Agilent Technologies 423B Service manual

The test & measurement
equipment you need at
the price you want.
Alltest carries the world’s largest selection of
used/refurbished benchtop test & measurement
equipment at 50% the price of new.
O
Our equipment is guaranteed working, warrantied, and
available with certified calibration from our in-house staff
of technicians and engineers.
• 10+ full time technicians with over 150 years of
specialization
• 90 day warranty & 5 day right of return on all
equipment
•
• 1-3 year warranties for new and
premium-refurbished equipment
• Every unit tested to OEM specifications
• Satisfaction guaranteed
You have plans, we will help you achieve them.
Any project. Any budget.
Get a quote today!
CCall (732) 919-3339 or email sales@alltest.net.
t
Alltest Instruments, Inc.
500 Central Ave.
Farmingdale, NJ 07727
P: (732) 919-3339
F: (732) 919-3332
alltest.net
s
sales@alltest.net

Agilent Technologies
423B and 8470B Detectors
Operating and Service
Manual
Part number: 00423-90103
Printed in USA
October 2003
Supersedes October 2002

ii Agilent 423B and 8470B Detectors
Notice
The information contained in this document is subject to change without
notice.
Agilent Technologies makes no warranty of any kind with regard to this
material, including, but not limited to, the implied warranties of
merchantability and fitness for a particular purpose. Agilent Technologies
shall not be liable for errors contained herein or for incidental or
consequential damages in connection with the furnishing, performance, or
use of this material.
Agilent Technologies assumes no responsibility for the use or reliability of
its software on equipment that is not furnished by Agilent Technologies.
This document contains proprietary information which is protected by
copyright. All rights are reserved. No part of this document may be
photocopied, reproduced, or translated to another language without prior
written consent of Agilent Technologies.
Restricted Rights Legend
Use, duplication, or disclosure by the U.S. Government is subject to
restrictions as set forth in subparagraph (c)(1)(ii) of the Rights in Technical
Data and Computer Software clause at DFARS 252.227-7013 for DOD
agencies, and subparagraphs (c)(1) and (c)(2) of the Commercial Computer
Software Restricted Rights clause at FAR 52.227-19 for other agencies.
Agilent Technologies
1400 Fountaingrove Parkway
Santa Rosa, CA 95403-1799, U.S.A.
© Copyright 1975, 1977, 1979, 2001-2003 Agilent Technologies, Inc.

Agilent 423B and 8470B Detectors iii
Warranty
Certification Agilent Technologies certifies that this product met its published
specifications at the time of shipment from the factory. Agilent Technologies
further certifies that its calibration measurements are traceable to the
United States National Institute of Standards and Technology (NIST,
formerly NBS), to the extent allowed by the Institute’s calibration facility,
and to the calibration facilities of other International Standards
Organization members.
Documentation
Warranty THE MATERIAL CONTAINED IN THIS DOCUMENT IS
PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED,
WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO
THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW,
AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS
OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY
INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT
LIMITED TO THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS
OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE FURNISHING, USE, OR
PERFORMANCE OF THIS DOCUMENT OR ANY
INFORMATION CONTAINED HEREIN. SHOULD AGILENT
AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT
WITH WARRANTY TERMS COVERING THE MATERIAL IN
THIS DOCUMENT THAT CONFLICT WITH THESE TERMS,
THE WARRANTY TERMS IN THE SEPARATE AGREEMENT
WILL CONTROL.
Assistance Product maintenance agreements and other customer assistance agreements
are available for Agilent Technologies products.
For assistance, call your local Agilent Technologies Sales and Service
Office.
Any adjustment, maintenance, or repair of this product must be performed
by qualified personnel. Contact your customer engineer through your local
Agilent Sales and Service Center.

iv Agilent 423B and 8470B Detectors
By internet, phone, or fax, get assistance with all your test & measurement needs.
Online assistance: www.agilent.com/find/assist
United States
(tel) 1 800 452 4844
Latin America
(tel) (305) 269 7500
(fax) (305) 269 7599
Canada
(tel) 1 877 894 4414
(fax) (905) 206 4120
Australia
(tel) 1 800 629 485
(fax) (61 3) 9210 5947
Europe
(tel) (31 20) 547 2323
(fax) (31 20) 547 2390
New Zealand
(tel) 0 800 738 378
(fax) 64 4 495 8950
Japan
(tel) (81) 426 56 7832
(fax) (81) 426 56 7840
Asia Pacific
(tel) (852) 3197 7777
(fax) (852) 2506 9284

