Radleys Carousel 6 Plus User manual

Instructions
Cooled Carousel 6 Plus™
Reaction Station
Your Local Distributor
February 2023 | Issue 5
Radleys, Shire Hill, Saffron Walden, Essex, CB11 3AZ. United Kingdom.
Tel: +44-1799-513320 Fax: +44-1799-513283
These instructions should be read in conjunction
with the Carousel 6 Plus and Stirring Hotplate Instructions.

Page 1
Cooled Carousel 6 Plus Reaction Station
Index
Page 1 Index
Page 2 1. Introduction
Page 3 2. Warranty
3. Safety Information
4. Important WARNINGS
Page 4 5. Product Component Guide
Page 5 6. Products & Accessories
Page 6 6. Products & Accessories - Continued
Page 7 6. Products & Accessories - Continued
Page 8 6. Products & Accessories - Continued
Page 9 7. Glassware
Page 10 7. Glassware - Continued
Page 11 8. Quick Start Guide
Page 12 9. Set-Up & Operation
Page 13 9. Set-Up & Operation - Continued
Page 14 9. Set-Up & Operation - Continued
Page 15 9. Set-Up & Operation - Continued
Page 16 9. Set-Up & Operation - Continued
Page 17 9. Set-Up & Operation - Continued
Page 18 10. Safe handling of cooling fluids
11. Cooling Mixture Appendix
Page 19 12. Radleys/Genevac Compatibility Chart
Page 20 WARRANTY - EMAIL BACK

Page 2
Cooled Carousel 6 Plus Reaction Station
1. Introduction
Thank you for purchasing your Cooled Carousel 6 Plus Reaction Station.
Please read this Instruction Manual thoroughly before operating your unit.
The Cooled Carousel 6 allows chemists to perform stirred, sub-ambient reactions in a range of
round bottom flasks from 5ml to 250ml with the option of an inert, moisture free atmosphere.
Features...
• Simultaneously performs up to six cooled and stirred reactions to -78˚C.
• Powerful, even stirring - reservoir fits on to a Carousel stirring hotplate.
• Robust HDPE cooling reservoir is compatible with a wide range of cooling mixtures, including
dry-ice/acetone for manually controlled cooling from ambient down to -78˚C.
• Insulated foam core maintains low temperatures for long periods, whilst protecting the stirrer
from freezing.
• Insulation also reduces condensation and ice formation on outer surfaces of reservoir.
• HDPE lid keeps your reaction cooler for longer, minimises ice formation on your flasks (maintaining visibility of the contents) and prevents
spitting from the cooling mixture.
• Round design makes all reaction flasks visible and allows easy addition of reagents and solvents, with no need to lean into the fumehood.
Whats NEW...
• Easy-on PTFE caps feature a ‘quick-thread’ for a gas tight seal to flasks.
• Easy to clean contoured surface.
• Exclusive wide neck flasks with quick coupling for improved sealing and ease of use.

Page 3
• Please read these instructions completely before using your Cooled Carousel 6 Plus Reaction Station.
• The Cooled Carousel is NOT suitable for heated reactions.
• Operate only in a fumehood with protective safety sash.
• The Carousel Reaction Flasks are not designed for pressurised reactions. DO NOT PRESSURISE ABOVE 1psi.
• Reactor Head - Chemical Resistance - the aluminium Carousel reactor head is resistant to the majority of solvents and splash resistant
to dilute acids and alkalis at room temperature. Extended exposure to acids or alkalis will attack the surface of the Carousel. The stronger
the concentration and the longer the exposure time the more chance and intensity of any attack. It is important to clean off any residual
chemical spills immediately after they occur.
• Cooling Reservoir - Chemical Resistance - The cooling reservoir is manufactured from a HDPE polymer which is compatible with the
freezing mixtures listed in Section 11. Cooling Mixture Appendix on page 18 of this manual.
• Ensure that the heating control on the stirring hotplate is switched off and that the hotplate is cool before using the Cooled Carousel.
• Cooling solvents such as acetone are extremely flammable. Always follow appropriate safety instructions and chemical handling
procedures.
Page 3
Warning
Applies when there is a possibility
of personal injury.
Important Note
Alerts the user to important facts.
!
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important Note
The Cooled Carousel should only be operated by trained and competent personnel. As with all chemistries, a full risk
assessment should be performed prior to starting an experiment, and care should be taken to monitor your reactions at all
stages. The Carousel should not be left unattended unless in a supervised area.
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
3. Safety Information
4. Important WARNINGS
The Cooled Carousel 6 Plus Reaction Station includes one year full parts and labour warranty from date of original purchase.
Warranty will only be valid if a completed Warranty Fax Back is returned within 1 month of date of purchase (see last page). In the event of
product failure please contact your local distributor.
Please do not return any goods without prior agreement.
2. Warranty
Cooled Carousel 6 Plus Reaction Station
The following symbols are intended to assist the user in the safe and efficient operation of the Cooled Carousel 6 Plus Reaction Station.

Page 4
Cooled Carousel 6 Plus Reaction Station
Easy-On PTFE Caps & Septa
Reactor Head
HDPE Reservoir
Central Gas Inlet
Stirring Hotplate
HDPE Stand
Easy-On Cap
Reflux Tube
Rodaviss
Connecting Set
Flask
Easy-On Cap
Flask with integrated
Reflux Tube
Easy-On Cap
Wide Neck Reflux Tube
Wide Neck Coupling
Wide Neck Flask
Stirring Bars
HDPE Lid
5. Product Component Guide
Hole for
temperature probe
Connector Pins

