Milwaukee MI 405 User manual

This manual suits for next models

1

Other Milwaukee Measuring Instrument manuals

Milwaukee MW306 MAX User manual

Milwaukee

Milwaukee MW306 MAX User manual

Milwaukee MI 106 User manual

Milwaukee

Milwaukee MI 106 User manual

Milwaukee M12 ICAV3 User manual

Milwaukee

Milwaukee M12 ICAV3 User manual

Milwaukee Mi 150 User manual

Milwaukee

Milwaukee Mi 150 User manual

Milwaukee PH55 User manual

Milwaukee

Milwaukee PH55 User manual

Milwaukee MW11 User manual

Milwaukee

Milwaukee MW11 User manual

Milwaukee MA881 User manual

Milwaukee

Milwaukee MA881 User manual

Milwaukee PRO MI415 User manual

Milwaukee

Milwaukee PRO MI415 User manual

Milwaukee MW10 User manual

Milwaukee

Milwaukee MW10 User manual

Milwaukee MI 806 User manual

Milwaukee

Milwaukee MI 806 User manual

Milwaukee MW150 User manual

Milwaukee

Milwaukee MW150 User manual

Milwaukee 2205-40 User manual

Milwaukee

Milwaukee 2205-40 User manual

Milwaukee PRO MW801 User manual

Milwaukee

Milwaukee PRO MW801 User manual

Milwaukee 48-22-9801 User manual

Milwaukee

Milwaukee 48-22-9801 User manual

Milwaukee SM101 User manual

Milwaukee

Milwaukee SM101 User manual

Milwaukee SMS510 User manual

Milwaukee

Milwaukee SMS510 User manual

Milwaukee CALIBRATION 2238-20 User manual

Milwaukee

Milwaukee CALIBRATION 2238-20 User manual

Milwaukee M12 3PL User manual

Milwaukee

Milwaukee M12 3PL User manual

Milwaukee MAX MW160 User manual

Milwaukee

Milwaukee MAX MW160 User manual

Milwaukee MI490 User manual

Milwaukee

Milwaukee MI490 User manual

Milwaukee MI 404 User manual

Milwaukee

Milwaukee MI 404 User manual

Milwaukee Mi 190 User manual

Milwaukee

Milwaukee Mi 190 User manual

Milwaukee SM700 User manual

Milwaukee

Milwaukee SM700 User manual

Milwaukee SM106 User manual

Milwaukee

Milwaukee SM106 User manual

Popular Measuring Instrument manuals by other brands

Powerfix Profi 278296 Operation and safety notes

Powerfix Profi

Powerfix Profi 278296 Operation and safety notes

Test Equipment Depot GVT-427B user manual

Test Equipment Depot

Test Equipment Depot GVT-427B user manual

Fieldpiece ACH Operator's manual

Fieldpiece

Fieldpiece ACH Operator's manual

FLYSURFER VIRON3 user manual

FLYSURFER

FLYSURFER VIRON3 user manual

GMW TG uni 1 operating manual

GMW

GMW TG uni 1 operating manual

Downeaster Wind & Weather Medallion Series instruction manual

Downeaster

Downeaster Wind & Weather Medallion Series instruction manual

Hanna Instruments HI96725C instruction manual

Hanna Instruments

Hanna Instruments HI96725C instruction manual

Nokeval KMR260 quick guide

Nokeval

Nokeval KMR260 quick guide

HOKUYO AUTOMATIC UBG-05LN instruction manual

HOKUYO AUTOMATIC

HOKUYO AUTOMATIC UBG-05LN instruction manual

Fluke 96000 Series Operator's manual

Fluke

Fluke 96000 Series Operator's manual

Test Products International SP565 user manual

Test Products International

Test Products International SP565 user manual

General Sleep Zmachine Insight+ DT-200 Service manual

General Sleep

General Sleep Zmachine Insight+ DT-200 Service manual

Sensa Core Lacto Spark user manual

Sensa Core

Sensa Core Lacto Spark user manual

VOLTCRAFT UTS-1980 operating instructions

VOLTCRAFT

VOLTCRAFT UTS-1980 operating instructions

EBCHQ 94915 Operation manual

EBCHQ

EBCHQ 94915 Operation manual

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Balluff

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Pulsar IMP Lite Series instruction manual

