FIELD INSTALLATION
INSTRUCTIONS

Wireless Remote Base Unit for VS Series and G8610, G8611, or G8612 Series Leak Detectors
This page intentionally left blank.

Agilent Field Service Sheet
Wireless Remote Base Unit for VS Series and G8610, G8611, or G8612 Series Leak Detectors
Page 1of 14
Document No.
699910002
Revision
D
Wireless Remote Base Unit for VS Series and G8610, G8611, or G8612 Series Leak Detectors
Preface
Document Standards
This manual uses the following documentation standards:
NOTE
Notes contain important information.
CAUTION
Cautions appear before instructions, which if not followed, could cause damage to
the equipment or data loss.
WARNING
Warnings appear for a particular procedure or practice which, if not followed correctly,
could lead to serious injury or death.
Hazard and Safety Information
The common international symbols used in this manual and on the equipment are defined below.
O
OFF Supply (Power)
Earth (Ground) Terminal
|
ON Supply (Power)
Caution, Hot Surface
AC –Alternating Current
Caution, Risk of Electrical
Shock
Warning, Risk of danger
Protective Conductor
Terminal
Frame or chassis Terminal

Wireless Remote Base Unit for VS Series and G8610, G8611, or G8612 Series Leak Detectors
Page 2of 14
Document No.
699910002
Revision
D
Operators and service personnel must be aware of all hazards associated with this equipment. They must know how to
recognize hazardous and potentially hazardous conditions, and know how to avoid them. The consequences of
unskilled, improper, or careless operation of the equipment can be serious. Every operator or service person must read
and thoroughly understand operation/maintenance manuals and any additional information provided by Agilent. All
warning and cautions must be read carefully and strictly observed. Consult local, state, and national agencies
regarding specific requirements and regulations. Address any safety, operation, and/or maintenance questions to your
nearest Agilent office.
NOTE
To clean the leak detector plastic enclosure, the LCD display and Front Panel
buttons, use only a soft cloth slightly dampened with water or a mild soap.
Do NOT use excess water or cleaning solvents of any kind.
Avoid splashing any liquids into the unit through the ventilation openings or
Front Panel buttons. Wipe the surface with a dry lint-free cloth.
CAUTION
The PCB’s are static sensitive devices. Wear a grounding strap when performing
any maintenance on these units and especially when performing maintenance of
static sensitive parts.
This manual includes two sections. The first gives instructions for installation with a VS series leak detector. The second
section gives instructions for operation with a new G8610, G8611, or G8612 series leak detector.

Wireless Remote Base Unit for VS Series and G8610, G8611, or G8612 Series Leak Detectors
Page 3of 14
Document No.
699910002
Revision
D
Section One: Installation on VS Series Leak Detector
Equipment Required
Kit Includes
Extended Length M5 Allen Wrench
Wireless Base PCB
Metric Allen Wrench Set
Wireless Remote Label
M3 Philips Head Screw Driver
RS232 Cable
Adjustable Wrench
Antenna
M3 Screw (Qty. 4)
Installation Procedure
For clarity, some items have been omitted from views.
Figure 1: Rear and Front Cover Screws
WARNING
Disconnect power from the unit before performing any maintenance
procedure that requires physically disconnecting any part of the system.
NOTE
Prior to wireless PCB installation, ensure that the software is revision
4.02 or higher by navigating from the Home screen menu to System
Information. If not, contact Agilent customer services, see the back
cover of this manual for a listing of our sales and service offices.

Wireless Remote Base Unit for VS Series and G8610, G8611, or G8612 Series Leak Detectors
Page 4of 14
Document No.
699910002
Revision
D
1) Turn off the power switch located on the back of the unit and unplug.
2) Wait 30 seconds for the high voltage to dissipate.
3) Using an extended length M5 Allen wrench, remove the four screws holding the rear plastic cover (Figure 1: Rear
and Front Cover Screws) and detach the rear plastic cover from the unit.
4) Remove the four screws holding the front plastic cover and detach the cover from the unit. Two screws are
situated at the front of the unit (not shown) and two screws are positioned inside the unit (Figure 1: Rear and
Front Cover Screws).
Figure 2: Fan Assembly and Socket Head Cap Screw
5) Disconnect the fan connector from the 24 V cable harness.
6) Detach the fan assembly by removing the M4 socket head cap screw and loosening the two wing nuts (Figure 2:
Fan Assembly and Socket Head Cap Screw). Place the fan assembly on a flat surface on its back.
Figure 3: Wireless Antenna
7) Hand tighten the antenna to the fan bracket, and then apply a ½ turn using an adjustable wrench (Figure 3:
Wireless Antenna).

