Airflow TA465 Operation manual

AIRFLOW™ INSTRUMENTS
AIR VELOCITY METER
MODEL TA465
OPERATION AND SERVICE MANUAL
P/N 6005359, REV D
OCTOBER 2014


i
Copyright
TSI Incorporated / 2011-2014 / All rights reserved.
Address
TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No.
(651) 490-3824
Limitation Of Warranty And Liability (effective April 2014)
(For country-specific terms and conditions outside of the USA, please visit www.tsi.com.)
Seller warrants the goods, excluding software, sold hereunder, under normal use and
service as described in the operator's manual, to be free from defects in workmanship
and material for 24 months, or if less, the length of time specified in the operator's
manual, from the date of shipment to the customer. This warranty period is inclusive of
any statutory warranty. This limited warranty is subject to the following exclusions and
exceptions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other
components when indicated in specifications, are warranted for 90 days from the date of
shipment;
b.
manuals;
c. Parts repaired or replaced as a result of repair services are warranted to be free from
defects in workmanship and material, under normal use, for 90 days from the date of
shipment;
d. Seller does not provide any warranty on finished goods manufactured by others or on
any fuses, batteries or other consumable materials. Only the original manufacturer's
warranty applies;
e. This warranty does not cover calibration requirements, and seller warrants only that
the instrument or product is properly calibrated at the time of its manufacture.
Instruments returned for calibration are not covered by this warranty;
f. This warranty is VOID if the instrument is opened by anyone other than a factory
authorized service center with the one exception where requirements set forth in the
manual allow an operator to replace consumables or perform recommended
cleaning;
g. This warranty is VOID if the product has been misused, neglected, subjected to
accidental or intentional damage, or is not properly installed, maintained, or cleaned
according to the requirements of the manual.Unless specificallyauthorized in a
separate writing by Seller, Seller makes no warranty with respect to, and shall have no
liability in connection with, goods which are incorporated into other products or
equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS
stated herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR
PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE. WITH RESPECT TO
SELLER’S BREACH OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT,
SAID WARRANTY IS LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND
EXCLUDES CLAIMS OF CONTRIBUTORY OR INDUCED INFRINGEMENTS.
BUYER’S EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE
PRICE DISCOUNTED FOR REASONABLE WEAR AND TEAR OR AT SELLER’S
OPTION REPLACEMENT OF THE GOODS WITH NON-INFRINGING GOODS.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER
OR BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES,
INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS
BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR

ii
OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND
OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR
REPLACEMENT OF THE GOODS. IN THE CASE OF SOFTWARE, SELLER WILL
REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO, WILL
REFUND THE PURCHASE PRICE OF THE SOFTWARE. IN NO EVENT SHALL
SELLER BE LIABLE FOR LOST PROFITS, BUSINESS INTERRUPTION, OR ANY
SPECIAL, INDIRECT, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER
SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING OR
REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be
brought against Seller more than 12 months after a cause of action has accrued. The
goods returned under warranty to Seller's factory shall be at Buyer's risk of loss, and
will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY
AND LIABILITY, which contains the complete and exclusive limited warranty of Seller.
This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or
its terms waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are
to our customers, our service policy is designed to give prompt attention to any
problems. If any malfunction is discovered, please contact your nearest sales office or
representative, or call Customer Service department at +44 (0) 149 4 459200 (UK),
(800) 874-2811 (USA), or (1) 651-490-2811 (International).

iii
CONTENTS
CHAPTER 1 UNPACKING AND PARTS IDENTIFICATION ............1
CHAPTER 2 SETTING-UP.................................................................3
Supplying Power to the Model TA465 Series.................................3
Installing the Batteries...............................................................3
DIP Switch Settings...................................................................3
Using the AC Adapter................................................................4
Connecting Ventilation or IAQ Probes............................................5
Using the Telescoping Thermoanemometer Probes......................5
Extending the Probe..................................................................5
Retracting the Probe..................................................................5
Using the Pressure Ports (TA465-P)..............................................6
Connecting the Static Pressure Probe......................................6
Connecting an Optional Pitot Probe or Airflow (straight
pitot) Probe..........................................................................7
Thermocouple Ports .......................................................................9
Connecting the Thermocouples ................................................9
Connecting the Optional Bluetooth Portable Printer Device ........10
Connecting to a Computer............................................................10
CHAPTER 3 OPERATION...............................................................11
Keypad Functions.........................................................................11
Common Terms............................................................................12
Menus...........................................................................................13
ZERO PRESSURE.............................................................................13
DISPLAY SETUP...............................................................................14
SETTINGS .......................................................................................15
FLOW SET UP ..................................................................................15
VOC SETUP.....................................................................................18
ACTUAL/STANDARD SETUP ..............................................................18
DATA LOGGING ...............................................................................19
Measurements.........................................................................19
Log Mode/Log Settings ...........................................................19
Choose Test............................................................................24
Name Test...............................................................................24
View Data................................................................................25
View Stats................................................................................25
View Samples..........................................................................26
Print Test .................................................................................27
Delete Data..............................................................................28
% Memory ...............................................................................29
ZERO CO ........................................................................................30
APPLICATIONS ................................................................................31

