GCM-12 / High Speed
Microcentrifuge
Quick Start Guide
For the full User Manual, please visit www.globescientic.com
CONTROL PANEL
Speed Display Time Display
1
2
4
3
5
IMPORTANT: Plug unit into appropriate main supply based
on the rated voltage.
Select� � � � � � � � � � � � � Press to choose the parameter you wish
to modify.
Pulse � � � � � � � � � � � � � Press to accelerate and hold at the set
speed. The centrifuge stops immediately
once this button is released.
Adjustment Knob Rotate clockwise to increase parameter
values. Rotate counterclockwise to
decrease parameter values. Press to
switch between speed and RCF. Radius
Door� � � � � � � � � � � � � � Press to open the lid. This button is not
enabled while the centrifuge is running.
Run / Stop � � � � � � � � Press to start the centrifuge. Press to
stop if centrifuge is already running.
OPERATION
1
2
3
4
5
The centrifuge rotor can separate samples whose density is lower than 2.0g/ml. If the separated sample’s density is over 2.0g/ml, please
calculate the allowable speed using the following formula:
Allowable Speed (rpm) = Maximum Speed × (2�0(g/ml) / Sample Density (g/ml))1/2
Rotor Tube Adapter Maximum Maximum
Allowable Imbalance
Type Type Required Speed (rpm) RCF (×g) Imbalance Volume Imbalance
1.5/2.0mL Tube — 15000 15100 2.0g/Tube 5mm/Tube
0.2mL PCR Tube MC-12-AD2 15000 11700
0.5mL PCR tube MC-12-AD5 15000 12780
GCM-12-R