Wireless Remote Base Unit for VS Series and G8610, G8611, or G8612 Series Leak Detectors
Page 5of 14
Document No.
699910002
Revision
D
Figure 4: Base Board Assembly with Posts
Figure 5: Antenna Connector
CAUTION
Use proper ESD safety techniques when performing a PCB card install.

Wireless Remote Base Unit for VS Series and G8610, G8611, or G8612 Series Leak Detectors
Page 6of 14
Document No.
699910002
Revision
D
8) Place the wireless board assembly on the four fan bracket standoffs. Note the correct orientation of the wireless
base board P300 connector (Figure 4: Base Board Assembly with Posts).
9) Connect the antenna cable to the U.FL (Color: silver) on the wireless base board (Figure 5: Antenna Connector).
Carefully center the connector on the jack and press the connector onto the jack until it snaps into place. Rotate
the antenna cable to ensure the connector is firmly seated in the jack.
10) Secure the wireless base board to the fan bracket using the four M3 Phillips screws provided with the kit.
11) Unfurl the black/red wire (P1-RF) with the 24 V three-pin connector and the ground lug (Figure 6: 24 V Power
Cable), which is tie rapped at the base of the leak detector, and:
• Plug the 24 V connector labeled P1-RF into the three-pin connector on the wireless board.
• Attach the 24 V ground lug to one of the bracket screws.
• Reinstall the fan assembly by tightening the M4 socket head cap screw and the two wing nuts (Figure 2: Fan
Assembly and Socket Head Cap Screw).
Figure 6: 24 V Power Cable
12) Connect the RS-232 cable end labeled J300-1 to P300 on the wireless board using its two slotted screws (Figure
5: Antenna Connector).

Wireless Remote Base Unit for VS Series and G8610, G8611, or G8612 Series Leak Detectors
Page 7of 14
Document No.
699910002
Revision
D
Figure 7: Front Panel Display
13) Dress the RS-232 cable through the leak detector to the front panel display.
14) Remove the M6 screw and washer from the front panel display and detach the display (Figure 7: Front Panel
Display).
15) Connect the RS-232 cable end labeled J300-2 to P300-COM2 on the front panel display board using its two
slotted screws (Figure 8: RS232 Cable).
Figure 8: RS232 Cable
16) Reattach the display by inserting the tabs into their slotted positions and secure using the M6 screw and washer.
17) Attach the front cover and secure to the frame using existing hardware.

Wireless Remote Base Unit for VS Series and G8610, G8611, or G8612 Series Leak Detectors
Page 8of 14
Document No.
699910002
Revision
D
18) Rotate the antenna to its horizontal locked position.
19) Guide the antenna through its location hole in the rear cover. Then secure the rear cover to the frame using
existing hardware.
Figure 9: Label
20) Attach the label to the rear cover in location shown (Figure 9: Label).
21) Connect the power cord and power up the unit.
22) Watch the home screen to verify the Spectube Pressure Wait message progresses to Stabilization Wait and
System Ready within ten minutes.
Refer to the operator's manual if the system fails to reach the System Ready mode.
23) After installation refer to the VS Series leak detector manual (P/N # 699909942) Section 3, for leak detector
activation of the wireless remote.
24) Refer to the wireless remote user's manual (P/N # 699909945) for operational instructions.
25) Agilent recommends a full calibration of the unit prior to leak test operations.