Agilent 423B and 8470B Detectors v
Safety and Regulatory Information
Review this product and related documentation to familiarize yourself with
safety markings and instructions before you operate the instrument. This
product has been designed and tested in accordance with international
standards.
WARNING The WARNING notice denotes a hazard. It calls attention to a procedure,
practice, or the like, that, if not correctly performed or adhered to, could result
in personal injury. Do not proceed beyond a WARNING notice until the
indicated conditions are fully understood and met.
CAUTION The CAUTION notice denotes a hazard. It calls attention to an operating
procedure, practice, or the like, which, if not correctly performed or adhered
to, could result in damage to the product or loss of important data. Do not
proceed beyond a CAUTION notice until the indicated conditions are fully
understood and met.
Instrument Markings
When you see this symbol on your instrument, you should refer to the instrument’s
instruction manual for important information.
This symbol indicates hazardous voltages.
The laser radiation symbol is marked on products that have a laser output.
This symbol indicates that the instrument requires alternating current (ac) input.
The CE mark is a registered trademark of the European Community. If it is
accompanied by a year, it indicates the year the design was proven.
The CSA mark is a registered trademark of the Canadian Standards Association.
1SM1-A This text indicates thatthe instrument is an Industrial Scientific and Medical Group
1 Class A product (CISPER 11, Clause 4).
ICES/NMB-001 This ISM device complies with Canadian ICES-001. Cet appareil ISM est
conforme a la norme NMB-001 du Canada.
This symbol indicates that the power line switch is ON.
This symbol indicates that the power line switch is OFF or in STANDBY position.
!

Contents
Agilent 423 and 8470B Detectors vi
Notice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii
Restricted Rights Legend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Documentation Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Safety and Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Initial Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Storage and Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Operating Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Operator’s Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Performance Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Frequency Response Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Higher Level Sensitivity Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Low Level Sensitivity Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Match Test (SWR) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Service Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Detector Element Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Load Resistor Parts Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
7-mm Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

Agilent 423B and 8470B Detectors 1
Description
Description
The Agilent Technologies 423B and 8470B detectors are 50 Ω(nominal)
Schottky diode devices designed for measurement use in coaxial systems.
Features •The instruments convert the RF power levels applied to the 50 Ωinput
connector into proportional values of dc voltage.
•The instruments measure relative power up to 200 mW and have a BNC
female connector for the output jack which allows the detected output to
be connected to a SWR meter.
•The output voltage polarity is negative, unless Option 003 is selected.
•The frequency range of the 423B is 10 MHz to 12.4 GHz.
•The frequency range of the 8470B is 10 MHz to 18 GHz.
•AM modulation can be detected up to about 1 MHz.
Options The 423B and 8470B detectors are available with the following
non-exclusive options.
Option 001: Matched pair of detectors
Option 002: Furnished with matched load resistor for optimum
square law characteristics
Option 003: Positive polarity output
Option 012: Furnished with stainless steel type-N male
connectors (8470B only)
Connectors •The mating RF input connectors used with the 423B and 8470B
Option 012 must be type-N female connectors which comply with
U.S. military standard MIL-C-39012.
•The 8470B (standard) RF input connector must be an 7-mm connector.
•The mating output connector for the 423B and 8470B must be a BNC
male.

2 Agilent 423B and 8470B Detectors
Installation
Installation
Initial Inspection 1. Inspect the shipping container for damage. If the shipping container or
cushioning material is damaged, it should be kept until the contents of
the shipment have been checked for completeness and the instrument
has been checked mechanically and electrically. Procedures for
checking electrical performance are given under “Performance Tests” on
page 6.
2. If the contents are incomplete, if there is mechanical damage or defect,
or if the instrument does not pass the electrical performance test, contact
the nearest Agilent Technologies office. Any adjustment, maintenance,
or repair of this product must be performed by qualified personnel.
Contact your customer engineer through your local Agilent Sales and
Service Center. Refer to “Assistance” on page iii. Agilent Technologies
will arrange for repair or replacement of the damaged or defective
equipment. Keep the shipping materials for the carrier’s inspection.
3. If you are returning the instrument for service, repackaging the
instrument requires original shipping containers and materials or their
equivalents. Agilent can provide packaging materials identical to the
original materials. Any adjustment, maintenance, or repair of this
product must be performed by qualified personnel. Contact your
customer engineer through your local Agilent Sales and Service Center.
Refer to “Assistance” on page iii. Attach a tag indicating the type of
service required, return address, model number and serial number. Mark
the container FRAGILE to insure careful handling. In any
correspondence, refer to the instrument by model number and serial
number.
Storage and Shipment The instrument should be stored in a clean, dry environment. The following
environmental limitations apply to both storage and shipment:
a. Temperature: −54 to +85 °C
b. Altitude < 7620 meters (25,000 feet)
c. Humidity < 95% relative
d. Shock 100 G for 11 ms
e. Vibration 20 G from 80 to 2000 Hz