iv
CALIBRATION ..................................................................................32
BLUETOOTH FUNCTIONS ..................................................................32
Discover Devices.....................................................................32
Discoverability..........................................................................33
PINcode...................................................................................33
# AutoConnects.......................................................................33
Printing Data Using the Portable Printer.......................................33
Downloading Software................................................34
CHAPTER 4 MAINTENANCE..........................................................35
Recalibration.................................................................................35
Cases............................................................................................35
Storage..........................................................................................35
CHAPTER 5 TROUBLESHOOTING................................................37
APPENDIX A SPECIFICATIONS.....................................................39
APPENDIX B OPTIONAL PLUG-IN PROBES ................................43
These Application Notes can be found under
http://www.tsi.com
AF-106
Traversing a Duct to Determine Average Air Velocity or Volume
AF-107
Using the Models TA465, PVM610, and PVM620 to Measure
Pressure
AF-109
Converting Standard Velocity to Actual Velocity
AF-114
Use of K-Factors with TA465, PVM610, and PVM620
AF-124
Heat Flow Calculations
AF-138
Percent Outdoor Air Calculation and Its Use
AF-141
Turbulence Intensity Measurements
AF-142
Draft Rate: A Determining Factor in the Quantification of Human
Comfort
AF-147
Photo-Ionization Detection (PID) Technology
AF-150
Using Bluetooth®Communications

1
Chapter 1
Unpacking and Parts Identification
Carefully unpack the instrument and accessories from the shipping
container. Check the individual parts against the list of components
below. If anything is missing or damaged, notify TSI immediately.
1. Carrying case
2. Instrument
3. Static pressure tip (TA465-P model only)
4. Rubber tubing (TA465-P model only)
5. USB cable
6. LogDat2 CD-ROM with downloading software
7. AC adapter

Chapter 1
2
(This page intentionally left blank)

3
Chapter 2
Setting-up
Supplying Power to the Model TA465 Series
The Model TA465 AIRFLOWInstruments Air Velocity Metercan be
powered in one of two ways: four size AA batteries or the AC adapter.
Installing the Batteries
Insert four AA batteries as indicated by the diagram located on
the inside of the battery compartment. The Model TA465 is
designed to operate with either alkaline or NiMH rechargeable
batteries. Battery life will be shorter if NiMH batteries are used. If
NiMH batteries are used the DIP switch will need to be changed.
Refer to the next section, DIP Switch Settings. Carbon-zinc
batteries are not recommended because of the danger of battery
acid leakage.
DIP Switch Settings
To access the DIP switch, remove the batteries from the battery
compartment. On the inside of the battery compartment, there is
a window with a single DIP switch (see figure below). The table
below shows the functions for the switch.
Caution: Make certain that power is turned off before changing the
DIP switch settings.
Switch
Function
Settings
1
NiMH
OFF: Alkaline Batteries
ON: NiMH Rechargeable Batteries
AIRFLOW is a trademark of TSI Incorporated.

