4 SMU Link Adaptor Instruction Manual
3. SYNCHRONISING 2 SMUS
A fully functioning two channel SMU can be created with two SMUs via a handshake triggering
system using the Link Adaptor to connect the trigger DIO lines.
NOTE
When using the Link adaptor to connect two SMUs, the adaptor internally connects the
‘Trigger In’ of each SMU to the ‘Trigger Out’ of the other. 'Trigger Out' will be set once all
measurements associated with the level / shape are complete.
This combination of 'Trigger in' and 'Trigger Out' allows for the handshaking of multiple
instruments without further wiring.
For full synchronisation of two SMUs, consider the following settings with the 'Manual Setup'
of the SMU:
To initialise the test, both SMUs must be ‘RUN’.
The DIO must all be set to either active high or active low on both instruments.
CNFG > [System] Interfaces > [DIO] Pin Action
A trigger (either level or shape) must be set on both instruments, A shape other than steady
must be selected to set the trigger.
CNFG > [Source Measure Action] Manual Setup > [Timing] Trigger