Page 5
Cooled Carousel 6 Plus Reaction Station
6. Products & Accessories
Heated System - Basic
System 34
RR99943 Carousel 6 Plus System
1 x RR99916 Carousel 6 Plus Reaction Station
1 x RR91070 Easy-On PTFE Cap, pk 6
1 x RR98076 Silicone Septa for PTFE Caps, pk 100
1 x RR98094 PTFE Magnetic Stirring Bar Retriever 350mm
1 x RR99917 250ml Reaction Flask + Reflux Tube + Connecting Set, pk 6
1 x RR99043 B24/29 Rodaviss Sealing Cap, pk 10
1 x RR99046 Rotary Evaporator Adapters Ordinary B29 Socket to B24 Cone, pk 2
1 x RR99061 Reaction Flask Support Ring, pk 6
1 x RR99064 Elliptical Stirring Bar 25mm RE, pk 10
1 x RR99067 Tubing for Inert Gas or Reflux Cooling 2 metres
Heated System - Additions
System 35
RR99944 Carousel 6 Plus System
1 x RR99916 Carousel 6 Plus Reaction Station
1 x RR91070 Easy-On PTFE Cap, pk 6
1 x RR91202 Carousel Tech Stirring Hotplate 230v UK Plug
1 x RR91210 Carousel Temperature Controller
1 x RR98076 Silicone Septa for PTFE Caps, pk 100
1 x RR98094 PTFE Magnetic Stirring Bar Retriever 350mm
1 x RR99917 250ml Reaction Flask + Reflux Tube + Connecting Set, pk 6
1 x RR99043 B24/29 Rodaviss Sealing Cap, pk 10
1 x RR99046 Rotary Evaporator Adapters Ordinary B29 Socket to B24 Cone, pk 2
1 x RR99061 Reaction Flask Support Ring, pk 6
1 x RR99064 Elliptical Stirring Bar 25mm RE, pk 10
1 x RR99067 Tubing for Inert Gas or Reflux Cooling 2 metres
1 x RR99077 B24/29 250ml Reaction Flasks with Sidearm + Septa Port, pk 6
1 x RR99078 B24/29 Liquid Additions Dropping Funnel with Pressure Equalising Arm, pk 3
1 x RR99078 B24/29 Liquid Additions Dropping Funnel with Pressure Equalising Arm, pk 3
1 x RR99079 B24/29 Solid Additions Funnel, pk 3
1 x RR99082 PTFE/Silicone Septa 29mm for B24/29, pk 50
Heated System - Basic + Hotplate
System 33
RR99942 Carousel 6 Plus System
1 x RR99916 Carousel 6 Plus Reaction Station
1 x RR91070 Easy-On PTFE Cap, pk 6
1 x RR91202 Carousel Tech Stirring Hotplate 230v UK Plug
1 x RR91210 Carousel Temperature Controller
1 x RR98076 Silicone Septa for PTFE Caps, pk 100
1 x RR98094 PTFE Magnetic Stirring Bar Retriever 350mm
1 x RR99917 250ml Reaction Flask + Reflux Tube + Connecting Set, pk 6
1 x RR99043 B24/29 Rodaviss Sealing Cap, pk 10
1 x RR99046 Rotary Evaporator Adapters Ordinary B29 Socket to B24 Cone, pk 2
1 x RR99061 Reaction Flask Support Ring, pk 6
1 x RR99064 Elliptical Stirring Bar 25mm RE, pk 10
1 x RR99067 Tubing for Inert Gas or Reflux Cooling 2 metres
180˚C
180˚C
180˚C

Page 6
Carousel 6 Plus Reaction Station
6. Products & Accessories
Cooled System - Comprehensive
System 39
RR99950 Cooled Carousel 6 Plus System
1 x RR99947 Cooled Carousel 6 Plus (Reservoir, Head, Stand + Cover)
1 x RR98024 Protective Cold Temperature Gloves
1 x RR91070 Easy-On PTFE Cap, pk 6
1 x RR91202 Carousel Tech Stirring Hotplate 230v UK Plug
1 x RR98076 Silicone Septa for PTFE Caps, pk 100
1 x RR98094 PTFE Magnetic Stirring Bar Retriever 350mm
1 x RR99917 250ml Reaction Flask + Reflux Tube + Connecting Set, pk 6
1 x RR99043 B24/29 Rodaviss Sealing Cap, pk 10
1 x RR99046 Rotary Evaporator Adapters Ordinary B29 Socket to B24 Cone, pk 2
1 x RR99054 100ml Reaction Flask, pk 6
1 x RR99061 Reaction Flask Support Ring, pk 6
1 x RR99064 Elliptical Stirring Bar 25mm RE, pk 10
1 x RR99067 Tubing for Inert Gas or Reflux Cooling 2 metres
1 x RR99077 B24/29 250ml Reaction Flasks with Sidearm + Septa Port, pk 6
1 x RR99078 B24/29 Liquid Additions Dropping Funnel with Pressure Equalising Arm, pk 3
1 x RR99078 B24/29 Liquid Additions Dropping Funnel with Pressure Equalising Arm, pk 3
1 x RR99079 B24/29 Solid Additions Funnel
1 x RR99905 Digital Thermometer (-250˚C to +400˚C) + 200mm Probe
1 x RR99908 Dry Ice Scoop
1 x RR99909 Cold Temp. Apron 1060mm long Waterproof
1 x RR99910 Protective Face Shield
-78˚C
Cooled System - Upgrade
System 38
RR99949 Cooled Carousel 6 Plus System
1 x RR99947 Cooled Carousel 6 Plus (Reservoir, Head, Stand + Cover)
1 x RR98024 Protective Cold Temperature Gloves
1 x RR98076 Silicone Septa for PTFE Caps, pk 100
1 x RR99067 Tubing for Inert Gas or Reflux Cooling 2 metres
1 x RR99905 Digital Thermometer (-250˚C to +400˚C) + 200mm Probe
1 x RR99908 Dry Ice Scoop
1 x RR99909 Cold Temp. Apron 1060mm long Waterproof
1 x RR99910 Protective Face Shield
Heated System - Comprehensive
System 36
RR99945 Carousel 6 Plus System
1 x RR99916 Carousel 6 Plus Reaction Station
1 x RR91070 Easy-On PTFE Cap, pk 6
1 x RR91202 Carousel Tech Stirring Hotplate 230v UK Plug
1 x RR91210 Carousel Temperature Controller
1 x RR98076 Silicone Septa for PTFE Caps, pk 100
1 x RR98094 PTFE Magnetic Stirring Bar Retriever 350mm
1 x RR99917 250ml Reaction Flask + Reflux Tube + Connecting Set, pk 6
1 x RR99043 B24/29 Rodaviss Sealing Cap, pk 10
1 x RR99046 Rotary Evaporator Adapters Ordinary B29 Socket to B24 Cone, pk 2
1 x RR99054 100ml Reaction Flask, pk 6
1 x RR99920 250ml Azeotropic Reaction Flask, pk 3
1 x RR99920 250ml Azeotropic Reaction Flask, pk 3
1 x RR99058 Aluminium Insert for 100ml Reaction Flask, pk 6
1 x RR99061 Reaction Flask Support Ring, pk 6
1 x RR99064 Elliptical Stirring Bar 25mm RE, pk 10
1 x RR99067 Tubing for Inert Gas or Reflux Cooling 2 metres
1 x RR99077 B24/29 250ml Reaction Flasks with Sidearm + Septa Port, pk 6
1 x RR99078 B24/29 Liquid Additions Dropping Funnel with Pressure Equalising Arm, pk 3
1 x RR99078 B24/29 Liquid Additions Dropping Funnel with Pressure Equalising Arm, pk 3
1 x RR99079 B24/29 Solid Additions Funnel, pk 3
1 x RR99082 PTFE/Silicone Septa 29mm for B24/29, pk 50
This system does not include reaction flasks, stirring hotplate etc and
is intended for those users who already have a heated Carousel 6 Plus.
-78˚C
180˚C