Agilent 423B and 8470B Detectors 3
Installation
Specifications
Frequency Range:Options:
423B: 10 MHz to 12.4 GHz
8470B: 10 MHz to 18 GHz
Note:RFmayleakthroughthe outputconnectorbelow1GHz. Itcan
be reduced, if objectionable, with a suitable low pass filter.
423B and 8470B:
Option 002: Furnished with matched load resistor for optimum
square law characteristics of 25 °C, within ±0.5 dB from square
law over a range of at least 30 dB up to 10 mV peak output,
working into an external load > 8 k Ω. Sensitivity typically is
greater than 0.1 mV/µW when load resistor is used.
Option 003: Positive polarity output
423B:
Option 001: Matched detector pair. Frequency response
characteristics (exclusiveof basic sensitivity) track within
±0.2 dB from 10 MHz to 12.4 GHz at −30 dBm.
8470B:
Option 001: Matched detector pair. Frequency response
characteristics (exclusiveof basic sensitivity) track within
±0.2 dB from 10 MHz to 12.5 GHz;
±0.3 dB from 12.4 GHz to 18 GHz at −30 dBm.
Option 012: Furnished with stainless steel type-N male connector.
Frequency Response:4, 5
423B and 8470B:
Octave flatness:
±0.2 dB over any octave 10 MHz to 8 GHz
Broadband flatness:
±0.3 dB 10 MHz to 12.4 GHz
8470B only:
Broadband flatness:
±0.5 dB 10 MHz to 15 GHz
±0.6 dB 10 MHz to 18 GHz
SWR1,2 (Max):
423B and 8470B:
10 MHz to 4 GHz, 1.15
4 GHz to 12.4 GHz, 1.3
8470B only:
12.4 GHz to 15 GHz, 1.3
15.0 GHz to 18 GHz, 1.7
Environmental:
423B:
Operating temperature: 0 C to +55 °C
Humidity: < 95% relative
Vibration: 20 G from 80 to 2000 Hz
Shock: 100 G for 11 ms
Altitude: < 4570 meters (15,000 feet)
8470B:
Operating temperature: −20 to +85 °C
Humidity: < 95% relative
Vibration: 20 G from 80 to 2000 Hz
Shock: 100 G for 11 ms
Altitude: < 4570 meters (15,000 feet)
Maximum Operating Input Power:
200 mW, peak or average
Maximum Short Term Input Power:
1 watt (typical) peak or average for < 1 minute
Sensitivity:1, 3, 5
High level: < 0.35 mW produces 100 mV output
Low level: > 0.5 mVdc/µW CW
Input Impedance: 50 Ω(nominal)
Output Impedance: 1 to 2 k Ω(typically 1.3 k Ω) shunted by
35 to 65 pF (typically 50 pF).
Output Polarity: Negative (refer to options for positive polarity units)
Detector Element: Supplied (refer to “Replacement Parts Table” on
page 12 for replacement elements)
Bias: Not required
Noise: < 50 µVpp with CW applied to produce 100 mV output
General:
Weight:
423B and 8470B: Net 114 g (4 oz.)
Dimensions:
423B: 63 mm long, 20 mm diameter (2.47 in. long,
0.78 in. diameter)
8470B: 64 mm long, 19mm diameter (2.50 in. long,
0.75 in.diameter)
1. Specifications given for +25 °C unless otherwise noted.
2. Measurement made at −20 dBm.
3. Sensitivity decreases with increasing temperature, typically: 0.5 dB from −20 to +25 °C; 0.5 dB from +25 to +40 °C;
1 dB from +40 to +55 °C; 1.25 dB from +55 to +75 °C; 1 dB from +75 to +85 °C;
4. Measured at −30 dBm. Flatness is the peak-to-valley range of the response.
5. External load resistance > 50 k Ω