Wireless Remote Base Unit for VS Series and G8610, G8611, or G8612 Series Leak Detectors
Page 9of 14
Document No.
699910002
Revision
D
Section Two: Installation on G8610, G8611, or G8612 Series Leak Detector
Equipment Required
Kit Includes
Slotted Screw Driver
Wireless Base PCB
Metric Allen Wrench Set
Wireless Remote Label
M3 Philips Head Screw Driver
RS232 Cable
Adjustable Wrench
Antenna
M3 Screw (Qty. 4)
Installation Procedure
For clarity, some items have been omitted from views.
WARNING
Disconnect power from the unit before performing any maintenance
procedure that requires physically disconnecting any part of the system.
1) Turn off the power switch located on the back of the unit and unplug.
2) Wait 30 seconds for the high voltage to dissipate.
Figure 1: Orientation
3) Using slotted screw driver, unfasten the four screws holding the front, rear and right side covers (Figure 1:
Covers) and detach. The left side cover is secured with hex-head screws.
Front
Right Side
Left Side (Not Shown)
Rear (Not Shown)

Wireless Remote Base Unit for VS Series and G8610, G8611, or G8612 Series Leak Detectors
Page 10 of 14
Document No.
699910002
Revision
D
Figure 2: Covers
4) Rotate the lid 90° to give better access to the area where the antenna will be installed (Figure 2: Lid).
Figure 2: Removing Valve
5) If the leak detector configuration has a valve in front of the bottom left post, move that valve aside by unscrewing
the two screws circled (Figure 2: Removing valve). Keeping the wiring connected, set valve aside. This valve is
only used on certain models of leak detectors. If the valve is not present, ignore this step.
PCB to mount to
opposite side of this
aluminum support

Wireless Remote Base Unit for VS Series and G8610, G8611, or G8612 Series Leak Detectors
Page 11 of 14
Document No.
699910002
Revision
D
6) Mount the printed circuit board to the four posts, noting the correct orientation. Fasten with screws (Figure 3. Base
Board Assembly with Posts).
Figure 3: Base Board Assembly with Posts
7) Using the provided washer and nut, hand tighten the antenna to the outside of the leak detector wall, and then
apply a ½ turn using an adjustable wrench (Figure 4: Wireless Antenna).
Figure 4: Wireless Antenna
O-ring on Outside (Not Shown)
Washer
Nut

Wireless Remote Base Unit for VS Series and G8610, G8611, or G8612 Series Leak Detectors
Page 12 of 14
Document No.
699910002
Revision
D
Figure 5: Antenna Connector
CAUTION
Use proper ESD safety techniques when performing a PCB card install.
8) Connect the antenna cable to the U.FL on the wireless base board (Figure 5: Antenna Connector). Carefully
center the connector on the jack and press the connector onto the jack until it snaps into place. Rotate the
antenna cable to ensure the connector is firmly seated in the jack.
9) Connect the RS-232 on the wireless board using its two slotted screws (Figure 6: RS232 Connection, left side).
Plug in the other side of the ribbon cable to the green port on the CPU board, located on the front side of the leak
detector, away from the antenna's printed circuit board (Figure 6: RS232 Connection, right side).
Figure 6: RS232 Connection
U.FL Jack

Wireless Remote Base Unit for VS Series and G8610, G8611, or G8612 Series Leak Detectors
Page 13 of 14
Document No.
699910002
Revision
D
10) Unfurl the black/red wire (P1-RF) with the 24 V three-pin connector and the ground lug (Figure 7: 24 V Power
Cable), which is tie rapped at the base of the leak detector, and:
•Plug the 24 V connector labeled P1-RF into the three-pin connector on the wireless board.
•Attach the 24 V ground lug to one of the bracket screws.
Figure 7: 24 V Power Cable
11) Reinstall the valve and dress internal wires inside of leak detector (Figure 2: Removing Valve).
12) Reattach front and side (Figure 1: Covers).
13) Rotate the antenna to its horizontal locked position.
14) Guide the antenna through its location hole in the rear cover. Then secure the rear cover to the frame using
existing hardware.
Base PCB 24 V
Power Cable
Ground Screw
Power In