Page 7
Cooled Carousel 6 Plus Reaction Station
6. Products & Accessories - Continued
Carousel 6 Plus Reaction Station
Carousel 6 Plus
RR99916 Carousel 6 Plus Reaction Station
Carousel 6 Plus Systems
RR99942 Carousel 6 Plus System 33 - Heated System - Basic + Hotplate
RR99943 Carousel 6 Plus System 34 - Heated System - Basic
RR99944 Carousel 6 Plus System 35 - Heated System - Additions
RR99945 Carousel 6 Plus System 36 - Heated System - Comprehensive
Glassware
Standard Flasks & Reflux Tubes
RR99151 5ml Reaction Flask with B14/23 Neck, pk 6
RR99148 10ml Reaction Flask with B14/23 Neck, pk 6
RR99923 B14/23 Reflux Tube + Connecting Set, pk 6
RR99145 25ml Reaction Flask, pk 6
RR99070 50ml Reaction Flask, pk 6
RR99054 100ml Reaction Flask, pk 6
RR99052 170ml Reaction Flask, pk 6
RR99041 250ml Reaction Flask, pk 6
RR99917 250ml Reaction Flask + Reflux Tube + Connecting Set, pk 6
RR99918 B24/29 Reflux Tube + Connecting Set, pk 6
RR99919 250ml Long Neck Reaction Flask, pk 6
Standard Flasks with Sidearms
RR99071 50ml Reaction Flask with B14/23 Sidearm + Septa Port, pk 6
RR99074 100ml Reaction Flask with B14/23 Sidearm + Septa Port, pk 6
RR99077 250ml Reaction Flask with B24/29 Sidearm + Septa Port, pk 6
RR99047 250ml Reaction Flask with B14/23 Sidearm + Septa Port, pk 6
RR99087 250ml Reaction Flask with 2x B14/23 Sidearms + Septa Port, pk 6
RR99088 250ml Reaction Flask with 2x B24/29 Sidearms + Septa Port, pk 6
RR99089 250ml Reaction Flask with 1x B14/23 + 1x B24/29 Sidearm + Septa Port, pk 6
Wide Neck Flasks with Coupling
RR99924 Wide Neck Reflux Tube + Coupling
RR99925 Coupling for Wide Neck Flasks
RR99940 PTFE Wide Neck Support Collar
RR99941 Nitrile O-Ring for PTFE Wide Neck Support Collar
RR99926 50ml Wide Neck Flask
RR99928 100ml Wide Neck Flask
RR99933 250ml Wide Neck Flask
Wide Neck Flasks with Sidearms
RR99927 50ml Wide Neck Flask with B14/23 Sidearm + Septa Port
RR99929 100ml Wide Neck Flask with B14/23 Sidearm + Septa Port
RR99930 100ml Wide Neck Flask with 2x B14/23 Sidearms + Septa Port
RR99934 250ml Wide Neck Flask with B24/29 Sidearm + Septa Port
RR99935 250ml Wide Neck Flask with 2x B24/29 Sidearms + Septa Port
Wide Neck Flasks with Baffles & Sidearms
RR99931 100ml Baffled Wide Neck Flask with B14/23 Sidearm + Septa Port
RR99932 100ml Baffled Wide Neck Flask with 2x B14/23 Sidearms + Septa Port
RR99936 250ml Baffled Wide Neck Flask with B24/29 Sidearm + Septa Port
RR99937 250ml Baffled Wide Neck Flask with 2x B24/29 Sidearms + Septa Port
Azeotropic Glassware
RR99922 100ml Azeotropic Reaction Flask, pk 3
RR99920 250ml Azeotropic Reaction Flask, pk 3
RR99939 100ml Tornado Azeotropic Reaction Flask, pk 3
RR99938 250ml Tornado Azeotropic Reaction Flask, pk 3
Liquid & Solid Addition Funnels
RR99049 B14/23 Solid Additions Funnel, pk 3
RR99079 B24/29 Solid Additions Funnel, pk 3
RR99048 B14/23 Liquid Additions Dropping Funnel + Pressure Equalising Arm, pk 3
RR99078 B24/29 Liquid Additions Dropping Funnel + Pressure Equalising Arm, pk 3
Evaporation Adapters, Splash Heads & Evaporation Flasks
RR99045 Rotary Evaporator Adapters Rodaviss B29 Socket to B24 Cone, pk 2
RR99046 Rotary Evaporator Adapters Ordinary B29 Socket to B24 Cone, pk 2
RR99055 Rotary Evaporator Adapters USA 24/40 Socket to B24 Cone, pk 2
RR99083 Splash Head Rodaviss B29 Socket to B24 Cone
RR99084 Splash Head Ordinary B29 Socket to B24 Cone
RR99085 Splash Head USA 24/40 Socket to B24 Cone
RR99053 250ml EZ-2/HT Evaporation Flask (Compatible with Genevac Evaporators), pk 4
Flasks ar NOT suppli d with r flux tub s or PTFE caps, th s must b purchas d s parat ly.
Carousel 6 Plus Accessories
Aluminium Inserts
RR99142 Aluminium Insert for 5ml Reaction Flask, pk 6
RR99141 Aluminium Insert for 10ml Reaction Flask, pk 6
RR99140 Aluminium Insert for 25ml Reaction Flask, pk 6
RR99060 Aluminium Insert for 50ml Reaction Flask, pk 6
RR99058 Aluminium Insert for 100ml Reaction Flask, pk 6
RR99057 Aluminium Insert for 170ml Reaction Flask, pk 6
PTFE Magnetic Stirring Bars
RR98075 Cross Shape Stirring Bar 10mm, pk 40
RR98091 Cross Shape Stirring Bar 16.5mm RE, pk 20
RR98096 Elliptical Stirring Bar 10mm RE, pk 40
RR98097 Elliptical Stirring Bar 15mm RE, pk 20
RR99064 Elliptical Stirring Bar 25mm RE, pk 10
RR98070 Octagonal Stirring Bar 13mm, pk 20
RR98071 Pivot Ring Stirring Bar 12 x 6mm, pk 40
RR98095 PTFE Magnetic Stirring Bar Evaluation Kit, pk 30
RR98094 PTFE Magnetic Stirring Bar Retriever 350mm
RR98114 Magnetic Stirring Bar Restrainer
PTFE Caps & Accessories
RR91070 Easy-On PTFE Cap, pk 6
RR91072 Easy-On PTFE Storage Cap, pk 6
RR98076 Silicone Septa for PTFE Caps, pk 100
RR98176 Viton Septa for PTFE Caps, pk 100
RR98060 O-Rings for Caps - Nitrile 24mm, pk 100
RR98160 O-Rings for Caps - Viton 24mm, pk 100
Replacement O-Rings & Quick Release Couplings
RR91060 Nitrile O-Rings 4mm Gas Outlet - Bottom, pk 50
RR91061 Nitrile O-Rings 3mm Gas Outlet - Top, pk 50
RR91062 Viton O-Rings 4mm Gas Outlet - Bottom, pk 50
RR91063 Viton O-Rings 3mm Gas Outlet - Top, pk 50
RR91065 Qui ck Release Male Body Coupling No Shut-off, pk 2
RR91066 Quick Release Barbed Coupling No Shut-off
RR99062 Quick Release Barbed Coupling + Shut-off, pk 2
RR99063 90˚ Elbow Quick Release Barbed Coupling + Shut-off, pk 2
RR99065 Quick Release Male Body Coupling + Shut-off, pk 2
Rodaviss Caps & Connecting Sets
RR99051 B14/23 Rodaviss Sealing Cap, pk 10
RR99043 B24/29 Rodaviss Sealing Cap, pk 10
RR99955 B14/23 Rodaviss Connecting Set, pk 10
RR99044 B24/29 Rodaviss Connecting Set, pk 10
RR99080 PTFE/Silicone Septa 20mm - B14/23, pk 50
RR99081 PTFE/Silicone Septa 24.5mm - B19/26, pk 50
RR99082 PTFE/Silicone Septa 29mm - B24/29, pk 50
RR99068 B14/23 Rodaviss Connecting Cap, pk 10
RR99090 B24/29 Rodaviss Connecting Cap, pk 10
Other Accessories
RR98906 Black Lab Marker, pk 10
RR99061 Reaction Flask Support Ring, pk 6
RR99067 Tubing for Inert Gas or Reflux Cooling 2 metres