4 Agilent 423B and 8470B Detectors
Operating Information
Operating Information
The detector can be used as a demodulator to obtain a pulse envelope which
can then be observed on an oscilloscope. It can also be used as a general
purpose detector.
CAUTION Static discharge can damage the detector element. A 100 pF capacitor
(1.2 m [4 ft] of coax cable) charged to 14 volts stores .1 erg, the maximum
pulse rating of the detector element.
•Connect cables to test equipment and discharge the center conductor
before connecting to the detector.
•DO NOT NEEDLESSLY HANDLE THE DETECTOR ELEMENT
USED IN CRYSTAL DETECTOR. Static electricity which builds up on
a person, especially on a cold dry day, must never be allowed to
discharge through the detector.
•Avoid exposed leads to or from the detector, since these are often
touched accidentally.
The power applied to the detector can be either modulated or continuous
wave (CW). If modulated at a 1000 Hz rate, an SWR meter can be used as an
indicator. For CW detection, a dc milliammeter or millivoltmeter can be
used as the indicator.
NOTE When using the detector with an oscilloscope, and the waveshapes to be
observed have rise times of less than 5µs, the coaxial cable connecting
oscilloscope and detector should be as short as possible and shunted with a
resistor.
Ideally, this resistor should be 50 Ω to terminate the coaxial cable properly.
However, with 50 Ωresistance, the output video pulse may be too small to
drive some oscilloscopes. Therefore, the cable should be shunted with the
smallest value of resistance that will obtain suitable deflection on the
oscilloscope; typically the value will lie between 50Ωand 2kΩ. The larger
the resistance the more degradation of rise time.

Agilent 423B and 8470B Detectors 5
Operating Information
Operator’s Check Peak Power Measurement
The procedure for peak power measurement involves calibration of an
oscilloscope which, in turn, is used to calibrate a CW generator. The output
of the calibrated CW generator is measured with a power meter; the peak
power of a pulse is thereby measured. The procedure is as follows:
1. Connect equipment as shown in Figure 1, step A. Observe the pulse on a
de-coupled oscilloscope. Using a marking pencil, mark on the graticule
the base-to-peak amplitude of the pulse envelope.
2. Replace the pulse source with a CW generator. While observing the
oscilloscope trace, adjust the amplitude of CW generator output to make
detector’s output equal to that of pulse generator, as indicated by
markings on the graticule (step A).
3. Leave the CW generator at setting obtained in step B. Disconnect the
detector from CW generator.
4. Connect the output of CW generator to the power meter (step C).
Measure the adjusted levels (set in step B) of the CW generator output.
The peak power of the pulse envelope observed in step A is equal to the
output power of the CW generator.
Figure 1 Peak Power Measurement
Reflectometer Application
For information about reflectometer systems and measurements refer to
Agilent Technologies Application Note Index. Copies are available upon
request.
Harmony Frequency Comparison Measurement Application
The detector can be used as a mixer in harmonic-frequency comparison
measurements. Refer to Agilent Technologies Application Note Index for
further information.

6 Agilent 423B and 8470B Detectors
Performance Tests
Performance Tests
Methods for testing detector specifications are given below. Refer to the
manuals of the equipment involved for operating instructions.
NOTE Multiple mismatch errors caused by attenuator SWR, power meter SWR,
and detector SWR should be taken into account, as well as the accuracy of
the indicator used to measure the detector’s output.
Frequency
Response Test 1. Using signal sources covering 10 MHz to 18 GHz with a 10 dB isolating
attenuator and a power meter, connect the power sensor to the attenuator.
Adjust the RF power level to −20 dBm input to the power sensor.
2. Without changing the RF power level of signal source, disconnect the
power sensor.
3. Connect the detector to the attenuator. Measure the dc voltage output
and record the measurement.
4. Change the frequency of the signal source and repeat steps 1 through 3.
5. Since the detector follows a square-law response at this power level, its
output is proportional to power (PdB = 10 log Vo). Total variation of
detector readings should meet specifications for all frequencies of
interest across the band.
Higher Level
Sensitivity Test 1. Using signal sources covering 10 MHz to18 GHz and a dc voltmeter or
oscilloscope as the indicator, connect the detector to the signal source.
Adjust the RF power level for a 100 mV detected output from the
detector, using a CW signal.
2. Disconnect the detector from the signal source and measure the RF
output level. The RF output level should be ≤0.35 mW.
3. Repeat steps 1 and 2 for all frequencies of interest across the band.
Low Level
Sensitivity Test 1. Using a signal source (covering 10 MHz to 18 GHz), a 10 dB attenuator,
and a power meter, connect the attenuator to the signal source and power
sensor to the attenuator. Adjust the RF power level for −20 dBm output
from the attenuator. Verify the ambient temperature.
2. Disconnect the power sensor from the attenuator and connect the
detector. Measure the dc voltage output from the detector. The output
shouldbe>5.0mVat25×°C.Between20×°Cand30×°C,thesensitivity
slope is typically −0.04 dB/× °C.
Match Test (SWR) 1. To verify the detector’s SWR specifications, use any system whose
measurement accuracies for SWR (residual SWR) are known.

