
Tip 2: Triggering
The 34450A has a triggering function
that allows you to initiate a measure-
ment at desired conditions. The
triggering feature is used through the
front panel during manual operation,
while for automated operation the
DMM is controlled through the USB,
GPIB, or serial connector. Setting the
34450A to be triggered must be done
in the proper sequence to ensure
the DMM starts measuring once
the triggering signal is received. The
DMM’s measurement function, range,
resolution, and speed are among the
first few settings that need to be
set. Leaving the DMM in auto-range
reduces the overall speed of each
measurement because an additional
algorithm is used to search, compare,
and set the right measurement range.
Manually setting the measurement
range is always a good practice when
the measurement values are known.
When the DMM is to be used in auto-
mated or remote mode, it can be set
to receive a triggering signal either
through the software or hardware.
Software triggering (also known as
bus triggering) is achieved by sending
standard commands for programma-
ble instruments (SCPI) commands to
the DMM over a USB, GPIB, or serial
connection. This triggering method
is preferred in software-controlled
environments where the test program
determines the measurement timing
by sending a *TRG command to the
DMM. If your application requires
data analysis or the comparison of
values before making a decision
to continue a measurement on the
DMM, using a software trigger is the
best approach.
Using hardware triggering is recom-
mended when using the DMM to
make one or a specified number of
measurements upon receiving a pulse
at the TRIG IN pin on the serial con-
nector. Hardware triggering through
the serial connector is best suited for
manufacturing environments where
instruments or mechanical parts, such
as handler and PLC, provide a pulse to
the DMM to trigger a measurement.
Using the TRIG IN pin of the 34450A
is a good choice if a triggering pulse
is available in the test environment
and throughput is one of the major
concerns. Figure 2 refers to the
serial connector at the rear panel of
the 34450A. The triggering signal is
applied to the TRIG IN and Ground
pin of the serial connector as denoted
by a positive and negative sign
respectively.
Figure 3. Types of triggering methods on 34450A.
In manual operation or local mode,
the triggering is meant to initiate a
single measurement or start the data
logging using the timing you define.
This triggering method is initiated by
pressing [Trig]. To manually trigger a
single measurement, set the DMM’s
triggering source to Manual/External
in the Utility menu. The triggering
method for data logging is explained
in the Tip 3 section of this application
note. Figure 3 outlines the triggering
operations on 34450A with criteria for
your selection reference.
Triggering Operation
Manual
Software
Automated
Hardware
• Measure upon receive
*TRG command.
• Best for software-
controlled environment.
• Use with data logging
or initiate a single
measurement.
• Control via front
panel.
• Required pulses to
be triggered.
• Utilize TRIG IN pin
at the serial port.
• Higher throughput.
Tx Rx Trig In Trig Out
Figure 2. TRIG IN pin on the serial connector of the 34450A.
3