Page 8
Cooled Carousel 6 Plus Reaction Station
6. Products & Accessories - Continued
Storm & Breeze Heating/Cooling Work Stations
Storm & Breeze Work Stations
RR96200 Storm Work Station + Carousel 6 PTFE Insulating Plates
(M24 hose connections)
RR96210 Breeze Work Station + Integral Stand
(M16 hose connections)
RR96220 PTFE Insulating Plates for Carousel 6 Plus
Insulated Hoses
HB6084 Insulated Hose - 100cm long, with M16 thread
HB6085 Insulated Hose - 150cm long, with M16 thread
HB6136 Insulated Hose - 200cm long, with M16 thread
HB6255 Insulated Hose - 300cm long, with M16 thread
HB6784 Insulated Hose - 100cm long, with M24 thread
HB6785 Insulated Hose - 150cm long, with M24 thread
HB6786 Insulated Hose - 200cm long, with M24 thread
HB6787 Insulated Hose - 300cm long, with M24 thread
M16 Hose Adapters & Valves
HB6945 M16 Female to M24 Male Adapter
HB6431 M16 Female to M30 Male Adapter
HB6195 M16 90 Degree Adapter
HB6091 M16 Ball Valve
RR96316 M16 Thread Protection Cap, pk 10
M24 Hose Adapters & Valves
HB6724 M24 Female to M16 Male Adapter
HB6723 M24 Female to M30 Male Adapter
HB9256 M24 90 Degree Adapter
HB9236 M24 Ball Valve
RR96336 M24 Thread Protection Cap, pk 10
M30 Hose Adapters & Valves
HB6454 M30 Female to M16 Male Adapter
HB9268 M30 Female to M24 Male Adapter
HB6461 M30 90 Degree Adapter
HB6451 M30 Ball Valve
RR96348 M30 Thread Protection Cap, pk 10
Thermofluids
HB6164 Silicone Oil -40˚C to +165˚C - 10 litres
HB6162 Silicone Oil -20˚C to +235˚C - 10 litres
HB6479 DWTherm -90˚C to +200˚C - 10 litres
Cooled Carousel 6 Plus Reaction Station
Cooled Carousel 6 Plus
RR99947 Cooled Carousel 6 Plus (Reservoir, Head, Stand + Cover)
Cooled Carousel 6 Plus Systems
RR99949 Cooled Carousel 6 Plus System 38 - Cooled System - Upgrade
RR99950 Cooled Carousel 6 Plus System 39 - Cooled System - Comprehensive
Cooled Carousel 6 Plus Components
RR99501 Cooled Carousel 6 Plus Cooling Reservoir
RR99948 Cooled Carousel 6 Plus Head
RR99503 Cooled Carousel 6 Plus Stand
RR99515 Cooled Carousel 6 Plus Cover (with gas connectors)
Cooled Carousel 6 Plus Accessories
RR99905 Digital Thermometer (-250˚C to +400˚C) + 200mm Probe
RR99906 Digital Thermometer (-250˚C to +400˚C)
RR99907 200mm Temperature Probe
RR99908 Dry Ice Scoop
RR99909 Cold Temperature Apron 1060mm long, Waterproof
RR99910 Protective Face Shield
RR98024 Protective Cold Temperature Gloves
RR71505 Cooling Protection Kit
1 x RR99909 Cold Temperature Apron 1060mm long, Waterproof
1 x RR99910 Protective Face Shield
1 x RR98024 Protective Cold Temperature Gloves
Tornado Plus Overhead Stirring System
Tornado Plus
RR99951 Tornado Plus Overhead Stirring System
RR99230 Universal Support Stand
RR99237 Carousel Temperature Controller with extended 345mm Probe
RR71127 Temp Sensor Support Rod (13mm x 500mm)
RR71120 Support Rod Hotplate Adapter (extension plate)
PTFE Stirring Paddles
RR99240 Centrifugal PTFE Stirrer Paddle, pk 6
RR99241 Centrifugal PTFE Stirrer Paddle, pk 1
RR99244 Anchor PTFE Stirrer Paddle for 100ml Flasks, pk 6
RR99245 Anchor PTFE Stirrer Paddle for 100ml Flasks, pk 1
RR99248 Anchor PTFE Stirrer Paddle for 250ml Flasks, pk 6
RR99249 Anchor PTFE Stirrer Paddle for 250ml Flasks, pk 1
RR99251 Propellor PTFE Stirrer Paddle, pk 6
RR99252 Propellor PTFE Stirrer Paddle, pk 1
RS Overhead Stirrers
Please note that other overhead stirrer options are available on request, please contact Radleys
for further information.
RR91302 RS27 Standard Overhead Stirrer 230v UK Plug
RR91302/EURO RS27 Standard Overhead Stirrer 230v Euro Plug
RR91306 RS37 Digital Plus Overhead Stirrer 230v UK Plug
RR91306/EURO RS37 Digital Plus Overhead Stirrer 230v Euro Plug
RR91308 RS50 Control Overhead Stirrer 230v UK Plug
RR91308/EURO RS50 Control Overhead Stirrer 230v Euro Plug
RR91428 Remote Control for RS50 Control
RR91430 RS232 Interface Cable Stirrer-PC
Spare Parts
RR99952 Quick-Thread Stirrer Guide Assembly
RR99953 Replacement Tubing for RR99952 110mm (Stirrer Guide to gas manifold)
Pl as contact Radl ys for d tails of r plac m nt parts for th Stirr r Guid and Tornado Syst m.
Stirring Hotplates
Tech Stirring Hotplates
RR91202 Carousel Tech Stirring Hotplate 230v UK Plug
RR91202/EURO Carousel Tech Stirring Hotplate 230v Euro Plug
RR91202/JAP Carousel Tech Stirring Hotplate 110v Japanese Plug
RR91202/SWISS Carousel Tech Stirring Hotplate 240v Swiss Plug
RR91202/USA Carousel Tech Stirring Hotplate 115v US Plug
Advanced Stirring Hotplates
RR91204 Carousel Advanced Stirring Hotplate 230v UK Plug
RR91204/EURO Carousel Advanced Stirring Hotplate 230v Euro Plug
RR91204/JAP Carousel Advanced Stirring Hotplate 100V Japanese Plug
RR91204/SWISS Carousel Advanced Stirring Hotplate 230v Swiss Plug
RR91204/USA Carousel Advanced Stirring Hotplate 115V US Plug
Oth r Stirring Hotplat mod ls ar availabl on r qu st...
Temperature Controllers & Sensors
RR91210 Carousel Temperature Controller
Includ s RR71125 T mp S nsor Support Rod and RR71120 Support Rod Hotplat Adapt r
RR91226 Pt1000 S/S Sensor for Carousel Advanced Stirring Hotplate
RR91227 Pt1000 Glass Coated Sensor for Carousel Advanced Stirring Hotplate
RR91228 Temperature Sensor Holder
RR91229 Temp Sensor Support Rod (13mm x 425mm)
RR71127 Temp Sensor Support Rod (13mm x 500mm)
RR71125 Temp Sensor Support Rod (13mm x 340mm)
RR71120 Support Rod Hotplate Adapter (extension plate)
RR91234 RS232 Interface Cable - 15 Pin to 9 Pin