Wireless Remote Base Unit for VS Series and G8610, G8611, or G8612 Series Leak Detectors
Page 14 of 14
Document No.
699910002
Revision
D
Figure 9: Label
15) Attach the label to the rear cover in location shown (Figure 9: Label). This new label will help service identify the
equipment used. Do not cover the existing label on the leak detector.
16) Connect the power cord and power up the unit.
17) Watch the home screen to verify the Spectube Pressure Wait message progresses to Stabilization Wait and
System Ready within ten minutes.
Refer to the operator's manual if the system fails to reach the System Ready mode.
18) After installation refer to the Leak Detector manual section for leak detector activation of the wireless remote.
19) Refer to the wireless remote user's manual (P/N # 699909945) for operational instructions.
20) Agilent recommends a full calibration of the unit prior to leak test operations.
New Label Location

Page 1of 3Agilent VPD Request for Return
EUROPE:
Fax: 00 39 011 9979 330
Fax Free: 00 800 345 345 00
Toll Free: 00 800 234 234 00
NORTH AMERICA:
Fax: 1 781 860 9252
Toll Free: 800 882 7426
PACIFIC RIM:
please visit our website for individual
office information
http://www.agilent.com
Vacuum Products Division
Instructions for returning products
Dear Customer,
Please follow these instructions whenever one of our products needs to be returned.
Complete the attached Request for Return form and send it to Agilent Technologies (see below), taking particular care to include the
completed Health and Safety declaration Section. No work can be started on your unit until we receive a completed copy of this
form.
After evaluating the information, Agilent Technologies will provide you with a Return Authorization (RA) number via email or fax, as
requested. Note: Depending on the type of return, a Purchase Order may be required at the time the Request for Return is submitted. We will
quote any necessary services (evaluation, repair, special cleaning, eg).
Product preparation
oRemove all accessories from the core product (e.g. inlet screens, vent valves).
oPrior to shipment and if applicable for your product, drain any oils or other liquids, purge or flush all gasses, and wipe off any
excess residue.
oIf ordering an Advance Exchange product, please use the packaging from the Advance Exchange to return the defective
product.
oSeal the product in a plastic bag, and package product carefully to avoid damage in transit. You are responsible for loss or
damage in transit.
oInclude a copy of the Health and Safety Declaration in the shipping documentation on the outside of the shipping box of your
returning product.
oClearly label package with RA number. Using the shipping label provided will ensure the proper address and RA number are on
the package. Packages shipped to Agilent without a RA clearly written on the outside cannot be accepted and will be returned.
oReturn only products for which the RA was issued.
Shipping
oShip to the location specified on the printable label, which will be sent, along with the RA number, as soon as we have
received all of the required information. Customer is responsible for freight charges on returning product.
oReturn shipments must comply with all applicable Shipping Regulations (IATA, DOT, ADR, etc.) and carrier requirements.
RETURN THE COMPLETED REQUEST FOR RETURN FORM TO YOUR NEAREST LOCATION:

Page 2of 3
Vacuum Products Division
Terms and conditions
TERMS AND CONDITIONS
Please read the terms and conditions below as they apply to all returns and are in addition to the Agilent Technologies
Vacuum Product Division – Products and Services Terms of Sale.
oUnless otherwise pre-negotiated, customer is responsible for the freight charges for the returning product. Return shipments
must comply with all applicable Shipping Regulations (IATA, DOT, etc.) and carrier requirements.
oAgilent Technologies is not responsible for returning customer provided packaging or containers.
oCustomers receiving an Advance Exchange product agree to return the defective, rebuildable part to Agilent Technologies
within 15 business days. Failure to do so, or returning a non-rebuildable part (crashed), will result in an invoice for the non-
returned/non-rebuildable part.
oReturns for credit toward the purchase of new or refurbished Products are subject to prior Agilent approval and may incur a
restocking fee. Please reference the original purchase order number.
oUnits returned for evaluation will be evaluated, and a quote for repair will be issued. If you choose to have the unit repaired,
the cost of the evaluation will be deducted from the final repair pricing. A Purchase Order for the final repair price should be
issued within 3 weeks of quotation date. Units without a Purchase Order for repair will be returned to the customer, and the
evaluation fee will be invoiced.
oProducts returned that have not been drained from oil will be disposed.
oA Special Cleaning fee will apply to all exposed products
oIf requesting a calibration service, units must be functionally capable of being calibrated.

