Agam AIT 501 User manual

It is recommended to select the specific 8pcs
When charging battery is applied, the charging battery
work mode should be selected at the startup:
Insulation Resistance Tester
1 pc
1 pc
1 pc
1 pc
1 pc
1 pc
1 pc
1 pc
8 pcs
One-plug
Two-plug
chargeable batteries (LR14) and a charger.
damaged part,
measurement requirement: Pollution Degree 2,
Overvoltage Category(CAT. III 600V) and Double
Insulation.
alerts the user to avoid electric shock.
affect
test
leads and power adaptor with the same
model or identical electrical specifications.
Low Battery Indication
after 15 minutes' inactivity.
There is a battery indicator shown on the upper left
corner of the display. Please refer to Table 3 for
detailed explanation.
almost
necessary. Under this status, the Meter
can still output 500V and 1000V to
measure, the measured accuracy will
not be affected.
hold USB button prior to startup, then press down ON/
OFF, LCD screen will display CHA or GEN, and select
to display CHA by pressing the up/down key, after
pressing USB key to confirm, the Meter successfully
enters the charging battery work mode. GEN means
the general alkaline battery work mode.
Press and
Two-plug
One-plug
One-plug
(Connected to one-plug test lead)
(Connected to one-plug black test lead)
( Connected to two-plug red test lead )
(Connected to two-plug red test lead)
voltage input terminal
Arrow Button
Arrow Button
Arrow Button
Arrow Button
The Meter defaults at 500V range and
under continuous measurement of
insulation resistance when turned on.
LIGHT Press to turn on/off the backlight.
Press to clear the saved data.
value. The Meter can save up to 18 sets.
previous
voltage range.
next voltage
range.
setting the timer for the
When compare function is enabled for
insulation resistance measurement,
values
setting the timer for the
When compare function is enabled for
insulation resistance measurement,
values
Press
measurements
This
The Meter defaults at
500V range and under continuous measurement of
insulation resistance when turned on.
AIT 501 USERS MANUAL
AIT 501 Smart Digital Insulation Tester
AIT 501 Smart Digital
This operating manual is subject to change without notice.
*END*
Insert the red test lead into two LINE terminals,
the black one into GUARD and the green one
into EARTH.
-114
Caution
test leads into EARTH
and two LINE terminals.
present on the red test lead,
if negative voltage is
output
cause
measuring resistance <2MΩ with use of 500V.
measuring resistance <5MΩ with use of 1000V.
measuring resistance <10MΩ with use of 2500V.
measuring resistance <20MΩ with use of 5000V.
voltage range.
green
outputs
Choose one of insulation resistance measurement
modes shown as below:
personal injury.
carry out begin and output insulation resistance
test voltage
turn off the voltage output,
the
the test voltage
output will be turned off,
the test
voltage output will be turned off and the
step
readings.
Calculation Tips:
PI = 3-minute ~ 10-minute resistance/30-second ~
1-minute resistance
You can choose compare from:
compare value.
See Figure 6 for the use of power adaptor.
Using Power Adaptor
Figure 6. Using Power Adaptor
Connecting USB Interface
See Figure 7 for USB interface connection.
General Specifications
18 sets
Test Voltage Automatically source the voltage.
Preset the timer for
two points and the Meter will carry out the measurement automatically.
B. Insulation Resistance Measurement
less
Marketed by:
RISHABH INSTRUMENTS PVT.LTD.
F-31, MIDC, Satpur, Nashik-422 007,India.
Tel.: +91 253 2202028, 2202202 Fax : +91 253 2351064
E-mail : India :- [email protected]
International :- [email protected]
www.rishabh.co.in
AIT 501 USERS MANUAL
AIT 501
AIT 501
AIT 501

Popular Test Equipment manuals by other brands

Fluke 714 instruction sheet

Fluke

Fluke 714 instruction sheet

Viavi HST-3000 Quick Card

Viavi

Viavi HST-3000 Quick Card

YOKOGAWA DLM2000 Series user manual

YOKOGAWA

YOKOGAWA DLM2000 Series user manual

iPA SMART MUTT 9004A Operator's manual

iPA

iPA SMART MUTT 9004A Operator's manual

Agilent Technologies 16197A Operation and service manual

Agilent Technologies

Agilent Technologies 16197A Operation and service manual

OfiTE 152-95 instruction manual

OfiTE

OfiTE 152-95 instruction manual

Mecmesin OmniTest-5.0 operating manual

Mecmesin

Mecmesin OmniTest-5.0 operating manual

Omega Engineering hot point CL900A Series user guide

Omega Engineering

Omega Engineering hot point CL900A Series user guide

V-TEK PT-55 user guide

V-TEK

V-TEK PT-55 user guide

National Instruments SCXI-116 Series quick start guide

National Instruments

National Instruments SCXI-116 Series quick start guide

Fnirsi SG-003A operating manual

Fnirsi

Fnirsi SG-003A operating manual

Siemens TD400 operating instructions

Siemens

Siemens TD400 operating instructions

Image Engineering LE7 x2 user manual

Image Engineering

Image Engineering LE7 x2 user manual

Bosch EPS 610 Original instructions

Bosch

Bosch EPS 610 Original instructions

SIGLENT SDS1000CFL Series user manual

SIGLENT

SIGLENT SDS1000CFL Series user manual

Atlona AT-DIS7-PROHD Q&A

Atlona

Atlona AT-DIS7-PROHD Q&A

Taylor Hobson Surtronic Duo 112-3115 manual

Taylor Hobson

Taylor Hobson Surtronic Duo 112-3115 manual

ZKETECH EBC-A40L user manual

ZKETECH

ZKETECH EBC-A40L user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.