Page 9
7. Glassware
Cooled Carousel 6 Plus Reaction Station
7.1 Standard Flasks...
• Constructed of heavy duty, borosilicate glass.
• Choice of sizes: 5ml, 10ml, 25ml, 50ml, 100ml, 170ml and 250ml.
• Precision engineered round bottom design ensures an excellent fit with the Carousel’s wells, maximising heat transfer.
• Flasks feature a B24/B29 Rodaviss joint for connection to the detachable Reflux Tube.
• The 5ml and 10ml flasks feature a B14/23 Rodaviss joint.
• Reflux Tubes feature a quick-thread, which when combined with PTFE Easy-On Cap, give a gas tight seal up to 1psi.
• Optional single piece 250ml Long Neck design.
• Compatible with magnetic stirring bars and the centrifugal stirrer paddles (Tornado).
• Flasks are NOT supplied with reflux tubes or PTFE caps, which must be ordered separately.
7.2 Standard Flasks with Sidearms...
• Sidearm flasks have all the features of the standard flasks with the addition of either one or two sidearms.
• Flasks are available with B14/23 and B24/29 sidearm options and include Rodaviss caps and sealing septa.
• Sidearms are compatible with Dropping Funnels and Solid Addition Funnels.
7.3 Azeotropic Flasks (Dean & Stark)...
• Unique single piece flasks fit directly into the Carousel 6 Plus, allowing chemists to perform up to six Azeotropic processes in parallel.
• Choice of sizes: 100ml and 250ml.
• Special Tornado version with wider neck is designed to accept centrifugal stirring paddle.
• Flasks feature a replaceable PTFE stopcock for removal of the aqueous phase.
• Feature B14/23 or B24/29 sidearms.
• Sidearms are compatible with Dropping Funnels and Solid Addition Funnels.
RR99145
25ml Flask
RR99151
5ml Flask
RR99148
10ml Flask
RR99070
50ml Flask
RR99054
100ml Flask
RR99052
170ml Flask
RR99041
250ml Flask
RR99918
B24/29 Reflux Tube
+ Connecting Set
RR99923
B14/23 Reflux Tube
+ Connecting Set
RR99917
250ml Flask
+ Reflux Tube
+ Connecting Set
RR99919
250ml Long
Neck Flask
RR99071
50ml Flask with
B14/23 Sidearm
RR99087
250ml Flask with
2 x B14/23 Sidearm
RR99088
250ml Flask with
2 x B24/29 Sidearm
RR99077
250ml Flask with
B24/29 Sidearm
RR99047
250ml Flask with
B14/23 Sidearm
RR99074
100ml Flask with
B14/23 Sidearm
RR99089
250ml Flask with
1 x B14/23 Sidearm
1 x B24/29 Sidearm
RR99922
100ml Azeotropic
Flask
RR99938
250ml Tornado
Azeotropic Flask
RR99920
250ml Azeotropic
Flask
RR99939
100ml Tornado
Azeotropic Flask

Cooled Carousel 6 Plus Reaction Station
7. Glassware - Continued
Page 10
7.4 Wide Neck flasks with quick coupling...
• Wide neck flasks allow easier removal of viscous or solid samples and facilitate the use of the anchor and propeller style PTFE stirrer
blade options (Tornado).
• These unique vessels feature a 50mm ID flat flange, which combines with a Nitrile O-ring and our unique self centering PTFE collar to
offer a leak-tight seal between the glass flask and wide neck reflux tube.
• The two flanges are compressed together using a two part threaded polymer coupling, which both holds the components rigid and offers
an excellent gas tight seal.
7.5 Wide Neck flasks with Sidearms...
• Wide neck sidearm flasks have all the features of the wide neck flasks with either one or two sidearms. Ideal for attaching dropping fun-
nels or powder funnels.
RR99933
250ml Wide
Neck Flask
RR99924
Wide Neck Reflux
Tube + Coupling
RR99925
Coupling for
Wide Neck Flasks
RR99926
50ml Wide
Neck Flask
RR99928
100ml Wide
Neck Flask
RR99941
Nitrile O-ring
for PTFE Wide
Neck Flask
RR99940
PTFE Wide
Neck Support
Collar
7.7 Sundry Glassware...
• Choice of 50ml Liquid Additions Dropping Funnels with B14/23 or B24/29 Rodaviss joints.
• The dropping funnels are ideal for the controlled addition of larger volumes of reagent directly into the reaction flask. Funnels feature a
drip cone and pressure equalising arm for ease of addition.
• Choice of Solid Additions Funnel with B14/23 or B24/29 Rodaviss joints; ideal for the addition of powders or solids into the reaction flask.
RR99930
100ml Wide Neck Flask
with 2 x B14/23 Sidearm
RR99929
100ml Wide Neck Flask
with B14/23 Sidearm
*RR99935
250ml Wide Neck Flask
with 2 x B24/29 Sidearm
RR99927
50ml Wide Neck Flask
with B14/23 Sidearm
RR99934
250ml Wide Neck Flask
with B24/29 Sidearm
7.6 Wide Neck flasks with Baffles & Sidearms...
• Wide necked baffled flasks improve the turbulence within the flask by disrupting the creation of a central vortex. Baffled flasks are rec-
ommended for use with stirrer paddles for maximum effect.
*RR99937
250ml Baffled Wide Neck
Flask with 2 x B24/29 Sidearms
RR99932
100ml Baffled Wide Neck Flask
with 2 x B14/23 Sidearms
RR99936
250ml Baffled Wide Neck
Flask with B24/29 Sidearm
RR99931
100ml Baffled Wide Neck Flask
with B14/23 Sidearm
RR99045
Rotary Evaporator Adapter
Rodaviss B29 Socket
to B24 Cone
RR99046
Rotary Evaporator Adapter
Ordinary B29 Socket
to B24 Cone
RR99055
Rotary Evaporator Adapter
USA 24/40 Socket
to B24 Cone
RR99048
B14/23 Liquid Additions
Dropping Funnel +
Pressure Equalising Arm
RR99078
B24/29 Liquid Additions
Dropping Funnel +
Pressure Equalising Arm
RR99079
B24/29 Solid
Additions
Funnel
RR99049
B14/23 Solid
Additions
Funnel
RR99053
250ml EZ-2/HT
Evaporation Flask
*The RR99935 and RR99937 250 l wide neck flasks with 2 x B24/29 sidear s (non-baffled and baffled) are not co patible with the RR99947 Cooled Carousel 6 Plus.