4 Chapter 2
Do not attempt to charge alkaline batteries.
Using the AC Adapter
The AC adapter can be used to power the instrument or to
charge the NiMH batteries when the DIP switch in the battery
compartment is set to NiMH. If the DIP switch is set to
Alkaline, and the AC power adapter is connected, then the
batteries will be bypassed and the meter will be powered by
the AC adapter. Be sure to provide the correct voltage and
frequency, which is marked on the back of the AC adapter.
AC Adapter input

Setting-up 5
Connecting Ventilation or IAQ Probes
e
mini-DIN connector which must align with the connector at the base
of the TA465 series meter. This will ensure the probe is properly
connected and remains so during use. Once connected and turned
on, refer to the DISPLAY SETUP for displaying the desired
measurements.
“D” Shaped
mini-DIN connector
Using the Telescoping Thermoanemometer Probes
The telescoping probe contains the velocity, temperature, and
humidity sensors. When using the probe, make sure the sensor
window is fully exposed and the orientation dimple is facing upstream.
N O T E
For temperature and humidity measurements, make sure that at
least 3 inches (7.5 cm) of the probe is in the flow to allow the
temperature and humidity sensors to be in the air stream.
Extending the Probe
To extend the probe, hold the handle in one hand while pulling on
the probe tip with the other hand. Do not hold the cable while
extending the probe as this prevents the robe from extending.
Retracting the Probe
To retract the probe, hold the handle in one hand while gently
pulling the probe cable until the smallest antenna section is
retracted.

6 Chapter 2
Do not use the instrument or probes near hazardous
voltage sources since serious injury could result.
Using the Pressure Ports (TA465-P)
The TA465-P includes pressure ports that can be used to measure
static and differential pressures in ductwork. For more information,
see Application Note AF-107.
Negative (-) Pressure Port Positive(+) Pressure Port
Connecting the Static Pressure Probe
The Static Pressure probe included with the TA465-P is
connected to the + port on the TA465-P using the included
tubing. The Static Pressure probe is used to measure the duct
static pressure and features a magnet which holds the probe to
the ductwork.

Setting-up 7
Connecting an Optional Pitot Probe or Airflow (straight
pitot) Probe
When connected to a pitot probe, air velocity or air volume can be
-
pressure ports on the Model TA465-P using two pieces of tubing
of equal length. The total pressure port of the pitot probe
-
For information on how to perform a duct traverse, refer to
Application Note AF-106.
N O T E
The pitot velocity needs a valid temperature to perform the
standard or actual velocity correction. This is accomplished
able of
measuring temperature (plug in probe or thermocouple) is
connected, the
The duct air temperature must then be manually inputted by
be or Thermocouple 1 or 2, and no
probe is connected, dashes (------) will appear on the
display.
For more information on entering the temperature manually,
refer to the Actual/Standard Setup section of this manual.

8 Chapter 2
Total pressure hole
Total pressure port (to
+ input on manometer)
Staticpressure
holes
Staticpressure port (to
input on manometer)
Do not use the instrument or probes near hazardous
voltage sources since serious injury could result.

Setting-up 9
Thermocouple Ports
The TA465 series includes two thermocouple ports at the base of the
meter. Any K-Alloy thermocouple with mini-connector can be
attached. See Display Setup for setting the thermocouple
temperature readings to be displayed as TC1, TC2, or TC1-TC2.
Connecting the Thermocouples
The K-Alloy thermocouple with mini-connector has one terminal
wider than the other. The wider terminal will be inserted into the
bottom of the TC1 or TC2 connector port.
K-alloy Thermocouple
TC1 TC2
Thermocouples from an alternate TSI supplier must have
the metal sheath electrically isolated from the wires
inside. Failure to meet this requirement may result in
false readings, electrical shock, or fire hazard.
Do not use the instrument or probes near hazardous
voltage sources since serious injury could result.

10 Chapter 2
Connecting the Optional Bluetooth®Portable Printer Device
To connect the Bluetooth printer to the Model TA465, power on the
unit and the printer. Then press the MENU soft key. From the Menu
use the and keys to highlight Bluetooth Functions and press the
key. Highlight Discover Devices and press the key. If other TSI
Bluetooth-printers are in the area, turn them off before searching. The
Model TA465 will then search for and list all available Bluetooth
devices.
For more information on establishing Bluetooth connections, refer to
TSI Applications Note AF-150.
Connecting to a Computer
Use the Computer Interface USB Cable provided with the
Model TA465 to connect the instrument to a computer for
downloading stored.
USB Communications Port
C a u t i o n
This symbol is used to indicate that the data port of the
Model TA465 is not intended for connection to a public
telecommunications network. Connect the USB data port
only to another USB port on a safety certified computing
device.
Protection provided by the instrument could be impaired
if used in a manner other than specified in this user
manual.
Bluetooth is a registered trademark of Bluetooth SIG.