Agilent 423B and 8470B Detectors 7
Service Instructions
Service Instructions
Repair and replacement information for the 423B and 8470B detectors and
Option 002 load resistor are given below. Any adjustment, maintenance, or
repair of this product must be performed by qualified personnel. Contact
your customer engineer through your local Agilent Sales and Service Center.
You can find a list of Agilent Sales and Service Centers on page iv. The
detectors have no internal adjustments.
Replaceable Parts Part numbers for replaceable parts are given in “Replacement Parts Table”
on page 12. To order a replacement part, address your order or inquiry to the
nearest Agilent Technologies office. Any adjustment, maintenance, or repair
of this product must be performed by qualified personnel. Contact your
customer engineer through your local Agilent Sales and Service Center.
Refer to “Assistance” on page iii. Include the model number, Agilent
Technologies part number, and a description.
Detector Element
Replacement The detector element assembly includes only a detector element. However,
Option 002 does include a resistor for the load resistor which is used to load
the diode for square-law operation.
CAUTION Read the following precautions carefully before performance of any
operation with the detector element when it is out of either the housing or the
detector element shipping container.
•Before installing the diode into the mount, touch the exposed metal on
the mount with your hand to discharge static electricity.
•When handing the diode to another person, first touch hands to ensure
that there is no difference in static electricity potential between you.
•Ohmmeters should not be used to measure forward- and back-resistance
since it is easy to damage these diodes. The ohmmeter’s open-circuit
voltages and short-circuit currents may cause problems.
1. Remove the female BNC connector and compression spring from inside
of the connector cap.
2. Remove the connector cap from the body. Use a pair of pliers with
plastic jaws or protect the body with heavy paper or tape.

8 Agilent 423B and 8470B Detectors
Service Instructions
CAUTION •Do not rotate the detector element while inserting or removing as
damage may result.
•When inserting the detector element, do not force the large pin end into
the center conductor.
•The detector element might be damaged if the detector element is not
centered.
3. Remove the old detector element, axial spacer, and RF washer and
discard the detector element.
4. Install the RF washer, axial spacer, and new detector element. Refer to
the illustrations on page 11 for proper orientation for the internal
components.
a. Insert the RF washer first.
b. Then carefully insert the large pin end of the detector element into
the center contact inside the detector body.
c. Place the axial spacer over the small pin end of the detector element.
5. Carefully place the connector cap over the body and assembled
components and tighten firmly in place.
6. Place the compression spring into the center conductor of the female
BNC connector. Carefully start the spring over the small pin of the
detector element through the hole in the connector cap. Keep the spring
in the BNC connector and screw the female BNC connector firmly into
place.
Load Resistor Parts
Replacement To replace the male BNC connector
Parts referred to in the following procedures are identified in Figure 2 on
page 9 and Figure 3 on page 10.
1. Remove the male BNC connector and lock washer from the housing.
Use a 3/8-inch open-end wrench and hold the housing either in a vise or
with pliers. Be sure to protect the housing of the load resistor with
material such as heavy paper or tape or use plastic jaws on the vise or
pliers.
2. Unsolder the resistor.
3. Solder the resistor to the new BNC connector.
4. Let the resistor cool and then check the resistance from male BNC pin
through the resistor. The resistance measurement should be ± 10% than
indicated by the color coding.
5. Replace the lockwasher and male BNC connector.