Page 3of 3
Vacuum Products Division
Request for Return Form
Customer information
Company : Contact Name:
Address: Tel: Fax:
Email:
Equipment
Product description Agilent PartNo Agilent Serial No Original Purchasing Reference
Failure description
Type of process (for which the equipment was used)
Type of return
Non Billable Billable New PO # (hard copy must be submitted with this form): ______________________________________
Exchange Repair Upgrade Consignment/Demo Calibration Evaluation Return for Credit
Health and safety
The product has been exposed to the following substances:
(by selecting ‘YES’ you MUST complete the table to the right)
Substances (please refer to MSDS forms)
*Agilent will not accept delivery of any product that is exposed to radioactive, biological,
explosive substances or dioxins, PCB’s without written evidence of decontamination.
Trade name Chemical name Chemical Symbol CAS Number
Toxic YES NO
Harmful YES NO
Corrosive YES NO
Reactive YES NO
Flammable YES NO
Explosive (*) YES NO
Radioactive (*) YES NO
Biological (*) YES NO
Oxidizing YES NO
Sensitizer YES NO
Other dangerous substances YES NO
Goods preparation
If you have replied YES to one of the above questions. Has the product been purged?
YES NO
If yes, which cleaning agent/method:
Has the product been drained from oil? YES NOT APPLICABLE
I confirm to place this declaration on the outside of the shipping box.
I declare that the above information is true and complete to the best of my knowledge and belief.
I understand and agree to the terms and conditions on page 2 of this document.
Name: Authorized Signature:
Position:
Date:
NOTE: If a product is received at Agilent which is contaminated with a toxic or hazardous material that was not disclosed, the customer will be held
responsible for all costs incurred to ensure the safe handling of the product, and is liable for any harm or injury to Agilent employees as well as to any third
party occurring as a result of exposure to toxic or hazardous materials present in the product.