11
Chapter 3
Operation
PressurePorts
Additional
Messages
Primary
Parameter
Secondary
Parameters
BatteryIndicator
ActivityArea
Soft Key
Indicators
Soft Keys
Keypad Functions
ON/OFF ( ) Key
Press the ON/OFF key to turn the Model TA465 on
and off. During the power up sequence the display
will show the following: Model Number, Serial
Number, and Software Revision. To turn the
instrument off, press and hold the ON/OFF Key for 3
seconds. The instrument will count down (OFF2,
OFF1, OFF). If AC Adapter is attached, the Battery
and ON/OFF Key is bypassed. If the ON/OFF Key is
pressed while the AC adapter is attached, the
attach the AC adapter or press the ON/OFF Key.
Arrow () Keys
Press to scroll through choices while setting a
parameter. Pressing the keys simultaneously
will lock the keypad to prevent unauthorized
adjustments to the instruments. To unlock the
keypad, press the keys simultaneously.

12 Chapter 3
Enter ( ) Key
Press to accept a menu selection, value or
condition. Press to Start or Stop datalogging when
in Continuous Key mode.
Arrow (or ) and
Menu Soft Keys
Press arrow keys to change choices while setting a
parameter. Press the Menu soft key to select the
Menu selections, which are Pressure Zero, Display
Setup, Settings, Flow Setup, VOC Setup,
Actual/Std Setup, Data Logging, Zero CO,
Applications, Calibration, and Bluetooth Functions.
Common Terms
In this manual there are several terms that are used in different
places. The following is a brief explanation of the meanings of those
terms.
Sample
Consists of all of the measurement parameters
stored at the same time.
Test ID
A group of samples. The statistics (average,
minimum, maximum, and count) are calculated for
each test ID. The maximum number of test IDs
is 100.
Time Constant
The time constant is an averaging period. It is used
to dampen the display. If you are experiencing
fluctuating flows, a longer time constant will slow
down those fluctuations. The display will update
every second, but the displayed reading will be the
average over the last time constant period. For
example, if the time constant is 10 seconds, the
display will update every second, but the displayed
reading will be the average from the last 10
seconds. This is also r
Log Interval
The logging interval is the period over which the
instrument will average the logged sample. For
example, if the logging interval is set to 30 minutes,
each sample will be the average over the previous
30 minutes.
Test Length
This is the time over which the data will be logged
-

Operation 13
Menus
The menu structure is organized to allow easy navigation and
instrument setup utilizing the arrow keys and button. To exit a
menu or menu item, press the ESC key.
To access the Menu items, press the Menu soft key.
To select a parameter, use the Arrow keys to highlight the
selection and press the button.
MENU
Zero Press
Display Setup
Settings
Flow Setup
VOC Setup
Actual/Std Setup
Data Logging
Zero CO
Applications
Calibration
Bluetooth Functions
ZERO PRESSURE
To zero the pressure reading, leave the positive and negative ports
open to atmosphere and select the Zero Press. The pressure sensor
should be zeroed after initial setup and checked periodically. The
instrument will indicate if the pressure zero was successful.

14 Chapter 3
DISPLAY SETUP
Display Setup menu is where you will setup the desired parameters to
be displayed on the instrument screen. With a parameter highlighted
you can then use the ON soft key to have it show up on the instrument
screen or select the OFF soft key to turn off the parameter. Use
PRIMARY soft key to have a parameter show up on the instrument
screen in a larger display. A total of 5 parameters can be shown on
the display, 1 primary (large font) and 4 secondary. Parameters
shown in the Display Setup screen are dependent on the type of
probe currently connected.
When set to PRIMARY, measurement will be the large font on the
display.
When set to ON, measurement will be displayed as a secondary
parameter (up to 4 can be displayed).
When set to OFF, measurement will not be displayed.
MENU
Zero Press
Display Setup
Settings
Flow Setup
VOC Setup
Actual/Std Setup
Data Logging
Zero CO
Applications
Calibration
Bluetooth Functions
DISPLAY SETUP
Velocity *ON
Flow OFF
Pitot Velocity OFF
Pressure ON
Temperature ON
Baro Press OFF
ON PRIMARY OFF
N O T E
Pitot Velocity and AFProbe Velocity cannot both be ON at the same
time, nor can one be set to PRIMARY and the other to ON. Only one
can be ON at any given time.
Table of contents
Other Airflow Measuring Instrument manuals