Agilent 423B and 8470B Detectors 9
Service Instructions
To replace the female BNC connector
1. Remove the female BNC connector. Refer to Figure 3 for parts
identification.
To remove or install the BNC connector, use a BNC wrench or use a
male BNC connector as a wrench to prevent damage to the connector.
2. Unsolder the contact spring.
3. Prepare the replacement female BNC connector:
a. Cut the center conductor lead to approximately 0.79 mm (1/32 in.)
b. With a flat file, smooth the end of the lead; remove burr with
tweezers of similar metal instrument.
4. Slip the contact spring over the center conductor lead, and solder.
CAUTION Use solder sparingly or it will creep back on the spring. Solder on the spring
will destroy its usefulness and it is difficult to remove.
5. Let the contact spring cool and then screw the connector into the mount.
To replace the 7-mm connector center contact
1. Refer to “7-mm Connectors” on page 13 for assembly and disassembly
instructions.
2. Refer to “Replacing the 7-mm Center Contact” on page 14 for the
center contact replacement procedure.
Figure 2 Cutting Center Conductor Lead to Accommodate Contact Spring

10 Agilent 423B and 8470B Detectors
Service Instructions
Figure 3 Load Resistor, Cutaway View

Agilent 423B and 8470B Detectors 11
Service Instructions
Figure 4 8470B and 8470B Option 012 Detector Assemblies
Figure 5 423B Detector

12 Agilent 423B and 8470B Detectors
Service Instructions
Replacement Parts Table
Description Part Number Description Part Number
423B Assembly Connector, RF 7 mm 1250-0909
Contact, RF connector 1250-0014 Spring, compression 1460-2606
Gasket, RF connector 1250-0015 Washer, lock internal 2190-0016
Ring, RF connector 1250-0016 Cap, connector 5020-0210
Connector, female BNC 1250-0212 Washer, cap 5021-0127
Nut, RF connector 1250-0918 Insulator 5040-0306
Spring, compression 1460-2606 Conductor, center 08470-20002
Washer, lock internal 2190-0016 Ring, RF connector (Opt 012) 1250-0016
Bead, connector 5020-0207 Connector, RF APC (Opt 012) 1250-0916
Cap, connector 5020-0210 Contact, RF connector (Opt 012) 1250-0917
Washer, cap 5021-0127 Nut, RF connector (Opt 012) 1250-0918
Conductor, center 00423-201 Replacement Diode Assemblies1
Body, crystal mount 00423-202
Replacement Diode Assemblies1
Single diode negative polarity 00423-60003
Single diode negative polarity
Single diode positive polarity (Opt 003)
Single diode negative polarity with
matching load resistor (Opt 002)
Single diode positive polarity with
matching load resistor (Opt 002)
08470-60012
08470-60013
08470-60014
08470-60015
Single diode positive polarity (Opt 003) 00423-60004
Single diode negative polarity with
matching load resistor (Opt 002) 00423-60005
Matched pair diodes negative polarity
(Opt 001) 00423-60007
Load Resistor Assembly 8470B 7-mm
Connector Assembly
Connector, male BNC 1250-0045 Inner conductor contact mechanism
(assembled contact and outer body) 1250-0816
Connector, female BNC 1250-0251 Coupling nut 1250-0918
Spring, contact 5000-0234 Coupling sleeve 1250-0820
Housing 5020-3215 Inner conductor contact 1250-0907
8470B and 8470B-012 Assemblies1
Connector, female BNC 1250-0212
Support bead 5040-0306
Tools
7-mm contact extractor
7-mm spanner wrench
Open-end wrench
5060-0370
5060-0237
8710-0877
Contact, RF connector 1250-0816
1. Refer to page 1 for a description of options.

Agilent 423B and 8470B Detectors 13
Service Instructions
7-mm Connectors
Figure 6 Use and Care of 7-mm Connectors
To Connect 1. On one connector, retract the coupling sleeve by turning the coupling
nut counterclockwise until the sleeve and nut disengage.
2. On the other connector, fully extend the coupling sleeve by turning the
coupling nut clockwise. To engage coupling sleeve and coupling nut
when the sleeve is fully retracted, press back lightly on the nut while
turning it clockwise.
3. Push the connectors firmly together, and thread the coupling nut of the
connector with retracted sleeve over the extended sleeve.
4. Close the gap between coupling nuts with the nut on the extended-sleeve
connector.
To Disconnect 1. Loosen the coupling nut of the connector showing the wider gold band.
2. IMPORTANT: Part the connectors carefully to prevent striking the inner
conductor contact.
This manual suits for next models
1
Table of contents