Cooled Carousel 6 Plus Reaction Station
8.1 Assemble the reactor head and cooling reservoir, so that they fit securely
together. Make sure that the reactor head is properly seated on the central
pillar of the reservoir.
8.2 Locate the Cooled Carousel assembly on top of the stirring hotplate.
8.3 Place a magnetic stirring bar into your chosen reaction flask.
8.4 If you are using a flask which does not have an integrated reflux
neck you will need to attach an appropriate reflux tube at this time
or skip to step 8.5.
See 9.5. Assemble the Reaction Flask and Reflux Tube.
8.5 Assemble the glassware and Easy-On PTFE Cap by screwing together.
This process requires a 1/4 turn of the cap, hand tighten until the cap feels
secure. Ensure the tube is properly located with the O-Ring seal in the cap.
8.6 If required fit a new Silicone Septa to the central hole in the cap, and ensure
that the cap valve is in the “open” position.
8.7 Place the flask into the Cooled Carousel at the required location, push the
cap down onto the connector pin until the flask is fully located in the base.
8.8 Select a suitable cooling mixture depending on the minimum temperature
desired. We recommend using a dry ice/acetone cooling mixture for best
cooling results.
8.9 Using a suitable wash bottle or similar, fill the reservoir with solvent to a
depth of approximately 2cm.
8.10 Using the RR99908 scoop, carefully add dry ice in small portions until the
desired temperature is reached.
8.11 Once the desired temperature is reached air sensitive liquid reagents may
be carefully added via the Silicone Septa under an inert atmosphere.
8.12 Locate the optional lid onto the cooling reservoir.
8.13 Connect the central quick fit gas connector to a regulated inert gas supply
and switch the gas supply on.
8.14 If a gas connector position is not in use it can be blocked off with an inverted
Silicone Septa.
8.15 Set the stirring hotplate to the required stirring speed.
8. Quick Start Guide
Warning
The cooling reservoir is not suitable for heated reactions.
Always ensure that the heating control on the stirring
hotplate is switched off and that the hotplate is cool before
using the cooling reservoir.
!
Important Note
Check that the reservoir can be removed from the top plate
without sticking before you fill the reservoir.
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Page 11
8.1
8.2 8.4
8.5 8.6
8.6 8.7
8.7
8.12
8.13
8.10
Open position
8.2
8.9
8.14

Cooled Carousel 6 Plus Reaction Station
9. Set-Up & Operation
Page 12
Important Note
Check that the reservoir can be removed from the top plate without sticking
before you fill the reservoir with a cooling mixture.
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important Note
The Carousel Stirring Hotplate RR98072 is recommended but any
stirring hotplate can be used if the top plate diameter does not exceed
135mm.
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Warning
The Cooled Carousel should always be used in a fumehood with protective
safety sash.
!
Warning
The cooling reservoir is not suitable for heated reactions. Always ensure that
the heating control on the stirring hotplate is switched off and that the hotplate
is cool before using the cooling reservoir.
!
9.1
9.2
For reactions using the cooling reservoir
9.1 Assemble reactor head and cooling reservoir
Place the reactor head on top of the central pillar of the cooling reservoir. The circular recess
in the reactor head accepts the circular protrusion on the central pilar. The Cooled Carousel’s
modular design allows the head to be easily lifted on and off the reservoir as required.
9.2 Locating the cooling reservoir onto the hotplate stirrer
The top surface of your stirring hotplate should be cleaned prior to use. Any small particles
on the surface may affect the fit of the Cooled Carousel reservoir and have an adverse affect
on the performance. Wipe the surface with a cloth or tissue, dampened with an appropriate
solvent (e.g. acetone) and check for any signs of contamination or obstruction.
The undersurface of the Cooled Carousel should also be cleaned prior to use. Wipe the
surface with a cloth or tissue, dampened with an appropriate solvent (e.g. acetone) and
check for any signs of contamination or obstruction.
Position the Cooled Carousel onto the stirring hotplate, making sure that it is secure and
properly seated. The circular recess in the base of the unit is designed to fit snugly around
the top plate of the stirring hotplate (maximum diameter 135mm).
The Cooled Carousels modular design allows it to be easily lifted on and off the hotplate
stirrer as required. Being circular, it can be rotated when in place to facilitate access to all
reaction positions. This removes the need to lean into the fumehood during operation.

9.3 Selection of Glassware
The Cooled Carousel is available with a range of unique glassware tailored for the specfic
requirements of parallel synthesis.
You should select the glassware for your application based on volume, side arm configuration,
neck size and if you need baffles or an azeotropic set-up.
For details of the full range see section 7.
9.4 Select an appropriate stirring bar
Select a suitable magnetic stirrer bar and place in the reaction flask.
We recommend the RR99064 Rare Earth - 25mm eliptical PTFE
Magnetic Stirring Bars.
Cooled Carousel 6 Plus Reaction Station
9. Set-Up & Operation - Continued
Warning
Always take care when loading stirring bars not to drop them, as this may frac-
ture the glass flask. Use RR98094 Stirring Bar Retriever where appropriate.
!
Important Note
All PTFE magnetic stirring bars with time and use, lose their magnetism.
Therefore to optimise stirring performance always replace bars regularly.
I portant, I portant,
I portant, I portant,
I portant, I portant,
I portant, I portant,
I portant, I portant,
I portant, I portant,
I portant, I portant,
Page 13
9.5.1
9.5.2 9.5.3
ab
c
9.5
9.5.2
9.5 Assemble the Reaction Flask and Reflux Tube
All glassware to be used in the Carousel requires a reflux neck. The next step will depend
on the glassware you have selected. If you have chosen...
a. ...a round bottom flask with an integrated reflux neck - see 9.5.1
b. ...a round bottom flask which requires a reflux tube be attached to it - see 9.5.2
c. ...a wide neck round bottom flask - see 9.5.3
9.5.1 Glassware with integrated reflux neck
As you have selected glassware with an integral reflux neck you will not need to add a
separate reflux tube and can skip directly to step 9.6.
9.5.2 Assembling a round bottom flask requiring a reflux neck
You will need to add a reflux tube to your round bottom flask using a Rodaviss®Ground Glass
Joint. Rodaviss is an extra-safe borosilicate glass joint for connecting laboratory glassware.
Rodaviss is safe under vacuum or pressure, grease free, interchangeable with standard 1:10
tapered ground joints (including A, B and C lengths), will not stick or jam, is extremely strong
and can be used up to 200˚C.
Place the cone of the reflux tube through the red connecting cap and then roll the O-Ring up
over the cone until it drops past the shoulder of the cone. Clip the black loosening ring
between the top of the connecting cap and glass rim on the cone. Insert the cone into the
socket on the flask. Screw down the connecting cap onto the thread on the socket to com-
press the O-Ring, thus ensuring a perfect seal and a rigid assembly.
In the event of any difficulty in undoing the joint because the surfaces are jammed, simply
unscrew the connecting cap back onto the loosening ring, which will push the joints easily
apart. This system will release even the most 'frozen' joints.