Request for Return Form
Agilent Vacuum Products Division/Sales and Service Offices
© Agilent Technologies, Inc. 2017
United States
Agilent Technologies
121 Hartwell Avenue
Lexington, MA 02421 - USA
Tel. +1 781 861 7200
Fax. +1 781 860 5437
Toll-Free: +1 800 882 7426
Belgium
Agilent Technologies Belgium S.A./N.V.
Customer Contact Center
Pegasus Park
De Kleetlaan 12A bus 12
B-1831 Diegem
Tel. +32 2 404 92 22
Fax. +32 2 626 46 30
customercare_belgium@agilent.com
Brazil
Agilent Technologies Brasil
Avenida Marcos Penteado de Ulhoa
Rodrigues, 939 - 6° andar
Castelo Branco Office Park
Torre Jacarandá - Tamboré
Barueri, Sao Paulo CEP: 06460-040
Toll-Free: +0800 728 1405
China
Agilent Technologies (China) Co. Ltd
No.3, Wang Jing Bei Lu, Chao Yang District
Beijing, 100102, China
Tel. +86 (0)10 64397888
Fax. +86 (0)10 64392765
Toll free: 800 820 6778 (Phone)
Toll free: 400 820 6778 (Cell Phone)
vacuum.cnm[email protected]
France
Agilent Technologies
Parc Technopolis - Z.A. de Courtaboeuf
3, avenue du Canada - CS 90263
91978 Les Ulis cedex, France
Tel. +33 (0) 1 64 53 61 15
Fax. +33 (0) 1 64 53 50 01
Germany and Austria
Agilent Technologies
Sales & Services GmbH & Co. KG
Lyoner Str. 20
60 528 Frankfurt am Main
GERMANY
Ph. +49 69 6773 43 2230
Fax. +49 69 6773 43 2250
India (Sales)
Agilent Technolo ies India Pvt. Ltd.
Unit Nos 110-116, & Part of 101 & 109
First Floor, Splendor Forum,
Plot No. -3, District Centre, Jasola
New Delhi-110025
Tel. +91 11 4623 7100
Fax. +91 4623 7105
India (Service)
Agilent Technologies India Pvt. Ltd.
C-Block, RMZ Centennial Plot Number -8A, 8B,
8C, 8D,
Doddanakundi Industrial Area, ITPL Road,
Mahadevapura Post, Bangalore- 560048
Tel. +91 80 40614000
Fax. +91 80 40148991
Italy
Agilent Technologies Italia S.p.A.
Via F.lli Varian, 54
10040 Leini, (Torino) - Italy
Tel. +39 011 9979 111
Fax. +39 011 9979 350
Toll free: 00800 234 234 00
Japan
Agilent Technologies Japan, Ltd.
9-1 Takakura-cho Hachioji-city,
Tokyo, Japan
Tel. +81- 3-5232-1253
Fax. +81-120-565-154
Toll-Free: +81-120-477-111
jp-vvt[email protected]
Korea
Agilent Technologies Korea, Ltd.
Ilshin Builiding 4F
Yongsan-gu Hannam-daero
Seoul Korea 04418
Tel. +82 (0)2 2194 9449
Fax. +82 (0)2 2194 9853
Toll free: 080 222 2452
Mexico
Agilent Technologies
Concepcion Beistegui No. 109
Col Del Valle
C.P. 03100 –Mexico, D.F.
Tel. +52 5 523 9465
Fax. +52 5 523 9472
Netherlands
Agilent Technologies Netherlands B.V.
Customer Contact Center
Laan van Langerhuize 1, toren A-8
1186 DS Amstelveen
Tel. +31 020 547 2600
Fax. +31 020 654 5748
customercare_netherland[email protected]
Singapore
Agilent Technologies Singapore Pte. Ltd,
1 Yishun Avenue 7,
Singapore 768923
Tel. (65) 6215 8045
Fax. (65) 6754 0574
Toll-Free: 1 800 2762622
Southeast Asia
Agilent Technologies Sales Sdn Bhd
Unit 201, Level 2 uptown 2,
2 Jalan SS21/37, Damansara Uptown
47400 Petaling Jaya, Selangor, Malaysia
Tel. +603 7712 6181
Fax. +603 7727 1239
Toll-Free: +1 800 880 805
Taiwan
Agilent Technologies Taiwan Limited
No. 20 Gao-shuang Road, Ping-zhen Dis
Tao-Yuan City
32450 Taiwan, R.O.C.
Tel. +8863 3 4959004
Toll-Free: 0800 018 768
UK and Ireland
Agilent Technologies LDA UK Limited
Lakeside Cheadle Royal Business Park
Cheadle, Cheshire SK8 3GR,
United Kingdom
Tel. +44 01865291570
Fax. +44 01865291571
Toll-Free: 00 800 234 234 00
Toll-Free fax.: 00 800 345 345 00
Other Countries
Agilent Technologies Italia S.p.A.
Via F.lli Varian, 54
10040 Leini, (Torino) - Italy
Tel. +39 011 997 9111
Fax. +39 011 997 9350
Toll-Free: 00 800 234 234 00
Customer Support & Service
NORTH AMERICA:
Toll Free: 800 882 7426
Lexington-service@agilent.com
EUROPE:
Toll-Free: 00 800 234 234 00
PACIFIC RIM:
please visit our website for individual office
information
http://www.agilent.com/chem/vacuum
Worldwide Web Site, Catalog and Order Online:
www.agilent.com/chem/vacuum
Representatives in most countries
Table of contents
Other Agilent Technologies Remote Control manuals
Popular Remote Control manuals by other brands

tws
tws TX-3S instruction manual

Flying Industry Development
Flying Industry Development S1PX-DC-ANT2 quick guide

Bang & Olufsen
Bang & Olufsen Beolink 1000 Service manual

GERA
GERA Z LED 60 Instructions for commissioning and operating

Origin Acoustics
Origin Acoustics OV100 installation manual

Redarc
Redarc REMOTE-RS manual