12 The Max/Min/Average Measurement Cycle
Pressing scrolls between Start and Stop
Press to scroll to „EXIT‟
Press to Exit out of Menu Setup. This will also Start or
Stop the measurement cycle, depending on your menu selection.
Note: There is a quick key option to Start or Stop the „Max Min
Average‟ measurement cycle, as follows:
From the Display Screen
Press and hold for two seconds until the monitor beeps.
This will start the measurement cycle, and the following display
will appear: -
S300 and S500 S305 and S505
CONC UNIT:
ppm (or)
mg/m3
TEMP UNIT:
ºC (or)
ºF
O3 LOW
0.024
O3 LOW
MIN 0.005
MAX 0.024
AVE 0.015
ST 0.013
O3 LOW TEM RH
PPM C %
MIN 0.005 22 50
MAX 0.024 25 56
AVE 0.015 24 54
ST 0.013
Other manuals for Series 300
1
This manual suits for next models
1
Table of contents
Popular Laboratory Equipment manuals by other brands

Belden
Belden HIRSCHMANN RPI-P1-4PoE installation manual

Koehler
Koehler K1223 Series Operation and instruction manual

Globe Scientific
Globe Scientific GCM-12 quick start guide

Getinge
Getinge 86 SERIES Technical manual

CORNING
CORNING Everon 6000 user manual

Biocomp
Biocomp GRADIENT MASTER 108 operating manual