9. Set-Up & Operation - Continued
Cooled Carousel 6 Plus Reaction Station
Page 14
9.5.3 Assembling Wide Neck glassware
• Position the Lower Coupling (2) around the
neck of the Flask (1) by prising it apart at
the open edges and pushing around the
neck of the Flask.
• The Nitrile O-Ring (4) fits around the
circumference of the Support Collar (3).
• Place the adjoined Support Collar (3) and
O-Ring (4) on to the flat flange of the Flask.
• Then place the Reflux Tube (5) on to the
Flask and O-Ring assembly. (Note: The
Reflux Tube is not held in place until the
next step)
• Slide the Upper Coupling (6) over the top of
the Reflux Tube (5) and screw onto the
Lower Coupling (2).
1. Flask
2. Lower
Coupling
3. Support
Collar
4. Nitrile
O-Ring
5. Reflux
Tube
6. Upper
Coupling
9.5.3
Wide Neck Component Guide
=
=
=
=
2
1
3&4
5
6
Important Note
Please note that the Easy-On Caps ‘seal’ on the Nitrile O-Ring before they are
fully tightened. Fully tightening the caps will effect a double seal on the PTFE
inner of the cap and Nitrile O-Ring.
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
9.6
9.6
9.6 Fitting the Easy-On Caps to the Reflux Neck
Assemble the flask and Easy-On Cap by screwing together. This process requires a ¼ turn
of the cap, hand tighten until the cap feels secure. Ensure that the reflux tube is properly
located within the O-Ring seal in the cap. The Nitrile O-Ring forms a gas tight seal with the
outside of the reflux tube.
Caps feature a replaceable Nitrile O-Ring as standard with an optional Viton replacement.
These O-Rings will be subject to chemical attack and will require periodic replacement.
• RR98060 O-Rings for Caps - Nitrile 24mm (standard), pk 100
• RR98160 O-Rings for Caps - Viton 24mm, pk 100
A Septa is located in the top of each Easy-On Cap, which permits either reaction monitoring
through the withdrawal of aliquots or the addition of reagents during synthesis. Septa require
periodic replacement.
• RR98076 Silicone Septa for PTFE caps, pk 100
Warning
Do not over tighten the Easy-On Caps, as this may damage the cap and
fracture the reflux tube.
!

9.10 Adding dry ice to the reservoir
Using the RR99908 scoop, carefully add dry ice in small portions until the desired tempera-
ture is reached.
See 9.16 Monitoring the coolant temperature.
Once the desired temperature is reached air sensitive liquid reagents may be carefully added
under an inert atmosphere via the Suba-Seals using a syringe.
Page 15
9. Set-Up & Operation - Continued
Cooled Carousel 6 Plus Reaction Station
9.8.1
9.8.2
Warning
Do not initially fill the reservoir with solvent beyond a 2cm depth, as bumping
(splashing) can occur and solvent may be ejected from the reservoir. In
particular, as the cooling mixture contacts the warmer surface of the tubes,
rapid evaporation of CO2can occur, causing bubbling and spitting.
!
Important Note
If you wish to cool down to the minimum temperature of you cooling mixture
it is preferable to put the dry ice in first and carefully add the solvent via a
wash bottle afterwards.
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important Note
By carefully controlling the amount of dry ice being added you can regulate
the temperature of the mixture more accurately.
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
9.9
9.10
Warning
Cooling mixtures can cause severe frost burns. Always wear suitable protec-
tive gloves, aprons and faceshield.
!
9.7 Using Easy-On Caps with Valve & Septum for Inerting & Additions
When using the Easy-On Caps, to ensure the flow of inert gas, the valve must be in the
‘OPEN’ position (see 9.7A). A quarter turn (anti-clockwise or clockwise) will ‘CLOSE’ the
valve, isolating the Reaction Flask (see 9.7B)
The Flasks can be isolated or removed during synthesis by simply closing the valve on the
Easy-On PTFE Cap ensuring the other reactions remain under controlled inert conditions.
9.8 Inserting Flasks
9.8.1 Insert the flask into the required location.
9.8.2 Push the Easy-On Cap onto the connector pin, until the cap is fully seated on the
connector. Repeat for each flask.
9.9 Adding cooling fluid to the reservoir
Select a suitable cooling mixture depending on the minimum temperature desired.
(Examples are given in Appendix see page 18). We recommend using a dry ice and acetone
cooling mixture for best cooling results.
Open Closed
9.7A 9.7B
Warning
Do not over tighten the Easy-On Caps, as this may damage the cap and
fracture the reflux tube.
!

9.11 Using the Cooled Carousel 6 Plus Lid
The HDPE lid is included with your cooling reservoir. The lid is designed to keep your reaction
cooler for longer, minimises condensation and ice formation on your flasks (maintaining
visibility of the contents) and prevents spitting from the cooling mixture.
The lid locates onto the cooling reservoir, resting on the edge of the reservoir and the
reactor head.
9.12 Connect Gas/Vacuum supply
The central gas inlet and radial distribution system combined with gas tight Easy-On PTFE
Caps allows reactions to be performed under an inert (nitrogen/argon) atmosphere.
For use under an inert atmosphere attach tubing to the central quick-release coupling and
connect via a 3-way tap or stopcock to a vacuum source and inert gas supply (recommended
maximum pressure 1psi).
Then, by alternately evacuating the system and filling it with a suitable inert gas (repeating
2 to 3 times) will achieve an inert atmosphere within the flasks.
The Reaction Flasks can be isolated or removed during synthesis by simply closing the valve
on the Easy-On PTFE Cap and removing the flask.
9.13 Stainless Steel Gas Outlets
The stainless steel Gas Outlets on top of the reactor head feature an upper and lower Nitrile
O-Ring as standard, an optional Viton replacements.
These O-Rings may be subject to chemical attack and will require periodic replacement.
RR91060 Nitrile O-Rings 4mm Gas Outlet - Bottom, pk 50
RR91061 Nitrile O-Rings 3mm Gas Outlet - Top, pk 50
RR91062 Viton O-Rings 4mm Gas Outlet - Bottom, pk 50
RR91063 Viton O-Rings 3mm Gas Outlet - Top, pk 50
9.14 Unused positions
If all Gas Outlets are not in use, unused outlets can be blocked off with a inverted Silicone
Septa.
9. Set-Up & Operation - Continued
Cooled Carousel 6 Plus Reaction Station
Page 16
9.12
Warning
When applying gas to the reaction station gas inlet system do not exceed 1
(one) psi as the Reaction Flask and Easy-On PTFE Caps are not rated for
pressure, (they are however suitable for applying a vacuum).
!
9.11
9.139.14