Pulsar

Pulsar IMP Lite Series instruction manual

ETCR 6800 user manual

ETCR

ETCR 6800 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

USER MANUAL
MI 405
Ammonia Medium
Range Meter
Dear Customer,
Thank you for choosing a Martini product.
This manual will provide you with the
necessary information for the correct use of
the instrument.
Please read it carefully before using the
meter. This instrument is in compliance with
directives.
SPECIFICATIONS:
Range 0.00 to 9.99 mg/L (NH3-N)
Resolution 0.01 mg/L
Precision ±0.10 mg/L @ 5.00 mg/L
Light Source Blue LED 466 nm
Light Detector Silicon Photocell and
466 nm narrow band interference filter
Method Adaptation of Nessler method
Environment 0 to 50°C (32 to 122°F);
100% RH max.
Battery Type 1 x 9 volt
Auto-Shut off After 10' of non-use
Dimensions 192 x 104 x 52 mm
(7.5 x 4.1 x 2”)
Weight 380 g
MEASUREMENT PROCEDURE:
1• Turn the meter on by pressing
ON/OFF.
2• When the LCD displays “- - -”, the
meter is ready.
3• Fill the cuvet with 10
mL of sample, up to the
mark, and replace the
cap.
4• Place the cuvet into the
holder and ensure that
the notch on the cap is
positioned securely into
the groove.
5• Press ZERO and “SIP”
will blink on the display.
6• After a few seconds the
display will show “-0.0-
”. The meter is now
zeroed and ready for
measurement.
7• Remove the cuvet. Open it.
8• Add 4 drops of NH3-1 (10 drops in
case of seawater analysis) and swirl
gently.
9• Add 4 drops of NH3-2 (10 drops in
case of seawater analysis). Replace
the cap and swirl gently.
10•Immediately reinsert
the cuvet into the
holder and ensure
that the notch on the
cap is positioned
securely into the
groove.
11•Press and hold
READ for 3 seconds
and the display will
show the countdown
prior to the
measurement.
Alternatively, wait
for 3 minutes and 30
seconds and just press READ. In
both cases “SIP” will blink during
measurement.
12•The instrument directly displays the
concentration in mg/L of ammonia
nitrogen (NH3-N).
13•To convert the reading to mg/L of
ammonia (NH3), multiply by a factor
of 1.216.
Interferences
• Organic compounds like:
chloramines, various aliphatic and
aromatic amines, glycine, (positive
error).
To eliminate these interferences
distillation is required.
• Organic compounds like: aldehydes,
alcohols (e.g. ethanol) or acetone
(negative error).
To eliminate these interferences
distillation is required.
• Sulfide (S2-): may cause turbidity.
• Hardness above 1 g/L as Calcium
Carbonate (CaCO3).
GUIDE TO DISPLAY CODES:
This prompt appears for 1 second each
time the instrument is turned on.
The dashes “- - -”
indicates that the
meter is in a ready
state and zeroing can
be performed.
Sampling In Progress.
Flashing “SIP” prompt
appears each time the
meter is performing a
measurement.
“-0.0-”, the meter is in
a zeroed state and
measurement can be
performed.
The blinking "BAT"
indicates that the
battery voltage is
getting low and the
battery needs to be
replaced.
“-bA-”, the battery is
dead and must be
replaced. Once this
indication is displayed,
the meter will lock up.
Change the battery
and restart the meter.
“Conf”, the meter has
lost its configuration. Contact your
dealer or the nearest Martini Customer
Service Center.
ERROR MESSAGES
On zero reading
• Blinking “-0.0-” indicates that the
zeroing procedure failed due to a low
signalto-noise ratio.
In this case
press ZERO again.
• “no L”, the
instrument can not
adjust the light
level. Please check
that the sample
does not contain
any debris.
• “L Lo”, there is
not enough light to
perform a
measurement.
Please check the
preparation of the
zero cuvet.
• “L Hi”, there is too
much light to
perform a
measurement.
Please check the preparation of the zero
cuvet.
On sample reading
• “-SA-”, there is too
much light for the
sample
measurement.
Please check if the
right sample cuvet is
inserted.
• “Inv”, the sample
and the zero cuvet
are inverted.
• “ZErO”, a zero
reading was not
taken. Follow the instruction in the
measurement procedure for zeroing the
meter.
• Under range. A blinking “0.00”
indicates that the sample absorbs less
light than the zero
reference.
Check the procedure
and make sure you
use the same cuvet for
reference (zero) and
measurement.
• A flashing value of
the maximum
concentration
indicates an over
range condition.
The concentration of
the sample is beyond
the programmed
range: dilute the sample and re-run the
test.
BATTERY REPLACEMENT
Battery replacement must only take
place in a non-hazardous environment.
Simply rotate the battery cover on the
back of the meter.
Detach the
battery from
the terminals
and attach a
fresh 9V
battery while
paying
attention to
the correct
polarity.
Insert the battery and replace the cover.
ACCESSORIES:
MI505-100 Ammonia MR Reagent
(100 tests)
MI0001 Glass cuvets (2 pcs)
MI0002 Caps for cuvets (2 pcs)
MI0003 Stoppers for cuvets (2 pcs)
MI0004 Tissue for wipping cuvets
(4 pcs)
MI0005 9V battery (1 pc)
WARRANTY
This instrument is warranted against defects
in materials and manufacturing for a period of
two years from the date of purchase.
If during this period the repair or replacement
of parts is required, where the damage is not
due to negligence or erroneous operation by
the user, please
return the meter to either distributor or our
office in the original packing and the repair
will be free of charge.
Damages due to accidents, misuse,
tampering or lack of prescribed maintenance
are not covered.