9.15 Temperature control
9.15.1 Set the stirring speed of the stirring hotplate to the desired level.
9.16 Monitoring the coolant temperature
For low temperature measurement we recommend use of the RR99905 Digital Thermometer
(-250°C to +400°C) and 200mm Probe.
The probe may be placed directly into the reservoir bowl to monitor the cooling mixture - by
sliding the probe through the hole in the reactor head. Alternatively insert the probe in to one
of the reaction tubes (through the Silicone Septa) to directly monitor the sample temperature.
9.17 Using the Carousel 6 Plus Stand
The Carousel 6 Plus stand is designed to support the reactor head either with or without
reaction flasks. The HDPE stand and integral drip tray catches any dripping condensation
from flasks and gives excellent stability.
• Use between reactions.
• Use when loading flasks.
• Use for reactions that require rapid air warming.
9.18 Once your synthesis is complete
• Turn off stirring.
• Turn off gas supply and disconnect gas inlet.
• Remove the lid.
• Remove the Carousel reactor head from reservoir and place on the Carousel Stand.
• The Reaction Flasks can now be removed.
9.19 Disposing of used cooling fluid
Cooled Carousel 6 Plus Reaction Station
9. Set-Up & Operation - Continued
Warning
Always wear suitable protective gloves, aprons and face shield when handling
cooling mixtures as extreme cold can cause severe frost burns.
!
Warning
Do not attempt to pour cold mixtures into containers that can be sealed. There
is a danger of splashing and pressure build up. Preferably allow the cooling
mixture to warm to room temperature before disposing of it according to
appropriate local regulations.
!
Page 17
9.15
Important Note
Please read the separate RR98072 Carousel Stirring Hotplate instructions
thoroughly before operation.
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Important, Important,
Warning
The cooling reservoir is not suitable for heated reactions. Always ensure that
the heating control on the stirring hotplate is switched off and that the hotplate
is cool before using the cooling reservoir.
!
9.16
9.16
9.16
9.17

10. Safe handling of cooling fluids
11. Cooling Mixture Appendix
Cooled Carousel 6 Plus Reaction Station
Page 18
9.1 Use protective equipment and follow the instructions
The use of manual cooling mixtures offers considerable savings in time, convenience and
cost when compared with automatic chillers or cryostats. However, extreme care should be
taken when using cooling mixtures.
• Cooling mixtures can unexpectedly spit and splash. Therefore it is important to cover all
exposed skin by wearing suitable protective gloves, aprons and face shield.
• Do not initially fill the reservoir with solvent beyond the recommended depth as vigorous
bumping (splashing) can occur and solvent may be ejected from the reservoir. In particu-
lar, as the cooling mixture contacts the warmer surface of the flasks rapid evaporation of
CO2can occur, causing bubbling and spitting.
• Take care when disposing of dry ice/acetone or other solvent mixtures! Do not attempt to
pour cold mixtures into containers that can be sealed. There is a danger of splashing and
pressure build up. Preferably allow the cooling mixture to warm to room temperature
before disposing of it according to appropriate local regulations
The following protective equipment is available from Radleys:
•RR99909 - Cold Temp. Apron 1060mm long Waterproof
•RR99910 - Protective Face Shield
•RR98024 - Protective Cold Temperature Gloves
•RR71505 - Cooling Protection Kit, includes
1 x RR99909 Cold Temp. Apron 1060mm long Waterproof
1 x RR99910 Protective Face Shield
1 x RR98024 Protective Cold Temperature Gloves
Dry Ice/Solvent Cooling Mixtures
Minimum Flash Point Inhalation Skin Explosive/
Solvent Temperature (°C) (°C) Toxicity Toxicity Fire Hazard
Ethylene glycol -15 110 low low low
Butyl ethyl ketone -38 41 moderate low high
Acetonitrile -42 5 high high high
Cyclohexanone -46 46 moderate low moderate
Di(ethylene glycol) diethyl ether -52 71 moderate moderate moderate
Ethanol -72 8 low low high
Methanol -78 11 high high high
Acetone -78 -17 low low high
A Phipps & D N Hum J Ch m Educ, 45, 664, 1968
Other Cooling Mixtures
Minimum
Temperature (°C) Comments
Sodium chloride/ice -21 33g salt/81g ice
Calcium chloride/ice -40 100g salt/81g ice

Cooled Carousel 6 Plus Reaction Station
Page 19
70-0061
RR91081 - Quick-Thread Glass
Reaction Tube 24mm x 150mm 70-0670
70-0657
70-0581 70-0581 70-0581 70-0581
70-0581
70-0581 70-0581 70-0581
70-0706 70-0706 70-0706
70-0658
70-0061 70-0061
Radleys/Genevac Compatibility
Carousel 12 Plus
Cooled Carousel 12 Plus
Metz Syn10 Reaction Station
Carousel 6 Plus
Cooled Carousel 6 Plus
Carousel 6 Plus
Cooled Carousel 6 Plus
Not compatible with Radleys
Products
Vessels
Per Rack
Racks
Per Evap
Total Vessels
Per Evap
Vessels
Per Rack
Racks
Per Evap
Total Vessels
Per Evap
Vessels
Per Rack
Racks
Per Evap
Total Vessels
Per Evap
Vessels
Per Rack
Racks
Per Evap
Total Vessels
Per Evap
RR99053 - EZ-2/HT 250ml
Evaporation Flask
RR99041 - 250ml Reaction
Flask (Flask only)
RR99052 - 170ml Reaction Flask
EZ-2 HT-4 HT-8 HT-12
RR99054 - 100ml Reaction Flask
8 2 16 6 4 24 6 8 48 6 12 72
3 2 6 2 4 8 2 8 16 2 12 24
2 2 4 2 4 8 2 8 16 2 12 24
1 2 2 1 4 4 1 8 8 1 12 12
3 2 6
N/A N/A N/A
12. Radleys/Genevac Rack Compatibility Guide
Other manuals for Carousel 6 Plus
1
Other Radleys Laboratory Equipment manuals

Radleys
Radleys GreenHouse Plus User manual

Radleys
Radleys Reactor-Ready User manual

Radleys
Radleys Reactor-Ready Pilot User manual

Radleys
Radleys Reactor-Ready Pilot User manual

Radleys
Radleys Carousel Standard User manual

Radleys
Radleys Carousel 6 Plus User manual

Radleys
Radleys Reactor-Ready User manual

Radleys
Radleys Heat-On RR68000 User manual

Radleys
Radleys Reactor-Ready Lab Reactor User manual

Radleys
Radleys Carousel 12 Plus User manual